BiPAP Vision Ventilator Testing with Various Patient Interfaces Report February 18, 2002

Similar documents
PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

AUTOVENT 4000 VENTILATOR

Unit 15 Manual Resuscitators

A bench study comparison of volume delivery in the presence of leak

Selecting the Ventilator and the Mode. Chapter 6

Astral in AirView: Improving patient care through connectivity. ResMed.com

Meet the changing needs of your patients. Trilogy Series, with advanced ventilator technology

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

S/5 Aespire. Quick Reference Guide

Astral in AirView: Improving patient care through connectivity. ResMed.com

Patient Setup Information

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Expanding versatility. The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients

Automatic Transport Ventilator

e360t and e360t+ Model

ASSISTANCE. Cardinal Health Pulmonetic Systems Medina Rd., Suite 100 Minneapolis, Minnesota


Average Volume Assured Pressure Support

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

CPAP Devices - Masks - Accessories for respiratory assistance in CPAP

Blease900 Ventilator PATIENT CENTERED VENTILATION

Leak compensation in positive pressure ventilators: a lung model study

Contents Warnings, Cautions, and Notes Contents of the Package Introduction to the Harmony S/T A Tour of the Harmony S/T...

First and Only Disposable Infant T-Piece Resuscitator. Because they are not miniature adults...

Health Professional Info

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Instructions for Use. OptiLife Nasal Lab Mask

ALLIED HEALTHCARE PRODUCTS, INC. Litholyme, GE Medisorb, Carbolime, and Amsorb Plus. Test Summary

C-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8

Oxygen injection sites: Important factors that affect the fraction of inspired oxygen

Operation and Maintenance of the EPV200 Portable Ventilator

pneuton Transport Ventilator

Mitos Fluika Pressure and Vacuum Pumps Datasheet

BiPAP Auto. M Series JR 1/22/08 EN-DOM

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

Software enhancements

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Notes on BIPAP/CPAP. M.Berry Emergency physician St Vincent s Hospital, Sydney

IMPACT. Uni-Vent. Eagle. Model 754 Portable Ventilator

Using Common Ventilator Graphics to Provide Optimal Ventilation

R409 Plus Veterinary Ventilator Machine

Panther 5 Acute Care Ventilator

NIOSH Task Number 18388

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

VPAP ST-A with ivaps. H5i. Clinical Guide. English. Making quality of care easy NONINVASIVE VENTILATOR HEATED HUMIDIFIER. Respiratory Care Solutions

Corporate Overview and Product Summary

Test your PICU/ NICU/ CHSU Knowledge

300 Held Drive Northampton Pa

LISTADO DE CIRCUITOS DE PACIENTE

Operation of Oxylog 3000 ventilator

Operator s Manual Addendum

A9600 Anesthesia Workstation. Your healthcare, we care.

2015 Hamilton Medical AG. All rights reserved. Printed in Switzerland.

SLE5000 Infant Ventilator with HFO

Simplified Automated Ventilator Training Slides

Mirage Vista Nasal Mask

Mass Flowmeters. Mass Flowmeters for Gases

3100A Competency Exam

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Airox Supportair Ventilator

1 of 9 10/20/2010 4:45 PM

A Bench Study of the Effects of Leak on Ventilator Performance During Noninvasive Ventilation

UNDERSTANDING, ACCELERATED

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

Flight Medical presents the F60

GE Healthcare. Centiva/5 Critical Care Ventilator. Meet a new level of expectations

HIGH FREQUENCY JET VENTILATION (HFJV): EQUIPMENT PREPRATION

TOPO. operating Manual

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

(Delivery System) Pocket Guide. Series 3 software

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE

VPAP IV ST POSITIVE AIRWAY PRESSURE DEVICE

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

IEC C LASSIFICATIONS:

Confidence. in your patient s comfort Philips Respironics patient interfaces and circuits for noninvasive ventilation

How to use the SAVe II Ventilator

What our equipment teaches us

BiPAP A30. Set-up guide. Connection to oximetry. Getting started. Guidelines for ventilation set-up. Main screen. Menus.

Adult/Pediatric Demonstration Lung Model. User s Manual

Oxylog 3000 plus Emergency & Transport Ventilation

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Hallowell EMC Anesthesia WorkStation Operating Manual

evolution 3e Ventilators

Pressure Controlled Modes of Mechanical Ventilation

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

Test lungs from IMT Analytics simulate a range of realistic ventilation situations during training.

NOTE. To enter the Extended Features menu (in normal ventilation mode): Push and hold the Select button for three seconds.

Draeger Fabius GS Pre-Use Check

Gas exchange measurement in module M- COVX

Oxygen Injection Site Affects F IO2 During Noninvasive Ventilation

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Device overview. 6 Breathing circuit connection. Note: When the remote control foot pedal is connected, the manual switch is disabled.

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

pro pro 2 AND User Manual

Electric Blower Based Portable Emergency Ventilator

The Merlin Small Animal Ventilator by Vetronic Services Ltd. Contents page. Introduction.2. Features of Merlin. 3

Aespire 7100 Essential performance Compact design

PATIENT SETUP QUICK REFERENCE GUIDE

Transcription:

BiPAP Vision Ventilator Testing with Various Patient Interfaces Report 48086 February 18, 2002 Prepared for ResMed by Valley Inspired Products, LLC inspiredrc.com

Valley Inspired Products, LLC Revision Date: February 18, 2002 Origination Date: February 18, 2002 Title: Vision Ventilator Interface Testing Document Number: 48086 Originator: PLB Revision Level: A Background: Interfaces Tested: The Respironics BiPAP Vision is a bilevel ventilator which is used with a passive exhalation port (leak). The Vision uses a calibration step during setup to determine the magnitude of this intentional leak. In this way, it can distinguish between the intentional leak and unintentional or patient leak. The Vision will display the estimated patient leak if the leak test has been performed. The purpose of this test is to evaluate the effect of not performing the leak test and/or performing the test with different exhalation ports combined with various mask interfaces. The ventilator reported tidal volume and leak flow, operation of low minute volume alarm and ability to deliver adequate pressure will be tested. Description with Full Face Mask with Integral Exhalation Port with Integral Exhalation Port Figure 1. Units Tested Page 2 of 11

Methods: Equipment Breathing Simulator (Hans Rudolph series 1100) Respironics BiPap Vision Ventilator Model 582059 Mass Flowmeter TSI Model 4040 Water Manometer Dwyer U-tube Procedure 1. Tidal Volume/Leak Estimation Test Install a flat plate to the patient side of each interface to be tested, with a 22 mm connector, so that the interface can be connected to the breathing simulator. Set up Equipment per figure 2. Hans Rudolph Series 1101 Breathing Simulator Ventilator Actuator Computer 6 ft - 22mm Patient Circuit Pressure Transducer Bellows Pressure Transducer Interface/ Exhalation Port Pneumotachometer Figure 2. Test Setup Calibrate all instruments. Set the breathing simulator with the following settings: Resistance 20 cmh2o/liter/sec Compliance 60 ml/cmh2o Frequency 20/min Amplitude 5 cmh2o Slope 10 Insp. % 34 Page 3 of 11

Set the Vision with the following settings IPAP 20 cmh2o EPAP 4 cmh2o Freq. 12/min Rise Time 0.1 sec Perform the leak test after turning on the Vision. Perform the tidal volume/leak test by recording 10 consecutive tidal volumes as reported by the breathing simulator and by the Vision. Record the estimated leak as reported by the Vision. Test the following combinations of interface used for ventilation/exhalation and interface used for the leak test. In some cases, the leak test is not performed at all. Exhalation port used for Leak Test Interface used for ventilation + Full Face Mask No test performed + Full Face Mask No test performed No test performed Elbow Elbow Record patient flow, effort, and airway pressure for three breaths in each test condition. Add an Rp 20 resistor in line between the interface and the breathing simulator. Change amplitude to 10. For each combination of interface and exhalation port, record three breaths data. Calculate the inspiratory volume generated by patient effort before the ventilator is triggered from EPAP to IPAP. This method of trigger evaluation is chosen because the ventilator is designed to trigger when the measured inspiratory volume is 6 ml. Page 4 of 11

2. Low Minute Ventilation Alarm Test Set the breathing simulator with the following settings: Resistance 30 cmh2o/liter/sec Compliance 40 ml/cmh2o Frequency 12/min Amplitude 4 cmh2o Slope 10 Insp. % 34 Set the Vision with the following settings IPAP 10 cmh2o EPAP 4 cmh2o Freq. 4/min Rise Time 0.1 sec Low Minute Ventilation Alarm 3 liters/min Reduce breathing frequency on the breathing simulator to assure that the low MV alarm sounds. Perform for the following combinations of leak. Exhalation port used for Leak Test Elbow Elbow Interface used for ventilation + Full Face Mask Page 5 of 11

3. High Flow/Pressure Test Set the breathing simulator with the following settings: Resistance 15 cmh2o/liter/sec Compliance 40 ml/cmh2o Frequency 8/min Amplitude 6 cmh2o Slope 10 Insp. % 34 Set the Vision with the following settings IPAP 35 cmh2o EPAP 4 cmh2o Freq. 4/min Rise Time 0.1 sec Run the leak test on the Vision using the Elbow. Install the between the breathing simulator and the ventilator. This interface was selected because testing showed it to have the highest exhalation port flow. Record patient flow and airway pressure during 3 breaths. Repeat after installing an additional.190 leak between the patient circuit and the mask, to simulate a significant unintentional mask leak. 4. Leak Flow Characterization Connect a flow generator to each interface with the TSI flowmeter and monometer interposed. Plug the patient side of the interface. Vary pressure from 4 to 36 cmh2o, recording resultant leak flow. Page 6 of 11

Results: 1. Tidal Volume / Leak Estimation / Triggering Exhalation Port used for Leak Test Respironics Circuit Respironics Circuit Respironics Circuit No test performed No test performed No test performed ResMed Full Face Mask ResMed Nasal Mask Elbow ResMed Nasal Mask Elbow Interface used for ventilation Reported Leak lpm Vision Tidal Volume Simulator Tidal Volume Deviation % + Full Face Mask 2 695 695 0.0% ResMed Full Face 18 714 712 0.3% Mask 21 696 695 0.1% + Full Face Mask 19 702 704-0.3% ResMed Full Face 35 712 719-1.0% Mask 38 691 699-1.1% ResMed Full Face Mask 1 707 714-0.9% ResMed Full Face 2 702 690 1.8% Mask 7 684 705-2.9% Tidal volumes shown are the average of 10 breaths for both the Vision and the Breathing Simulator. Page 7 of 11

The following graph shows the leak reported by the ventilator. When no leak test was performed, the Vision s reported leak is labeled Total Leak, when the leak test is performed it is labeled Patient Leak. Ventilator Reported Leak 40 35 30 ResMed Full ResMed Nasal Respironics Full 25 lpm 20 15 10 5 0 None ResMed Nasal Elbow Leak Test Done with - ResMed Full The following graph shows the inspiratory volume required from patient effort before the ventilator cycled from EPAP to IPAP. Results are the average of 3 breaths. Inspiratory Trigger Volume 12 10 ResMed Full ResMed Nasal Respironics Full 8 ml 6 4 2 0 None ResMed Nasal Elbow Leak test done with - Page 8 of 11

2. Low Minute Volume Alarm Test In all combinations of interfaces and exhalation ports, the low minute volume alarm was triggered with simulator reported minute volumes in the range of 2.5 to 3 liters per minute. The Vision reports minute volume as an integer (no data to the right of the decimal), so it is difficult to judge the accuracy of the minute volume indicator in this range, but the indicator showed either 2 or 3 liters per minute during all tests. The effect of not performing the leak test on the low minute volume alarm was not evaluated. 3. High Flow/Pressure Test In both test conditions (using, with and without added unintentional leak), the ventilator achieved the desired pressure. The following is a graph of the patient flow and the airway pressure for both conditions. Airway Pressure, cmh2o 40 35 30 25 20 15 10 Pressure with intentional leak only Pressure with unintentional leak Patient Flow 120 90 60 30 0-30 -60 Patient Flow, lpm 5-90 0 0 1 2 3 4 5 6 sec -120 Page 9 of 11

4. Leak Characterization The following graph shows the measured leak flow at different pressures. The maximum intentional leak was achieved when using the. The lowest leak was with the and Full Face Mask. Also included in the chart are the values from the combined with the.190 diameter unintentional patient leak, which was used in the high flow/pressure test. 140 Flow, lpm 120 100 80 60 40 + Leak ResMed Full Face ResMed Nasal Elbow 20 + Full Face 0 0 5 10 15 20 25 30 35 40 Pressure, cmh2o Page 10 of 11

Observations Use of various patient interfaces with the BiPAP Vision ventilator during this bench comparison did not affect its ability to accurately measure tidal volumes, trigger breaths, trigger the low minute volume alarm or deliver adequate pressure. It appears that the leak test has the principal purpose of allowing measurement of patient leak separate from the intentional leak. Caution should be used in extrapolating these bench results into the clinical setting, with a much wider variety of patient conditions. Measures should always be taken to assure that patients are receiving adequate ventilation. Testing was completed at the offices of Valley Inspired Products, LLC in Burnsville MN February 5 th through 15 th 2002. Signed: Peter L. Bliss Technical Director February 19, 2002 Page 11 of 11