Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump

Similar documents
The Principles of Vacuum Technology

A novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties

In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks

Development of a high vacuum sample preparation system for helium mass spectrometer

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES

Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

Practical approach and problems in in-situ RGA calibration

Gun Vacuum Performance Measurements February 2003

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

The Helium Leak Detector

Matching Vacuum Pump to Process

Total Pressure Measurement Using Residual Gas Analyzers

School of Chemistry SOP For Operation Of Glove Boxes

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

E SC 412 Nanotechnology: Materials, Infrastructure, and Safety Wook Jun Nam The Pennsylvania State University

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

2 Switching off ECRIS 3. 4 Acquire Beam Spectrum 4. 6 Vent Instrument Chamber 5. 7 Pump down Chamber 5. 8 Baking out Chamber 5.

The use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

The Broadest Gauging Line in the Industry

Selecting the right pressure sensor for your application

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Chapter 8: Cryo-sorption pumps

How To Use Getters and Getter Pumps

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation

Choosing a Quadrupole Gas Analyzer Application Note #9

ATH 500 M. Pumping speed ATH 500 M. 3.1 Vacuum generation / Turbopumps / Magnetically levitated / ATH 500 M. Page 398 / Part 3.1

Hiden QIC Series Gas Analysers Real-time gas analysis systems

Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

MATRIX-MG Series. Innovation with Integrity. Automated High-Performance Gas Analyzers FT-IR

High Performance Mass Spectrometers

How To Choose a Roughing/Backing Pump for the Turbo and Drag Family

APPLICATION OF A CHANNEL PLATE IN FIELD-ION MICROSCOPY

Leak Detection. Kirsten Zapfe, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, Special thanks to S. Holm, N. Mildner and A.

Vacuum System at IUAC

MagneticTurbo TM. TurboDrag pumps with pumping speed up to 1500 l/s TMH/U 1000 M / TMH/U 1600 M. Pumping speed. Start / Chapter content

Underwater cryotrap - membrane inlet system (CT-MIS) for improved in situ analysis of gases by mass spectrometry.

cooled, Heated ATH 2800 MT, DN 250 ISO-F, External drive electronics, Water cooled, Heated

Operating Procedures for the. SAMCO ICP RIE System

1)! DO NOT PROCEED BEYOND THIS MARK

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels

Traysealer T 100. Tray packaging in the smallest space

S.A. Klein and G.F. Nellis Cambridge University Press, 2011

DC gun high voltage conditioning with Krypton gas Carlos Hernandez-Garcia October 2008

Residual Gas Analysis Systems for Industry

ISO/TC 112/WG 2 N 49. ISO TS QMS WD v6. Date of document Expected action Comment Due Date Background

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Plasma Sources and Feedback Control in Pretreatment Web Coating Applications

METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS

Chapter 6 Leak detection

NGN Series nitrogen generator

This educational seminar discusses creating, measuring, and troubleshooting Ultra-High Vacuum ( Torr) systems.

AC : ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET

Series 203 Granville-Phillips Variable Leak Valve

INSTRUCTIONS. VP-100 Ion Pump. Model / R001

Technical Specifications of Hydrogen Isotope Handling and Recovery System

Operating Procedures for Metal Evaporator I

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer

Detecting Hydrogen in Helium Streams. Samuel Goodman. Pittsford Mendon High School. Rochester, New York. Advisor: Dr.

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Granville-Phillips Series 370 Stabil-Ion Vacuum Gauge Controller

SAES Packaging service. making innovation happen, together

Application Note. ASTRON Remote Plasma Source Ignition Best Practices PROBLEM

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

Best Practice for Identifying Leaks in GC and GC/MS Systems

PHOENIX L300i Best-in-class Helium Leak Detectors

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Proportional valves with piezo technology in medical technology

Leak Checking Large Vacuum Chambers

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

Major Design Features

Some tentative AMS improvements (and a note on IE calibrations)

Quadruple mass spectrometers (transducer type)

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

[Graphic Division] 10

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Vacuum Pumps. Turbomolecular Pumps F31 F32 F33 F34 F35 F36 F37 F38 F39 F40

Gases. Properties of Gases Gas Pressure

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

STANDARD OPERATING PROCEDURES

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

The latest automotive systems require innovative leak test methods and fixturing.

A Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide

DIAPHRAGM PUMPS WITH KNF STABILIZATION SYSTEM

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

/ /

ZUERCHER TANK REGULATORS PRINCIPLES OF TANK VENTILATION

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

DRAFT. Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35

VACUUM REGULATORS CONTENTS

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Transcription:

8 th Harsh Environment Mass Spectrometry Wokshop Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump Sept 22, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips Vacuum Instrumentation, Brooks Automation, Inc. Longmont, CO Bob Garcia SAES Getters Group Colorado Springs, CO 1

Less is More... The Workshop on Harsh-Environment Mass Spectrometry (HEMS) was created in 1999 as a means of encouraging interaction among people involved in deploying mass spectrometers outside of the typical laboratory setting. These environments are diverse, ranging from volcanoes and battlefields, to ocean depths, outer space and other rugged locales. Building mass spectrometers to withstand the rigors of such harsh and remote environments places a unique burden on engineering design and science objective planning, where operational requirements for power, size and durability must be met while achieving the goals of the scientific mission. Source: http://www.hems-workshop.org/aboutus.html 2

Win some. Loose the rest! ART MS Sensors are characterized by their excellent sensitivity, fast scan rates, rugged design but particularly by their small size (scalable) and low power requirements. Mass spectrometers must operate under high vacuum - Vacuum pumps are required system components for atmospheric gas samplers relying on ART MS sensors. Vacuum pump stations are generally bigger and/or require more power than the ART MS sensor itself. There is a need to identify/develop high vacuum pumps that match or beat the size and power requirements of ART MS sensors. 3

The Power/Size Conundrum.. ART MS Single Board Controller Courtesy: Gary McMurtry, Pace Tech, Hawaii VQM 830 Sensor (7.5W) Turbo Pump Station (50W) 4

High Vacuum Pump for ART MS High Pumping Speed >5L/s Low Power Small size Light Weight Large Gas Capacity Inert Gas Pumping (for air sampling) Low base pressure (for low background) Species Independent Pumping Speed 5

High Vacuum Pump Technologies Three basic technologies considered: Turbo Pump Ion Pump Non-evaporable Getter Pump 6

High Vacuum Pump Technologies Pump Speed Power Size Weight Capacity Back Pump Turbo >40 Species independ ence Inert Gas Ion >5 30Lb @ 60L/s NEG >100 CH 4 NEG + ION Do we need to build the NEG-Ion pumping system? 7

NEXTorr - NEG Ion Pump SAES Getters NEG = Non-Evaporable Getter Sorption Pump NEXTorr D 100-5 2010-2012 SAES Getters S.p.A. NEXTorr and SAES are registered trademarks property of SAES Getters S.p.A High Pumping Speed for all active gases: NEG 10-100L/s Adequate pumping speed for Inert gases and methane: Noble ION >6L/s Adequate capacity (gas and pressure dependence) Minimal Power requirement during operation (5KV, mw) Extremely compact and light (0.5L, 2.2Kg) UHV ultimate pressure limit (E-9 Torr) Available controller provides automated NEG activation (500ºC, 1 hour) and independent ion pump control. Multiple Reactivations (>50) Total Pressure measurement 8

(NEXTorr VQM 830) Gas Analyzer Single-Stage Atmospheric Gas Sampling System Gate Valve (to TP system) NEG-Ion Pump NEXTorr D 100-5 Variable Leak Valve G-P Series 203 ART MS Sensor VQM 830 Gas Inlet Cross: 6-way Full Range Pressure Gauge G-P Series 390 9

(NEXTorr VQM 830) Gas Analyzer Single-Stage Atmospheric Gas Sampling System TP Station Gate Valve Gas Sample NEG-Ion Pump NEXTorr D 100-5 6-way Cross High Vacuum ART MS Sensor VQM 830 ATM Gas Inlet Variable Leak Valve G-P Series 203 Pressure Gauge G-P Series 390 10

NEXTorr Activation NEXTorr s NEG must be activated in order to provide full pumping speed and capacity: Gate Valve open, Leak Valve closed Ion Pump Off SAES NIOPS-02 Power Supply automates Activation (9V, 5A, 1 hour) Built in heating element (500ºC) TP provides High Vacuum conditions Close Gate Valve at end of activation Turn ION Pump on at end of activation NEXTorr is ready! Start NEG Only NEG+ION NEXTorr 5e-8 Torr NEG Only Base pressure 6.5E-9 Torr Stop 11

Base Pressure NEXTorr 1400 1200 1000 H 2 H 2 O Post Activation Base Pressure Spectrum NEXTorr (700mW) - P= 6.5e-9 Torr Very Clean Base Pressure Spectrum Ar and CH 4 are efficiently pumped out. 800 600 400 200 CO Ar CO 2 No Hydrocarbons 0 0 20 40 60 80 100 120 12

NEXTorr - NEG Only Operation = No Power 1800 Start 1600 1400 1200 CH 4 Post Activation Base Pressure Spectrum NEXTorr NEG Only P= 1.3E-7 Torr Ar ION Off Total Pressure CH 4 Ar 1000 800 600 400 200 0-5 15 35 55 75 95 115 135 13

NEXTorr - NEG Only Operation NEXTorr NEG Only Operation requires no external power but provides no pumping speed for inert gases and reduced speed for methane. NEXTorr NEG Only operation not recommended for Air sampling- Ar and methane build up in the chamber-overpressure. Start Air Flow NEG Only Total Pressure Ar CH 4 14

Air Sampling- Gas-Dependent Pump Speed Air sampling not recommended with NEXTorr NEG Only Operation- Ar buildup. Air Sampling is possible with NEXTorr. Gas dependency of pumping speeds must be considered for gas compositional analysis Gas Oxygen 140 Nitrogen 40 Ar 6 Speed (L/s) 2000 1500 1000 500 Air Sampling - P= 5.4E-7 Torr NEXTorr (184 mw) N 2 O 2 Ar Gas Dependent Pumping Speeds: Oxygen underestimated Argon overestimated 0-5 15 35 55 75 95 115 135 15

Direct Sampling Mass Spectrometry NEXTorr /ART MS combination can be used for direct sampling Mass Spectrometry. The buffer/carrier gas must be chosen carefully- i.e. no Argon! Choose high pumping speed and high capacity gas. 2500 2000 1500 1000 500 Changes in background gas composition must be expected 0 Nitrogen Gas sampling N 2 Argon levels rise as Nitrogen is introduced. Ar -5 15 35 55 75 95 115 135 16

Organic Gas Sampling Acetone in N 2 Volatile organic compounds can be sampled. Fragmentation patterns do not suggest any cracking at Ion Pump s high energy discharge. Need to understand species dependence of capacity, differential pumping speed and # reactivation cycles. 1400 1200 1000 800 600 400 200 0 Acetone in Nitrogen NEXTorr P = 2E-7 Torr 15 43 IPA impurity 58 0 10 20 30 40 50 60 70 17

Transportation under High Vacuum A sealed High Vacuum system with an activated NEXTorr will remain at HV for an extended period of time with Ion Pump off ( Tested, 10 days): 1. Activate NEG 2. Seal the vacuum system- 3. Ship - 4. Turn on ion pump. 15 hour trend: sealed sampling system with NEG Only- Total Pressure holds steady at 2.3e-7 Torr. NEG Only- Sealed System Ion Pump On NEG Only P= 2.3e-7 Torr-15 hours NEG + ION NEG+ ION P= 1.3e-9 Torr 18

Build or Buy? Price Synergy (*) Build Buy Size Magnetic Field Flexibility Connectivity (vacuum and electrical) Control (*) Getter pump must intercept impurities from Ion Pump. 19

Final Comments/Recommendations NEXTorr Pump viable for MS Gas Sampling- pumping speed, capacity and reactivation. NEXTorr Pump matches size, power and weight of ART MS sensors. NEXTorr pump provides continuous power free pumping during transportation. Gas specific pumping speeds and capacities must be considered/understood. Choose carrier gas wisely (pumping speed and capacity) NEXTorr NEG Only Operation incompatible with Air sampling. NEG temperature provides additional variable to control gas species pumping speed. 20

Questions? Visit the adjacent Brooks Automation or SAES Getters booths for a live demo of Low-Power Atmospheric Gas Sampling System Based on VQM 830 (ART MS) Sensor and NEXTorr (NEG+Ion) Pump. 21

Technology [pumping speed] Turbo Pump [40-100L /s] Ion Pump [5-60 L/s] Non Evaporable Getter (NEG) [100-1000 s L/s] Pro High Pumping speeds High Compression ratios Throughput pump w/out capacity limits Large flow Shrinking product Rugged (dual bearings) Fairly species independent Large pressure range Low power [??W] No exhaust Decent pumping speed UHV base pressure Vibration free Does not require Backing pump Good for H2 Bakeable Rugged No power requirement after activation High pumping speed Small size Light Dose not require backing pump Cons Large dimensions Heavy High power consumption (>10W) Require backing pump (more power, 35-50W) Light gas problems (H2 and He) Exhaust problems Heavy @ high pump speed (30Lb for 60L/s) Bulky @high pump speed Magnetic fields High Voltage Discharge Capacity limitation Requires HV to start (primer pump) Species dependent Inert gases require noble design ($$) Gas cracking UHV pumps Low flows Requires activation under HV UHV pump Limited capacity Species dependent Low pumping speed for Inert gases and methane Capacity limit 22