Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Similar documents
Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Technical protocol of APMP key comparison for measurement of absorbed dose to water for 60 Co (APMP.RI(I)-K4)

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Bi-lateral Comparison (APMP.L-K ) Calibration of Step Gauge

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Maintenance and general care of secondary standard and field instruments

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

CORESTA RECOMMENDED METHOD N 6

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC #

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

Relative Dosimetry. Photons

Constancy checks of well-type ionization chambers with external-beam radiation units

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

arxiv: v1 [physics.ins-det] 16 Aug 2014

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

FUGITIVE EMISSIONS EXPERIMENTAL MEASUREMENTS AND EQUIVALENCY

THE development of more advanced techniques in radiotherapy,

Review of fundamental photon dosimetry quantities

Organisation Internationale de Métrologie Légale

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

at NIST: ultra-low outgassing rates

Withdrawn at December 31, Guideline DKD-R 6-2. Calibration of Measuring Devices for Vacuum DEUTSCHER KALIBRIERDIENST. Part 3

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

Abstract: Introduction:

Introductory Lab: Vacuum Methods

Pressure Metrology in Mexico

A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF / / / / User Manual

General Accreditation Guidance. User checks and maintenance of laboratory balances

Setting up and carrying out a check source measurement

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

arxiv: v1 [physics.ins-det] 6 May 2016

ISO Vacuum gauges Calibration by direct comparison with a reference gauge

Proton Therapy QA Tools

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

TECHNICAL SPECIFICATION

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

Total Ionization Dose (TID) Test Results of the RH1028MW Ultralow Noise Precision High Speed Operational Low Dose Rate (LDR)

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

WP2 Fire test for toxicity of fire effluents

BRML INM. National Metrology Institute. COOMET Comparison COOMET.L-S20 676/MD/15. Calibration of gauge blocks by comparison. Technical protocol

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Sartorius CCL kg Prototype Mass Comparator

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Technical Information Conducal CLY421

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers

Interlaboratory comparison in the pressure range from 0 to 2 MPa for accredited calibration laboratories

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

PRESSURE TRANSDUCERS WITH UNIFIED (4-20) ma OR (10-50) ma OUTPUT SIGNAL

Determination of Air Density with Buoyancy Artefacts

ON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE*

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Thimble Ionization Chambers

Influence of Reynolds number on Cd of Critical flow venturi nozzle and validation with ISO standards

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES

Please note that there was an error in the initial proposal: samples should be nominally 1 inch in diameter (see below).

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

Final Report of APMP.QM-K46 Ammonia in Nitrogen at 30 µmol/mol Level

FPG8601 Force Balanced Piston Gauge

Test report for UN 38.3 Lithium ion cells

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

Sangil LEE, Mi Eon KIM, Sang Hyub OH, Jin Seog KIM. Center for Gas Analysis Korea Research Institute of Standards and Science (KRISS) Daejeon, Korea

GMC-I Service GmbH Calibration Center

COOMET TECHNICAL COMMITTEE 1.8 PHYSICAL CHEMISTRY

Lab # 03: Visualization of Shock Waves by using Schlieren Technique

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

Performance of the Liquid Hydrogen Target

Commissioning and periodic tests of the Esteya electronic brachytherapy system

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Guide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

ISO INTERNATIONAL STANDARD. Natural gas Measurement of properties Volumetric properties: density, pressure, temperature and compression factor

Gas mixtures. Individual solutions specifically for your application

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Date of Shipment: Xxxxx 00, 20xx 3402c National Institute of Advanced Industrial Science and Technology National Metrology Institute of

Underwater measurement of photosynthetically active radiation

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

PTB Ex PT Scheme. Procedure Instruction of program Explosion Pressure - Test Round 2017

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

Report on Co-production of CH 4 /air within ACRM framework

Characteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

Transcription:

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation This intercomparison will be carried out under the lead of KRISS with the cooperation of ARPANSA. This comparison has been approved by CCRI (I) in May, 2003. 1. Circulate three transfer chambers (two NE2571 from INER and ARPANSA and one PTW 30001 from KRISS), with build-up caps (One of the reasons for circulating three chambers is to be able to assign smaller uncertainty by repeating the same measurement. The other reason is that if one chamber changes during transport or handling, the two that agree can be taken as correct. With only two chambers, if one changes it may be difficult to use either result). 2. The initial stability check of the transfer chambers will be done by KRISS before the circulation of the chambers using 90 Sr source. Each NMI is required to return the chambers to KRISS for the stability check when the measurement is done. The chambers will be forwarded to the next NMI by KRISS after the stability check. 3. Perform calibration for air kerma in a 60 Co beam, at 1 m with a size of 10 cm 10 cm in rectangular beam or = 10 cm in circular beam (however, the beam size doesn t have to be exact. In that case, specify the beam size). 4. The charge collected by the chamber is measured using NMI s electrometer following a pre-irradiation of at least 1 hour. All the measurements are corrected for current leakage. The measurement of result is expressed in mgy/nc. 5. The experimental details regarding the measurement using the chamber is as follows; a) Chamber Positioning : The axis of the chamber is placed in the reference plane at 1 m from the radiation source. The chamber is oriented with the straight line inscribed on the stem facing the radiation source. The chamber center is taken to be aligned with a black line on the build-up cap. No correction is applied for the radial non-uniformity of the beam because an effect for this chamber type is estimated to be less than 2 10-4 in relative value. b) Collecting Voltage : A collecting voltage of 250 V (negative polarity) is applied to the chamber at least 1 hour before any measurement. Collecting voltage should be applied to outer shell of the chamber and circuit diagram for the measurement is described in the figure. c) Ambient Condition : During the calibration the surrounding air temperature is to be stable to better than 0.05 o C. The ionization current is normalized to the reference conditions 20 o C and 101.325 kpa. The relative humidity within the range from 40 to 70 % is highly recommended. d) Reproducibility of measurement : For the determination of chamber calibration factor, a set of ten measurements for each chamber is recommended. 6. Each participant must include the percentage relative standard uncertainty of their air kerma standard as outlined in the ISO document Guide to the Expression of Uncertainty in Measurement, 2 nd edition, 1995. As in this document, uncertainties

should be classified as either type A (determined by statistical methods, with associated degrees of freedom) or type B (determined by other than statistical). All contributing uncertainties should be separately identified and their values tabulated. In particular, uncertainties for the following quantities must be included: For the participants from the primary standard institutions a) Primary standard cavity chamber current measurement. b) Measurement time, electrometer (voltage readout, internal or external capacitance), temperature, pressure and humidity. c) Conversion of primary standard current measurement to air kerma. d) Each correction factor applied in the absolute measurement. e) Transfer chamber ionization current measurement. f) Same as b) but for the transfer chamber measurement. g) Position of the transfer chamber relative to that of the primary standard chamber. Uncertainty component Type A Type B Air kerma rate Ionization current measured by primary standard chamber Sensitive volume of the primary standard chamber Air density under the measurement condition Physical constants used in the primary laboratory - Average energy spent by an electron of charge to produce an ion pair in dry air - fraction of energy loss by bremsstrahlung - Mean mass-energy absorption coefficient ratio between air and graphite - Mean stopping power ratio between graphite and air Correction factors to be applied to the standard - Correction for attenuation and scatter in the wall - Correction for Mean center of electron production - Correction for loss of ionization due to recombination - Correction for scattering from the chamber stem - Correction for humidity - Correction for radial and axial non-uniformity of the beam - Correction for polarity effect Current measurement system Environmental condition under which the measurement is done (temperature and pressure) Position of the primary standard chamber Ionization current of transfer chamber Environmental condition for the measurement with transfer chamber (temperature and pressure) Position of transfer chamber Quadratic summation Combined Standard Uncertainty

For the participants from the secondary standard institutions h) Secondary standard thimble chamber current measurement. (identify the traceability) i) Same as b), plus uncertainty of the calibration factor of the secondary standard chamber. j) Same as b) but for the transfer chamber measurement. k) Position of the transfer chamber relative to that of the secondary standard chamber. Uncertainty component Type A Type B Air kerma rate Calibration factor of secondary standard chamber Current measurement system Environmental condition (temperature and pressure) Position of secondary standard chamber Ionization current of transfer chamber Environmental condition for the measurement with transfer chamber (temperature and pressure) Position of transfer chamber Quadratic summation Combined Standard Uncertainty 7. Each participant should specify the measurement conditions at their laboratory, i.e. distance from the source to the reference plane, central axis air kerma rate, beam size and shape in the reference plane at 50 % of the central axis air kerma rate. 8. Three chambers will be circulating in a single group of 10 participants. Stability check : The stability check of the transfer chambers will be done by KRISS (NMI 0) whenever each NMI finishes the measurement. The chambers will return to KRISS and will be sent to the next NMI after the stability check. Measurement : Three chambers will be circulated through the participating NMI s and will be calibrated by ARPANSA, AIST and BARC before, in the middle and after the circulation. The sequence is as follows; Three chambers calibrated by AIST (NMI 1) Three chambers will be circulated to the participating NMI 0, 2, 3. Three chambers calibrated by ARPANSA (NMI 4) Three chambers will be circulated to the participating NMI 5, 6. Three chamber calibrated by BARC (NMI 7) Three chambers will be circulated to the participating NMI 8, 9. 9. The types of the transfer chamber for the RMO Key Comparison will be NE2571 (#3072, #3259) and PTW 30001-2229.

10. The diagram describing the way the comparison will be conducted is shown below; NMI 1 NMI 2 NMI 9 NMI 8 NMI 3 NMI 0 NMI 7 NMI 4 NMI 5 NMI 6 For each transfer chamber, NMI 1 NMI 4 NMI 7 /3 1 = NMI 1, 2 = NMI 2,. 10 = NMI 0 APMP reference value = 1 2 10 3 10 11. Each NMI is required to notify the result and the condition of the chambers when the measurement is done. If something happens to the transfer chambers during the measurement, the corresponding NMI should return the chambers to KRISS

immediately. If this is prevented by unforeseen circumstances, the pilot laboratory must be informed without delay, as it may be necessary to reschedule the timetable. 12. Each NMI should be responsible for the safe transportation of the chambers as well as delivery fee back to KRISS and KRISS is also responsible for the delivery to the next NMI. It is strongly recommended that the allocated measurement time be 3 weeks. 13. KRISS, ARPANSA, BARC and NMIJ will collect the results and KRISS will be responsible for the summary and report of the final result to APMP. 14. Three BIPM linked NMI s (ARPANSA, AIST and BARC) are required to provide KRISS with the result of Key comparison with BIPM so that KRISS will use them to obtain APMP reference values for the transfer chambers. Figure : Circuit diagram of applying collecting voltage to the ionization chamber Chamber Outer Wall Electrode Input Low s Signal Ground