Scleral Fitting Guide

Similar documents
Fitting Guide. Minutes-to-Fit 2.0. by SynergEyes Moving Vision Forward

An introduction to modern scleral lenses

FITTING AND TROUBLESHOOTING SCLERAL LENSES

DISCLOSURES FITTING AND TROUBLESHOOTING SCLERAL LENSES WHO S A GOOD CANDIDATE SCLERAL LENS ADVANTAGES ANDREW J. BIONDO, OD, FSLS.

Acknowledgments. Paragon Vision Sciences Adoption of CRT certification slide for this generic talk. Material Considerations

NormalEyes Guidelines For Successful Fitting RIGID GAS PERMEABLE SCLERAL CONTACT LENSES. Manufactured in Paragon HDS 1OO (paflufocon D)

Sofmed breathables XW (comfilcon A) Sofmed breathables XW premium (comfilcon A) Sofmed breathables XW toric (comfilcon A)

Spherical Soft Contact Lens Fitting. J. Perrigin OPTO 6374 Spring 2017

The BE Retainer System Certification Course. Supplement

Dr. Thomas R. Reim President, DreimLens, Inc.

POLYMACON HYDROPHILIC CONTACT LENSES PRACTITIONER FITTING GUIDE

Pardon the Objection (PTO) Contact Lens Clinical Case Debate (2 hours)

PROFESSIONAL FITTING AND INFORMATION GUIDE. Paragon CRT Manufactured in Paragon HDS (paflufocon B)

Ophthalmology Clinical Pearls A Quick Guide To Crystalens

Bulletin NEXT MEETING

Orthokeratology. What is Orthokeratology? How does it work?

Appendix C: Skill Stations Examination - Marking Rubric

Biometry tips. Marilyn Watkins-Ramdin, COMT ROUB

NAVAL SUBMARINE MEDICAL RESEARCH LABORATORY

PRODUCT PORTFOLIO RIGID GAS PERMEABLE CONTACT LENSES SOFT CONTACT LENSES LENSCARE SOLUTIONS ACCESSORIES

LASIK Wavefront Guided idesign LASIK with IntraLase

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) ICO-OSCAR Instructor Directions

ARTICLE IN PRESS Contact Lens & Anterior Eye xxx (2012) xxx xxx

AIR OPTIX BRAND CONTACT LENSES LENS PARAMETER GUIDE November 2016

EVERY TREATMENT IS A TRUE ORIGINAL

OPERATION. Estimated kerf width compensation. HPR260 Manual Gas Instruction Manual 4-9

LENS EXCHANGE WITH IMPLANTATION OF A BITORIC INTRAOCULAR LENS IN EYES WITH SIGNIFICANT ASTIGMATISM AFTER PREVIOUS CORNEAL SURGERIES

Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650

Appendix C: Skill Stations Examination - Marking Rubric

Figure 1 Schematic of opposing air bearing concept

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

FRA Spacer Instruments. For anterior lumbar interbody fusion (ALIF) using straight anterior or anterolateral approach.

Mild steel. Stainless steel. HPR130XD Auto Gas Revision 2. Nozzle retaining cap. tube. Shield cap 30 A 50 A 80 A 130 A

Relative Dosimetry. Photons

Getting Started GUIDE LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL pg 1

Model 23H Hand Crank Seamer

PACKAGE INSERT RIGID GAS PERMEABLE CONTACT LENSES FOR ORTHOKERATOLOGY. (hexafocon B)

Crusher Bearings: Knowing the Basics Leads to Better Care

DIRECT DRIVE DIXIE DOUBLE SEAMER Model 25D

FITTING and ADJUSTING The Free n Easy SADDLE

How images are formed in the eye

Rejuvenating Aftercare for Contact Lens Patients

V E J A R A N O P R O T O C O L FOR T H E M A N A G M E N T O F Q U E R A T O C O N O SURGICAL T I PS

Kintrol Instructions for Use Product Number: VS4

Underwater Housing for Panasonic Lumix DMC-ZS60, DMC-TZ80

Underwater Housing for Canon G7X Mark II

Standard Operating Procedure - First Draft for Review Only- StarWeld Ultra DX

Contact lenses designed to fit more of your patients

TECH TIPS: ROPE DEFECTS

Vibration isolation system 1VIS10W. User manual

Inflatable Dock Slide Owner s Manual

SECTION 4 - POSITIVE CASTING


STRIP EDGE SHAPE CONTROL

Info Counting chamber (haemacytometer)

Assembling & Installing your Pool Cleaner

Front Cover Picture Mark Rasmussen - Fotolia.com

senofilcon A Soft (hydrophilic) Contact Lenses Visibility Tinted with UV Blocker for Daily Disposable Wear

Oxygen tension under a contact lens. Kenneth A. Poise and Marianne Decker

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

HCMTCB MATERIALS SAMPLING & TESTING PERFORMANCE CHECKLIST

METROLOGY LAB. I DEVICES BASED ON VERNIER SCALE

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR)

SUPER BLUE INSTRUCTIONS FOR TRIMMING AND INSTALLING ORIGINAL NETS WITH STRIPES

MODEL 200 KNIFE GATE VALVES INSTALLATION & MAINTENANCE MANUAL

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Ripple Tank Exploring the Properties of Waves Using a Ripple Tank

THE POWER IS ON RIGGING + TUNING GUIDE Specifications First Time Rigging 9.1. Rig Assembly. Tuning Tips.

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

CROSS COURSE INSTRUCTION DOCUMENT

Aeris Activity. Product Manual

RATLINER Assembly Instructions.

IN-PLACE DENSITY AND MOISTURE CONTENT OF SOIL AND SOIL- AGGREGATE BY NUCLEAR METHODS (SHALLOW DEPTH) FOP FOR AASHTO T 310

Outline Chapter 7 Waves

Martin Paired 13 Spading Closing Wheels

Ripple Tank P CONTENTS: DEMONSTRATION SETUP INSTRUCTIONS:

AllPro Foot. Product Manual

1.0 - OPENING AND CLOSING THE DOOR

Calibration Procedure

AllPro Foot. Posterior Mount Product Manual

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal

The Usage of Propeller Tunnels For Higher Efficiency and Lower Vibration. M. Burak Şamşul

How to ensure optimal functionality of Campagnolo drivetrains

UsER manual for Watersens ph -REDOX

Installation of Your SprayMaster System

DBML-60/80 Squeeze Tool

Pontoon Slide Owner s Manual


Your new Keystone Telebinocular

Kinterra Instructions for Use Product Number: ROM-00-XXXXX-XX

Aeris Performance 2. Product Manual

Deep water plume models - What s special about deep water

TEST PROTOCOL VERIFICATION OF THE BEHAVIOUR OF A MECHANICAL COMPRESSION COUPLING DURING SETTLEMENT

Dick Bowdler Acoustic Consultant

Self-Aligning Cylindrical Roller Bearings. Plummer Block Housing. Series ACB, sealed on both sides. Series SLG01

200 STRINGING MACHINE

Unity Sleeveless Elevated Vacuum System

ASICO Innovative I/A Tip System (Patented) Maximize your I/A efficiency for Today s MICS and FLACS procedure

CHANGE RECORDS TABLE OF CONTENTS

PERFORMANCE VS INTERCEPTOR EXTENSION

Transcription:

Scleral Fitting Guide

Scleral Lens Design...3 Fitting Philosophy...4 Select a Diagnostic Lens...5 Initial Lens Evaluation...6 Assess Scleral Landing Zone (SLZ), Sagittal Depth, Limbal Clearance, Lens Power...7-10 Troubleshooting Tips...11 Quick Reference Guide...Back Cover Ordering and Technical Support Hours: Monday-Friday: 9:00AM to 8:00PM (EST) To Place Orders Phone: 877.733.2012 option 1 Fax: 877.329.2012 Email: orders@synergeyes.com Online: SynergEyes.com/Professional Customer Care Representatives Phone: 877.733.2012 option 1 Email: customerservice@synergeyes.com Technical Consultation Phone: 877.733.2012 option 2 Email: consultation@synergeyes.com Returns Phone: 877.733.2012 option 1 Place orders, track shipments or view your account statements 24 hours a day on RxConnect SynergEyes.com/Professional Parameters: Sagittal Depth 3000 to 4600 µm in 100 µm steps Flat and Steep Meridian SLZ Both meridians available 28 to 50 in steps of 2 Base Curves 7.4 to 9.2 mm in 0.2 mm steps Diameter Power Materials 14.0 to 17.5 mm in 0.5 mm steps -25.00 to +25.00 D in 0.25 D steps Menicon Z and Contamac Optimum Extreme Front Toric Parameters: Axis 0 to 180 in 1 steps Cylinder Power -0.75 to -3.00 D in 0.25 D steps 2

SynergEyes VS Scleral Lens Design The SynergEyes VS lens is an innovative scleral lens with a distinctive bi-tangential periphery. The lens completely vaults the cornea and limbus landing entirely on the sclera. The toric periphery of the lens may be precisely controlled in both flat and steep meridians, aligning with the sclera for ease of landing and stability. This non-rotationally symmetric lens design combines two key innovations to maximize comfort and optimize lens performance: linear landing zones and bi-tangential toricity. The linear landing zones are designed to follow the linear (rather than curved) shape of the sclera accommodating a toric sclera, which is present in 95% of cases. The adjustable scleral landing zone (SLZ) in the flat or steep meridians aims to distribute the lens pressure more equally on the sclera to improve the scleral lens fit. Sagittal depth Front optic zone Mid-periphery Scleral landing zone Technology Inside Back optic zone Background Research data suggests that toricity is more pronounced in the scleral area than in the limbal area, irrespective of the toricity of the cornea. 1 Van der Worp, et.al found in most cases, the shapes of the limbus and anterior sclera were linear rather than curved. This information was the basis for designing a scleral lens with a bi-tangential (non-rotationally symmetrical) periphery. The objective of a scleral lens with tangential fitting is to reduce the adhesive pressure on the eye, to achieve greater tolerance and prolong the daily duration of use. 2 The ability to adjust the flat or steep meridian of a bi-tangential design distributes the lens pressure more equally on the sclera and improves the scleral lens fit. 3 A special thank you to Henny Otten of Visser Contact Lens Practice for many of the images. 1. Van der Worp E, Graf T, Caroline PJ. Exploring beyond the corneal borders. Contact Lens Spectrum 2010; 25 (6):26 32. 2.Van der Worp E. A Guide to Scleral Lens Fitting (monograph online). Pacific University Common Knowledge: Books and Monographs. 2010. Available at: http://commons.pacificu.edu/cgi/viewcontent.cgi?article=1003&context=mono. Accessed June 13, 2013. 3. Visser, Medical Applications and Outcomes of Bi-tangential Scleral Lenses. Optometry and Vision Science VOL. 90. NO.10, Oct. 2013 3

Fitting Philosophy The SynergEyes VS lens is designed to fully vault the cornea and the limbus and gently land on the sclera. A well-fit lens exhibits the following characteristics: The linear scleral landing zone has edge alignment without vessel blanching. Apical corneal clearance of approximately 200 µm after settling. Limbal clearance of about 100 µm. The design of the SynergEyes VS lens uses peripheral toricity and linear scleral landing zones as primary elements. Over 95% of the scleral lenses prescribed have a toric periphery. After a quick check for central touch or excessive clearance, observe the scleral landing zone. After 30 minutes, align the scleral landing zone with the sclera and then observe the sagittal depth. The following lens parameters affect how the lens fits: Lens Location Observation Adjust Scleral Landing Zone (SLZ) Central Optic Zone Mid-periphery Limbal Zone Edge Lift and Blanching Apical Clearance Limbal Clearance SLZ, Steep Meridian SLZ Sagittal Depth SLZ (first) Base Curve (second) Diagnostic Lenses The diagnostic lens set consists of 16 SynergEyes VS lenses with a diameter of 16.0 mm, a base curve of 8.4 mm and plano powers. All lenses have toric peripheries and the toricity is expressed in two meridians in the scleral landing zone (SLZ): a flat meridian landing zone and a steep meridian landing zone. In the fitting set, the landing zone difference for all lenses corresponds with a height difference between the flat and steep meridians of about 180 µm at the edge of the lens. Increase SLZ number to tighten (steepen) Scleral landing zone Central optic zone Mid-periphery limbal zone Decrease Apical Clearance Sagittal Depth 3400 32-38 3600 32-38 SLZ Flat meridian 3400 34-40 3600 34-40 3800 34-40 3400 36-42 3600 36-42 START HERE 3800 36-42 3400 38-44 3600 38-44 3800 38-44 3400 40-46 3600 40-46 Increase Apical Clearance Lens Identification: The diagnostic lenses are laser marked. SLZ Steep meridian 4000 34-40 4000 36-42 4000 38-44 Decrease SLZ number to loosen (flatten) Sagittal depth (μm) SLZ flat meridian Scleral Lens Markings SLZ steep meridian 4

Select a Diagnostic Lens There are three ways to choose an initial diagnostic lens: 1. Start with a 3600 µm sagittal depth, 36 flat meridian, 42 steep meridian. This is the most prescribed lens. Select 3800 µm, 36 flat meridian, 42 steep meridian for the other eye. 2. Use Rx Fitting Resource to determine a starting lens. 3. Select a lens based on experience and this fitting guide which is based on the fitting protocol of the Visser Contact Lens Practice. A. Selecting Initial Flat and Steep Meridian Scleral Landing Zones (SLZ): To select the initial flat meridian SLZ, start with 36 in the flat meridian or assess the scleral profile. A normal scleral profile ranges from 34 to 38. For a flatter scleral profile, use 28, 30 or 32. For a steeper scleral profile, use 40 and above. All diagnostic lenses have a steep meridian SLZ that is 180 µm greater than the flat meridian SLZ. Sclera Profile SLZ Percent Usage Flat Profile 28-32 8% Normal Profile 34-38 80% Steep Profile 40-50 12% Tip: There is a 60µm change for each SLZ number. Sagittal depth Front optic zone Mid-periphery Scleral landing zone Technology Inside Back optic zone B. Selecting Sagittal Depth: The most common sagittal depth of the SynergEyes VS scleral lens is 3600 µm. Similar to other scleral lenses, the desired sagittal depth varies with condition. A keratoconus patient may have a deeper cornea profile and require a greater increase in sagittal depth compared to a normal eye, while a post-corneal graft or post-refractive surgery patient may require a more shallow depth. Corneal profile may be assessed upon visual inspection. Another option is to use the Rx Fitting Resource which accesses corneal topographer measurements, estimates corneal height and recommends a starting sagittal depth. Cornea Profile Sagittal Depth Percent Usage Shallow 3000 µm to 3300µm 10% Normal 3400 µm to 3800µm 75% Deep 3900 µm to 4100 µm 10% Very Deep 4200 µm to 4400 µm 5% 5

Initial Lens Evaluation Clean and condition the diagnostic lens. Place the selected lens on eye filled with non-preserved saline and fluorescein. Upon insertion, perform a quick check for central touch or excessive clearance and then evaluate the scleral landing zone (SLZ) and look for scleral alignment. The lens should be well-balanced with gentle movement upon a push-up test. Hash Mark Steep Meridian Allow lens to settle for 30 minutes. The diagnostic lenses have hash marks to indicate the flattest meridian. Measure the axis by projecting a narrow beam from the slit lamp parallel to the engravings on the scleral lens. The axis can be read from the axis gauge. Make note of the position of the flat meridian. Hash marks will align with flattest meridian. Assess Scleral Landing Zone (SLZ) Assess the lens initially in primary gaze, holding lids away to assess the periphery. Shift gaze in all directions for final assessment for lens movement and potential binding. Extreme gaze shift may show false readings. Check scleral landing zone for alignment with the sclera by observing blanching or edge lift. If blanching exists decrease the SLZ number (flatten) in the corresponding meridian. If edge lift exists, increase the SLZ number (steepen) in the corresponding meridian. If SLZ is too tight (too steep) decrease the SLZ number (flatten) Lens edge too steep. Lens too tight. Decrease SLZ by 2. 6 Blanching. Lens too tight. Decrease SLZ by 4. Extreme edge tightening around entire lens. Decrease SLZ by 6 in both meridians.

Assess Scleral Landing Zone (SLZ) If SLZ is too loose (too flat) increase the SLZ number (steepen) Excessive edge lift in flat meridian. Increase SLZ by 2 in flat meridian. Excessive edge lift in both meridians. Increase SLZ In both meridians. Observation Suggestion Blanching: Lens edges in both meridians are too steep (too tight) Edge Lift: Lens edges in all meridians appear too flat (too loose) Too Steep Too Flat Too Steep Too Steep Too Flat Steep Meridian Steep Meridian Too Flat Too Steep Too Flat Decrease flat meridian SLZ by 4 and decrease steep meridian SLZ by 4. Example: Flat meridian SLZ: 36 to 32 Steep meridian SLZ: 42 to 38 Increase flat meridian SLZ by 4 and increase steep meridian SLZ by 4. Example: Flat meridian SLZ: 38 to 42 Steep meridian SLZ: 44 to 48 Edge Lift: Lens edges are too flat (too loose) in the steep meridian and ideal in the flat meridian Too Flat Steep Meridian Too Flat Keep the flat meridian SLZ the same. Increase steep meridian SLZ by 4. Example: Flat meridian SLZ: 36 stays at 36 Steep meridian SLZ: 42 to 46 Blanching: Lens edges are too steep (too tight) in the flat meridian and ideal in the steep meridian Too Steep Steep Meridian Too Steep Decrease flat meridian SLZ by 4 and keep the steep meridian SLZ the same. Example: Flat meridian SLZ: 38 to 34 Steep meridian SLZ: 44 stays at 44 7

Assess Sagittal Depth Assess the apical clearance at the highest point of corneal elevation using an optic section with a white light or an anterior segment OCT. The ideal apical clearance is approximately 200 µm. The diagnostic lens thickness is 350 µm. Apical clearance may be adjusted by increasing or decreasing the sagittal depth of the lens. In the case of inadequate apical clearance, increase the sagittal depth of the lens the appropriate amount. Tip: The clearance can be estimated by comparing the lens thickness (350 µm at plano) to the tear reservoir under the lens. This can be evaluated by a narrow slit lamp beam moved from limbus to limbus at an angle of 45. Excessive clearance will make it more difficult to insert the lens without air bubbles. In eyes that tend to accumulate debris behind the lens, a shallower sagittal depth should be chosen. A deeper sagittal depth may be necessary in eyes that are prone to progressive ectasia. Insufficient Clearance Lens 0.30 mm Tear Layer 0.11 mm Cornea 0.58 mm If insufficient apical clearance, increase sagittal depth Lens 0.30 mm Tear Layer 0.58 mm Cornea 0.59 mm Excessive Clearance If excessive apical clearance, decrease sagittal depth Tear Layer 0.20 mm Cornea 0.78 mm Lens 0.30 mm Ideal Clearance OCT image 8

Assess Limbal Clearance Tip: If the SLZ and the sagittal depth are correct, most lenses don t require a change in base curve. Assess the limbal clearance with fluorescein, as seen below, or by using white light or an OCT. It is important that there is no touch of the lens on the limbus. Ideal limbal clearance is about 100 µm. This may vary circumferentially. When landing zones and sagittal depths are correct, the limbal clearance is adjusted by changing the base curve. If there is bearing or less than 50 µm of clearance, it s important to re-assess the landing zone. If the SLZ is too flat, as demonstrated by edge lift, you will observe a decrease in limbal clearance. An increase in the SLZ should provide the ideal limbal clearance. Bearing in limbal area. Flatten base curve. Excessive clearance in limbal area. Steepen base curve. Ideal clearance in limbal area. Maintain base curve. Before changing the base curve, check the landing zone. If there is edge lift, increase SLZ number by 2. Tip: Changes in the base curve should always be accompanied by a lens power change. SynergEyes VS lenses are available in base curve increments of 0.2 mm. A 0.2 mm flattening of the base curve requires a power adjustment of +1.00. A 0.2 mm steepening of the base curve requires a lens power adjustment of -1.00. Lens Diameter The 16.0 mm diameter provided in the diagnostic set will suffice for a majority of patients with average to smaller corneas. The lens diameter should be approximately 1.5 to 2.5 mm beyond the limbus or 3.0 to 5.0 mm greater than HVID. Lens diameter is available from 14.0 to 17. 5 mm in 0.5 mm steps. 9

Determine Lens Power Perform an over-refraction after the optimal diagnostic lens has settled on the eye. The power in the diagnostic lenses is plano. The SynergEyes VS is a non-rotational lens that stabilizes over the flat meridian at the hash marks allowing for cylinder availability at any axis. An axis adjustment may be necessary to compensate for position. Hash Mark Hash Mark Tip: All diagnostic and toric lenses have hash marks. They align with the flattest meridian. Locate and assess the hash marks. They align with the flattest meridian. If the lens is correctly fit, the hash marks will be stable and remain in a constant position. To compensate for misalignment, note the axis location of the hash marks and the axis of the over-refraction. Subtract the hash mark axis from the over-refraction axis. If the resulting number is positive, that is the cylinder axis to order (see example). If the resulting number is negative, subtract that number from 180 and that is the axis to order (see formula). Formula: Over-refraction axis hash mark axis = + X Over-refraction axis hash mark axis = X X = Cylinder axis lens order; 180 X = Cylinder axis lens order Example: Hash mark settles at 30 180 axis 90 axis 30 To compensate for misalignment from 180 axis: Hash mark: 30 Over refraction is -1.00-1.00 x 140. 140-30=110 Order: SynergEyes VS: -1.00-1.00 x 110 Hash Mark Tip: SynergEyes Consultation is available to help with adjustments if the location of the hash marks and the toric over-refraction are provided. 10

Troubleshooting Tips Observation Tight lens fit: Blanching, compression or tight edge in landing zone in either the flat meridian or steep meridian Blanching is circular (around the entire lens) Loose lens fit: Too much edge lift or scleral standoff or too much movement Excessive movement Excessive central clearance Insufficient central clearance If mid-perpiheral or limbal clearance too thin or too deep Cylinder over-refraction Visual fluctuations Lens surface deposits Vision is blurry Debris in tear layer Inferior decentration Bubbles under the lens Lens discomfort Recommendation Mild: Decrease SLZ in the specific meridian by 2 (60 µm) Moderate: Decrease SLZ by 4 (120 µm). Note: The change of an SLZ will also change SLZ on the opposite side of the lens. Mild: Decrease SLZ in both meridians by 2 (60 µm). Moderate: Decrease SLZs in both meridians by 4 (120 µm). Mild: Increase SLZ in the specific meridian by 2 (60 µm). Moderate: Increase SLZ in the specific meridian by 4 (120 µm). Note: The change of an SLZ on one side of the lens will also change the SLZ on the opposite side of the lens. If landing zone looks good and the lens has too much movement, increase diameter. Reduce sagittal depth to allow for 200-250 µm of clearance. Increase sagittal depth to allow for 200-250 µm of clearance. Increase or decrease base curve and adjust lens power. Ideal limbal clearance is 100 µm. Note where hash marks line up, order a front toric. Check surface quality (wettability and deposits). Reduce sagittal depth, increase diameter, verify lens stability. Use appropriate cleaning solutions. If deposit is protein, use a protein remover. Check and treat ocular causes (GPC, MGD). If cleaning system is not sufficient, change material. Perform over-refraction to identify residual astigmatism. Order a front toric. Record lens stability in degrees at follow-up visits to follow lens stability over time. Remove, clean and rinse lens if vision is blurred. Improve SLZ alignment to prevent debris from getting under the lens. Check and treat ocular causes (GPC, MGD). Optimize lens solutions. Rinse the eye with saline or other rinsing solution before inserting the lens. Reduce clearance. Minor decentration is acceptable. First, check vertical scleral fit and SLZ alignment. Then, increase diameter. Reinsert lens to eliminate insertion bubbles. Consistent central bubbles can indicate over-vault. Decrease vault. Consistent bubbles in mid-periphery can indicate the base curve is too flat. Steepen base curve. If bubbles arise after correct insertion this may indicate SLZ is too flat. Initial lens will have some awareness, but should not be uncomfortable. Check edge lift. Increase SLZ in the specific meridian. Check central clearance. Increase sagittal depth if clearance is insufficient. 11

Select Lens and Perform Assessment Determine Flat and Steep Meridians Hash Mark Start with a 3600 µm, 36 flat, 42 steep. Observe hash marks which align Perform a quick check for central touch or with the flattest meridian. excessive clearance and SLZ alignment. The steep meridian is 90º away. Ste ep Me ridi an Start Here Quick Reference Guide Fla tm erid ian Settle 30 Minutes Scleral Landing Zone: Observe and Steep Meridian Landing Zones If blanching / no edge lift, decrease SLZ number to loosen lens If excessive edge lift, increase SLZ number to tighten lens Ideal landing zone Next Step Control Sagittal Depth: Observe Apical Clearance Lens 0.30 mm Lens 0.30 mm Tear Layer 0.58 mm Cornea 0.58 mm Cornea 0.59 mm Assess Tear Layer 0.11 mm Lens 0.30 mm Tear Layer 0.20 mm Cornea 0.78 mm If insufficient apical clearance, increase sagittal depth If excessive apical clearance, decrease sagittal depth Ideal clearance OCT Image White light Next Step Limbal Zone: Observe Limbal Clearance If bearing in limbal area, flatten base curve. If excessive clearance in limbal area, steepen base curve. Ideal clearance in limbal area. Maintain base curve. Before changing the base curve, check the landing zone. If there is edge lift, increase SLZ number by 2. Determine Power NOTE: A 0.2mm change in base curve requires a +/ 1.00 D power adjustment. Perform over-refraction. If front cylinder is required, an axis adjustment may be necessary to compensate for rotation. SynergEyes consultation is available to help with adjustments if the location of the hash marks and the toric over-refraction are provided. Lens diameter should be 1.5-2.5 mm beyond the limbus or 3.0-5.0 mm greater than HVID. Diagnostic lens diameter is 16.0 mm. 2017 SynergEyes, Inc. 75270-A