SOME OBSERVATIONS ON NITROUS OXIDE CYLINDERS DURING EMPTYING

Similar documents
PERFORMANCE OF NEEDLE VALVES

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Section 14: Equipment failure

DURING the course of certain investigations it became

Anaesthetic infrastructure and supplies

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS

Recommendations on Checking Anaesthesia Delivery Systems

Section 1 Questions. Physics and Clinical Measurement... 3 Clinical Anaesthesia Physiology Pharmacology... 79

Nitrous oxide and oxygen from a single cylinder

(fig. 3) must be at the same temperature as the water in this chamber CALORIMETRIC STUDIES OF THE EXTREMITIES

Anesthesiology Research and Scholarship. Follow this and additional works at:

Guidelines on Monitoring in Anaesthesia

USE OF TIRE PRESSURE TO IMPROVE THE CALIBRATION OF THE BICYCLE W...

A SIMPLE TITRATION METHOD FOR DETERMINING THE SPECIFIC GRAVITY ON ONE DROP OF URINE

MEDICAL GASES Manufacture, Storage, Transport & Delivery B.KIRUTHIGA LECTURER DEPT OF PHARMACEUTICAL CHEMISTRY

Keeping anaesthetic pollution to minimum in theatre environment

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY

Anaesthesia. Update in. Gases and Vapours. Ben Gupta Correspondence Summary

TRICHLOROETHYLENE CONCENTRATION FROM A " BOYLE " TYPE ANAESTHETIC APPARATUS. Department of Anaesthetics, Welsh National School of Medicine, Cardiff

VETROSON. OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

GASEOUS HOMEOSTASIS AND THE CIRCLE SYSTEM

VOLUMETRIC TECHNIQUES

INTERNATIONAL ISO STANDARD Anaesthetic machines for use with humans. Appareils d anesfhesie pour utilisation chez I homme

Characterizers for control loops

The right atmosphere for health. A system for medical gas supply

AND OPERATION INSTRUCTIONS

Figure 5. Nasal mask in place

Flow meter. bellow vaporizer. APL valve. Scavenging system

BASIC LABORATORY TECHNIQUES (Revised )

3 1 PRESSURE. This is illustrated in Fig. 3 3.

Sign up to receive ATOTW weekly -

Understanding vaporizers

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

IT is frequently desirable to be able to determine the

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Alkylphosphines. Storage and Handling Recommendations

CLOSED LOOP SOLVENT EMISSIONS CONTROL. David C.H. Grant

EXPERIMENTAL STUDY ON THE DISCHARGE CHARACTERISTICS OF SLUICE FOR TIDAL POWER PLANT

Experiment Instructions. Circulating Pumps Training Panel

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

The desflurane (Tec 6) vaporizer: design, design considerations and performance evaluation

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:

THE PERFORMANCE OF A WALTON FIVE ANAESTHETIC MACHINE

O-Two Self-Study Guide. Equinox Relieve

Dissolved Oxygen Guide

The Gas Laws: Boyle's Law and Charles Law

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Cover Page for Lab Report Group Portion. Pump Performance

COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP

Pressure Regulator. Instructions for Use

W. D. A. SMITH Research Department of Anaesthetics, Royal College of Surgeons of England, London

MEDAP-MOBILE GAS SUPPLY

Figure 3. Miniscav scavenging unit

CHEM 321 Experiment 1

RESPIRATION OF MUSCLE. By W. M. FLETCHER, M.A., M.B., Fellow of Trinity College, Cambrtidge. (Three Figures

Measuring Mass and Volume

SAFE AND ENVIRONMENTALLY FRIENDLY

BASIC LABORATORY TECHNIQUES (Revised )

J.A. BAIN AND W.E. SPOEREL~

The Discussion of this exercise covers the following points:

[2] After a certain time, the temperature of the water has decreased to below room temperature.

GEOTHERMAL WELL COMPLETION TESTS

METHODS TO PREVENT THE PREMATURE ACTIVATION OF RELIEF DEVICES ON TRANSPORT TANKS

Instrument Operating Manual

SOME THOUGHTS ON THE TEACHING OF INSTRUMENTATION TECHNIQUES AND ELECTRONICS IN MEDICINE

Welcome to the Specialized Medical Services Respiratory training webinar series!

SAFETY TRAINING LEAFLET 06 CARBON DIOXIDE

UNDERSTANDING A DIVE COMPUTER. by S. Angelini, Ph.D. Mares S.p.A.

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

ON-SITE OXYGEN GENERATION WHY PSA SYSTEMS ARE A GOOD CHOICE

Sample. HLTMS207C Handle Medical Gases Safely

and its weight (in newtons) when located on a planet with an acceleration of gravity equal to 4.0 ft/s 2.

Handling and Storage of Gases

ICHEME SYMPOSIUM SERIES NO. 144

EXPERIMENT 1 BASIC LABORATORY TECHNIQUES AND TREATMENT OF DATA MEASUREMENTS

Effect of oxygen flow on inspired oxygen and carbon dioxide concentrations and patient comfort in the Amron TM oxygen hood

General Accreditation Guidance. User checks and maintenance of laboratory balances

Journal of Applied Fluid Transients, Vol 1-1, April 2014 (3-1)

Truck-mounted Mass Flow Metering for LPG Delivery

THE DANGERS OF USING A THREE-WAY VALVE AS A STOPCOCK IN A TRANSDUCER SYSTEM FOR MEASURING VENOUS BLOOD PRESSURE

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SHORT COMMUNICATION A SIMPLE, SENSITIVE AND VERSATILE SOLID-STATE PRESSURE TRANSDUCER

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

Experiment 18 Properties of Gases

(Received 9 September 1940)

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

A Study on Oxygen Utilization Pattern And Costing In Critical Care Units And Operation Theatre oa Rural Tertiary Care Hospital.

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

Transcription:

r Br. J. Anaesth. (1974), 46, 534 SOME OBSERVATIONS ON NITROUS OXIDE CYLINDERS DURING EMPTYING P. L. JONES SUMMARY In an effort to correct misunderstandings and to establish the value of nitrous oxide pressure monitors on anaesthetic machines, the decrease of pressure within nitrous oxide cylinders has been studied under a variety of clinical conditions. Although the matter had been clarified by 1946, the teaching of several current anaesthetic textbooks is still fallacious since an almost linear fall in cylinder pressure under constant flow conditions has been demonstrated. In addition, premature failure of nitrous oxide cylinder pressure due to cooling of the liquid contents has been noted. The value of cylinder pressure monitors is discussed in the light of these findings. Many anaesthetic machines in Britain are not equipped with pressure gauges for registering nitrous oxide (N a O) cylinder pressure. Consequently, patients receiving anaesthetics in which N,0 is largely responsible for unconsciousness run the risk of awareness when the cylinder approaches exhaustion. The absence of pressure gauges appears to stem from a widely held misunderstanding concerning the pressure changes within N,O cylinders during use. The present study was undertaken to clarify this situation and to investigate the feasibility of providing a warning device for use with N 2 O cylinders. It is well recognized that the relationship between a N 3 O cylinder pressure and the contents hinges upon the fact that N 2 O may be present in both liquid and gaseous phases at normal ambient temperature. Texts related to anaesthesia still mistakenly indicate that when a cylinder of N 2 O is drained at a constant gas flow rate, the pressure within the cylinder is kept constant by vaporization of the liquid. When the last drop of liquid has evaporated, the pressure then falls linearly in a manner consistent with Boyle's law. It was in 1946 that Macintosh and Mushin first indicated the fallacy of this teaching in an account of an experiment conducted upon a 900-1. cylinder which was emptied at a constant rate of 8 l./min. They found that the pressure in the cylinder decreased progressively throughout the period of emptying despite the presence of liquid N,O. Furthermore, they demonstrated that, if liquid remained in the cylinder, the PETER L. JONES, F.F.A.R.C.S., University Hospital of Wales, Heath Park, Cardiff. pressure would eventually increase again to its original level when the cylinder was allowed to rewarm after discontinuing the flow of gas. These authors pointed out that this is because of the effect of temperature upon the vapour pressure of liquid N,O. Nevertheless, several authors have perpetuated the original erroneous interpretation (Minnitt and Gillies, 1948; Wylie and Churchill-Davidson, 1972; Lee and Atkinson, 1973). Hill (1972) makes the same assertion, even though he provides evidence which clearly supports the teaching of Macintosh and Mushin. Indeed, his contention can only be true for a cylinder emptied infinitely slowly. Macintosh and Mushin, on the other hand, had used a high flow with a small cylinder, thereby possibly exaggerating the effect of cooling. The experiments described below were designed to establish the relevance of their work to present-day clinical practice. Experimental method. Standard 900 and 1,800-1. pin index, N,O cylinders (British Oxygen Co. Ltd) were connected to a suitable yoke on an anaesthetic machine. The cylinder pressure was recorded using a calibrated Bourdon pressure gauge (0-1000 Lb/sq.in.). A continuous record of pressure was made by connecting a low-torque rotation transducer to the needle spindle of the Bourdon gauge, the electrical output being continuously recorded on a potentiometric chart recorder. The cylinder was connected via a 60 Lb/sq.in. (42 KN/m 2 ) reducing valve to a flowmeter and needle valve. This permitted flow of up to 10 l./min to be controlled. Weighed cylinders were allowed to come into thermal equilibrium with

OBSERVATIONS ON NITROUS OXIDE CYLINDERS DURING EMPTYING 535 room air and were then emptied at a constant flow rate in an ambient temperature of between 20 and 23 C until the pressure in the cylinders had decreased to 60 Lb/sq.in. Each cylinder was then turned off, warmed to room temperature in a water bath, and re-weighed to establish the amount of nitrous oxide remaining. It was then re-connected to the anaesthetic machine and allowed to run at the original gas flow once more until it was finally empty. In all, four 1800-1. cylinders and four 900-L cylinders were tested, one of each being tested at flow rates of 2, 4, 8 and 10 l./min. In addition, the development of frost upon the cylinder walls was noted and recorded. b/tn* BOO fitres ddvcrcd FIG. 3 HOO 20OO RESULTS Cylinder pressure recordings. The cylinder pressure tracings for 900-1. cylinders are summarized in figures 1 and 2 and those for 1800-1. cylinders in figures 3 and 4. The horizontal axis in figures 1 and 3 represents litres delivered. For comparison, time forms the horizontal axis in figures 2 and 4. The former correlation makes for more convenient comparison though, of course, the latter is more relevant clinically. In all cases, a progressive fall in pressure accompanied the release of gas from the cylinder at constant flow. The rate at which the pressure declined (i.e. the slope of the curve) was related to the flow rate. Flows of 10, 6, and 4 l./min from both 1800- and 900-1. cylinders lead to premature failure of the pressure and in these cases re-warming enabled a further volume of gas to be delivered. Only at a flow rate of 2 l./min did the cylinders not fail prematurely, and even at this flow, the moment of vaporization of the "last drop" of liquid N,0 was not clearly defined by a sharp change in slope. FIG. 1 bin 1 FIGS. 1 and 2. Pressure change in 900-L N,O cylinders emptying at various continuous flow rates. Broken lines indicate the effect of rewarming to ambient temperature. Ib/in* tin* (n*utc») FIG. 4 FIGS. 3 and 4. Pressure change in 1800-L N,O cylinders emptying at various continuous flow rates. Broken lines indicate effect of rewarming to ambient temperature. The development of surface frost. Frost developed on the outside surface of all the cylinders during the initial decompression. This was not related to the state of emptiness of the cylinders or to the duration of operation. Generally it developed below a demarcation line situated approximately half-way down the cylinder when the pressure within the cylinder reached approximately 300 Lb/sq.in. Once this line was established, the frost level did not fall more than 2 cm before the cylinder failed, though the thickness of the ice layer did increase. When those cylinders which had failed prematurely were inverted immediately after disconnection from the yoke, a tinkling sound was audible and the neck of the cylinder immediately became extremely cold. Frost promptly formed on the neck in these circumstances (see figs. 5A,B). It was originally intended to allow cylinders to re-warm at ambient temperature whilst attached to the anaesthetic machine. However, it was found that re-warming was incomplete after 6 hours with the cylinder valve closed and re-warming in a water bath which had been brought to ambient temperature was undertaken.

536 BRITISH JOURNAL OF ANAESTHESIA FIG. 5A FIG. 5B FIG. 5. The effect of inversion of a frosted cylinder which contains residual liquid NjO. (A) Cylinder immediately after removal from the anaesthetic machine; (B) 1 minute after inversion. Note the development of frost around the neck of the cylinder. -6O -4O -2O O degrees centigrade FIG. 6. The relationship between vapour pressure of N,O and temperature in the presence of liquid N,O. DISCUSSION From the evidence it is clear that, under clinical circumstances, the pressure within N,0 cylinders does not remain constant. On the contrary, particularly at high flow rates, an almost linear decline in pressure occurs. This is the result of a progressive decrease in temperature within the cylinder. The relationship between the vapour pressure of nitrous oxide and temperature in the presence of liquid N,0 is indicated in figure 6 which is taken from data published by Couch and Kobe (1961). Using the vapour pressure curve, the change in temperature that occurred in the 1800-L cylinder which had been emptied at 6 l./min has been extrapolated (fig. 7) and indicates that there is an increase in the rate of fall of temperature with time. The explanation for this depends upon several factors: (1) The latent heat of vaporization (which is the sole reason for heat consumption within the cylinder) increases as the temperature decreases. Thus, as the rate of vaporization is kept constant, so the rate of heat consumption increases. Nitron OiMa FIG. 7. Calculated temperature decay in the 1800-L N,O cylinder emptied at 6 l./min. This corresponds to the pressure curve shown in figure 2. (2) The first source of heat for this increase in consumption is derived from the thermal inertia of the liquid N,O itself. However, as the liquid evaporates, so this source of heat decreases resulting in a further acceleration in the rate of decline of temperature.

OBSERVATIONS ON NITROUS OXIDE CYLINDERS DURING EMPTYING 537 (3) As the cylinder cools it absorbs heat from its environment: the greater the temperature differential the more rapidly the heat is absorbed. This tends to limit the acceleration in decline of temperature and accounts for the difference in pattern between rapidly and slowly emptied cylinders. (4) Ice formation on the cylinder wall further complicates the situation. Its formation can occur only after the temperature of the cylinder has decreased below zero degrees centigrade thus implying that the pressure within the cylinder has decreased below 300 lb per sq. inch in the presence of liquid N = O. Thereafter, as ice forms, latent heat of formation of ice is delivered to the vaporizing N,O. However, the increasing thickness of the ice and snow layer acts as an insulator which limits the further transfer of heat from the environment. Warning of cylinder failure. Failure of the NjO gas flow during anaesthesia may have two consequences. In those anaesthetic techniques where it is of major significance in maintaining unconsciousness, as in neuroleptanaesthesia, temporary discontinuation of that gas might result in awareness. Of perhaps greater physiological significance is the influence of the total fresh gas flow rate upon the degree of re-breathing with some anaesthetic systems. Unnoticed failure of a N 5 O cylinder could reduce the fresh gas flow by 70%. The hypercapnia, which would ensue if this situation was to pass unheeded, could lead to serious consequences. Notwithstanding these considerations, a committee of the Ministry of Health set up to discuss warning devices and contents gauges for anaesthetic machines, did not consider it necessary to provide warning devices for N,0 cylinders (Hospital Technical Memorandum No. 15, 1965). In view of the fact that N.,0 cylinder failure does sometimes temporarily pass unnoticed, and that this is occasionally associated with the hazards detailed above, it seems imprudent not to adopt some form of system which gives warning of the failure of gas delivery from a N,0 cylinder. Though it is true that the pressure within a N,0 cylinder gives no indication of its total contents, this does not constitute an adequate indication for discarding pressure monitoring. In practice, these cylinders fail (i.e. achieve an internal pressure below that which is necessary to maintnin a constant pressure to the needle valves) for two reasons: either because the liquid N,0 has become too cold; or because the cylinder is empty. In both instances, cylinder pressure will give the most reliable warning of failure. The development of frost on a cylinder gives no indication of impending failure since it is purely temperature dependent and merely indicates, at the time of its development, that there is residual liquid NjO present and that the pressure within the cylinder is below 300 Lb/sqin. This situation may occur when a cylinder is still half full if high gas flow rates are employed. Alternatively, it may never develop at all if the emptying of the cylinder is interrupted by re-warming intervals. Although the simple test of inversion of a frosted failed N,0 cylinder will give some indication of whether liquid N 2 O remains within it (fig 5A,B), the only reliable, convenient method of establishing the contents of a cylinder, is to compare its weight with the tare weight stamped on the neck of the pin index valve body, taking care to ensure that ice remaining on the cylinder wall does not falsify the readings. Conventional N 3 O cylinder pressure gauges are unsatisfactory for giving advance warning of failure for a number of reasons. Firstly, since N 3 O cylinders may occasionally achieve temperatures in storage exceeding 65 C, pressures exceeding 2575 Lb/sq.in. might be encountered. Conventional gauges capable of withstanding such pressures are calibrated to 3000 Lb/sq.in. Under normal conditions of temperature such a gauge would be registering only quarter full-scale for a full cylinder, and the measurement of 70 or 80 Lb/sq.in. as an indication of impending cylinder failure is unlikely to be achieved with any degree of accuracy. Furthermore, such a system requires constant attention. It would appear that the most satisfactory warning system would be some form of pressure sensitive device introduced into the circuit and operating an audible alarm. Such a device sited between the cylinder and the reducing valve must be capable of withstanding pressures exceeding 2000 Lb/sq.in. and, possibly, temperatures as low as 60 or 70 C (depending on its precise location) yet be sufficiently sensitive to respond to a pressure difference of only a few Lb/sq.in. In addition, one would have to be provided for each cylinder yoke. A pressure sensitive device operating at slightly below the reducing valve pressure and located between it and the flowmeter block, designed to operate some form of audible warning device and powered by the pressure remaining in the N 5 O cylinder would constitute the most practicable sys-

538 BRITISH JOURNAL OF ANAESTHESIA tern. Such a device, the Ritchie Whistle, is available at moderate cost (Longworth Scientific Instruments Ltd), and is readily fitted to anaesthetic machines which are designed to operate at 35-60 Lb/sq.in. ACKNOWLEDGEMENTS I would like to thank Mr E. K. Hillard, Chief Technician, Mr Roy Saunders, Technician, Department of Anaesthetics, for their technical assistance, and several members of the Department of Anaesthetics for their helpful criticisms during the preparation of <*"*«paper. Finally, I would like to express my gratitude to Professor William W. Mushin, CB.B., without whose help and encouragement this paper would not have been written. REFERENCES Couch, E. J., and Kobe, K. A. (1961). Volumetric behaviour of nitrous oxide. J. Chem. Bng. Data, 6, 229. Hill, D. W. (1972). Physics Applied to Anaesthesia, p. 171. London: Butterworths. Hospital Equipment Information, No. IS, December, 1965. Warnmg Devices and Contents Gauges for Anaesthetic Machines. Lee, J. A., and Atkinson, R. S. (1973). A Synopsis of Anaesthesia, 7th cdn. Bristol: Wright. Macintosh, R. R., and Mushin, W. W. (1946). Physics for the Anaesthetist. Oxford: Blackwell. Minnitt, R. J., and Gillies, J. (1948). Text-book of Anaesthetics, 7th edn, p. 105. Edinburgh: Livingstone. Wylk, W. D., and Churchill-Davidson, H. C. (1972). A Practice of Anaesthesia (3rd edn), p. 281. London: Lloyd-Luke. UNIVERSITY OF BRISTOL DEPARTMENTS OF SURGERY AND ANAESTHETICS Day Release Course (Wednesdays) for F.R.C.S., FJJLR.C.S. and ALR.CO.G. Part 1 1914/15 A Day Release Course in Basic Medical Sciences for the Primary F.R.C.S., F.F.A.R.C.S. and M.R.C.O.G. Part I will be held each Wednesday from September 18 for two full terms. The course will consist of lectures, demonstrations and laboratory practicals, and will include instruction in Anatomy, Histology, Physiology, Pharmacology, Clinical Measurement, Clinical Chemistry and General Pathology, and will be held in the Bristol Royal Infirmary and the University of Bristol Medical SchooL Further information and application forms may be obtained from the Medical Postgraduate Dean, University of Bristol, Senate House, Tyndall Avenue, Bristol BS8 1TH.