Clinical and Statistical Evaluation of 1 st Automatic: A Pilot Study

Similar documents
A method for mathematical arterialization of venous blood gas

Converting venous acid base and oxygen status to arterial in patients with lung disease

Essential Skills Course Acute Care Module. Respiratory Day 2 (Arterial Blood Gases) Pre course Workbook

EMERGENCY MEDICINE AND THE LABORATORY. Ehsan Bolvardi(MD)

Solutions for transcutaneous monitoring

BLOOD GAS (ARTERIAL AND VENOUS

Preanalytical errors in blood gas testing

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Arterial Punctures Blood Gas Collections (ABG)

Product description concerning v-tac

Specifications SIMPLY DEDICATED

Section Two Diffusion of gases

Point-of-Care Testing: A Cardiovascular Perfusionist s Perspective

AQAI SIS MODULES AND MODELS PAGE 1. AQAI Simulator Interface Software AQAI SIS

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future

Blood Parameter Monitoring System 550

Entry to the premium class!

Anaesthetic infrastructure and supplies

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

CDI System 500. Blood Parameter Monitoring System. Continuous blood parameter monitoring for improved blood gas management

Analysis of the relationship between arterial and peripheral venous blood gasses. ABSTRACT

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Organis TestChest. Flight Simulator for Intensive Care Clinicians

Respiratory System Study Guide, Chapter 16

Gas Exchange in Animals. Uptake of O2 from environment and discharge of CO2. Respiratory medium! water for aquatic animals, air for terrestial

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

By: Aseel Jamil Al-twaijer. Lec : physical principles of gas exchange

SenTec Digital Monitoring System Illuminate Ventilation and Oxygenation in Neonatology PCO2 PO2

Q U A L I T Y C O N T R O L F o r u s e w i t h t h e R N A M e d i c a l, I n c. QC 823 Blood Gas/Electrolyte Controls

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:

Public Assessment Report Scientific discussion. Lung test gas CO (He) AGA, 0.28%, inhalation gas, compressed (carbon monoxide, helium) SE/H/1154/01/MR

Preanalytical issues related to blood sample mixing

New A7 Anesthesia Workstation

Subj: HYPERBARIC OXYGEN TREATMENT IN NAVY RECOMPRESSION CHAMBERS

Respiratory physiology II.

IEC C LASSIFICATIONS:

Blood Gas Interpretation

Cartridge Based & Point-Of-Care Blood Gas Analyser QAP Performances & Quality Management Systems.. Issues

Chest Drains. All Covenant Health Intermediate Care Nursery staff. Needle Aspiration

Avalanche Victim Resuscitation Checklist

TRANSCUTANEOUS CO2 MONITORING (TCpCO2) PURPOSE

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

ph is defined as the hydrogen ion concentration and its activity is measured by ion-selective electrode potentiometry.

Image-Based Measurement of Alveoli Volume Expansion in an Animal Model of a Diseased Lung

HemoCue Hb Procedure Template

ABL80 FLEX CO-OX Specifications

Urgent Hyperbaric Oxygen Therapy

Introducing Astral. Greater freedom. Confident care. Designed efficiency

Ventilating the Sick Lung Mike Dougherty RRT-NPS

See what really matters Award-winning Interface of Swisstom BB 2

49 Arterial Blood Gases

Oxygen status algorithm, version 3, with some applications

Lung Volumes and Capacities

Respiratory Physiology. Adeyomoye O.I

Committee on Issues in Organ Donor Intervention Research. Presented by: Richard D. Hasz Vice President, Clinical Services

fabian Therapy evolution Our 2-in-1 device

INTELLiVENT -ASV. The world s first Ventilation Autopilot

3100A Competency Exam

Fingertip Pulse Oximeter

The Clinical Significance of Patient Specimen Transport Modality: Pneumatic Tube System Impact on Blood Gas Analytes

Supporting you in saving lives!

Applied Physics Topics 2

Assessing the ABL 500 blood gas analyser

Standards and guidelines for care and management of patients requiring oxygen therapy.

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

Please, after a carefully reading of given references, answer the following questions:

Test lungs from IMT Analytics simulate a range of realistic ventilation situations during training.

Blood gas analysis: effect of air bubbles in syringe and delay in estimation

Oxygen and Carbon dioxide Transport. Dr. Laila Al-Dokhi

Collin County Community College. Lung Physiology

Stat Profile Prime Critical Care Analyser

Alveolus and Respiratory Membrane

Respiratory Medicine. A-A Gradient & Alveolar Gas Equation Laboratory Diagnostics. Alveolar Gas Equation. See online here

SEP-1 Additional Notes for Abstraction for Version 5.0b

Medical Instruments in the Developing World

CHAPTER 3: The cardio-respiratory system

ANZAI AZ-733VI. Respiratory Gating System. since 1976

NROSCI/BIOSC 1070 and MSNBIO 2070 Exam # 2 October 28, 2016 Total POINTS: % of grade in class

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

Introduction to ICD-10-CM. Improving the Financial Health of the Practices we Serve.

Rodney Shandukani 14/03/2012

CHAPTER 3: The respiratory system

Lab 17. The Respiratory System. Laboratory Objectives

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

S5 TM. Min.I. Mini Maxi Me. The proven safe and reliable, world-leading perfusion system, now optimized for minimally invasive and pediatric surgery

UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION MUST, AT ALL TIMES, ACCEPT THE ENTIRE CARDIAC OUTPUT

SUBCUTANEOUS GAS EQUILIBRATION IN

MEDAP-MOBILE GAS SUPPLY

ScaleChem Basics. Revised 29 May This material may not be printed without the following acknowledgement page.

ADVANCED SUBSIDIARY GCE G622 APPLIED SCIENCE

- How do the carotid bodies sense arterial blood gases? o The carotid bodies weigh 25mg, yet they have their own artery. This means that they have

VDU Lifting/Handling COSHH Spillage Disposal Sharps * * * *

HAMILTON-T1 Intelligent transport ventilation for armed forces

Mechanical Ventilation

Intelligent Ventilation solution from ICU to MRI

The Safe Use and Prescription of Medical Oxygen. Luke Howard

MINI- COURSE on Management of OXYGEN in babies with RESPIRATORY DISTRESS

Transcription:

Clinical and Statistical Evaluation of 1 st Automatic: A Pilot Study Anne Thilde ANDERSEN, Nikolaj Børty NIELSEN, Michael TRUDSLEV, Behamin BAKHSHAIE, Shona PEDERSEN, Ole HEJLESEN Department of Health Science and Technology, Aalborg University, Denmark Department of Clinical Biochemistry, Aalborg Hospital, Denmark

The patient s problem Patients with respiratory disease are at risk of inadequate lung ventilation and inadequate tissue oxygenation Patients with metabolic diseases are at risk of acid-base abnormalities An arterial blood gas analysis is a simple procedure that can be performed bedside and can provide information about lung ventilation, tissue oxygenation and acid-base status

The clinician s problem Medical personnel at the ICU are under considerable stress in their everyday clinical settings and therefore prone to make errors 75% of all laboratory medical errors occur during the pre-analytical phase An error in the pre-analytical phase of an arterial blood gas analysis can seriously influence the diagnosis and treatment

Errors Common pre-analytical errors Wrong patient identification Air bubbles in the specimen Insufficient mixing prior to analysis

1st Automatic Partial automation developed by Radiometer A partially automated bar code scanning system minimizes the risk of patient ID and specimen mix up significantly Automatic mixing of the specimen ensures a consistently homogeneous specimen prior to analysis

1 st Automatic SafePICO Needle cover Pre-heparinised syringe Integrated metal ball SafeTIPCAP Unique barcode Kilde: www.radiometer.com/1st

1 st Automatic FLEXQ 1. Scanning 2. Identification 3. Mixing 4. Analysing

Objective of the study To assess whether 1st Automatic is a suitable alternative to the conventional method To examine whether there is a difference between the two methods

Patients and Specimens Department of Anaesthesia and Intensive Care, Aalborg Hospital 5 patients randomly selected 21 pairs of specimens one specimen processed by the conventional method and one specimen processed by 1st Automatic

Blood Gas Analysis The nurse collected both specimens Both specimens in a pair was analyzed by the same analyzer (ABL825 / ABL837) One of the specimens in each pair was processed by the mounted FLEXQ module (performed by the observer) the other specimen was processed manually (performed by the nurse) Nine parameters: ph, pco2, po2, cna+, ck+, cca2+, cglu, clac and thb.

Comparison of Procedures Bland-Altman plots depict the difference in the blood gas parameters for each pair of specimens (conventional method 1st Automatic) as a function of the mean of the two specimens in a pair Paired-sample t-tests A mean difference of 10%, or less, of the normal range for each parameter was considered to be clinically insignificant

Results 21 paired samples from five patients. The table shows the p-values of the T tests. The blood gas pressures (pxx) are in kpa and the blood concentrations (cxx) in mmol/.

Discussion ph t-test gives p = 0.082, which indicates that there might be a bias However, the mean difference is around 1% of the normal range pco2 and po2 t-test shows p = 0.17 for pco2, and p = 0.94 for po2, which indicates that no bias Mean difference is less than 2% and 1% respectively

cna+ t-test shows p = 0.008, which is statistically significant The mean difference is 0.714, which is 7% of the normal range ck+ t-test shows p = 0.80 Mean difference of less than 1%, which indicates that that no bias is present

cca2+ and cthb t-test shows p = 0.002 for cca2+ and p = 0.010 for cthb the difference being 10% and 1% respectively cglu and clac t-test shows p = 0.065 for cglu and p = 0.25 for clac The difference is 5% and 3% respectively

Summary There was a good consistency between the results using 1st Automatic and the conventional method. It was not possible to show any clinically significant difference between the two procedures. The consequences on workflow using the two procedures have not been addressed in the present study 1st Automatic is now beeing used at the hospital

Acknowledgements The FLEXQ, safepico and 1st Automatic clients were provided by Radiometer Denmark Thanks for their help to Jan Pedersen at Department of Clinical Biochemistry, Aalborg Hospital, to the staff at Thorax Intensive Care Unit, Department of Anaesthesia and Intensive Care, Aalborg Hospital, and to Soren Lundbye-Christensen, Cardiovascular Research Centre, Aalborg Hospital