Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer

Similar documents
Variable Temperature Operation

Updated: Variable Temperature

DCIF Training Guide Variable Temperature Bruker B400 (last update )

PINMRF Checkout Quiz - Bruker/TopSpin Version

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Installation Instructions

Things to know before operating an NMR spectrometer André Boltjes, University of Groningen DD NMR Lab v

DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3

User guide for the NMR spectrometer AVA300

Bruker Sample Transport

1. Cryogen Tanks. N2 Fill Tutorial for Vanderbilt Magnets. Contents: CSB NMR Facility

Installation Instructions

[USING THE VITROBOT April 9, Using the Vitrobot

Installation Instructions

Typically NMR Sample Configuration

Shimming with VnmrJ 3.1 Software for Peak NMR Performance

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

Freedom8 ShoeBox Compressor Manual

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Table of Contents Want full page pictures with most important parts labeled in the beginning.

Modular training weapon Owners manual

FOURIER TRANSFORM INFRARED SPECTROSCOPY

5 Gallon Pressure Pot with HVLP Spray Gun and Hose

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES

Series Environmental Chambers

QUICK START GUIDE TO DRINKING WATER MONITORING

SOLUBILITY OF A SOLID IN WATER

General remarks regarding ALS robot system:

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Napa Technology Trouble Shooting. For Premier & Premier PLUS Models

Exploring the Properties of Gases

MODEL 269DF-LCF-LCF/WS FILLING SYSTEM

WATER CONTROL SYSTEM QUICK START

Bruker BioSpin NMR CASE. System Requirements Manual. Version. NMR Spectroscopy. think forward

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18

Met One E-BAM Particulate Monitor

User Instruction Manual

DD2 Site Preparation Checklist Hardware Site Preparation Specification

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

PLUMBING, FIRE SAFETY, AND RADIANT HEATING AND COOLING SYSTEMS INSTRUCTIONAL GUIDE. Making ProPEX Connections

User guide for the NMR spectrometer AVA400Stud

Procedure for Filling 2 T Bruker Magnet with Nitrogen

Mass Spec will not Autotune

CryoProbe Prodigy. Bruker User Manual. Innovation with Integrity. Version NMR

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Microscopy Cryogenic WORKSTATION. Performance by Design

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Vapor Sorption Analyzer

The Discussion of this exercise covers the following points:

FAULT CODE TROUBLESHOOTING INDEX

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Cryostream 800 Series

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Operation of the Perkin Elmer TGA-GC/MS

Welcome to your new Firepod It will only work properly on Propane Gas (Also known as LPG and Patio Gas) with the special regulator supplied

SOLUBILITY OF A SOLID IN WATER

222 Schwinn Recumbent Exercise Bike Parts List Full Size Hardware Chart Product Illustration Assembly Instructions

Heavy Duty Dissolved Oxygen Meter

Evaluation copy. Wind Chill. computer OBJECTIVES MATERIALS

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Complex Care Hub Manual: Oxygen Therapy

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

Spot Weld Drill Spot Weld Removal System Instruction Manual

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Aerobic Respiration. Evaluation copy

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

ALU-CD 12, ALU-CD 25, ALU-CD 35, ALU-CD

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Helium Level Measurement Unit

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

SERVICE BULLETIN. TOPIC: Cylinder NUMBER: SUPERSEDES: NEW DATE: May 15, Indexing Basic Cylinder Height On CFR F-1/F-2 Rating Units

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Chem Bio Liquid Nitrogen Facility

Regulators repair and maintenance. XS Compact 2nd stage. January 2014 Rev XSC /3 Ed. C /14 1

6800 Maintenance Instruction System Flush Procedure

FLUORESCENCE DETERMINATION OF OXYGEN

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

Automatic LN2 Fill Station Instructions

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

215 Liquid Handler Up-Grade Manual

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No

6900 Maintenance Instruction System Flush

Operating Manual ALU 7 - ALU 10 - ALU 20 - ALU 26 ALU 35 - ALU 60 - ALU 100

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer

BD Cytopeia Fluidic Kit User s Guide

Autopilot setup. VRF (Virtual Rudder Feedback) calibration. Software setup NSS evo2 Installation Manual

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Instruction Manual. B&O WITH SLOPE BACK TENDER Live Steam

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Using the Akta Prime plus October 22, 2012

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

8. Carefully layout all the parts of the reservoir setup and clean any dirt, grime or dust from the parts. ( image 8 ) 8

INSTALLATION, CARE & USE MANUAL. EMASMB & LMASMB Surface Mount Bottle Filling Stations

Transcription:

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer (1) Use the drop down menu to create a new experiment/dataset for acquiring the room temperature 1D spectrum. The name of this data set should be used for the rest of the VT experiments on the same sample for a specific nuclei. Be sure to set the experiment number to 1 and clearly indicate in the title that this spectrum was acquired at room temperature (22 o C). (2) Go through the normal steps for acquiring the 1D spectrum as outlined in the 1D training manual. (3) Ensure that the shims at room temperature are sufficient for the resolution you would like in your VT spectra. If you are unhappy with the shims you can either manually shim or go through another auto-shim process. NOTE: You will need to check the shim at each temperature and shim when required. A good starting room temperature shim ensures that this process is much easier. (4) Reference your spectra and then write down the value for the SR parameter. Referencing can be done by clicking on the calibrate spectrum option in the NMR step by step drop down menu. The SR parameter can be found in the ProcPars tab or by simply typing SR in the command line and pressing enter. You will need this number later when referencing your VT spectra at various temperatures. NOTE: this number should be recorded for your ROOM TEMPERATURE spectrum, as the shift of the reference line is also temperature dependant and will change at each temperature. (5) Create a new data set for each temperature you will be working with. These new data sets should use the same experiment name and the current parameters but a different experiment number for each temperature. Clearly indicate the temperature in the title of each experiment.

(6) Open the VT interface by typing edte in the command line and hitting enter. Please be patient and wait while the VT interface opens. The default settings should be the following: Target temp = 22 o C Probe heater = ON Max power = 5% Gas flow = 400 L/h Cooling = empty and off Be sure that when you finish your VT experiments these parameters are set back to the default values shown above (step 10 below). If you do not return the parameters to these values you could lose your VT privileges. If these values are not at the default setting when you open the VT interface please inform the facility manager immediately. (7) You are now ready to setup the VT hardware for your experiment. YOU MUST carefully follow the policies and procedures for VT use of the instrument AT ALL TIMES. The following three sections provide instructions for setting up the instrument to work properly inside the three VT ranges of the spectrometer: (A) Normal Operational Range: +20 o C to +70 o C For this VT range no setup is required and your experiments can be run with the console in its normal operation mode.

(B) High Temperature Range: +70 o C to +150 o C (a) For the high temperature range you must completely open the valve found behind the magnet (shown in figure 1 below). This valve controls the air flow through the shim supply and when it is completely open you should hear air flowing through the shims. If you do not hear the air flow you should contact the facility manager before proceeding. NOTE: Starting your experiment without this air flow can cause severe damage to the shim set. (b) When using this temperature range you should ONLY use the ceramic spinner (figure 2) and make sure to know the boiling point of your solvent. DO NOT exceed the boiling point of your solvent as doing so can cause severe damage to the probe head! Figure 1: Valve for air flow through the shim supply

Figure 2: The ceramic (left) and plastic (right) spinners (C) Low Temperature Range: -150 o C to +20 o C (a) Fill the LN2 holding dewar. Figure 3 LN2 Holding Dewar

(b) Turn off the air flow to the spectrometer by shutting valve 2 as indicated in figure 3 below. Note: the valve is closed when it is perpendicular to the air lines. Valve 1 should remain open at all times as this valve delivers air to the stabilizer legs on the magnet. Figure 4 Gas valves for the air flow to the spectrometer. (c) Open the Nitrogen tank in the instrument room. This can be done by opening valves 1 and 2 as indicated in the figure 4. DO NOT adjust the pressure regulator.

Figure 5 Nitrogen Gas Cylinder found in the instrument room outside of the NMR lab. (d) Disconnect the airline from the quick connect indicated in figure 3. The airline should now be attached to the nitrogen gas line (grey line) with the quick connect adaptor as indicated in figure 5. Figure 6 N2 gas line (grey) found in NMR lab.

(e) Open the valve on the N2 gas line completely. You should hear a hiss when the valve opens and the spectrometer is now operating with N2 gas. (f) Completely open the valve found behind the magnet (see figure 1). This valve controls the N2 gas flow through the shim supply and when it is completely open you should hear gas flowing through the shims. If you do not hear the gas flow you should contact the facility manager before proceeding. NOTE: Starting your experiment without this gas flow can cause severe damage to the shim set. (g) Place the LN2 dewar in the circle outlined on the floor behind the magnet (h) Disconnect the console gas line from the probe (see figure 7) Figure 7 Rear of probe illustrating location of VT air line connection. (i) The o-ring and the ring clamp can be found on top of the console and are shown in figure 8A. Take the LN2 VT line off of the wall beside the magnet (figure 8B) and slide the o-ring all the way up to the top of the LN2 VT line (figures 9B).

Figure 8 (A) O-ring and ring clamp (B) VT Line for LN2 use Figure 9 O-ring on the (A) bottom and the (B) top of the LN2 VT line (j) Place the LN2 VT line in the LN2 dewar and let the line cool down. The line is cold when the pressure release valve stops hissing.

(k) Secure the LN2 VT line in the LN2 dewar by attaching the ring clamp finger tight as shown in figure 10. Figure 10 LN2 Dewar with the LN2 VT Line in place and secured with the ring clamp (l) Place the white string on the VT line to provide the line with support. Figure 11 The white string attached to the VT line to provide support.

(m) Gently push forward on the white Teflon nozzle found at the end of the LN2 VT line and attach the line to the probe as shown in figure 12. NOTE: No grease is required as long as the white string is holding the line level and straight. Now that the LN2 VT line is securely connected you can begin your experiments in the low temperature range. Figure 12 Picture of the LN2 VT line attached to the probe.

(9) Run your spectra at each desired temperature and be sure to check both your tuning and shimming at each temperature as these are highly temperature dependant. The temperature is varied by opening up the VT interface by typing edte in the command line and hitting enter. High temperature and normal operating ranges: you just need to set your target temperature and wait for the sample and probe to warm up. If the probe is taking an excessive amount of time to warm up you can increase the max power setting for the heater but this setting should not exceed 25%. The gas flow should remain at 400 L/h and the cooling setting should remain at empty and off. Low temperature range: Turn the cooling setting on and set your target temperature. You should adjust/increase the cooling flow according to your needs. In other words, the flow setting should be increased to allow you to reach your target temperature but not go below it. (10) Once your VT experiments are completed open up the VT interface by typing edte in the command line and hitting enter. Please be patient and wait while the VT interface opens. Return the VT parameters to the default settings shown below: Target temp = 22 o C Probe heater = ON Max power = 5% Gas flow = 400 L/h Cooling = empty and off NOTE: Be sure that when you book your VT experimental time you allow enough time after you are completed for the system to return to 22 o C.

(11) Lastly, use the following procedures for properly removing your VT setup and returning the instrument to normal use: (A) Normal Operational Range: +20 o C to +70 o C For this VT range no setup was required and the system is already configured for normal use, therefore no removal is required. NOTE: please make sure you have set the EDTE settings to those given in step 10 above. (B) High Temperature Range: +70 o C to +150 o C Ensure that the EDTE settings have been returned back to the default values given in step 10 above. Once the system has dropped back down to a temperature below 50 o C you need to completely close the valve for air flow through the shim supply (see figure 1). Lastly, once the system is back at 22 o C, put the ceramic spinner back in the case where you found it. (C) Low Temperature Range: -150 o C to +20 o C Ensure that the EDTE settings have been returned back to the default values given in step 10 above. Once the system has warmed back up to 22 o C you need to complete the following steps to return the system back to normal operation: (a) Disconnect the LN2 VT line from the probe head and replace it with the gas line from the console that you disconnected in step h above. (b) Remove the white string from the VT line. (c) Remove the ring clamp from the top of the LN2 VT line and pull the line out of the dewar. Cap the dewar and hang the VT line back up on the wall beside the magnet and allow it to warm up. Once the line has warmed up remove the o-ring from the line and place the o-ring and ring clamp back on top of the console.

(d) Completely close the valve found behind the magnet (see figure 1). This is the valve used to control air flow through the shims. (e) Go to the N2 gas line (grey gas line in figure 6) and completely close the valve. (f) Remove the quick connect line from the N2 gas line and re-attach it to the quick connect on the air line shown in figure 4. (g) Open valve 2 (see figure 4) to return the system to normal air flow. (h) Close valves 1 and 2 on the nitrogen tank located in the instrument room. (12) The System is now back in normal operational mode and should be ready for the next user of the instrument! NOTE: Be sure that when you finish your VT experiments the edte settings are set back to the default Values shown in step 10 above!!! If you do not return the system to these values you could lose your VT privileges. BE SURE to allow adequate time following your last experiment for the system to return back to the default temperature. In other words YOU MUST book enough time that when the next scheduled user begins the temperature of the system is back at 22 o C. Failure to do so leads to wasted time for the next user and could result in the loss of your VT privileges.