Micro Motion Pressure Drop Testing

Similar documents
The Discussion of this exercise covers the following points:

MEMORANDUM. Investigation of Variability of Bourdon Gauge Sets in the Chemical Engineering Transport Laboratory

Rupture Disc Sizing for Micro Motion Coriolis Sensors

Experiment 8: Minor Losses

COMPUTATIONAL FLOW MODEL OF WESTFALL'S LEADING TAB FLOW CONDITIONER AGM-09-R-08 Rev. B. By Kimbal A. Hall, PE

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners

A centrifugal pump consists of an impeller attached to and rotating with the shaft and a casing that encloses the impeller.

TWO PHASE FLOW METER UTILIZING A SLOTTED PLATE. Acadiana Flow Measurement Society

Gerald D. Anderson. Education Technical Specialist

485 Annubar Primary Flow Element Installation Effects

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Equation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg

Applied Fluid Mechanics

This portion of the piping tutorial covers control valve sizing, control valves, and the use of nodes.

Device Description. Operating Information. CP Q (eq. 1) GT. Technical Bulletin TB-0607-CFP Hawkeye Industries Critical Flow Prover

Modeling a Pressure Safety Valve

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

TUTORIAL. NPSHA for those who hate that stuffy word. by Jacques Chaurette p. eng. copyright 2006

Case Study Results on Natural Gas Custody Transfer Measurement with Coriolis Meters in Saudi Arabia

Performance of an 8-path gas ultrasonic meter with and without flow conditioning. Dr Gregor Brown Caldon Ultrasonics Cameron

Drilling Efficiency Utilizing Coriolis Flow Technology

A REVIEW OF THE 2000 REVISIONS TO ANSI 2530/API MPMS 14.3/AGA REPORT NO. 3 - PART2 Paul J. LaNasa CPL & Associates

ONSITE PROVING OF GAS TURBINE METERS Daniel J. Rudroff Invensys Metering Systems

Sizing Pulsation Dampeners Is Critical to Effectiveness

Fluid Flow. Link. Flow» P 1 P 2 Figure 1. Flow Model

CHAPTER 6: PERMEABILITY MEASUREMENT

DEVELOPMENT OF A ROBUST PUSH-IN PRESSUREMETER

HOW TO NOT MEASURE GAS - ORIFICE. Dee Hummel. Targa Resources

the asset can greatly increase the success rate of Coriolis technology in this application. 3 Coriolis measurement in multiphase flow

Permeability. Darcy's Law

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cold-acclimatised pipelines

GOOD PRACTICE GUIDE AN INTRODUCTION TO FLOW METER INSTALLATION EFFECTS

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

SIZING AND CAPACITIES OF GAS PIPING Reserved

The Use of a Process Simulator to Model Aeration Control Valve Position and System Pressure

CENTER PIVOT EVALUATION AND DESIGN

Diagnostics for Liquid Meters Do they only tell us what we already know? Terry Cousins - CEESI

SIZING AND CAPACITIES OF GAS PIPING

INSTRUMENTATION EQUIPMENT

SIZING AND CAPACITIES OF GAS PIPING

****** * EX * ****** DWN W.W.POWELL CALCULATION OF FLOW LOSSES IN INLET CHK D.PAPA AND DISCHARGE HEADERS ASSOCIATED WITH

Steam generator tube rupture analysis using dynamic simulation

WaterSense Specification for Spray Sprinkler Bodies. Version 1.0

Pressure Measurement

ISS0065 Control Instrumentation Lecture 12

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

The University of Hong Kong Department of Physics Experimental Physics Laboratory

ULTRASONIC METER FLOW CALIBRATIONS CONSIDERATIONS AND BENIFITS. Joel Clancy

Multiphase Flow in Coriolis Mass Flow Meters Identifying Multiphase Conditions and Remediating Errors

Pressure Control. where: p is the pressure F is the normal component of the force A is the area

International Journal of Research in Advent Technology Available Online at:

Applying Hooke s Law to Multiple Bungee Cords. Introduction

T EK-COR 1100A. Coriolis Mass Flowmeter. FLOW. Technology Solutions

Peria G Mohamed, Abbas Al-Shawwaf, Mahdi Al-Qazweeni & Nasser Bu-Arki Kuwait Oil Company

ULTRASONIC FLOW METER CALIBRATIONS CONSIDERATIONS AND BENIFITS American School of Gas Measurement Technology, September 2007.

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Third measurement MEASUREMENT OF PRESSURE

Oil and Gas Custody Transfer

OIL AND GAS INDUSTRY

PMI Pulse Decay Permeameter for Shale Rock Characterization Yang Yu, Scientist Porous Materials Inc., 20 Dutch Mill Road, Ithaca NY 14850

PIG MOTION AND DYNAMICS IN COMPLEX GAS NETWORKS. Dr Aidan O Donoghue, Pipeline Research Limited, Glasgow

No E. General Information. Averaging Pitot Tube Flow Sensor HITROL CO., LTD.

Experiment (13): Flow channel

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

The capacity of the cargo tank venting system (46 CFR );

Installation Effects on Ultrasonic Flow Meters for Liquids. Jan Drenthen - Krohne

Lab Problems. Lab Problems for Chapter Fluid Characterization by Use of a Stormer Viscometer L-1

SFC. SKYLINE FLOW CONTROLS INC. The Leader of Accurate and Reliable Flow Measurement DESCRIPTION & APPLICATIONS: ADVANTAGES:

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Wet Gas Flowmetering Guideline

CHAPTER 2 EXPERIMENTAL SETUP AND PROCEDURE

Gas Measurement Fundamentals Certification. Curriculum

ME 333 Fluid Mechanics. Lab Session VISCOUS LOSSES IN PIPES

Calibrating and Monitoring MUSMs in Natural Gas Service

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler

International Journal of Technical Research and Applications e-issn: , Volume 4, Issue 3 (May-June, 2016), PP.

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

CONTROL VALVE WHAT YOU NEED TO LEARN?

UNUSUAL ASPECTS OF PUMP SYSTEMS. Jacques Chaurette p. eng. ww.lightmypump.com July 2003

Restriction Orifice. Single or Multi Stage Orifice to. Reduce Pressure or. Limit the Flow Rate

Optimizing Compressed Air Storage for Energy Efficiency

Smart Water Application Technologies (SWAT)

Coriolis Mass Flow Meter

Process Dynamics, Operations, and Control Lecture Notes - 20

H16 Losses in Piping Systems

Oil And Gas Office Houston Fax Test Separator / Off-Shore Metering

The Discussion of this exercise covers the following points:

Positive displacement proportioning of two-component epoxy intumescent coatings for passive fire protection

SIZING AND CAPACITIES OF GAS PIPING

Motion Graphing Packet

Gas Flow Calibration Basics

Application Worksheet

Performance Characteristics of Airlift Pumps with Vortex Induced by Tangential Fluid Injection

RCD - Ultra-Rugged Differential Pressure Flowmeter

THE INNER WORKINGS OF A SIPHON Jacques Chaurette p. eng. January 2003

Gas Physics Pressure and Flow Topics Covered:

1 PIPESYS Application

INTERACTION BETWEEN WIND-DRIVEN AND BUOYANCY-DRIVEN NATURAL VENTILATION Bo Wang, Foster and Partners, London, UK

ENSURING AN ACCURATE RESULT IN AN ANALYTICAL INSTRUMENTATION SYSTEM PART 1: UNDERSTANDING AND MEASURING TIME DELAY

Transcription:

12/2018 Micro Motion Pressure Drop Testing www.emerson.com/micromotion

Introduction Micro Motion has traditionally taken a very conservative approach to pressure drop, with single pressure measurements up- and downstream of the Unit Under Test (UUT). There are two issues with this: 1. Any asymmetric flow profile on the inlet side or outlet side is unquantified and can create large errors. 2. Pipe pressure loss is included in the meter loss. Depending on where the pressure taps are, this error can be large or small. To minimize errors associated with no. 1 above, it s common to place the taps approximately 10D along the length of the pipe away from the last disturbance. There is still no quantification of whether asymmetric flow is occurring at either tap, and it almost by definition includes a significant amount of pipe-length pressure loss in the measurement. This paper describes a methodology to: Quantify the extent of any asymmetry Measure and subtract the pipe pressure loss for a precise pressure loss between the flanges. Baseline Equation Micro Motion pressure drop on all products is predicted by the Darcy-Weisbach equation 1 : [1] Weight density is related to density by: Substituting equation [3] into equation [2] results in: The only difficulty in using equation [4] is that the term is not constant. At high Reynolds number, the term is asymptotic to a constant value, but at low Reynolds numbers, increases. Micro Motion sizing tools (Product Advisor and ToolKit) break into several components to make the curve-fit easier. This paper does not describe the curve-fit methodology. A typical relationship between number is shown in Figure 1. and Reynolds CMFHC2M Loss Factor vs Reynolds Number [3] [4] Where: Reynolds Number Note f is the Darcy Friction Factor. Pressure drop is equal to the head loss scaled by the fluid weight density: [2] Figure 1 - (pressure loss factor) for a CMFHC2M vs. Reynolds Number Note that the loss factor is asymptotic to approximately a value of 1.25. 1 Fluid Mechanics, Ninth Edition, page 284. Streeter, Wylie & Bedford, 1998. Page 2

Reynolds number is defined as: Where: [5] Multiple pressure measurements along the length of the pipe, shown in Figures 2 & 3. This helps to visualize the integrity of the data. Refer to the following section. The pressure tap into the pipe should be small (1/8 diameter is typical) and should be de-burred inside the pipe. The curve vs Re must be monotonic; that is, the slope is trending toward the asymptote (zero slope) but it is always negative and cannot have an inflection and switch to be positive. Baseline Equation Significant errors can be made when measuring pressure drop due to swirl and asymmetric flow profile. Although this statement is always true, it s especially important to mitigate the effects of an asymmetric flow profile when the Reynolds number is high. This is true because disturbances that are introduced to the flow by elbows, bends, tees, etc. travel for many pipe diameters downstream. Physically, Reynolds number describes the ratio of the inertial forces to the drag (or viscous) forces. A high Reynolds number therefore depicts a flow that is dominated by inertial forces. Therefore, a disturbance introduced to the flow carries for a long distance down the pipe because there is no drag to slow down the disturbance. Practical measures are common to ensure a good pressure drop measurement. They include: Long straight runs. In gas measurement it is common to require at least 10D. For precision orifice flow measurements 100D is sometimes required. Flow conditioning (plates, tube bundles, etc). Multiple pressure measurements radially around the diameter of the pipe, as shown in Figure 2. This helps ensure a good average pressure measurement; for instance, if the velocity is high on top of the pipe (due to swirl) and low on the bottom of the pipe, a pseudo-average measurement results when a radial manifold is used. Figure 2a -- A pressure tap is made up of 4 radial pipe penetrations, 90 apart Figure 2b Multiple taps make up the complete measurement Page 3

White Paper 12/2018 4 upstream taps Pressure Drop Testing 4 downstream taps Figure 3 4 taps upstream, 4 taps downstream Precision Pressure Drop Measurement Figure 4 shows diagrammatically how pressure drop data should be distributed along the pipe. Note that the slope of each line represents the pressure loss across the four measurement locations upstream and downstream of the meter (1 through 4 and 5 through 8, respectively). If the slope of these lines are linear and parallel, there is no asymmetry in the flow profile. Note that pressure can be absolute or gauge, as long as all measurements use consistent units. There need to be at least three measurement points upstream and three downstream to assess uncertainty (two points would always make a perfect line, regardless of flow profile). Micro Motion chose four measurement points to further increase confidence in the data. Measurements are made at locations 1 through 8. If all the data falls predictably on the downwardsloping lines, the data is good. If points are scattered, data integrity is poor. The most likely points to have problems are 1 and 5. The pressure at point 1 is a function of the upstream conditions; if there is significant flow disturbance upstream of the test section, point 1 may be suspect. In a similar way, if point 5 is too close to the meter under test, its pressure measurement will not be accurate. If all points have good data integrity, the inlet pressure of the meter is the forward extrapolation of data points 1 through 4; the outlet pressure is the reverse extrapolation of the data points 5 through 8. Data Collection and Analysis Data collection was made using a portable DAQ system reading at once per second for five minutes (approximately 300 data points). A sample of the output is shown on next page: Figure 4 Pressure drop measurement set-up Page 4

Figure 5 Pressure Drop Data Notice in this example that the pressure at the last tap (no. 8 at 69.75 inches) measures only 1 psig. Care must be taken to make sure pressure is high enough to eliminate flashing or air being pulled from solution in the water. The pressures (labelled P1, P2 etc) are plotted vs. Location in Figure 6 below. Figure 6 Pressure Drop Data from Figure 5, Plotted and Regressed The meter is located between 30 and 39.5 inches, resulting in a pressure drop of 29.2 psi. Note that the current sizing tools predict a pressure drop of 34.2psi, or 17% high. The over-estimate is likely a result of errors based on traditional pressure loss measurement as described at the beginning of this paper. Page 5

As discussed previously, the data in Figure 1 must be monotonic and decreasing vs. increasing Reynolds Number. Deviations indicate poor pressure measurements resulting from poorly calibrated transmitters, flow profile problems, and/or flashing due to low pressure. Note that flashing may occur when air is pulled from the solution and is not necessarily boiling of the water. A very important reason for normalizing the pressure drop as shown in Figure 7 is that it allows for any fluid to be used to establish the vs Re curve. Other Fluids Any fluid can be used to characterise pressure drop. In fact, it is difficult to predict pressure drop characteristics on low Reynolds numbers or high Reynolds numbers with water. Micro Motion uses mineral oil with viscosities between 30 and 90 cp and air at approximately 0.01 cp when additional characterization is required. These fluids are difficult to use because oil is difficult to handle for environmental reasons and air does not generate much pressure loss due to its low density. Note that if other fluids are used, viscosity is a critical parameter to understand. A Brookfield rheometer is used to measure the viscosity of liquids. Care should be taken to ensure any fluids other than water are Newtonian in nature. Example Use the data from Figures 5 & 6: Tube velocity is calculated by (remembering that there are two tubes): Reynolds number is then: Rearrange equation [4] to calculate fl/d: Summary Most Coriolis manufacturers have traditionally taken pressure measurements without attention to flow profile or pipeline pressure loss. If an upstream disturbance causes the flow to swirl, a pressure measurement at a single point may be inaccurate. Micro Motion recognizes that accurate pressure loss calculations are critical to Coriolis users. This test program considers the data integrity of both a multipoint flow profile and assymetry mitigation to provide a precise and thorough evaluation of pressure loss across a meter. Emerson Automation Solutions 7070 Winchester Circle Boulder, CO USA 80301 T +1 303 527 6277 F +1 303 530 8459 www.emerson.com 2018 Micro Motion, Inc. All rights reserved.