Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Similar documents
Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design

More Speed and Resolution for Current Methods

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

The instrument should have a mass range from 2 to 2000 m/z or better Interface

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think! There Are Many Reasons to Reduce Analysis Time

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Video Notes. LC Troubleshooting Series Retention Time Shifts

Agilent Dimension Software for ELSD User Manual

Retention Time Locking: Concepts and Applications. Application

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Gradients in ion chromatography. Optimal separation with shorter analysis time

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

LC, RSLC, Bio RS and Dual Gradient Systems

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions

Extended Application Note

HPLC. Solvent Delivery Module Systems

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

DOS in Food Science & Nutrition, Manasagangotri, Mysore

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Method Development Solutions

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

New improved robot for gas kinetics

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Agilent 1290 Infinity Pump Head Maintenance

AKTA pure 25 New Owner s Intro

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Running HPLC Methods on an ACQUITY UPLC System

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

The instrument will process the samples one by one till all the samples have been cleaned up.

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES

Carrier Gases in Capillary GC

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide

Experiment GC : Analysis of BTEX by GC-FID

Setting a New Standard in Flash Chromatography Performance

HPLC troubleshooting guide

User Manual. Agilent 1260 Infinity Analytical- and Preparative-scale Fraction Collectors

Agilent 1220 Infinity II LC Mobile Upgrade Kit

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Buck Scientific, Inc.

Introduction. Experimental

SOP: Buck Scientific BLC-20P HPLC Operation

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Carrier Gases in Capillary GC

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

6890 GC Site Preparation

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

Constant Pressure Inlet (CCN) Operator Manual

GCMSD-Headspace Analysis SOP

Fused-Core TM Particle Columns provide UHPLC performance on any HPLC

ISO 7941 First edition

Peak Biotech - Company & Product Presentation

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

APPLICATION NOTE. GC Integrated Permeation Device

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

May 1, By John Heim,Doug Staples

Prep LC Training Manual

User Manual. Agilent 1260 Infinity Micro-scale Fraction Collector/Spotter

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Your local gas generation partner. Precision series Modular gas generation solution for GC.

FROG-4000 TM Chemical Analysis System

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results

Rapid and Reliable Detection of Dissolved Gases in Water

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

LCMSMS Solutions For Steroid Analysis

Automated Determination of Dissolved Gases in Water Anne Jurek

acculink platinum zv zoning system

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

Measuring range Δp (span = 100%) Pa

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information

System Overview. TCD Detector temperature setpoint Regulator

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

HL2024 Constant Flow Solutions

Supporting information for J. Med. Chem., 1994, 37(5), , DOI: /jm00031a018

Transcription:

Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information contained herein is intended for use by informed individuals who can and must detere its fitness for their purpose. Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC.doc http:// www.chem.agilent.com Page 1 of 1

Evaluation of linear and step gradient performance and retention time precision of the Agilent 12 Series Rapid Resolution LC system Technical Note 3 4 5 Introduction A main performance criterion of any LC system is precision of retention times, which is influenced primarily by the pumping device. It is important that the selected flow is delivered precisely and that for gradient operation the mobile phases are mixed reliably over the complete gradient range. In this Technical Note the precision of retention times for gradient and isocratic applications is evaluated. Further, based on tracer experiments, it is demonstrated that linear and stepwise gradients are delivered with excellent accuracy and precision. The experiments also show that the variability of the delay volume meets the demands of applications using narrow-bore columns as well as standard-bore columns.

Experimental Equipment An Agilent 12 Series Rapid Resolution LC system was used with the following modules: Agilent 12 Series binary pump SL with vacuum degasser for applications using 1.8-µm particle columns up to 15-mm length and with internal diameters from 2.1 to 4.6 mm Agilent 12 Series high-performance autosampler SL for highest area precision Agilent 12 Series thermostatted column compartment SL with wide temperature range from 1 degrees below ambient up to 1 C Agilent 12 Series diode-array detector SL for 8-Hz operation, including new data protection tool ZORBAX SB C-18 columns with different internal diameters and lengths, packed with 1.8-µm particles Pump design The Agilent 12 Series binary pump SL is designed to meet the demands of high throughput and highest performance for optimum resolution and lowest pump ripple for a wide range of applications on narrow or standard-bore columns. Figure 1 shows a schematic of the 12 Series binary pump SL. The delay volume can be configured according to the application needs. The standard configuration is best suited for highest performance in terms of noise and resolution. In this standard configuration the mixer and damper are in the flow path and the delay volume is between 6 and 8 µl, depending on the pressure. This configuration is Figure 1 Standard and low delay volume configurations of the Agilent 12 Series binary pump SL. suited for applications where lowest noise and highest resolution for optimum quantification is needed. For applications where a low delay volume is needed the damper and the mixer can be switched out of the flow path, resulting in a delay volume of 12 µl. This configuration is best suited for application where speed and lowest peak dispersion are needed. To enable these changes the pressure transducer is separated from the damper. Both pump heads have an additional damping coil each of 5 µl volume and an additional outlet valve. This does not add delay volume during gradient analysis because mixing occurs later in the flow path. The pump can be used up to pressures of 6 bars, providing for the use of sub-2- micron particle columns up to 15-mm length. In the standard delay volume configuration the mixer and damper are in the flow path. To change to the low delay volume configuration, the flexible.17-mm id capillary between the mixer and the upper position of the pressure sensor is disconnected. Then, the capillary originally attached to the mixer is connected to the pressure sensor. The damper and mixer are now not in the flow path. Degassing is recommended for all application ranges to ensure optimum performance. The Agilent 12 Series micro degasser SL has 4 channels and an internal volume of 1 ml. Each channel degasses the solvents continuously. The following sections discuss the performance of both pump configurations; with and without mixer and damper in the flow path. The configuration with mixer and damper in the flow path delivers optimum resolution and lowest pump ripple and is best suited for the analysis of complex samples where highest 2

resolution and lowest system noise is needed. The configuration without damper and mixer in the flow path is best suited for LC/MS or LC/UV using 2.1-mm ID columns. Retention time precision of Agilent 12 Rapid Resolution LC system for gradient LC analysis Retention time precision was evaluated for both pump configurations under gradient and isocratic conditions using standard bore and narrow bore columns. In the first example the standard delay configuration was tested using a standard bore column.a phenone mix was analyzed using gradient conditions from 35 to 95 %. A 1 x 4.6 mm column and a flow rate of 1.5 ml/ were selected. These conditions are close to conditions used in conventional LC analysis. Figure 2 shows 1 runs overlaid, demonstrating the retention time precision obtained for this application. The run time was 6. The evaluation of the retention time precision was found to be less than.7 % RSD. ma 25 2 15 1 5 DAD1 DAD1 A, A, Sig=245,1 Sig=245,1 Ref=36,8 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575113.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575114.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575115.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575116.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575117.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575118.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575119.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\557512.D) DAD1 A, Sig=245,1 Ref=36,8 (E:\CHEM32\1\DATA\INJECTIONVOL2\5575121.D) (E:\CHEM32\1\DATA\INJECTIONVOL2\5575122.D) Overlay of 1 runs Injection volume: 5 µl Injected amount: 8 ng RSD RT <.7 % Figure 2 Overlay of 1 chromatograms. 1 2 3 4 5 Chromatographic conditions: Test sample: Set of 9 compounds, 1 ng/ul each, dissolved in water/acn (65/35) 1. Acetanilide, 2. Acetophenone, 3. Propiophenone, 4. Butyrophenone (2 ng/ml), 5. Benzophenone, 6. Valerophenone, 7. Hexanophenone, 8. Heptanophenone, 9. Octanophenone Column: 1 x 4.6 mm ZORBAX SB C-18, 1.8 µm for 6 bar operation Pump: Solvent: A = Water, B = ACN Gradient: 35 to 95 %B in 5, hold at 95 %B for 1 Stop Time: 6 Post Time: 2 Flow Rate: 1.5 ml/ Autosampler: Injection volume: 5 µl Wash 1 s for exterior of needle Thermostatted column compartment: Temperature: 5 C Detector: 13 µl cell 2-Hz data acquisition Peak width = <.1 Slit: 8 nm Signal: 245/1 nm, ref 36/8 nm 3

Retention time precision of Agilent 12 Rapid Resolution LC system using the low-delay volume configuration In the previous section the retention time precision of the standard delay configuration was evaluated. In this section the retention time precision of the low volume delay configuration will be evaluated. The analyses were run at high and low flow rates as well as at high and low temperatures. For all low flow rate applications the relative standard deviation for the retention times was typically less than.7 %. For high flow rate and ultra-fast applications the relative standard deviation for the retention times was typically less than.14 %, see figure 3 and also literature reference 1. Limit QA/QC =.5.45 % RSD.4.35.3.25.2.15.1.5. RT Precision (n=1) Flow=.35 ml/, T=4 C Flow=.35 ml/, T=8 C Flow=1.2 ml/, T=4 C Flow=2. ml/, T=8 C Acetanilide Acetophenone Propiophenone Butyrophenone Benzophenone Valerophenone Hexanophenone Heptanophenone Octanophenone.67 %RSD, SD=.17.7 %RSD, SD=.84.98 %RSD, SD=.48.134 %RSD, SD=.31 Average RSD Figure 3 Statistical evaluation of the retention time precision for the repeated analysis of a nine-compound mixture under various conditions using the low delay configuration of the Agilent 12 Series Rapid Resolution LC system. Chromatographic conditions: Solvent: A = Water, B = ACN Temp.: 4 C, 8 C Flow:.35 ml/, 1.2 ml/, 2. ml/ Gradient:. 35 %B 2.6 95%B 3.2 95%B 3.21 35%B Time values for F =.35 ml/. For all other flow rates times are scaled so that (time x flow) =.9 ml Stop time: 3.2 Post time: 2. Wavelength: 245 nm (8), Ref. 45 nm (1) Slit: 8 nm Peak width: >.25 (.5 s response time), 8 Hz Spectra: All, 19-5 nm, bandwidth = 1 nm Injection vol.: 1. µl Injector: Overlapped injection, Automatic delay volume reduction, Sample flush out factor = 1 Needle wash = 5 s Sample: Phenone mix (.1 µg/µl each,.2 µg/µl of Butyrophenone) Low flow Low temp. mau F=.35 ml/ T=4 C 4 35 3 25 2 15 1 5.5 1 1.5 2 2.5 3. High flow Low temp. mau F=1.2 ml/ T=4 C 3 25 2 15 1 5.2.4.6.8 Low flow High temp. mau F=.35 ml/ T=8 C 4 35 3 25 2 15 1 5.5 1 1.5 2 2.5 3 mau High flow High temp. F=2. ml/ T=8 C 35 3 25 2 15 1 5.1.2.3.4.5 Figure 4 Overlaid chromatograms of the repeated analysis of a 9-compound mixture under various conditions. 4

Norm. 12 1 8 6 4 2 2.75 2.8 2.85 2.9 2.95 3 Chromatographic conditions: Sample: Isocratic standard sample, 1. dimethylphthalate.15 wt%, 2. diethylphthalate.15 wt%, 3. biphenyl.1 wt%, 4. o-terphenyl.3 wt% Column: 2.1 x 1 mm, ZORBAX SB C-18, 1.8 µm Mobile phase: Water/acetonitrile = 3/7 Injection vol.: 3 µl Flow rate:.5 ml/ ma 2 15 1 Overlay of 1 runs Injection volume: 3 µl RSD for RT <.7 % Retention time precision of Agilent 12 Rapid Resolution LC system for isocratic LC analysis In this example isocratic conditions were selected, see conditions in figure 5. A narrow bore column was chosen and the flow rate was set to.5 ml/. In figure 5 an overlay of 1 runs is shown for the analysis of test mixture. Typically the relative standard deviation for this type of application is less than.7 % for retention times. Delay volume, system ripple, precision and accuracy of gradients To evaluate pump performance tracer experiments are a frequently used tool to evaluate the system ripple at different gradient mixtures. The delay volume and the accuracy and the precision of gradients are also evaluated using step gradients and linear gradients. Figure 6 shows a step gradient from to 1 % in 1 % steps using the Agilent 12 Series Rapid Resolution LC system configured for standard delay volume. 5.5 1 1.5 2 2.5 3 3.5 Figure 5 Precision of retention times on 2.1-mm column, isocratic elution, overlay of 1 runs, RSD of RT less than.7 %. Norm. 2 15 1 5 Overlay of theoretical and real step gradient from to 1 % in 1 % steps 21 Bar backpressure Ripple on step 1: =.1 % Ripple on step 3: =.7 % Ripple on step 5: =.6 % Ripple on step 8: =.1 % Delay volume with ALS in the flow path = 86 µl 1 2 3 4 5 6 7 Figure 6 Accuracy, delay volume and ripple using the standard delay configuration of the Agilent 12 Series Rapid Resolution LC system at 21 bar backpressure, to 1 %B gradient in 1 % steps. Chromatographic conditions: Column: Restriction capillaries Mobile phases: A=water, B=water/.5 % acetone Flow rate: 1 ml/ Step gradients: %B to 1 %B in 1 % steps, each step held for 5 %B to 1 %B in 1 % steps, each step held for 5 Linear gradient: 1 % to % in 5 Temperature: 36 C Detection: Signal 265/2 nm, reference 36/1 nm Slit 4 nm Peak width.1 13 µl cell 5

With the autosampler in the flow path the delay volume is about 86 µl. The system ripple is less than or equal to.1 % related to the 1 % step. The overlay with theoretical step gradient shows excellent agreement with the real step gradient with regard to the step heights. Figure 7 shows an overlay of 3 step gradients using the standard delay volume configuration of the Agilent 12 Series Rapid Resolution LC system. The steps are 1 % from to 1 %. The precision for the 3-step gradient runs is excellent and the system ripple for the 1 % steps is less than or equal to.11 % related to the 1 % step. Figure 8 shows an overlay of 3 linear gradients using the low delay volume configuration of the Agilent 12 Series Rapid Resolution LC system. The linear gradient was run from 1 % to % tracer. The overlay shows that precision for a linear gradient is also excellent for the low delay volume configuration. The delay volume in this configuration is about 42 µl with autosampler in the flow path. If the autosampler is taken out of the flow path the delay volume is about 12 µl. The system ripple with the autosampler in the flow path is typically less than.5 % related to the 1 % step. mau 2 17.5 15 12.5 1 7.5 5 2.5 Overlay of 3 runs Step gradient from 1 to 1% in 1% steps Delay volume with ALS in the flow path = 945 µl 21 Bar Backpressure Ripple on step 1: =.11 % Ripple on step 3: =.9 % Ripple on step 5: =.11 % Ripple on step 6: =.1 % Ripple on step 9: =.1 % 1 2 3 4 5 Figure 7 Precision and ripple using the standard delay configuration of the Agilent 12 Series Rapid Resolution LC system at 21 bar backpressure, to 1 % in 1 % steps. mau 2 15 1 5 62 64 66 68 7 72 74 Figure 8 Precision using the low delay volume configuration of the Agilent 12 Series Rapid Resolution LC system at 2 bar backpressure, linear gradient from 1 to %, 3 overlays. 6

Conclusion The data presented in this note show that the precision of retention time of the Agilent 12 Series binary pump SL pump is excellent for a wide range of LC applications using either narrowbore or standard-bore columns. Typically a precision for retention times of less than.7 % relative standard deviation can be expected for isocratic and gradient applications using the standard delay volume configuration of the pump. For the low delay volume configuration the retention time precision is typically less than.7 % relative standard deviation for more conventional flow rates and is typically less than.14 % for ultra-fast applications with run times less than 1. References 1. Performance characteristics of the Agilent 12 Series LC system, Agilent Application Note, Publication Number 5989-4489EN, 26. Accuracy and precision of linear and step gradients are excellent. Ripple for the standard delay volume configuration is typically less than or equal to.1 %. The low delay volume configuration has a system ripple typically less than.5 %. The delay volume variability supports the needs for low delay volume applications as well as the demands for delay volume needed for more conventional validated methods. The standard delay volume configuration with the autosampler in the flow path has a delay volume of 86 µl at 2 bar backpressure. The low delay volume configuration has a delay volume of 12 µl when the autosampler is switched out of the flow path. 7

Angelika Gratzfeld-Huesgen and Michael Frank are Application Chemists at Agilent Technologies, Waldbronn, Germany. www.agilent.com/chem/12rrht Copyright 26 Agilent Technologies All Rights Reserved. Reproduction, adaptation or translation without prior written permission is prohibited except as allowed under the copyright laws. Published April 1, 26 Publication Number 5989-531EN