SERVICE MANUAL MEDIVATORS SCOPE BUDDY ENDOSCOPE FLUSHING AID. (For Serial Numbers ) Endoscope reprocessing the way it should be

Similar documents
SCOPE BUDDY USER MANUAL ENDOSCOPE FLUSHING AID Rev D

Rev G Medivators Inc.

REV A International ( ) 2013 MEDIVATORS Inc.

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

PERSONAL AIR BREATHING UNIT

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

KJ4000 Operating Instructions & Parts Manual NSN

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

3800 SERIES END SUCTION PUMP REPAIR PARTS INDEX. Model Model 3804

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408)

TL-12 FLUSH TANK. Manual Ref. # MN-??? Polyurethane Machinery Corp. REVISION 1.0

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

OPERATION. Zinc Sprayer. For Maximum Working Pressure see page 2. Rev. E

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

Revolution Gurney Attachment

CSA Sample Draw Aspirator Adapter Operator s Manual

ROPV R40 E Series User Manual

Vessel Installation and Operating Manual

PC1131 Electric Air Compressor

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

3 Scotch-Weld Polyurethane Reactive Adhesive Applicator Replacement Part Installation Guide

Model 687 Mouse Ventilator User s Manual. Model 687 Mouse Ventilator

INDUSTRIAL VALVES MODELS: C62-A; C62-D. INSTRUCTION MANUAL Installation Operation Parts Service DIAPHRAGM BYPASS PRESSURE REGULATING VALVES

92831 TEL: (714) FAX:

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

INSTALLATION GUIDE Gas Rangetops

KJ1000 Operating Instructions & Parts Manual NSN

TrekMill USER'S MANUAL MM5050

EQUIPMENT OPERATION MANUAL. Loctite Dispense Valve Manual

ATEX ATEX

TOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual

PCS 340 Core Saturator : 115 Volt : 230 Volt. Instruction Manual

Operating Instructions for BAIR22-6 AIR-POWERED CRIMPING TOOL

INSTRUCTIONS FOR MODEL SG6500 GAS HEATERS

INFUSION SYSTEM (110V) (220V) USER S MANUAL

Installation and Operating Manual

ATD /8 x 50 Retractable Air Hose Reel Owner s Manual

Owners Manual Models: 075-AQF-CX. Aquatrol Backwashing Catalytic Carbon Filter. Visit us online at.

Color/Catalyst Dispense Valves

Models: C62/63/64-A/D

FLANGED TWO-PIECE BALL VALVES

INSTRUCTIONS PARTS LIST U.S. PAT: 6,874,404; 6,516,707; 6,223,645

HAS2400 Hepa Air Scrubber

For Models: 55A00-2 (HAI UPB 1500W Wall Switch Dimmer) 55A00-3 (HAI UPB 2400W Wall Switch Dimmer) 37A00-1 (HAI Auxiliary Switch)

RK-IR Sample Draw Aspirator Adapter Operator s Manual

J Air and Water Kit Instructions Part# 02584

Sweeper Dual Control. Automatic Sweeper Switching System SW SERIES SWEEPERS SW-4815 SW-4820 SW-4825 MAGUIRE PRODUCTS INC.

SC505 Sof Care Inflator. Operating Instructions and Service Manual

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

HAI UPB 15A Relay Switch and Auxiliary Switch Installation and Operating Instructions

Acid- Rite ph Adjustment System 2500 by Axiall, a Westlake Company

RPS900W Redundant Power Supply. Installation Guide.

Operating Manual CBA Series Cascade Breathing Air Assemblies

SIX R (P ) AND SEVEN R (P ) POSITION CYLINDERS Service Information

RAM 4021 Operation Manual

FLANGED TWO-PIECE BALL VALVES

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

EQUIPMENT Operation Manual

4 and 8 Station PDS Control Installation & Operators Manual

Accu-Tab Systems 1000 Series by Axiall Corporation

Model 6812C Dual Lane Scoreboard Owner s Manual

GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised )

MODEL CALIBRATION GAS DELIVERY SYSTEM

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

OPERATION / 3Z9305G 1:1 RATIO TRITON / Spray Packages / / / List of Models, page 2 / / / 100 psi (7 bar, 0.7 MPa) Maximum Working Pressure

Digital Melting Point Apparatus

WATER HEATER THERMAL EXPANSION TANKS Owner s Manual. Safety Instructions Installation Maintenance Warranty. Models: 2-5 Gallon Capacity

Freedom8 ShoeBox Compressor Manual

Adapter Kit 5WAR RA NT Y. For use with EZ-Up Gravity Boots and select Teeter Inversion Tables* NEW! Assembly Instructions FULL YEAR

Model ASSEMBLY and OPERATING INSTRUCTIONS

User Instruction Manual

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

Model 6810A Dual Lane Scoreboard Owner s Manual Rev B

Yoke Block Instruction Manual

Budget Range Operators Handbook

Iron Filter Installation / Operation Manual

Cocoa Patio Pond with Lit Spillway

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

Yoke Block Instruction Manual

Research Grade Blood Pressure Transducer User's Manual

SASK-A-POLE OWNERS AND USERS MANUAL

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Operating Manual. Model BAC-30RT30/BAC-30DC

60 farm jack. Distributed exclusively by Harbor Freight Tools Mission Oaks Blvd., Camarillo, CA 93011

User s Guide. VetPro 1000 Mobile Dental Delivery System. For Models:

CHEMICAL INDUCTION KIT (Pt.No Issue 6, October 2015)

Wallmax Professional Aneroid Sphygmomanometer Model 222 User Manual

MANUAL BE SERIES Test Benches

Operating Instructions Part No

453 Series Steam Heated Vaporizing Regulator

IMPACT WRENCH ASSEMBLY AND OPERATING INSTRUCTIONS

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

The Univentor 1250 Anaesthesia Unit

Accu-Tab Systems 2000 P Series by Axiall Corporation

Dual Release Walkers 6291 Series. 1 General. 1.1 Symbols WARNING

Transcription:

SERVICE MANUAL SCOPE BUDDY ENDOSCOPE FLUSHING AID (For Serial Numbers 201345) Endoscope reprocessing the way it should be

is a registered trademark of Minntech Corporation. SCOPE BUDDY is a registered trademark of Minntech Corporation. 50097-067 Rev C 2010, Minntech Corporation All rights reserved. This publication is protected by copyright. Copying, disclosure to others, or the use of this publication is prohibited without the express written consent of Minntech Corporation. Minntech reserves the right to make changes in the specifications shown herein without or obligation. Contact your Medivators representative or Medivators customer service for more information.

TABLE OF CONTENTS PAGE 1 INTRODUCTION Using this Manual 5 Product Description 5 Product Specifications 6 2 QUALITY ASSURANCE TEST Flow Validation Setup and Tubing Prime 8 Performing a Flow Validation Test 10 3 PUMP REPLACEMENT Replacement Instructions 11 4 PARTS LIST Illustration 15 Parts List 16 5 Warranty 17

INTRODUCTION Using this Manual This Service Manual is to be used for servicing the Medivators Scope Buddy Endoscope Flushing Aid having a serial number 201345. A SCOPE BUDDY unit having a serial number 201344 is not serviceable, and should be discarded and replaced with a new unit. The information contained within this manual allows individual service part component identification, and verifies proper Scope Buddy performance and operation through the Quality Assurance Flow Validation Test. It is important to follow the provided instructions to keep the device in good operating condition, delivering proper fluid flows, and to ensuring long product life. Contact the manufacturer s representative if you have any questions regarding this product, its use, operation, or servicing. Product Description The Medivators SCOPE BUDDY Endoscope Flushing Aid is designed to facilitate the flushing of rinse water and cleaning solutions through the channels of flexible, immersible endoscopes. This device is to be used during the manual pre-cleaning phase of endoscope reprocessing, prior to the endoscope being placed in an automated endoscope reprocessor (AER) for high level disinfection. The SCOPE BUDDY combines a diaphragm pump, digital timer, and electronic control circuitry to pump fluids through tubing connections to the channels of the endoscope. Three different tubing sets allow connection to most types of endoscopes. The Elevator Wire Channel Hookup allows connection to the Fujinon endoscope s elevator wire channel, or to the Fujinon endoscope s Forward Water Jet. The DSD Extension Line allows connection to the DSD hook-up which attaches to the endoscope. The Universal Cleaning Adapter Extension Line and multi-way adapters allow connection to the Universal Cleaning Adapter Hook-up provided with the endoscope. The SCOPE BUDDY is designed to be used as a flushing aid only during the manual cleaning process of a flexible endoscope. This device is not an endoscope disinfector. All endoscopes must be high-level disinfected prior to use on a patient. 5

Product Specifications Chassis Dimensions (height - width - depth) Weight (approx.) Power Cord and Connector Electrical Requirements Current Draw Breaker and Protection Maximum Fluid Delivery Maximum Pump Pressure Cycle Time Range Setting Cycle Completion Indicators Electrical Safety Certifications 9 1/2 H x 6 W x 8 D (inches) 24.13 H x 15.24 W x 2032 D (cms) 10 lbs. (4.5 kg) Medical Grade - 6 feet (2.83 m) 18AWG, # 3 Connector 100-250 VAC ± 10% 47-63 HZ 1Ф < 1.0 Amperes 1 AMP max. at 90 VAC input, Non-replaceable, internal primary current fuse, inrush limiting, thermal shutdown. 1 Liter per minute 29 PSI w/safety Bypass 0 to 99 minutes/59 seconds Audible and Visual ETL and cetl Listed Conforms to UL 61010-1 Certified to CAN/CSA C22.2 No. 61010-1 Electromagnetic Interference FCC Title 47, Part 15, Class A, EN 55011 and EN 50082-1 Environmental Rating Designed for Use Installation / Over Voltage Category Classification Standard Indoor Use Only II 1, Ordinary protection Pollution Degree 2 Operating Temperature Range 15ºC to 35ºC Operating Humidity Range 10-75% non-condensing Storage Temperature Range -29ºC to 57ºC Storage Humidity Range Warranty Period 5-90% non-condensing One (1) year 6

QUALITY ASSURANCE The operation and performance of Scope Buddy must be verified anytime a part or component is replaced, after all servicing, and before the first use of the day. The Quality Assurance Test verifies correct Scope Buddy operation, and identifies if an acceptable amount of fluid is being delivered in the specified time. The detailed Quality Assurance Test instructions must be followed to ensure the device is operating properly and delivering the correct volume of fluid. The SCOPE BUDDY Quality Assurance Test is performed using the Flow Verification Tube, Fluid Intake Line, the Universal Cleaning Adapter Extension Line, the SCOPE BUDDY device, and a graduated cylinder. To perform the test, SCOPE BUDDY is preset to pump for a specific period of time and then after pumping, the dispensed fluid is measured to identify if it is within an acceptable volume range. WARNING: Failure to perform the quality test can result in the device not delivering the proper amount of fluid through the channels of an endoscope. Note: A 200cc graduated cylinder is required to perform this test. 7

Flow Validation Setup and Tubing Prime 1. Plug the unit in and turn it on by pressing the power switch located on the side of the unit. The timer display will now illuminate. See Figure 1. Power Switch Figure 1 2. Locate the Fluid Intake Line and attach it to the front panel tubing connection port labeled IN. Place the screened end into the sink or container of clean water. Verify the screened end is fully submerged into the water. See Figure 2. Figure 2 8

3. Locate the Universal Cleaning Adapter Extension Line and attach it to the front panel tubing connection port labeled OUT. See Figure 3. Figure 3 4. Locate the Flow Verification Tube and attach the plastic Luer connector to the remaining end of the Universal Cleaning Adapter Extension Line. Place the open end of the Flow Verification Tube into the sink. See Figure 4. Scope Buddy Unit Fluid Intake Line Universal Cleaning Adapter with Flow Verification Tube Figure 4 Universal Cleaning Adapter with Flow Verification Tube 9

5. Using the SCOPE BUDDY front panel arrow keys, set the timer display to one (1) minute. 6. To prime the device, select the START/STOP button to begin fluid delivery. Watch until all air has been purged through the device and a steady column of water is exiting the flow verification tube. 7. Press START/STOP button to halt fluid flow. Performing a Flow Validation Test 1. Place the open end of the pre-primed Flow Verification Tube into an empty graduated cylinder. 2. Using the SCOPE BUDDY front panel arrow keys, set the timer display to ten (10) seconds. 3. Press the START/STOP button to begin the test and the timer display will now begin counting down. Water will begin flowing into the graduated cylinder. 4. Once the timer counts down to zero, fluid flow will stop and a five-second audible tone will be heard. During this time, the timer display will flash all zeros and then remain with a zero display providing an indication that the cycle is complete. 5. Remove the Flow Verification Tube from the graduated cylinder and read the volume of fluid dispensed into the cylinder. Be sure to read the graduated cylinder directly from the side to avoid parallax, and also read the cylinder from the bottom of the water meniscus. The dispensed volume should be greater than or equal to 100 ml s. 6. If the dispensed volume is above or equal to 100 ml s, then SCOPE BUDDY is ready for use. If the dispensed volume is below 100 ml s, then contact your local distributor or Medivators Technical Support. Do not use SCOPE BUDDY if the dispensed volume is below 100 ml s. WARNING: Do not use SCOPE BUDDY if it fails the Quality Assurance Test. Doing so can result in endoscopes which are not flushed with the correct volume of detergent or rinse water solutions. 10

PUMP REPLACEMENT Follow these step-by-step instructions to replace the SCOPE BUDDY pump. SCOPE BUDDY servicing is to be carried out by a qualified technician. The Quality Assurance Test, as detailed in Section 2, must be performed immediately following any SCOPE BUDDY service work to verify proper system operation, and that SCOPE BUDDY is delivering the correct volume of fluid over time. If SCOPE BUDDY does not operate correctly or deliver the correct volume of fluid during the Quality Assurance Test, then contact your local Medivators representative or distributor for further assistance. Note: Do not use SCOPE BUDDY if it fails the Quality Assurance Test. WARNING: Failure to perform the Quality Assurance Test can result in the device not delivering the proper amount of fluid through the channels of an endoscope. SCOPE BUDDY Replacement Instructions 1. Turn SCOPE BUDDY off by pressing the power switch located on the side of the unit, and then disconnect the power cord from the AC wall outlet and the SCOPE BUDDY unit. 2. Remove the five (5) screws from the back of the unit. See Figure 5. Figure 5 11

3. Lay the SCOPE BUDDY on the back plate and then lift/tip the blue housing forward. See Figure 6. Figure 6 4. Follow the red and black wires from the back of the pump motor to the connector labeled PMP and then disconnect the plug. See Figure 7. Figure 7 12

5. Remove the two elbow connections from each side of the pump head. See Figure 8. Press the retaining collar portion inwards towards the pump while gently pulling the elbow away from the pump. Figure 8 6. Using a 3/8 wrench, remove the four (4) nuts that hold the pump to the back plate. See Figure 9. Figure 9 7. Remove pump from back plate. 13

8. Using a 5/16 wrench, remove the two (2) screws that retain the mounting bracket to the original pump, and then reattach the same bracket to the new pump. See Figure 10. Figure 10 9. Reassemble the SCOPE BUDDY in reverse order. First mount the new pump onto the back plate, and then route the two (2) wires between the back plate and the motor bracket. Reattach and tighten the nuts so they are snug enough to hold the pump motor in place, but not too tight that the rubber mounts are crushed. 10. Connect the pump power connection to the connector labeled PMP. 11. Return the back plate into the blue SCOPE BUDDY housing. Verify that the wires and hoses are not pinched. 12. Reattach the five (5) screws that hold the back plate to the blue SCOPE BUDDY housing. 13. Reattach the power cords and turn system power back on. 14. Run SCOPE BUDDY for several minutes using water to verify there are no fluid leaks. 15. Perform the Quality Assurance Test as detailed in Section 2. 14

PARTS LIST 15

Scope Buddy Service Parts List No. Part Number Description Unit Qty 1 72233-213 Chassis, Scope Buddy ea 1 2 78399-658 Cover, Scope Buddy Medivators ea 1 3 78400-538 Pump Assembly, Service Part ea 1 4 39500-046 Controller, Scope Buddy ea 1 5 37590-241 Harness, Power Jack ea 1 6 37590-242 Harness, Rocker Switch ea 1 7 47509-454 OUT Connector ea 1 8 47509-456 IN Connector ea 1 9 47509-350 Clamp, Hose, #4, Snapper ea 2 10 47509-439 Tubing, 3/16 x 3/8 PVC cm 61 11 37590-390 Clamp, Hose #4, Snapper ea 2 12 47509-447 Foot, Rubber Isolation ea 4 13 43580-154 10-32x3/8 SEMS, PPH, SSTL ea 5 14 34500-074 Resistor 3K 1/4W 5% ea 1 15 CN03-0000 Nut, 10-32 Nylon Lock ea 4 16 CW02-0004 Washer, #10 Flat SS ea 4 37590-321 Main Wire Harness ea 1 18 4-1-532 Connector Male ea 1 19 30597-282 Socket, INIV Mate-N-Lock ea 2 22 17599-490 Foam Tape, Die Cut 1 x 1 rl 1 23 47509-546 Elbow Tube/Hose 3/8 x 1/4 HB ea 2 24 70209-122 PWB Scope Buddy Buzzer ea 1 25 35548-054 Buzzer 24V Pulse PWB MTG ea 1 72233-214 Bracket, Scope Buddy ea 1 38010-017 Power Supply, SB IEC Connector ea 1 50096-381 Manual, Scope Buddy User ea 1 37590-391 Cordset, 120VAC Blk 6 FT ea 1 37590-403 Cordset, AUS/NEW ZEALAND ea 1 37590-370 Cordset, Europe ea 1 50096-814 Scope Buddy Quick Start Chart ea 1 78400-207 CD, Scope Buddy Inservice ea 1 78399-669 Tubing Kit, Scope Buddy ea 1 16

WARRANTY Limited Warranty Medivators Reprocessing Systems, a division of Minntech Corporation ( Medivators ), warrants that the endoscope reprocessor it manufactures (the Product ) will conform to Medivators written specifications and will be free from defects in material and workmanship under normal use and service for a period (the Warranty Period ) of fifteen (15) months from date of shipment from Medivators or one (1) year from the date of installation, whichever occurs first. Notwithstanding anything herein to the contrary, the Warranty Period for consumables and accessories supplied by Medivators, including, but not limited to, endoscope hook-ups, filters, printers, printer supplies, test strips, and accessory bags, is ninety (90) days from the date of installation or 120 days from date of shipment, whichever occurs first. The warranty does not cover, and Medivators will have no warranty obligation whatsoever with respect to, any damage to the Product caused by or associated with: (i) external causes, including without limitation, accident, vandalism, acts-of-god, power failure or electric power surges, (ii) abuse, misuse or neglect of the Product by the customer or use of unauthorized third party filters or other consumables and accessories, (iii) usage not in accordance with product instructions, (iv) the customer s failure to perform required preventive maintenance, or (iv) servicing or repair not authorized by Medivators. Limitation of Remedy The warranty obligation of Medivators hereunder is limited to the repair or replacement (at its option) of the Product or any parts it deems defective. This will be customer s exclusive remedy for a covered defect. In order to recover under the warranty, the customer must notify Medivators if in the U.S.A., or country of installation of the defect (describing the problem in reasonable detail) prior to the expiration of the warranty period and within thirty (30) days of discovery of the defect. Upon receiving Medivators official return goods authorization (RGA), the customer must promptly return the defective part or Product to Medivators (or the service center indicated on the RGA), freight and insurance prepaid. Medivators will not be responsible for any damage during shipment. Warranty Disclaimer THE WARRANTY ABOVE IS ENTIRE WARRANTY OBLIGATION TO PURCHASERS OF THE PRODUCT. IT IN LIEU OF ALL OTHER WARRANTIES OF, EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION, WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, AND DOES NOT REPRESENT OR WARRANT THAT THE PRODUCT WILL MEET CUSTOMER S REQUIREMENTS. RESPONSIBILITY FOR DEFECTS IN THE PRODUCT IS LIMITED SOLELY TO REPAIR AND REPLACEMENT AS SET FORTH IN THIS WARRANTY STATEMENT. TO THE EXTENT PERMITTED BY LAW, MINNTECH CORPORATION SHALL NOT, UNDER ANY CIRCUMSTANCES, BE LIABLE TO CUSTOMER FOR CONSEQUENTIAL, INCIDENTAL, INDIRECT OR SPECIAL DAMAGES OR LOSSES, INCLUDING WITHOUT LIMITATION, DAMAGES ARISING OUT OF OR IN CONNECTION WITH ANY MALFUNCTIONS, DELAYS, LOSS OF PROFIT, INTERRUPTION OF SERVICE, OR LOSS OF BUSINESS OR ANTICIPATORY PROFITS, EVEN IF MINNTECH CORPORATION HAS BEEN APPRISED OF THE LIKELIHOOD OF SUCH DAMAGES OCCURRING. This Warranty gives customer of the Product specific legal rights, and customers may also have other rights which vary from jurisdiction to jurisdiction. In no event shall Medivators liability exceed the original purchase price of the covered Product. No representative or agent of Medivators has any authority to bind Medivators to any other representation or warranty with respect to the Product, and the customer accepts the Product subject to all of the terms above. 17

Medivators Website Resource Center Go to: www.minntech.com/medivators, Select Resource Center and User Library for detailed user guides and hook up matrices, report forms and logs, and product bulletins Medivators Customer and Technical Support Toll Free: Phone: 1-800-444-4729 FAX: 1-866-421-7696 Minntech Corporation 14605 28th Avenue North Minneapolis, MN 55447 USA Toll Free: 1.800.328.3340 Minntech BV Sourethweg 11 6422PC Heerlen The Netherlands Tel: +31.455.471.471 Minntech Japan Corporation 4F Shinjuku Yochomachi Bldg. 10-10 Yochomachi, Shinjuku-ku Tokyo 162-0055 Japan Tel: +81.03.3225.8680 Minntech Asia Pacific Pte Ltd. 1A International Business Park #05-01 Singapore 609933 Tel: +65.6227.9698