UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS

Similar documents
Justification and use of hand-held X-ray equipment

[Type here] MANAGEMENT OF X-RAY GENERATING EQUIPMENT

HEALTH AND SAFETY SERVICES

Guidance on the Ionising Radiations Regulations 1999 (IRR99)

Local Rules for Radiography: Ionising Radiation Regulations Working Instructions Staff MUST Follow

Reviewed: DD Month Management of Closed Radioactive Sources

MIT Analytical X-Ray Safety Policies and Procedures

MANAGING INDUSTRIAL SITE RADIOGRAPHY WORK

Newsletter - November 2017

DEPARTMENT OF CHEMISTY, PHYSICS AND GEOLOGY RADIATION SAFETY PLAN

Minnesota Rules, Chapter 4732 X-ray Revision

Ionising Radiations Code of Practice & Guidance Contingency Arrangements

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Diamond Light Source. Accelerator. (Linac vault, Booster ring, Storage Ring and Zone 13 excluding RF test facility)

WORKING WITH IONISING RADIATION

University of Aberdeen

RP COP006- Working with Unsealed Radioactive material

SAFETY QUALITY TECHNOLOGY. Guidance on Safe Isolation Procedures

RADIATION PROTECTION SERVICE DEPARTMENT OF WELLBEING, SAFETY & HEALTH

LENS Radiation Safety Manual

MSE CONTROLLED DOCUMENT. Decontamination of NORM Contaminated Equipment

Identification of Controlled and Supervised Areas and Designated Sinks

Confined Spaces. This procedure applies to all workers of the Diocese of Maitland Newcastle (DoMN).

Local Rules: Ionising Radiation Regulations Working Instructions Staff MUST Follow

To comply with the OHS Act, the responsible manager must carry out and document the following:

Low Voltage Electricity System Safety Rules & Associated Safety Guidance

3 Philosophy of use. Leaded Gloves. Leaded Apron. Respiratory Protection (RPE) Whole body protective suit

School of Biological Sciences: Zoology Building

PNRA-RG (Rev. 0) October, 2017 RADIATION SAFETY IN INDUSTRIAL RADIOGRAPHY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

Risk Analysis and Hazard Reduction Report V1.0

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Requirements for Reduced Supervision of Power Plants, Thermal Liquid Heating Systems, and Heating Plants

Removal of Lead-Based Paint

PRO Lifting Operations

SOAS LONE WORKING GUIDANCE

Safe management of industrial steam and hot water boilers A guide for owners, managers and supervisors of boilers, boiler houses and boiler plant

Lifting Operations and Lifting Equipment Regulations (LOLER) Policy. Version No. 2. Review: May 2019

Guidance. Ionising Radiations

Hazardous Energy Control (Lockout-Tagout)

CONTROL OF SUBSTANCES HAZARDOUS TO HEALTH PROCEDURE

Golden Rule #8: Equipment Safeguarding

Contractor Control Procedures. Contractor Control Procedures. Working Together. November Borders College 24/11/2014.

PRESSURE SYSTEMS POLICY

Risk Assessment Form Cardiff School of Biosciences

SAFETY AND HEALTH INDUSTRIAL HYGIENE Hazard Communication Program Effective Date: 08/07/14 Standard: 10.1 Document Number: KUSHIH0008 Rev: 01

WORKING AFTER-HOURS OR ALONE GUIDELINE

SEVERE INJURY WITH HAND AMPUTATION AND SKULL FRACTURE

ELECTRICAL (COMPREHENSIVE) SAFETY PROGRAM REGULATORY STANDARD: OSHA - 29 CFR CFR , ,

OBJECTIVE 22: EMERGENCY WORKERS, EQUIPMENT, AND VEHICLES - MONITORING AND DECONTAMINATION

SAFE WORK METHOD STATEMENT SWMS No Version

ONR Transport Permissioning Stakeholder Event. 20% NCT Regulation

A MESSAGE FROM THE MD

RISK ASSESSMENT Installation of electrically operated service lifts (non passenger type)

ISOLATION ISSUE 2 1 AIM 2 4 REASONS FOR INCLUSION 3 6 PLANT AND EQUIPMENT REQUIREMENTS 3 7 SYSTEM & PROCEDURAL REQUIREMENTS 4 8 PEOPLE REQUIREMENTS 6

Document Control Summary Status: Replacement for Electricity High Voltage SOP v1.0 Version: V1.1 Date: July 2017 Author/Title: Owner/Title:

Vermin, Insects and Scavenging Birds Management Plan

Ground Fault Circuit Interrupter(GFCI) Policy

unable to summon help as a result of injury, ill health or an emergency. unable to carry out a task safely while alone.

in84061.txt at Page 1 of 6 2/25/00

Hazard & Risk Management

SOP. Working in Confined Spaces

HEALTH AND SAFETY SERVICES

Hazard Communication Program

MISSOURI UNIVERSITY OF SCIENCE & TECHNOLOGY Procedure: 3.6

ESTATES OFFICE OPERATIONS WORKS INSTRUCTIONS WI 6. Fume Cupboards, Ventilated Cabinets, Local Exhaust Ventilation

Lockout/Tagout Plan Environmental Health & Safety Office July 2017

Mining and Quarrying Safety and Health Regulation 2001

MINE SAFETY TARGETED ASSESSMENT PROGRAM. Ground or strata failure NSW metalliferous mines. April

ARKANSAS TECH UNIVERSITY FACILITIES MANAGEMENT HEALTH AND SAFETY MANUAL (LOCKOUT/TAGOUT) 30.0

General Requirements of the NEC Advanced Course. Building Officials and Inspectors Educational Association

Lockout / Tag out Program

Confined Spaces PTW Document Suite DOCUMENT CONTROL. Document Details. Documentation Status: Next Scheduled Review Date: July 2008

3. PURPOSE: To establish the minimum safety content of swimming pool Operations Manuals.

The University of Georgia Automated External Defibrillator (AED) Program Guidelines

Health and Safety Executive. Key aspects of HS(G) 253. Andrew Hall HID CI 1G. HM Specialist Inspector (Mechanical)

Pressure Systems Safety Regulation

Planned reference and intervention levels in Finland

Health & Safety Policy and Procedures Manual SECTION 38 RADIATION SAFETY PROGRAM

Safety Manual > Safety Checklist for Labs

Control of Hazardous Energy Lockout / Tagout Program

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

Newtown Neighbourhood Centre

Hazard Communication Plan (HCP) 29 CFR

NZQA expiring unit standard version 6 Page 1 of 5. Demonstrate knowledge of electrical safety and safe working practices for electrical workers

Personal Protective Equipment

University College Dublin Machinery/Equipment Risk Assessment Template

Regulation of radiation doses received during high altitude mountaineering expeditions

2. The purpose of this program is to achieve the following objectives:

Status Date Prepared Reviewed Endorsed Approved

Department of Materials. Department Safety Policy October 2015 (modified 16/2/2016) Contents

Scope: This plan applies to all personnel, including contractors, who enter or work in confined spaces, or supervise such activities.

CONFINED SPACE WRITTEN PROGRAM

Maintenance and general care of secondary standard and field instruments

Electrical Safety in the Food Industry

INDUSTRIAL HYGIENE GROUP Standard Operating Procedure. TVA-1000 PID/FID Portable Toxic Vapor analyzer. 1 of 9. Contents

Instructions in the Use of Radiation Protection. Equipment and Radiation Measuring Instruments by. Disaster Response Personnel in a Nuclear Emergency

Management Plan for Electrical Safety ISD #535

Design & Technology Department Health & Safety Policy. Outwood Academy Ripon

Status Date Prepared Reviewed Endorsed Approved

CONTRACTOR WHS HAZARD STANDARD CONFINED SPACE EXTERNAL USE ONLY

Transcription:

UNIVERSITY OF EDINBURGH GENERIC RISK ASSESSMENT - RADIATION RP/RA8 - WORK WITH X-RAY FLUORESCENCE ANALYSERS Foreword This is a generic risk assessment for use by persons involved in or effected by the application of radiation outlined in the Scope section below. It must be read together with the relevant completed and countersigned Proposed Scheme of Work form. Together they make up the radiation risk assessment of each application for each person working with radiation in the University. The specific control measures for each application are recorded in the relevant Local Rules. Where risks are recognised with an application that are not included in this assessment the University Radiation Protection Adviser must be informed. [Note that this risk assessment refers only to the radiation risks arising from work with X- Ray Fluorescence (XRF) Analysers. There are other more general risks arising from work with XRF Analysers and these must be properly assessed using the University s general risk assessment approach. Guidance can be obtained from the Health and Safety Department.] Fully enclosed XRF Not enclosed but not handheld Typical handheld XRF Scope This risk assessment applies to routine operations regarding the University s work with portable XRF Analysers used for material or elemental analysis. Typical types of portable XRF analysers are shown above and can roughly be categorised into 3 types: fully enclosed, not enclosed but not hand-held, and not enclosed but held in the hand during operation. Page 1 of 6

Hazards - Irradiation of part or whole of the body to external x radiation during normal use. - Irradiation of part or whole of the body to external x radiation in case of an accident; Persons likely to be exposed to the hazard: University staff, research staff, students, visiting workers, other workers (e.g. interns) and members of the public. Pregnant women and persons less than 18 years old would be at particular risk. Risk before the implementation of control measures: Category of Person High Medium Low University staff, research staff, students, visiting workers and other workers (e.g. interns) Members of the public Persons particularly at risk: Pregnant women and young persons The radiation levels in the main beam of an XRF analyser can be very high and can exceed 1Sv/h at the aperture and can still exceed 1mSv/h at 50cm. Scattered radiation levels are very much dependant on the sample being analysed and the type of XRF equipment used. Typically, scattered radiation levels are very low for fully enclosed XRF equipment or for equipment where the x-ray beam is focussed onto a small area of the sample (e.g. around 1 mm 2 ). Scattered radiation levels from hand-held XRF devices can typically be up to 500 µsv/h at 10cm around the front of the instrument near the aperture/sample but are generally less than 1 µsv/h at the trigger and at other areas around the unit away from the aperture. Reasonably foreseeable accidents: Type of Incident X-rays don t switch off when asked to do so. Equipment begins generating x-rays when it has not been asked to. A failure (i.e. mechanical, electrical or accidental damage) resulting in a person or persons being exposed to the x-ray beam Possible Effects Page 2 of 6

Type of Incident Procedural failure resulting in an unintended exposure Accident leading to an overexposure Unauthorised use of the equipment Theft / loss of the equipment Possible Effects Skin reddening unlikely to occur unless exposure in excess of 1hr; Exposure to the main beam resulting in an unnecessary radiation exposure potential restriction of work duties following the overexposure Exposure to the main beam or scattered radiation resulting in an unnecessary radiation exposure above Exposure to the main beam or scattered radiation resulting in an unnecessary radiation exposure above Justification and Optimisation It is assumed that the analysis of materials by XRF is needed to fulfil the desired work. The use of radiation generators for detection and analysis is an existing justified practice 1. Control Measures At the design or procurement stage for a project utilising an XRF analyser, consideration should first be given to whether the x-ray hazard can be engineered out; for example, by carrying out the work using a fully-enclosed piece of equipment or in an interlocked enclosure. It is always better to minimise the risk of exposure to x radiation using engineering controls than to minimise the risk using procedural and behavioural controls. Technical All XRF analysers, must: o Have a clear warning label on them to indicate that they are capable of emitting ionising radiation. This should be in a prominent position on the front of the instrument. o Have some form of key-operation or password protection to prevent unauthorised use of the equipment. o Have sufficient shielding to ensure leakage dose rates are not greater than 3 µsv/h around the equipment (except at the aperture); o Have a warning light to indicate when x-rays are being generated; 1 https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/717288/180618_justi ficationguidancedoc_updatedannex.pdf Page 3 of 6

o Have a maintenance contract in place, or if this is not required for warranty purposes, at least access to the manufacturer who can check that the equipment is performing as expected. o Only be repaired/maintained by properly trained and authorised persons (i.e. by the manufacturer or an agent of the manufacturer). o Be taken out of service if they are thought to be damaged or if any of the safety features are not working. XRF analysers that are fully-enclosed must also have: o Adequate shielding such that the radiation dose rate at any point on contact with the surface of the equipment does not exceed 3 µsv/h. o Any access points (e.g. door(s), hatch(es), panel(s)) either interlocked or fixed/bolted down to prevent access to the main x-ray beam. Opening of any of these access points must close the shutter on the x-ray tube or de-energise the x-ray generator. XRF analysers where access to the main beam is permitted under normal conditions of use must also: o If hand-held, have a two-handed option which ensures the user has both hands on the equipment during measurements; o If hand-held, have an exposure control trigger on the instrument that if released, stops the x-rays being generated; o Ideally have both of the safety features outlined below, but where this is not practicable, at least one of them must be fitted: A proximity sensor which is enabled under normal conditions of use to prevent x-rays being generated when there is not a sample held against the aperture; A low-count backscatter detector to turn off the x-rays if at any point in the measurement it detects a low count (e.g. if the sample has been moved away from the aperture/sample area); o Not permit reinstatement of the x-ray emission if the x-ray emission is interrupted at any point during the measurement (e.g. if the sample is moved away from the aperture and the x-rays stop, the mere act of moving the sample back in front of the aperture must not re-start the measurement). A manual intervention must be made to restart x-ray emission. Procedural The procedures for safe operation of XRF analysers must be written down in Local Rules. Where access to the primary x-ray beam is possible under normal conditions of use, a Controlled Area must be set up whenever the equipment is in a ready-state such that it is capable of producing x-rays. Page 4 of 6

Where access to the primary x-ray beam is possible under normal conditions of use, a suitable real-time radiation monitor must be used to detect enhanced levels of radiation in the area. This can be achieved by either placing a hand-held radiation monitor such as the Mini Instruments 900 type E in the Controlled Area near the sample or by the operator wearing a suitable Electronic Personal Dosimeter (EPD). Where practicable, operate the XRF analyser from outside the designated Controlled Area. Only essential personnel must remain in the Controlled Area during measurements. If a person s presence in the Controlled Area is essential, they should position themselves as far away from the x-ray equipment or sample as practicable (note, the operator may need to be close to the equipment if they are operating a hand-held XRF analyser). The shortest measurement time to achieve the desired result must be used. Where practicable, e.g. for small samples, carry out measurements in a shielded enclosure which is interlocked to prevent access to the main x-ray beam. Operation of XRF equipment must only be by persons trained in the correct use of the equipment. The equipment must be kept secure when not in use. The operator must wear a whole-body radiation dosemeter or extremity dosimeter during measurements if this advised in the Local Rules or PSoW form. Visual inspection of the equipment for damage must be carried out by the operator before use. If the XRF equipment is to be left unattended for any length of time or when it is not being used, it must be made safe. For example, by disconnecting the battery pack, by disconnecting cables that permit x-ray generation or by activating password protection on the equipment or enabling software. Plans must be drawn up and included in the Local Rules for procedures in case of the accidents listed above. Behavioural Persons operating fully-enclosed XRF equipment, where there is NO practical access to the main x-ray beam, do not require specific radiation safety training and can achieve adequate information and instruction by: o Reading the relevant University Radiation Protection Unit Guidance Note on enclosed x-ray apparatus; o Completing the relevant PSoW for enclosed x-ray apparatus; and, o Receiving local instruction on the general use of the equipment Persons operating XRF equipment where there IS access to the main beam under normal conditions of use must be authorised using the University s radiation authorisation arrangements, which includes the radiation protection training arrangements. Page 5 of 6

A Radiation Protection Supervisor must be appointed to provide suitable supervision of the use of XRF equipment where there is access to the main x-ray beam under normal conditions of use. For XRF equipment in which there is no access to the main x-ray beam under normal conditions of use (i.e. for fully-enclosed XRF equipment), the following is advised: o If an RPS is appointed for the area, appoint the RPS to carry out local equipment training and to countersign PSoW forms. o If no RPS is appointed for the area, the default position is that the Line Manager takes responsibility for local equipment training and to countersign PSoW forms. The appointed RPS must be trained in the measures required to ensure compliance with the controls outlined in this risk assessment and with the Local Rules. For XRF analysers that are not fully enclosed, the main x-ray beam must never deliberately be pointed at any part of the body. Samples must never be held in the hand during measurements. Dose Constraint No special dose constraint is required for work with XRF analysers. Dose Investigation Level In view of the fact that the majority of the risks can be controlled by technical means and scattered radiation levels are generally low, classified workers should not be required for work with XRF analysers. The need for personal dosimetry for XRF analysers depends on the type of analyser (e.g. fully enclosed or not) and is discussed in the relevant Local Rules for the XRF equipment. The need for personal dosimetry is then communicated to the users by return of their Proposed Scheme of Work form. Dose investigation levels are set out in the relevant Local Rules for the XRF equipment but will not exceed a cumulative dose of 6mSv over the year. Risk after the implementation of control measures: Category of Person High Medium Low University staff, research staff, students, visiting workers and other workers (e.g. interns) Members of the public Persons particularly at risk: pregnant women and young persons Page 6 of 6