Taro Takahashi, Colm Sweeney, S. C. Sutherland. Lamont-Doherty Earth Observatory of Columbia University Palisades, NY

Similar documents
Methods Used in Measuring Surface Seawater pco 2 Aboard RVIB Nathaniel B. Palmer

REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING February - March (RVIB Nathaniel B. Palmer Cruise 01/2)

REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING July (RVIB Nathaniel B. Palmer Cruise 00/5)

F:\Ships\NBPalmer\2001\nbp01_3\surface\NBP01_3_data_report.doc Page 104/26/17 4:13 PM

F:\Ships\NBPalmer\2003\nbp03_4a\surface\NBP03_4A_data_report.doc

F:\Ships\NBPalmer\2002\nbp02_2\surface\NBP02_2_data_report.doc Page 104/25/17 3:33 PM

F:\Ships\NBPalmer\2001\nbp01_4\surface\NBP01_4_data_report.doc Page 106/02/15 11:21 AM

F:\Ships\NBPalmer\2002\nbp02_1\surface\NBP02_1_data_report.doc Page 103/19/15 11:16 AM

REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING February - April RVIB Nathaniel B. Palmer Cruise 03/4

Part 1 - Data Processing and Quality Control.

F:\Ships\lmgould\2003\lmg03_2\surface\LMG03_2_data_report.doc Page 107/27/17 12:55 PM

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

Part 1 - Data Processing and Quality Control.

F:\Ships\NBPalmer\2003\nbp03_1a\surface\NBP03_1a_data_report.doc Page 104/24/17 2:54 PM

Part 1 - Data Processing and Quality Control.

COMPARISON OF SURFACE WATER pco 2 AND SST DATA DURING AN ENCOUNTER OF RVIB PALMER AND RV GOULD ON MARCH 13, 2009

Under way pco 2 System Users Manual

Generating Calibration Gas Standards

WHP Operations and Methods { July William M. Smethie, Jr. Lamont Doherty Geological Observatory Palisades, New York U.S.A.

Uniform format surface fco 2 database Benjamin Pfeil and Are Olsen Bjerknes Centre for Climate Research, Bergen, Norway

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Technical Report by Peter Guenther and Charles D. Keeling Carbon Dioxide Research Group Scripps Institution of Oceanography La Jolla, CA

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

Sensor Stability. Summarized by Breck Owens David Murphy, Sea-Bird Electronics AST-14 Wellington March 18-21

This test shall be carried out on all vehicles equipped with open type traction batteries.

CTD Data Documentation

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Columbus Instruments

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Air Pycnometer. Operating Manual. Umwelt-Geräte-Technik GmbH

Accuracy and Stability of Sea-Bird s Argo CTD and Work Towards a Better CTD for Deep Argo. David Murphy Sea-Bird Electronics

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Best Practice Guide, Servomex 2700

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Under pressure pushing down

Cover Page for Lab Report Group Portion. Pump Performance

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

Lung Volumes and Capacities

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

(12) United States Patent

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

LEAP CO 2 Laboratory CO 2 mixtures test facility

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

MS.RAJA ELGADY/PRESSURE PAPER 3

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Experiment. THE RELATIONSHIP BETWEEN VOLUME AND TEMPERATURE, i.e.,charles Law. By Dale A. Hammond, PhD, Brigham Young University Hawaii

HS Bubbler Technology. HyQuest Solutions New Zealand User Group August 2017

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

What I have learned about SF 6 gas testing.a Practical explanation

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Chapter 10 Gases. Characteristics of Gases. Pressure. The Gas Laws. The Ideal-Gas Equation. Applications of the Ideal-Gas Equation

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels

Operation manual Oxygen sensor

Transfeu WP2 Subtask Development of small-scale test method for fire effluents

Laser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C

APPLICATION NOTE. GC Integrated Permeation Device

GCMSD-Headspace Analysis SOP

Helium Mass Spectrometric Leak Detection In Large Size Process Plants

Module 9. Advanced Data Processing

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

Kinetic-Molecular Theory

CORESTA RECOMMENDED METHOD N 6

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Figure Vapor-liquid equilibrium for a binary mixture. The dashed lines show the equilibrium compositions.

TR Test Report Complementary Testing RKN e1 Container (+5 C) Protection by Insulation Version: 01 Page: 2 (22)

Engineering Note. Algorithms. Overview. Detailed Algorithm Description. NeoFox Calibration and Measurement. Products Affected: NeoFox

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Experiment 8 GAS LAWS

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Fall 2004 Homework Problem Set 9 Due Wednesday, November 24, at start of class

Transmitter CS 21 Operation Manual

Standard Test Methods for Water Vapor Content of Electrical Insulating Gases by Measurement of Dew Point 1

Application Note AN-107

Boyle s Law: Pressure-Volume Relationship in Gases

Sampling Considerations for Equilibrium Dissolved Oxygen [DO] Sensors

Oxygen Contamination

Ideal gas law. Introduction

States of Matter Review

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Straw CO 2 Leak Testing

Gases and Pressure SECTION 11.1

Procedures for correcting in situ CTD data and results obtained during the NEAR-GOOS Cross-Basin Climate Monitoring Section project

Transcription:

1 REPORT OF UNDERWAY pco 2 MEASUREMENTS IN SURFACE WATERS AND THE ATMOSPHERE DURING THE ANTARCTIC PACK ICE SEALS EXPEDITION, DECEMBER 1999 JANUARY 2000, IN THE ROSS AND AMUNDSEN SEAS - (RVIB Nathaniel B. Palmer Cruise 99/9) 1. General 1-a) The Cruise: Taro Takahashi, Colm Sweeney, S. C. Sutherland Lamont-Doherty Earth Observatory of Columbia University Palisades, NY 10964 23 May, 2017 RVIB Nathaniel B. Palmer departed Lyttleton, New Zealand on 20 December 1999, (yearday 354) and arrived at the ice-edge off McMurdo Station, Antarctica on 10 February 2000 (yearday 41). Surface pco 2 data was collected from the time the ship reached 200 miles off New Zealand until arrival at McMurdo. Our pco2 system was energized, and began recording data at 0855 UTC on 21 December (yearday 355) and shut down at 1803 UTC on 9 February 2000 (yearday 40). A total of 33,050 seawater, and 986 air VCO 2 observations were recorded during this time. After editing, these 28,923 seawater values remained. The ship s thermosalinograph (TSG) was operating during this time, and the 1- minute sampled data were used to measure surface temperature and salinity. Shipboard measurements of wind speed, wind direction, barometric pressure, position, and heading were also obtained from the ship s main data acquisition system (RVDAS). This cruise crossed the Antarctic Circumpolar current along 175 E longitude, and entered the Ross Sea near 65 S and 175 E. The bulk of the cruise covered the Amundsen Sea between 160 W and 120 W, remaining south of 65 S. The last part was a transit across the Ross Sea near latitude 77 S. In May 2017 we completely re-edited this file with the skills developed in the year before 2017. We have applied the revised algorithm for correcting for the time lag between water passing the remote temperature probe and arriving at our equilibrator. We move the pco 2 data one scan earlier. 1-b) The pco 2 data: The primary source of data are files created by the ships data acquisition system, (RVDAS). A program runs to combined pco 2 data, TSG data, and several other parameters, such as latitude, longitude, windspeed, water-flow in the uncontaminated seawater line, etc. For this cruise there were 2 days (yeardays 357 and 358, 23 & 24 December 1998), where TSG data was missing from the RVDAS files. We used raw pco 2

2 and TSG data files for this period, and the position is from our GPS receiver in the pco 2 system computer. Thirty-three seawater and one air observation are made between standard sets. Each observation involves flowing the equilibration gas for 180 seconds. We also flow the air for 180 seconds before stopping for an observation. Details on these and other aspects of the analysis system are explained later. Standards: Cylinder Concentration (ppm) Nitrogen (UHP) 0.0 CC02231 236.39 CC02218 109.95 CC15551 362.8 CA02205 495.18 A 4 th order calibration curve is used with all 5 standards, as described in Section 2. Standards flow for about 70 seconds before an observation is made. After rather extensive editing, 27,547 seawater observations were retained. A FoxPro program (editpco2.prg) performed the editing process. 1-c) AIR data: The raw VCO 2 of air shows many positive excursions, and also a latitude gradient across the Antarctic Front. North of 58 N we estimated a constant value of 368.2 from the raw data. South of 67 S we estimated a constant value of 367.2. In between across the Antarctic Front, we used the equation: VCO 2 = 0.13 * Latitude + 376.1. Then using the SST, salinity (34.0 where no salinity is available), and barometric pressure, we calculate pco 2 of air, and pco 2 (seawater air). 1-d) Editing: Nitrogen flow dropped to near zero for an extended period from yearday 23.82140 (23 Jan) through yearday 24.78520. Because this flow also passes through the drying column, we had to reject the data for the entire period. The gas could not be assumed to be dry. We rejected the data where the equilibrated gas flow dropped below 27 ml/min. This occurred several times, for extended periods: yearday 359.62110 to yearday 360.00092 yearday 360.69882 to yearday 361.13248 Also for shorter periods during the time from yearday 3 through yearday 15.

3 Thermosalinograph and SST: Generally the thermosalinograph temperature is used as the sea surface temperature. From yearday 1.98348 until yearday 2.03165 we used the equilibration temperature and a constant offset of 0.61 C using the observations before and after the data dropout. Another period from yearday 37.00114 until 37.99903 an offset of 0.51 C was used, again using the observations before and after the data dropout. Waterflow less than 40 l/min also resulted in data being rejected. This occurred for only brief periods, less than a few minutes duration. There is a period where Equilibration temperature was faulty. This was apparently due to a short across the leads of the electronic temperature probe. It began on yearday 20.74886, and was corrected at 26.24412. During this period, equilibration temperature is estimated using the TSG temperature and a constant offset of +0.42 C. We use the period before and after the TSG temperature dropout to estimate the offset. After all these procedures, there still about 59 periods that were rejected by inspecting the data. The criteria were generally that the system had not returned to equilibrium as observed by the VCO 2 value settling to a more constant value. Further, there are periods where the temperature spikes up, indicating an interruption in the waterflow. While not observed in the waterflow data, they were still enough to effect the VCO 2 data, and it is not clear whether the system was in equilibrium during these brief periods. They are generally 5-10 observations wide, but some are as narrow as a single observation. The program editpco2.prg contains all of these edits. Salinity was edited in only a rudimentary manner. The system here is very sensitive to bubbles in the water stream, and even when all the temperature anomalies are eliminated, there may be some salinity spikes remaining. I did apply a simple filter rejecting values below 33.25 and above 34.50, but that is the only salinity-specific editing that was performed. 1-f) Temperature Calibration: Jay Ardai did some simple calibration experiments with the various thermometers with the following results. These experiments were performed on 29 March 2000 after cleaning and reinstalling the RTD sensor into the pco 2 equilibrator. Ice-water Bath (29 Mar 2000): Platinum Therm. Glass (#62774) RTD TSG 0.007 to 0.011 0.020 0.25 to 0.26 0.005 0.015 0.26 At date_time 2000088.4541: 0.78 0.96 0.63 (1105 UTC, day 088) Across Drake Passage: 2.54 2.69 2.391

4 2. MEASUREMENTS OF pco 2 IN SURFACE WATERS 2-a) The LDEO Underway System for Surface Water pco 2 Measurements: The system for underway measurements of pco 2 in surface waters consists of a) a water-air equilibrator, b) a non-dispersive infra-red CO 2 gas analyzer and c) a data logging system. The measurement system is schematically shown in Fig. 1, and is similar with the one described in Bates et al. (1998). Each of these units and the data reduction procedures used will be described below. Figure 1 - The underway pco 2 system used for the measurements of pco 2 in surface waters during the Southern Ocean JGOFS (AESOP) Program. 2-b) Water-air Equilibrator: The equilibrator has a total volume of about 30 liters and is equipped with a specially designed drain which maintains automatically the level of water in the equilibrator at a constant level at about half the height of the equilibrator leaving about 15 liters of headspace. Seawater from the ship s uncontaminated water line is continuously

5 pumped into the equilibrator at a rate of about 10 liters/min, giving a mean residence time of water in the equilibrator of about 1.5 minutes. The headspace above the water serves as an equilibration chamber. A carrier gas (commonly marine air) is drawn into the chamber by a diaphragm pump, and exchanges CO 2 with a continuous flow of seawater sprayed into the chamber through a shower head. Because of large gas-water contact areas created by fine water droplets as well as gas bubbles in the pool of water, CO 2 equilibration between the carrier gas and seawater is achieved rapidly with a e-folding time of 2 to 3 minutes. Under normal operating conditions, the carrier gas in the equilibration chamber is pumped into the infra-red gas analyzer at a rate of about 50 ml/min. At this rate, the residence time of the carrier gas in the equilibration chamber is about 300 minutes, that is about 100 times as long as the equilibration time. Therefore, the carrier gas in the head space is always in equilibrium with water. The over all response time of the equilibrator system has been estimated to be of an order of several minutes. The large volume of water in the equilibrator is chosen in order to have a large thermal inertia of the equilibrator, so that the effects of room temperature changes on the equilibration temperature may be minimized. The temperature of water in the equilibrator is monitored continuously using a Guildline platinum resistance thermometer (readable to 0.05 o C) and recorded on the data logging computer. A calibrated mercury thermometer is also inserted in the equilibrator for testing the performance of the platinum thermometer. At the gas intake end of the equilibrator, a flow indicator based on U-tube manometer is attached. This gives a visual confirmation for the fact that marine air is taken into the equilibration chamber at a desired flow rate. Since we operate the system with the equilibration chamber at the same pressure as the ambient room pressure, the total pressure, at which the gas was equilibrated, is measured using a precision electronic barometer (Setra Model 270, Action, MA) outside the equilibrator. This equilibration pressure is also logged on the computer. The temperature and salinity of seawater at the in situ conditions were measured using a SeaBird Model SBE-21 thermosalinograph aboard the N. B. Palmer and a SIO/ODF thermosalinograph unit based on Neil Brown sensors aboard the R. Revelle. The precision of the report temperature data has been estimated to be about 0.005 o C. 2-c) Infra-red CO 2 Gas Analyzer: The equilibrated gas was passed through a water trap (to collect aerosols and condensates), mass flow controller and a reverse flow naphion dryer (PermaPure flushed with pure nitrogen gas) to remove water vapor (to a level of 20 o C), and was introduced into the IR sample cell at a rate of about 50 ml/min for CO 2 determinations. A LI-COR infra-red gas analyzer (Model 6251, Lincoln, NB) was used. After about 3 minutes of purging period, the gas flow was stopped and readings were recorded on the computer. Although an electronic circuit was provided by the manufacturer in order to linearize the CO 2 response, it exhibited a few inflexions that deviated from linearity by a few ppm. Therefore, we chose not to use the outputs from the linearization circuit supplied by the manufacturer. Instead, we used five standard gas mixtures (one pure nitrogen and four CO 2 -air mixtures) during the expeditions, and established response curves using the raw output from the analyzer. The CO 2 concentrations in the gas mixtures were calibrated

6 using the SIO standards determined by C. D. Keeling s group using the manometric method. The concentrations of CO 2 in the standard gas mixtures were summarized above. During normal operations, each of the standard gas mixtures was passed through the analyzer for 70 to 90 seconds at a rate of about 60 ml/min. This replaced the IR analyzer cell completely with the new gas. The flow was stopped for 5 seconds and then a millivolt reading from the analyzer was taken and recorded. Samples of equilibrated air and marine air were pumped through the analyzer for 180 seconds (3 minutes) at a rate of about 50 ml/min to purge the previous sample in the IR cell. The flow was stopped for 5 seconds and a reading for the analyzer output was recorded. This procedure was intended to eliminate errors due to fluctuations of the dynamic pressure within the IR cell by irregular gas flow rates. The slow flow rates used for samples were required for the removal of water vapor using the PermaPure membrane dryer. Between two sets of calibration runs using the five standard gases, 6 to 20 samples were analyzed depending upon the stability of the IR analyzer. 2-d) Data Logging System: The following values were recorded on a laptop computer. The sample locations were derived from a GPS positioning unit that is a part of our surface water pco 2 system. The CO 2 readings for samples were recorded once every 3 minutes (180 seconds), and those for the standard gas mixtures once every 1.5 minutes. Date, Time (GMT), Latitude, Longitude, Sample ID (standard gas cylinder numbers, seawater CO 2, atmospheric CO 2 ) Barometric pressure in the laboratory (to 0.1 mb) IR cell temperature, Gas flow rate in the IR cell (to 0.1 ml/min), Temperature of equilibration (to 0.01 o C), Analyzer output (millivolts to 0.1 mv) CO 2 concentration in dry gas sample (preliminary based on the last response curve), and pco 2 (preliminary value based on the last response curve). 2-e) Data Deduction Procedures: The concentration of CO 2 in the sample was computed by the following way based on the millivolt reading and time of the reading. The millivolt reading taken for each of the five standard gases at the time of sample measurement was computed by linearly interpolating as a function of time using the readings taken before and after the respective standard gases were analyzed. This yields millivolt reading for each of the five standard gases at the time when the sample was analyzed. These five values were fit to a fourth-order polynomial equation (with five constants to be determined). This serves as the response curve. The CO 2 concentration in the sample was computed using the

7 response curve that was established at the time of each sample analysis. This method has been demonstrated to yield more reliable CO 2 values compared with those computed, for example, using a least-squares fit of a quadratic or cubic functions to the five calibration points. The method described above yields atmospheric CO 2 values that are consistent with those reported for the South Pole and the Cape Grim by the Climate Monitoring and Diagnostics Laboratory/NOAA in Boulder, CO. The partial pressure of CO 2 in seawater, (pco 2 )sw, at the temperature of equilibration, Teq, in the unit of microatmospheres (µatm) was computed using the expression: (pco 2 )sw @ Teq = (Vco 2 )eq x (Pb Pw),......[1] (Vco 2 )eq Pb Pw = the mole fraction concentration (ppm) of CO 2 in the dried equilibrated carrier gas; = the barometric pressure (that is equal to the total pressure of equilibration) in atmospheres; and = the equilibrium water vapor pressure at Teq ( o C) and salinity. The water vapor pressure was computed using the following formulation; Pw (atm) = (1/760)x(1 5.368x10-4 x Sal) x EXP{[0.0039476 (1/TK)]/1.8752x10-4 },... [2] where Sal is salinity in PSU measured using the ship s thermosalinograph, and TK is the temperature of equilibration in o K. The (pco 2 )sw at the in situ temperature, T in situ, was computed using a constant value of 0.0423 % per o C for the effect of temperature (Takahashi et al., 1993): (pco 2 )sw @ Tin situ = (pco 2 )sw @ Teq x EXP[0.0423 x (Tin situ Teq)]. The value for Tin situ is taken to be the seawater temperature measured by the ship s thermosalinograph at the time of pco 2 measurements. Teq is generally warmer than Tin situ by 0.5 ~ 0.8 o C. Hence the temperature correction is normally less than 3% of pco 2 values. The over all precision of the reported pco 2 )sw values has been estimated to be about +1.5 uatm. 3. MEASUREMENTS OF pco 2 IN THE ATMOSPHERE 3-a) Measurements: The air measurement system is shown schematically in Fig. 1. Uncontaminated marine air samples were collected about 10 m above the sea surface using a DEKORON tubing (1/4 i.d., Calco Inc., PA), a thin-wall aluminum tubing protected by plastic casing. The intake was located at the middle of the foremast about 10 m above the sea

8 surface. A KNF Neuberger air pump that was located near the IR analyzer was used to pump air through the tubing and into the IR analyzer. Even when air samples were not analyzed, the pump was on all the time to keep the air flowing through the sampling line. For the analysis, the air sample was passed through a water trap and a drying column to remove water vapor (the same PermaPure column as used for the equilibrated gas) and introduced into the IR cell for CO 2 analysis at a rate of about 50 ml/min. After 3 minutes of purging the cell, the flow was stopped for 5 seconds and the IR millivolt output reading was recorded. 3-b) Data Processing: The partial pressure of CO 2 in the air, (pco 2 )air, was computed in the unit of microatmospheres (µatm) in the same way as that for seawater using Eq. [3] below: (pco 2 )air = (Vco 2 )air x (Pb Pw),......[3] (Vco 2 )air Pb Pw = the mole fraction concentration (ppm) of CO 2 in the dried air sample; = the barometric pressure at sea surface in atmospheres; and = the equilibrium water vapor pressure at Tin situ ( o C) and salinity given by Eq. [2]. The precision of the atmospheric pco 2 values have been estimated to be about + 1 µatm. 4. REFERENCES CITED Bates, N. R., Takahashi, T., Chipman, D. W. and Knapp, A. H. (1998). Variability of pco 2 on diel to seasonal time scales in the Sargasso Sea. Jour. Geophys. Res., 103, 15567-15585. CO 2 Group, Lamont-Doherty Earth Observatory. (1999) pco 2 Equilibrator Users Manual, LDEO of Columbia University, Palisade, NY, pp.10. Takahashi, T., Olafsson, J., Goddard, J., Chipman, D. W. and Sutherland, S. C., (1993). Seasonal variation of CO 2 and nutrients in the high-latitude surface oceans: A comparative study. Global Biogeochemical Cycles, 7, 843-878.