UL TEST REPORT AND PROCEDURE

Similar documents
Guidelines to the standard IEC60601

Further technical information is available on

Models EXN and OLG. General Description. Features and Benefits. Pressurizer for Vertical Air Conditioners Increased Safety for Hazardous Locations

Models EXN and OLG. General Description. Features and Benefits. Pressurizer for Vertical Air Conditioners Increased Safety for Hazardous Locations

COVER PAGE FOR TEST REPORT

Key Comparison of Safety Requirements in Europe, North America and Australia for Electronic Products

Leakage Current Testing Is it right for your application?

Zelio Relays RSL Slim Interface Plug-In Relays

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

Consumer Products Services Germany GmbH Türkheim Germany

Operating Instructions

Certificate of Compliance

Field Certification Certificate of Compliance

IECEx Certificate of Conformity

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI

Pressure Measurement Single-range transmitters for general applications

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/16 Siemens FI US Edition

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

Gas density monitor With integrated transmitter Model GDM-100-TI

Supplementary Operator s Manual 42/24-14 EN Rev. 3

Technical Data. Dimensions

Tightness controls TC 1 3 and TC 4

C&G Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations

Technical Data. General specifications. Rated operating distance s n 5 mm

2-1- Shipping Receiving Lifting Location on Site Installation Stages 6

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

Technical Data. Dimensions

C&G 2395 Exam Paper April Section A - All questions carry equal marks. Answer all three questions. Show all calculations.

METROSIL CURRENT TRANSFORMER PROTECTION UNIT (CTPU) ENCLOSURES IEC COMPLIANT

Technical Data. Dimensions

Inspection and Testing Pocket Notes

C&G 2395 Exam Paper June Section A-All questions carry equal marks. Answer all three questions. Show all calculations.

IEC April Authorized Sole Agent : PT. Wisma Niagatama Perkasa. Phone : (031) Fax. : (031)

PRODUCT SPECIFICATION

Compact pressure switches for gas and air

Technical Information Conducal CLY421

TURQUOISE SWITCHES ABJ (BJ) SWITCHES

Dimensions. Technical Data General specifications Switching element function Rated operating distance s n 5 mm

ServicePlus S21R & S27R Series 2 Installation Instructions

Prewired position switches FA series

Differential Pressure Control Types MP 54, MP 55 and MP 55A

Project: Consultant : MEP Contractor: PREPARED BY: DORMAN SMITH SWITCHGEAR - DUBAI REVIEWED BY CONSULTANT : Page 1 of 15

BGA158 Series CE Approved Class A Shutoff Gas Valve

GAS FUEL VALVE FORM AGV5 OM 8-03

MODELS (2 foot Version) (4 foot Version)

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

Galaxy Shore Connection 500 kva Installation 04/

Pressure Measurement Single-range transmitters for general applications

Technical Data. General specifications Switching element function Rated operating distance s n 5 mm

Differential pressure controls, Type MP 54, 55 and 55A

Meta-MEC Manual Motor Starters

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

Karandikar Laboratories Pvt. Ltd., Gat No. 142, Boisar Chilhar Road, At Betegaon, Boisar (E), Taluka Palghar, Maharashtra

It is essential that the maintenance staff is qualified for electrical works and follows safety procedures.

Dimensions. Technical Data General specifications Switching element function Rated operating distance s n 10 mm

User s Manual EJX/EJA-E Series NEPSI Certification [Option code: /NF2, /NS21, /NS24 and /NS25] IM 01C25A00-12E

Reg No: 1999/027771/07 ACCREDITED AND APPROVED TEST LABORATORY IN TERMS OF ARP 0108: REGULATORY REQUIREMENTS FOR EXPLOSION PROTECTED APPARATUS

CERTIFICATION IMPACT ANALYSIS: IEC 60065, SEVENTH EDITION vs. GB Audio, video and similar electronic apparatus Safety requirements

SITRANS P measuring instruments for pressure

Electrical Shore Connections

Operating Instructions

Florida Building Code 6 th Edition (2017) Residential. Broward County Edition Loose-leaf Supplement

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

Technical Data. General specifications Switching element function Rated operating distance s n 2 mm

Bulletin Swimming pool, hot tub and spa installations Rules , , , , , and

Operating Instructions

Pressure Measurement Single-range transmitters for general applications

The Limit Switch with Better Seal, Shock Resistance, and Strength. General-purpose Limit Switch. Model Number Structure

Differential pressure switch MP 54, MP 55 and MP 55A

Electropneumatic Positioner Type 4763 Pneumatic Positioner Type 4765

Technical Data. Dimensions

Wing Union/Hammer Union Pressure Sensors. Datasheet. Models 435/437

RPS900W Redundant Power Supply. Installation Guide.

BGA158 Series CE Approved Class B Shutoff Gas Valve

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Protection of motor circuits

Certificate of Compliance

Technical Data. Dimensions

BGA244 Binary Gas Analyzer

Dimensions. Technical Data General specifications Switching element function DC Dual NC Rated operating distance s n 3 mm

Gas Furnaces 95AF2V. Warranty Heat Exchanger - 20 Year Limited Parts - 5 Year Limited

Modulating Valves for Atmospheric, Infrared, and Direct Fired Burners

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

VLT 2000 Series. How to size your VLT

IMPACT ANALYSIS: GB Safety of Information Technology Equipment

Controller for boilers Galan - regulator for management of heating elements and circuit SolarSentinel-DBTW User guide

Pressure switches for air and water CS

EX-RECESSED CEILING LIGHT FITTINGS

Ventam 85 Installation & Commissioning Instructions

Technical Data. General specifications Switching element function DC Dual NC Rated operating distance s n 3 mm. Short-circuit protection

Data sheet Pressure switch type CS December 2001 DKACT.PD.P10.A B1119

GM Series Dual-Block Multi-Function Gas Control Valves

SMOKE EVACUATION FANS OUTSIDE THE HAZARDOUS AREA

Certificate of Compliance

Pressure Automated Calibration Equipment

To Loughborough University (LU) Facilities Management (FM) Health and Safety Policy

Inductive slot sensor

Miniature controls. Contactors. Overload relays. Starters. Control relays

INSTALLATION GUIDE Gas Rangetops

Transcription:

Issue Date: 2015-01-30 Page 1 of 24 Report Reference # E186249-A270-UL UL TEST REPORT AND PROCEDURE Standard: Certification Type: CCN: Product: Model: Rating: UL 60950-1, 2nd Edition, 2014-10-14 (Information Technology Equipment - Safety - Part 1: General Requirements) CAN/CSA C22.2 No. 60950-1-07, 2nd Edition, 2014-10 (Information Technology Equipment - Safety - Part 1: General Requirements) Component Recognition QQGQ2, QQGQ8 (Power Supplies for Information Technology Equipment Including Electrical Business Equipment) Switching power supply LCM1500Q-T, LCM1500W-T, LCM1500U-T, LCM1500L-T, LCM1500N-T, LCM1500R-T, LCM1000Q-T, LCM1000W-T, LCM1000L-T, LCM1000N-T, LCM1000U-T, LCM1500W-T-5-411 AC Input: 100-240V, 19.0AMax, 50/60Hz For model: LCM1500Q-T +24V, 63.0A Max, 1500W Max; +24V, 31.25A Max, 750W Max; (reverse fan airflow only) For model: LCM1500W-T +48V, 32.0A Max, 1500W Max; forward fan airflow only) +48V, 20.83A Max, 1000W Max; (reverse fan airflow only) For model: LCM1500U-T +36V, 42.0A Max, 1500W Max; For model: LCM1500L-T +12V, 125.0A Max, 1500W Max; For model: LCM1500N-T +15V, 100.0A Max, 1500W Max; For model: LCM1500R-T +28V, 53.6A Max, 1500W Max; For model: LCM1000Q-T +24V, 42.0A Max, 1000W Max; +24V, 31.25A Max, 750W Max; (reverse fan airflow only) For model: LCM1000W-T +48V, 21.0A Max, 1000W Max; Copyright 2017

Issue Date: 2015-01-30 Page 2 of 24 Report Reference # E186249-A270-UL +48V, 20.83A Max, 1000W Max; (reverse fan airflow only For model: LCM1000L-T +12V, 84.0A Max, 1000W Max; For model: LCM1000N-T +15V, 67.0A Max, 1000W Max; For model: LCM1000U-T +36V, 28.0A Max, 1000W Max; For model: LCM1500W-T-5-411 +48V, 32.0A Max., 1500W Max; +5Vsb, 1.0A Max (Optional) Applicant Name and Address: ASTEC INTERNATIONAL LIMITED 16TH FLOOR, LU PLAZA 2 WING YIP STREET, KWUN TONG, KOWLOON, HONG KONG This is to certify that representative samples of the products covered by this Test Report have been investigated in accordance with the above referenced Standards. The products have been found to comply with the requirements covering the category and the products are judged to be eligible for Follow-Up Service under the indicated Test Procedure. The manufacturer is authorized to use the UL Mark on such products which comply with this Test Report and any other applicable requirements of UL LLC ('UL') in accordance with the Follow- Up Service Agreement. Only those products which properly bear the UL Mark are considered as being covered by UL's Follow-Up Service under the indicated Test Procedure. The applicant is authorized to reproduce the referenced Test Report provided it is reproduced in its entirety. UL authorizes the applicant to reproduce the latest pages of the referenced Test Report consisting of the first page of the Specific Technical Criteria through to the end of the Conditions of Acceptability. Any information and documentation involving UL Mark services are provided on behalf of UL LLC (UL) or any authorized licensee of UL. Prepared by: Brian Wong / Project Handler Reviewed by: Steve Chiu / Reviewer Copyright 2017

Issue Date: 2015-01-30 Page 3 of 24 Report Reference # E186249-A270-UL Supporting Documentation The following documents located at the beginning of this Procedure supplement the requirements of this Test Report: A. Authorization - The Authorization page may include additional Factory Identification Code markings. B. Generic Inspection Instructions - i. Part AC details important information which may be applicable to products covered by this Procedure. Products described in this Test Report must comply with any applicable items listed unless otherwise stated in the body of this Test Report. ii. Part AE details any requirements which may be applicable to all products covered by this Procedure. Products described in this Test Report must comply with any applicable items listed unless otherwise stated in the body of each Test Report. iii. Part AF details the requirements for the UL Certification Mark which is not controlled by the technical standard used to investigate these products. Products are permitted to bear only the Certification Mark(s) corresponding to the countries for which it is certified, as indicated in each Test Report. Product Description The equipment is switching power supply, intended for building in as a component used in information technology equipment which employs isolating transformers. Reinforced insulation is provided between primary and secondary. Basic insulation is provided between primary and PE (Protective Earth). Model Differences Models LCM1500W-T, LCM1500U-T are identical to previous tested model LCM1500Q-T except for model number, input / output ratings, transformer's T801, Choke (L1, L2), Fan, heatsink (HTSK1, HTSK3, HTSK4) and minor secondary component changes. Model LCM1500L-T is identical to previous tested model LCM1500Q-T except for model number, output ratings, Power Transformer (T801), Fan, heatsink (HTSK1, HTSK3, HTSK4), Base and Cover Enclosures and minor secondary component changes. Model LCM1000Q-T, LCM1000W-T, LCM1000L-T, LCM1000N-T and LCM1000U-T are identical to previous tested models LCM1500Q-T, LCM1500W-T LCM1500L-T, LCM1500N-T and LCM1500U-T respectively except for model number, output rating and output power. Model LCM1500N-T is identical to previous tested model LCM1500L-T except for model number, output ratings, Power Transformer (T801) and fan source. Model LCM1500R-T is identical to previous tested model LCM1500Q-T except for model number, output ratings and Power Transformer (T801). Model LCM1500W-T-5-411 is identical to previous tested model LCM1500W-T except for model number, 5Vsb current rating with additional 5Vsb auxiliary board, auxiliary transformer (T607) and bridging capacitors (C136, C137) and bridging capacitor on 5Vsb board (C636). Technical Considerations Equipment mobility : for building-in Connection to the mains : To be determined in end system Operating condition : continuous Access location : operator accessible Over voltage category (OVC) : OVC II Mains supply tolerance (%) or absolute mains supply values : +10%, -10%

Issue Date: 2015-01-30 Page 4 of 24 Report Reference # E186249-A270-UL Tested for IT power systems : Yes (for Norway) IT testing, phase-phase voltage (V) : 230 Class of equipment : Class I (earthed) Considered current rating of protective device as part of the building installation (A) : 30 A Pollution degree (PD) : PD 2 IP protection class : IP X0 Altitude of operation (m) : 5000 meters (3048 meters only for Model LCM1500W-T-5-411) Altitude of test laboratory (m) : less than 2000 meters Mass of equipment (kg) : less than 18 The product was submitted and evaluated for use at the maximum ambient temperature (Tma) permitted by the manufacturer s specification of: 50 C at full load up to 70 C at derated output power for the main output (50%). Derated load up to 70deg C is applicable to forward fan airflow only. The means of connection to the mains supply is: To be determined in end system The product is intended for use on the following power systems: TN and IT The equipment disconnect device is considered to be: for consideration at end system The product was investigated to the following additional standards: EN 60950-1:2006 + A11:2009 + A1:2010 + A12:2011 + A2:2013 (which includes all European national differences, including those specified in this test report). The following were investigated as part of the protective earthing/bonding: Printed wiring board trace (refer to Enclosure - Schematics + PWB for layouts) The following are available from the Applicant upon request: Installation (Safety) Instructions / Manual The power supply in this equipment was: Investigated to IEC 60950-1. As part of the investigation of this product, the power supply and its test report were reviewed and found to comply with IEC 60950-1. The Clearances and Creepage distances have additionally been assessed for suitability up to 5000m elevation except for model LCM1500W-T-5-411 which was evaluated for use of up to 3048m only. Engineering Conditions of Acceptability For use only in or with complete equipment where the acceptability of the combination is determined by UL LLC. When installed in an end-product, consideration must be given to the following: The following Production-Line tests are conducted for this product: Electric Strength, Earthing Continuity The end-product Electric Strength Test is to be based upon a maximum working voltage of: Primary- SELV: 391 Vrms, 716 Vpk, Primary-Earthed Dead Metal: 390.6 Vrms, 698 Vpk,, For Model LCM1500L-T,, Primary-SELV: 396.2 Vrms, 716 Vpk, Primary-Earthed Dead Metal: 393.5 Vrms, 725 Vpk,, For Model LCM1500N-T,, Primary-SELV: 394.7 Vrms, 734 Vpk, Primary-Earthed Dead Metal: 395.1Vrms, 734 Vpk.,, For model LCM1500R-T, Primary-SELV: 247.4 Vrms, 460 Vpk, Primary-Earthed Dead Metal: 248.8 Vrms, 457 Vpk. The following secondary output circuits are SELV: +12V/ +15V/ +24V/ +48V/ +36V/ +28V, +5vsb The following secondary output circuits are at hazardous energy levels: +12V/ +15V/ +24V/ +48V/ +36V/ +28V The following secondary output circuits are at non-hazardous energy levels: +5Vsb The power supply terminals and/or connectors are: Not investigated for field wiring The maximum investigated branch circuit rating is: 30 A

Issue Date: 2015-01-30 Page 5 of 24 Report Reference # E186249-A270-UL The investigated Pollution Degree is: 2 Proper bonding to the end-product main protective earthing termination is: Required An investigation of the protective bonding terminals has: Been conducted The following magnetic devices (e.g. transformers or inductor) are provided with an OBJY2 insulation system with the indicated rating greater than Class A (105 C): T101, T801, T302 (ClassF), T607 (Class B, for LCM1500W-T-5-411 only) The following end-product enclosures are required: Mechanical, Fire, Electrical The equipment is suitable for direct connection to: AC mains supply Fans: The fan provided in this sub-assembly is provided with a fan guard to reduce the risk of operator contact with the stator. The secondary output connector has not been evaluated for field connections. The power supply has been evaluated for use in Class 1 equipment as defined in UL 60950-1 Second edition and CAN/CSA C22.2 No. 60950-1-07. An additional evaluation shall be made if the power supply is intended for use in other than Class 1 equipment. This equipment was not evaluated for end system mounting. When installed in the end system, the proper evaluation should be considered. Compliance to the temperature limits of user touchable parts and surfaces of the power supply shall be considered at the end system. For Model LCM1500Q-T: Additional evaluations have been considered for the +24V +/- 10% output voltage adjustability limited to the following combined conditions: maximum allowed 63.0 A output current and 1500W output power. Output voltage adjustability and limited output current/power is also applicable to Model LCM1000Q-T. For Model LCM1500W-T: Additional evaluations have been considered for the +48V -12.5% / +10% output voltage adjustability limited to the following combined conditions: maximum allowed 32.0A output current and 1500W output power. Output voltage adjustability and limited output current/power is also applicable to Model LCM1000W-T. For Model LCM1500U-T: Additional evaluations have been considered for the +36V +/- 10% output voltage adjustability limited to the following combined conditions: maximum allowed 42.0A output current and 1500W output power. Output voltage adjustability and limited output current/power is also applicable to Model LCM1000U-T. For Model LCM1500L-T: Additional evaluations have been considered for the +12V output voltage adjustability from 10V to 13.2V limited to the following combined conditions: maximum allowed 125.0A output current and 1500W output power. Output voltage adjustability and limited output current/power is also applicable to Model LCM1000L-T. For operating temperature up to 50 C, the +12V / +24V / +48V/ +36V/ +28V output of the equipment can be loaded full loading. For operating temperature over 50 C and up to 70 C, the output loading should be decreased by 2.5% per 1 degree rise. The following are the output loading conditions used in the entire testing of model LCM1500Q-T:, Loading 1: +23.81V, 63.0A; +5Vsb, 2.0A, Loading 2: +26.4V, 56.82A; +5Vsb, 2.0A, Loading 3: +24.0V, 62.5A; +5Vsb, 2.0A, Loading 4: +23.81V, 31.5A; +5Vsb, 2.0A (50% load of Loading 1 at ambient temp 70 C), Loading 5: +23.81V, 47.25A; +5Vsb, 2.0A (75% load of Loading 1 at ambient temp 60 C). The following are the output loading conditions used in the entire testing of model LCM1500W-T:, Loading 1: +46.88V, 32.0A; +5Vsb, 2.0A, Loading 2: +52.80V, 28.41A; +5Vsb, 2.0A, Loading 3: +48.0V, 31.25A; +5Vsb, 2.0A, Loading 4: +46.88V, 24.0A; +5Vsb, 2.0A (75% load of Loading 1 at ambient temp 60 C), Loading 5: +46.88V, 16.0A; +5Vsb, 2.0A (50% load of Loading 1 at ambient temp 70 C).

Issue Date: 2015-01-30 Page 6 of 24 Report Reference # E186249-A270-UL The following are the output loading conditions used in the entire testing of model LCM1500U-T:, Loading 1: +35.72V, 42.0A; +5Vsb, 2.0A, Loading 2: +39.60V, 37.88A; +5Vsb, 2.0A, Loading 3: +36.0V, 41.67A; +5Vsb, 2.0A, Loading 4: +35.72V, 31.5A; +5Vsb, 2.0A (75% load of Loading 1 at ambient temp 60 C), Loading 5: +35.72V, 21.0A; +5Vsb, 2.0A (50% load of Loading 1 at ambient temp 70 C). The following are the output loading conditions used in the entire testing of model LCM1500L-T:, Loading 1: +12.0V, 125.0A; +5Vsb, 2.0A, Loading 2: +13.2V, 113.64A; +5Vsb, 2.0A, Loading 3: +12.0V, 93.75A; +5Vsb, 2.0A (75% load of Loading 1 at ambient temp 60 C), Loading 4: +12.0V, 62.5A; +5Vsb, 2.0A, (50% load of Loading 1 at ambient temp 70 C). For Model LCM1500N-T: Voltage adjustability of +/- 10% is applicable for +15V output but limited to the following combined conditions: maximum allowed 100A output current and 1500W output power. Output voltage adjustability and limited output current/power is also applicable to Model LCM1000N- T. The following are the output loading conditions used in the entire testing of model LCM1500N-T:, Loading 1: +15V, 100A; +5Vsb, 2.0A, Loading 2: +16.5V, 90.91A; +5Vsb, 2.0A, Loading 3: +15V, 75A; +5Vsb, 2.0A (75% load of Loading 1 at ambient temp 60 C), Loading 4: +15V, 50A; +5Vsb, 2.0A (50% load of Loading 1 at ambient temp 70 C), Loading 5: +15.0V, 0A; +5Vsb, 2.0A. For Model LCM1500R-T: Additional evaluations have been considered for the +28V +/- 10% output voltage adjustability limited to the following combined conditions: maximum allowed 53.6 A output current and 1500W output power. The following are the output loading conditions used in the entire testing of model LCM1500R-T:, Loading 1: +25.2V, 53.6A; +5Vsb, 2.0A, Loading 2: +28.0V, 53.6A; +5Vsb, 2.0A, Loading 3: +30.8V, 48.71A; +5Vsb, 2.0A, Loading 4: +28.0V, 40.2A; +5Vsb, 2.0A (75% load of Loading 2 at ambient temp 60 C), Loading 5: +28.0V, 26.8A; +5Vsb, 2.0A (50% load of Loading 2 at ambient temp 70 C), Loading 6: +28.0V, 0A; +5Vsb, 2.0A. Additional evaluation has been considered for adding reverse fan airflow direction (air from input/output side going out towards the fan side) at specific loading condition applicable for models LCM1500Q-T, LCM1500W-T, LCM1000Q-T, LCM1000W-T Reverse fan airflow was evaluated at the following limited loading conditions: for model LCM1500Q-T and LCM1000Q-T: +24V, 31.25A max; 5Vsb, 2A, 750W max output power at AC 90 264V at 70 deg C max ambient temperature; for model LCM1500W-T and LCM1000W-T: +48V, 20.83A max; 5Vsb, 2A, 1000W max output power at AC 100 264V at 50 deg C ambient temperature, 900W max output power at AC 90 264V at 70 deg C ambient temperature. The following are the output loading conditions used in the entire testing of model LCM1500W-T-5-411: Loading 1: +48.0V, 31.25A; +5Vsb, 1.0A, Loading 2: +46.88V, 32A; +5Vsb, 1.0A, Loading 3: +46.88V, 24.0A; +5Vsb, 1.0A (75% load of Loading 2 at ambient temp 60 C), Loading 4: +46.88V, 16.0A; +5Vsb, 1.0A (50% load of Loading 2 at ambient temp 70 C). Additional evaluation was considered for model LCM1500W-T-5-411 for power supply inhibit mode condition at 50 C ambient temp using loading 5: +48V (OFF); +5Vsb, 0.5A max only.