Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer

Similar documents
Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Practical approach and problems in in-situ RGA calibration

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Gas Sampling in the DST

Major Design Features

Mass Spec will not Autotune

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Quadruple mass spectrometers (transducer type)

Gun Vacuum Performance Measurements February 2003

Columbus Instruments

The Principles of Vacuum Technology

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

Residual Gas Analysis Systems for Industry

The latest automotive systems require innovative leak test methods and fixturing.

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Constant Pressure Inlet (CCN) Operator Manual

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

Datasheet: K-30 ASCII Sensor

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts

Leak Checking Large Vacuum Chambers

8.1 Calibration of gas sensors

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

USER GUIDE. Fog Generator. ATMOSPHERE APS Series Haze Generator. MAX APS Series. No: , Revision A/d

HumiSys HF High Flow RH Generator

Zebron Gas Management

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

APPLICATION NOTE. GC Integrated Permeation Device

MV5: Automated concentration system

Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump

DPC-30 DPC-100. Reference Manual

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Degassing ERC KK Saitama, Japan

Total Pressure Measurement Using Residual Gas Analyzers

FTC130 Transmitter. Operating Manual

The New PHD 4 Portable Helium Detector

Thermo Scientific Model 146i Principle of Operation

Atlas 6TM. Indoor Air Quality Monitor Instruction Manual. Measures CO 2, Temperature and Humidity DISTRIBUTED BY

Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.

Choosing a Quadrupole Gas Analyzer Application Note #9

ACCELERATOR VACUUM PROBLEMS: IN-LINE OIL TRAPPING BETWEgN THE STANFORD LINEAR ACCELERATOR AND THE BEAM SWITCHYARD

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

Journal of Applied Fluid Transients, Vol 1-1, April 2014 (3-1)

Vapor Intrusion. Sampling Solutions VAPOR INTRUSION. Sampling Solutions

A New GC/MS System Designed for Helium Carrier Gas Conservation

GMA FUNCTIONAL TEST SETTING & TESTING OF SPIN UP PARAMETERS. P0586 REV -- 20, June 2000 GPB SCIENCE MISSION PROCEDURE

FPG8601 Force Balanced Piston Gauge

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007

APPLICATION OF A CHANNEL PLATE IN FIELD-ION MICROSCOPY

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Application Note AN-107

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

Technical Report. Comparison of Ventilation Standard Calibration Methods

DISCLAIMER. Government. Neither the United States Government nor any agency thereof, nor any of their

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

Chapter 6 Leak detection

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

Calibration of a GC955

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Results of mathematical modelling the kinetics of gaseous exchange through small channels in micro dischargers

Measuring range Δp (span = 100%) Pa

Application Note. ASTRON Remote Plasma Source Ignition Best Practices PROBLEM

Gas and Moisture Transport through Non-Hermetic Glass-to-Metal Bridge- Wire Initiator Seals

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Cover Page for Lab Report Group Portion. Head Losses in Pipes

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

OOC RGA. Ion Pump 400 l/s. 8" ports with baked viton o-ring seals. Welsh rotary roughing pump 650 l/s manual isolation valve.

* Analysis of TCG ( Total Combustible Gas ) and 6 kinds of Gases * High Sensitivity for C 2

MACH ONE MASS FLOW CONTROLLER. MACH ONE SERIES flow control. MASS FLOW CONTROLLERS at the speed of sound.

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Detecting Hydrogen in Helium Streams. Samuel Goodman. Pittsford Mendon High School. Rochester, New York. Advisor: Dr.

DSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications

Paper Session II-C - Verification of International Space Station Component Leak Rates by Helium Accumulation Method

Operation of the Perkin Elmer TGA-GC/MS

INTRODUCTION. The Quantum Technology system has the following advantages:, as it does not need plastic gas-bags which are volume-consuming,

Digital Vacuum Regulator

A Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide

Procedure of Xenon Transfer

Transcription:

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand Space Systems International Presented at the 7 th Workshop on Harsh Environment Mass Spectrometry, Santa Barbara, CA September 22, 2009

The Major Constituent Analyzer Major Constituent Analyzer (MCA) is a magnetic-sector mass spectrometer based system for monitoring ISS air. Six simultaneous detection channels; for H 2, CH 4, H 2 O, N 2, O 2, and CO 2 which are the major on-board air constituents of ISS. Is self-calibrating. Includes active built-in-test (BIT) capability. Is comprised of seven orbitreplaceable units (ORUs) for easy maintenance. Page 2

International Space Station Page 3

MCA Functionality Of the six major constituents that MCA is designed to measure, five (H 2, CH 4, N 2, O 2, and CO 2 ) are currently being measured as planned. Measurement of the sixth constituent, H 2 O, was descoped during MCA development. At that time, the issue of transporting humid air through the ISS plumbing was not completely understood and the effect of H 2 O surface adsorption on the sampling lines had not been characterized. Recently the MCA program has been authorized to revisit the issue through NASA change request CR0773A. Page 4

The Calibration Sequence Zero Cal (Offset) Beam Centering Full Cal w/ VGA (Gain) Page 5

MCA Normal Operation The plumbing schematic for MCA is shown on the right. Under normal operation, the incoming air flows through an inlet valve of the MCA manifold and into the analyzer. Page 6

MCA Calibration Start (Zero Cal) The plumbing schematic for MCA is shown on the right. Under normal operation, the incoming air flows through an inlet valve of the MCA manifold and into the analyzer. At the beginning of the zero and full calibration routines, the inlet valve at the analyzer entrance is closed. The analyzer then drops to a baseline level to establish background signal amplitudes. Page 7

The Problem with Measuring H 2 O Measuring p(h 2 O) has been problematic because, unlike the other constituent gases, H 2 O adsorbs onto the interior surfaces of the transport tubing. Measured H 2 O levels do not reflect the actual gas-phase environment until equilibrium is achieved. As shown in the figure, reaching equilibrium can require considerable time (>00 minutes), which in not compatible with MCA operation requirements. The decay characteristics are not strictly exponential. This phenomenon effects MCA capacity to accurately calibrate the H 2 O channel and to predict ISS humidity. Intensity (Volts) 0 9 8 7 6 5 4 3 2 0 O 2 Close SV0 N 2 H 2 O ITU Data Manual Mode Room Air to Vacuum Room Temperature Slow decay Rapid drop in partial pressure for O 2 and N 2 0 20 40 60 80 00 20 Time (Min) Page 8

Zero Calibration Zero Calibration establishes the electrometer background average (EBk). Full Calibration then uses EBk to adjust electrometer correction values (ECVs). Any error in EBk generates an incorrect calibration. As shown in the Figure, EBk is acquired before the H 2 O signal is stable. ITU Data Automatic Mode Page 9

What is responsible for the slow decay? The primary suspect component responsible for the slow decay rate is the analyzer inlet valve, SV0. SV0 has a high internal surface area, low gas conductance, and is unheated. Laboratory experiments using the MCA Integration and Test Unit (ITU) revealed that the shape of the decay curve for H2O can be changed by heating the valve. However, the slow decay is not eliminated. Furthermore, the baseline levels do not converge even after an extended time period. Intensity (V) 9 8 7 6 5 4 3 2 60% RH 2% RH ~00 C SV0 Valve 0 500 000 500 2000 2500 3000 Time (sec) 20 C Black 00 C Gray Page 0

Simulink Modeling of the Decay SV0 Valve PP SV0 valve closed. switch N 2 SV0 leak 2 Ctrl s+ t(n2) O 2 4e-7 s+ t(o2) s+ t(cd4) 0.5 20 s+ t(h2o) PPout Intensity (Volts) H 2 O H2O Pressure 0.5 300 s+ H20 SV 3 H2O Product t2(h2o) s+ t(co2) s+ t(h2) s+ t(ch4) 0 20 40 60 80 00 20 Time (Min) Page

Simulink Modeling of the Decay Analyzer SV0 valve closed. N 2 N2 torr to Amp 0. CO2 torr to CO CO torr to Amp PP O2 torr to Amp CD4 torr to Amp gain 2 Fil Product I O 2 H2O from inlet 0.2 36000 s+ H2O background from Housing & Ion Pump (ts = 0 hr) 0.04 604800 s+ H2O background from Housing & Ion Pump (ts = week) H2O to O 2 H2O to CO2 H2O torr to Amp CO2 torr to Amp H20+CD4 MS Intensity (Volts) H 2 O H2 torr to Amp CH4 torr to Amp 0 20 40 60 80 00 20 Time (Min) Page 2

Improving Zero Calibration The experimental thermal and modeling data confirm that the SV0 valve is primarily responsible for the observed decay process. With MCA validated and on-orbit, heating the SV0 valve is not an option. The modeling suggests that by collecting decay curve data for a short time after one timeconstant, the final baseline can be extrapolated. This will require a slight increase in the zero calibration time. Page 3

Full Calibration Full Calibration includes the Zero Calibration, and occurs every 6 weeks on orbit. Full Calibration uses a verification gas mixture containing CD 4 to determine the EAv. The EAv and the EBk are used to determine the electrometer correction values. The CD 4 signal rides an unknown amount of H 2 O. Page 4

Full Calibration Sample pathway is closed at 200 torr and uncontrolled when the verification gas inlet is opened, raising the pressure to 400 torr. Verification gas is thus forced into the side ports of the manifold, diluting trapped gasphase H 2 O which can diffuse into the verification gas flow path. The decay observed is a function of this process as well as desorption processes including those described in the zero calibration. This is not a problem for normal MCA operation, where the flow throughput greatly dominates any residual diffusion. Page 5

An Alternate Amplitude Method for CD 4 Switch the manifold valve TSV02 off and the Verification gas valve on simultaneously to preserve controlled flow. H 2 O level will decay from a known initial level. CD 4 amplitude can be extrapolated accurately. Page 6

Full Calibration Full Calibration includes the Zero Calibration, and occurs every 6 weeks on orbit. Full Calibration uses a verification gas mixture containing CD 4 to determine the EAv. The EAv and the EBk are used to determine the electrometer correction values. The CD 4 signal rides an unknown amount of H 2 O. Page 7

Taking advantage of what we ve learned Beam Centering Full Cal w/ VGA (Gain) Zero Cal (Offset) We are altering the calibration sequence to perform beam centering first using atmospheric gas to reduce consumption of verification gas. The full cal step will be performed next to take advantage of the logic previously described. The zero cal will follow, allowing additional time to establish an accurate baseline for H 2 O. Page 8

Summary The improved water accuracy logic is currently being programmed into the MCA firmware as part of NASA CR0773A. Testing is nearly complete. The firmware upgrade is scheduled for delivery on ULF3 in November, 2009. It may be possible to extend the extrapolation logic of the zero calibration to the sampling of humid air from long lengths of sample tubing. This is under review. Page 9

Acknowledgements MCA Development Team Hamilton Sundstrand John Granahan - Boeing John Cover - NASA Page 20