Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Similar documents
Recommendations. For Standards of. Monitoring and Safety during. Cardiopulmonary Bypass

Recommendations for Standards of Monitoring and Alarms During Cardiopulmonary Bypass

Guidelines on Monitoring in Anaesthesia

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

Point-of-Care Testing: A Cardiovascular Perfusionist s Perspective

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Recommendations on Checking Anaesthesia Delivery Systems

Anaesthetic infrastructure and supplies

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

Perfusion Safety Review in Japan

Clinical Engineering. Equipment (Heart-lung machines)

Recommended Minimum Facilities for Safe Anaesthetic Practice in Operating Suites

Recommendations for the minimal monitoring of the patient during anaesthesia.

Section 14: Equipment failure

Blood Parameter Monitoring System 550

S5 TM. Min.I. Mini Maxi Me. The proven safe and reliable, world-leading perfusion system, now optimized for minimally invasive and pediatric surgery

American Society of ExtraCorporeal Technology Standards and Guidelines For Perfusion Practice

Min.I. Mini Maxi Me. The proven safe and reliable, world-leading perfusion system, now optimized for minimally invasive and pediatric surgery

Minimum equipment list for cardiopulmonary resuscitation Primary dental care

Equipment and monitoring for cardiopulmonary bypass

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Using the Lifebox oximeter in the neonatal unit. Tutorial 1 the basics

healthy volunteers were included in the study.

I. Subject: Medical Gas Systems Utility Failure Interim Procedure III. Purpose:

Keeping anaesthetic pollution to minimum in theatre environment

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

A V A L A N C H E R E S C U E

Location Publication / Responsible And Number. Controller Standards Australia

THEORY AND PRACTICE IN THE USE OF A PUMP- OXYGENATOR FOR OPEN INTRACARDIAC SURGERY *

Monitoring, Ventilation & Capnography

APPENDIX 5. Donor Heart Perfusion & Preservation Protocol for NORS Teams in the UK

Medical Instruments in the Developing World

Clinical Skills. Administering Oxygen

PRACTICE GUIDELINE EM004 EMERGENCY CENTRE EQUIPMENT

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

HAEMOFILTRATION GUIDELINES

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

undergoing hypothermic CPB encountered difficulties elevating PaCO 2 and ph using

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

EMERGENCY MEDICINE AND THE LABORATORY. Ehsan Bolvardi(MD)

AQUARIUS Continuous Renal Replacement Therapy Competency

Anesthesia Machine Intro

First Class. Monitoring. solutions for. Range of monitors CO 2 O 2 SpO 2 ECG NIBP. More Choices to meet your needs

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

ENT 318/3 Artificial Organ. Artificial Lung. Lecturer Ahmad Nasrul bin Norali

Hartle et al. Checking anaesthetic equipment 2012 Anaesthesia 2012, 67, The pre-use check to ensure the correct functioning of anaesthetic equ

A complete family of Pediatric Perfusion Systems

SEMEM3-65 Servicing anaesthetic and ventilation equipment

INSPIRE Min.I. Redefining minimally invasive approach to perfusion. Expand your choices with INSPIRE Min.I.

Greater Manchester, High Peak and East Cheshire Neonatal Transport Service

SOME THOUGHTS ON THE TEACHING OF INSTRUMENTATION TECHNIQUES AND ELECTRONICS IN MEDICINE

CDI System 500. Blood Parameter Monitoring System. Continuous blood parameter monitoring for improved blood gas management

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Nitrous Oxide Sedation

Apollo Safety Features

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

Oxygen Saturation Monitors & Pulse Oximetry

Essential Performance for MED rd ed. PSES San Diego chapter meeting December 11, 2012

ekom Compressor for Medical Air Sonata with Ekom

Reviewing Hyperbaric Oxygen Services: Consultation Guide

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems

SimNewB. User Guide.

Mechanical Ventilation

Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia

LQG Based Robust Tracking Control of Blood Gases during Extracorporeal Membrane Oxygenation


McHENRY WESTERN LAKE COUNTY EMS SYSTEM FALL 2014 CONTINUING EDUCATION MANDATORY FOR ALL PRIMARY AND PROBATIONARY ALS SYSTEM PROVIDERS.

Cardiac Catheterisation Instrumentation. Lancashire & South Cumbria Cardiac Network

Shared Haemodialysis Care Handbook. Name Hospital number: Shared HD Care Nurse: Date: Machine type: Dialysis Unit:

OXYGEN FOR ADULTS IN ACUTE CARE

Considerations in Circuit Miniaturization

Flow meter. bellow vaporizer. APL valve. Scavenging system

Vital Signs Monitoring

ECMO Levitronix CentriMag Protocols Neonatal & Infant

Section 1 Questions. Physics and Clinical Measurement... 3 Clinical Anaesthesia Physiology Pharmacology... 79

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

Selecting and Connecting Breathing Systems

Engineering World Health Youth EWH Design Projects That Matter

Monitoring. Suzie Ward March 2013

CARDIOVIT AT-104 ergospirometry

PROTOCOL January 25, 2008

Blood Pressure Monitoring: Arterial Line

UNIQUE CHARACTERISTICS OF THE PULMONARY CIRCULATION THE PULMONARY CIRCULATION MUST, AT ALL TIMES, ACCEPT THE ENTIRE CARDIAC OUTPUT

(12) United States Patent (10) Patent No.: US 6,524,267 B1

PACKAGE LEAFLET: INFORMATION FOR THE USER. Neophyr 225 ppm mol/mol medicinal gas, compressed

PERFUSION SYSTEMS MECC SMART SMART SUCTION SYSTEM

QUESTIONNAIRE FOR MEMBERS OF THE PUBLIC

INSPIRE TM C The integrated closed system oxygenator for gentle perfusion

Troubleshooting Guide Aquarius

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Section Three Gas transport

Solutions for transcutaneous monitoring

SenTec Digital Monitoring System Illuminate Ventilation and Oxygenation in Neonatology PCO2 PO2

Orpheus Cardiopulmonary Bypass Simulation System

Update on Thoracic Anesthesia: Perioperative Lung Protective Strategies in One- and Two-Lung Ventilation

TRANSCUTANEOUS CO2 MONITORING (TCpCO2) PURPOSE

London Respiratory Team

General Information regarding ACVAA Abstract Submission and Presentation

Transcription:

Recommendations for Standards of Monitoring during Cardiopulmonary Bypass

Membership of the working party (2006) Ian Curle Fiona Gibson Jonathan Hyde Alex Shipolini David Smith J P van Besouw Kate Wark Mike Weatherall David Whitaker Society of Clinical Perfusion Scientists of Great Britain & Ireland Society for Cardiothoracic Surgeons in Great Britain and Ireland Society for Cardiothoracic Surgeons in Great Britain & Ireland Society of Clinical Perfusion Scientists of Great Britain & Ireland This document is available on the following websites: Society of Clinical Perfusion Scientists of Great Britain & Ireland www.scps.org www.acta.org.uk Society for Cardiothoracic Surgeons in Great Britain & Ireland www.scts.org

Introduction The aim of this document is to determine standards of monitoring during cardiopulmonary bypass for adult and paediatric surgery. These standards are considered by the Society of Clinical Perfusion Scientists of Great Britain and Ireland, the and the Society for Cardiothoracic Surgeons in Great Britain and Ireland to be the minimal monitoring required during cardiopulmonary bypass. This includes monitoring for the onset of and weaning from cardiopulmonary bypass, and for confirmation of anticoagulation and ventilation of the lungs. These standards are for use in conjunction with the Society of Clinical Perfusion Scientists of Great Britain and Ireland Standards of Practice document 1 and local protocols. Sources of reference include publications from the Society of Clinical Perfusion Scientists of Great Britain and Ireland 1, the Association of Anaesthetists of Great Britain and Ireland 2 and the American Society of Extra-Corporeal Technology 3. (Within this document on site facility is defined as on the hospital site, near patient facility is defined as within or in close proximity to the cardiac theatre.) All centres undertaking cardiac surgery involving cardiopulmonary bypass should plan to institute these recommendations of monitoring by 6 months from the date of publication.

The safe conduct of cardiopulmonary bypass is the joint responsibility of surgeons, anaesthetists and clinical perfusionists and requires a high level of c ommunication between the team members. Whilst it is considered best practice during the conduct of cardiopulmonary bypass for a surgeon and an anaesthetist to be present in the operating room at all times during cardiopulmonary bypass, it is recognised that there are circumstances where this may not be essential. However, safety of the cardiopulmonary bypass remains the primary responsibility of the perfusionist who must be present at all times. These situations should be managed using locally agreed clinical governance guidelines and protocols should be in place to ensure patient safety. (Examples of such locally agreed guidelines, in relation to anaesthetic practice, can be found on the ACTA website.) Only an accredited clinical perfusionist registered with the College of Clinical Perfusion Scientists of Great Britain and Ireland can undertake or supervise the conduct of cardiopulmonary bypass 1,4,5. A named and accredited clinical perfusionist not distracted by other clinical commitments, in close proximity and freely available must supervise a trainee undertaking a cardiopulmonary bypass 1. This document is a review of the original guidelines it will be reviewed regularly and may be revised or updated before the formal publication of a new edition. For the latest version, please refer to the ACTA website; www.acta.org.uk

General Recommendations Adequate records of monitoring should be kept at all times and where a variable is monitored it should be regularly recorded. All units should have electronic acquisition and storage of this data, including the ability to produce a printout 2. Records should be retained in the patient s notes and electronically. All monitors and alarms used should be calibrated and maintained regularly according to the manufacturer s instruction and the recommended service schedule. All equipment must be checked before use. During cardiopulmonary bypass, the electrocardiogram (ECG), intravascular pressures and core body temperature should be continuously displayed and clearly visible to the clinical perfusionist, surgeon and anaesthetist. (Ideally, three separate screens)

Monitoring of clinical parameters acquired directly from the patient The following should be monitored continuously, with local protocols dictating the frequency of recording: Electrocardiogram (ECG) Systemic arterial pressure Central venous pressure Core body temperature Urine output should be monitored using a freely draining urinary catheter. Pulse oximetry should be continuously displayed when there is a spontaneous pulsatile circulation. Expired carbon dioxide tension/concentration should be continuously displayed when the lungs are being ventilated. It is accepted that special clinical circumstances (for example emergency surgery or failure to insert a urinary catheter) may on some occasions preclude complete monitoring.

Monitoring associated with the cardiopulmonary bypass circuit The following should be monitored continuously: Oxygen saturation of the blood in the venous return line. Oxygen tension or saturation of the blood in the arterial line. Continuity of the fresh gas flow to the oxygenator using an in-line flow meter or rotameter 2. Oxygen concentration in the gas circuit to the oxygenator using an oxygen analyser with alarms, sited after the oxygen blender and vaporiser if used 2. Blood flow rate generated by the arterial pump. Arterial line pressure. Cardioplegia delivery line pressure when cardioplegia is delivered using the heart lung machine. Temperature of the blood in the arterial limb of the cardiopulmonary bypass circuit. Temperature of water in the heater/cooler system. Fluid record; all fluids, drugs and blood products added to the extra corporeal circuit during bypass should be recorded on the perfusion chart. Filtrate volume should be measured and recorded when a haemofilter/ concentrator is being used. Continuous in line monitoring screens (e.g. for potassium, haemoglobin, ph, arterial po2 and pco2, arterial and venous Oxygen saturation,) should be available.

The following measurements should be available at a near patient facility; (Local protocols should dictate the frequency of measurements) Anticoagulation should be confirmed by an acceptable method e.g. activated clotting time (ACT), which should be measured after heparinisation and before cardiopulmonary bypass and should be measured at regular intervals to ensure adequate anticoagulation. Other parameters that should be measured are; Blood gases Red cell concentration (haemoglobin or haematocrit). Potassium Glucose The following measurements should also be available at an on site facility: Clotting studies Calcium Lactate

Safety devices Local protocols for the conduct of cardiopulmonary bypass should be formulated by all hospitals undertaking cardiac surgery using cardiopulmonary bypass. The following are considered best practice: Power failure alarm with a battery powered back-up unit for the cardiopulmonary bypass machine. Battery powered Torch sited near to the bypass machine. Bubble detector on the arterial line of a roller pump cardiopulmonary bypass circuit with an alarmed automatic pump cut out facility. Level sensor on a hard shell venous reservoir system in the cardiopulmonary bypass circuit with an alarmed automatic pump cut out facility. A Retrograde Flow Alarm or an occlusion device is essential when using a Centrifugal Pump. Anaesthetic gas-scavenging apparatus whenever volatile agents are used in the cardiopulmonary bypass circuit 6. Out of range temperature alarm on the heater/cooler unit.

References 1. Standards of Practice. Society of Perfusionists of Great Britain and Ireland, London 1999. www.scps.org.uk 2. Recommendations for Standards of Monitoring during Anaesthesia and Recovery. The Association of Anaesthetists of Great Britain and Ireland, London 2007 www.aagbi.org 3. AmSECT Guidelines for Perfusion Practice. The American Society of Extra- Corporeal Technology, 1998. www.amsect.org 4. Guidelines for the Provision of Anaesthetic Services. The Royal College of Anaesthetists, London. 1999. www.rcoa.ac.uk 5. The Employment of Clinical Perfusionists in the NHS, The Department of Health s guidance on best practice for the employment of clinical perfusionists in the NHS, December 1999. 6. Health Services Advisory Committee. Anaesthetic Agents: Controlling Exposure under COSHH. London: HMSO, 1995.