Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Similar documents
no peaks missing peaks too small peaks increasing retention decreasing retention

GCMSD-Headspace Analysis SOP

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

6890 GC Site Preparation

FROG-4000 TM Chemical Analysis System

Rapid and Reliable Detection of Dissolved Gases in Water

Mass Spec will not Autotune

FROG-4000 TM Chemical Analysis System

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Best Practice for Identifying Leaks in GC and GC/MS Systems

5890II GC Standard Operating Procedure 9/2/2005

FROG-5000TM Chemical Analysis System User s Manual

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Headspace Troubleshooting Made Easy. Presented today by: Susan Scheib Sr. Technical Support Specialist Joe Barta FSE Chicago & Escalation Engineer

This GC has two columns (I think new columns can be ordered from Restek)

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Automated Determination of Dissolved Gases in Water Anne Jurek

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent G Rebuilt Electron Capture Detector

Experiment GC : Analysis of BTEX by GC-FID

Multiple Gas#5 GC configuration Jan 2016

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Agilent 7697A Headspace Sampler

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved

Quick Reference Guide. OPTIC 3 - PTV on Shimadzu GC-2010 Gas Chromatograph

Zebron Gas Management

Agilent 7820A Gas Chromatograph

Figure 1: Plug in ion inlet of mass spec

A SCENT OF FUTURE. Master DHS. Dynamic Head Space Sampler. Operating Manual

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

CP-3800 GC Operator s Manual

AKTA pure 25 New Owner s Intro

System Overview. TCD Detector temperature setpoint Regulator

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Easy Nest Kits. Installation Suggestions

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

CORESTA RECOMMENDED METHOD Nº 67

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

The Tekran Series 2600-IVS Difference!

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

POWER FLUSH KIT INSTRUCTIONS

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Retention Time Locking: Concepts and Applications. Application

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

3 Flow and Pressure Control

Instant Connect Helium Saver Injector Module

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

AcornVac Vacuum Plumbing Systems - Trouble Shooting Guide

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

2B Technologies, Inc.

FAULT CODE TROUBLESHOOTING INDEX

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Automated Liquid Handling Station

CP-3800 GC Getting Started Manual

Buck Scientific, Inc.

Autosampler problems!

AQUAMATIC EASY NEST KITS INSTALLATION SUGGESTIONS

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

User manual. TC-20 Sample conditioning and dry-purge rig. QUI-0003 Version 6.3 September

6890 GC Not Ready Messages (A16010)

Mass Spec will not pump down

Running HPLC Methods on an ACQUITY UPLC System

Prep LC Training Manual

STANDARD OPERATING PROCEDURES

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

New improved robot for gas kinetics

Using the Akta Prime plus October 22, 2012

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Prep LC Training Manual S.V.

The instrument will process the samples one by one till all the samples have been cleaned up.

Simple and Worry Free Gas Filters

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

Solvent System Walkthrough

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Sorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors, Odors

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Transcription:

Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct various sources of contamination How to prevent future hauntings! 2

What is a Ghost Peak? Simply put: any peak that does not belong in your chromatogram. 3

What is a Ghost Peak? Simply put: any peak that does not belong in your chromatogram. 4

What is a Ghost Peak? Less simply put: contamination can present in several ways Elevated baseline Contribution to target peaks Tailing/ splitting 5

What is a Ghost Peak? Less simply put: contamination can present in several ways Elevated baseline Contribution to target peaks Tailing/ splitting 6

What is a Ghost Peak? Less simply put: contamination can present in several ways Elevated baseline Contribution to target peaks Tailing/ splitting Contamination Contribution Clean Peak 7

What is a Ghost Peak? Less simply put: contamination can present in several ways Elevated baseline Contribution to target peaks Tailing/ splitting 8

How to Identify a Ghost Peak Visually detect stray peaks in chromatogram Helps to have an example of a clean chromatogram, or old chromatograms from when the system was clean Helps identify extra peaks, elevated baselines and potentially tailing/split peaks Poor calibration results Can help ID contribution to peaks and split/tailing peaks 9

Troubleshooting Tips Take a moment to review manuals Examine the facts and use valid reasoning Identify the root cause of the problem Avoid quick fixes and shotgun approach BE PATIENT! 10

Exorcising the Demon Step 1: locate the source of the contamination Break system down into major components: Autosampler components, Detector, GC inlet, GC column 11

A Word on Method Parameters GC cycle time Heavy compounds missing from the end of chromatogram Show up in following blank Fix: increase final temperature and/or hold time 12

A Word on Method Parameters Inlet temperature Typically heavy compound carryover, potentially others Check temperature of inlet and/or any transfer lines Fix: ensure temperature is adequate to prevent condensation of compounds, increase if necessary 13

Exorcising the Demon System blanks Run a full system blank Run GC only blank (no autosampler) Run Detector only Blank (no column connected if possible) 14

Solvents/Standards Solvent Methanol, acetone, etc. Direct inject of solvent used to prepare standards Compare results to a second source of solvent Don t forget that syringe could contribute Second GC system to confirm 15

Solvents/Standards Solvent ruled out Standards Crack new ampules of internal and surrogate standards Direct inject Crack new ampules of calibration standard Direct inject Do not forget syringes as source of contamination 16

Carrier Gas Gas filters Gas supply tubing Gas regulators Carrier gas supply http://www.restek.com/catalog/view/3445/22019 17

Carrier Gas Allow GC oven to sit at or below starting temperature for up to several hours Run GC blank (no injection) If carrier gas contamination is focusing on cold column, peaks will be larger If no change in peak size, continue with trouble shooting 18

Carrier Gas Verify filter is not expired (color change indicator) Check for contamination Replace if uncertain Gas supply Tank tubing and regulator Replace all, verify issue resolved, then isolate issue to source If gas contaminated, other components may be contaminated as well 19

Detectors Mass Spectrometers Check the tune report, and/or air-water report Is the detector working properly? Is there sufficient vacuum? Is the source clean? 20

Detectors Other detectors (ECD, TCD) Remove column and cap detector Run a blank and compare detector response to that of blank with column installed Improved: contamination from GC or autosampler Same: contamination from make up gas, detector itself 21

Inlet Maintenance Detector has been ruled out Replace inlet liner, septum Inspect seals for damage/discoloration and replace Clip the column Verify correct column depth and re-install Re-run system blank 22

Other Inlet Issues Split ratio Verify that split ratio valve is functioning properly Sticking or non-opening valve can cause overloading of inlet, column and detector Septum purge Faulty septum purge can allow septum off-gassing to reach detector. Tubing can collect contamination 23

Column Issues Detector and inlet issues ruled out Column clipped during inlet maintenance Swap column with a new one or other column of similar phase Issue resolved, replace column 24

Autosamplers Liquid Injection Headspace Purge and Trap 25

Liquid Injection Autosamplers Syringe Replace/clean- verify no change Cleaning solvent Verify adequate number of cleans between samples with uncontaminated solvent Vials Verify contamination-free sample vials 26

Headspace Autosamplers Needle Transfer line Sample loop Vials 6-port valve 27

Headspace Autosamplers Vials Verify clean, compare to a different lot/manufacturer Needle Often the source of contamination Clean/replace and re-run Trap In dynamic systems, replace trap and re-run to verify 28

Headspace Autosamplers Transfer line, valve oven, sample loop Verify temperatures sufficient to prevent condensation of compounds If condition persists, active site or more persistent contamination 29

Purge and Trap Systems Two main components Concentrator Sparger, sample mount, analytical trap, condensate trap, 6-port valve, transfer line Autosampler Sample needle, syringe/sample loop, transfer tubing, standards injector 30

Concentrator Typically carryover is just late eluters, but contamination can be anything Check temps during Desorb and Bake Ensure desorb of trap and cleanup between samples 31

Concentrator Desorb only Rules out gas pathway If contaminated, change trap and re-run Still have issues, likely an active site in gas pathway, including tubing, 6 port, transfer line, and/or condensate trap 32

Concentrator Disconnect autosampler/bypass autosampler functions Liquid pathway Sparger, sample introduction valve, drain valve Sparger Check for discoloration. Clean or replace, then run blank Check drain valve for proper function 33

Autosampler Check hot water and/or MeOH Rinse Ensure cleanup of liquid path between samples Possible sources of contamination Sample vials Multi-stage needle Syringe/sample loop Standard addition system 34

Autosampler Soil autosampler Contamination in water mode or soil mode? Run each to verify Soil mode only: needle, soil valve, transfer line Water mode only: liquid handling pathway Both: rinse water, syringe, needle 35

Autosampler Syringe based system Bypass needle by placing transfer tubing directly into vial of water Clean/replace syringe Bypass standard addition system Still have issues, multi-port valve 36

Autosampler Loop and valve system Verify loop rinse times, ensure adequate to rinse loop between samples Clean/replace loop Bypass standard addition system Clean/replace transfer tubing 37

Preventative Actions Maintain proper methods Before making method changes, as self Why? Retain records of changes made and why Keep a copy of last known good method If its not broke, don t fix it! 38

Preventative Actions Regular preventative maintenance GC maintenance Inlet maintenance, clip column, regular detector cleaning/tuning Column life Condition columns properly and replace on a regular basis as needed Don t wait until problems arise, waste valuable time troubleshooting! 39

Preventative Actions Autosampler system maintenance Clean/replace syringes, rinse solvents/water and other consumables on a regular basis Analytical traps Same as columns, condition properly and replace regularly as needed Monitor purge and bake pressures, spike in pressure may signal time to replace 40

Preventative Actions Consumables Purchase standards, solvents and vials from reputable manufacturers Refresh standards and solvents regularly to prevent contamination Use good practice when cleaning syringes, flasks and other glassware 41

Preventative Actions Know your samples Consider pre-screening for P&T samples Headspace, prevents contamination from high conc. samples Consider in-vial purge for difficult matrices Consider headspace to prevent column fouling Increase maintenance frequency if/when necessary 42

As Always, We re Here to Help! Don t let troubleshooting scare you, we re here to help! Tech Support 1-800-874-2004 Tekmar_IntlTech@Teledyne.com tekmarsupport@teledyne.com www.teledynetekmar.com Jaocb.Rebholz@Teledyne.com 43