Calibration Procedure

Similar documents
Calibration Procedure

Calibration Procedure. PosiTest DFT Combo

PosiTector UTG, UTG-STD & UTG-ME

DeFelsko PosiTector CMM IS Probe

Calibration Procedure

Calibration Procedure

DeFelsko PosiTector Dew Point Meter DPM & DPMD Probes

Calibration Procedure

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Operating Manual. BACVis. Manual BACVis. for. Sensors and MilliGascounter. Rev

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Standard Operating Procedure

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

TECHNICAL MANUAL CALIBRATION INSTRUCTIONS FUEL PRESSURIZING & DUMP VALVE TEST STAND 1TL

Cygnus Underwater. UT Meters & Spare Parts Sales. Searchwise Aberdeen Ltd Unit 6, Broomiesburn, Ellon Ind Estate, Ellon, Aberdeenshire, UK, AB41 9RD

Heavy Duty Dissolved Oxygen Meter

Manual Weighingblock VB2 series and Uniscale

8.1 Calibration of gas sensors

ali-q TM Gravimetric Verification Procedure

Standard Operating Procedure Measuring & Testing Equipment

! Warning, refer to accompanying documents.

AIR FLOW ANEMOMETER INSTRUCTION MANUAL

DIGITAL AUTOMATIC TYRE INFLATOR

CCPaSEC. Procedure for the Quality Control Yearly Equipment Check. July 2018 CCPaSEC Quality Control Team

Dissolved Oxygen/Temp Waterproof Tester

PIPELINE AND HAZARDOUS MATERIAL SAFETY ADMINISTRATION CYLINDER REQUALIFICATION FACILITY INSPECTION REPORT GENERAL INFORMATION

PQI Model 301 & 300Test Block Procedure

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Pneumatic high-pressure controller Model CPC7000

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

DO NOT PRINT THIS PAGE

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

M104 Series: Digital Manometer: 0.1 % of Full Scale

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

Manual Crane Scale PCE-CS 300

BAPI Pressure Line of Products - FAQs

BI-680 Online Dissolved Oxygen Controller Instruction Manual

Operating Instructions Part No

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007)

Model of a Standard Operating Procedure (SOP) for SO 2 measurements with a UV fluorescence analyzer

6 digital caliper with case

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Introduction. The Shearwater Petrel is an advanced technical diving computer for open and closed circuit divers.

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

Digital Pressure Gauge Main features: Data Sheet: DFP.418.R4.EN

PTG100 Precision Test Gauge

GE Measurement & Control

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Location of CC Number if not located with the identification:

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR PRESSURE GAUGE. This publication replaces T.O. 33K dated 30 January 2012.

Technical Procedure for Instrument Calibration and Maintenance

Do Not Print This Page.

J12 TICKET DISPENSER HOPPER ALL STOP STOP 1 STOP 2 STOP 3 BIG DOUBLE SMALL TAKE PLAY START CURRENT PRODUCTION BOARD TYPE

CÜBE. User Manual. American Weigh Scales. CUBE-100 (100g x 0.01g) CUBE-650 (650g x 0.1g)

Technical Procedure for General Laboratory Equipment

Simrad GD10P Extended On-site Maintenance

HEATEC TEC-NOTE. Setting Siemens Pressure Transmitter. Used on Heatec Vertical Asphalt Tanks. Publication No , Revised

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Rejuvenation Instructions

Procedure for Field Calibration Check of Digital Pressure Gauges (Using a DG-700 or DG-500 as the Reference Standard)

Owner s Manual Humiport 10/20

CALIBRATING THE GMA PRESSURE SENSORS AND THE ECU ENGINEERING UNIT

Ambient Weather WS-03 Thermo-Hygrometer

User-Calibration of Mettler AT200 Analytical Balance

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

Liquid Handling Systems

Digital Melting Point Apparatus

UNDERSTANDING, ACCELERATED

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

MANUAL: SHO-FLOW Flow Meter Operations

Tex-914-K, Calibrating Motorized Gyratory-Shear Molding Asphalt Press

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Pneumatic high-pressure controller Model CPC7000

DQM Annual Hopper QA Checks

Model 130M Pneumatic Controller

Digital Vacuum Regulator

Model SM/PL Series. Calibration Guide. Precision Loads. Anritsu Company 490 Jarvis Drive Morgan Hill, CA USA

Bristol Fire Department Standard Operating Procedures. Multi Gas Detector

Optical Time Domain Reflectometer. Operating System 2 Windows XPe Windows XP Pro with desktop option 256 MB

IN-PLACE DENSITY OF BITUMINOUS MIXES USING THE NUCLEAR MOISTURE-DENSITY GAUGE FOP FOR WAQTC TM 8

DG100 Digital Depth Gauge and Timer. Training Module, Rev B

Safety Alert. Repeating Symbols: 1 Planning Information

Installation and Operation Manual

FRDS GEN II SIMULATOR WORKBOOK

American Weigh Scales, Inc. BLADE Series D i g i t a l P o c k e t S c a l e BLADE-550. User Manual BLADE-400 BLADE-1000

American Weigh Scales, Inc. LB User Manual

FAULT CODE TROUBLESHOOTING INDEX

LM6 LEATHER MOISTURE METER

Document Number: 302/34000/76 SV/sv Revision: (B) 2009/02/05

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Operating Instructions Part No

Elcometer ISO Groove Depth Checker. Operating Instructions

INSTALLATION & OPERATION MANUAL. Dual Duct VAV Terminals. Redefine your comfort zone.

Document No.: PZ _00. Safety Alert

750 Cable Cutter INSTRUCTION MANUAL

E Lightband Compression Tool Operating Procedure

MP15 Jockey Pump Controller

Transcription:

Management Procedure 2535 Revision: E Date Issued: October 23, 1998 Date Revised: March 6, 2012 Calibration Procedure DeFelsko Corporation PosiTector 6000 N PosiTector 6000 NS PosiTector 6000 NRS Coating Thickness Probes Table of Contents 1 Introduction and UUC Performance Requirements... 2 Table 1-1 Measurement Ranges... 2 2 Measurement Standards and Support Equipment Performance Requirements... 2 Table 2-1 UUC Accuracy Requirements and Description... 2 Table 2-2 Minimum Use Specification... 2 Table 2-3 Actual Equipment Specification... 2 Table 2-4 Calibration Environmental and Warm-up Requirements... 3 3 Preliminary Operations... 3 4 Calibration Process... 4 Figure 4-1 Measurement Area for Square or Round Reference Standards... 4 5 Performance Requirements... 5 Table 5-1 Performance Requirements and Calibration Data for PosiTector 6000 N, NS & NRS... 5 Management Procedure Change Notice... 6 Management Procedure 2535 Rev. E Page 1 of 6

1 Introduction and UUC Performance Requirements 1.1 This procedure describes the calibration of DeFelsko Corporation PosiTector 6000 N, 6000 NS and 6000 NRS probes with the following specifications: Table 1-1 Measurement Ranges Probe Measurement Range 6000 N 6000 NS 0-1500 microns 6000 NRS (0-60 mils) 1.2 The unit being calibrated will be referred to as the UUC (unit-under-calibration). 2 Measurement Standards and Support Equipment Performance Requirements 2.1 The UUC accuracy requirements are based upon the published UUC performance specifications. 2.2 The test uncertainty ratio applied in this Calibration Procedure is 4:1 unless otherwise stated. 2.3 The Minimum-Use-Specifications are the minimum test equipment specifications required to meet all the UUC accuracy requirements and the test uncertainty ratio applied Table 2-1 UUC Accuracy Requirements and Description UUC Performance Specifications Test Method 0-50 microns (1 microns + 1% of reading) 6000 N (0-2 mils) (0.05 mils + 1% of reading) Compared to 6000 NS >50 microns (2 microns + 1% of reading) Reference Standards 6000 NRS (>2 mils) (0.1 mils + 1% of reading) Table 2-2 Minimum Use Specification Range Accuracy 0-50 microns + 0.25 microns (0-2 mils) (+ 0.013 mils) >50 1500 microns + 0.62 microns (>2 60 mils) (+ 0.035 mils) Equipment Generic Name Coating Thickness Reference Standards Table 2-3 Actual Equipment Specification Range Accuracy Manufacturer/Model # s Applicable 75-1500 microns 0.43 microns DeFelsko Corporation, (3-60 mils) ( 0.017 mils) STD-A1 Management Procedure 2535 Rev. E Page 2 of 6

Caution: The instructions in this Calibration Procedure relate specifically to the equipment and conditions listed in Section 2. If other equipment is substituted, the information and instructions must be interpreted accordingly. Table 2-4 Calibration Environmental and Warm-up Requirements Measurement Standards & Support Equipment Temperature: 23 5 C. Environmental Requirements: Relative Humidity: Less than 95% Measurement Standards & Support Equipment Warm-up and Stabilization Requirements: Not Required 3 Preliminary Operations Note: Review the entire document before starting the calibration process. 3.1 Visual Inspection 3.1.1 Visually inspect the UUC for: Damaged LCD readout probe wear or coating cracked or broken case missing probe cover, battery door or other parts proper identification 3.1.2 Damage or excess wear shall be repaired prior to beginning the calibration process. 3.2 Gage Reset 3.2.1 For bodies with serial numbers after 700000; when the unit is powered down, simultaneously hold the + and middle buttons until the reset symbol (2 arrows) appears. All other bodies press and hold the + button. Caution: Be sure to keep the probe well away from any metal surface during the RESET process. 3.3 Probe Zero 3.3.1 After Reset, select the Main Menu ZERO function and measure the uncoated Reference Standard. One measurement is sufficient. 3.3.2 Perform a zero check by measuring the same standard. If the probe does not read within tolerance, repeat the Main Menu ZERO function. Management Procedure 2535 Rev. E Page 3 of 6

4 Calibration Process Note: Whenever the test requirement is not met, verify the results of each test and take corrective action before proceeding. 4.1 Review the Performance Requirements Table 5-1. Note: The probe calibration may be verified in either normal or high-resolution mode. Accuracy is the same for both modes. 4.2 Using the appropriate Certificate of Calibration template for the UUC, record the thickness from the Reference Standard labels. 4.3 Determine the allowed range of readings for the UUC using the calculation methods shown in Table 5-1. 4.4 Use the UUC to take readings of all the reference standards. Verify that the readings are within the allowable limits determined in 4.3. Record the readings on the Certificate of Calibration. Note: Record all digits displayed on the LCD. This may vary depending on the resolution mode. 4.5 In taking readings the probe tip shall be centered on point A of the Reference Standard as shown in Figure 4-1. Figure 4-1 Measurement Area for Square or Round Reference Standards Management Procedure 2535 Rev. E Page 4 of 6

5 Performance Requirements Note: The technician shall collect the data needed to complete columns A and B of the appropriate table below. Do not write in this procedure. Table 5-1 Performance Requirements and Calibration Data for PosiTector 6000 N, NS & NRS Thickness on Standard Label Min. Reading Allowed Max. Reading Allowed Actual Probe Measurement A B 0-1 +1 Calculation <50 microns: (A times 0.99) - 1. Round up to the nearest 1 micron. > 50 microns: (A times 0.99) - 2. Round up to the nearest 1 micron. Calculation < 50 microns: (A times 1.01) + 1. Round down to the nearest 1 micron. > 50 microns: (A times 1.01) + 2. Round down to the nearest 1 micron. * For imperial/metric readings convert using 1 mil = 25.4 microns Management Procedure 2535 Rev. E Page 5 of 6

Management Procedure Change Notice Procedure Number: 2535 Revision Level: E Date of Change: March 6, 2012 Title: Calibration Procedure for PosiTector 6000 N, NS & NRS Reason for Change: Rev E gage updates Description of Change: Standardized probe/gage verbiage Changed section 3.2.1 Removed gage serial number reference in the note of section 4.1 I confirm I have read and understand the procedure and the change described above. Printed Name Signature Date Management Form 0010.02-05/1998 Management Procedure 2535 Rev. E Page 6 of 6