Series 7250 Ruska High-Speed Digital Pressure Controller. GE Sensing. Features

Similar documents
Series Ruska Digital Pressure Indicator. GE Sensing & Inspection Technologies. Features

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Additel 761 Automated Pressure Calibrators

Additel 761 Automated Pressure Calibrators Selection Guide

Additel 761 Automated Pressure Calibrators

Additel 761 Automated Pressure Calibrators

Additel 761 Automated Pressure Calibrators Selection Guide

Pressure calibration. 20

Additel 780 Series Pressure Controller

Additel 780 Series Pressure Controller

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

Pneumatic High-Speed Pressure Controller Model CPC3000

ADT 761 Automated Pressure Calibrator

Additel 780 Series Pressure Controller

Pneumatic high-pressure controller Model CPC7000

Pneumatic high-pressure controller Model CPC7000

Industrial pressure controller Model CPC4000

DPI 610/615 Series. Druck Portable. Pressure Calibrators. GE Sensing. Features. Setting the Standard for Portable. Pressure Calibrators

628 Differential Pressure Decay Leak Tester

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

PACE Modular Pressure Controller

Industrial pressure controller Model CPC4000

Hand-held pressure calibrator with integrated pump Model CPH6600

The PACE6000. Modular Pressure Controller/Indicator. GE Measurement & Control Solutions. Features

High-speed pressure sensor Model CPT6140

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

PACE1000 Precision Pressure Indicator

Precision digital pressure gauge Model CPG1000

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Digital deadweight tester Model CPD8500

Additel. Overvie. vacuum. Standalone. to 200 bar. products for. to use. The establishes. the future. Calibration Experts EN ISO / IEC

Modular pressure controller Model CPC6050

FPG8601 Force Balanced Piston Gauge

DO NOT PRINT THIS PAGE

ADT. Series. oller. Overvie. vacuum Removable. Standalone. Easy-to-use. products for. to use. The establishes

HumiSys HF High Flow RH Generator

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

DPC-30 DPC-100. Reference Manual

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05


Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

PPC4 Pressure Controller/Calibrator

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Precision pressure sensor Basic version Model CPT6020

PPC2+ BG0002. Operation and Maintenance Manual (Use with PPC2+ Operation and Maintenance Manual) 2001 DH Instruments, Inc.

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Digital dead-weight tester Model CPD8500

PPC4 Pressure Controller/Calibrator Operation and Maintenance Manual

Four common PRESSURE CALIBRATION PAIN POINTS

PPC2 AF. Operation and Maintenance Manual

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

PRESSURE MEASUREMENT DPI 515 PRECISION PRESSURE CONTROLLER/CALIBRATOR OPERATOR'S MANUAL K245. K245 Issue No.2

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Pressure & Temperature Calibration Equipment

E-DWT-H Electronic Deadweight Tester

陕西创威科技有限公司. Shaanxi Creat Wit Technology Co., Ltd. Precision pressure calibrator-cwy300 Operating Manual

Automatic Isokinetic Sampler. ST5 Evo

PTG100 Precision Test Gauge

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

High Precision Gauge Pressure Indicators

BAPI Pressure Line of Products - FAQs

Pressure Calibrators. Series PC..., PM...

Pressure Calibration Solutions

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

SPD Pressure Sensor Families

Pressure Calibration Equipment

Single Pressure/ vacuum decay leak rate, loss, rate of loss, or occlusion tests with standard Cv valves.

Automated Mass Handling for High Performance Pressure Balances

Oxidation Stability of Gasoline and Aviation Fuels

Design Features. General Description

Pressure Automated Calibration Equipment

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Pneumatic service kit Model CPG-KITP

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

WTSeries C R Y S T A L. Digital Pressure Gauge. Operation Manual. engineering corporation

TOPIC: KNOWLEDGE: K1.01 [2.5/2.7] An atmospheric pressure of 15.0 psia is equivalent to... A psig. B psig. C psig.

UNIVERSAL MASK TEST SYSTEM

Cecomp. Electronics. Test. Measure. Monitor. Transmit. Alarm. Cecomp High Perfor. formance Digital Pressur

Mitos Fluika Pressure and Vacuum Pumps Datasheet

UNIK 5600/5700 Marine Certified Pressure Sensing Platform

Powers Controls RC 195 Single Input Receiver-Controller

EPR High Precision 3000 psi Electronic Pressure Controller Accurate to 0.25% of full scale

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Pressure Technical Information

PPC4 Pressure Controller/Calibrator

Definition - Pressure

Digital pressure gauges

Selecting the right pressure sensor for your application

Type 4660 High-Low Pressure Pilot

86BSD Backside Digital Output

Digital Pressure Gauge Main features: Data Sheet: BIT02B.417.R6.EN COVER on request

Auto-Zero Calibration Technique for Pressure Sensors

Fisher 4660 High-Low Pressure Pilot

Transcription:

Features Pressure ranges from 0 to 5 and 0 to 3000 psi (0 to 400 mbar and 0 to 210 bar) xi and 7250i provide advanced precision of 0.005% of reading provides 0.003% of full scale precision Stability: 0.0075% of reading per year Time to setpoint: 20 seconds with no overshoot Control stability: 10 ppm Active matrix color screen with enhanced navigation menus The Series 7250 high speed digital pressure controllers represents the fifth generation of automatic pressure controllers from GE and offers unmatched performance, an active matrix color screen and enhanced control stability. Utilizing multiple sensors and ranges in a single instrument, the Series 7250 combines precision, stability, speed and affordability. Models 7250xi, 7250i and 7250, up to 2500 psi (172 bar), use GE s unique quartz sensor, the most accurate pressure sensing technology in a pressure calibrator. Each quartz sensor is manufactured and tested to provide the ultimate performance required by a GE pressure calibrator, ensuring that every customer receives quality, precision and stability in their instrument. Series 7250 Ruska High-Speed Digital Pressure Controller Series 7250 is a Ruska product. Ruska has joined other GE high-technology sensing businesses under a new name _ GE Industrial,. g

Advanced Precision Models 7250xi and 7250i offer advanced percent of reading precision for increased capability with a single instrument, reducing the investment required to calibrate a wide variety of pressure devices and ranges. The xi provides 0.005% of reading precision from 5% to 100% of the instrument s range. This unmatched precision is achieved through unique quartz pressure sensing technology and multiple quartz sensors in a single instrument. Various ranges from 0 to 20 and 0 to 2500 psi (0 to 1.6 and 0 to 172 bar) are available. The i provides 0.005% of reading from 25% to 100% of range and is available in various ranges from 0 to 5 and 0 to 1000 psi (0 to 400 mbar and 0 to 69 bar). The Series 7250 features GE s unique fused-quartz sensor. This rugged transducer offers unequalled precision and a stability of 0.0075% of reading per year. For pressure below the lower threshold of 5% for the 7250xi and 25% for the 7250i, the precision becomes 0.005% of the lower value. For example, a 1000 psi (69 bar) xi provides 0.005% of reading from 50 to 1000 psi (3.44 to 69 bar); the precision for pressures from 0 to 50 psi (0 to 3.44 bar) is 0.005% of 50 psi (3.44 bar). xi Precision Stability (1 year) Calibration Standard Environmental: Temperature (included in precision) Uncertainty (2 sigma) 0.005% of reading 0.0075% of reading 0.0010% of reading 0.000% of reading Selecting the full scale range of the 7250xi and 7250i to purchase is simplified with the percent of reading capability. Simply determine the highest pressure required and then decide if 0.005% of reading is required down to 25% or 5% to maintain the appropriate calibration ratio. The complex analysis associated with selecting the appropriate single or multirange instruments with percent of full scale performance is no longer required. And, since the performance of the 7250xi and 7250i is continuous throughout the range, the time consuming task of controlling to atmosphere and switching ranges is eliminated. The 7250xi and 7250i not only provide unequalled precision, but also excellent long term stability due to the inherent properties of quartz in combination with several enhancements GE has implemented over the last few years. The total uncertainty of the 7250xi and 7250i over a one year calibration interval is 0.009% of reading down to 5% and 25%, respectively (see the table, above right). GE s simplified specifications eliminate guesswork and specmanship to allow you to determine the actual performance delivered with every instrument. Head Pressure Control (passive mode) Expanded Uncertainty (2 sigma) Standard Precision 0.001% of reading 0.000% of reading 0.009% of reading For applications that do not require the level of performance provided in the 7250xi or 7250i, the Model 7250 offers an economical approach to automatic pressure testing and calibration with a precision of 0.003% of full scale for ranges to 2500 psi (172 bar). A 3000 psi (210 bar) full scale range is available with 0.01% of full scale precision, or select the triple scale option that provides 0.01% of 1000, 2000 and 3000 psi (70, 140 and 210 bar). High-Speed Pressure Control All Series 7250 instruments up to 2500 psi (172 bar) reach setpoint in 20 seconds or less into a 15 cubic inch (245 cc) volume, with no overshoot, to allow high-speed pressure test and calibration.

Dual Control Modes The Series 7250 also provides two user-selectable control modes: active and passive mode. In active mode, the 7250 is continually maintaining the setpoint and can compensate for small leaks and pressure changes due to temperature. In passive mode, the user defines a control band and the 7250 will turn off the controller once it achieves the setpoint within the control band. In a leak-free and temperature-stable system, passive mode contributes no additional uncertainty, providing optimum performance. Onboard programs _ for frequently used or lengthy calibrations, the Series 7250 can store up to 20 user-defined programs/profiles with up to 1000 steps total in internal memory. Computer interface _ every Series 7250 is provided with both an RS232 and IEEE-488 interface, and Series 7250 syntax follows SCPI protocol for easy programming. Intecal, an off-the-shelf software package is available in addition to the LabVIEW driver, a free download and optional MET/CAL driver. As a standard feature, software written for GE s previous generation Series 7215, 7010 and 6000 instruments is fully supported by the Series 7250. The Series 7250 can also be set to 510 emulation mode to use software originally written for the DPI 515. Firmware updates can also be performed over the RS232 interface (updates can be downloaded from the web site). The Series 7250 features multi-lingual menus and displays. Automating Pressure Test and Calibration The 7250xi, 7250i and 7250 are easy to use and can automate your calibrations in several ways: Step up/down _ for calibrations where the increments are fixed intervals, enter a user-defined step value. The Series 7250 increases or decreases the pressure by the step amount with the jog dial _ no more lengthy keystroke sequences to program. Sweep test _ for simple exercising routines, as with dial gauges, enter a start value, a stop value, and number of times to repeat the cycle. The Series 7250 will automatically exercise the device under test prior to the calibration run. The Series 7250 can be used with Intecal software to perform fully automated calibrations. Intecal includes drivers for popular DMMs, multiplexers and environmental chambers, allowing a variety of configurations and options for maximum flexibility and capability. Please refer to the Intecal datasheet for additional information. Versatility to Handle Any Pneumatic Pressure Calibration The Series 7250 is versatile enough to handle almost any type of pneumatic pressure calibration. Wide pressure range _ available in a variety of standard or custom full scale pressure ranges from 5 to 3000 psi (400 mbar to 210 bar)

Pressure units/scales _ select from over twelve standard units of measure, including inhg at 0 C and 60 F, kpa, bar, psi, inh 2 O at 4 C, 20 C and 60 F, kg/cm 2, mmhg at 0 C, cmhg at 0 C, and cmh 2 O at 4 C, and two user-defined units The Series 7250 high speed digital pressure controllers can easily automate your test and calibration workload. All are easy to use, easy to maintain, and have the reliability, performance and features that you want. Head pressure _ automatic correction for head pressure differences Absolute mode _ the 7250i and 7050 offer three different methods to make absolute pressure measurements. The barometric reference option provides the most convenient method and is available on ranges 15 psi (1.0 bar) and higher. Alternatively, the vacuum reference option allows the connection of an external vacuum pump to the reference port of the instrument. An onboard vacuum sensor monitors the reference vacuum and allows for automatic zeroing in absolute mode. This option provides the lowest overall uncertainty since it does not include the additional uncertainty of a secondary barometric reference sensor. For pressure to 50 psia (4 bar) permanent absolute models are also available. The 7250xi is only available with the barometric reference option for absolute measurement. The Series 7250 features an easy-to-navigate menu structure with full text descriptions for menus and commands. The large color display allows the pressure value to be displayed even when viewing a submenu selection such as the units selection screen shown above. Autovent and autozero _ with few keystrokes, the Series 7250 will vent the test port to atmosphere or automatically zero itself (autovent is not applicable to permanent absolute models) Protection of the device under test _ set upper and lower pressure limits to ensure protection of the device under test Options The Series 7250 can be provided for gauge mode operation, or with: Optional vacuum (negative gauge) mode for bidirectional devices Optional barometric reference for absolute mode calibrations Permanent absolute ranges to 50 psia (4 bar) full scale which include a tare feature for simulated gauge mode operation Triple scale option for 3000 psia (210 bar) range () NVLAP accredited calibration report Intecal software for Windows All Series 7250s are fully programmable and can store up to 20 separate programs with up to 1000 steps.

Series 7250 Specifications Performance Precision xi From 5% to 100% full scale (FS): Below 5% FS: 0.005% of reading 0.005% of 5% FS Pressure Ranges Standard Pressure Ranges xi 20, 50, 100, 300, 500, 1000 and 2500 psig (1.6, 4, 6, 25, 40, 60 and 160 barg) i 5, 10, 20, 50, 100, 300, 500 and 1000 psig (400 mbar, 1.0, 1.6, 4, 6, 25, 40 and 60 barg) 15, 20, 30 and 50 psia (1.0, 1.6, 2.5 and 4 bara) 5, 10, 20, 30, 50, 100, 150, 300, 500, 1000, 1500, 2000 and 2500 psig (400 mbar, 600 mbar, 1.0, 1.6, 2.5, 4, 6, 10, 16, 25, 40, 60, 100 and 160 barg) 15, 20, 30 and 50 psia (1.0, 1.6, 2.5, 4 and 210 bara) Optional Pressure Ranges xi Any full scale range from 20 to 2500 psig (1.6 to 172 barg) i Any full scale range from 5 to 1000 psig (400 mbar to 69 barg) Any full scale range from 5 to 2500 psig (400 mbar to 172 barg) to 3000 psia (210 bara) Optional Modes Absolute using barometric reference sensor for ranges from 15 to 2500 psig (1.0 to 172 barg) Vacuum (negative gauge) for ranges 15 to 2500 psig (1.0 to 172 barg) Vacuum reference (requires external vacuum pump) i From 25% to 100% full scale (FS): 0.005% of reading Below 25% FS: 0.005% of 25% FS Ranges to 2500 psi (172 bar): 0.003% of FS 3000 psi (210 bar): 0.01% of FS With triple scale: 0.01% of 1000, 2000 and 3000 psia (70, 140 and 210 bara) Stability Ranges 2500 psi (172 bar): 0.0075% of reading per year 3000 psia (210 bar) range: 0.01% of FS Control Stability Active mode: 0.001% FS (10 ppm) Passive mode: no additional uncertainty Display Resolution User selectable to 1:1,000,000 Control Response 20 seconds or less with zero overshoot into a 15 in 3 volume (245 cc) for ranges to 2500 psi (172 bar) 60 seconds for 3000 psia (210 bar) range Negative Gauge Precision (optional) xi 0.005% of 5% FS or 0.00075 psi (0.05 mbar), whichever is greater i 0.005% of 25% FS or 0.0075 psi (0.5 mbar), whichever is greater 0.003% of FS Barometric Reference (optional) 0.002 psi (0.1379 mbar) maximum error per year Vacuum Reference (optional) 0.0002 psi (0.014 mbar) maximum error per year (Only available on 7250i and 7250)

Series 7250 Specifications Calibration A calibration report with traceability to National Institute of Standards and Technology (NIST) is provided. GE calibrates all Series 7250s with the Model 2465 (0.0010% of reading) to 1000 psi (69 bar) and the Model 2470 (0.0011% of reading) gas piston gauge above 1000 psi (69.4 bar). A NVLAP-accredited calibration is available. Total Uncertainty The maximum deviation from the true value of pressure including precision, stability, temperature effects and the calibration standard is: xi _ 5% to 100% FS 90 day 0.006% reading 1 year 0.009% reading i _ 25% to 100% FS 90 day 0.006% reading 1 year 0.009% reading Ranges to 2500 psi (172 bar) 90 day 0.003% FS + 0.002% reading 1 year 0.003% FS + 0.0075% reading 3000 psi (210 bar) 90 day 0.01% of each range 1 year 0.014% of each range Control Parameters Volume 5 to 60 cubic inches (82 to 980 cc) Low Control (absolute) 0.15 psia (10 mbar) Communications RS232 and IEEE-488, SCPI syntax. Series 7215, Model 7010, Series 6000 and DPI 510 emulation are standard. LabVIEW driver _ download from www.gesensing.com MetCAL driver _ optional Firmware updates are performed via RS232 interface Languages The Series 7250 is capable of displaying menus and functions in: English, French, Chinese, German, Japanese, Spanish and Italian Options Barometric reference (absolute and vacuum) Vacuum only (negative gauge) Vacuum reference (requires external vacuum pump) Intecal software NVLAP-accredited calibration Triple scale for 3000 psia (210 bar) 7250 Rack mount kit Liquid trap assembly General Display TFT, VGA, active matrix, 6.4 in (162.5 mm) 640 x 480 resolution, 65,000 colors Temperature Operating: Storage: 64 F to 96 F (18 C to 36 C) -4 F to 158 F (-20 C to 70 C) Humidity 5% to 95% relative humidity, non-condensing Dimensions (h x w x d) All versions: 7 in x 16.5 in x 19 in (178 mm x 419 mm x 483 mm) Weight /7250i: xi: 17 lb (7.7 kg) 20 lb (9 kg)

Series 7250 Specifications Other Products and Services In addition to a wide range of digital pressure controllers and indicators from 1 to 40,000 psi (25 mbar to 2750 bar), GE manufactures primary standard piston gauges from 0.2 to 72,5000 psi (14 mbar to 5000 bar). Test Port and Supply Connection 1/4 in NPT female Warm Up Time 2 to 3 hours; may be left on indefinitely Pressure Medium Nitrogen or clean dry air Precision is defined as the combined effects of linearity, repeatability and hysteresis throughout the operating temperature range. Expression of uncertainty conforms with the recommendations of the ISO Guide to the Expression of Uncertainty in Measurement.

g 2005 GE. All rights reserved. 920-143A All specifications are subject to change for product improvement without notice. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies. GE is a registered trademark of General Electric Co. www.gesensing.com