MEC OEM. User Manual. Commercial in Confidence

Similar documents
ANALOX O2 EII Oxygen Analyser User Manual for your Personal Protection

O2 EII Pro. Technical Manual

ANALOX ATA - Trimix Analyser. Technical Manual. Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK

ANALOX O2 PORTABLE OPERATORS MANUAL

O2 Portable. User Manual

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

O3 3E 1 F Gas Sensor Module

Operating Instructions

Technical Data Sheet MF010-O-LC

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

OXY 3690 MP. Measuring transducer for air oxygen

Pressure Automated Calibration Equipment

Operation and Maintenance Manual

Columbus Instruments

Instructions for Use

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

MODEL GT820 OXYGEN SENSOR

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

High-performance submersible pressure transmitter For level measurement Model LH-10

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer

Connections The C20B can be connected either using the Molex locking socket, or by connection to four plated through holes. The plated through

Expert Hydrostatic Level Transmitters

Air Check Lite Toxic Gas Monitor

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

RAM Operation Manual

Operation Manual. O2 Quickstick. Oxygen Analyzer

RAM 4021-DPX Operation Manual

High-performance submersible pressure transmitter For level measurement Model LH-10

GasSense NDIR User Manual

User Manual. Quantos Automated Dosing Liquid Module

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

Sensoric 4-20 ma Transmitter Board Operation Manual

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit:

User s Guide. Pressure Controller for liquid delivery systems

RAM 4021 Operation Manual

User Manual. Pro CO LP Alarm. Carbon Monoxide Analyzer and Low Pressure Monitor with Alarm REV: 03.19

Digital Melting Point Apparatus

Instruction Manual. CG16K Barometrically Compensated Capsule Dial Gauge. CG16K Capsule Dial Gauge, 0 to 25 mbar

Pneumatic high-pressure controller Model CPC7000

Transmitter CS 21 Operation Manual

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Model PDT Dewpoint Transmitter

Precision pressure sensor Basic version Model CPT6020

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

ATS430 turbidity sensor Retractable insertion assembly

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Safety data sheet NAVITAS 33 Revision date : 2009/03/26 Page: 1/5

ANALOX MRC MkII USER INSTRUCTION MANUAL

Domiciliary Oxygen Switch. Instructions for Use

Operation Manual. Pro O 2 Pro O 2 Remote. Oxygen Analyzers. Rev 08.17

High-performance submersible pressure transmitter For level measurement Model LH-20

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

Operation Manual. Pro O 2 Alarm TM. Oxygen Analyzer. Rev

MATERIAL SAFETY DATA SHEET

6900 Maintenance Instruction System Flush

Instruction and Maintenance Manual

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

High-performance submersible pressure transmitter For level measurement Model LH-20

Material Safety Data Sheet

User Manual. µ-gard MD Electrochemical Carbon Monoxide Transmitter Serial No. MD April, 2011

Fixed Ozone Monitor User Manual December, Automatikprodukter

3G Operating instructions. Digital Purge Gas Valve 3G Type: Document no.: D0001 Version: 01. July 2011/Rev.

Operating Instructions Part No

HBLT-A1 LIQUID LEVEL SENSOR

Handi+ OPERATING MANUAL & INSTRUCTIONS FOR USE. R218P15 Industrial. R218M15 Rev. D

Hand-held pressure calibrator with integrated pump Model CPH6600

Met One E-BAM Particulate Monitor

User manual CF8-D/W-IN

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Gas density monitor With integrated transmitter Model GDM-100-TI

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

Material Safety Data Sheet

LogIT Dissolved Oxygen Probe

Material Safety Data Sheet CoolingCare 8141

AutoChanger Installation & User Guide Issue 2

Instruction Manual Dräger MSI P7 and MSI P7 plus

QPL... Pressure Switches. Siemens Building Technologies HVAC Products

LK-SX CO 2 +VOC. Application. Security Advice Caution. Notes on Disposal

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

User Manual. MagnaMaster and ChloroMaster Electrolytic Chlorinator.

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

PEUGEOT PROTECT Tyre & Trim Dressing. PEUGEOT PROTECT Tyre & Trim Dressing, 400ml

Datasheet: K-30 ASCII Sensor

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

Transcription:

Analox Limited 15 Ellerbeck Court, Stokesley Business Park, North Yorkshire, TS9 5PT, UK UK/RoW T: +44 (0)1642 711400 F: +44 (0)1642 713900 US T: (714) 891-4478 W: www.analox.net E: info@analox.net

List of Contents List of Contents... 3 1 Safety information... 6 2 About the MEC... 8 2.1 MEC variants available... 9 3 Installation... 10 3.1 Physical mounting... 10 3.2 Gas connections... 12 3.3 Electrical connections (External)... 12 3.4 Electrical connections (Internal)... 13 3.5 Power/Comms connections... 13 3.6 4-20mA MEC... 14 3.7 Sensor connections... 14 3.8 RS485 communication... 15 3.9 Operation... 15 4 Sensor replacement... 16 4.1 Oxygen sensor replacement... 16 4.2 Toxic gas sensors... 19 4.3 Helium sensor replacement... 21 5 Calibration... 22 5.1 Performing a calibration... 23 6 Spares and Accessories... 26 7 Specifications... 27 7.1 MEC-O2... 27 7.2 MEC-CO (20ppm range)... 27 7.3 MEC-CO (500ppm range)... 27 7.4 MEC-H2S... 28 7.5 MEC-NO2... 28 7.6 MEC-NO... 28 7.7 MEC-SO2... 29 7.8 MEC-He... 29 7.9 MEC 4-20mA... 29 8 General care and disposal... 30 Document ref: MEC-802-12 August 2018 Page 3 of 30

List of figures Figure 1 Mounting Drawing... 10 Figure 2 Electrical Connections... 13 Figure 3 Calibration Equipment... 23 List of tables Table 1 MEC Variants... 9 Table 2 Electrical Connections... 12 Table 3 Internal Wiring Termination... 13 Table 4 Sensor Connections... 14 Table 5 Toxic Gas Sensors... 19 Table 6 Accessories... 26 Table 7 Replacement Cells... 26 Table 8 Recommended Calibration Gas... 26 Document ref: MEC-802-12 August 2018 Page 4 of 30

Warnings, Cautions and Notes Warnings and Cautions are used in this Manual to highlight potential hazards and safety risks. Notes are used to provide supplementary information that is not hazard-related. WARNING: THIS INDICATES A POTENTIALLY HAZARDOUS SITUATION THAT, IF NOT AVOIDED, COULD RESULT IN DEATH OR SERIOUS INJURY. CAUTION: THIS INDICATES A POTENTIALLY HAZARDOUS SITUATION THAT, IF NOT AVOIDED, COULD RESULT IN EQUIPMENT DAMAGE OR LOSS OF DATA. NOTE: THIS INDICATES INFORMATION THAT IS CONSIDERED IMPORTANT BUT IS NOT HAZARD RELATED. Document ref: MEC-802-12 August 2018 Page 5 of 30

1 Safety information WARNING: READ THE SAFETY INFORMATION FULLY BEFORE USING THE ANALOX MEC. Electrochemical sensors The sensor used in particular variants of MEC is an electrochemical sensor which contains toxic compounds. Under normal conditions the sensor will be safely sealed. To prevent leakage, the unit must not be exposed to temperatures outside the specified range, or be exposed to organic vapours, which may cause physical damage to the body of the sensor. The unit must not be stored in areas containing organic solvents or in flammable liquid stores. When the life of the sensor has expired or it is leaking or otherwise damaged it must be disposed of safely in accordance with local regulations. Electrochemical sensors contain caustic electrolyte (acidic or alkali depending on type) which is hazardous. In the event of an accident, use the following first aid procedures Skin Body Part Effect Contact could result in a chemical burn. First Aid Procedures Immediately flush the skin thoroughly with water for at least 15 minutes. Persons with pre-existing skin disorders may be more susceptible to the effects of the substance. Remove contaminated clothing and wash before re-use. Ingestion Corrosive. May cause sore throat, abdominal pain, nausea, and severe burns of the mouth, throat, and stomach, and may be fatal. Obtain medical advice if continued irritation. If swallowed DO NOT INDUCE VOMITING. Wash out mouth thoroughly with water and give plenty of water to drink. Eye Persons with pre-existing eye problems may be more susceptible to the effects of the substance. Obtain medical advice immediately Irrigate thoroughly with water for at least 15 minutes. Document ref: MEC-802-12 August 2018 Page 6 of 30

Body Part Effect First Aid Procedures Obtain medical advice immediately. Corrosive. May cause redness, pain, blurred vision, and eye burns. Inhalation Contact can result in the permanent loss of sight. Persons with pre-existing impaired respiratory function may be more susceptible to the effects of the substance. Inhalation is not an expected hazard unless heated to high temperatures. Remove to fresh air. Rest and keep warm. Obtain medical advice if applicable. Mist or vapour inhalation can cause irritation to the nose, throat, and upper respiratory tract. Document ref: MEC-802-12 August 2018 Page 7 of 30

2 About the MEC The Analox MEC is a small, robust sensor module available in a range of gas variants. All variants are housed in an IP65 enclosure and conform to the same physical dimensions, power requirements and communication protocol. An integral pressure sensor provides compensation for the effects of atmospheric pressure where appropriate. Digital communication via RS485 permits mounting the sensor remotely from the monitoring equipment The Analox MEC allows for easy replacement of sensor cells. Document ref: MEC-802-12 August 2018 Page 8 of 30

2.1 MEC variants available The Analox MEC is available in the variants shown in below: Table 1 MEC Variants Gas Range Part Number Node Address *Oxygen 0-3000mBar MECO2ABBP $40 *Oxygen 0-100% MECO2DBBP $40 *Carbon monoxide 0-20ppm MECCOEBAP** $50 *Carbon monoxide 0-500ppm MECCORBAP $50 *Hydrogen sulphide 0-50ppm MECUJBAP $30 Helium 0-100% MECHEMBAP $80 Helium 1 0-100% (mbar output) MECHEMBDP $80 *Nitrogen dioxide 0-10ppm MECPGBAP $52 *Nitric oxide 0-100ppm MECQKBAP $51 *Sulphur dioxide 0-20ppm MECTHBAP $53 4-20mA 0-60000mbar MECVTBBP $F8 * A hydrator should be installed if the MEC is going to be continuously connected to a dry gas source with no exposure to ambient air. The hydrator is required to pull moisture from the ambient air to re-hydrate the electrochemical cells within the MEC. If dry gas is constantly being passed over the sensors their life will be dramatically shortened. The hydrator is available from Analox Ltd, part number is listed in section 6. ** When ordering a spare carbon monoxide MEC (Part Number MECCOEBAP) it will leave the factory after being tested with a nitrogen balance span gas. The MEC CO should have a span done prior to use with a balance gas to match the gas being analysed (Either a nitrogen or helium balance). 1 NOTE: Helium MEC (MECHEMBDP) requires a remote pressure source to work correctly. Document ref: MEC-802-12 August 2018 Page 9 of 30

3 Installation 3.1 Physical mounting The sensor is housed in an IP65 ABS box with integrated mounting holes. The instrument should be screwed to a suitable surface using the two mounting holes indicated in the drill drawing in (all dimensions in mm). Figure 1 Mounting Drawing The enclosure lid holes are located in the top left and bottom right of the enclosure. These are captive screws. Document ref: MEC-802-12 August 2018 Page 10 of 30

The mounting holes of the enclosure are located at the top right and bottom left of the enclosure. A 4mm hole is pre-drilled at these corners that will accept an M4 screw thread. For the initial installation the enclosure lid should be removed to access the mounting hole chambers. The mounting screws are fitted in to these and the enclosure can then be secured to a suitable surface using either tapped holes or nuts. The maximum allowable diameter of the screw head that can be fitted in to the mounting hole chambers is 6.5mm. Mounting hole chambers In the lid at the mounting hole corners there are 4mm access holes predrilled to allow the mounting screws to be accessed. Once mounted, this enables the MEC to be removed from its mounting position without needing to remove the enclosure lid. The mounting screws will then remain captive as long as the lid remains fitted. Mounting screw access holes Document ref: MEC-802-12 August 2018 Page 11 of 30

3.2 Gas connections The sensor may either monitor gas in the surrounding atmosphere or a flow adapter can be inserted into the gas port inlet to allow monitoring of pumped sample gas from a remote location. When used to monitor a pumped sample, Analox recommend that a short length (10-30cm) of tubing should be attached to the flow adapter exhaust. Care should be taken to ensure that the sample flow rate is within specification and that the exhaust line is not restricted, otherwise gas pressure within the sensor may be increased, resulting in false, elevated measurement readings or damage to the sensor. Monitoring local atmosphere Monitoring a remote atmosphere via a sample line Note: If the application in which the MEC is being used is monitoring via a sample line the humidity of the gas must be taken in to consideration as long term exposure of electrochemical cells to low humidity gas (<15%RH) will result in a degrading of the sensor performance. Also, the material that the pipe-work is made from must also be considered. Certain gasses can exhibit a property by which they stick to inside of the pipe which results in inaccurate measurement of the target gas. In such cases it is advised to use pipe-work that prevents the gas from sticking such as PTFE. Please consult Analox for guidance. 3.3 Electrical connections (External) Electrical connections with the sensor are made via a short screened cable. The cable screen is made off into a green/yellow wire terminated with an M4 ring terminal. This wire is un-terminated inside the MEC enclosure. Table 2 Electrical Connections CORE COLOUR SIGNAL DETAILS Red +SUPPLY Power Supply Blue -SUPPLY 5V DC Yellow RS485A Green RS485B RS485 communications Green/Yellow Earth Screen Document ref: MEC-802-12 August 2018 Page 12 of 30

Use of the screen will depend on the particular installation. It is best connected to a clean Earth to form a shield around the sensor cable. Note that it is not recommended for the screen to be connected to the negative supply line. Figure 2 Electrical Connections 3.4 Electrical connections (Internal) The electrical connections made to the internal electronics are made via clamp terminals. It is important to ensure that each core of the cable is connected to the correct terminal. Each MEC has a Power/Comms terminal block and a sensor terminal block. 3.5 Power/Comms connections Table 3 details the internal wiring termination for the MEC connection cable. This is standard for all variants of MEC. Table 3 Internal Wiring Termination CORE COLOUR SIGNAL TERMINAL Red +SUPPLY J2.1 Blue -SUPPLY J2.2 Yellow RS485A J2.3 Green RS485B J2.4 Document ref: MEC-802-12 August 2018 Page 13 of 30

3.6 4-20mA MEC The 4-20mA MEC is used to connect a 4-20mA active or passive device to an Analox protocol device (such as an SDA). For details of how to connect to a 4-20mA MEC, see the quick-start guide supplied with the 4-20mA MEC (MEC-822) or ask Analox for the latest wiring diagram. 3.7 Sensor connections Table 4 details the internal wiring termination for the MEC sensors. Each MEC variant has a sensor termination that is particular to the type of sensor that is used in that variant. Table 4 Sensor Connections MEC VARIANT WIRE COLOUR TERMINAL Oxygen Red J3.1 Black J3.2 Carbon monoxide Red J3.1 Black J3.2 Yellow J3.3 Hydrogen sulphide Nitrogen dioxide Nitric oxide See Carbon monoxide Sulphur dioxide Helium Red J3.1 Not connected J3.2 Blue J3.3 Yellow J3.4 4-20mA Red J3.1 Black J3.2 Document ref: MEC-802-12 August 2018 Page 14 of 30

The example shown below is of an MEC-Oxygen, showing the Power/comms and sensor connections 3.8 RS485 communication Using RS485 communication data from the sensor can be obtained, and calibration commands may be sent to the sensor. The hardware protocol is 9600 baud, 1 start bit, 2 stop bits, 8 data bits. Please see P0075-805-01 ACG OEM Communication Manual for a description of the serial communication protocol, with code and message examples. 3.9 Operation After switch on, the sensor takes about thirty seconds to warm up. The MEC will then continuously monitor the gas levels. During operation the unit continuously runs various self-checks and sets fault flags accordingly. The status of these flags is available via the command G flags. Please see P0075-805-01 ACG OEM Communication Manual for details of sensor fault flags. Document ref: MEC-802-12 August 2018 Page 15 of 30

4 Sensor replacement The sensor replacement procedure is specific to the sensor type that is used for that particular variant of MEC. The following sections detail the procedure for replacement of the different sensor types used. 4.1 Oxygen sensor replacement Replacement part number for your sensor is: 9100-9212-9HM 1] Ensure all electrical connection between the MEC and monitor equipment are disconnected. 2] Loosen the corner screws of the enclosure and remove the lid. Document ref: MEC-802-12 August 2018 Page 16 of 30

3] Using a terminal screwdriver push down the clamp release button to release the clamp and gently pull the cell wire from the clamp. Repeat this for both the red and black wires. 4] Unscrew the green lock ring from the chimney of the cell housing and remove from the enclosure lid. 5] Fit the new cell in to the enclosure lid and secure in place using the lock ring. Document ref: MEC-802-12 August 2018 Page 17 of 30

6] Using a terminal screwdriver, push down the clamp release button to release the clamp. Push the ferrule of the cell wire in to the clamp and remove the screwdriver from the clamp release button. Gently pull on the wire to ensure that the wire is held by the clamp. Repeat this for both the red and black wires, taking note of the correct polarity as shown below. 7] Refit the enclosure lid in place and secure in place using the corner screws. 8] Following a sensor replacement a full calibration MUST be performed Document ref: MEC-802-12 August 2018 Page 18 of 30

4.2 Toxic gas sensors CAUTION: A NITRIC OXIDE (NO) MEC SENSOR NEEDS A MINIMUM OF 8 HOURS FROM POWER UP TO SETTLE BEFORE A VALID READING CAN BE TAKEN. Toxic gas sensor type covers a range of sensors of particular target gas. These are shown in Table 5 with the corresponding part number for the replacement sensor. NOTE: TOXIC GAS SENSORS ARE NOT DESIGNED TO BE CONTINUOUSLY EXPOSED TO THEIR TARGET GAS, THEY ARE BEST TO BE USED TO DETECT ELEVATED LEVELS OF TARGET GAS. Table 5 Toxic Gas Sensors Gas Replacement cell part number Carbon monoxide 9100-2030S Hydrogen sulphide 9100-2041 Nitrogen dioxide 9100-2044 Nitric oxide 9100-2045 Sulphur dioxide 9100-2042 CAUTION: A NITRIC OXIDE (NO) MEC SENSOR NEEDS A MINIMUM OF 8 HOURS FROM POWER UP TO SETTLE BEFORE A VALID READING CAN BE TAKEN. Document ref: MEC-802-12 August 2018 Page 19 of 30

1] Ensure all electrical connection between the MEC and monitor equipment are disconnected. 2] Use a flat blade screwdriver to pop out the cell holder from the sensor bulkhead fitting. 3] Rotate the cell PCB connector clockwise to release from the cell holder. 4] Disconnect the cell from the PCB connector. Document ref: MEC-802-12 August 2018 Page 20 of 30

5] Connect the new cell to the PCB connector (Note that it will only connect in one orientation). 6] Fit the cell in to the housing in the reverse action of step 3] 7] Refit the cell holder in to the sensor bulkhead. 8] Following a sensor replacement a full calibration MUST be performed. 4.3 Helium sensor replacement Replacement part number for your sensor is: 9100-4535/MEC 1] Ensure all electrical connection between the MEC and monitor equipment are disconnected. 2] Loosen the corner screws of the enclosure and remove the lid. 3] In the same way as shown in section 4.1 for the Toxic sensor, using a terminal screwdriver push down the clamp release button to release the clamp and gently pull the cell wire from the clamp. Repeat this for the red, blue and yellow wires. 4] Following a sensor replacement a full calibration MUST be performed. Document ref: MEC-802-12 August 2018 Page 21 of 30

5 Calibration Whilst in use, an MEC sensor should be periodically calibrated at intervals deemed necessary for the monitoring application by exposing the sensor to gas of a known concentration. Calibration adjustments of the sensor s output can be made where necessary as follows. Each sensor s output is defined by two calibration points, one low and one high. To perform a successful calibration adjustment, both low and high calibration points should be adjusted. NOTE: THE HIGH AND LOW CALIBRATIONS PERFORMED ON A SENSOR SHOULD BE SELECTED APPROPRIATELY. AN MEC DOES NOT REQUIRE THAT THE LOW CALIBRATION IS A TRUE ZERO CALIBRATION (I.E. ZERO CONCENTRATION OF THE TARGET GAS), BUT FOR GREATEST ACCURACY ACROSS THE WHOLE SENSOR RANGE IT IS RECOMMENDED THAT A GAS WITH A KNOWN ZERO CONCENTRATION OF THE TARGET GAS IS USED. No additional calibration should be required for the 4-20mA MEC. See section 6 for list the recommended calibration gasses for all MEC variants. Document ref: MEC-802-12 August 2018 Page 22 of 30

5.1 Performing a calibration The following equipment is required to perform a calibration on an MEC: Figure 3 Calibration Equipment 1] Push one end of the tubing onto one of the barbs of the flow adaptor and the other onto the outlet of the flow regulator. Document ref: MEC-802-12 August 2018 Page 23 of 30

2] Screw the flow regulator into the valve of the low calibration gas bottle. 3] Insert the flow adaptor into the gas port of the MEC sensor. 4] Open the flow regulator valve to deliver a flow of between 0.2 and 1.0 litres per minute. Document ref: MEC-802-12 August 2018 Page 24 of 30

5] Allow the calibration gas to flow for a few minutes or until the reading has stabilised. 6] Perform the calibration by sending the MEC sensor the appropriate calibration command. Please see P0075-805-01 ACG OEM Communication Manual for details of the calibration command using the serial communication protocol. 7] Close the flow regulator valve once complete. 8] Repeat steps 2] to 7] with the high calibration gas. 9] Once complete remove the flow adaptor from the gas port. 6 Document ref: MEC-802-12 August 2018 Page 25 of 30

6 Spares and Accessories Table 6 Accessories Item Description Part no. 1 Calibration flow adaptor 8000-0910A 2 0.5LPM 110L bottle regulator SA7L705XFLIND 3 Hydrator P0149-613 4 Bulgin 3 way plug connector (For 4-20mA MEC s only) PXP6010/03S/ST/0507 Table 7 Replacement Cells Item Description Part no. 1 MEC O2 cell (electrochemical) 9100-9212-9HM 2 MEC CO cell (electrochemical) 9100-2030S 3 MEC H2S cell (electrochemical) 9100-2041 4 Nitrogen dioxide 9100-2044 5 Nitric oxide 9100-2045 6 Sulphur dioxide 9100-2042 7 MEC He cell (pellistors) 9100-4535/MEC Table 8 Recommended Calibration Gas Item Description Part no. MEC-O2 1 Zero calibration gas (N2) SA7L10501 2 Span calibration gas (20.9% O2 balance N2) SA7L11009 MEC-CO 1 Zero calibration gas (20.9% O2 balance N2) SA7L11009 2 Span calibration gas (18ppm CO, 20.9% O2 balance N2) SA7L110176 MEC-H2S 1 Zero calibration gas (20.9% O2 balance N2) SA7L11009 2 Span calibration gas (50ppm H2S, 20.9% O2 balance N2) SA7L110820 MEC-He 1 Zero calibration gas (N2) SA7L10501 2 Span calibration gas (100% He) SA7L110120 Document ref: MEC-802-12 August 2018 Page 26 of 30

7 Specifications 7.1 MEC-O2 Range 0 to 100% (Atmospheric pressure) 0 to 3000mBar (Hyperbaric) Temperature range -5 C to 55 C (23 F to 131 F ) Accuracy (over ±10 C range) Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) ±(0.035% O 2 + 1% of reading + Temp coefficient) Temperature coefficient = 0.4% of reading/ C or 0.22% of reading/ F Electrochemical 2 years for the electrochemical cell in atmospheric air Supply voltage 5.0V dc (±0.5V) RS485 using Analox protocol 65 x 50 x 35 mm 7.2 MEC-CO (20ppm range) Range 0 to 20ppm Temperature range -5 C to 55 C (23 F to 131 F ) ±(1ppm CO + 5% of reading + Temp coefficient) 0-20ppm Accuracy (over ±10 C range) Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) Temperature coefficient = 0.1ppm CO/ C + 0.5% of reading/ C or 0.056ppm CO/ F + 0.278% of reading/ F Electrochemical 2 Years in operation Supply voltage 5.0V dc (±0.5V) RS485 using Analox protocol 65 x 50 x 35 mm 7.3 MEC-CO (500ppm range) Range 0 to 500ppm Temperature range -20 C to 40 C (-4 F to 104 F ) ±(2ppm CO + 5% of reading + Temp coefficient) 0-500ppm Accuracy (over ±10 C range) Temperature coefficient = 0.4% of reading / C or 0.222% of reading / F Detection mode Electrochemical Sensor life 2 Years in operation Supply voltage Supply voltage 5.0V dc (±0.5V) Data output RS485 using Analox protocol Enclosure dimensions (H x W x D) 65 x 50 x 35 mm Document ref: MEC-802-12 August 2018 Page 27 of 30

7.4 MEC-H2S Range 0 to 50ppm Temperature range -5 C to 55 C (23 F to 131 F ) ± (2ppm + Temp coefficient) Accuracy Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) Temperature coefficient = 0.075 ppm/ C or 0.042 ppm/ F Electrochemical 2 Years in operation Supply voltage 5.0V dc (±0.1V) RS485 using Analox protocol 65 x 50 x 35 mm 7.5 MEC-NO2 Range 0 to 10ppm Temperature range -5 C to 55 C (23 F to 131 F ) ± (1ppm + Temp coefficient) Accuracy Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) Temperature coefficient = 0.05 ppm/ C or 0.028 ppm/ F Electrochemical 2 Years in operation Supply voltage 5.0V dc (±0.5V) RS485 using Analox protocol 65 x 50 x 35 mm 7.6 MEC-NO Range 0 to 100ppm Temperature range -5 C to 45 C (23 F to 113 F ) ± (4ppm + Temp coefficient) Accuracy Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) Temperature coefficient = 0.17 ppm/ C or 0.094 ppm/ F Electrochemical 2 Years in operation Supply voltage 5.0V dc (±0.5V) RS485 using Analox protocol 65 x 50 x 35 mm CAUTION: A NITRIC OXIDE (NO) MEC SENSOR NEEDS A MINIMUM OF 8 HOURS FROM POWER UP TO SETTLE BEFORE A VALID READING CAN BE TAKEN. Document ref: MEC-802-12 August 2018 Page 28 of 30

7.7 MEC-SO2 Range 0 to 20ppm Temperature range -5 C to 55 C (23 F to 131 F ) ± (2ppm + Temp coefficient) Accuracy Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) Temperature coefficient = 0.1 ppm/ C or 0.056 ppm/ F Electrochemical 2 Years in operation Supply voltage 5.0V dc (±0.5V) RS485 using Analox protocol 65 x 50 x 35 mm 7.8 MEC-He Range 0 to 100% Temperature range -5 C to 55 C (23 F to 131 F ) ± (2%FS + Temp coefficient) Accuracy Detection mode Sensor life Supply voltage Data output Enclosure dimensions (H x W x D) Temperature coefficient = 0.05 %FS/ C or 0.028 %FS/ F Pellistor Up to 5 Years Supply voltage 5.0V dc (±0.1V) RS485 using Analox protocol 65 x 50 x 35 mm 7.9 MEC 4-20mA Range 0 to 60000mBar Temperature range -5 C to 55 C (23 F to 131 F ) Accuracy ± 0.15% FS Detection mode Supply voltage Data output Enclosure dimensions (H x W x D) 4-20mA Supply voltage 5.0V dc (±0.5V) RS485 using Analox protocol 65 x 50 x 35 mm Document ref: MEC-802-12 August 2018 Page 29 of 30

8 General care and disposal Although designed to be water resistant the MEC should not be intentionally immersed in liquid or left outside unprotected. To clean the MEC use a damp soft cloth. According to WEEE regulation this electronic product cannot be placed in household waste bins. Please check local regulations for information on the disposal of electronic products in your area. Document ref: MEC-802-12 August 2018 Page 30 of 30