Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

Similar documents
AREA VALVE SERVICE UNITS

MANUAL GAS MANIFOLDS

AUTOMATIC GAS MANIFOLDS

DUPLEX PRESSURE REDUCING SET

Zeus Area Valve Service Unit (AVSU) and AVSU Module. Installation, Operation and Maintenance Manual

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Technical Specification

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Installation and Maintenance Manual. Compact Medical Gas Outlets

Flexible Pendant. Installation, Operation and Maintenance Manual

Product Description 3 Cleaning and Lubricating 4 Inspection and Testing 4

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

ATS430 turbidity sensor Retractable insertion assembly

Manifolds & pressure reducing sets

Zeus Vertical Headwall

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

Engineering Data Sheet

K13.12 FACE FIXED PANIC EXIT DEVICE K13 DOORS

BSA6T and BSA64T Stainless Steel Bellows Sealed Stop Valves Installation and Maintenance Instructions

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant

Auto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL

1.0 - OPENING AND CLOSING THE DOOR

Tri-Tech Medical Inc.

INSTALLATION INSTRUCTIONS PUSH BAR PULLMAN PANIC BOLT DEVICE TO EN 1125 : 2008

L230 LED 40 Watt Searchlight

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

k valve 2 (50)HF 2½ (65)SF installation guide aylesbury For valve sizes (DN): tel fax

CRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES

Installation, Operation and Maintenance Manual for Back Pressure Regulator

Manual Actuated Boiler Blowdown Valves

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

SCA Series Inverted Bucket Steam Traps

East SP and Zeus SP Terminal Units Installation, Operation and Maintenance Manual

book : t95575.fm Seite 1 Mittwoch, September 20, :10 AM. Avensys. Exposed Single Lever Mixer

Model VR6 System. Installation, Operation & Maintenance

System Pressure Manager Standard & System Pressure Manager Plus

USM21 Sealed Bimetallic Steam Trap for use with Pipeline Connectors Installation and Maintenance Instructions

AIR DISTRIBUTOR. PART NUMBER (Including, but not inclusive) AD28483-_ALB, AD28563-_ALB, AD36097-_ALB, AD36098-_ALB, AD28485-_ALB

Flexfence Swaging Unit. Product Manual. Release 02/16.

M45 ISO Ball Valve DN25 to 150 Installation and Maintenance Instructions

Pressure Dump Valve Service Kit for Series 2300 Units

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

PO Box 370 Phone: Swansea NSW 2281 Fax: AUSTRALIA Mobile: Web:

Damao Range. Single & Double Line Pressure Regulator - 4 or 8 bar

UBP32 Sealed Balanced Pressure Thermostatic Steam Trap for use with PC_ Pipeline Connectors Installation and Maintenance Instructions

Working Environment Safety Assessment

Contents. 1. General information BROEN BALLOMAX Steel Ball Valves Approvals Quality Management...2

Mounting and Operating Instructions EB 8546 EN. Supply Pressure Regulator Type Fig. 1 Supply pressure regulators

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions

Inverse Proportional Pressure-Relief Cart., Size 2 4

Inverse Proportional Pressure-Relief Cart., Size 2 4

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

Domiciliary Oxygen Switch. Instructions for Use

Proportional Pressure-Relief Cartridge Valve, Size 2...4

Fig 1, Fig 12, Fig 13, Fig 14, Fig 16 and Fig 16L Strainers Installation and Maintenance Instructions

APCO ARV CLEAN WATER AIR RELEASE VALVES. Model 50A

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

Overview of types. T5-R B2/R B2 en v Subject to changes 1 / 4. k vs (Sequence 2)

SERIES-O REGULATOR. Instructions for use

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

TANDEM MODULATOR FOR OUTLET GAS FLOW: STEPPED (836 TANDEM) - CONTINUOUS (837 TANDEM) SERVO-CONTROLLED PRESSURE REGULATOR

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Knob Adjustment)

Mounting and Operating Instructions EB 8546 EN. Type 4708 Supply Pressure Regulators. Translation of original instructions. Type Type

Pressure Dump Valve Service Kit for Series 3000 Units

Installation Instructions for NP100 Series Nitrogen Control Panels

ACCESSORY KIT INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS 1/2 THERMOSTATIC VALVE AND TRIM

Operating manual. for. tapping points of the EM15/EE15 series

Warning: NEVER ALTER TEST EQUIPMENT

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1)

Spiratec ST14, ST16 and ST17 Sensor Chambers and sensors

WARNING: FOR OUTDOOR USE ONLY

Operating Manual. R280 Pressure regulator made of brass. 1. Intended Use

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

Proportional Pressure-Relief Cartridge Valve, Size 2...4

Rp [ ] Sv min. [ ] C215QP-B 15 1/ C215QP-D 15 1/ C220QP-F 20 3/ [ ]

1. General safety information. 2. General product information. 3. Installation. 4. Commissioning. 5. Operation. 6. Maintenance. 7.

OPERATING AND MAINTENANCE MANUAL

Instructions for Use

Installation and Maintenance Manual. Dual Medical Gas Outlets

Inverse Prop. Pressure-Relief Cart., Size 2 4

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

Operating Instructions

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

MUELLER A A Non-Adjustable. Vertical Indicator Posts. Reliable Connections. General Information 2. Technical Data 3.

TEMPERATURE STABILISED THERMOSTATIC SHOWER PANEL

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES

Reduce pressure zone device suitable for high and medium hazard rated applications Flanged end connections

AutoChanger Installation & User Guide Issue 2

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

Transcription:

The Area Value Service Units and Module format (AVSU) provide a local gas isolation facility for use during normal installation and maintenance or in the event of an emergency. They are built in accordance with the HTM2022, HTM02-01, C11 and EN737 standard s and are suitable for Oxygen, Nitrous Oxide, Oxy-Nitrous Oxide, MA4, MA7 and Vacuum. Features: The AVSU can be installed as a flush or surface mounted. Housed in a steel white epoxy coated box and a key operated 90 lock for security. A breakable glass front to the door gives an immediate indication as to the gas type, the valve position and the gas flow. In an emergency the glass front can be broken thus allowing access to the valve. The valve type is a 90 shut off ball valve, which is operated via a handle, and each unit comes complete with a blanking plate, which can be fitted, to either side of the valve. Non-interchangeable NIST connectors are fitted upstream and downstream of the valve. The NIST connectors provide the facility to purge a system before the working gas is introduced and also following any downstream interruption of the supply. The NIST connectors also allow for the pipeline pressure to be tested and provide an easily accessible point for gas sampling. Each NIST include non-return valves with 100% sealing. Side entry sizes available 22mm, 28 and 42mm (other entry versions available) Pressure switches if required can be fitted inside the box to enable local monitoring-see the following picture. All Gas ID labelling is fixed internally and viewed through the door window. AVSU-Pressure switch location Pressure Sensor-Right to Left Flow Pressure Sensor-Left to Right Flow

Operation The AVSU is used to isolate the gas supply in a department/area for maintenance or in an emergency. The gas supply is isolated by rotating the valve handle thro 90 degrees to the closed position. To reconnect the gas supply the valve handle is turned back through 90 degrees to the open position as indicated. The AVSU is used to physically isolate the gas supply, during periods of modification works for example, by inserting a blank spade between the valve and the pipeline, these are located within the box. The NIST connectors situated upstream and downstream of the valve can be used for purging, testing, gas sampling and for the connection of an emergency supply. To connect to a NIST, unscrew the NIST cap and screw on the connecting equipment. It is only possible to connect the correct gas or vacuum NIST fitting. Upon completion of the procedure, the equipment should be removed, the NIST cap replaced, the door closed/locked and the key returned. Authorised Access To gain access to the valve assembly, the key for the door should be obtained from the responsible person. A permit to work may be required depending on the nature of the work to be carried out. It is advisable to close and lock the door of the AVSU if you leave the AVSU during completion of works. Emergency Access If access is required to the valve assembly in an emergency, this can be achieved without having the key to the door by breaking the glass window preferably with a hard object. The key will be required to replace the breakable glass window and return the AVSU to its original secure condition. Safety This equipment should be operated and maintained by personnel who are suitably trained and fully conversant with HTM2022, HTM02-01 and BS EN 737. This equipment should be kept clean and free from oil and grease at all times. Oxygen will ignite spontaneously in the presence of oil and grease. If you suspect that any equipment is contaminated, do not use it. No attempt should be made to modify equipment for use with a gas other than that identified. This equipment should not be operated at pressures exceeding those stated in HTM 2022 and HTM 02-01. Installation

The AVSU may be installed in horizontal or vertical positions. Either way it is advisable to maintain the following sequence from left to right or top to bottom. 02, N2O, 50% O2/50% N2O, Med AIR 4, Surgical AIR, Vacuum The AVSU should be installed at a height that is easily accessible by hand. Consideration must be given to furniture or crash rails that may be installed at a later stage of the construction. Assembly-AVSU Singular units Open the door, remove valve handle and valve cover plate from the AVSU assembly and store in a safe place. Ensure that a cable entry hole is drilled in the box if a pressure sensor is fitted within the box.*ensure the pressure sensors are on the downstream of the unit. Determine the position of the AVSU and secure to the wall utilising the mounting holes on the rear of the box. Braze the copper stub to the fixed pipeline system; take care not to burn the enclosure by taking appropriate means. Pressure test as per the contract conditions eg.htm2022 guidelines. Install the cover plate and valve handle. Assembly-AVSU Module Remove the 2 nd fix form the completed assembly, by unscrewing the 4 screws on the side. Unplug the power cable and pressure sensor leads. Offer up the 1 st fix to the required position on the wall, taking note of the floor to ceiling height as specified on the original specification sheet. Utilising the mounting holes on the 1 st fix Backplate secure to the wall using suitable fixings. Connect to the mains power on a 13amp fused spur. *only suitably trained personnel should undertake this procedure. Ensure the fitted pressure sensors are downstream of the pipeline system. Braze the copper stubs to the fixed pipeline system; take care not to burn the enclosure by taking appropriate means. Pressure test as per the contract conditions eg.htm2022 guidelines. Connect the power and pressure sensor leads from the 2 nd fix to the 1 st fix. Insert the 4 screws on the side of the 1 st fix loosely. Offer up the 2 nd fix fascia cover locating the 4 screws on the side of the fascia into the relevant mounting keys holes on the 2 nd fix.

Flush Mounting Surface Mounting

AVSU Module Mounting Detail

1. Fix the 1 st fix plate to the wall at the required height as confirmed on the original order specification sheet. 2. Insert supplied M5 Domed head screws into one side of the 1 st fix plate ensuring to leave enough thread for the 2 nd fix cover to slide on the slot provided. 3. Using large holes in cover locate the screws and slide cover over the screw head. 4. Slide the cover flush with the wall and tighten M5 Dome head screws. 5. For opposite side slide the cover back to the wall and repeat 4. Maintenance The AVSU is used mainly to assist in isolating the gas supply during maintenance procedures. However, some minimal maintenance may be required on the unit as detailed below: Monthly; Visually inspect the enclosure for signs of damage and check the door is securely locked. Annually; Open the door and clean any debris from the mounting box. Check the valve operates freely. Check the blank spade is in the box for future use. Spare Parts List Item No. Description Part Number 1 22mm ball valve AVSU-22BV 2 28mm ball valve AVSU-28BV 3 22mm Thru Plate 22THRU 4 28mm Thru Plate 28THRU 5 22mm Blank Plate 22BLANK 6 28mm Blank Plate 28BLANK 7 O-Ring KIT AVSU-ORK 8 Blank NIST Cap NISTCAP Due to our policy of continual improvement PRECISION UK Ltd reserve the right to alter dimensions and or specification without notice although every effort will be made to advise of any such modifications.