CCPaSEC. Procedure for the Quality Control Yearly Equipment Check. July 2018 CCPaSEC Quality Control Team

Similar documents
Measuring Flow with. Centre County PaSEC. QC Team July P a g e

QUICK START GUIDE TO DRINKING WATER MONITORING

Euclid Creek Volunteer Watershed Monitoring Program Directions for Performing Physical/Chemical Tests

v /07/17 Page 1 of 18

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

PREPARED BY: Marshall K. Cheung, Ph.D., Laboratory Director. REVISED BY: Marshall K. Cheung, Ph.D., Laboratory Director

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

YSI Model 550a Dissolved Oxygen (DO) Meter

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Optical Dissolved Oxygen Meter

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Optical Dissolved Oxygen Meter

Heavy Duty Dissolved Oxygen Meter

Calibration Procedure

Low-Flow Purging Procedure with MicroPurge basics Equipment. MP10 Bladder Pump Controller MP15 Control. MP30 Drawdown Meter.

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Annual Inspection and Calibration Checklist VAF Oilcon MK-6(M) MEPC.108(49) Page 1 of 5

Bante904 Benchtop Conductivity/Dissolved Oxygen Meter Instruction Manual

Calibration Procedure

UsER manual for Watersens ph -REDOX

Calibration Procedure

Oxygen Sensor with Electrode

FIELD PROCEDURE: WATER CHEMISTRY

Blue Thumb Test Procedures

Three Lakes Association Hydrolab for Dummies. Table of Contents:

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Dissolved Oxygen/Temp Waterproof Tester

Experiment C-6 Gas Solubility

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control

WAV Level 2 Stream Monitoring Program

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

MEASUREMENT OF TOTAL PACKAGE OXYGEN (TPO) AT BOTTLING

ADVANCED PORTABLE LAUNCH MONITOR. ES14 User Manual

COOPERATIVE LAKES MONITORING PROGRAM TRAINING FOR. Dissolved Oxygen and Temperature

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

YSI-85_maint-cal Lab Version Page 1 of 13

STANDARD OPERATING PROCEDURE MAINTENANCE/CALIBRATION CHECKS OF DISSOLVED OXYGEN PROBES

DeFelsko PosiTector Dew Point Meter DPM & DPMD Probes

Met One E-BAM Particulate Monitor

DO_maint-cal Page 1 of 13

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit:

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Introduction to Chlorosense

PosiTector UTG, UTG-STD & UTG-ME

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

BAUSCH & LOMB ENGINEERING ROCHESTER. Title: Membrane Press Procedure Page 1 of 14

YSI ProDSS calibration and routine maintenance rev 5/29/2018

NEULOG OXYGEN LOGGER SENSOR GUIDE

Do Not Print This Page.

WELCOME TO THE REVOLUTION

PRODUCT SHEET. Order probe only as RXPROBE02

DT 630 ALTIMETER, BAROMETER AND COMPASS WATCH OPERATING INSTRUSTIONS

Altimeter and Compass Watch Instruction Manual

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Hygea Ultrasonic Validation Kit

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

OMNISPORT 2000 SWIMMING QUICK REFERENCE

D-OptoLogger Dissolved Oxygen Logger Operation Manual

AquaChek Select 7 in 1 Total Hardness, Total Chlorine, Total Bromine, Free Chlorine, ph, Total Alkalinity, Cyanuric Acid

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

Overview. Front Panel: Keypad and Display

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Troxler Electronic Laboratories, Inc.

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

D-Opto. Dissolved Oxygen Sensor. Operation Manual (Software Version 1.0)

SS32L Dissolved O 2 Probe

Fish Farming Water Quality Test Kit

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

User Manual GRI- 1500Li

P-TRAK Ultrafine Particle Counter Model 8525

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

2005 CERTIFICATE OF ACCEPTANCE (Part 1 of 3) MECH-1-A

Pool and Spa. Water Testing

4.0 Audit Procedure for the MetOne SASS Speciation Sampler

LogIT Dissolved Oxygen Probe

222 Schwinn Recumbent Exercise Bike Parts List Full Size Hardware Chart Product Illustration Assembly Instructions

OXYGEN MINERALE ABSOLUTE RANGE OWNER S MANUAL OXY MINERALE 600/1000 SERIES

WIND CLIPPER KTS ILLUM SCALE INC DEC CLIPPER WIND SYSTEM

Presented by ARAMSCO

MAGNA-CHEM AUTOMATIC CHLORINATOR - MODEL: CSA-2P MAGNA ORP Chlorine Pump

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

SPUTTER STATION STANDARD OPERATING PROCEDURE

AquaChek Select 7 in 1 Total Hardness, Total Chlorine, Total Bromine, Free Chlorine, ph, Total Alkalinity, Cyanuric Acid

TABLE OF CONTENTS INTRODUCTION 3 SAFETY PRECAUTIONS 3 PACKAGE CONTENTS 4 DEVICE OVERVIEW 4 BUTTON OPERATION SUMMARY 5 BASIC OPERATION 6

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

ACV-10 Automatic Control Valve

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

OWNER S MANUAL NMB SERIES

Introduction. The Shearwater Petrel is an advanced technical diving computer for open and closed circuit divers.

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

01 TM SuperModified Testing with RS485 <> USB

Aerobic Respiration. Evaluation copy

RAM Operation Manual

Transcription:

CCPaSEC Procedure for the Quality Control Yearly Equipment Check July 2018 CCPaSEC Quality Control Team 1

ph and Conductivity Equipment Check. Battery check: replace the batteries. Inspect the conductivity probes and clean it if necessary. When complete, check the appropriate Physical Check box on the table. Procedure Calibrate 2 or 3 meters 24 hrs. before we do an equipment check (Note: We allow users to celibate their meters up to 24 hours before use and they may drift a little so it is best to calibrate the meters a day ahead of preforming the equipment check to determine the percent recovery PR). We will not check all the meters during the yearly equipment check nor need to calibrated them, because we would be using the same standards used to calibrate them as to check them. The duplicate field tests we do for the relative percent difference (RPD) is a good indicator of the meters performance since we compare the field teams with our newly calibrated Quality team meter. 2

Kit Physical check Pass/Fail 4.01 ph 7.0 ph 10.0 ph conductivity ph & conductivity datasheet Check only 2 or 3 meters, 24 hours after calibration for percent recovery 3

Colorimeter Equipment Check.. The colorimeter check can be done using one standard for each test. * The Nitrate results should be factored per the HACH Differences for Cadmium Reduction Method of Nitrate Analysis procedure. Colorimeter Physical condition: Battery check: replace the batteries. Inspect the sample cell cavity and clean it if necessary. Verify the colorimeter kit sample cells appear clean. Verify the kit is complete with unexpired reagent chemicals, wipes, scissors, sample cell holder etc. Replenish as necessary. When complete, check the appropriate Physical Check box on the table. 4

Colorimeter Comparison Function and Range Check. Purpose: We will not attempt to calibrate the colorimeters. Our purpose is to determine the differences between them to establish an average percent recovery (PR) for them. Record the value of the standard on the datasheet. 1. Turn on the Colorimeter and program each colorimeter for the Nitrate test. 2. Assure the window shows NO3-N (if not press the conc button) 3. Wipe the Blank sample cell, insert it in the colorimeter. Orientate the diamond toward the keypad and press Zero for each colorimeter. 4. Wipe the standard sample cell, insert it in the colorimeter. Read the Nitrate standard value for each colorimeter and record each value. 5. Reprogram the colorimeter for Sulfate and Phosphate. Repeat steps 5 & 6 for each reagent and each colorimeter. A correction factor is not needed for them. 5

Formula for Correction of Differences for Cadmium Reduction Method of Nitrate Analysis: Because we measured the Nitrate results using a sample prepared using distilled water and the NitraVer 5 Nitrate Powder pillow (the cadmium reduction method) the value you record for each team colorimeter will not be the same as the number shown on NO3-N on the bottle of the high nitrogen standard (4.43 mg/l). The correction factor is the ratio of the colorimeter reading of your prepared sample made with distilled water to the value listed on the bottle of the standard (4.43 mg/l). Correction factor = Colorimeter reading / Standard (4.43) Example: 9.5 4.43 = 2.14 Do not apply the correction factor to the colorimeter readings you log on the datasheet. Our Percent Recovery spread sheet will automatically apply the correction factor to each of the recorded readings and determine the mean for all colorimeters. A correction factor for Sulfate and Phosphate is determined in the same way. 6

Test NITRATE SULFATE PHOS- PHATE Colorimeter Kit number Physical Check Pass/Fail Program 51 91 79 Standard Value Colorimeter datasheet 7

Dissolved Oxygen Meters Equipment Check Dissolved Oxygen Meter Physical condition: Verify that they appear clean and working. The kit has calibration instructions. Battery check: replace the batteries. Inspect the probe cell cavity and clean it or replace if necessary. When compete, check the appropriate Physical Check box on the table. Dissolved Oxygen Meter Measurement: Purpose: Our purpose is to determine the differences between them and establish an average and a tolerance range we can expect from them. Test: 1. Calibrate the meter for the altitude of the test. 2. Prepare a container with tap water. Insert the meter probe and while moving it back and forth in the water note the measurement. 3. All meters should show a similar result. If not, repeat the calibration and retest. Alternate with 0 oxygen standard 1. Prepare the 0 zero dissolved oxygen standard. 2. Measure and record the reading of each meter. 8

Kit Serial Number Physical check Pass/Fail Tap water dissolved oxygen mg/l dissolved oxygen with zero std. Dissolved Oxygen (YSI Meter) worksheet 9

FloWatch Equipment Check Purpose: To determine physical condition. We do not have a means of calibrating for velocity. FloWatch Meter: Physical condition: 1. Verify that they have not sustained physical damage, appear clean and the kit is complete with instructions. 2. Verify that the cable is attached to the hand unit and connector is properly seated. (we are advising the volunteers to not disassemble that connector). 3. Replace the batteries. 4. Verify the setup for AV (average), Time mode, Units as cm/sec and degrees centigrade - see illustration below. Reset it if needed. 5. Inspect the turbine. It should spin easily with just a breath of air. All four of the blades must appear uniform in shape and pitch (the angle they are set at). 6. The pitch determines the velocity. One revolution is equal to one centimeter per second. Please remember to turn the unit OFF after use. When complete, check the appropriate Physical Check box on the table. Replace the unit in the packaging do not disassemble the cable connector from the unit. 10

Kit Physical Pass/Fail Measure cm/sec @ frequency 60 cycles/sec. FloWatch flowmeter worksheet 11

Calibrating the meter You need to provide an alternating magnetic field to calibrate the FloWatch. This can be done by holding probe close to a small motor running at 60 Hz single phase. Turn the FloWatch on. Hold the probe near the motor to pickup a reading. It should read 205 cm/sec. If not, it will need recalibration. Probe 1. Turn meter on.. 2. Press both bottom buttons simultaneously. A count down will start on the screen. 3. Before the countdown ends release the bottom left button (do not release the bottom right) and press the top right button. 4. Release both buttons and use the arrows to set the value on the screen to 123. 5. The meter will return to normal mod several seconds after no buttons have been pushed. Step 1. To calibrate the meter, press these two buttons simultaneously. Step 3. While holding the lower right button - press the top right arrow button to enter the setup 4. The screen should show 123 If not. Adjust the number with the up/down arrows (buttons). Step 2. A countdown begins to clear the memory. You need to release the left button before it stops. Do not release the bottom right button. 12