Relationship Between Dynamic Expiratory Time Constant τ edyn and Parameters of Breathing Cycle in Pressure Support Ventilation Mode

Similar documents
Clinic of Cardiac Surgery, East Slovakian Institute of Cardiovascular Diseases and Faculty of Medicine Safarik University, Košice, Slovakia.

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Mechanical Ventilation

Mechanical Ventilation

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Using Common Ventilator Graphics to Provide Optimal Ventilation

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Pressure Controlled Modes of Mechanical Ventilation

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

Flight Medical presents the F60

ASV. Adaptive Support Ventilation

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

Technical Data and Specifications

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Selecting the Ventilator and the Mode. Chapter 6

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

Minimum size for maximum performance

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Potential Conflicts of Interest Received research grants from Hamilton, Covidien, Drager, General lel Electric, Newport, and Cardinal Medical Received

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Automatic Transport Ventilator

APRV: Moving beyond ARDSnet

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB

Objectives. Respiratory Failure : Challenging Cases in Mechanical Ventilation. EM Knows Respiratory Failure!

AQAI SIS MODULES AND MODELS PAGE 1. AQAI Simulator Interface Software AQAI SIS

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Organis TestChest. Flight Simulator for Intensive Care Clinicians

Physiological based management of hypoxaemic respiratory failure

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

Introduction to Conventional Ventilation

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

Lung recruitment maneuvers

Key words: intrahospital transport; manual ventilation; patient-triggered ventilation; respiratory failure

Mechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"

Mechanical Ventilation

SLE5000 Infant Ventilator with HFO

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

Guide to Understand Mechanical Ventilation Waveforms

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

VENTILATORS PURPOSE OBJECTIVES

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

Neonatal Assisted Ventilation. Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI.

Driving Pressure. What is it, and why should you care?

Equipment options. Chassis: With bottle holder Without bottle holder. Measuring modules: Agas with CO2 Agas with CO2 + O2

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

Managing Patient-Ventilator Interaction in Pediatrics

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Supporting you in saving lives!

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02

RESPIRATORY REGULATION DURING EXERCISE

Operating Instructions for Microprocessor Controlled Ventilators

HAMILTON-G5. The modular high-end ventilation solution

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Mechanical Ventilation. Flow-Triggering. Flow-Triggering. Advanced Concepts. Advanced Concepts in Mechanical Ventilation

Disclosures. The Pediatric Challenge. Topics for Discussion. Traditional Anesthesia Machine. Tidal Volume = mls/kg 2/13/14

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Medical Ventilators. Ryan Sanford Daniel Tajik

Accumulation of EEV Barotrauma Affect hemodynamic Hypoxemia Hypercapnia Increase WOB Unable to trigger MV

Intelligent Ventilation solution from ICU to MRI

Why we should care (I)

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

3100A Competency Exam

Physiological Basis of Mechanical Ventilation

Health Professional Info

Spontaneously Breathing Lung Model Comparison of Work of Breathing Between Automatic Tube Compensation and Pressure Support

HAMILTON-C6. The next generation of intelligent ICU ventilators

COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL

INTENSIVE CARE VENTILATORS

Mechanical power and opening pressure. Fellowship training program Intensive Care Radboudumc, Nijmegen

QUICK REFERENCE GUIDE

AUTOVENT 4000 VENTILATOR

Panther 5 Acute Care Ventilator

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Medical Instruments in the Developing World

INTELLiVENT -ASV. The world s first Ventilation Autopilot

The Basics of Ventilator Management. Overview. How we breath 3/23/2019

Classification of Mechanical Ventilators

Ventilating the Sick Lung Mike Dougherty RRT-NPS


Gas exchange measurement in module M- COVX

New Frontiers in Anesthesia Ventilation. Brent Dunworth, MSN, CRNA. Anesthesia Ventilation. New Frontiers in. The amount of gas delivered can be

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

4/2/2017. Sophisticated Modes of Mechanical Ventilation - When and How to Use Them. Case Study 1. Case Study 1. ph 7.17 PCO 2 55 PO 2 62 HCO 3

PROBLEM SET 9. SOLUTIONS April 23, 2004

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

NSQIP showed that the University of Utah was a high outlier in for patients receiving >48 cumulative hours of mechanical ventilation.

Conflict of Interest Disclosure Robert M Kacmarek

PROCEDURE (TASK): ROUTINE VENTILATOR CHECK. 5. Verifies current ventilator Insures correspondence between physician's

MV-TRAINER USER MANUAL

PROBLEM SET 8. SOLUTIONS April 15, 2004

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

Transcription:

Physiol. Res. 67: 875-879, 2018 Relationship Between Dynamic Expiratory Time Constant τ edyn and Parameters of Breathing Cycle in Pressure Support Ventilation Mode P. ČANDIK 1, D. RYBÁR 1, F. DEPTA 1, F. SABOL 2, A. KOLESÁR 2, K. GALKOVÁ 5, P. TÖRÖK 1,5, V. DONIČOVÁ 4, Š. IMRECZE 1, M. NOSÁĽ 1, V. DONIČ 3 1 Clinic of Anesthesiology and Intensive Medicine, East Slovakian Institute of Cardiovascular Diseases and Faculty of Medicine, Safarik University, Košice, Slovakia, 2 Clinic of Cardiac Surgery, East Slovakian Institute of Cardiovascular Diseases and Faculty of Medicine, Safarik University, Košice, Slovakia, 3 Department of Human Physiology, Faculty of Medicine, Safarik University, Košice, Slovakia, 4 Department of Pathophysiology, Faculty of Medicine, Safarik University, Košice, Slovakia, 5 Faculty of Social Sciences and Health Care, University of Constantine Philosopher, Nitra, Slovakia Received August 24, 2017 Accepted June 21, 2018 On-line September 11, 2018 Summary Study of the relationship between ventilation parameters: monitored expiratory time constant τ edyn and breathing trigger frequency (f trig) and time of breathing cycle (T cy) are main goals of this article. Parameters were analyzed during last 4±2 h before weaning from ventilation in 66 patients ventilated in pressure support mode (PSV). We have found out, that there exist mathematical relationships, observed during adequate gas exchange, yet not described. Monitored parameters are represented by τ edyn, f trig and T cy. The analysis showed close negative correlation between T cy and f trig (R 2 =0.903). This implies that each increasing of τ edyn causes decreasing of f trig and vice versa. The calculation of regression equation between τ edyn and T cy outlined that T cy = 5.2625 * τ edyn + 0.1242 (R 2 =0.85). Regulation of respiratory cycles by the respiratory center in the brain is probably based on evaluation of τ edyn as the τ edyn probably represents a regulatory element and T cy regulated element. It can be assumed, that respiratory center can optimize the work of breathing in order to minimize energy in system patient + ventilator. The unique relationship, described above could be useful in clinical practice for development of new ventilation modes. Key words Time constant Weaning from ventilator Time of breathing cycle Breathing cycle Corresponding author V. Donič, Department of Human Physiology, Faculty of Medicine, Safarik University, Trieda SNP 1, Košice, Slovakia. E-mail: viliam.donic@upjs.sk Introduction The most important part of weaning from artificial lung ventilation (ALV) is the last four hours before patient disconnection from ventilator. Patient ventilator synchrony during this time is crucial. Ventilation parameters are usually changed according to physician personal experiences (Boles et al. 2007, Petter et al. 2003, Tassaux et al. 2002). Synchrony between spontaneous breathing activity (trigger) and ventilator settings is essential for successful weaning. The f trig is controlled by patient s spontaneous breathing activity. According to our experiences the best and mostly used weaning ventilation mode is pressure support ventilation (PSV) (Solsona et al. 2009, Tassaux et al. 2002). The patient s respiratory center keeps respiratory rhythm (frequency and inspiratory/expiratory time ratio) to maintain adequate minute ventilation. The settings of pressure support (by physician) are essential for adequate tidal volume (VT) in nonsufficient spontaneous ventilation. The exact, yet simple indicator for correct PHYSIOLOGICAL RESEARCH ISSN 0862-8408 (print) ISSN 1802-9973 (online) 2018 Institute of Physiology of the Czech Academy of Sciences, Prague, Czech Republic Fax +420 241 062 164, e-mail: physres@fgu.cas.cz, www.biomed.cas.cz/physiolres

876 Čandik et al. Vol. 67 pressure support settings does not exist. We have studied the relationship between parameters of ventilation and physical properties of lungs. Our previous study discovered significant difference (150-700 %) between measured Tau (τ meas ) and Tau calculated (τ calc ). By definition τ calc = airway resistance (Raw) * total lung compliance (Ctot). Therefore calculated τ is inappropriate for evaluation of lung mechanics. In this study we are using a measured expiratory time constant termed as dynamic time constant (τ edyn ). Measured τ edyn shows real time changes in pneumatic system properties. System consists of patient s lungs, airways, endotracheal tube (ET) and ventilator with actual ventilation parameters settings. The goal was to clarify relationship between breathing parameters and τ edyn (in system lungs + ventilator), which could be usable in clinical practice for ventilator setting. Material and Methods The retrospective study was done in 66 patients with following demographics parameters: weight 82±17 kg, height 168±12.5 cm and age 59±12 years. The patients have been ventilated for various diseases including pulmonary edema, ARDS, postoperative respiratory insufficiency and convulsions. Duration of artificial lung ventilation (ALV) was 4.5±2.3 days, and then the patients were observed during the last 4±2 h before disconnection from ALV in PSV mode. Average ventilation parameters are in Table 1. Table 1. Mean value of ventilation parameters: pressure of pressure support (Pps), positive end expiratory pressure (PEEP), minute ventilation (MV), breathing frequency (f), switching from inspirium to expirium (PSi/e) in percentage of maximal flow (Qi/e), f trig. Pps (cm H 2 O) 6.8 ± 1.5 PEEP (cm H 2 O) 5.2 ± 1.3 MV (l/min) 7.9 ± 1.4 f (breath/min) 17.5 ± 2.4 Qi/e (%) 5 Trigger sensitivity (l/min) 1 (Fresenius, Germany) titration. During the last two hours of measurement the Propofol infusion was stopped. Aerosol of Marcaine 0.5 % (Janssen, Sweden) in dose 2 ml/2 h was applied intratracheally for minimizing patient ventilatorasynchrony. After the last measurement we disconnected the patient via T-piece. Inclusive criteria were hemodynamic stability, PaO 2 /FiO 2 >300 mm Hg, no pain using the Critical Care Pain Observation Tool (CPOT 0-1), no severe brain damage or other pathology potentially affecting ventilation. Only PSV mode was used for weaning from ALV. We have used the servo ventilator (Aura-V, Chirana a.s., Slovakia) with lung mechanics monitor and data collection with ProfiLung software. The data were collected by computer every 3 min followed by multiparametric analysis of dependency of monitoring parameters showed relationship between τ edyn, f trig and time of breathing cycle (T cy ). Relationships between T cy, τ edyn and spontaneous breathing frequency (f trig ) were further evaluated. Methodology of measurement of τ edyn We measured τ edyn by using the iterative method. After recording the expiratory flow curve in time, the computer algorithm calculated the tidal volume (VT flow integral over time) from patient breath. Then, the program searched time values, when expiratory volume reached 63 %, 85 %, and 96 % of expiratory volume. Because gas flow curve during expirium is degressive (exponential), the time required to reach 63 %, 85 % or 96 % of expiratory volume, represent the time constant τ edyn1, τ edyn2, or τ edyn3. Generally for further calculation the first expiratory dynamic time constant value is used τ edyn1, indicated further as τ edyn. Why, we choose τ edyn1? In the first phase of expirium, when flow is highest also resistance is highest. During next phases of expirium represented by decreased flow, τ edyn2 and τ edyn3 are shorter, therefore no important and without big influence on our measurement. Therefore we choose value of τ edyn1 as value which we used. Described procedure has been repeated for each breath cycle. The sampling frequency of expiratory flow was 1 khz. Data are expressed as mean ± SD, from 66 patients during 4±2 h of weaning. During the first 2±1 h from the beginning of measurement, patients were lightly sedated Richmond Agitation-Sedation Scale (RASS -1+1) using Propofol Calculated and the measured value of Tau (τ) explanation of differences Tau calculated τ calc could be expressed as τ calc = Raw * Ctot. In this formula only the airway resistance (Raw) and the total compliance of the lungs and thorax (Ctot) are considered. On the other hand during measured

2018 Relationship Between Time Constant and Breathing Cycle Time 877 Tau τ meas the flow sensor is connected inside ventilator circuit, therefore the measured value of τ edyn reflects properties of all pneumatic elements including breathing circuit and mechanical properties of the lungs. In our patient group we use ALV using endotracheal tube (ET). The difference between the calculated and the measured value can be significantly different, which make the τ calc unusable in clinical practice. Because of retrospective study, the Ethical committee approval was not necessary. However our data comes from previously study The computer assisted ALV, which was approved by Ethical committee of the East Slovakian Institute of Cardiovascular Diseases: EK no. VZ/7 / KardO/2011. Results The presented results describe the new relationship of τ edyn, f trig and time of breathing cycle parameters T cy. The Figures 2 and 3 showed relationship between τ edyn and f trig as well as T cy during weaning. The results come from 66 patients during last four hours of ALV before disconnection form ventilator. Figure 1 shows changes of indicated ALV parameters during the weaning procedure lasting 4 h. We can see that the parameters of ventilator support pressure (Pps), peak airway pressure (Paw), positive end expiratory pressure (PEEP) decreased during weaning procedure (adjusted by physician), but minute ventilation (MV) remained stable. According to our experience, this pattern represents proper weaning (Tables 1 and 2). This was necessary to obtain for correct data collection. Table 2. Blood gas exchange parameters during weaning. FiO 2 0.35 ± 0.07 PaO 2 / FiO 2 (mm Hg) 378 ± 69 PaCO 2 (mm Hg) 42 ± 4.9 PaO 2 (mm Hg) 132 ± 34 ph 7.41 ± 0.08 Lactate (mmol/l) 1.6 ± 0.43 Data are expressed as mean ± SD, from 66 patients during 4±2 h of weaning. It indicates adequate gas exchange; therefore correction of ventilator settings during weaning procedure was not necessary. Fig. 1. ALV ventilation parameter settings. Pps pressure of pressure support, PEEP positive end expiratory pressure, Paw airway pressure, MV minute ventilation, expressed in mean (±SD) in 66 patients during 4 h of weaning (parameters Pps, PEEP was setting by physician during weaning).

878 Čandik et al. Vol. 67 Fig. 2. Average values of f trig and τ edyn during weaning. More precise analysis shows strong negative correlation between f trig and τ edyn. R 2 =0.903. Increasing of τ edyn leads to decrease in f trig and vice versa. Fig. 3. Linear regression between T cy and τ edyn during weaning (mean values, n=66). T cy = 5.2625 * τ edyn + 0.1242 (R 2 =0.858). It shows new interesting relationship between time of breathing cycle (T cy=60/ ftrig) and τ edyn. Discussion Discovery of relationship between ventilation parameters influenced by breathing regulation center are important. Understanding of principle of regulatory mechanism of breathing is crucial for setting of ventilation parameters during ALV. Weaning from ALV represents the procedure in which patient is completely disconnected from ventilator and ET tube is withdrawn (Boles et al. 2007, Meade et al. 2001). The exact indicator for correct weaning from ALV does not exist (Boles et al. 2007, Petter et al. 2003). Weaning is suitable model for study of interaction between patient natural breathing pattern and ventilator parameter settings. Our goal was to find possible new patients + ventilator relationship as well as internal principles, which plays a major role, which was not yet published. Monitoring of τ edyn as presented in Figures 2 and 3 provides information about dynamics of gas exchange. Parameter of τ edyn determines relationship between ventilation work and gas volume necessary for adequate ventilation (CO 2 elimination and O 2 supply). Measured τ edyn varies

2018 Relationship Between Time Constant and Breathing Cycle Time 879 during each breath cycle. τ edyn depends on flow, inspiratory and expiratory time, frequency of breathing, resistance of lungs-ventilator system (R sys ), compliance and inhomogenity of gas distribution in the lungs. It is impossible to easily assess τ edyn, but it is possible to measure it by described iteration method using ventilator computer. τ edyn value reflects all influences of ventilation settings, mechanical properties of the lungs as well as whole lung + ventilator system resistance (R sys ). The prolongation of τ edyn leads to drop in the f trig and vice versa as can be seen in Figure 2. The reason for this is probably to assure lowest energy consumption for gas exchange. The close relationship between τ edyn and f trig on Figure 3, could be also explained, that the breathing center optimizes ventilation to assure the lowest energy consumption. Work of breathing is reduced by lowering the resistance of airways by decreasing flow of gas. During stable clinical condition τ edyn represent probably regulatory component and T cy regulated component. The indirect indicator of ventilator pattern is f trig. Changes in the duration of τ edyn depend mainly on changes in the parameters of the pulmonary mechanics. To achieve effective gas exchange, the other ventilator parameters must probably adapt to τ edyn. Those parameters can be changed by patient s natural breathing regulatory mechanism or during ALV by ventilator parameter settings. However in the PSV mode, the only adjustable parameter is Pps and PSi/e. Conclusions We revealed and described a possible internal relationship between τ edyn and T cy in the system patient + ventilator. We observed negative correlation between f trig and τ edyn. According to our results, could be assume, that the respiratory center is probably reacting to changes of resistance and compliance by adjustment of breathing pattern according to changes in τ edyn. The respiratory center is able to set the optimal breathing pattern, even in the conditions, when patient is connected to ventilator by tubing, ET, connectors etc. τ edyn probably represents integrated value of the lung mechanics. Parameters T cy, VT and f trig, probably continuously adapt to the τ edyn with aim to reach the most effective gas exchange. The relationship described above T cy = 5.2625 * τ edyn + 0.1242, may have practical implication for adjusting ventilator settings during artificial lung ventilation and creation of better ventilator algorithm. Only using of iteration methods of measurement allowed monitoring of τ edyn during artificial lung ventilation in the system patient lungs + ventilator. This is a novelty of this study and brings practical benefit for clinical use. According to our knowledge described relationship was not published yet. Conflict of Interest There is no conflict of interest. References BOLES M, BION J, CONNORS A, HERRIDGE M, MARSH B, MELOT C, PEARL R, SILVERMAN H, STANCHINA M, VIEILLARD-BARON A, WELTE T: Weaning from mechanical ventilation. Eur Respir J 29: 1033-1056, 2007. MEADE M, GUYATT G, COOK D, GRIFFITH L, SINUFF T, KERGL C, MANCEBO J, ESTEBAN A, EPSTEIN S: Predicting success in weaning from mechanical ventilation. Chest 120 (6 Suppl): 400S-424S, 2001. PETTER AH, CHIOLERO RL, CASSINA T, CHASSOT P-G, MULLER XM, REVELLY J-P: Automatic "respirator/weaning" with adaptive support ventilation: the effect on duration of endotracheal intubation and patient management. Anesth Analg 97: 1743-1750, 2003. SOLSONA JF, DIAZ Y, VAZQUEZ A, PILAR GRACIA M, ZAPATERO A, MARRUGAT J: A pilot study of a new test to predict extubation failure. Crit Care 13: R56, 2009. TASSAUX D, DALMAS E, GRATADOUR P, JOLLIET P: Patient-ventilator interactions during partial ventilatory support: a preliminary study comparing the effects of adaptive support ventilation with synchronized intermittent mandatory ventilation plus inspiratory pressure support. Crit Care Med 30: 801-807, 2002.