CGuard MicroNet-Covered Embolic Prevention Stent System A Game Changer in Carotid Revascularization

Similar documents
CARDIONOVUM GmbH Am Bonner Bogen Bonn Germany. CARDIONOVUM Sp. z o.o. ul. Panska Warsaw Poland

CoreValve in degenerated CoreValve 1st Re do CoreValve

SeQuent Please NEO. Clinically proven Polymer-free Drug Coated Balloon Catheter. Vascular Systems

Aortic valve replacement: what s the future. Experience in aortic valve replacement with Trifecta bioprosthesis

Challenges in Measuring CSF Flow with MRI

Crosser CTO Recanalization Catheter High Frequency Mechanical Recanalization System

Update on Acute Stroke Management

NAVA Neurally Adjusted Ventilatory Assist In Neonates

Air Embolism During TEVAR: Carbon Dioxide Flushing Decreases the Amount of Gas Released From Thoracic Stent-Grafts During Deployment

ICE & TEE Monitoring in the Cath Lab TEE for ASD/PFO Closure

Lead Extraction Overview

imágenes en el Hospital Clinic

Patent Foramen Ovale and Diving

SITE NAME: ISF DOCUMENT VERSION CHECKLIST Complete yes or no to confirm the presence or absence of each document specified.

Urgent Medical Device Recall Medtronic StrataMR Adjustable Valves & Shunts

S.I.T.E.M.S.H Congress

Use of Oral Contrast Material in Whole Body Trauma CT

AutoPulse. Mechanical CPR. in the Hospital

HUD and HDE Topics. Humanitarian Use Device and. Humanitarian Use Device. Humanitarian Device Exemption

SUPPLEMENTARY APPENDIX. Ary Serpa Neto MD MSc, Fabienne D Simonis MD, Carmen SV Barbas MD PhD, Michelle Biehl MD, Rogier M Determann MD PhD, Jonathan

1 out of every 5,555 of drivers dies in car accidents 1 out of every 7692 pregnant women die from complications 1 out of every 116,666 skydives ended

The Crosser CTO Recanalization Device (Bard

Remodeling or Reimplantation?

*

Lightning Mass Casualty Incident at 13,000 feet. Will Smith, MD, Paramedic, FAWM

Inspired by the challenge to create a better solution for the thoracic aorta, RelayPlus was specifically designed to:

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

The Art of. Flow Restoration. Raiden3. Non-Compliant PTCA Balloon

The Safe Use and Prescription of Medical Oxygen. Luke Howard

DCS CPG: H02.01C TRANSCUTANEOUS PO2 (PtcO2) INTERPRETATION

Hyperbaric Oxygen and TBI: What Does Science Tell Us. Kathleen Bell, MD Department of Rehabilitation Medicine

Acute phase form Year Version

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Adverse effects of anabolic androgenic steroids abuse on gonadal function, glucose homeostasis and cardiovascular function

Hyperbaric Oxygen Therapy

Invasive imaging follow- up of BRS in bifurca8on lesions

Draeger Neonatal Conference Dallas, Texas May 30, Biomedical Applications of PFC. Thomas H. Shaffer, MS.E., PhD

Case Iatrogenic venous air embolism

Self-fulfilling prophecy? Current affairs. Reality check 11/29/2011

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant

Remodeling or Reimplantation?

Diffusion MRI: Basic Principles and Clinical Applications

ASAHI Neurovascular Guide Wire

Institutional Review Board - Restriction s Impo sed

PERFORMANCE ZONES

3100A Competency Exam

Kids save lives in Italia e in Europa. Federico Semeraro Ospedale Maggiore, Bologna,

+ Competitive Intelligence Report: Final Expense Life 3rd Quarter 2010

Inspections, Compliance, Enforcement, and Criminal Investigations

APRV: Moving beyond ARDSnet

How might GRADE work for ILCOR? Summary of specific components of GRADE using example worksheet

Intelligent Ventilation solution from ICU to MRI

UNIVERSITY OF GEORGIA Institutional Review Board

Drug eluting balloon catheter

Humanitarian Use Devices Made Simple

McHENRY WESTERN LAKE COUNTY EMS SYSTEM FALL 2014 CONTINUING EDUCATION MANDATORY FOR ALL PRIMARY AND PROBATIONARY ALS SYSTEM PROVIDERS.

CASE REPORT. Subacute L5 Paresis Following a Capoeira Presentation

Accident Investigation

한국학술정보. Educational Effect on Prehospital Personnel for Prehospital Stroke Management

Why we should care (I)

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

APPENDIX 1 DIVING MEDICAL EXAM OVERVIEW FOR THE EXAMINING PHYSICIAN. Diving Safety Officer

Diagnosis Players, coaches, parents and heath care providers should be able to recognize the symptoms and signs of a concussion:

JUN L WARNING LETTER VIA FEDERAL EXPRESS

Basic - Peripheral Vascular Assessment VASCULAR DOPPLER PHOTO-PLETHYSMOGRAPHY PULSE VOLUME RECORDING ANKLE AND TOE BRACHIAL INDEX

INTERSECTION WITH RIGHT TURN TAPER

Sprinter Family. balloon dilatation catheters. Sprinter Legend OTW Sprinter Legend RX Sprinter OTW

Vancouver 2010 Medical Services Presentation to EMS Chiefs Conference Dr. Jack Taunton May 2008

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

INTERVENTIONAL CARDIOLOGY

Patent Foramen Ovale and Fitness

SCUBA - self contained underwater breathing apparatus. 5 million sport scuba divers in U.S. 250, ,000 new certifications annually in U.S.

Operating Instructions for Microprocessor Controlled Ventilators

Respiratory Signs: Tachypnea (RR>30/min), Desaturation, Shallow breathing, Use of accessory muscles Breathing sound: Wheezing, Rhonchi, Crepitation.

Current Status: Active PolicyStat ID: Origination: 04/2018 Effective: 04/2018 Approved: 04/2018 Last Revised: 04/2018 Expiration: 04/2021

NEUROLOGICAL INSIGHTS FOR TEACHING GOLF TO TODAY S FITNESS CHALLENGED

Audiovisual (AV) biofeedback improves anatomical position management in breath-hold

Instructions/Procedure for Use of the Speaking Valve

Workers Compensation Accident Investigations

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Instructions for Submitting an Exam to Cengage for AAPC CEU Approval

BALLOON. 1. PTA Balloon Catheter 1.1 ALLUNGA Description Technical Data Materials used Compliance Chart

Canadian and Provincial Deceased Donation Data Update

Immersion pulmonary oedema: is it confused with drowning?

G622. APPLIED SCIENCE Monitoring the Activity of the Human Body ADVANCED SUBSIDIARY GCE. Thursday 27 May 2010 Afternoon. Duration: 1 hour 30 minutes

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Cardiac Arrest General

MALL CROSSING STUDY. Testing the Effectiveness Of the 4th Street East Crossing. For: City of Charlottesville Neighborhood Development Services

OHRP Guidance on Written IRB Procedures

UTMB RESPIRATORY CARE SERVICES GUIDELINES Respiratory Care Services Medical Gas Cylinder Storage, Use and Transport

Testimony. The Impact of Hyperbaric Medicine on Government Health Care, Disability and Education Expenditures

New Home Oxygen Contract

3 Revising Diagnostic Imaging PARs

AVALANCHE RESUSCITATION. ALICIA PETERSON, MD Emergency and Wilderness Medicine

Patrick Bitter, Jr., MD, FAAD

Walking for Heart Health in Rural Women

The aorto-ventricular junction in aortic repair

EMERGENCY USE 03/02/2016

You ain t seen nothin yet Critical care informatics

SECOND EUROPEAN CONSENSUS CONFERENCE ON HYPERBARIC MEDICINE THE TREATMENT OF DECOMPRESSION ACCIDENTS IN RECREATIONAL DIVING

Transcription:

CARENET J Am Coll Cardiol 2015;66:B33 CGuard MicroNet-Covered Embolic Prevention Stent System A Game Changer in Carotid Revascularization Piotr Musialek, MD DPhil FESC Jagiellonian University Dept. of Cardiac & Vascular Diseases John Paul II Hospital, Krakow, Poland

CARENET J Am Coll Cardiol 2015;66:B33 CGuard MicroNet-Covered Embolic Prevention Stent System A Game Changer in Carotid Revascularization Piotr Musialek, MD DPhil FESC Jagiellonian University Dept. of Cardiac & Vascular Diseases John Paul II Hospital, Krakow, Poland

CARENET J Am Coll Cardiol 2015;66:B33 CGuard MicroNet-Covered Embolic Prevention Stent System A Game Changer in Carotid Revascularization Piotr Musialek, MD DPhil FESC Jagiellonian University Dept. of Cardiac & Vascular Diseases John Paul II Hospital, Krakow, Poland

Disclosures ABs / Research Support / Consulting / Speaker Bureau Abbott, Balton, InspireMD, Medtronic, Penumbra NB. The PARADIGM study has been Invesigator-Initiated and Investigator-Executed ( no industry support )

CGuard embolic prevention stent P. P. Musialek Musialek @ ICI TCT 2015 2015

P. Musialek @ ICI 2015

Mandatory DW-MRI P. Musialek @ ICI 2015

Mandatory DW-MRI CARENET 03-007, PJ (Krakow) P. Musialek @ ICI 2015 Prior to CAS 24h after 30 d after CAS

Mandatory DW-MRI CARENET 03-007, PJ (Krakow) P. Musialek @ ICI 2015 Prior to CAS Rec.Symptomatic LICA 24h after 30 d after CAS THR Note self-tapering

Mandatory DW-MRI CARENET 03-007, PJ (Krakow) P. Musialek @ ICI 2015 Prior to CAS Rec.Symptomatic LICA 24h after 30 d after CAS PARADIGM & PARADIGM-EXTEND STUDIES THR Note self-tapering

Mandatory DW-MRI CARENET 03-007, PJ (Krakow) Prior to CAS Rec.Symptomatic LICA 24h after 30 d after CAS PARADIGM & PARADIGM-EXTEND THR Note self-tapering Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Mandatory DW-MRI CARENET 03-007, PJ (Krakow) Prior to CAS Rec.Symptomatic LICA 24h after 30 d after CAS PARADIGM & PARADIGM-EXTEND THR Note self-tapering Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Mandatory DW-MRI CARENET 03-007, PJ (Krakow) Prior to CAS Rec.Symptomatic LICA 24h after 30 d after CAS PARADIGM & PARADIGM-EXTEND THR R O U T I N E CLINICAL PRACTICE 2015 + Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

R O U T I N E CLINICAL PRACTICE 2015 + P. Musialek @ ICI 2015

P. Musialek @ TCT 2015

CAS (and CEA) are and will remain emboli-generating procedures P. Musialek @ ICI 2015

CAS (and CEA) are and will remain emboli-generating procedures P. Musialek @ ICI 2015

CAS (and CEA) are and will remain emboli-generating procedures P. Musialek @ ICI 2015

CAS (and CEA) are and will remain emboli-generating procedures P. Musialek @ ICI 2015

CAS (and CEA) are and will remain emboli-generating procedures P. Musialek @ ICI 2015

Δ P. Musialek @ TCT 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

Conventional Carotid Stent K. Mathias 2013 J. Schofer, P. Musialek et al. TCT 2014 P. Musialek @ TCT 2015

Conventional Carotid Stent Human Carotid OCT Image Courtesy Dr Juan Rigla, MD PhD Perceptual Imaging Lab, Univerity of Barcelona P. Musialek @ ICI 2015

P. Musialek @ ICI 2015 ANY data on the incidence of PLAQUE PROLAPSE in conventional carotid stents?

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015 Post-procedural PLAQUE PROLAPSE through conventional stent struts De Donato et al. Eur J Vasc Endovasc Surg 2013;45:579-587.

Conventional Carotid Stent K. Mathias 2013 J. Schofer, P. Musialek et al. TCT 2014 P. Musialek @ TCT 2015

Anti - Embolic Carotid Stent J. Schofer, P. Musialek et al. TCT 2014 P. Musialek @ TCT 2015

P. P. Musialek @@ TCT ICI 2015 CGuard embolic prevention system

Specific, carotid-dedicated design NB. CGuard EPS is not yet available in the US P. Musialek @ ICI 2015

P. Musialek @ TCT 2015

P. Musialek @ TCT 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015 DW-MRI: the unforgiving testimony of what you ve done to the TARGET ORGAN...

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

Anti - Embolic Carotid Stent K. Mathias 2013 J. Schofer, P. Musialek et al. TCT 2014 P. Musialek @ TCT 2015

Anti - Embolic Carotid Stent CGuard Embolic-Prevention Stent OCT Image (human, iv vivo) Courtesy Dr Juan Rigla, MD PhD Perceptual Imaging Lab, Univerity of Barcelona J. Schofer, P. Musialek et al. TCT 2014 P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66(suppl):B33 P. Musialek @ ICI 2015

* Emboshield (n=7); FilterWire (n=14); Spider (n=25) ** Gore FlowReversal (n=4) or flow reversal with MoMa (n=21) (NB. mean flow reversal time was 6min 48s, from 5min 18s to 11min 2s) # ø 4.5mm (n=5); ø 5.0mm (n=36); ø 5.5mm (n=29); ø 6.0mm (n=1); 30mm in 51 lesions; 40mm in 18 lesions (2 other lesions required two stents each) P. Musialek @ ICI 2015

Musialek P et al. Impact of routine micronet-covered embolic prevention stent system use on contemporary carotid revascularization: All-comer PARADIGM Study. JACC 2015;66:B33 P. Musialek @ ICI 2015

Evolving L Haemisph stroke Case # 063 (Krakow)

Case # 063 (Krakow) NO new brain lesions M. Urbanczyk, P. Banys, Dept. Radiology, JP2 Hospital, Krakow, Poland

R Haemisph minor stroke 4 d before, now recurrent TIAs Note self-tapering Case # 067 (Krakow)

R Haemisph minor stroke 4 d before, now recurrent TIAs Case # 067 (Krakow) NO new brain lesions NO new brain lesions M. Urbanczyk, P. Banys, Dept. Radiology, JP2 Hospital, Krakow, Poland

L Haemisph minor stroke 5 d before Case # 068 (Krakow)

L Haemisph minor stroke 5 d before Case # 068 (Krakow) NO new brain lesions NO new brain lesions M. Urbanczyk, P. Banys, Dept. Radiology, JP2 Hospital, Krakow, Poland

L Haemisph minor stroke 5 d before Case # 068 (Krakow) NO new brain lesions NO new brain lesions M. Urbanczyk, P. Banys, Dept. Radiology, JP2 Hospital, Krakow, Poland

L Haemisph minor stroke 5 d before Case # 67 (Krakow) M. Urbanczyk, P. Banys, Dept. Radiology, JP2 Hospital, Krakow, Poland

RICA 6.2/1.5 m/s Highly-calcific I Predilatation 3.0x20mm followed by NC 4.5x15/20atm CGuard 9.0x30mm, postdilated ø5.5x20mm/16atm A. Mazurek, P. Musialek epcr2015

Highly-calcific II NO brain lesions with CAS Predilatation 2.0x20 followed by NC 4.0x15, CGuard 8.0x40mm, postdilated ø 5.0mm/16 atm A. Mazurek, P. Musialek epcr2015

Highly-calcific III Note self-tapering Predilatation 2.5x15mm followed by 4.0x15, CGuard 9.0x30mm, postdilated ø5.0mmx20/24atm A. Mazurek, P. Musialek epcr2015

CGuard P. Musialek @ ICI 2015

CGuard P. Musialek @ ICI 2015

P. Musialek @ ICI 2015 PARADIGM EXTEND

30d data P. Musialek @ ICI 2015 Z E R O Stroke/ MI/death

30d data P. Musialek @ ICI 2015 Z E R O Stroke/ MI/death 12mo data

30d data P. Musialek @ ICI 2015 Z E R O Stroke/ MI/death 2 asymptomatic self-withdrawals @ 30 days 12mo data

30d data P. Musialek @ ICI 2015 Z E R O Stroke/ MI/death 2 asymptomatic self-withdrawals @ 30 days 100% follow up of the remaining patients 12mo data

30d data P. Musialek @ ICI 2015 Z E R O Stroke/ MI/death 2 asymptomatic self-withdrawals @ 30 days 100% follow up of the remaining patients 12mo data Z E R O Stroke Deaths Z E R O Strokes @ 12mo Per-Protocol independent neurological assessment

30d data P. Musialek @ ICI 2015 Z E R O Stroke/ MI/death 2 asymptomatic self-withdrawals @ 30 days 100% follow up of the remaining patients Z E R O Stroke Deaths Z E R O Strokes 1 pulmonary embolism death @ 5 mo 1 respiratory failure death @ 8 mo 1 malignant tumor death @ 9 mo @ 12mo Per-Protocol independent neurological assessment

2 0 1 5 P. Musialek @ ICI 2015.. NO device-related adverse events NO procedure-related events CARENET Multicenter Trial 12 mo data

Peak Systolic Velocity (cm/sec) CARENET in-stent Peak Systolic Velocities 300 70% in-stent stenosis threshold* 250 200 2 0 1 5 150 100 50 0 30 d 6 mo 12 mo ECA patency 100% 100% 100% * Setacci et. Al.. Grading Carotid Intrastent Restenosis of 814 CAS patients Stroke 2008 P. Musialek @ ICI 2015

Peak Systolic Velocity (cm/sec) * Setacci et. Al.. Grading Carotid Intrastent Restenosis of 814 CAS patients Stroke 2008 P. Musialek @ VEITH 2015 CARENET in-stent Peak Systolic Velocities 300 70% in-stent stenosis threshold* 250 200 2 0 1 5 150 100 50 0 30 d 6 mo 12 mo ECA patency 100% 100% 100%

Peak Systolic Velocity (cm/sec) * Setacci et. Al.. Grading Carotid Intrastent Restenosis of 814 CAS patients Stroke 2008 P. Musialek @ ICI 2015 CARENET in-stent Peak Systolic Velocities 300. 250 70% in-stent stenosis threshold* NO in-stent restenosis concern 200 150 100 50 0 30 d 6 mo 12 mo ECA patency 100% 100% 100%

Peak Systolic Velocity (cm/sec) * Setacci et. Al.. Grading Carotid Intrastent Restenosis of 814 CAS patients Stroke 2008 P. Musialek @ ICI 2015 CARENET in-stent Peak Systolic Velocities 300. 250. 200 150 70% in-stent stenosis threshold* NO in-stent restenosis concern NO CGuard ECA patency concern 100 50 0 30 d 6 mo 12 mo ECA patency 100% 100% 100%

CAS (and CEA) are and will remain emboli-generating procedures amenable to elimination with MicroNet P. Musialek @ ICI 2015

P. Musialek @ ICI 2015 Endovascular Solution for All-Comers Note self-tapering Endovascular Reconstruction of the Carotid Bifurcation

P. Musialek @ ICI 2015

DW-MRI Evidence (CARENET) 2015 + Clinical Evidence (CARENET, PARADIGM, PARADIGM-EXTEND) P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

P. Musialek @ ICI 2015

CGuard Embolic-Prevention Stent Image Courtesy Dr Juan Rigla, MD PhD Perceptual Imaging Lab, Univerity of Barcelona P. P. Musialek @@ TCT ICI 2015

CAS 2010 VISION CGuard Embolic-Prevention Stent Image Courtesy Dr Juan Rigla, MD PhD Perceptual Imaging Lab, Univerity of Barcelona P. P. Musialek @@ TCT ICI 2015

CAS 2010 VISION Carotid Revascularization 2015 + R E A L I T Y CGuard Embolic-Prevention Stent Image Courtesy Dr Juan Rigla, MD PhD Perceptual Imaging Lab, Univerity of Barcelona P. P. Musialek @@ TCT ICI 2015

CGuard Embolic-Prevention Stent Image Courtesy Dr Juan Rigla, MD PhD Perceptual Imaging Lab, Univerity of Barcelona P. P. Musialek @@ TCT ICI 2015

Back-up slides

CREST 30 d P. Musialek @ ICI 2015

CREST CREST P. Musialek @ ICI 2015

CREST CREST P. Musialek @ ICI 2015

What would CREST show today CREST with 30 d P. Musialek @ ICI 2015

What would CREST show today CREST with instead of 30 d? P. Musialek @ ICI 2015

Feeding today s CGuard CAS data to CREST CREST 30 d would indicate CAS superiority P. Musialek @ ICI 2015