Supplementary Certificate of Approval. No S433

Similar documents
Certificate of Approval. No 5/6B/211

All Purpose AP-Series Operation Manual

PRODUCT CONFORMITY CERTIFICATE

Organisation Internationale de Métrologie Légale. Direct mass flow measuring systems for quantities of liquids Annex C: Test report format

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

General Accreditation Guidance. User checks and maintenance of laboratory balances

Laboratory Manual Non Automatic Weighing Devices Page: 1 of 15

Precision pressure sensor Basic version Model CPT6020

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

INTERNATIONAL OIML R 32 RECOMMENDATION. Rotary piston gas meters and turbine gas meters ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

WELMEC 6.4, Guide for packers and importers of -marked prepacked products

Manual Weighingblock VB2 series and Uniscale

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

UNDERSTANDING, ACCELERATED

KERN EG/EW Version /02

Vehicle-mounted meters, pump supplied

Mass Flowmeters. Mass Flowmeters for Gases

OEM Manual. MODEL ½ Digit DRUM SCALE

Guidelines to the standard IEC60601

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e

OXFORD CALIBRATION WEIGHTS CLASS E1, E2, F1

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

MTB Load Cell. Globally Approved for Accuracy and Safety. MTB Load Cell Fully Approved

IA CERTIFICATE (Annual review required by MASC covered by an additional letter) (Revision 2- Review & Review)

Operating Instructions Part No

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Application Note AN-107

Vehicle- or rack-mounted liquefied gas meters, pump supplied

BRISBANE BASKETBALL 2013/14 GBL RULES Championship Season: Updated 16/12/2015

This test shall be carried out on all vehicles equipped with open type traction batteries.

These Terms and conditions apply to audit trails incorporated in weighing and measuring devices and systems 1.

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Organisation Internationale de Métrologie Légale

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

GEMS SENSORS & CONTROLS

Pneumatic high-pressure controller Model CPC7000

Any use of this material except in accordance with a written agreement with Western Power is prohibited.

Pneumatic high-pressure controller Model CPC7000

Supplementary Operator s Manual 42/24-14 EN Rev. 3

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

Specification For Water Meters. Pr10068

Model 601CV. Mass Flow Controller. Product Features:

Australian Standard. Bourdon tube pressure and vacuum gauges AS

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

BA307C & BA308C Intrinsically safe loop-powered 3½ digit panel mounting indicators issue 9

MIL-STD-883G METHOD

Pressure Measurement Single-range transmitters for general applications

SAUTER FK Version /2016 GB

Engineering and Design Rules for Hopper Scales

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

High-speed pressure sensor Model CPT6140

Scoreboard Operator s Instructions MPC Control

TXDOT DESIGNATION: TEX-922-K

TXDOT DESIGNATION: TEX-922-K

Sensoric 4-20 ma Transmitter Board Operation Manual

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

FAS-W. Features. Applications. Automatic condensation hygrometer

Automatic Control Unit. Controlmatic E. Type Series Booklet

Mass Flow Controller (MFC) for Gases

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

Preparation and Installation of the Sanitary BDI-FLX Sensor and Connection to the BDI-FLX Interface Cable

Ultima. X Series Gas Monitor

9700 Transmitter Submersible hydrostatic level transmitter

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

KENYA ACCREDITATION SERVICE

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

Unlimited possibilities. Get started in no time

Hand-held pressure calibrator with integrated pump Model CPH6600

Location of CC Number if not located with the identification:

Grading Policy and Procedures & Rules for State Championships, State Age, Night Inter District and Association Carnivals

PRESCODEM400, PRESCODEM400/220

United Kingdom of Great Britain and Northern Ireland

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

Guideline No.M-05(201510) M-05 AIR COMPRESSOR. Issued date: 20 October China Classification Society

High-performance submersible pressure transmitter For level measurement Model LH-20

Mass Flow Controller (MFC) for Gases

Application Notes. SLP85xD Load Cells

Level transmitter LT 100

2014 Programme of Training Courses in Measurement & Calibration

RESISTORS, HEATERS, FLEXIBLE, SINGLE AND DOUBLE LAYER. ESCC Detail Specification No. 4009/002

I/P Transducer Positioner Module

628 Differential Pressure Decay Leak Tester

Field Certification Certificate of Compliance

FINAL CERTIFICATION RETURN YEAR ENDED ON JUNE 2017 NSW Harness Race Field Information Use

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

Smart Water Application Technologies (SWAT)

Inspection and Testing

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

WELMEC European Cooperation in Legal Metrology

AMB MOISTURE BALANCE AMB 50, AMB 110 and AMB 310. Operating Manual

This manual provides necessary requirements for meeting the IEC or IEC functional safety standards.

No. 2 TEST WEIGHT ASSESSMENT

ARTICLES GOVERNING THE Toyota 86 Championship

U.S. CONSUMER PRODUCT SAFETY COMMISSION Office of Compliance

Transcription:

12 Lyonpark Road, North Ryde NSW 2113 Australia Supplementary Certificate of Approval No S433 Issued under Regulation 60 of the National Measurement Regulations 1999 This is to certify that an approval for use for trade has been granted in respect of the Cardinal Model 220 Digital Indicator submitted by Cardinal Scale Manufacturing Co 203 East Daugherty Street Webb City MO 64870 USA. NOTE: This Certificate relates to the suitability of the pattern of the instrument for use for trade only in respect of its metrological characteristics. This Certificate does not constitute or imply any guarantee of compliance by the manufacturer or any other person with any requirements regarding safety..../2

Supplementary Certificate of Approval No S433 Page 2 CONDITIONS OF APPROVAL This approval becomes subject to review on 1 April 2009, and then every 5 years thereafter. Instruments purporting to comply with this approval shall be marked NSC No S433 and only by persons authorised by the submittor. Instruments incorporating a digital indicator purporting to comply with this approval shall be marked NSC No S433 in addition to the approval number of the instrument. It is the submittor s responsibility to ensure that all instruments marked with this approval number are constructed as described in the documentation lodged with the Commission and with the relevant Certificate of Approval and Technical Schedule. Failure to comply with this Condition may attract penalties under Section 19B of the National Measurement Act and may result in cancellation or withdrawal of the approval, in accordance with the Commission s Document NSC P 106. The values of the performance criteria (maximum number of scale intervals etc.) applicable to an instrument incorporating the pattern approved herein shall be within the limits specified herein and in any approval documentation for the other components. The Commission reserves the right to examine any instrument or digital indicator of an instrument purporting to comply with this approval. Auxiliary devices used with this instrument shall comply with the requirements of General Supplementary Certificate No S1/0/A. DESCRIPTIVE ADVICE Pattern: approved 2 March 2004 A Cardinal model 220 digital indicator. Variants: approved 2 March 2004 1. Model 200 digital indicator. 2. Model 205 digital indicator. 3. Model 210 digital indicator. Technical Schedule No S433 describes the pattern and variants 1 to 3. FILING ADVICE The documentation for this approval comprises: Supplementary Certificate of Approval No S433 dated Technical Schedule No S433 dated (incl. Table 1 and Test Procedure) Figures 1 to 6 dated Signed by a person authorised under Regulation 60 of the National Measurement Regulations 1999 to exercise the powers and functions of the Commission under this Regulation.

TECHNICAL SCHEDULE No S433 Pattern: Submittor: Cardinal Model 220 Digital Indicator Cardinal Scale Manufacturing Co 203 East Daugherty Street Webb City MO 64870 USA 1. Description of Pattern A Cardinal model 220 digital mass indicator (Table 1 and Figure 1) which may be configured to form part of: A weighing instrument with a single weighing range of up to 10 000 verification scale intervals; or A multi-interval weighing instrument with two partial weighing ranges (each with its own verification scale interval) in which case it is approved for use with up to 10 000 verification scale intervals per partial weighing range. Instruments may be fitted with output sockets (output interfacing capability) for the connection of auxiliary and/or peripheral devices. Instruments are powered directly by mains AC power. This approval does not include the use of the indicator as an automatic weighing instrument, unless specifically mentioned in a certificate of approval for such an instrument. TABLE 1 Specifications Maximum number of verification scale intervals 10 000 or 10 000 per range Minimum sensitivity 0.9 µv/scale interval Excitation voltage 12 V DC Maximum excitation current 480 ma 1.1 Zero Zero is automatically corrected to within ±0.25e whenever the instrument comes to rest within 0.5e of zero. Note: For multi-interval operation, zero is automatically corrected to within ±0.25e 1 whenever the instrument comes to rest within 0.5e 1 of zero. The instrument has a semi-automatic zero-setting device (to set the instrument to within ±0.25e of zero) with a nominal range of not more than 4% of the maximum capacity of the instrument. The instrument has an initial zero-setting device with a nominal range of not more than 20% of the maximum capacity of the instrument. 1.2 Tare A semi-automatic subtractive taring device of up to the maximum capacity of the instrument may be fitted. A pre-set taring device of up to the maximum capacity (or of up to the Max 1 for multi-interval instruments) may also be fitted..../2

Technical Schedule No S433 Page 2 1.3 Display Check A display check is initiated whenever power is applied. 1.4 Additional Features The pattern also has certain additional functions (totalising, counting, set point controls (including for filling purposes), under/over checkweighing, weight grading, batching). However this approval does not include the use of the indicator as an automatic weighing instrument, unless specifically mentioned in a certificate of approval for such an instrument. The additional functions (other than the indications of measured mass, i.e. gross, tare, net, totals, displayed either on the indicator or on an auxiliary or peripheral device) are not approved for trade use. 1.5 Markings and Notices Instruments carry the following markings: Manufacturer s mark, or name written in full Cardinal Scale Manufacturing Co Name or mark of manufacturer s agent... Indication of accuracy class Maximum capacity Max... kg #1 Minimum capacity Min... kg #1 Verification scale interval e =... kg #1 Maximum subtractive tare T = -... kg #2 Serial number of the instrument... Pattern approval mark for the indicator NSC No S433 Pattern approval mark for other components... #3 #1 These markings are also shown near the display of the result if they are not already located there. #2 This marking is required if T is not equal to Max. #3 May be located separately from the other markings. In addition, instruments not greater than 100 kg capacity shall carry a notice stating NOT TO BE USED FOR TRADING DIRECT WITH THE PUBLIC, or similar wording. Note: For multi-interval instruments the markings shall be as above, with the exception of the following: Maximum capacity Max.../... kg * Verification scale interval e =.../... kg * 1.6 Verification/Certification Provision Provision is made for the application of a verification/certification mark..../3

Technical Schedule No S433 Page 3 1.7 Sealing Provision Provision is made for the calibration adjustments to be sealed by use of a lead and wire seal to prevent removal of the calibration access screw on the back of the indicator (Figure 2). 2. Description of Variants 2.1 Variant 1 The Cardinal model 200 indicator is a single interval instrument (Figure 3). This model is similar to the pattern but has basic functions only and is in a different style enclosure. This model does not have the pre-set tare facility. 2.1.1 Sealing Provision Provision is made for the calibration adjustments to be sealed by use of one or more lead and wire seals to prevent removal of the two calibration access screws on the back of the indicator (Figure 4). 2.2 Variant 2 The Cardinal model 205 indicator (Figure 5) is a single interval instrument which is similar to the pattern but has basic functions only. This model does not have the pre-set tare facility. 2.2.1 Sealing Provision As described in clause 1.7 for the model 220 (the pattern). 2.3 Variant 3 The Cardinal model 210 indicator (Figure 6) is a single interval instrument which is similar to variant 2 (model 205) but has a different keypad including numeric keys, and has some additional functions (e.g. counting, time). 2.3.1 Sealing Provision As described in clause 1.7 for the model 220 (the pattern)..../4

Technical Schedule No S433 Page 4 TEST PROCEDURE Instruments should be tested in conjunction with any tests specified in the approval documentation for the instrument to which the pattern is connected, as appropriate, and in accordance with any relevant tests specified in the Uniform Test Procedures. Maximum Permissible Errors at Verification/Certification For single range instruments, the maximum permissible errors for increasing and decreasing loads on initial verification/certification for loads, m, expressed in verification scale intervals, e, are: ±0.5e for loads 0 < m < 500; ±1.0e for loads 500 < m < 2 000; and ±1.5e for loads 2 000 <m < 10 000. For multi-interval instruments with verification scale intervals of e 1, e 2, apply e 1 for zero adjustment, and for maximum permissible errors apply e 1, e 2,, as applicable for the load.

FIGURE S433 1 Cardinal Model 220 Digital Indicator

FIGURE S433 2 To Be Sealed Sealing of Cardinal Model 220 Digital Indicator

FIGURE S433 3 Cardinal Model 200 Digital Indicator

FIGURE S433 4 To Be Sealed To Be Sealed Sealing of Cardinal Model 200 Digital Indicator

FIGURE S433 5 Cardinal Model 205 Digital Indicator

FIGURE S433 6 Cardinal Model