GasLab SAN-102 Micro Oxygen Monitor User Manual

Similar documents
GasLab SAN-102 Micro Oxygen Monitor User Manual

RAM 4021-DPX Operation Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021 Operation Manual

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Phosgene Sensor. Sensoric COCl2 3E 1

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

User Manual. Pro CO LP Alarm. Carbon Monoxide Analyzer and Low Pressure Monitor with Alarm REV: 03.19

O3 3E 1 F Gas Sensor Module

Operation Manual. Pro Trimix Alarm TM. Helium & Oxygen Gas Analyzer. Rev

COMBAT SWIM BOARD. RJE International, Inc. RJE International Feb 1, TAC-300 Revision 6

In Vivo Scientific, LLC INSTRUCTION MANUAL

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL EM Washington Street Melrose, MA Phone Toll Free

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

EM151/EM201. Electronic Manometer - + INSTRUCTION MANUAL. EM201 PSI

Advanced Test Equipment Rentals ATEC (2832)

COMBAT SWIM BOARD. RJE International, Inc. RJE International January 16, TAC-300 Revision 4

Operating Manual. Check and Go Disposable Medical Depot, Inc Port Washington N.Y

Operation Manual. O2 Quickstick. Oxygen Analyzer

Verification Of Calibration for Direct-Reading Portable Gas Monitors

CALIBRATION KIT. Instruction Manual

DIVER NAVIGATION BOARD. RJE International, Inc. RJE International Feb 1, TAC-100D Revision 10

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Atmospheric Hazards. Hazardous Atmosphere 3/29/2017. Oxygen deficiency and enrichment Flammable gases and vapors Toxic contaminants

RAM 9025-XP Operation Manual

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

DEEPSEA. LED Micro SeaCam POWER & LIGHT. Operator s Manual. Original Instruction Ruffin Road San Diego, CA USA

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements

Ambient Weather WS-03 Thermo-Hygrometer

CSA Sample Draw Aspirator Adapter Operator s Manual

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. C

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL RLD10

Overview. Front Panel: Keypad and Display

Columbus Instruments

MODEL RP050BMST RESPIRATORY PROTECTOR MANUAL

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

BI-680 Online Dissolved Oxygen Controller Instruction Manual

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

SDM-2012 Docking Station Standalone Configuration Operator s Manual

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

HI 2314 HI 2315 HI 23151

PERSONAL AIR BREATHING UNIT

PS-425 Pitot-Static/Air Data Tester

GasSense NDIR User Manual

MODEL CALIBRATION GAS DELIVERY SYSTEM

Ozone. Sensoric O3 3E 1

Owner s Manual Humiport 10/20

WARRANTY, QUICK START GUIDE

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors

Return Policy / Procedures

UltraMaxO OXYGEN ANALYZER OPERATING MANUAL & INSTRUCTIONS FOR USE R221P11. R221M11 Rev. K

SPECIFICATIONS APCEPH1

Atlas 6TM. Indoor Air Quality Monitor Instruction Manual. Measures CO 2, Temperature and Humidity DISTRIBUTED BY

User Instruction Manual

MODEL 100 NITROGEN INFLATION CART

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007)

Dual-gas Oxygen and Carbon Monoxide SMART Analyzer User Manual

453 Series Steam Heated Vaporizing Regulator

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual

Portable oxygen concentrator

GAS DETECTION: CALIBRATION GAS

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

DG100 Digital Depth Gauge and Timer. Training Module, Rev B

MODEL RP050BMST-S1 RESPIRATORY PROTECTOR MANUAL

EC214 EC215 - EC215R Bench Conductivity Meters

User Manual. PolyGard AT-1130 V3. Electrochemical Nitrogen Dioxide Transmitter. July, 2011

RGC-IR Remote Gas Calibrator for IR400

Oxygen Concentrator. Installation and Operation Manual. ClearWater Tech, LLC. Integrated Ozone Systems

Nitrogen Supply System

COMBAT SWIM BOARD. RJE International, Inc. TAC-300 Revision C. RJE International August 3, TAC-300 Revision C

ENVIRONMENTAL INSTRUMENTS, LLC

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

Describe Flammable Gas Measurement

SMART Oxygen Analyzer. User Manual

Operating Instructions Part No

Manufactured by: AAA

700 SERIES BMS II BUBBLER MONITORING SYSTEM

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

GE Measurement & Control

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

Safety Meeting Topic #11 Hazards of Confined Space

Transcription:

GasLab SAN-102 Micro Oxygen Monitor User Manual The GasLab SAN-102 is a wearable, personal safety meter designed to monitor ambient oxygen levels in real time. It is designed to protect workers in confined spaces. Features Audible and visual low / high oxygen level alarm Silicone boot and metal clip are light and comfortable to wear Electrochemical sensor, no consumable parts Includes ambient temperature and % relative humidity sensors Lithium-ion rechargeable battery. USB Charging port Runs 4+ days per charge Applications Stored Gases in Confined Spaces. Industrial use of stored or piped gases creates the possibility that oxygen in breathable air could be displaced by gases such as nitrogen, helium, argon or carbon dioxide resulting in a hazardous condition. Underground Tunnels, Caves or Mines. Over time oxygen can be displaced by other naturally occurring gases like carbon dioxide, carbon monoxide, ammonia, methane and the oxidation of sulfides resulting in lower oxygen levels than is required for personal safety. Combustion in Confined Spaces. Combustion first uses up the available oxygen, then produces partial oxidation of carbon (carbon monoxide), hydrogen cyanide, and nitrogen oxides. Hydrogen gas can also be produced and displace oxygen. Please take a moment to read these instructions before use. They will provide you with all the necessary information for the correct use of your oxygen monitor. Copyright 2018, CO2Meter, Inc. All Rights Reserved. Last modified November 2, 2018 1

Theory of Operation The SAN-102 Oxygen Monitor uses an electrochemical sensor to determine the concentration of oxygen in air samples. Air reaches the sensor by diffusing through the openings on the back of the unit. Normal air movement is enough to carry the sample to the sensor. Device 1. OLED display 2. Power button 3. Fahrenheit and Celsius switching button 4. Alarm LED 5. Micro USB charge port 6. Vent 7. Clip 8. Silicone boot OLED Display 1. Oxygen concentration 2. Air Temperature ( F or C) 3. % Relative Humidity 4. Oxygen Concentration Units (%) 5. Battery Charge Indicator OPERATION Plug the USB cable into the meter and a 120VAC wall adapter. It will take approximately 2 hours to receive a full charge. Once the meter is charged, it can run 4+ days before recharging. Press the power button to turn on the meter. When the meter is first turned on, it performs a 10 second countdown for device warm up, then enters normal display with current oxygen level readings displayed. The device starts taking measurements when powered on and updates readings every 3 seconds. The OLED display will turn off automatically after 120 seconds of inactivity to conserve battery power. However, the sensor will continue to monitor the O2 level. Briefly press the Power button to view the display. Hold down the Power button for 3 seconds to turn the meter off. This will conserve the battery s charge until you need to use the meter again. Copyright CO2Meter, Inc. All Rights Reserved PAGE 2 of 6

Low and High Oxygen Level Alarm The SAN-102 is equipped with audible and visual alarms to alert you when the ambient oxygen concentration exceeds either of the two factory preset alarm levels: Danger Low Oxygen Alarm: 19.5% - LED will flash and audible alarm will sound 3x / sec. Warning High Oxygen Alarm: 23.5% - LED will flash and audible alarm will sound 2x / sec. Normal atmosphere contains between 20.8% and 21% oxygen. OSHA defines any atmosphere that contains less than 19.5% oxygen as oxygen deficient and any atmosphere that contains more than 23.5 percent oxygen as oxygen enriched. These levels are factory set and cannot be changed. Fahrenheit and Celsius Switching Press the F/ C button to switch between the two temperature units. Press the Power button briefly to save the temperature setting and return to normal display. Calibration The SAN-102 comes pre-calibrated from the factory. However, the O2 sensor should be calibrated at least once a year, or as described in your company's safety procedures. You can perform the calibration yourself, or you can return it to CO2Meter for factory calibration at a nominal fee. SAN-102 can be atmospherically calibrated. By pressing the Fahrenheit and Celsius switching button for 4 seconds, the monitor will enter O2 atmospherically calibration mode. While in calibration mode, press the Fahrenheit and Celsius switching button quickly to switch between atmospherically calibration or Exit. Press the Power button to confirm. The atmospherically calibration will not require additional materials but must be performed in outdoor (fresh air) during the entire procedure. The temperature and humidity sensors do not require calibration and should remain accurate for the life of your unit. Low Battery Alarm The meter continuously measures the rechargeable battery s voltage. If it drops to a critical level, the low battery alarm will sound until the battery is recharged or the battery power is depleted. Note that the meter can continue to monitor oxygen levels while charging. CLEANING AND STORAGE The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow drying completely before using. Do not use aromatic hydrocarbons or chlorinated solvents for cleaning. ATTENTION! - If a Warning or Danger alarm is triggered while using the instrument as a safety monitor, leave the area and seek fresh air immediately. Remaining on site under such circumstances can cause serious impairment or even lead to death. Copyright CO2Meter, Inc. All Rights Reserved PAGE 3 of 6

SPECIFICATIONS Oxygen Sensor Specifications Measurement Range 0 to 30.0% vol. Resolution 0.1% vol. Temperature Range -20 to 50 C (-4 F ~ 122 F) Humidity Range 0 to 99%RH Baseline Drift (-10-50 C) <1% change in output @ 3 months Expected Lifespan 24 months Temperature Sensor Specifications Temperature Range -20.0 to 60.0 C (-4 to 140 F) display Display Resolution 0.1 C (0.1 F) Display Options C/ F(switchable) Accuracy ±0.5 C( ±0.9 F) Response Time 5~30 seconds(device must equilibrate with environment) Humidity Sensor Specifications Measurement Range 0.0 to 99.9%RH Display Resolution 1%RH Accuracy ±4.5%RH Response Time <8 seconds for 63% of step change General Operating Storage Power Supply Dimensions Weight -20 C to 50 C (-4 F to 122 F),0~99% RH non-condensing -20 C to 60 C(-4 F to 140 F),<99% RH non-condensing Li-ion battery(4.2v, 400mAh) 60x35x25mm (2.36x1.38x0.98 ) 45 grams (1.58oz.) TROUBLESHOOTING GUIDE Symptom / Issue Possible Cause / Resolution Device does not power ON Reading does not change Make sure the power supply is connected properly and that there is power at the outlet. Use the supplied adapter. Verify that your meter has had enough time to charge before turning it on. In fresh air, you will rarely see oxygen levels change. This is normal. Copyright CO2Meter, Inc. All Rights Reserved PAGE 4 of 6

OXYGEN RECOMMENDED LEVELS > 23.5% Oxygen enriched, presents significant fire and explosion risk 20.8-21% Normal oxygen level in fresh air 19.5% Lowest safe oxygen level limit (OSHA) < 16% Headache, nausea, difficulty performing manual tasks < 12% Difficulty breathing, impaired coordination and perception < 10% Poor judgement and bluish lips < 8% Fatal within minutes if not treated < 6% Fatal in 40 seconds SUPPORT The quickest way to obtain technical support is via email. Please send all support inquires to support@co2meter.com. Please include a clear, concise definition of the problem and any relevant troubleshooting information or steps taken so far, so we can duplicate the problem and quickly respond to your inquiry. WARRANTY This meter comes with a 1 YEAR (warranty period) limited manufacturer s warranty, starting from the date the meter was shipped to the buyer. During this period of time, CO2Meter.com warrants our products to be free from defects in materials and workmanship when used for their intended purpose and agrees to fix or replace (at our discretion) any part or product that fails under normal use. To take advantage of this warranty, the product must be returned to CO2Meter.com at your expense. If, after examination, we determine the product is defective, we will repair or replace it at no additional cost to you. This warranty does not cover any products that have been subjected to misuse, neglect, accident, modifications or repairs by you or by a third party. No employee or reseller of CO2Meter.com s products may alter this warranty verbally or in writing. LIABILITY All liabilities under this agreement shall be limited to the actual cost of the product paid to CO2Meter.com. In no event shall CO2Meter.com be liable for any incidental or consequential damages, lost profits, loss of time, lost sales or loss or damage to data, injury to person or personal property or any other indirect damages as the result of use of our products. Copyright CO2Meter, Inc. All Rights Reserved PAGE 5 of 6

RETURNS If the product fails under normal use during the warranty period, a RMA (Return Material Authorization) number must be obtained from CO2Meter.com. After the item is received CO2Meter.com will repair or replace the item at our discretion. To obtain a RMA number, call us at (386) 256-4910 or email us at support@co2meter.com. When requesting a RMA please provide reason for return and the original order number. If we determine that the product failed because of improper use (water damage, dropping, tampering, electrical damage etc.), or if it is beyond the warranty date, we will inform you of the cost to fix or replace the product. For additional warranty information visit our website: www.co2meter.com/pages/terms-conditions CONTACT US We are here to help! For information or technical support, please contact us. support@co2meter.com (386) 256-4910 (Technical Support) (386) 872-7665 (Sales) www.co2meter.com CO2Meter, Inc. 131 Business Center Drive Ormond Beach, FL 32174 USA Copyright CO2Meter, Inc. All Rights Reserved PAGE 6 of 6