RESISTANCE OF HUMIDIFIERS, AND INSPIRATORY WORK IMPOSED BY A VENTILATOR-HUMIDIFIER CIRCUIT

Similar documents
EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS

COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Medical Instruments in the Developing World

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

A MODEL LUNG WITH DIRECT REPRESENTATION OF RESPIRATORY MUSCLE ACTIVITY!


Selecting the Ventilator and the Mode. Chapter 6

AUTOVENT 4000 VENTILATOR

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

Classification of Mechanical Ventilators

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

Mechanical Ventilation

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Operation and Maintenance of the EPV200 Portable Ventilator

The comparison of Work Of Breathing methodologies on a patient model

VENTILATORS PURPOSE OBJECTIVES

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

How Ventilators Work. Chapter 3

Gas exchange measurement in module M- COVX

PROBLEM SET 8. SOLUTIONS April 15, 2004

Recommendations on Checking Anaesthesia Delivery Systems

Automatic Transport Ventilator

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

Technical Data and Specifications

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

SLE5000 Infant Ventilator with HFO

Expanding versatility. The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients

Medical Ventilators. Ryan Sanford Daniel Tajik

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

Mechanical Ventilation

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Mechanics of proportional-assist ventilation

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

APRV: Moving beyond ARDSnet


UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

Organis TestChest. Flight Simulator for Intensive Care Clinicians

Mechanical Ventilation

The aim of this guideline is to describe the indications and procedure for using high flow nasal prong oxygen

VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE

Breathing Circuits. Product training

The Merlin Small Animal Ventilator by Vetronic Services Ltd. Contents page. Introduction.2. Features of Merlin. 3

A bench study comparison of volume delivery in the presence of leak

(Received 9 September 1940)

Solutions. Digital pressure management solutions. Copyright 2013 GaleMed Corporation

Operating Instructions for Microprocessor Controlled Ventilators

Spontaneously Breathing Lung Model Comparison of Work of Breathing Between Automatic Tube Compensation and Pressure Support

Flight Medical presents the F60

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Pressure Controlled Modes of Mechanical Ventilation

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

Hallowell EMC Anesthesia WorkStation Operating Manual

Using Common Ventilator Graphics to Provide Optimal Ventilation

Neonatal tidal volume targeted ventilation

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

RESPIRATORY CARE POLICY AND PROCEDURE MANUAL. a) Persistent hypoxemia despite improved ventilatory pattern and elevated Fl02

Concentration and second-gas effects in the water analogue

Software enhancements

Guide to Understand Mechanical Ventilation Waveforms

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV)

Complex Care Hub Manual: Pressure Ventilation on the LTV 1200

Flow meter. bellow vaporizer. APL valve. Scavenging system

Potential Conflicts of Interest Received research grants from Hamilton, Covidien, Drager, General lel Electric, Newport, and Cardinal Medical Received

CORESTA RECOMMENDED METHOD N 6

Unit 15 Manual Resuscitators

Test your PICU/ NICU/ CHSU Knowledge

FRESH GAS UTILIZATION OF EIGHT CIRCLE SYSTEMS

Hospital and Transport for Controlled Breathing

Introduction to Conventional Ventilation

Blease900 Ventilator PATIENT CENTERED VENTILATION

Operator s Manual Addendum

Mechanical ventilation: simplifying the. terminology. Rapidly changing technology has unfortunately resulted in ambiguity in the terminology

Principles of Mechanical Ventilation: A Graphics-Based Approach

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

Puritan Bennett 840 Universal Ventilator

The Manley ventilator

Panther 5 Acute Care Ventilator

CATALOGUE. Adult and Infant Respiratory Care Catalogue

Health Professional Info

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

(Delivery System) Pocket Guide. Series 3 software

The Time Constant of The Respiratory System

MACS Mask CPAP System

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems

Performance of Ventilators Compatible With Magnetic Resonance Imaging: A Bench Study

Transcription:

British Journal of Anaesthesia 1991; 66: 258-263 RESISTANCE OF HUMIDIFIERS, AND INSPIRATORY WORK IMPOSED BY A VENTILATOR-HUMIDIFIER CIRCUIT T. E. OH, E. S. LIN AND S. BHATT SUMMARY The pressures and resistances of a bubble humidifier (Bennett Cascade) and a blow-by humidifier (Fisher and Paykel) were measured and computed at gas flow rates from 4.5 to 100 litre min~ 1. Pressures increased with flows, with the Bennett pressures being greater at all flows. The resistance of the Fisher-Paykel increased with flows, but remained less than that of the Bennett. An inverse resistance-flow relationship was seen with the Bennett up to a flow of 35 litre min~ 1. The work of breathing through a Servo 900C ventilator-humidifier circuit was computed, using a lung model. Work was performed by the Servo 900C on the lung, especially with the Fisher-Paykel circuit. The Bennett circuit required considerably greater (3.7 times more) inspiratory work. Thus the Bennett Cascade humidifier may present an unacceptable 10-litre min" 1 inspiratory load during spontaneous breathing. KEY WORDS Airway, resistance to breathing. Equipment, humidifiers The need for airway humidification in long-term mechanical ventilation is unquestioned. Heated water reservoir humidifiers are used in ventilator circuits to provide water vapour with little or no aerosol. There are two basic types of heated humidifiers: bubble or cascade, and pass-over or blow-by [1]. In bubble humidifiers, gas is dispersed through heated water, whereas in blow-by humidifiers, gas passes over the water surface. The Bennett Cascade (Bennett Medical Equipment, U.S.A.) and the Fisher Paykel humidifiers (Fisher and Paykel Medical, New Zealand) are the most commonly used heated bubble and blow-by heated humidifiers, respectively. It is recognized that successful weaning from mechanical ventilation may be determined by the heavy inspiratory load of some breathing systems [2, 3], but there is little published literature on the contribution of humidifiers to the work of spontaneous breathing. This study was undertaken to compare the flow resistances of the Bennett Cascade and Fisher-Paykel humidifiers, and the inspiratory work of a ventilator circuit with each humidifier. MATERIALS AND METHODS The resistances of a Bennett Cascade humidifier and a Fisher-Paykel humidifier (MR 500) were studied under constant flow conditions. Both humidifiers were filled with water, but remained unheated. Wall source air was passed into the inlet hose of each humidifier via a RT-200 Calibration Analyzer (Timeter Instrument Corp., PA, U.S.A.). The gas flow rate was set successively at 4.5, 7, 10, 15 and 20 litre min" 1, and thereafter at increments to a maximum of 100 litre min" 1. At each constant gas flow rate, pressures were measured at both the inlet and outlet, using a differential strain gauge pressure transducer (Motorola MPX 10DP). Resistive pressure at each flow rate was given by the pressure decrease. Three sets of pressure and flow measurements were recorded with each humidifier. With the Fisher-Paykel circuit, a new disposable water chamber was used for each set of measurements. Work of breathing was studied using a Servo 900C ventilator (Siemens Elema, Sweden). The Servo 900C was set to cater for spontaneous breathing (continuous positive airway pressure T. E. OH, D.A., F.C.ANAES., F.F.A.R.A.C.S.; E. S. LlN, M.SC, M.R.C.P., F.CANAES.; S. BHATT, M.D., F.C.ANAES.; Department of Anaesthesia and Intensive Care, Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, N.T., Hong Kong. Accepted for Publication: August 17, 1990. Correspondence to T. E.O.

RESISTANCE AND WORK OF HUMIDIFIERS 259 Modified Cape ventilator (TC50) Gould oscilloscope recorder Servo 900C ventilator Pressure-flow transducers Humidifier FIG. 1. Schematic diagram of model lung to simulate spontaneous breathing. The Servo 900C ventilator was set to the "CPAP" mode. (CPAP) mode), with a triggering sensitivity marginally greater than 0 cm H 2 O. Spontaneous ventilation was simulated using a modified Cape TC 50 ventilator (Cape Engineering, England), adjusted to simulate a tidal volume of 600 ml at a rate of 15 b.p.m. and with a peak inspiratory flow of about 45 litre min" 1 (fig. 1). Airway pressures and flows were measured at the patient airway, using a strain gauge pressure transducer (Motorola MPX 10DP) and a platinum hot wire flow transducer (Ohmeda SE 302T). The pressure and flow transducers were calibrated with the RT-200 Calibration Analyzer. Simultaneous pressure and flow signals were digitized at a rate of 200 s" 1 and recorded and displayed on an oscilloscope (Gould DSO 1604). Work of breathing imposed by the whole breathing system was derived from the pressureflow values. The product of pressure (P) and flow (V) gives instantaneous power (W) at that moment. Power over a defined time interval (T 0 -T T ) is given by integration of P. V over that time. Work during that time interval is then given by W T 1 W.dt This method of determining Jr. work done has been described previously [2, 4]. Inspired tidal volume (Vr) was derived from integration of flow during the inspiratory phase Ti. Work of breathing was converted to J/litre of ventilation. Data were analysed by linear and polynomial regression and a comparison between the two humidifiers was made using analysis of variance; P < 0.01 was considered significant. RESULTS The three sets of pressure recordings with each humidifier were nearly identical. The coefficient of variation for the pressure recordings ranged from 0.1 to 2.8% except at the smallest flows. At a flow rate of 4.5 litre min" 1, the coefficient of variation was 5.9% for the Fisher-Paykel and 8.7 % for the Bennett Cascade. Resistive pressures increased with greater flows (fig. 2, table I). The pressure:flow curves were fitted with a second order polynomial with r = 0.999 for the Bennett a Q_ J ID essui Q. Resistive 1.2 0.8 0.6 0.4 0.2 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 Flow (litres' 1 ) FIG. 2. Pressure-flow curves of Bennett Cascade humidifier (O) and Fisher-Paykel humidifier (#). Pressures of the Bennett are greater at all gas flows.

260 TABLE I. Pressures and resistances of Fisher-Paykel (F-P) and Bennett Cascade (Bennett) humidifiers related to gas flow rates BRITISH JOURNAL OF ANAESTHESIA 0.8 0.6 Mean flow (litre min" 1 ) Mean pressure (kpa) Mean resistance (kpa litre" 1 s) 0.4 F-P 4.5 7.0 10.0 15.3 20.8 30.0 40.0 50.5 60.6 70.2 80.5 90.5 100.7 2.5 Bennett 4.5 8.0 10.5 15.1 20.1 29.8 41.9 51.4 62.5 70.5 80.9 90.1 100.5 F-P 0.003 0.006 0.011 0.025 0.042 0.084 0.147 0.221 0.326 0.463 0.621 0.779 0.968 Bennett 0.200 0.231 0.242 0.263 0.295 0.347 0.452 0.558 0.673 0.779 0.905 1.010 1.157 F-P 0.045 0.053 0.063 0.097 0.120 0.168 0.221 0.263 0.322 0.395 0.463 0.516 0.571 Bennett 2.665 1.736 1.383 1.045 0.884 0.699 0.640 0.650 0.645 0.663 0.670 0.673 0.673 7~ 0.2 1 0 o -0.2 LL. -0.4-0.6-0.8 100 200 300 400 600 Time (0.005 s) FIG. 4. Flow curves of a simulated spontaneous breath from a Servo 900C ventilator with a Fisher-Paykel (dotted line) and a Bennett Cascade (solid line) humidifier in circuit. A-B = Expiratory phase; expiratory flows are identical. B-C = Inspiratory phase shown with a negative polarity; peak inspiratory flow was set at 0.75 litre s" 1 (45 litre min" 1 ). Q_ a 1.5 ID - 1 0) ' CD 0.5 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 Flow (litres' 1 ) FIG. 3. Resistance-flow curves of Bennett Cascade humidifier (O) and Fisher-Paykel humidifier (#). The Bennett has an inverse relationship for flows up to 35 litre min" 1. Cascade and r= 1.000 for the Fisher-Paykel. Resistive pressure at each flow rate with the Bennett Cascade was always greater compared with the Fisher-Paykel (P < 0.001). Calculated resistances (table I) were plotted against flow rate (fig. 3). The Bennett Cascade's resistance was high initially, but decreased sharply with increasing flows up to 35 litre min" 1. There- after, it remained relatively constant at a mean of 0.66 kpa litre" 1 s. With the Fisher-Paykel, resistance showed a small linear increase with increasing flow (r = 0.998) but was always less than the Bennett's resistance (P < 0.001). Resistances at 45 litre min" 1 were 0.22 kpa litre" 1 s for the Fisher-Paykel, and 0.64 kpa litre" 1 s for the Bennett Cascade. In the work of breathing experiments, the lung model produced identical curves in each ventilatory cycle for both flow and pressure. The flow curves of the Servo 900C system with each humidifier are shown in figure 4. Simulated inspiration was accorded a negative polarity for easier visualization. The expiratory flow curves were identical and the inspiratory curves almost identical for the two humidifiers. However, corresponding inspiratory pressure curves (fig. 5) were different. The Bennett Cascade generated larger negative inspiratory pressures (maximum of -0.43 kpa) than the Fisher-Paykel (maximum of 0.24 kpa). Computed inspired FT was 680 ml. Power curves for both humidifiers are shown in figure 6. Positive values denote the power required to overcome the resistance of the system, whereas negative values are the power inherent in the

RESISTANCE AND WORK OF HUMIDIFIERS 261 TABLE II. Work done (/Wire" 1 ) during a simulated spontaneous breath using a Servo 900C ventilator system with the specified humidifier in-line Inspiratory Expiratory Humidifier phase phase Total Fisher-Paykel Bennett Cascade -0.032 0.087 0.140 0.137 0.108 0.224-0.6 300 400 Time (0.005 s) FIG. 5. Pressure curves of a simulated spontaneous breath from a Servo 900C ventilator with a Fisher-Paykel (dotted line) and a Bennett Cascade (solid line) humidifier in circuit. Expiratory curves are identical and shown here to precede inspiration. Inspiratory curves show extremely rapid pressure changes. The Bennett Cascade generated greater negative pressures. -0.4 100 200 300 400 Time (0.005 s) 500 600 FIG. 6. Computed power curves of a simulated spontaneous breath from a Servo 900C ventilator with a Fisher and Paykel (dotted line) and a Bennett Cascade (solid line) humidifier in circuit. Expiration is represented here to precede inspiration. Negative power values denote work done on the lung by the breathing system. The Bennett Cascade required greater power during inspiration. system because of an assist effect of the ventilator (see below). Work done during a simulated spontaneous breath is given in table II. Inspiratory work (Wi) with a Bennett Cascade in the circuit was 0.087 J litre" 1 ; with the Fisher- Paykel circuit it was 0.032 J litre" 1. DISCUSSION Previous studies have reported values for work of breathing in ventilators and tracheal tubes [2, 3, 5-15], but reports on humidifiers are few. This is somewhat surprising, as humidifiers were designed originally for mechanical and assisted ventilation, and not specifically for spontaneous breathing. Dennison and co-workers [5] studied the resistances of eight circuits, including two with a Bennett Cascade humidifier, but tracheal tubes were attached. Mecklenburg and colleagues [2] reported on the resistance and work of breathing of the Kontron ventilator and its humidifier, a bubble device. Our present study compared the Bennett Cascade with the Fisher- Paykel, a blow-by humidifier. With both humidifiers, increasing resistive pressures were produced by increasing flows. Greater resistive pressures were seen with the Bennett Cascade as expected, as it requires gas to be drawn through water. The curvilinear pressure-flow curve obtained with the Bennett Cascade was similar to that reported by Dennison and co-workers for a Bennett Cascade circuit with a 9-mm tracheal tube [5]. When resistances were computed, the Bennett Cascade demonstrated an inverse relationship between resistance and flow for flow rates less than 35 litre min" 1 (fig. 3). This phenomenon was reported also by Dennison and co-workers [5] and may be caused by the complex passage of gas through the Bennett Cascade. Gas is drawn through water in its chamber and then through a grid, creating froth [1] which decreases re-

262 BRITISH JOURNAL OF ANAESTHESIA sistance to gas flow. Low gas flows may produce proportionately less froth and thus meet with greater resistance. As a corollary, if sufficient froth is produced by adequate flows, resistance remains relatively constant. This resistance was demonstrated in our study to be 0.66 kpa litre" 1 s with flows greater than 35 litre min" 1 (fig. 3). The Fisher-Paykel humidifier, on the other hand, has gas passing over the water surface [1]. Resistance would be expected to be less than that of bubble humidifiers. In addition, it incorporates filter paper wicks to increase water surface area effectively. Thus increased flow rates are likely to increase turbulence within the unit, giving rise to increases in resistance. The resistance increases were small in this study. There are no specifications for a maximum resistance to flow applicable to humidifiers, as far as we know. Nunn recommended an upper limit of about 0.6 kpa litre" 1 s at a flow rate of 30 litre min" 1 for a breathing system [16]. The resistance of the Fisher-Paykel alone at the same flow was 0.17 kpa litre" 1 s (with a maximum of 0.57 kpa litre" 1 s at 100 litre min" 1 ). However, the resistance of the Bennett Cascade was considerably greater than the recommended upper limit at low flows, and approximated that limit with flows greater than 30 litre min" 1. Consequently, the resistance of a ventilator circuit with a Bennett Cascade humidifier would be considerably greater than 0.6 kpa litre" 1 s. The humidifiers were studied also in a simulated clinical setting using a popular modern intensive care ventilator, the Servo 900C. A simulated peak inspiratory flow of 45 litre min" 1 was chosen, which approximates to those of stable weaning patients [5, 11]. However, our lung model was not able to produce physiological expiratory flows. Hence computed work done on expiration might be expected to be artificially high. Nevertheless, this did not affect the experiment, as differences in pressure and work of breathing were produced in the inspiratory phase. Our study was not designed to examine natural work against the patient's own compliance and resistance. The work of breathing for a single breath is expressed appropriately in terms of joules or J/litre of inspiration. Thus the Servo 900C- Bennert Cascade system demanded 3.7 times more work than the Fisher-Paykel system for an inspiratory breath with a peak flow of 45 litre min" 1. Nevertheless, the Bennett Cascade's inspiratory work (0.087 J litre l ) is much less than the 0.544 J litre" 1 (0.272 J for 500 ml FT) reported for the Kontron ventilator with its humidifier [2]. "Normal" work in breathing directly from atmosphere at rest is about 0.5 J litre" 1 [8, 15]. The Servo 900C-Bennett Cascade system thus demanded 17.4% additional work. On the other hand, the Servo 900C-Fisher-Paykel system performed work on the lung. This "reduced work" characteristic of the Servo 900C was reported also by bench [8, 10, 11] and clinical [9] studies on the work of breathing imposed by demand valves of different ventilators. Comparison with our study is difficult, as the circuits in these studies either excluded in-line humidifiers [9] or included tracheal tubes [8, 11] or use of CPAP [10]. The experimental model of Katz, Kraemer and Gjerde included unspecified in-line humidifiers and found that the Servo 900C reduced work by -0.025 to 0.035 J litre" 1 at peak flows of 20-60 litre min" 1 [8]. A reduced work of -0.020 J litre" 1 at aflowof 20 litre min" 1 was reported by Bersten and colleagues [11] with a Fisher-Paykel humidifier and a 7-mm tracheal tube. Our value of 0.032 J litre" 1 without a tracheal tube, at a flow of 45 litre min" 1, is consistent with the values reported above. The larger negative inspiratory pressures seen with the Servo 900C-Bennett Cascade system confirmed the Bennett's greater resistance. For both humidifiers, an assist effect from the ventilator during the inspiratory phase was obvious. The Servo 900C directs gas into the circuit during inspiration from a pressurized concertina reservoir bag [1, 17]. Hence inspiratory flow can exceed demand, as represented by the increase in airway pressure to greater than that of endexpiration [8] (fig. 5). This would explain the "reduced work" effect of the Servo 900C ventilator. The pressure flutter during inspiration (fig. 5) can be explained by the inspiratory demand valve of the Servo 900C. This scissors valve is activated by a servo-controlled, rapid cycling stepper motor [17]. A delay of about 0.15 s follows the start of a patient breath before the occluded valve opens [18]. As gas rapidly enters the circuit, a positive pressure is recorded. The gas delivery, patient inspiratory effort, circuit resistance and servo valve result in an inspiratory pressure flutter. This flutter can be significant if the patient's inspiratory flow demands are not met. Excessive circuit resistance can cause respir-

RESISTANCE AND WORK OF HUMIDIFIERS 263 atory muscle fatigue and prolong weaning from mechanical ventilation. The humidifier and the ventilator's inspiratory demand valve are the major contributors to the inspiratory resistance of a breathing system. Higher circuit resistances require greater negative intrapulmonary pressures to initiate each spontaneous breath, thus increasing the work of breathing. The peak inspiratory flows of weaning patients are usually about 40-50 litre min" 1, but greater flows may be required under stress [5]. If flow from the breathing system is insufficient to meet inspiratory demands, the resistive load on the patient is further increased. This is especially pertinent to the Bennett Cascade humidifier. In conclusion, the resistance and inspiratory work imposed by the blow-by Fisher-Paykel humidifier is acceptable over the relevant range of patient inspiratory flows. However, the Bennett Cascade humidifier requires much more work for a spontaneous breath, and even by itself, has considerably greater resistances at the same flow rates. These findings confirm our clinical observation that patients are difficult to wean with the Bennett Cascade humidifier. Thus the Bennett Cascade humidifier should not be used for weaning, unless the ventilation mode or gasflowis able to reduce the work of breathing, such as with a Servo 900C. Indeed, the use of bubble humidifiers for spontaneous breathing is questioned. REFERENCES 1. McPherson SP. Respiratory Therapy Equipment, 3rd Edn. St Louis: C. V. Mosby & Co, 1985. 2. Mccklcnburgh JS, Latto IP, Al-Obaidi TAA, Swai EA, Mapleson WW. Excessive work of breathing during intermittent mandatory ventilation. British Journal of Anaesthesia 1986; 58: 1048-1054. 3. Shapiro M, Wilson RK, Casar G, Bloom K, Teague RB. Work of breathing through different sized endotracheal tubes. Critical Care Medicine 1986; 14: 1028-1031. 4. Engstrom CG, Norlander OP. A new method for analysis of the actual power as a function of gas flow, pressure and time. Acta Anaesthcsiologtca Scandinavica 1962; 6: 49 55. 5. Dennison FH, Taft AA, Mishoe SC, Hooker LL, Eatherly SB, Beckham RW. Analysis of resistance to gas flow in nine adult ventilator circuits. Chest 1989; 96: 1374-1379. 6. Gibney RTN, Wilson RS, Pontoppidan H. Comparison of work of breathing on high gas flow and demand valve continuous positive airway pressure systems. Chest 1982; 82: 692-695. 7. Gherini S, Peters RM, Virgilio RW. Mechanical work on the lungs and work of breathing with positive endexpiratory pressure and continuous positive airway pressure. Chest 1979; 76: 251-256. 8. Katz JA, Kraemer RW, Gjerde GE. Inspiratory work and airway pressure with continuous positive airway pressure delivery systems. Chest 1985; 86: 519-526. 9. Beydon L, Chasse M, Harf A, Lemaire F. Inspiratory work of breathing during spontaneous ventilation using demand valves and continuous flow systems. American Review of Respiratory Diseases 1988; 138: 300-304. 10. Samodelov LF, Falke KJ. Total inspiratory work with modern demand valve devices compared to continuous flow CPAP. Intensive Care Medicine 1988; 14: 632-639. 11. Bersten AD, Rutten AJ, Vedig AE, Skowronski GA. Additional work of breathing imposed by endotrachcal tubes, breathing circuits, and intensive care ventilators. Critical Care Medicine 1989; 17: 671-677. 12. Sassoon CSH, Kees Mahutte C, Simmons DH, Light RW. Work of breathing and airway occlusion during assist-mode mechanical ventilation. Chest 1988; 93: 571-576. 13. Marini JJ, Capps JS, Culver BH. The inspiratory work of breathing during assisted mechanical ventilation. Chest 1987; 87: 612-618. 14. Marini JJ, Rodriguez RM, Lamb V. The inspiratory workload of patient initiated mechanical ventilation. American Review of Respiratory Diseases 1986; 134: 902-909. 15. Bolder PM, Healy TEJ, Bolder AR, Beatty PCW, Kay B. The extra work of breathing through adult endotracheal tubes. Anesthesia and Analgesia 1986; 65: 853-859. 16. Nunn JF. Applied Respiratory Physiology, 2nd Edn. London: Bunerworths, 1977; 103. 17. Spearman CB, Sanders HG jr. Physical principles and functional designs of ventilators. In: Kirby RK, Smith RA, Desautels DA, eds. Mechanical Ventilation. New York: Churchill Livingstone, 1985; 59-114. 18. Op't Holt TB. Work of breathing and other aspects of patient interaction with PEEP devices and systems. Respiratory Care 1988; 33: 445-453.