Dialflow Regulator. Instructions for Use

Similar documents
Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use

Pressure Regulators. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use

Pressure Regulator. Instructions for Use

Oxygen Demand Valve. Instructions for Use

Analgesic Demand Valve. Instructions for Use

Domiciliary Oxygen Switch. Instructions for Use

Firesafe Flowmeter. Instructions for Use

Service Manual. Regulators for Medical Gases

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

Service Manual. Dial Flow Regulators

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

BRITISH COMPRESSED GASES ASSOCIATION Registered office: 4a Mallard Way, Pride Park, Derby, UK. DE24 8GX Company Number: 71798, England

Medical oxygen. LIV IQ Digital therapy device.

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

Rev. B. Operating Instructions

BOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare

Peninsula Dental Social Enterprise (PDSE)

Analgesic and Oxygen Demand Valve. Service Manual

SERIES-O REGULATOR. Instructions for use

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

Medical Gas Data Sheet (MGDS) Medical air

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

MANUAL GAS MANIFOLDS

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

User manual. ZX trolley.

- GAS REGULATOR - COMPACT STYLE

U S E R M A N U A L. (CGA 870 Connection)

General Safety Information

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

Presence PR, PRC & PRE

AUTOMATIC GAS MANIFOLDS

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

Home Oxygen Products. > Innovation > Quality > Safety

Medical Gas Data Sheet (MGDS) Medical air. BOC: Living healthcare

Rejuvenation Instructions

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

SAFETY PRINCIPLES FOR PRESSURE REGULATORS FOR MEDICAL OXYGEN CYLINDERS

APC Pressure Reducer with Sensor

Home Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

Plano Independent School District Health Service Oxygen in Use Administrative Guideline

Equinox II NITROUS OXIDE / OXYGEN ADMINISTRATION EQUIPMENT 01EQ7000E

Welcome to the Specialized Medical Services Respiratory training webinar series!

User Instruction Manual

BROCKTON AREA MULTI-SERVICES, INC. MEDICAL PROCEDURE GUIDE. Date(s) Reviewed/Revised:

Synthetic medical air.

AUTOVENT 4000 VENTILATOR

Safe Work Instructions For Oxygen Acetylene Equipment

TactAir PCP Fill Kit/Adapter. Operation Manual

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

U S E R M A N U A L CAUTION. Models: 19MFA Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

Technical Specification

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

Lung Function medical gas mixture 1.

Operating Instructions Part No

Paradise Pump Series Operator s Manual

Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet.

Medical Gas Data Sheet (MGDS) Medical helium. BOC: Living healthcare

PRESSURE REDUCTION FLOWMETERING, QUICK COUPLING PROBES AND ACCESSORIES

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to:

USER MANUAL. AIR-OXYGEN BLENDER (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS

USER MANUAL Series (CGA 870 Connections) SAVE THESE INSTRUCTIONS

Model PDT Dewpoint Transmitter

PRESSURE REGULATORS AND FLOWMETERS, ALL MEDICAL GASES

Safety and operating instructions

Welker Constant Pressure Cylinders With Welker Solid Indicator, Syringe T-Handle, or Vortex Mixer

User Operation Manual and Safety Instructions

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

Lung Function medical gas mixture 4.

Installation, Operation, and Maintenance Manual

Model ASSEMBLY and OPERATING INSTRUCTIONS

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

stormrider Instructions for Use Version 08/2017

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

I. Subject: Compressed gas cylinders, regulators, and portable liquid systems.

PRODUCT INFORMATION AND INSTRUCTIONS OM-810

TEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125

OXY FUEL SYSTEM SET UP & SHUT DOWN PROCEDURE CHECKLIST ACETYLENE FUEL GAS

80 Litre Suction Oil Drainer

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

Operating Instructions Part No

Pressure Regulators. Operating Instructions. Instrumentation

Operating Manual. R280 Pressure regulator made of brass. 1. Intended Use

REGULATOR OWNER S GUIDE

CS150 CAP STAPLER OWNER S MANUAL

Tri-Safem Model I1 Regulator. INSTRUCTIONS and PARTS

Transcription:

Dialflow Regulator Instructions for Use 702-0030.11 May 2014

1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user or others Indicates a potentially hazardous situation which, if not avoided, could result in damage to the equipment or property Attention, consult accompanying documents Use no oil 0120 CE Marked to Medical Device Directive 93/42/EEC as amended by directive 2007/47/CE No smoking Service due date 1304X12345 YYMM Date of manufacture identification 2. Warnings and Cautions 2.1. Warnings! Read through this entire instruction manual before using or showing others how to use a Dialflow Regulator. As with all medical equipment, attempting to use this device without a thorough understanding of its operation may result in patient or user injury. Medical gases are, or should be considered, drugs and may be used in critical treatments. Medical gases should only be used for medical purposes on the authority of a physician or authorised clinician and then strictly in accordance with their instructions. Oxygen is not flammable; however an oxygen enriched atmosphere will drastically increase the rate and severity of combustion. Oil and/or grease in the presence of an oxygen enriched atmosphere will become highly combustible. Oxygen must never be allowed to come into contact with oil, grease or other hydrocarbon or petrochemical based substances. Do not use oil or grease on Dialflow Regulators. Never administer oxygen while smoking or when near an open flame. Oxygen therapy may be a critical treatment. A regulated flowmeter should be used in strict accordance with the prescription and instructions of a physician. The effectiveness of supplemental oxygen therapy can only be determined by continuous monitoring of blood oxygen levels. It is essential that PaO 2 or SpO 2 monitoring is carried out when supplemental oxygen is prescribed.

The use of the Dialflow Regulators for gases other than those they are designated is expressly prohibited. Never permit compressed medical gases to enter a Dialflow Regulator suddenly. Always open the cylinder valve slowly. Never use medical gases from a cylinder without reducing the pressure through a suitable regulator intended for that gas. Always close the cylinder valve when the Dialflow Regulator is not in use. Never use oxygen as a pressure medium to purge obstructed pipelines or equipment, operate pneumatic tools or to build up any pressure in tanks. Never install a pin index Dialflow Regulator with more than one yoke seal between the valve and the regulator. Before attaching the regulator, verify that the post valve is not already fitted with a yoke seal. Ensure that you have the correct regulator for the type of cylinder you are intending to use. Never attempt to fit a regulator to an incompatible cylinder. Ensure that the threaded fittings or indexing pins on the regulator are correctly mated and tightened before opening the cylinder valve. Do not stand in front of a regulator outlet when opening the cylinder valve. Before removing a regulator from a cylinder, fully close the cylinder valve and release all gas from the regulator. Secure cylinders to a wall, stand or cart. Do not use or store oxygen equipment near excessive heat (>50 C or 125 F) or open flame. Only appropriately trained personnel working in controlled conditions must carry out disassembly, assembly and testing of regulators and flowmeters. Do not submerge the device in any fluid. Ensure that no fluid is allowed to enter the inlet valve or the vent holes. Dialflow Regulators are not MRI compatible. Flowmeters may be affected if the cylinder pressure is other than stated in the specifications. The threaded bull nose cylinder connection for medical oxygen and medical air may be the same, ensure that you have the correct cylinder for the application. This regulator is only intended for use with oxygen. Do not obstruct the vent holes or interfere with the relief valve in any way. These are for venting gas in the event that the relief valve is activated. This regulator is designed for use with cylinder pressures up to 20,000 kpa (3,000 psi/ 200 bar). Do not attempt to connect BPR Medical Ltd regulators to cylinders having fill pressures in excess of this value. Ensure that the device is not set between flow settings. If set between flow settings failure to provide therapy may occur. Display of flow value does not indicate flow is occurring. 2.2. Cautions! Device performance may be affected if it is stored or transported in temperature outside of the range -20 C to +60 C (-4 F to +140 F). The Dialflow Regulator is not suitable for autoclaving.

3. Functional Description Flow Selection Dial Flow Indication Window Outlet Barb Vent Cylinder Connection Cylinder Contents/ Pressure Gauge 3.1. Intended Use The Dialflow Regulator is intended to deliver medical gas at one of eleven pre-set rates directly from a high pressure cylinder. Flow is indicated in litres per minute (l/min) and is visible through a window adjacent to the flow selection dial. Medical gas is or should be considered a drug and should only be used for medical purposes on the authority of a physician and then strictly in accordance with their instructions. Always refer to the product labelling. 3.2. Technical Description The Dialflow Regulator reduces the high pressure of the cylinder to a safe, more manageable pressure for application to patients. There are many types of cylinder connection used and these may be different from one country to another, ensure that this regulator is the correct connector for the cylinder to be used. Once connected to a cylinder, and with the cylinder valve opened, the user may select the flow rate prescribed by rotating the gas flow selection dial until the correct flow rate is seen in the viewing window. The rate at which gas flows from the regulator is controlled by a hole of known size and accuracy. Each flow rate has a different size hole. The output connector allows the user to connect an oxygen facemask, cannula or other application device as appropriate. 4. Operating Instructions 4.1. Preparation and Connection Check that the cylinder type and regulator inlet connector are compatible. Check the presence and condition of the input connector seal, which for pin index regulators is a washer type seal (Bodoc) and for bullnose regulators an O-ring. Fit the regulator to the cylinder ensuring that it is properly mated and the connectors are done up hand tight.

Turn the flow selection dial fully clockwise and check that the indication in the flow meter window is zero. Open the cylinder valve slowly with a suitable cylinder key. Check the regulator cylinder contents gauge to verify that there is sufficient gas available. Connect a sufficient length of tubing (not supplied) to the outlet barb. 4.2. Testing Prior to Use Turn the flow selection dial to its highest setting and check that gas flow can be felt at the patient end of the supply tubing. If no medical gas flow is sensed refer to the troubleshooting guide in section 7 of this manual. Turn off the gas flow by turning the flow selection dial fully clockwise. 4.3. Operation Connect the free end of the tubing to the patient or patient enclosure using the appropriate connector (not supplied). Determine the appropriate flow rate in litres per minute (l/min) required for the patient. Turn the flow selection dial until the required rate is visible in the flow indication window; the dial clicks into place and is obvious to the touch. Continue to monitor the patient as advised by the clinician. Check the regulator contents gauge regularly. 4.4. After Use Upon completion of the therapy, turn the flow selection dial to zero, close the cylinder valve and disconnect the outlet hose from the regulator and the patient. To remove the regulator from the cylinder, first ensure the cylinder valve is closed. Bleed off any residual gas pressure in the regulator by turning the flow selection dial to any flow setting until flow ceases. Remove the regulator from the cylinder. 5. Maintenance 5.1. Interim Inspection All Dialflow Regulators should be inspected and tested on an annual basis to ensure correct performance. 5.1.1. Cleaning Wipe down the outside of the device with an alcohol wipe. Do not allow the ingress of water or other solution into the device. Check the exterior condition of the device. Pay particular attention to the input connector seals, which should be replaced if damaged or missing. Check that the holes in the side of the device, which are designed to vent gas in the event of relief valve activation, are not obstructed or have otherwise been tampered with. 5.1.2. Internal Leak Test Connect the regulator to a cylinder in accordance with this Instructions for Use. Set the flow selection dial to zero and open the cylinder valve. Connect a hose to the outlet barb and immerse the other end of the hose into water. Check for leak through (internal leak). Remove the hose.

5.1.3. External Leak Test Attach a fresh, dry hose to the outlet barb and connect the other end of the hose to a calibrated flowmeter in the correct range for the device under test. 5.1.4. Flow Test Verify flow rates at all flow settings against those given in the device specification. 5.2. Service It is recommended ranges A and E be serviced every 2 years, ranges C and D every 4 years to ensure the device continues to perform in accordance to its specification. All Dialflow Regulators have a Service Due date on their label, indicating when the next service is due. Warning! Servicing must be carried out by a suitably qualified person working in a controlled environment. Full details of the recommended servicing requirements can be found in the Service Manual. The Service Manual can be obtained from your local BPR Medical distributor, details of which can be found at www.bprmedical.com. Complete the service in accordance with the instructions given in the Service Manual or return the device to a recognised BPR Medical Service Centre on or before the date shown. 6. Specification Specification Flow Accuracy Input Pressure Range Value ±10 % of setting at 1 l/min and above, ±20 % of setting below 1 l/min 1,000 to 20,000 kpa (10 to 200 bar) Range A: 0, 0.02, 0.03, 0.05, 0.08, 0.12, 0.2, 0.3, 0.5, 0.75, 1.0, 3.0 Flow Settings (l/min) Range C: 0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.5 Range D: 0, 0.5, 1.0, 1.5, 2.0, 3.0, 4.0, 5.0, 6.0, 8.0, 10.0, 15.0 Range E: 0, 0.01, 0.02, 0.03, 0.04, 0.05, 0.06, 0.07, 0.08, 0.09, 0.1, 1.0 Operating, Storage and Transport Temperature Range -20 C to 60 C (-4 F to +140 F) Effects on Accuracy Varying Inlet Pressure: Varying Outlet Pressure: Varying Temperature: Typically less than 15 % change across the input pressure range: Minimum: 1,000 kpa Maximum: 20,000 kpa Less than 1 % of reading up to 5 kpa (50 cmh 2 O) back pressure Typically 0.2 % of reading per degree Celsius Environmental Regulatory Transport/Storage/Operation: -20 C to +60 C (-4 F to +140 F) Humidity: 0-100 % RH non-condensing CE: Medical Device Directive 93/42/EEC - Class llb

Applied Standards BS EN ISO 10524-1 BS EN ISO 14971 BS EN ISO 15001 BS EN ISO 15002 BS EN ISO 15223-1 BS EN 13544-2+A1 BS 5682 FDA: Class 1 Health Canada: Class 2 - Licence No: 83602 Pressure regulators for use with medical gases. Pressure regulators and pressure regulators with flow-metering devices Medical devices. Application of risk management to medical devices Anaesthetic and respiratory equipment. Compatibility with oxygen Flow-metering devices for connection to terminal units of medical gas pipeline systems Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. General requirements Respiratory therapy equipment. Tubing and connectors Specification for probes (quick connectors) for use with medical gas pipeline systems 7. Troubleshooting Fault Possible Cause Solution No gas flow Gas cylinder turned off Dialflow Regulator not connected properly Gas cylinder empty Filter blocked Flow selection dial set to 0 Check gas supply Check gas specific probe is correctly connected Replace gas cylinder Service or repair required Select a positive flow rate Internal/external leak Seal failure Service or repair is required Insufficient gas flow Filter partially occluded Supply pressure too low and/or gas cylinder nearly empty Service is required Check gas supply and/or replace gas cylinder 8. Parts and Spares List Part Number Description 818-0001 Dialflow Regulator - O 2 Range A - UK Bullnose/Barb 818-0002 Dialflow Regulator - O 2 Range A - Pin Index/Barb 818-0005 Dialflow Regulator - O 2 Range C - Pin Index/Barb 818-0006 Dialflow Regulator - O 2 Range C - UK Bullnose/Barb 818-0009 Dialflow Regulator - O 2 Range D - Pin Index/Barb 818-0010 Dialflow Regulator - O 2 Range D - UK Bullnose/Barb 818-0017 Dialflow Regulator - O 2 Range D - Pin Index/Barb/Aux Schrader

818-0018 Dialflow Regulator - O 2 Range D - UK Bullnose/Barb/Aux Schrader 818-0024 Dialflow Regulator - Air Range D - UK Bullnose/Barb 818-0025 Dialflow Regulator - Air Range D - UK Bullnose/Barb 818-0028 Dialflow Regulator - Air Range D - Pin Index/Barb/Aux Schrader 818-0030 Dialflow Regulator - O 2 Range E - UK Bullnose/Barb 818-0031 Dialflow Regulator - O 2 Range E - Pin Index/Barb 818-0037 Dialflow Regulator - Air Range D - UK Bullnose/Barb/Aux Schrader Spare Parts and Servicing 303-0030 Cylinder Contents Gauge 303-0060 Cylinder Contents Gauge Boot 212-0012 Pin Index Yoke Seal 212-0048 Benelux Bullnose O-ring 212-0094 UK Bullnose O-ring 610-0056 Dialflow Regulator Service Kits (Pack of 10) 610-0057 Bodok Bonded Seal (Pack of 50) 610-0058 Bullnose O-ring (Pack of 50) 9. Distributor Details United Kingdom Canada BOC Healthcare BOMImed Inc. Customer Service Centre 96 Terracon Place Priestley Road Winnipeg Worsley MB R2J 4G7 Manchester Canada M28 2UT T: +1 204.633.2664 T: +44 (0)8456 447 764 F: +1 204.632.1064 F: +44 (0)8450 870 062 www.bomimed.com www.bochealthcare.co.uk BPR Medical Ltd License No: 86841 BPR Medical Limited T: +44 (0)1623 628 281 22 Hamilton Way, Mansfield Nottinghamshire NG18 5BU, United Kingdom F: +44 (0)1623 628 289 cs@bprmedical.com www.bprmedical.com