Table 11 Provision of terminal units, AVSUs and local alarms

Similar documents
Medical gas pipeline systems

ARCHIVE. Scottish Health Technical Memorandum 2022 (Part 1 of 2) Medical gas pipeline systems

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:

Manifolds & pressure reducing sets

Medical gases Health Technical Memorandum Medical gas pipeline systems Part A: Design, installation, validation and verification

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

Technical Specification

BOC Healthcare Training

BOC Healthcare training

1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)

Anaesthetic infrastructure and supplies

ISO INTERNATIONAL STANDARD. Medical gas pipeline systems Part 1: Pipeline systems for compressed medical gases and vacuum

UTMB RESPIRATORY CARE SERVICES GUIDELINES Respiratory Care Services Medical Gas Cylinder Storage, Use and Transport

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

The right atmosphere for health. A system for medical gas supply

MEDAP-MOBILE GAS SUPPLY

Keeping anaesthetic pollution to minimum in theatre environment

This document is a preview generated by EVS

MANUAL GAS MANIFOLDS

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

Flow meter. bellow vaporizer. APL valve. Scavenging system

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1)

USE A RELIABLE A INNOVATIVE A COST EFFICIENT GAS EQUIPMENTS AND GAS SYSTEMS

ARCHIVE. Scottish Health Technical Memorandum 2022 (Part 2 of 2) Operational management. Medical gas pipeline systems

Figure 3. Miniscav scavenging unit

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

C O R P O R AT E B R O C H U R E

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant


o An experienced and trained assistant is available to help you with induction.

Pure Gases Liquid & High Pressure High Pressure Gas Mixtures Equipment Safety MEDICAL GASES

Recommendations on Checking Anaesthesia Delivery Systems

Optimum oxygen supply for your patients! ATMOS Oxygen Supply

Medical gas pipeline systems

INTERNATIONAL ISO STANDARD Anaesthetic machines for use with humans. Appareils d anesfhesie pour utilisation chez I homme

Part 1: Medical gas pipeline systems. Pipeline systems for compressed medical gases and vacuum

Medical Instruments in the Developing World

MEDICAL GAS CYLINDERS AND MEDICAL PIPELINE SYSTEMS (MGPS) POLICY

AUTOVENT 4000 VENTILATOR

FLOWMETER UNITS FLOWMETER DEVICES, BY-PASS SYSTEMS

Gilbert Bain Hospital. Medical Gas Pipeline Systems (MGPS): Operational Policy Document. Version number: Final

DESCRIPTION AND OPERATION OXYGEN SYSTEM OXYGEN SYSTEM

AUTOMATIC GAS MANIFOLDS

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

The centerpiece of a comprehensive anaesthesia workstation

ekom Compressor for Medical Air Sonata with Ekom

RESOLUTION A.567(14) adopted on 20 November 1985 REGULATION FOR INERT GAS SYSTEMS ON CHEMICAL TANKERS

Medical Gas Policy. Staff/Groups Consulted

Recommended Minimum Facilities for Safe Anaesthetic Practice in Operating Suites

Summary of the MRA LPG Codes of Practice

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY

Perfect Medical Product

Liquid Medical Oxygen Medicinal gas, cryogenic Package Leaflet: Information for the User

O-Two Self-Study Guide. Equinox Relieve

AREA VALVE SERVICE UNITS

Pressure Equipment Directive (PED) 97/23/EC Page 033 of 124

INTERNATIONAL ISO STANDARD Anaesthetic machines for use with humans. Appareils d anesfhesie pour utilisation chez I homme

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

MEDICAL EQUIPMENT CATALOG

Guidelines on Monitoring in Anaesthesia

GASES FIRE & RESCUE SPECIALTY. Gases & Equipment. Medical Gases Liquid & High Pressure Oxygen U.S.P. Nitrogen N.F. Carbon Dioxide U.S.P.

Compressed Medical Oxygen Medicinal gas, compressed

Breathing Air Purifiers. Comply with European Pharmacopoeia and International Breathing Air Standards

Anesthesia Solutions. Prima SP and Prima SP2 Anesthesia Systems. The Future of Anesthesia. Partnership for Life

This test shall be carried out on all vehicles equipped with open type traction batteries.

ON-SITE OXYGEN GENERATION WHY PSA SYSTEMS ARE A GOOD CHOICE

Definition An uninterrupted path between the atmosphere and the alveoli

Section 14: Equipment failure

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

User Manual PTW-DensiX. D / Di/Zi

ISO INTERNATIONAL STANDARD. Flow-metering devices for connection to terminal units of medical gas pipeline systems

MEDAP-FLOWMETERS FOR OXYGEN AND COMPRESSED AIR

Apollo Safety Features

BY GREGGERSEN CATA LOG

QUESTIONNAIRE FOR MEMBERS OF THE PUBLIC

Version: 1.0. Medical Gas Policy. Name of Policy: Effective From: 21/07/2011

SAFETY FEATURES OF PORTABLE CRYOGENIC LIQUID CONTAINERS FOR INDUSTRIAL AND MEDICAL GASES

HOSPITAL STRUCTURAL PRODUCTS 12. MEDICAL GAS SYSTEMS LUMINAIRES AND OTHER EQUIPMENTS WALL CORNER PROTECTION SYSTEMS A. B. C.

American National Standard

Oxygen Dialflow Meter. Instructions for Use

Standard Specification for Anesthetic Gas Scavenging Systems Transfer and Receiving Systems 1

Medical Air. Mark Allen and Paul Edwards. Continuing Education Publication

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

Respiratory Protective Equipment

ISO 5359 INTERNATIONAL STANDARD. Low-pressure hose assemblies for use with medical gases

Review Date 01/01/2016 Director of Estates and Facilities Expiry Date 15/01/2017 Withdrawn Date

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

SERIES-O REGULATOR. Instructions for use

State-of-the-art anesthesia technology

Recommendations - USA

Trifecta Systems. Cryogenic Equipment for High Performance Gas Supply. Trifecta Product Catalog

Oxygen Dialflow Meter. Instructions for Use

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

Class 2 Gases Learning Tool

Emergency Mobile Unit Medicinal gases everywhere, always on time

About us. We supply Industrial gases, calibration gases, Medical gas Fireman protection and equipment

VETROSON. OXY-GEN SYSTEMS (Oxygen Generating Systems) Question & Answer Brochure

Transcription:

Table Provision of terminal units, AVSUs and local alarms Department O N O N O/O MA4 SA7 VAC AGSS He/O AVSU Alarm Accident and Emergency set () set hp/lp (9) Resuscitation room, per trolley space Note: One set either side of the trolley space, if installed in fixed location, eg trunking; or both sets in an articulated supply pendant that can be positioned either side of the bed space. sets* 5 Major treatment/plaster room per trolley space Post-anaesthesia recovery per trolley space Note: One set either side of the trolley space, if installed in fixed location, eg trunking; or both sets in an articulated supply pendant that can be positioned either side of the bed space. Treatment room/cubicle set/8 TUs Operating department set () Anaesthetic rooms (all) Operating room, orthopaedic: For anaesthetist For surgeon Note: Orthopaedic surgery is normally performed in operating rooms provided with ultra-clean systems. Such systems are much more effective in terms of airflow when provided with partial walls. These walls may be effectively used to include terminal units that can be supplied by rigid pipework. Such installations do not suffer from excessive pressure loss when surgical air is required at high flows. Operating room, neurosurgery Surgeon Note: If multi-purpose pendants are used, there may be some loss of performance of surgical tools because of bore restrictions and convolution of the flexible connecting assemblies at the articulated joints. p p 4 set/8 TUs sets* set per suite ()(3) set per suite ()(3) set per suite hp/lp (0) set per suite hp/lp (0) 3 Provision of terminal units, and the location of AVSUs, local alarm indicator panels and LVAs

6 Department O N O N O/O MA4 SA7 VAC AGSS He/O AVSU Alarm Operating room, general surgery etc /surgeon (see (4) regarding CO ) Note: Terminal units installed in separate pendants: p = project team option where some orthopaedic overspill surgery may be performed. Post-anaesthesia recovery, per bed space Note: One set either side of the bed space, if installed in fixed location, eg trunking; or both sets in an articulated supply pendant that can be positioned either side of the bed space. p set per suite ()(3) set per suite hp/lp (0) sets* (5) alarm for both sets of AVSUs () Equipment service room per work space set set hp/lp () Maternity department set () set (9) LDRP room (normal/abnormal) Mother Baby (per cot space) (allow for cots only) Operating suite: Obstetrician Paediatrician (per cot space) (allow for cots only) set per 68 rooms (5) set set hp/lp (0) Post-anaesthesia recovery (per bed space) set set () Equipment service room** per work space set set () Neonatal unit, per cot space Note: One set either side of the bed space, if installed in fixed location, eg trunking; or both sets in an articulated supply pendant that can be positioned either side of the bed space. sets* for both sets of AVSUs () Equipment service room** per work space set set () In-patient accommodation: Single bed room Multi-room, per bed space Nursery, per cot space set for ward unit set () Medical gases HTM 0-0 Medical gas pipeline systems Part A: Design, installation, validation and verification Provision for cots only irrespective of number of cot spaces

7 Department O N O N O/O MA4 SA7 VAC AGSS He/O AVSU Alarm Diagnostics departments set () set (9) Special procedures room set set (0) Anaesthetic room set Holding and recovery p set Ultrasound set Fluoroscopy set Urography set Tomography set Magnetic resonance imaging (MRI) suite set set (0) CAT room set set (0) Angiography set set (0) Endoscopy set set (0) Lineac bunkers set set (0) General purpose rooms In-patient accommodation Single-bed room Multi-bed room, per bed space Treatment room (Appropriate for adult acute, children and the elderly) Renal Per dialysis station Per bed space set for the ward unit () set () Critical care area set () set () Per bed space Note: One set either side of the bed space, if installed in fixed location, eg trunking; or both sets in an articulated supply pendant that can be positioned either side of the bed space. Equipment service room, per work space Coronary care unit (CCU) 4 p p p p 4 4 p p p sets* (4) set set () Per bed space 4 4 4 sets* (4)(6)(7) for both sets of AVSUs () 3 Provision of terminal units, and the location of AVSUs, local alarm indicator panels and LVAs

8 Department O N O N O/O MA4 SA7 VAC AGSS He/O AVSU Alarm High dependency unit (HDU) Per bed space 4 4 4 sets* (4)(6)(7) for both sets of AVSUs () Burns unit p p p sets* (4)(6)(7) for both sets of AVSUs () Adult mental illness accommodation set () Electro-convulsive therapy (ECT) room set () set hp/lp (0) Post-anaesthesia recovery, per bed space set () set () Adult acute day care accommodation Treatment room: p p set p set set hp/lp Surgeon if (p) (0) Post anaesthesia recovery, per bed space p set set () Day patient accommodation set () Single bed room Multi-bed room, per bed space Treatment room p set ()(3) set hp/lp () Endoscopy room p p p set ()(3) set (0) Fracture clinic Plaster room p p set () set (9) Oral surgery, orthodontic department set () set hp/lp if (p) (9) Consulting/treatment room, type Consulting/treatment room, types and 3 Recovery room, per recovery position Appliance laboratory, per workstation Out-patient department p p Dental air will be required p p set/46 rooms set/46 rooms set set set hp/lp () Treatment room/cubicles p p p set ()(8) set hp/lp if (p) (9) Medical gases HTM 0-0 Medical gas pipeline systems Part A: Design, installation, validation and verification

9 Sterile services department Wash room Inspection, assembly and packing (IAP) room set () set (9) Other departments (as required by project team) set () set located (9) as appropriate Notes: * Dual circuits. ** Where the delivery and neonatal units are in close proximity, the equipment service room can be shared. Dental vacuum only. p = Project team option. hp/lp = high-pressure and low-pressure alarms for oxygen, medical air and nitrous oxide when installed together. All other local alarms, low pressure only. () Departmental AVSUs installed on the hospital street side of fire compartment doors. () Installed immediately outside the room. (3) Where air is used to control movable pendant fittings, it should be taken from the 7 bar surgical air system. (4) In addition to the dual circuits, additional AVSUs will be required to sub-divide the number of terminal units controlled. This subdivision should be based on the layout of the accommodation; for example, if the recovery area is divided into a number of separate room/areas, each would have a separate sub-set (see Figures 4 and 5). (5) This is intended to provide some flexibility and the exact number will depend on the total number of rooms within the department. (6) If a high-dependency unit is included within general in-patient accommodation, a separate set of AVSUs should be provided for the unit. In addition to the departmental valves or the ward as a whole, an additional set will be required to control the single-bed, multi-bed and treatment rooms. (7) Department AVSUs may be required if the units are large and separate from, for example, the critical care area. (8) Additional AVSUs may be required in a large unit: the aim should be to have about 8 rooms controlled by a set of valves discretion is required to arrive at the logical number. (9) Installed in reception area. (0) Installed in the operating room in the main panel or within the room, or an ante-room, eg control room of an MRI device. () Installed at the main staff base (nurses station). () Installed in the room space with the AVSUs. (3) Separate AVSUs will be required if endoscopy room is included. (4) Carbon dioxide is used for insufflation during some surgical procedures. A pipeline installation is a project team option and is covered in Chapter. Two NIST connector bodies units should be installed. General: Normally, departmental AVSUs would be installed at the hospital street side of the entrance doors to a department and would reflect the method of horizontal evacuation in the event of an emergency. In some large departments, for example an operating department, the clean-service corridor is likely to cross one or more fire compartment walls. Additional AVSUs may therefore be required to reflect the evacuation route. If a department includes one or more floors, a set of AVSUs should be provided for each floor, which will act as emergency overall fire valves. AVSUs for zones within critical care areas should be located where they can be seen by staff not necessarily at the staff base. Local alarms within critical care areas should be provided for the individual space; that is, if a critical care area of, say, 8 beds is sub-divided into three separate six-bed wards, there should be one alarm only for each space (not one for each of the dual circuits). 3 Provision of terminal units, and the location of AVSUs, local alarm indicator panels and LVAs

Medical gases HTM 0-0 Medical gas pipeline systems Part A: Design, installation, validation and verification Table 3 Oxygen: design and diversified flows Department In-patient accommodation (ward units): Single 4-bed rooms and treatment room Ward block/department Accident & emergency: Resuscitation room, per trolley space Major treatment/plaster room, per trolley space Post-anaesthesia recovery, per trolley space Treatment room/cubicle Operating: Anaesthetic rooms Operating rooms Post-anaesthesia recovery Maternity: LDRP rooms: Mother Baby Operating suites: Paediatrician Post-anaesthesia recovery In-patient accommodation: Single/multi-bed wards Nursery, per cot space Special care baby unit Radiological: All anaesthetic and procedures rooms Design flow for each terminal unit (L/min) 0 0 00 0 0 0 0 0 00 0 0 0 0 0 00 00 Diversified flow Q (L/min) Q w = 0 + [(n )6/4] Q d = Q w [ + (nw )/] Q = 00 + [(n )6/4] Q = 0 + [(n )6/4] Q = 0 + [(n )6/8] Q = 0 + [(n )6/0] Q = no addition made Q = 00 + (nt )0 Q = 0 + (n )6 Q = 0 + [(n )6/4] Q = 0 + [(n )3/] Q = 00 + (ns )6 Q = 0 + (n )3 Q = 0 + [(n )3/4] Q = 0 + [(n )6/6] Q = 0 + [(n )3/] Q = 0 + (n )6 00 Q = 0 + [(n )6/3] Critical care areas 0 Q = 0 + [(n )6]3/4 Coronary care unit (CCU) 0 Q = 0 + [(n )6]3/4 High-dependency unit (HDU) 0 Q = 0 + [(n )6]3/4 Renal 0 Q = 0 + [(n )6/4] CPAP ventilation 75 Q = 75n 75% Adult mental illness accommodation: Electro-convulsive therapy (ECT) room Post-anaesthesia, per bed space Adult acute day care accommodation: Treatment rooms Post-anaesthesia recovery per bed space Day patient accommodation (as In-patient accommodation ) Oral surgery/orthodontic: Consulting rooms, type Consulting rooms, types & 3 Recovery room, per bed space 0 0 0 0 0 0 0 Q = 0 + [(n )6/4] Q = 0 + [(n )6/4] Q = 0 + [(n )6/4] Q = 0 + [(n )6/4] As In-patient accommodation Q = 0 + [(n )6/] Q = 0 + [(n )6/3] Q = 0 + [(n )6/6] Out-patient: Treatment rooms 0 Q = 0 + [(n )6/4] Equipment service rooms, sterile services etc 00 Residual capacity will be adequate without an additional allowance 6

4 Gas flow Legend for Tables 3: Q = diversified flow for the department; Q w = diversified flow for the ward; Q d = diversified flow for the department (comprising two or more wards); n = number of beds, treatment spaces or single rooms in which the clinical procedure is being performed, not the individual numbers of terminal units where, in some cases, more than one is installed; ns = number of operating suites within the department (anaesthetic room and operating room). nw = number of wards nt = number of theatres supply. Care should be taken when calculating air exchange rates in wards/rooms in which large numbers of CPAP machines may be in use simultaneously and where failure of mechanical ventilation could result in raised ambient oxygen concentrations. Consideration should be given to installation of systems to warn of ventilation failure and oxygen concentrations above 3%. Maternity 4.4 For LDRP (labour, delivery, recovery, post-partum) rooms, the diversified flow is based on 0 L/min for the first terminal unit and 6 L/min for 5% of the remainder. Two cot spaces may be provided, each with a terminal unit. Only one will be considered to be in use. The diversified flow for cot spaces is based on 0 L/min for the first and 50% of the remainder at 3 L/min. 4.5 In the event of multiple births, the additional gas usage will have negligible overall effect on the total flow. 4.6 Maternity department operating rooms are designed as a suite; that is, it is presumed that oxygen will be provided either in the anaesthetic room or in the operating room. In post-anaesthesia recovery, it is assumed that 75% of beds will require oxygen to be delivered. Hyperbaric oxygen chambers 4.7 Hyperbaric oxygen chambers should be supplied from a separate branch from the main riser/ distribution pipe: the pipeline system should be from a liquid supply source. Typical flows for a single patient chamber are as shown in Table 4. Table 4 Gas flow hyperbaric chambers O atmosphere and recirculation: On open circuit On recirculation Max. time for one complete treatment hours hours Total consumption for max. treatment time (L) 30,000 7,50 Consumption for each additional minute (L/min) O only, no recirculation hours 30,000 50 O delivery by built-in breathing mask and overboard pump O delivery by built-in breathing hood and overboard pump 50 hours,00 0 hours 7,50 60 7

Medical gases HTM 0-0 Medical gas pipeline systems Part A: Design, installation, validation and verification Notes a. The flows for a recirculating unit assume the standard method of operation is recirculation throughout the treatment. It is recommended that the pipeline should be designed for open circuit operation to ensure adequate flow under all conditions. b. Clinical practice may require the inclusion of air during the treatment; it may also be necessary to switch to air in the unlikely event of an oxygen convulsion. Therefore consideration should be given to the provision of medical air from a separate dedicated medical air plant in accordance with Chapter 7. c. Some hyperbaric chambers use air as a buffer and consequently less oxygen is consumed. The advice of the manufacturer should be sought. Where this is the case, the air should be supplied from a separate supply system complying with the requirements for medical air systems. Nitrous oxide 4.8 Nitrous oxide is provided for anaesthetic purposes and occasionally for analgesic purposes. In all cases, each terminal unit should be capable of passing 5 L/min, but in practice the flow is unlikely to exceed 6 L/min. 4.9 When calculating diversities in a department, 5 L/ min is allowed for the first and 6 L/min for the remainder, subject to the appropriate diversity factor being applied (see Table 5). 4.30 It is assumed that, for an operating department, nitrous oxide may be in use simultaneously in all operating rooms. As it is unlikely that a patient would be anaesthetised in the anaesthetic room at the same time that a patient in the associated operating room was continuing to be treated under an anaesthetic (and because the duration of induction is short), no additional flow is included. Nitrous oxide/oxygen mixture 4.3 All terminal units should be capable of passing 75 L/min for a very short period (normally of five seconds duration) to supply inhalationary gasps by the patient, and a continuous flow of 0 L/min. The actual flow would not normally exceed 0 L/ min. 4.3 The diversified flow in delivery rooms is based on 75 L/min for the first bed space and 6 L/min for each of the remainder, of which only half of the women in labour will be using gas for 50% of the time. (The peak inhalationary gasp is 75 L/min, whereas the respirable minute volume will be catered for with a flow of 6 L/min it should also be borne in mind that a woman in labour would not continuously breathe the analgesic mixture.) For larger maternity departments with twelve or more LDRP rooms, two peak inhalationary gasps are included. 4.33 Nitrous oxide/oxygen mixture may be used in other areas for analgesic purposes. The diversified flow is based on 0 L/min for the first treatment space, and 6 L/min for a quarter of the remainder for 5% of the time. 4.34 Design and diversified flows for nitrous oxide/ oxygen mixtures are given in Table 6. Table 5 Nitrous oxide: design and diversified flows Department Design flow for Diversified flow Q (L/min) each terminal unit (L/min) Accident & emergency: resuscitation room, per trolley space 0 Q = 0 + [(n )6/4] Operating 5 Q = 5 + (nt )6 Maternity: operating suites 5 Q = 5 + (ns )6 Radiological: all anaesthetic and procedures rooms 5 Q = 0 + [(n )6/4] Critical care areas 5 Q = 0 + [(n )6/4] Oral surgery/orthodontic: consulting rooms, type 0 Q = 0 + [(n )6/4] Other departments 0 No additional flow included Equipment service rooms 5 No additional flow included 8

4 Gas flow Table 6 Nitrous oxide/oxygen mixtures design and diversified flows Department Maternity: < LDRP room(s), mother > LDRP rooms Design flow for each terminal unit (L/min) Diversified flow Q (L/min) 75 Q = 75 + [(n )6/] Q = 75 x + [(n )6/] Other areas 0 Q = 0 + [(n )0/4] Equipment service rooms 75 No additional flow included Air 4.35 Air is used to provide power for several types of equipment including surgical tools, ventilators and nebulisers. Oxygen should be avoided as a power source because of fire risk and cost, and should not be used where medical air is available, unless specifically recommended by the device manufacturer. 4.36 Air should be provided at two different pressures but to the same Ph. Eur. standard: a. a pressure of 0 kpa is required for medical air to drive ventilators and for other respiratory applications; b. a pressure of 700 kpa or higher is required for surgical air to drive surgical tools. Medical air 0 kpa General 4.37 The use of medical air, particularly for respiratory use and during anaesthesia, has increased markedly in recent years. This service is the most critical of the medical gas services, since air-powered ventilators cease to operate in the event of failure of the supply. 4.38 Medical air is also directly inhaled by patients during ventilation. It may also be used to dilute oxygen before administration because of the potentially toxic effects of pure oxygen. 4.39 The supply system for medical air 0 kpa may be a manifold system, a compressor system or a proportioning system (synthetic air), and includes an emergency reserve manifold. A compressor plant, or synthetic air supply, should always be specified where air-powered ventilators are to be used. 4. One of the major uses of medical air is for patients ventilators, which fall into two main categories those used during anaesthesia and those used during critical care. Pneumatically-powered ventilators can use up to 80 L/min free air continuously. The exact flow requirements will depend on the design of the ventilator. The flow and pressure requirements for some typical ventilators are given in Table 7. 4.4 Current models of anaesthetic ventilator are very similar to critical care models, and may require peak flows of up to 80 L/min and average flows of 0 L/min. Almost all such units are pneumatically driven and electronically controlled. 4.4 Medical air 0 kpa is also used for other equipment such as anaesthetic gas mixers, humidifiers and nebulisers. The flow rates normally required would not exceed 0 L/min, and this flow is always in excess of the actual volume respired. Pressure requirements 4.43 A minimum pressure required at terminal units for respiratory use is 370 kpa. 4.44 Medical air should not be used to supply mechanical services (see paragraph 7.30). 4.45 Some medical gas pendants use the medical air supply for operating the control/retraction system. This is permitted, provided that: a. a flow limiting device is provided to protect the medical air system in the event of failure of any downstream component; b. a non-return valve is incorporated to protect the system integrity; c. appropriate AVSU arrangements are in place (see Chapter 3). (The surgical air supply should be used to provide the power source whenever possible.) 4.46 The flow requirements should be ascertained and taken into account prior to the installation of the equipment. 9

Medical gases HTM 0-0 Medical gas pipeline systems Part A: Design, installation, validation and verification Flow requirements 4.47 Flow requirements for medical air are given Table 8. In ward areas and treatment rooms, the use of medical air is most likely to be for nebulisers. In these areas they should be capable of passing 0 L/min, although typically 0 L/min will be required in in-patient accommodation where air is used for nebulisers. Table 7 Typical pressure and flow requirements for ventilators and nebulisers Ventilator type Pressure (kpa) Flow (L/min) Anaesthesia, typically gas-driven, electronically controlled Nominally 0. Max 600 () Pneumatically driven ventilators use up to 80 max. 0 continuous Critical care, electrically controlled, gas-powered Nominally 0. Max 600 () 80 peak () 80 continuous Neonatal, gas-driven, electronically controlled Nominally 0. Max 600 () 80 peak () continuous Nebulisers 0 0 Notes:. It is strongly recommended that ventilators are not connected to the 700 kpa system since their blenders only work satisfactorily with a tolerance of about 0%: with high differential pressures for air and oxygen an incorrect mixture could be obtained.. These flows can be achieved under certain clinical conditions. The peak flows are usually of very short duration. 30

4 Gas flow Table 8 Medical air 0 kpa design and diversified flows Department In-patient accommodation (ward units): Single/multi-bed and treatment rooms () Ward block/department Accident & emergency: Resuscitation room, per trolley space Major treatment/plaster room, per trolley space Post-anaesthesia recovery, per trolley space Operating: Anaesthetic rooms Operating rooms Post-anaesthesia recovery Maternity: LDRP rooms: Baby () Operating suites: Post-anaesthesia recovery Neonatal unit (SCBU) Design flow for each terminal unit (L/min) 0 0 Diversified flow Q (L/min) Q w = 0 + [(n )0/4] Q d = Q w [ + (nw )/] Q = + [(n )0/4] Q = + [(n )0/4] Q = + [(n )/4] No additional flow included Q = + [(nt )/4] Q = + [(n )0/4] Q = + [(n )/4] Q = + [(n )/4] Q = + [(ns )0/4] Q = + [(n )/4] Q = n Radiological: All anaesthetic and procedures rooms Q = + [(n )/4] Critical care areas (3) 80 Q = 80 + [(n )80/] High-dependency units 80 Q = 80 + [(n )80/] Renal 0 Q = 0 + [(n )0/4] Oral surgery/orthodontic: Major dental/oral surgery rooms Q = + [(n )/] All other departments No additional flow allowance to be made Equipment service rooms No additional flow included Notes:. It is assumed that a patient will use oxygen in a ward or in the treatment room.. Where two cot spaces have been provided in an LDRP room, assume only one will require medical air. 3. This diversified flow is also used for helium/oxygen mixture (see paragraph 4.73). Surgical air 700 kpa 4.48 The pressure requirements of surgical tools are between 600 and 700 kpa and flows may vary between 00 and 350 L/min (STP; see Table 9). Most surgical tools are designed to operate within this pressure range. Higher pressures are likely to cause damage to tools. Inadequate tool performance, however, is likely to result from the lack of flow at the specified pressure. 4.49 The introduction of synthetic air (from on-site blending of oxygen and nitrogen) leads to the possibility of using nitrogen as the power source for surgical tools. 4.50 The pipeline systems should be designed to provide a flow of 350 L/min at 700 kpa at the outlet from the terminal unit. Existing systems may not meet this requirement (but should be capable of delivering 50 L/min at the terminal unit). Note Some surgical tools require up to 500 L/min at up to 0 kpa. These will require a separate supply, normally from cylinders. 3

Medical gases HTM 0-0 Medical gas pipeline systems Part A: Design, installation, validation and verification Table Vacuum design and diversified flows Department In-patient accommodation: Ward unit Multiple ward units Accident & emergency: Resuscitation room, per trolley space Major treatment/plaster room, per trolley space Post-anaesthesia recovery, per trolley space Treatment room/cubicle Operating: Anaesthetic rooms Operating rooms: Surgeon Operating suites Post-anaesthesia recovery Maternity: LDRP rooms: Mother Baby Operating suites: Obstetrician Operating suites Post-anaesthesia recovery In-patient accommodation: Ward unit comprising single, multi-bed and treatment room Multi-ward units Nursery, per cot space SCBU Design flow for each terminal unit (L/min) Diversified flow Q (L/min) Q = Q d = + [(n )/4] Q = + [(n )/4] Q = + [(n )/4] Q = + [(n )/4] Q = + [(n )/8] No additional flow included Q = Q = Q s = 80 + [(ns )80/] Q = + [(n )/4] Q = + [(n )]/4 No additional flow included Q = Q = Q s = 80 + [(ns )80/] Q = + [(n )/4] Q = Q = + [(n )/] No additional to be included Q = + [(n )/4] Radiology/diagnostic departments: All anaesthetic and procedures rooms Q = + [(n )/8] Critical care areas Q = + [(n )/4] High-dependency units Q = + [(n )/4] Renal Q d = + [(n )/4] Adult mental illness accommodation: ECT room Post-anaesthesia, per bed space Adult acute day care accommodation: Treatment rooms Post-anaesthesia recovery per bed space Day patient accommodation (as In-patient accommodation ) Oral surgery/orthodontic: Consulting rooms, type Consulting rooms, types & 3 Recovery room, per bed space Q = + [(n )/4] Q = + [(n )/4] Q = + [(n )/4] Q = + [(n )/8] As In-patient accommodation Dental vacuum only Dental vacuum only Q = + [(n )/8] Out-patient: Treatment rooms Q = + [(n )/8] Equipment service rooms, sterile services etc Residual capacity will be adequate without an additional allowance 34