Butterfly control & isolation valve with extended control range with RS232 interface

Similar documents
Butterfly control & isolation valve with extended control range with DeviceNet interface

Installation, Operating & Maintenance Instructions. Variable leak valve with manual actuator. Series 590 DN 16 mm (I. D. ⅝") E

Installation, Operating & Maintenance Instructions. All-metal gate valve with extended pneumatic actuator. Series 48 DN mm (I.D.

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

BUBBLER CONTROL SYSTEM

FTC130 Transmitter. Operating Manual

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

ACV-10 Automatic Control Valve

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Instruction Manual Contact Pressure Vacuum Gauge

SPECIFICATIONS ATTENTION

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Installation and Operation Manual

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

Manual Weighingblock VB2 series and Uniscale

Mass Flow Controller (MFC) for Gases

Technical Data Sheet MF010-O-LC

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

GAS FUEL VALVE FORM AGV5 OM 8-03

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

BUBBLER CONTROL SYSTEM

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Mass Flow Controller (MFC) for Gases

Pneumatic high-pressure controller Model CPC7000

TR Electronic Pressure Regulator. User s Manual

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

Digital Vacuum Regulator

GM Series Dual-Block Multi-Function Gas Control Valves

C-Flow Coriolis Mass Flow Meters

Cover Page for Lab Report Group Portion. Head Losses in Pipes

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

How to specify a product. Process Sensors and Mechanical Instruments

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

In Vivo Scientific, LLC INSTRUCTION MANUAL

Ambient Capacitance Gauge

Instruction Manual Differential Pressure Transmitter

Hand lever valves VHER

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

Mass Flow Controller (MFC) for Gases

OC Panel High Limit Aquastat Kit, Manual Reset p/n

SonoMeter 30 Energy Meters

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

DPC-30 DPC-100. Reference Manual

1. Functional description. Application program usage. 1.1 General. 1.2 Behavior on bus voltage loss and bus voltage. 1.

Cover Page for Lab Report Group Portion. Pump Performance

REASONS FOR THE DEVELOPMENT

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Proportional pressure regulators VPPE

Gas Module for Series GMS800

Datasheet: K-30 ASCII Sensor

SDI mA Pressure Transducer (Submersible + Dry) Model WL2100 Operating Manual

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

Pneumatic high-pressure controller Model CPC7000

Technical Data Sheet High Performance Digital Gas Mass Flow Meters and Controllers Features Description

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

Measuring range Δp (span = 100%) Pa

O.K. Safety first CDN USA. Table of Contents. Approvals. Attention IFGC CSA UL ANSI NFPA

High-performance submersible pressure transmitter For level measurement Model LH-10

Mass Flow Meter (MFM) for gases

Design Features. General Description

Bypass Mass Flow Controller (MFC) for gases

Transmitter CS 21 Operation Manual

[Instruments for vacuum measurement, checking and adjustment] 3

Capacitance Diaphragm Gauge

Valve Proving System VDK 200 A S06*

Pressure Automated Calibration Equipment

Model 601CV. Mass Flow Controller. Product Features:

628 Differential Pressure Decay Leak Tester

Controlling the prefeeder

User s Guide Temperature Sensor Converter TSC-599

Mass Flow Controller (MFC) for Gases

Pneumatic Outflow Control

4400 TwinHybrid Gas Seal

200 PSI COMPRESSORS - MODEL NUMBERS

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

TH-1800-T TRX-700-CNG Converter

444C DUAL PERFORMANCE VALUE PACK

AMS 2710 PCB pressure sensor module with V output

2/2-Way Proportional Valve (motor-driven)

Pressure Switches Application Guideline

Automatic Valve Proving Control

1)! DO NOT PROCEED BEYOND THIS MARK

Overview. Front Panel: Keypad and Display

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Mass Flow Meter (MFM) for Gases

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

MANUAL KPS Pressure Control Valve

Series MTS300 Motorized Test Stands

Transcription:

Butterfly control & isolation valve with extended control range with RS232 interface This manual is valid for the valve ordering number(s): 61534-KEAG-0003 configured with firmware 615P.1E.44 The fabrication number is indicated on each product as per the label below (or similar): made in Switzerland Fabrication No.: 615.......... /.... A...... Fabrication number sample picture Explanation of symbols: Read declaration carefully before you start any other action! Keep body parts and objects away from the valve opening! Attention! Hot surfaces; do not touch! Loaded springs and/or air cushions are potential hazards! Wear gloves! Disconnect electrical power and compressed air lines. Do not touch parts under voltage! Read these «Installation, Operating & Maintenance Instructions» and the enclosed «General Safety Instructions» carefully before you start any other action! 1/88

Imprint Manufacturer Website www.vatvalve.com Phone +41 81 771 61 61 Fax +41 81 771 48 30 Email CH@vatvalve.com Publisher Editor Print Copyright VAT Vakuumventile AG 2012 No part of these Instructions may be reproduced in any way (photocopies, microfilms or any other reproduction processes) nor may it be manipulated with electronic systems, duplicated or distributed without written permission from VAT. Offenders are liable to pay damages. The original VAT firmware and updated state of the art versions of the VAT firmware are intended for use with VAT products. The VAT firmware contains a limited, time unlimited user license. The VAT firmware may not be used for purposes other than those intended nor is it permitted to make copies of the VAT firmware. In particular, it is strictly forbidden to give copies of the VAT firmware to other people. The use of trade names, brand names, trademarks, etc. in these Instructions does not entitle third parties to consider these names to be unprotected and to use them freely. This is in accordance with the meaning of the laws and acts covering brand names and trademarks. 2/88

Contents: 1 Use of product... 5 1.1 Technical data... 6 2 Installation... 8 2.1 Unpacking... 8 2.2 Installation into the system... 8 2.3 Tightening torque... 10 2.3.1 Mounting of ISO-KF flanges... 10 2.4 Admissible forces... 10 2.4.1 Admissible forces at controller... 11 2.5 Requirements to sensor connection... 12 2.6 Electrical connection... 13 2.6.1 Ground connection... 13 2.6.2 Sensor supply concepts... 14 2.6.3 Power and sensor connection ( 15 VDC sensors) using SPS module... 15 2.6.4 RS232 interface connection... 16 2.6.5 Service port connection... 16 3 Operation... 17 3.1 Introduction... 17 3.1.1 Local operation... 18 3.1.2 Remote operation... 18 3.1.3 Safety mode... 19 3.1.4 Service indication... 19 3.2 Operation under increased temperature... 19 3.3 Behavior during power up... 19 3.4 Behavior in case of power failure... 20 3.5 Display information... 20 3.6 Setup procedure... 22 3.6.1 Interface configuration... 22 3.6.2 Valve and sensor configuration... 23 3.6.3 ZERO... 23 3.6.4 Pressure control configuration... 24 3.6.5 Control algorithms... 26 3.7 Close valve... 30 3.8 Open valve... 30 3.9 Position control... 31 3.10 Pressure control... 31 3.11 Tuning of pressure control performance... 31 3.11.1 Tuning of pressure control (adaptive)... 32 3.11.2 Tuning of pressure control performance with fixed PI control... 37 3.11.3 Tuning of pressure control (soft pump)... 42 3.12 RS232 interface... 46 3.12.1 Schematics... 46 3.12.2 Digital inputs... 48 3.12.3 RS232 Command syntax... 49 3.12.4 Control commands... 49 3.12.5 Inquiry commands... 50 3.12.6 Setup commands... 55 3.12.7 Error messages... 61 4 Trouble shooting... 62 5 Maintenance & repairs... 65 5.1 Maintenance procedures... 65 5.1.1 Replacement of plate o-ring... 67 5.1.2 Replacement of shaft feedtrough seals and plate o-ring... 70 5.2 Option board... 79 5.2.1 Durability of power fail battery... 79 5.2.2 Retrofit / replacement procedure... 80 3/88

6 Drawing... 83 7 Spare parts... 84 7.1 Control and actuating unit... 84 7.2 Valve unit... 85 7.3 Seal kit... 85 7.4 Accessories... 86 7.4.1 Centering ring with Viton o-ring... 86 8 Warranty... 87 4/88

1 Use of product This product is a Butterfly control valve with isolation functionality for downstream pressure control in vacuum systems. Use product for clean and dry indoor vacuum applications under the conditions indicated in chapter «Technical data» only! Other applications are only allowed with the written permission of VAT. 5/88

1.1 Technical data Control and actuating unit Input voltage 1) +24 VDC (±10%) @ 0.5 V pk-pk max. [connector: POWER] Power consumption 74 W [connector: POWER] Sensor power supply output 2) ±15 VDC (±5%) / 1000 ma max. [connector: SENSOR] Sensor input Signal input voltage / Input resistance ADC resolution Sampling time 0-10 VDC / Ri>100 kω (linear to pressure) 0.23 mv 10 ms [connector: SENSOR] Digital inputs 3) ±24 VDC max. [connector: INTERFACE] Digital outputs 3) Input voltage Input current Breaking capacity 70 VDC or 70 V peak max. 0.5 ADC or 0.5 A peak max. 10 W max. [connector: INTERFACE] Analog outputs 3) (optional) 0-10 VDC / 1 ma max. [connector: INTERFACE] PFO 4) battery pack Charging time Durability 2 minutes max. up to 10 years @ 25 C ambient refer to «Durability of power fail battery» for details Ambient temperature 0 C to +50 C max. (<35 C recommended) Pressure control accuracy ±5 mv or ±0.1% of setpoint, whichever is higher Position resolution / position control capability >28000 (full stroke) Closing time throttling only 0.5 s (full stroke) Opening time throttling only 0.5 s (full stroke) Closing time throttling & isolation 0.6 s (full stroke) Opening time throttling & isolation 0.6 s (full stroke) 1) Internal overcurrent protection by a PTC device. 2) Refer to chapter «Sensor supply concepts» for details. 3) Refer to chapter «Schematics» for details. 4) PFO = Power Failure Option. Refer to «Behavior in case of power failure» for details. 6/88

Valve unit Pressure range at 20 C Leak rate to outside at 20 C Leak rate valve seat at 20 C 1 x 10E-8 mbar to 1.2 bar (abs) 1 x 10E-9 mbar l/s 1 x 10E-9 mbar l/s Cycles until first service: - Isolation cycles (open - closed - open) 250 000 (with FKM (e.g. Viton ) seals, under clean and unheated conditions) - Throttling cycles (open - max. throttle - open) 2 000 000 (with FKM (e.g. Viton ) seals, under clean and unheated conditions) Admissible operating temperature Mounting position Wetted Materials 10 C to +120 C Valve seat must face chamber side. Control unit for ISO-KF version needs support, when mounted on horizontal piping and control unit does not hang. - Body Stainless steel 316L (1.4404 or 1.4435) - Plate Stainless steel 316L (1.4404 or 1.4435) - Shaft Stainless steel 316L (1.4404 or 1.4435) - Plate screws Stainless steel 316L (A4) - Plate seal FKM (e.g. Viton ) - Rotary feedthrough seals atmosphere side FKM (e.g. Viton ) - Rotary feedthrough seals vacuum side FKM (e.g. Viton ) - Slide bearing for shaft iglidur X Other parts Stainless steel 316L (1.4404 or 1.4435), 1.4122, 301 (1.4310), 304 (1.4303), 1.4571, 304 (A2) Max. differential pressure on plate during isolation Max. differential pressure on plate during opening and throttling Min. controllable conductance [N 2 molecular flow] Conductance in open position [N 2 molecular flow] Weight about Dimensions 1000 mbar 1000 mbar 0.1 l/s 120 l/s 3.6 kg Refer to dimensional drawing of specific valve ordering number (available on request) 7/88

2 Installation 2.1 Unpacking Valves DN63 / 2½" and larger must not be lifted solely by the actuator. 2.2 Installation into the system Fingers and objects must be kept out of the valve opening and away from moving parts. The valve plate may start to move just after power is supplied. Do not connect or disconnect sensor cable when device is under power. Connection overview Connectors at controller panel Ground connection 8/88

1. Install valve [1] into the vacuum system. Valve seat side must faced process chamber. The valve seat side is indicated by the symbol " " on the valve flange. Caution: Do not tighten the flange screws stronger than indicated under «Tightening torque». Caution: Do not admit higher forces to the valve than indicated under «Admissible forces». Note: Make sure that enough space is kept free to do preventive maintenance work. The required space is indicated on the dimensional drawing. Note: Control unit of valves with ISO-KF flanges (615.. K... ) needs support when mounted on horizontal piping and control unit does not hang. 2. Install a ground connection cable at controller. Refer to «Electrical connection» 3. Install sensor(s) [2] according to the recommendations of the sensor manufacturer and directives given under «Requirements to sensor connection». 4. Connect pressure sensor cable [3] to sensor(s) and then to valve (connector: SENSOR). Refer to chapter «Electrical connection» for correct wiring. Note: 61534-KEAG-0003 supports 1 sensor(s). 5. Connect valve to remote control unit [4] (connector: INTERFACE). Refer to «RS232 connection» for correct wiring. 6. Connect power supply [5] to valve (connector: POWER). Refer to chapter «Electrical connection» for correct wiring. Note: To provide power to the valve motor pins 4 and 8 must be bridged, otherwise motor interlock is active and the valve enters the safety mode and is not operative. Refer also to «Safety mode». 7. This valve may optionally be equipped with a heating device. Connect VAT heating device according to manual of respective heating device. 8. Perform «Setup procedure» to prepare valve for operation. Note: Without performing the setup procedure the valve will not be able to do pressure control. 9/88

2.3 Tightening torque 2.3.1 Mounting of ISO-KF flanges Tightening torques for ISO-KF flange connections depends on the type of seal which is used. Follow recommendations of seal manufacturer. 2.4 Admissible forces Forces from the weight of other components can lead to deformation of the valve body and to malfunction of the valve. The stress has to be relieved by suitable means. The following forces are admissible: Valve size Axial tensile or compressive force «F A» Bending moment «M» mm inch N lb. Nm lbf. 50 2 150 34 11 8 M FA 10/88

2.4.1 Admissible forces at controller Attention! Do not pushing, shocking load, or stressing the valve controller! Do not deposit anything at valve controller! The admissible force at valve controller in regards to the pedestal is shown in table below. In case higher force is applied, the pedestal could be permanently damaged! Overview Admissible force «F» F = Force a = middle of aluminium part of controller (b / 2) c = pedestal a F a 400 N c b a F a c b 11/88

2.5 Requirements to sensor connection To achieve fast and accurate pressure control a fast sensor response is required. Sensor response time: < 50ms The sensor is normally connected to the chamber by a pipe. To maintain that the response time is not degraded by this connection it needs to meet the following requirements: Inner diameter of connection pipe: Length of connection pipe: > = 10 mm < = 300 mm These conductance guidelines must include all valves and limiting orifices that may also be present. Make also sure that there is no obstruction in front of sensor connection port inside the chamber. The sensor should also be mounted free of mechanical shock and vibration. Dynamic stray magnetic fields may introduce noise to sensor output and should be avoided or shielded. 12/88

2.6 Electrical connection 2.6.1 Ground connection Recommendation for ground strap between controller and system (chassis) Material L (Length max.) B1 (min.) B2 (min.) d1 ( ) d2 ( ) copper tinned 200 mm 25 mm 25 mm 4.5 mm customized Recommended torque: 1,3 1,7Nm B2 L System Ground strap PE B1 d1 d2 Ground connection Chassis (FE) (Functional Earth) Valve controller Note: Connection plates of ground strap must be total plane for a good electrical contact! Note: The connection point at chassis (FE) must be blank metal (not coated). It is also possible to connect the ground strap at system chamber if it is well connected to PE. Note: Avoid low chassis cross section to the system PE connection. (Min. same cross section as ground strap) 13/88

2.6.2 Sensor supply concepts This valve has a built in sensor power supply (SPS). It supplies 15 VDC. Concept: Valve versions: External + 24 VDC supplied to POWER connector is converted into 15 VDC by the valve internal SPS and supplied to SENSOR connector to supply 15 VDC sensors. Refer to chapter «Power and sensor connection ( 15 VDC sensors) with optional SPS module» for schematic and correct wiring. 615.. -.. A. -.... / 615.. -.. C. -.... SPS module included Note: The SPS module can be retrofitted. Refer to chapter «Retrofit / replacement procedure» for instruction. 14/88

2.6.3 Power and sensor connection ( 15 VDC sensors) using SPS module [615.. -.. A. -.... / 615.. -.. C. -.... versions only] Pins 4 and 8 must be bridged for operation! An optional switch would allow for motor interlock to prevent valve from moving. Note: Use shielded sensor cable(s). Keep cable as short as possible, but locate it away from noise sources. Connect the 15 VDC sensors at DB 15 female sensor connector exactly as shown in the drawing above. Do not connect other pins that may damage sensor, power supply or controller! Connector: Use only screws with 4-40UNC thread for fastening the connectors! 15/88

2.6.4 RS232 interface connection Refer to «Schematics» for wiring information. 2.6.5 Service port connection The service port (connector: SERVICE) allows to connect the valve to a RS232 port of a computer. This requires a service cable and software from VAT. You can either use our freeware 'Control View', which can be downloaded from www.vatvalve.com or purchase our 'Control Performance Analyzer'. Alternatively the VAT Service Box 2 can be connected to the service port for setup and local operation. The service port is not galvanic isolated. Therefore we recommend using this only for setup, testing and maintenance and not for permanent control. Refer also to «Local Operation» for details and to «Spare parts / Accessories» for ordering numbers of service cable, software and Service Box 2. Connector: Use only screws with 4-40UNC thread for fastening the service port connector! 16/88

3 Operation Valve opening Fingers and objects must be kept out of the valve opening and away from moving parts. Risk of injury. Do not connect the controller to power before the valve isn t installed complete into the system. 3.1 Introduction This valve is designed for downstream pressure control in vacuum chambers. It can be employed in a pressure control mode or a position control mode. In both cases local or remote operation is possible. 17/88

3.1.1 Local operation Local operation means that the valve is operated via the service port using a computer or the Service Box 2. When using a computer, a service cable and a software from VAT are required. You can either download our freeware 'Control View' from www.vatvalve.com or purchase our 'Control Performance Analyzer'. These software are beneficial especially for setup, testing and maintenance. How to start: Connect service cable, start software and push button LOCAL to enable for operation. Then enter menu Setup/Sensor and do sensor configuration according to your application to make sure that you get the correct pressure displayed. Control view supports: - parameter setup - manual control - numeric monitoring - basic diagnostic 'Control Performance Analyzer' supports: - parameter setup - manual control - sequence control - numeric and graphical monitoring - data recording - data analysis - advanced diagnostic When communication to service port is interrupted the valve will change to remote operation. So when service cable will be disconnected or software will be shut down, the valve returns automatically to remote operation. This may result in an immediate movement of the valve depending on remote control. Refer to «Spare parts / Accessories» for ordering numbers of service cable, software and Service Box 2. 3.1.2 Remote operation This product is equipped with a RS232 interface to allow for remote operation. See section «RS232 Interface» for details. Control View software, 'Control Performance Analyzer' software or 'Service Box 2' may be used for monitoring during remote control. Note: In case Control View or Control Performance Analyzer software is connected to valve make sure REMOTE button is pushed to enable for remote operation. In case Service Box 2 is connected to valve make sure the LED on button LOCAL is OFF for remote operation. 18/88

3.1.3 Safety mode By means of an external switch (see connection diagrams «Electrical connection») the motor power supply can be interrupted. In this case the valve enters the safety mode. This motor interlock prevents the valve from moving (e.g. maintenance work). Data reading from the control unit remains possible. When motor interlock is active during power up the valve directly enters the safety mode and is not able to synchronize. Display shows D C or D999. In this case synchronization cycle will be done when motor interlock is deactivated. Then Display shows INIT for a moment followed by SYNC. When safety mode is entered from operation (i.e. pressure control mode), the unit will automatically switch to position control mode and remain at current position. Once motor interlock is deactivated the unit remains in position control mode. 3.1.4 Service indication This product is able to indicate that the valve unit needs to be cleaned, or an obstruction is present. A service request is indicated when the control unit detects that motor steps are apparently not effective. This may happen when the valve unit is heavily contaminated. These lost steps are recognized and will be repeated to attempt target position in the short term. But in the medium term the valve unit requires cleaning or inspection. Service request (SR) would be indicated on the display or could be read via remote operation. Refer to «Display information» for details. 3.2 Operation under increased temperature Hot valve Heated valve may result in minor or moderate injury. Do not touch valve and heating device during operation. Once heating is switched off (valve and system) await until the valve is cooled down complete before doing any work This valve may be operated in the temperature range mentioned in chapter «Technical data». 3.3 Behavior during power up Valve position before power up: Closed (isolated) All other than closed (not isolated) Refer also to chapter «Display information». Reaction of valve: Valve power up configuration = closed (default) Valve remains closed. Display shows alternately C C and INIT. Synchronization will be done when first movement command is received. Valve power up configuration = open Valve runs to max. throttle position to detect the limit stops to synchronize. Display shows configuration of product resp. SYNC until synchronization is done. Valve position after power up is open. Valve runs to max. throttle position to detect limit stop for synchronization. Display shows configuration of product resp. SYNC until synchronization is done. Valve position after power up is closed Valve position after power up is open 19/88

3.4 Behavior in case of power failure Valve position before power failure: Any Reaction of valve: Without Power Failure Option (PFO) 615.. -.. G. -....(No PFO, no SPS) 615.. -.. A. -....( No PFO, SPS) Valve remains at current position. All parameters are stored in a power fail save memory. For PFO retrofit and other options refer to chapter «Retrofit». 3.5 Display information 1 2 3 4 There is a 4 digit display located on the panel. It displays configuration, status and position information. For details refer to following tables. Display Power up: Description Digit 1 Digit 2 Digit 3 Digit 4 At first all dots are illuminated then configuration is displayed: Firmware version [e.g. 1E00] (1 st information for about 3s) Controller configuration (2 nd information for about 3s) In case D999 is displayed, motor interlock is active. Refer to «Safety mode» for details. SYNC indicates that powerup synchronization is running. 1 E 0 0 2 = RS232 interface 3 = RS232 interface with analog outputs 0 = basic 1 = with SPS 1) 2 = with PFO 2) 3 = with SPS 1) and PFO 2) 1 = 1 sensor version 2 = 2 sensor version S Y N C 1) SPS = optional ±15 VDC Sensor Power Supply module 2) PFO = Power Failure Option 20/88

Operation: Description / Mode Digit 1 Digit 2 Digit 3 Digit 4 PRESSURE CONTROL mode POSITION CONTROL mode Valve closed Valve open Closed / open interlock (Valve closed / open by digital input) HOLD (position frozen) activated ZERO running LEARN running Safety mode established. Refer to «Safety mode» for details. Power failure Service request 1) (valve requires cleaning) P V C O I H Z L D F 0... 100 = valve position (%, 0 = closed / 100 = open) 1) If SR is blinking alternatively with the actual mode display (e.g. P.11..SR) the valve requires cleaning. S R Note: RxD / TxD activity of RS232 communication is displayed by 2 blinking dots in digit 2. The lower dot indicates RxD activity where the upper dot indicates TxD activity. The indication is not real time. Fatal error: Description Digit 1 Digit 2 Digit 3 Digit 4 Fatal error occurred E Error code. Refer to «Trouble shooting» for details 21/88

3.6 Setup procedure To enable this valve for pressure control setup steps 1 to 5 must be performed. In case position control is required only it s sufficient to perform steps 1 to 3. Setup step 1 POWER UP 2 3 INTERFACE CONFIGURATION VALVE CONFIGURATION 4 ZERO 5 PRESSURE CONTROL COFIGURATION Description Turn on external + 24VDC power supply (and external 15 VDC for sensor power supply if required). Refer to chapter «Behavior during power up» for details. RS232 baudrate, parity, data length and number of stop bits for valve must be selected. Refer to chapter «Interface configuration» for details. Basic configurations of the valve must be adapted according to application needs. Refer to chapter «Valve configuration» for details. Compensation of the sensor offset voltage. Refer to chapter «ZERO» for details. Accommodation of PID controller to the vacuum system characteristic. Refer to chapter: «Pressure Control configuration» for details. Note: Without PRESSURE CONTROL COFIGURATION the valve is not able to run pressure control. 3.6.1 Interface configuration Interface configuration must be adapted according to application needs. The factory default setting of the interface is shown in the table below. Baud rate Data bits Stop bits Parity Digital input OPEN Digital input CLOSE 9600 7 1 EVEN not inverted Inverted Functionality of digital interlock inputs CLOSE VALVE and OPEN VALVE. These may be configured as not inverted, inverted or disabled. Default is not inverted. Refer also to «Digital inputs». Pressure and position range for RS232 communication must be selected. Default for pressure is 0-1 000 000. Default for position is 0-100 000. Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Do configuration in menu Setup / Interface. Remote operation: (Refer to chapter «setup commands» for details) 1. Send INTERFACE CONFIGURATION 2. Send RANGE CONFIGURATION 22/88

3.6.2 Valve and sensor configuration Basic valve configuration must be adapted according to application needs. Definition of valve plate position (CLOSE or OPEN) after power up sequence. Default is. Definition of valve plate position (CLOSE or OPEN) in case of a power failure. Default is not defined. Only for versions that have Power Fail Option equipped [615.. -.. H. -....; 615.. -.. C. -....]. ZERO function. This may be disabled or enabled. Default is enabled. Refer also to «ZERO». Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) 1. Do power up configuration in menu Setup / Valve. 2. Do power fail configuration in menu Setup / Valve. 3. Enable or disable ZERO function in menu Setup / Sensor. 4. Do sensor configuration in menu Setup / Sensor. Remote operation: (Refer to chapter «RS232 setup commands» for details) 1. Send VALVE CONFIGURATION 2. Send SENSOR CONFIGURATION 3.6.3 ZERO ZERO allows for the compensation of the sensor offset voltage. When ZERO is performed the current value at the sensor input is equated to pressure zero. In case of a 2 sensor system both sensor inputs will be adjusted. A max. offset voltage of +/- 1.4 V can be compensated. The offset value can be read via local and remote operation. Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to menu Zero / ZERO and follow instructions. Remote operation: (Refer to chapter «Control commands» resp. «Setup commands» for details) 1. Send OPEN VALVE 2. Wait until process chamber is evacuated and sensor signal is not shifting anymore. 3. Send ZERO Note: Do not perform ZERO as long as pressure gauge voltage is shifting otherwise incorrect pressure reading is the result. Refer to manual of sensor manufacturer for warm up time. Note: Do not perform ZERO, if the base pressure of your vacuum system is higher than 1 of sensor full scale. We recommend disabling ZERO function in this case; refer to «Valve and sensor configuration» of the setup procedure. Otherwise incorrect pressure reading is the result. 23/88

3.6.4 Pressure control configuration Local operation: ( Control View or Control Performance Analyzer ) 1. Open CV or CPA 2. Go to «Tools» > «Terminal» and send setup command s:02 according to application needs. (possibility of adjustment see below) Set Get Command s:02abcdefgh configuring Control mode a s:02a switch to Control mode a without configuring i:02 get configuration actual Control mode i:02a get configuration Control mode a Describtion Acknowledgement (within 10ms after reception of command) s:02 s:02 i:02abcdefgh i:02abcdefgh This command selects control mode, gain factor, sensor response time, setpoint ramp time, p-gain, i-gain and pump down for the PID controller. Remote operation: Note: Refer to chapter: «RS232 Command syntax» > «RS232 commands» > PID CONTROLLER CONFIGURATION. data length 8 characters a Control algorithm: 0 = adaptive downstream (default) 1 = fixed PI downstream control 2 = fixed PI upstream control 3 = soft pump Note: For adjustment of Control algorithm see table below. a (control mode) 0 adaptive downstream parameter description b Gain Factor: (default is 8 = 1.00) 0 = 0.10, 1 = 0.13, 2 = 0.18, 3 = 0.23, 4 = 0.32, 5 = 0.42, 6 = 0.56, 7 = 0.75, 8 = 1.00, 9 = 1.33, A = 1.78, B = 2.37, C = 3.16, D = 4.22, E = 5.62, F = 7.50, G = 0.0001, H = 0.0003, I =0.001, J = 0.003, K = 0.01, L = 0.02, M = 0.05 c Sensor Response Time (s): (default is 0 = 0.00) 0 = 0.00, 1 = 0.02, 2 = 0.04, 3 = 0.06, 4 = 0.08, 5 = 0.10, 6 = 0.15, 7 = 0.20, 8 = 0.25, 9 = 0.30, A = 0.35, B = 0.4, C = 0.50, D = 0.60, E = 0.80, F = 1.00 d Setpoint Ramp Time (s): (default is 0 = 0.0) 0 = 0.0, 1 = 0.5, 2 = 1.0, 3 = 1.5, 4 = 2.0, 5 = 2.5, 6 = 3.0, 7 = 3.5, 8 = 4.0, 9 = 4.5, A = 5.0, B = 5.5, C = 6.0, D = 6.5, E = 7.0, F = 7.5, G = 8.0, H = 8.5, I = 9.0, J =9.5, K = 10.0 e, f, g, h are not used. Set 0000 24/88

a (control mode) 1 Fixed PI downstream control and 2 Fixed PI upstream control parameter description b, c are not used. Set 00 d Setpoint Ramp Time (s): 0 = 0.0, 1 = 0.5, 2 = 1.0, 3 = 1.5, 4 = 2.0, 5 = 2.5, 6 = 3.0, 7 = 3.5, 8 = 4.0,9 = 4.5, A = 5.0, B = 5.5, C = 6.0, D = 6.5, E = 7.0, F = 7.5, G = 8.0, H = 8.5, I = 9.0, J =9.5, K = 10.0 ef P-Gain: 00 = 0.0010, 01 = 0.0013, 02 = 0.0018, 03 = 0.0024, 04 = 0.0032, 05 = 0.0042, 06 = 0.0056, 07 = 0.0075, 08 = 0.010, 09 = 0.013, 10 = 0.018, 11 = 0.024, 12 = 0.032, 13 = 0.042, 14 = 0.056, 15 = 0.075, 16 = 0.10, 17 = 0.13, 18 = 0.18, 19 = 0.24, 20 = 0.32, 21 = 0.42, 22 = 0.56, 23 = 0.75, 24 = 1.0, 25 = 1.3, 26 = 1.8, 27 = 2.4, 28 = 3.2, 29 = 4.2, 30 = 5.6, 31 = 7.5, 32 = 10, 33 = 13, 34 = 18, 35 = 24, 36 = 32, 37 = 42, 38 = 56, 39 = 75, 40 = 100 gh I-Gain: 00 = 0.0010, 01 = 0.0013, 02 = 0.0018, 03 = 0.0024, 04 = 0.0032, 05 = 0.0042, 06 = 0.0056, 07 = 0.0075, 08 = 0.010, 09 = 0.013, 10 = 0.018, 11 = 0.024, 12 = 0.032, 13 = 0.042, 14 = 0.056, 15 = 0.075, 16 = 0.10, 17 = 0.13, 18 = 0.18, 19 = 0.24, 20 = 0.32, 21 = 0.42, 22 = 0.56, 23 = 0.75, 24 = 1.0, 25 = 1.3, 26 = 1.8, 27 = 2.4, 28 = 3.2, 29 = 4.2, 30 = 5.6, 31 = 7.5, 32 = 10, 33 = 13, 34 = 18, 35 = 24, 36 = 32, 37 = 42, 38 = 56, 39 = 75, 40 = 100 3 soft pump d Setpoint Ramp Time (s): (default is 0 = 0.0) 0 = 0.0, 1 = 0.5, 2 = 1.0, 3 = 1.5, 4 = 2.0, 5 = 2.5, 6 = 3.0, 7 = 3.5, 8 = 4.0, 9 = 4.5, A = 5.0, B = 5.5, C = 6.0, D = 6.5, E = 7.0, F = 7.5, G = 8.0, H = 8.5, I = 9.0, J =9.5, K = 10.0 ef P-Gain: 00 = 0.0010, 01 = 0.0013, 02 = 0.0018, 03 = 0.0024, 04 = 0.0032, 05 = 0.0042, 06 = 0.0056, 07 = 0.0075, 08 = 0.010, 09 = 0.013, 10 = 0.018, 11 = 0.024, 12 = 0.032, 13 = 0.042, 14 = 0.056, 15 = 0.075, 16 = 0.10, 17 = 0.13, 18 = 0.18, 19 = 0.24, 20 = 0.32, 21 = 0.42, 22 = 0.56, 23 = 0.75, 24 = 1.0, 25 = 1.3, 26 = 1.8, 27 = 2.4, 28 = 3.2, 29 = 4.2, 30 = 5.6, 31 = 7.5, 32 = 10, 33 = 13, 34 = 18, 35 = 24, 36 = 32, 37 = 42, 38 = 56, 39 = 75, 40 = 100 bc and gh are not used. Set 00 and 00 example: s:0230010100 (s:02abcdefg) s:02 = command a bc d ef gh 3 = soft pump 00 = not used 1 = 0.5 (ramp time) 01 = 0.0013 (p-gain) 00 = not used Note: To optimize P-Gain, I-Gain and soft pump of the pressure controller refer to chapter «Tuning of control performance». 25/88

3.6.5 Control algorithms This valve is equipped with 3 different types of control algorithms. These are: Adaptive algorithm (automatic PI adjustment) To use the adaptive algorithm the valve must undergo the LEARN procedure prior to pressure control operation. Refer to chapter: 3.6.5.1 Adaptive algorithm. Fixed PI algorithm To use the fixed PI algorithm the PI values must be calculated according to the process condition. Refer to chapter: 3.6.5.2 Fixed PI. Soft pump To use the P control algorithm the P values must be calculated according to the process condition. Refer to chapter: 3.6.5.3 Soft pump. Prior to pressure control operation the control type must be selcted and the valve must be prepared accordingly. 26/88

3.6.5.1 Adaptive algorithm LEARN adapts the PID controller of the valve to the vacuum system and its operating conditions. LEARN must be executed only once during system setup. The LEARN routine determines the characteristic of the vacuum system. Based on this, the PID controller is able to run fast and accurate pressure control cycles. This characteristic depends on various parameters such as chamber volume, conductance and flow regime. Therefore it must be performed with a specific gas flow according to instruction below. The result of LEARN is a pressure versus valve position data table. This table is used to adapt the PID parameters. The data table is stored in the device memory which is power fail save. The data table can be up-/downloaded via Control Performance Analyzer software or remote interface. Due to encoding the data may not be interpreted directly. By an OPEN VALVE, CLOSE VALVE, POSITION CONTROL or PRESSURE CONTROL command the routine will be interrupted. Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to Learn / LEARN menu and follow instructions. Note: Gasflow calculation according to recommendation below is done automatically based on inputs. Remote operation: (Refer to chapter «RS232 control commands» resp. «RS232 setup commands» for details) 1. Send OPEN VALVE 2. Set specific gas flow according to calculation below and wait until flow is stable. LEARN does not need to be performed with the process gas. Instead N 2 or Ar may be used. 3. Send LEARN (with pressure limit set to full scale) Note: Sensor signal must not shift during LEARN. Wait until sensor signal is stable before LEARN is performed. Note: Learn may take several minutes. Do not interrupt the routine as a single full run is required to ensure fast and accurate pressure control. The PID controller covers 5% to 5000% of the gas flow which was used for learn. 27/88

Gasflow calculation for LEARN: Do not apply a different gasflow for learn than determined below. Otherwise pressure control performance may be insufficient. Note: Required pressure / flow regime must be known to calculate the most suitable learn gas flow for a specific application. 1. At first it is necessary to find out about the required control range respectively its conductance values. Each working point (pressure / flow) must be calculated with one following formulas. Choose the applicable formula depending on units you are familiar with. 1000 q WP C WP = p WP C WP q WP p WP required conductance of working point [l/s] gasflow of working point [Pa m 3 /s] pressure of working point [Pa] C WP = q WP p WP C WP q WP p WP required conductance of working point [l/s] gasflow of working point [mbar l/s] pressure of working point [mbar] C WP = q WP 78.7 p WP C WP q WP p WP required conductance of working point [l/s] gasflow of working point [sccm] pressure of working point [Torr] 2. Out of these calculated conductance values choose the lowest. C R = min(c WP1, C WP2,..., C WPn) C R required lower conductance [l/s] C WPx required conductance of working points [l/s] Note: To make sure that the valve is capable to control the most extreme working point verify that C R C min of the valve (refer to «Technical data»). 3. Calculate gasflow for learn. Choose the applicable formula depending on units you are familiar with. q L = p max 2000 C R q L gasflow for learn [Pa m 3 /s] p max max. pressure to control [Pa] C R required lower conductance [l/s] q L = p max 2 C R q L gasflow for learn [mbar l/s] p max max. pressure to control [mbar] C R required lower conductance [l/s] q L = 39.4 p max C R q L gasflow for learn [sccm] p max max. pressure to control [Torr] C R required lower conductance [l/s] 28/88

3.6.5.2 Fixed PI This valve may be used for downstream or upstream pressure control depending on configuration. The PI parameters of the pressure controller require correct adjustment. These parameters must be set once during system setup and are stored in the device memory which is power fail save. Based on the PI controller configuration, the valve is able to run fast and accurate pressure control cycles. The PI parameters can be calculated approximately with the formulas below. On request VAT can provide a parameter calculation tool. Note: In downstream control mode valve will move towards open when current pressure is higher then setpoint. In upstream control mode valve will move towards close when current pressure is higher then setpoint. P-Gain Step 1 (choose the applicable formula depending on units you are familiar with): d Valve diameter [mm] d V p SFS V Chamber volume [l] K = q max 1.737 10 6 p SFS Sensor full scale 1) [Pa] q max max. gasflow to control [Pa m 3 /s] Step 2: K = K = d V p SFS q max 1.737 10 3 d V p SFS 1.151 q max d Valve diameter [mm] V Chamber volume [l] p SFS Sensor full scale 1) [mbar] q max max. gasflow to control [mbar l/s] d Valve diameter [inch] V Chamber volume [l] p SFS Sensor full scale 1) [Torr] q max max. gasflow to control [sccm] 1) In case of 2 sensor operation use high range sensor for calculation (615.. -... Q -.... version only) P-Gain = K (but max. 10) K Gain I-Gain Step 1 (choose the applicable formula depending on units you are familiar with): T N1 = 66 V d d V Valve diameter [mm] Chamber volume [l] T N1 = 2.6 V d d V Valve diameter [inch] Chamber volume [l] Step 2: T N2 = T N1 (but max. 3) T N1 Reset time Step 3: I-Gain = K T N2 K T N2 Gain limited reset time 29/88

3.6.5.3 Soft pump This valve may be used for downstream control depending on configuration. The P parameters of the pressure controller require correct adjustment. These parameters must be set once during system setup and are stored in the device memory which is power fail save. Based on the P controller configuration, the valve is able to run fast and accurate pressure control cycles. The P parameters can be calculated approximately with the formulas below. On request VAT can provide a parameter calculation tool. Note: In downstream control mode valve will move towards open when current pressure is higher then setpoint. P-Gain Step 1 (choose the applicable formula depending on units you are familiar with): d Valve diameter [mm] d V p SFS V Chamber volume [l] K = q max 1.737 10 6 p SFS Sensor full scale 1) [Pa] q max max. gasflow to control [Pa m 3 /s] Step 2: K = K = d V p SFS q max 1.737 10 3 d V p SFS 1.151 q max d Valve diameter [mm] V Chamber volume [l] p SFS Sensor full scale 1) [mbar] q max max. gasflow to control [mbar l/s] d Valve diameter [inch] V Chamber volume [l] p SFS Sensor full scale 1) [Torr] q max max. gasflow to control [sccm] 2) In case of 2 sensor operation use high range sensor for calculation (615.. -... Q -.... version only) P-Gain = K (but max. 10) K Gain 3.7 Close valve Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Push CLOSE button Remote operation: (Refer to chapter «RS232 control commands» for details) Send CLOSE VALVE 3.8 Open valve Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Push OPEN button Remote operation: (Refer to chapter «RS232 control commands» for details) Send OPEN VALVE 30/88

3.9 Position control The valve position is directly controlled according to the position setpoint. Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Select or enter position setpoint Remote operation: (Refer to chapter «RS232 control commands» for details) Send POSITION CONTROL 3.10 Pressure control To prepare valve for PRESSURE CONTROL perform complete «Setup procedure». The valve has parameters that may be modified to tune pressure control performance. Refer to «Tuning of control performance». The included PID controller controls the chamber pressure according to the pressure setpoint by means of the valve position. The PID controller works with an adaptive algorithm to achieve best results under altering conditions (gasflow, gas type). Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Select or enter pressure setpoint Remote operation: (Refer to chapter «RS232 control commands» for details) Send PRESSURE CONTROL 3.11 Tuning of pressure control performance Tuning of pressure control performance with adaptive control, refer to chapter: 3.11.1 Tuning of pressure control (adaptive) Tuning of pressure control performance with fixed PI control, refer to chapter: 3.11.2 Tuning of pressure control performance with fixed PI control Tuning of pressure control performance with soft pump control, refer to chapter: 3.11.3 Tuning of pressure control (soft pump) 31/88

3.11.1 Tuning of pressure control (adaptive) Normally the default settings will result in good pressure control performance. For some applications tuning may be required to improve performance. The tuning procedures for each parameter (grey boxes) and its default values are described separately below. Strictly keep the order of the procedure. Start Set all paramters to default Pressure control ok? No Reset gain factor to 1.0 Gain factor adjustment Sensor delay adjustment Setpoint ramp adjustment Valve speed adjustment Gain factor adjustment Pressure control ok? No Pressure control ok? No Pressure control ok? No Pressure control ok? No End Contact VAT 32/88

3.11.1.1 Gain factor adjustment The gain factor effects: Stability Response time Default value is 1. Adjustment range is from 0.0001 to 7.5. Higher gain results in: faster response higher over- / undershoot of pressure Lower gain results in: slower response lower over- / undershoot of pressure Adjustment procedure: 1. Start with gain factor 1.0 2. Open valve. 3. Control a typical pressure / flow situation. 4. Repeat from step 2 with lower (higher) gain factors until optimal pressure response is achieved and stability is ok. Note: Normally adjustments down to gain factors of 0.42 should lead to good results. Otherwise you may need to improve sensor connection. Refer to «Requirements to sensor connection». Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Remote operation: (Refer to chapter «RS232 setup commands» for details) Set gain factor in menu Setup / Control Parameter Send PID CONTROLLER CONFIGURATION 33/88

3.11.1.2 Sensor delay adjustment Sensor delay adjustment effects: Stability Default value is 0.0s. Adjustment range is from 0 to 1.0s. Pipes and orifices for sensor attachment delay response time and so badly impact pressure control stability. By adapting this parameter to the approximate delay time stability problems can be reduced. But control response time will be slowed down by this measure. Note: Whenever possible sensors should be attached to the chamber according to «Requirements to sensor connection». This is the most effective measure against stability issues. If your gauge attachment fulfills these criteria do not use this parameter. Adjustment procedure: 1. Start with gain factor 1.0 and sensor delay 0s. 2. Open valve. 3. Control a typical pressure / flow situation. 4. Repeat from step 2 with higher sensor delays until best possible stability is achieved. 5. Adjustment gain factor again. Refer to «Gain factor adjustment». Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to Setup / Controller menu. Select sensor delay. Remote operation: (Refer to chapter «setup commands» for details) Send PID CONTROLLER CONFIGURATION 34/88

3.11.1.3 Setpoint ramp adjustment Setpoint ramp effects: Undershoot of pressure Response time Default value for Setpoint Ramp is 0.0s. Adjustment range for Setpoint Ramp is from 0 to 10 s. This parameter defines the time that is used to decrease / raise pressure between 2 setpoints. Especially in pressure decrease situations at low flows pressure response can be improved much by adapting setpoint ramp time. Pressure chart Without setpoint ramp optimizing With setpoint ramp optimizing P SFS P START P END t Choose the applicable formula depending on units you are familiar with. t = Setpoint Ramp Adjustment procedure: 1. Start with optimal gain factor and sensor delay time according to preceding tuning steps. 2. Control a typical pressure / flow situation. 3. Control a lower pressure. 4. Repeat from step 2 with longer setpoint ramps until best response is achieved. 5. Verify pressure control response for a setpoint raise situation. Note: In case a long ramp time is required to get optimal performance for pressure decrease situations it may be of advantage to apply different settings for decrease / raise control situations. Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to Setup / Control Parameter menu. Select setpoint ramp. Remote operation: (Refer to chapter «setup commands» for details) Send PID CONTROLLER CONFIGURATION 35/88

3.11.1.4 Valve speed adjustment Valve speed effects: Response time Default value is 1000. Adjustment range is from 1 to 1000. This parameter effects valve plate actuating speed. Speed adjustment is effective for PRESSURE CONTROL and POSITION CONTROL. Note: Normally best pressure control response is achieved with max. valve speed. In particular applications it may be of advantage to have a slower valve response. Note: OPEN and CLOSE are always done with max. speed. Adjustment procedure: 1. Use optimal gain factor, sensor delay time and setpoint ramp according to preceding tuning steps. 2. Open valve. 3. Control a typical pressure / flow situation. 4. Repeat from step 2 with slower valve speed until required response is achieved. Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to Setup / Control Parameter menu. Select valve speed. Remote operation: (Refer to chapter «RS232 setup commands» for details) Send VALVE SPEED 3.11.1.5 Required information for support (in case of unsuccessful tuning procedure): Go to Tools / Create Diagnostic File in Control View resp. Control Performance Analyzer and save file Pressure / flow / gas conditions to be controlled Chamber volume Pumping speed (l/s) and pump type (e.g. turbo pump) System description Problem description Send diagnostic file with and all required information to tuning-support@vat.ch 36/88

3.11.2 Tuning of pressure control performance with fixed PI control Normally the default settings will result in good pressure control performance. For some applications tuning may be required to improve performance. The tuning procedures for each parameter (grey boxes) and its default values are described separately below. Strictly keep the order of the procedure. Start Set P-Gain / I-Gain to default value and valve speed to maxium Yes Pressure control ok? No P-Gain to adjustment / optimizing I-Gain to adjustment / optimizing Pressure control ok? No Yes End Contact VAT 37/88

3.11.2.1 P-Gain adjustment The P-gain effects: Stability Response time Adjustment range is from 0.001 to 100. Higher gain results in: faster response higher over- / undershoot of pressure Lower gain results in: slower response lower over- / undershoot of pressure Adjustment procedure: 1. Start with P-Gain according to «Pressure controller configuration». 2. Open valve. 3. Control a typical pressure / flow situation. 4. Repeat with lower (higher) gain factors until optimal pressure response is achieved and stability is ok. Note: In case it s not possible to achieve stable pressure control you may need to improve sensor connection. Refer to «Requirements to sensor connection». Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to Tools / Terminal menu and send configuration command for adjustment. Refer to «Pressure controller configuration» for command details. Remote operation: (Refer to chapter «RS232 setup commands» for details) Send PID CONTROLLER CONFIGURATION 3.11.2.2 I-Gain adjustment The I-gain effects: Stability Response time Adjustment range is from 0.001 to 100. Higher gain results in: faster response higher over- / undershoot of pressure Lower gain results in: slower response lower over- / undershoot of pressure Adjustment procedure: 1. Start with I-Gain according to «Pressure controller configuration». 2. Open valve. 3. Control a typical pressure / flow situation. 4. Repeat with lower (higher) gain factors until optimal pressure response is achieved and stability is ok. Note: In case it s not possible to achieve stable pressure control you may need to improve sensor connection. Refer to «Requirements to sensor connection». Local operation: ( Control View, Control Performance Analyzer or Service Box 2 ) Go to Tools / Terminal menu and send configuration command for adjustment. Refer to «Pressure controller configuration» for command details. Remote operation: (Refer to chapter «RS232 setup commands» for details) Send PID CONTROLLER CONFIGURATION 38/88

3.11.2.3 Optimizing P gain and I gain Introduction PI controller mode is used if for any reason (e.g. too long system time constant) the adaptive control mode does not provide satisfying control performance. In PI controller mode the parameters P gain and I gain have to be set according to the systems characteristics. The best set of parameters can be found by using the empiric method below. 1. Optimizing P gain and I gain 1.1 Pressure and gas flow for optimization A PI controller delivers the best results for a certain working point (pressure/gas flow). If there is only one working point, this pressure and gas flow has to be used for optimizing P and I gain. If there are several working points that have to be covered, the pressure for optimizing is the medium pressure between highest and lowest pressure to be controlled, the gas flow for optimizing is the highest flow out of all working points. Two different pressure set points are necessary for optimization. Set point 1 (SP1) is the pressure for optimizing as determined above. Set point 2 (SP2) is about 10-20% lower than SP1. Example: pressure range: 4 10 Torr Flow range: 2 4 slm Pressure set points and gas flow for optimization: SP1 = 7 Torr SP2 = 6 Torr Gas flow = 4slm 39/88

1.2 Optimizing P gain While optimizing P gain, the gas flow determined above has to be constant all the time. Start optimization with P gain set to 1.0 and I gain set to 0.0. Set chamber pressure to SP2, wait until the pressure is stable. Set pressure to SP1. If the transition from SP2 to SP1 results in a significant pressure over shoot or even does not stabilize at all, the P gain is too high. If there is no over shoot and the pressure reaches SP1 asymptotically and very slow, P gain is too low. The optimal P gain value is found if the transition from SP2 to SP1 results in a slight pressure over shoot. It does not matter if there is still a deviation between SP1 and actual pressure. Example: 40/88

1.3 Optimizing I gain While optimizing I gain, the gas flow determined above has to be constant all the time. Start with P gain set to half of the value found when optimizing P gain and set I gain to 1.0. Keep the P gain constant. Set chamber pressure to SP2, wait until the pressure is stable. Set pressure to SP1. If the transition from SP2 to SP1 results in a significant pressure over shoot or if the valve position does not stabilize, I gain is to high. If the transition results in a slow asymptotical pressure rise and there is still a constant deviation to SP2, the I gain is too low. The optimal value for I gain is found if the transition from SP2 to SP1 result in just a slight pressure over shoot, a stable valve position and the actual pressure matches SP2 exactly. Example: Note: Check control performance over the whole control range with parameters above. 41/88

3.11.3 Tuning of pressure control (soft pump) Normally the default settings will result in good pressure control performance. For some applications tuning may be required to improve performance. The tuning procedures for each parameter (grey boxes) and its default values are described separately below. Strictly keep the procedure order. Start P-Gain to calculated value and valve speed to maxium Pressure control ok? No P-Gain to adjustment Valve speed adjustment Pressure control ok? No Pressure control ok? No End Contact VAT 42/88

3.11.3.1 Optimizing P gain This valve may be used to control pressure ramps during pump down. The P parameter of the pressure controller requires correct adjustment. This parameter must be set once during system setup and is stored in the device memory which is power fail save. Based on the soft pump controller configuration, the valve is able to run fast and accurate pressure control cycles. The P parameter can be evaluated using below instruction. Local operation: ( Control View resp. Control Performance Analyzer ) Go to Tools / Terminal menu and do the: Control cofiguration Control mode + P-Parameters Refer to chapter «Pressure control configuration» for details. Remote operation: (Refer to chapter «setup commands» for details) Send PID CONTROLLER CONFIGURATION Introduction Pump down control mode allows a completely user-defined pressure profile, usually from atmosphere down to some process pressure 1. Optimizing P gain The P gain value evaluated for soft pump control mode might be different than the P gain value evaluated for PI controller mode. When switching to pump down control mode the P gain value evaluated fort he PI controller has to be send to the valve controller. When switching back into PI controller mode the respective P gain value has to be send again. Adaptive pressure control mode ignores any P gain value. 1.1 Basic settings The pump down characteristic is determined by start pressure, end pressure and pump down time. This straight line from start pressure to end pressure. The VAT soft pump controller requires a pump down time shorter than 10 sec. for good control results. If the required pump down time is longer than 10 sec., the pump down curve has to be partitioned into sections shorter than 10 sec. with corresponding end pressure. Example: Start pressure: End pressure: Pump down time: 760 Torr 10 Torr 30 sec. Here the pump down time and the corresponding pressure is being divided into three sections. The host sends a new pressure set point every 10 sec.: Time Set point 0 sec. 760 Torr 10 sec. 510 Torr 20 sec. 260 Torr 30 sec. 10 Torr 43/88

1.2 Optimizing P gain WE start by setting the P gain to 1.0 as a trial value and adjust according to the response. The pump down routine has to be controlled as follows: Move control valve into close position Start pump down by opening the pump isolation valve or starting the pump and sending the first pressure set point to the valve controller. With the example above, the first pressure set point is 510 Torr. At each new interval (exceeding 10 sec) send the new pressure set point. Repeat until process pressure is achieved. While pumping down chamber pressure and valve position should be data logged to compare the actual pump down curve with the ideal straight pump down line. If the pressure follows the ideal pump down line with significant delay, the P gain is too low. If the pressure oscillates around the ideal pump down line or if the valve position oscillates, P gain is too high. P gain is optimized if the pressure follows the ideal pump down line closely and the valve position is not oscillating at all. Example: 44/88

3.11.3.2 Required information for support (in case of unsuccessful tuning procedure): Go to Tools / Create Diagnostic File in Control View resp. Control Performance Analyzer and save file Pressure / flow / gas conditions to be controlled Chamber volume Pumping speed (l/s) and pump type (e.g. turbo pump) System description Problem description Send diagnostic file with and all required information to tuning-support@vat.ch 45/88

3.12 RS232 interface 3.12.1 Schematics This interface allows for remote operation by means of a command set based on the RS232 protocol. In addition there are 2 digital inputs and 2 digital outputs. Digital inputs may be operated either by switches or by voltage sources. Note: Optional analog outputs are available on 615.. -... V -.... and 615.. -... W -.... versions only. Active digital inputs have higher priority than RS232 commands. a) Configuration with switches for digital inputs: Note: Do not connect other pins than indicated in the schematics above! Connector: Use only screws with 4-40UNC thread for fastening the DB-25 connector! 46/88

b) Configuration with voltage source for digital inputs: Note: Do not connect other pins than indicated in the schematics above! Connector: Use only screws with 4-40UNC thread for fastening the DB-25 connector! 47/88

3.12.2 Digital inputs Pin Function Signal type Description Priority 15 CLOSE VALVE 17 OPEN VALVE 23 DIGITAL GROUND 25 DIGITAL COMMON Digital input 1) Digital input 1) Digital ground Digital common This function will close the valve. Valve will be in interlock mode as long as function is activated. After deactivation of function it will remain effective until - OPEN valve digital input is active - converse RS232 control command have been received The function is activated when optocoupler is on in non inverted configuration. The function is activated when optocoupler is off in inverted configuration. Configuration can be done in local operation via service port or in remote operation. This function will open the valve. Valve will be in interlock mode as long as function is activated. After deactivation of function it will remain effective until converse RS232 control command have been received. The function is activated when optocoupler is on in non inverted configuration. The function is activated when optocoupler is off in inverted configuration. Configuration can be done in local operation via service port or in remote operation. Ground for all digital inputs. Ground is used when digital inputs are operated by switches. Connect switches to ground. See also «3.12.2 Schematics» configuration a). Common for all digital inputs. Common is used when digital inputs are driven by voltage sources. Connect + or terminal of source with common (optocoupler inputs are capable of bidirectional operation). See also «3.12.2 Schematics» configuration b). 1 2) 2 2) 1) All digital inputs are digitally filtered. Filter delay is 50ms. This means that digital signals must be applied for at least 50ms to be effective. Refer to «3.12.2 Schematics» for details about input circuit. 2) Highest priority is 1. Functions with lower priorities will not be effective as long as higher priority functions are active. These digital inputs have higher priority than all RS232 commands. RS232 commands will not be accepted while digital inputs are active. 48/88

3.12.3 RS232 Command syntax [function][value][cr][lf] Each element is separated with square brackets for clarity. Square brackets are not part of command syntax. Unless otherwise specified all elements are ASCII characters. There are no spaces between the elements necessary. Commands and values are case sensitive. Data length of value depends on command. Number of characters is specified in the description. Some commands do not require the value element. [CR] is Carriage Return (0D hexadecimal). [LF] is Linefeed 3.12.4 Control commands Control function POSITION CONTROL HOLD CLOSE VALVE OPEN VALVE PRESSURE CONTROL Command Description Set [R:][xxxxxx][CR][LF] [R:][CR][LF] Acknowledgement (within 10ms after reception of command) Get [i:38][cr][lf] [i:38][00xxxxxx][cr][lf] data length 6 characters for writing 8 characters starting with double zero for reading xxxxxx position SETPOINT, value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details Change to POSITION CONTROL mode and transfer of position SETPOINT value resp. reading of position SETPOINT. Note: Reading returns position setpoint only in case pressure control is not selected. Set [H:][CR][LF] [H:][CR][LF] This function stops the valve at the current position. It is effective in PRESSURE CONTROL and POSITION CONTROL. The function can be revoked by a POSITION CONTROL, PRESSURE CONTROL, OPEN VALVE or CLOSE VALVE command. Set [C:][CR][LF] [C:][CR][LF] Valve will close. Set [O:][CR][LF] [O:][CR][LF] Valve will open. Set [S:][0xxxxxxx][CR][LF] [S:][CR][LF] Get [i:38][cr][lf] [i:38][0xxxxxxx][cr][lf] data length 8 characters starting with a zero xxxxxxx pressure SETPOINT, value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details Change to PRESSURE CONTROL mode and transfer of pressure SETPOINT resp. reading of pressure SETPOINT. Note: Reading returns pressure setpoint only in case pressure control is selected, otherwise position setpoint is returned. 49/88

3.12.5 Inquiry commands Inquiry function ASSEMBLY POSITION PRESSURE Command Description Acknowledgement (within 10ms after reception of command) Get [i:76][cr][lf] [i:76][xxxxxxsyyyyyyyabc][cr][lf] data length 17 characters xxxxxx position, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details s sign, 0 for positive pressure readings, - for negative pressure readings yyyyyyy pressure, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details a 0 = local operation, 1 = remote operation, 2 = locked remote operation b 0 = Initialization (Refer to chapter: «Behavior during power up» 1 = synchronization, 2 = POSITION CONTROL, 3 = CLOSED 4 = OPEN, 5 = PRESSURE CONTROL, 6 = HOLD, 7 = LEARN 8 = INTERLOCK (OPEN by digital input) 9 = INTERLOCK (CLOSED by digital input) C = power failure, D = safety mode E = fatal error (read «FATAL ERROR STATUS» for details) c 0 = no warning, 1 = warning present (read «WARNINGS» and «ERROR STATUS» for details) This function returns an assembly consisting of POSITION, PRESSURE and main status information for the valve. Get [A:][CR][LF] [A:][xxxxxx][CR][LF] data length 6 characters xxxxxx position, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details This function returns the current valve position. Note: When motor interlock is active during power up the valve enters the safety mode and is not able to recognize position. In this case position 999 999 is returned. Get [P:][CR][LF] [P:][sxxxxxxx][CR][LF] data length 8 characters s sign, 0 for positive readings, - for negative readings xxxxxxx pressure, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details This function returns the actual pressure. 50/88

Inquiry function SENSOR 1 READING SENSOR 2 READING PRESSURE CONTROL STATUS Command Description Get [i:64][cr][lf] [i:64][sxxxxxxx][cr][lf] Acknowledgement (within 10ms after reception of command) data length 8 characters s sign, 0 for positive readings, - for negative readings xxxxxxx sensor 1 reading, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details This function returns direct reading from sensor 1 input. Get [i:65][cr][lf] [i:65][sxxxxxxx][cr][lf] data length 8 characters s sign, 0 for positive readings, - for negative readings xxxxxxx sensor 2 reading, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details This function returns direct reading from sensor 2 input. Get [i:36][cr][lf] [i:36][abcdefgh][cr][lf] data length 8 characters a 0 = no pressure control (e.g. if position control is selected) 1 = wide range control (PD control) 2 = close up control (PID control) cdefgh reserved, do not use The controller distinguishes 2 control ranges and acts accordingly. DEVICE STATUS Get [i:30][cr][lf] [i:30][abcdefgh][cr][lf] data length 8 characters a 0 = local operation, 1 = remote operation, 2 = locked remote operation b 0 = Initialization (Refer to chapter: «Behavior during power up» 1 = synchronization, 2 = POSITION CONTROL, 3 = CLOSED 4 = OPEN, 5 = PRESSURE CONTROL, 6 = HOLD, 7 = LEARN 8 = INTERLOCK (OPEN by digital input) 9 = INTERLOCK (CLOSED by digital input) C = power failure, D = safety mode E = fatal error (read «FATAL ERROR STATUS» for details) c 0 = Power Failure Option (PFO) disabled 1 = Power Failure Option (PFO) enabled d 0 = no warning, 1 = warning present (read «WARNINGS» and «ERROR STATUS» for details) efg reserved, do not use h 0 = normal operation, 1 = simulation running This function returns status information about the valve. Note: In simulation mode the valve can demonstrate pressure control capability independent of other equipment such as vacuum chamber, flow controller and gauge. Normal operation is not possible when simulation is running. 51/88

Inquiry function WARNINGS SENSOR OFFSET SENSOR 1 OFFSET SENSOR 2 OFFSET Command Description Acknowledgement (within 10ms after reception of command) Get [i:51][cr][lf] [i:51][abcdefgh][cr][lf] data length 8 characters a 0 = no service required 1 = service request, it is indicated when the control unit detects that motor steps are apparently not effective. This may happen when the valve is heavily contaminated or the gate seal is heavily sticking. These lost steps are recognized and will be repeated to attempt target position in the short term. But in the medium term the valve requires cleaning or inspection. b 0 = LEARN data set present, 1 = LEARN data set not present c 0 = power failure battery ready 1 = power failure battery not ready d 0 = compressed air supply ok 1 = compressed air supply not ok efgh reserved, do not use This function returns warning information about the valve. If a warning is present countermeasure should be taken. Use RESET command to delete service request bit. Note: Without LEARN the valve is not able to run pressure control Get [i:62][cr][lf] [i:62][aaaabbbb][cr][lf] data length 8 characters aaaa offset sensor 1 (-140... 0140 = -1.40V... +1.40V) bbbb offset sensor 2 (-140... 0140 = -1.40V... +1.40V) This function returns the sensor offset voltages for both sensors (adjusted by ZERO). Get [i:60][cr][lf] [i:60][xxxxxxxx][cr][lf] data length 8 characters xxxxxxxx offset sensor 1 (-1400000... 01400000 = -1.400000V... +1.400000V) This function returns the sensor 1 offset voltage (adjusted by ZERO). Get [i:61][cr][lf] [i:61][xxxxxxxx][cr][lf] data length: 8 characters xxxxxxxx offset sensor 2 (-1400000... 01400000 = -1.400000V... +1.400000V) This function returns the sensor 2 offset voltage (adjusted by ZERO). 52/88

Inquiry function LEARN STATUS LEARN PRESSURE LIMIT ERROR STATUS FATAL ERROR STATUS Command Acknowledgement (within 10ms after reception of command) Description Get [i:32][cr][lf] [i:32][abcdefgh][cr][lf] data length 8 characters a 0 = LEARN not running, 1 = LEARN running b 0 = LEARN data set present, 1 = LEARN data set not present c 0 = ok 1 = last LEARN interrupted by user (control command) 2 = last LEARN interrupted by control unit (valve open pressure > sensor full scale) d e f g h 0 = ok 1 = valve open pressure > 50% sensor full scale (gasflow too high) 2 = valve open pressure < 0 (sensor offset present) 0 = ok 1 = valve max. throttle pressure < 10% sensor full scale (gasflow too low) 0 = ok 1 = pressure not raising during LEARN (gasflow missing) 0 = ok 1 = sensor unstability during LEARN reserved, do not use This function checks the status of LEARN and indicates if the conditions during LEARN were ok. Get [i:34][cr][lf] [i:34][0xxxxxxx][cr][lf] data length 8 characters starting with a zero xxxxxxx pressure limit for LEARN, return value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details This function returns the pressure limit applied for LEARN. Get [i:52][cr][lf] [i:52][abcdefgh][cr][lf] data length 8 characters a reserved, do not use b 1 = sensor 1 signal converter failure c reserved, do not use d 1 = firmware memory failure efgh reserved, do not use This function returns an error code in case of any malfunction of the device otherwise 0 is returned. Get [i:50][cr][lf] [i:50][abc][cr][lf] data length 3 characters abc error code = 000 (no error) or 020 (E. 20) or 022 (E. 22) or 040 (E. 40) See in chapter «Trouble shooting» for details. This function returns an error code in case of any malfunction of the device. 53/88

Inquiry function THROTTLE CYCLE COUNTER ISOLATION CYCLE COUNTER POWER UP COUNTER Command Acknowledgement (within 10ms after reception of command) Description Get [i:70][cr][lf] [i:70][xxxxxxxxxx][cr][lf] data length 10 characters xxxxxxxxxx number of throttle cycles This function returns the number of throttle cycles. A movement from max. throttle position to open back to max. throttle position counts as one cycle. Partial movements will be added up until equivalent movement is achieved. Get [i:71][cr][lf] [i:71][xxxxxxxxxx][cr][lf] data length 10 characters xxxxxxxxxx number of isolation cycles This function returns the number of isolation cycles. Each closing of the sealing ring counts as one cycle. Get [i:72][cr][lf] [i:72][xxxxxxxxxx][cr][lf] data length 10 characters xxxxxxxxxx number of power ups This function returns the number of control unit power ups. HARDWARE CONFIGURATION FIRMWARE CONFIGURATION IDENTIFICATION Get [i:80][cr][lf] [i:80][abcdefgh][cr][lf] data length 8 characters a 0 = Power Failure Option (PFO) not equipped 1 = Power Failure Option (PFO) equipped b 0 = ±15V sensor power supply (SPS) not equipped 1 = ±15V sensor power supply (SPS) equipped c 2 = RS232 Interface without analog outputs 3 = RS232 Interface with analog outputs d 1 = 1 sensor version, 2 = 2 sensor version efgh reserved, do not use This function returns the hardware configuration of the device. Get [i:82][cr][lf] [i:82][xxxxxxxx][cr][lf] data length 8 characters xxxxxxxx firmware version, e.g. 650P1D00 This function returns firmware version of the device. Get [i:83][cr][lf] [i:83][xxxxxxxxxxxxxxxxxxxx][cr][lf] data length 20 characters xxx...xxx identification code, e.g. 61534-KEAG-0003/0001/, unused digits are filled up with spaces (20 hexadecimal) This function returns an identification code. This code is unique for each valve and allows tracing. 54/88

3.12.6 Setup commands Setup function ACCESS MODE INTERFACE CONFIGURATION VALVE CONFIGURATION Command Description Set [c:01][xx][cr][lf] [c:01][cr][lf] Acknowledgement (within 10ms after reception of command) data length: 2 characters xx 00 = local operation (service port) 01 = remote operation, change to local enabled 02 = locked remote operation, change to local not possible via service port This function selects the access authorization to the valve. To read access mode use inquiry command DEVICE STATUS. Note: Local operation is only possible when either Control View or Control Performance Analyzer software is running. When communication to service port is interrupted the valve will automatically change to remote operation. Set [s:20][abcdefgh][cr][lf] [s:20][cr][lf] Get [i:20][cr][lf] [i:20][abcdefgh][cr][lf] data length 8 characters a baud rate: 0 = 600, 1 = 1200k, 2 = 2400, 3 = 4800, 4 = 9600 5 = 19.2k, 6 = 38.4k, 7 = 57.6k, 8 = 115.2k b parity bit: 0 = even, 1 = odd, 2 = mark, 3 = space, 4 = no c data length: 0 = 7 bit, 1 = 8 bit d number of stop bits: 0 = 1, 1 = 2 e 0 (reserved, do not change) f digital input OPEN VALVE: 0 = not inverted, 1 = inverted, 2 = disabled g digital input CLOSE VALVE: 0 = not inverted, 1 = inverted, 2 = disabled h 0 (reserved, do not change) This function does the RS232 and digital input configuration. Note: Digital outputs are always enabled. Set [s:04][abcdefgh][cr][lf] [s:04][cr][lf] Get [i:04][cr][lf] [i:04][abcdefgh][cr][lf] data length 3 characters a valve position after power up: 0 = closed, 1 = open b valve position after power failure: 0 = closed, 1 = open c 0 (reserved, do not change) d 0 (reserved, do not change) e 0 (reserved, do not change) f 0 (reserved, do not change) g 0 (reserved, do not change) h 0 (reserved, do not change) This function does the valve configuration. 55/88

Setup function SENSOR CONFIGURATION Command Description Set [s:01][abcdefgh][cr][lf] [s:01][cr][lf] Acknowledgement (within 10ms after reception of command) Get [i:01][cr][lf] [i:01][abcdefgh][cr][lf] data length 8 characters a 0 = no sensor 1 = 1 sensor operation (sensor 1 input) 2 = 2 sensor operation with automatic changeover (low range = sensor 2 input, high range = sensor 1 input) 3 = 1 sensor operation (sensor 2 input) 4 = 2 sensor operation with automatic changeover (low range = sensor 1 input, high range = sensor 2 input) Note: Sensor operation modes 2, 3 and 4 are possible with 2 sensor hardware (615.. -... H -.... and 615.. -... W -....) only. Note: For applications where the high range sensor is used for for monitoring purpose only, select sensor operation modes 1 or 3 for pressure control with low range sensor and read high range sensor from «SENSOR 2 READING» resp. «SENSOR 1 READING». b 1 = ZERO enabled, 0 = ZERO disabled cdefgh High range / Low range sensor full scale ratio * 1 000 (1000 100000). In case of a 1 sensor valve use any value within the valid range. This function does the sensor configuration for pressure control. 56/88

0-10V 0-10V Installation, Operating & Maintenance Instructions Setup function RANGE CONFIGURATION Command Description Set [s:21][abcdefgh][cr][lf] [s:21][cr][lf] Acknowledgement (within 10ms after reception of command) Get [i:21][cr][lf] [i:21][abcdefgh][cr][lf] data length 8 characters a range for POSITION: 0 = 0 1 000, 1 = 0 10 000, 2 = 0 100 000 bcdefgh upper value for PRESSURE and SENSOR READING: 1000... 1000000 e.g. 0010000 -> pressure range 0 10 000 This function defines the communication range between the valve and the host computer for POSITION, PRESSURE and SENSOR READING. Note: In case ZERO has been performed, gauge offset for PRESSURE and SENSOR READING is compensated. Note: In case 2 sensor operation for pressure control is selected, PRESSURE covers high range gauge because switchover between sensors is done automatically. SENSOR 1 READING and SENSOR 2 READING always return full scale values according to selected range. High range sensor 1 Torr PRESSURE 0 10 000 SENSOR 1 READING 0 10 000 Low range sensor 100 mtorr 0 mtorr SENSOR 2 READING 0 10 000 Above picture shows a 2 sensor system. In this configuration sensor 2 covers low range (100 mtorr) and sensor 1 covers high range (1 Torr). RANGE CONFIGURATION for PRESSURE resp. SENSOR READING is set to 10 000. Switchover between sensors is done automatically. 57/88

Setup function ZERO PRESSURE ALIGNMENT LEARN DOWNLOAD LEARN DATA UPLOAD LEARN DATA Command Description Set [Z:][CR][LF] [Z:][CR][LF] This command initiates ZERO to compensate for offset of gauge(s). Note: Refer to «ZERO» for correct zero procedure. Set [c:6002][xxxxxxxx][cr][lf] [c:60][cr][lf] Acknowledgement (within 10ms after reception of command) data length: 8 characters xxxxxxxx System base pressure, value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details. Alignment range is equivalent to max. +/-1.4V sensor signal. This command aligns PRESSURE to a certain value. Also SENSOR READING will be aligned accordingly. It might be used instead of ZERO in case base pressure is not low enough. Set [L:][0xxxxxxx][CR][LF] [L:][CR][LF] data length 8 characters starting with a zero xxxxxxx Pressure limit for LEARN, value depends on configuration, refer to «RS232 setup commands, RANGE CONFIGURATION» for details This command starts LEARN. By OPEN VALVE, CLOSE VALVE or POSITION CONTROL commands the routine may be interrupted. Note: Without LEARN the PID controller is not able to perform pressure control. Refer to «LEARN» for correct learn gas flow and procedure. Set [d:][pppdddddddd][cr][lf] [d:][ppp][cr][lf] data length 3 + 8 characters ppp pointer, 000... 103 dddddddd single data set This command downloads the LEARN data sets from the host computer to the valve. There are a total number of 104 data sets. Each data set consists of 8 data bytes and needs to be uploaded separately. Note: Make sure that all 104 data sets will be downloaded. Get [u:][ppp][cr][lf] [u:][pppdddddddd][cr][lf] data length 3 + 8 characters ppp pointer, 000... 103 dddddddd single data set This command uploads the LEARN data sets from the valve up to the host. There are a total number of 104 data sets. Each data set consists of 8 data bytes and needs to be uploaded separately. Note: Make sure that all 104 data sets will be uploaded. 58/88

Setup function PID CONTROLLER CONFIGURATION Command Description Set [s:02][abcdefgh][cr][lf] [s:02][cr][lf] Acknowledgement (within 10ms after reception of command) Get [i:02][cr][lf] [i:02][abcdefgh][cr][lf] data length 8 characters a Controller algorithm: 0 = adaptive downstream (default) 1 = fixed PI downstream control 2 = fixed PI upstream control 3 = soft pump b Gain Factor: 0 = 0.10, 1 = 0.13, 2 = 0.18, 3 = 0.23, 4 = 0.32, 5 = 0.42, 6 = 0.56, 7 = 0.75, 8 = 1.00, 9 = 1.33, A = 1.78, B = 2.37, C = 3.16, D = 4.22, E = 5.62, F = 7.50, G = 0.0001, H = 0.0003, I =0.001, J = 0.003, K = 0.01, L = 0.02, M = 0.05 c Sensor Response Time (s): 0 = 0.00, 1 = 0.02, 2 = 0.04, 3 = 0.06, 4 = 0.08, 5 = 0.10, 6 = 0.15, 7 = 0.20, 8 = 0.25, 9 = 0.30, A = 0.35, B = 0.4, C = 0.50, D = 0.60, E = 0.80, F = 1.00 d Setpoint Ramp Time (s): 0 = 0.0, 1 = 0.5, 2 = 1.0, 3 = 1.5, 4 = 2.0, 5 = 2.5, 6 = 3.0, 7 = 3.5, 8 = 4.0, 9 = 4.5, A = 5.0, B = 5.5, C = 6.0, D = 6.5, E = 7.0, F = 7.5, G = 8.0, H = 8.5, I = 9.0, J =9.5, K = 10.0 ef P-Gain: 00 = 0.0010, 01 = 0.0013, 02 = 0.0018, 03 = 0.0024, 04 = 0.0032, 05 = 0.0042 06 = 0.0056, 07 = 0.0075, 08 = 0.010, 09 = 0.013, 10 = 0.018, 11 = 0.024 12 = 0.032, 13 = 0.042, 14 = 0.056, 15 = 0.075, 16 = 0.10, 17 = 0.13 18 = 0.18, 19 = 0.24, 20 = 0.32, 21 = 0.42, 22 = 0.56, 23 = 0.75 24 = 1.0, 25 = 1.3, 26 = 1.8, 27 = 2.4, 28 = 3.2, 29 = 4.2 30 = 5.6, 31 = 7.5, 32 = 10 gh I-Gain: 00 = 0.0010, 01 = 0.0013, 02 = 0.0018, 03 = 0.0024, 04 = 0.0032, 05 = 0.0042 06 = 0.0056, 07 = 0.0075, 08 = 0.010, 09 = 0.013, 10 = 0.018, 11 = 0.024 12 = 0.032, 13 = 0.042, 14 = 0.056, 15 = 0.075, 16 = 0.10, 17 = 0.13 18 = 0.18, 19 = 0.24, 20 = 0.32, 21 = 0.42, 22 = 0.56, 23 = 0.75 24 = 1.0, 25 = 1.3, 26 = 1.8, 27 = 2.4, 28 = 3.2, 29 = 4.2 30 = 5.6, 31 = 7.5, 32 = 10, 33 = 13, 34 = 18, 35 = 24 36 = 32, 37 = 42, 38 = 56, 39 = 75, 40 = 100 Note: In adaptive downstream mode ef (set 00) and gh (set 00) are not used. In fixed PI modes b (set 0) and c (set 0) are not used. In pump down mode bc (set 00) and gh (set 00) are not used. This command selects gain factor, sensor response time and setpoint ramp for the PID controller. Note: Refer to «Tuning of control performance» for details. 59/88

Setup function VALVE SPEED RESET Command Description Set [V:][00xxxx][CR][LF] [V:][CR][LF] Acknowledgement (within 10ms after reception of command) Get [i:68][cr][lf] [i:68][0000xxxx][cr][lf] data length 6 characters starting with double zero for writing 8 characters starting with quadruple zero for reading xxxx valve speed, 1... 1000 (1 = min. speed, 1000 = max. speed) This command allows changing the actuating speed of the valve plate. Speed selection is effective for pressure control and position control. Open valve and close valve are always done with max. speed. Note: Refer to «Valve speed adjustment» for details. Set [c:82][xx][cr][lf] [c:82][cr][lf] data length 2 characters xx 00 = reset service request bit from WARNINGS 01 = reset FATAL ERROR (restart control unit) This function resets warnings and errors. 60/88

3.12.7 Error messages Description Protocol Parity error Framing error (data length, number of stop bits) Input buffer overflow (to many characters) Error message [E:][000001][CR][LF] [E:][000003][CR][LF] [E:][000002][CR][LF] <CR> or <LF> missing Commands [E:][000010][CR][LF] : missing [E:][000011][CR][LF] Unknown command Invalid value Value out of range Invalid number of characters (between : and [CR][LF]) [E:][000020][CR][LF] [E:][000021][CR][LF] [E:][000022][CR][LF] [E:][000023][CR][LF] [E:][000030][CR][LF] [E:][000012][CR][LF] ZERO disabled Setup [E:][000060][CR][LF] Device Status Command not accepted due to local operation Command not accepted due to synchronization, CLOSED or OPEN by digital input, safety mode or fatal error [E:][000080][CR][LF] [E:][000082][CR][LF] Hardware Command not applicable for hardware configuration [E:][000041][CR][LF] 61/88

4 Trouble shooting Failure Check Action No dots lighted on display 24 V power supply ok? Connect valve to power supply according to «Electrical connection» and make sure that power supply is working. Remote operation does not work Display shows «E 20»and position is 009999 (fatal error - limit stop of valve unit not detected) Display shows «E 21»and position is 009999 (fatal error - rotation angle of valve plate limited during power up) - Local operation via service port active - Safety mode active, check for D on display? Clamp coupling screw not fastened? - Valve plate centric adjusted? - Valve unit heavy contaminated? - Valve plate mechanically obstructed? - Switch to remote operation. - Provide power to motor to allow for operation. - Refer to «Electrical connection» for details. Tighten clamp coupling screw according to «Maintenance procedure». RESET or restart of valve is necessary. - Adjust valve plate according to «Maintenance procedure». - Clean valve unit according to «Maintenance procedure». - Resolve obstruction. - Reset control unit. Cycle power (OFFàON) or - Send reset command: local via service port with CV/CPA/Service Box2 Display shows «E 22»and position is 009999 (fatal error - rotation angle of valve plate limited during operation) - Valve unit heavy contaminated? - Valve plate mechanically obstructed? - Clean valve unit according to «Maintenance procedure». - Resolve obstruction. - Reset control unit. Cycle power (OFFàON) or - Send reset command: local via service port with CV/CPA/Service Box2 Display shows «E 40»and position is 009999 (fatal error - motor driver failure detected) Display shows «D 0» Motor Interlock is open Motor power supplied? Replace control and actuating unit according to «Maintenance procedure». - Provide power to motor to allow for operation. - Refer to «Electrical connection» for details. 62/88

Failure Check Action Display shows «SR» (Service Request) Valve unit heavy contaminated? Or gate seal is sticking. - Clean valve unit according to «Maintenance procedures». - Reset control unit. Cycle power (OFFàON) or - Send reset command: local via service port with CV/CPA/Service Box2 CLOSE VALVE does not work OPEN VALVE does not work Display shows «M C» Maintenance mode active Display shows «M100» Maintenance mode active Pressure reading is wrong or pressure reading is negative - Safety mode active, check for D on display? - Maintenance mode active - Safety mode active, check for D on display? - Maintenance mode active - Sensor(s) connected? - 2 sensor version present at valve controller? - Provide power to motor to allow for operation. - Refer to «Electrical connection» for details. - Refer to Display shows «M C» in this table - Provide power to motor to allow for operation. Refer to «Electrical connection» for details. - Refer to Display shows «M100» in this table Pin 14 of service connector is connected to ground. Plate will close. Further movement of plate is blocked. Note: Priority of pin 14 is higher than pin 13. If pin 14 is connected to ground after pin 13 the valve will close. Ground of service connector is at pin 4 and 8. Pin 13 of service connector is connected to ground. Plate will open. Further movement of plate is blocked. - Refer to «Electrical connection». - Check valve version on page 1. Verify configuration. Refer to «Setup procedure». ZERO does not work - ZERO done? - Does sensor power supply provide enough power for sensor(s)? - Valve in open position, check for O on display? - ZERO disabled? - Perform ZERO when base pressure is reached. Refer to «ZERO» for details. - Verify sensor supply voltage. - OPEN VALVE and bring chamber to base pressure before performing ZERO. - Enable ZERO. Refer to «Valve and sensor configuration» for details. 63/88

Failure Check Action Pressure is not 0 after ZERO - Sensor voltage shifting? - System pumped to base pressure? - Sensor offset voltage exceeds ±1.4V - Wait until sensor does not shift any more before performing ZERO. - OPEN VALVE and bring chamber to base pressure before performing ZERO. - Replace pressure gauge. PRESSURE CONTROL does not work - Safety mode active, check for D on display? - PRESSURE CONTROL selected, check for P on display? - Provide power to motor to allow for operation. Refer to «Electrical connection» for details. - Select PRESSURE CONTROL mode. Refer to «Pressure control» for details. PRESSURE CONTROL not optimal - LEARN done? - Setup done completely? - LEARN done? - ZERO performed before LEARN? - LEARN interrupted? - Was gas flow stable during LEARN? - Tuning done? - Is sensor range suited for application? - Perform LEARN. Refer to «Setup procedure» for details. - Perform «Setup procedure» completely. - Perform LEARN. Refer to «LEARN» for details. - Perform ZERO then repeat LEARN. Refer to «Setup procedure» for details. - Repeat LEARN. Refer to «LEARN» for details. - Repeat LEARN with stable gas flow. Refer to «LEARN» for details. - Tune valve for application. Refer to «Tuning of control performance» for details. - Use a sensor with suitable range (controlled pressure should be >3% and < 98% of sensor full scale). - Noise on sensor signal? - Make sure a shielded sensor cable is used. If you need any further information, please contact one of our service centers. You can find the addresses on our website: http://www.vat.ch 64/88

5 Maintenance & repairs Under clean operating conditions, the valve does not require any maintenance during the specified cycle life. Contamination from the process may influence the function and requires more frequent maintenance. Before carrying out any maintenance or repairs, please contact VAT. It has to be individually decided whether the maintenance/repair can be performed by the customer or has to be carried out by VAT. The fabrication number on the valve made in Switzerland Fabrication No.:............ /.... A...... Fabrication number has always to be specified. All supplies (e. g. compressed air, electrical power) must be disconnected for removal/installation of the valve from/into the system. Even with disconnected supply, loaded springs and/or air cushions in cylinders can be potential hazards. Keep fingers and objects away from the valve opening! Products returned to VAT must be free of harmful substances such as e.g. toxical, caustic or micro-biological ones. If products are radioactively contaminated, fill in the VAT form «Contamination and Radiation Report» and send it with the product. The form is available at VAT. The maximum values indicated in the form must not be exceeded. 5.1 Maintenance procedures Keep fingers out of the valve during maintenance work. Use cleanroom gloves during maintenance work. Two preventive maintenance procedures are defined for this valve. This are: Replacement of plate o-ring (without controller and valve separation). Refer to chapter: «Replacement of plate o-ring». Replacement of shaft feedthrough seals and plate o-ring. Refer to chapter: «Replacement of shaft feedtrough seals and plate o-ring». 65/88

Required frequency of cleaning and replacement of seals is depending on process conditions. A critical factor influencing the maintenance period is the lifetime of the vacuum grease, being limited under increased temperature. In this case grease will separate to PTFE and oil. The oil may flow and contaminate the valve parts. VAT can give the following recommendations for preventive maintenance with FKM seals under clean and unheated condition: Seals Plate o-ring Shaft feedthrough seal Maintenance recommendation Exchange after 250 000 isolation cycles or 2 000 000 control cycles Exchange after 2 000 000 control cycles These values do not include any impact of the process. Therefore preventive maintenance schedule has finally to be checked for the actual process conditions. 66/88

Replace o-rings at plate Remove the plate Preparation Installation, Operating & Maintenance Instructions 5.1.1 Replacement of plate o-ring Description 1. Vent vacuum system on both sides of the valve. 2. Move the plate to position 50% open. 3. Disconnect electrical POWER connector at valve and remove valve from vacuum system. Pay attention: Electrical power is needed for plate adjustment in step 19. Take care to the sealing surface! Required tool 4. Unfasten and remove the 2 fastening screws Allen Wrench 3mm 5. Push the plate a little down 6. Remove the plate on the other side of shaft 7. Remove the plate o-ring 8. Replace the plate if necessary Soft tool 67/88

Replace o-rings at plate Installation, Operating & Maintenance Instructions Description Required tool 9. Clean the plate and o-ring groove Cleanroom wipes soaked with isopropyl alcohol 10. Lubricate the new o-ring with 0.0125 ml vacuum grease For new o-ring refer to chapter spare parts. Vacuum grease 11. Pay attention that grease is distributed constantly over the whole circumference 12. Place the new o-ring at o-ring groove at one side 13. Move the o-ring in the o-ring groove on the other side 68/88

Assemble and adjust plate Assemble and adjust plate Replace o-rings at plate Installation, Operating & Maintenance Instructions Description Required tool 14. Push in the o-ring equally around the plate into o-ring groove 15. Plate with equal mounted o-ring 16. Place the plate on the shaft in the valve body 17. Tighten plate screws to block. 18. Loosen the plate screws a quarter turn counter clockwise. Fact: Plate is now movable on shaft and ready to sliding into valve seat. Proceed with step 19. Allen Wrench 3mm 19. Connect electrical POWER connection at valve Attention! Valve does synchronize automatically and moves the plate into sit of valve body (close position). 20. Disconnect electrical POWER connection Attention! Keep fingers out of the valve opening during plate movement! 69/88

Remove control and actuating unit from valve unit Preparation Install valve into system Installation, Operating & Maintenance Instructions Description Required tool 21. Fasten the plate screws with 2.5 Nm 22. If some grease on plate surface or valve body clean it with cleanroom wipes. Allen torque wrench 3mm Cleanroom wipes 23. Reinstall valve into vacuum system according to chapter «Installation» of valve manual 5.1.2 Replacement of shaft feedtrough seals and plate o-ring Description 1. Vent vacuum system on both sides of the valve. 2. Move the plate to position 50% open. 3. Disconnect electrical POWER connector at valve and remove valve from vacuum system. Pay attention: Take care to the sealing surface! Required tool 4. Loosen clamp coupling screw Allen Wrench 2mm 70/88

Remove plate Installation, Operating & Maintenance Instructions Description Required tool 5. Unfasten and remove the 3 fastening screws Allen Wrench 3mm 6. Remove control and actuating unit from mechanical valve unit Pay attention: If clamp coupling is separated, assemble them at control and actuating unit. 7. Unfasten and remove the 2 fastening screws Allen Wrench 3mm 71/88

Replace o-rings from plate Installation, Operating & Maintenance Instructions Description Required tool 8. Push the plate a little down 9. Remove the plate on the other side of shaft 10. Remove the plate o-ring 11. Replace the plate if necessary Soft tool 12. Clean the plate and o-ring groove Cleanroom wipes soaked with isopropyl alcohol 13. Lubricate the new o-ring with 0.0125 ml vacuum grease For new o-ring refer to chapter spare parts. Vacuum grease 72/88

Replace o-rings from plate Installation, Operating & Maintenance Instructions Description Required tool 14. Pay attention that grease is distributed constantly over the whole circumference. 15. Place the new o-ring at o-ring groove at one side 16. Move the o-ring in the o-ring groove on the other side 17. Push in the o-ring around the plate into o- ring groove equally 18. Plate with equal mounted o-ring 19. Deposit the plate on a clean surface 73/88

Seal removal and valve cleaning Seal removal and valve cleaning Installation, Operating & Maintenance Instructions Description Required tool 20. Unfasten both screws complete Allen Wrench 3mm 21. Remove mechanical unit from valve body 22. Clean shaft Clean room wipes soaked with isopropyl alcohol 23. Remove both o-rings Soft tool 74/88

Lubricate and install seals Installation, Operating & Maintenance Instructions Description Required tool 24. Clean shaft feed through Cleanroom wipes soaked with isopropyl alcohol 25. Clean the valve body Cleanroom wipes soaked with isopropyl alcohol 26. Lubricate seal contact surface of valve body with a slight film of 0.0125 ml vacuum grease Vacuum grease 27. Lubricate seal contact surface of shaft with a slight film of 0.0125 ml vacuum grease Vacuum grease 28. Lubricate each o-ring with a slight film of 0.0125 ml vacuum grease. Pay attention: If required use new o-rings Vacuum grease 75/88

Assemble and adjust plate Mount mechanical unit to valve body Installation, Operating & Maintenance Instructions Description Required tool 29. Slide both o-rings onto shaft till the end 30. Deposit 0.05 ml vacuum grease between the o-rings 31. Clean shaft from vacuum grease Vacuum grease Cleanroom wipes 32. Install mechanical unit into valve body, see picture 33. Fasten and tighten the 2 mounting screws with 2.5 Nm Allen torque wrench 3mm 34. Place the plate on the shaft in the valve body 76/88

Mount control and actuating unit to valve unit Installation, Operating & Maintenance Instructions Description Required tool 35. Tighten plate screws to block. 36. Loosen the plate screws a quarter turn counter clockwise so that plate is still movable at shaft. Allen Wrench 3mm A 37. Press the plate (A) into valve seat. 38. Press the lever by hand to axis (C). Make sure that lever (B) is engaged at axis (C) completely. (mechanical close position). C B 39. Fasten the plate screws with 2.5 Nm 40. If some grease on plate surface or valve body clean it with lint and dust free cloth Allen torque wrench 3mm Cleanroom wipes 41. Assemble valve unit with control and actuating. Push axis (C) into clamp coupling C 77/88

Install valve into system Installation, Operating & Maintenance Instructions Description Required tool 42. Tighten mounting screws adequately Allen Wrench 3mm 43. Tighten clamp coupling screw with 1.1 Nm Allen torque wrench 2mm 44. Reinstall valve into vacuum system according to chapter «Installation». 78/88

Durability [months] Installation, Operating & Maintenance Instructions 5.2 Option board The option board may or may not be equipped in your valve depending on the order. Refer to page 1 of this manual to check valve version. This board includes the optional modules for the valve which are: ±15VDC sensor power supply (SPS) Power failure option (PFO) It is available in 3 versions. These are: SPS module only PFO module only SPS and PFO module The modules may be retrofitted or replaced easily. The battery lifetime of the PFO module depends on the ambient temperature (see below). To assure PFO function the option board must be replaced after battery life has expired. For ordering number of the modules refer to «Spare parts and accessories». 5.2.1 Durability of power fail battery The curves in the graph show the estimated life of UltraCap PFO in the worst condition (max. sensor load = 1 A, valve heating temperature = 150 degree C). If the SPS is not fully loaded (< 1 A) or heating temperature of valve body is lower than 150 degree C, the corresponding life time curve will be somewhere in between the upper and the lower curve. Therefore please determine the equivalent maintenance period for replacing the UltraCap battery (Option board). 120 100 80 60 40 w/o SPS module, valve unheated with SPS module, valve unheated w/o SPS module, valve heated with SPS module, valve heated 20 0 20 25 30 35 40 45 50 Ambient temperature [ C] Note: This graph shows estimated life of UltraCap PFO for reference and not as guaranteed value. 79/88

5.2.2 Retrofit / replacement procedure ESD Precaution! All work on the control and actuating unit has to be done under ESD protected environment to prevent electronic components from damage! Top view on control and actuating unit with panel removed: Master board Motor driver board Option board Interface board Note: All boards have a fixed position into control and actuating unit. It is not possible to fit a board in other position as shown in picture above. Do not try out other positions, that may be destroy the socket of boards! 80/88

Description Required tool 1. Remove the panel screws. Pozidriv screw driver size 1 2. Remove female screw locks from POWER, SENSOR and INTERFACE connectors. Open end wrench 4.5 mm 3. Lift the panel carefully. 4. Disconnect fan cable from board. 81/88

Description Required tool 5. Remove or replace master board. 6. Remove or replace interface board. 7. Remove or replace option board. 8. Insert master board and interface board in reverse order as disassembled at correct positions (see steps 6 5). 9. Reconnect fan cable (see step 4). 10. Place the panel and tighten screws with 1.1 Nm (see steps 3 2). 11. Tighten female screw locks from POWER, SENSOR and INTERFACE connectors with 1.1 Nm (see step 1). Pozidriv screw driver size 1 Open end wrench 4.5 mm 82/88