DXH Troubleshooting FlowCharts

Similar documents
Quick Reference Guide

COULTER A C T diff Analyzer. Operator s Guide. diff. PN DB (March 2016) Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA U.S.A.

COULTER LH 500 HEMATOLOGY ANALYZER

COULTER A C T diff 2 Analyzer. Operator s Guide. PN BA (August 2010) Beckman Coulter, Inc. 250 S. Kraemer Blvd.

COULTER LH 500 HEMATOLOGY ANALYZER

COULTER LH 750 HEMATOLOGY ANALYZER COULTER LH 780 HEMATOLOGY ANALYZER COULTER LH SLIDEMAKER COULTER LH SLIDESTAINER Training Guide

Automated Liquid Handling Station

Hemoglobin Reference Failure Troubleshooting

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

HumaCount 30 TS and 60 TS

Reference method for the enumeration of erythrocytes and leucocytes

Observations Possible Causes Recommended Actions. Turquoise color

Navigator 600 Silica analyzers

Proficiency Testing Corrective Action Checklist: PT PROVIDER: PT EVENT: TEST:

Mass Spec will not pump down

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

ABX Micros CRP 200. Technical manual. P/n: RAA028A. HORIBA ABX B.P MONTPELLIER Cedex 4 - FRANCE

Hemoglobin g/dl Female Male 12 years and older 11.9 to to 17.5

Quality Control of Pulmonary Diagnostic Systems

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

AFDXXX(X)AC Series Operators Manual Please read this manual thoroughly before attempting to operate your water maker.

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

6800 Maintenance Instruction System Flush Procedure

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

Overview. SEMEMI216 SQA Unit Code HF4G 04 Carrying out fault location on fluid power equipment and circuits

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

OAI Model 200 Tabletop Mask Aligner Portland State University

HemoCue Hb Procedure Template

Rotary Abreast Automatic Teat Sprayer User Manual. Technical Manual Revision 0.01

Balloon Gastrostomy (BG) Tube

Medonic M-series Art no May 2006

SURVEYING THE HAZARDOUS MATERIALS INCIDENT

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Please refer F 90 of the present manual for the identification of the valves used in the detector.

UsER manual for Watersens ph -REDOX

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER

Gas Network Craftsperson

Quality Management Determined from Risk Assessment

Mass Spec will not Autotune

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Convergent Technologys product overview

BD Cytopeia Fluidic Kit User s Guide

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Liquid Handling Systems

Multi-Probe Sampling System User Guide & Manual

Guidelines. Milan Special School District. Automated External Defibrillator (AED) Program

SEMPEO SQA Unit Code FP2P 04 Maintaining fluid power equipment

6900 Maintenance Instruction System Flush

no peaks missing peaks too small peaks increasing retention decreasing retention

PositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible

Technical Procedure for General Laboratory Equipment

PH01 Perfusion Cannula Manual

Mix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink).

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

EAS Basic Soil Gas Sampler

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Using the Akta Prime plus October 22, 2012

Agilent Auxiliary Gas Module AGM 2. User s Guide

PP-1, PP-1W, PP-2, & PP-3

Cartridge Based & Point-Of-Care Blood Gas Analyser QAP Performances & Quality Management Systems.. Issues

CBMTG Laboratory Committee-Laboratory Guidance

Thermo Environmental 48C Carbon Monoxide QA Plan Section I Electronic Calibration Branch Responsibilities

The benefits of the extended diagnostics feature. Compact, well-proven, and flexible

UNIVERSITY OF ROCHESTER MEDICAL CENTER

Glove Box Installation Manual

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

1. Introduction Why use the Flow-Through Cell? Component Identification... 1

Powers Controls RC 195 Multiple Input Receiver- Controller

Agilent 1290 Infinity Pump Head Maintenance

SEMEM SQA Unit Code HWAW 04 Carrying out maintenance activities on fluid power equipment

Headspace Troubleshooting Made Easy. Presented today by: Susan Scheib Sr. Technical Support Specialist Joe Barta FSE Chicago & Escalation Engineer

Data Sheet T 8389 EN. Series 3730 and 3731 Types , , , and. EXPERTplus Valve Diagnostic

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

PerkinElmer Aanalyst100

Cover Page for Lab Report Group Portion. Pump Performance

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

LAMINATED SAND FILTERS Filtering Equipment. Manufactured in Spain for PQS Piscinas y Consumo SA

PROCESS CONTROL HANDBOOK Z-6

OPERATING INSTRUCTIONS Sherlock Chamber Testing System

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Rhino VE CE Bulk Unloader Operator s Card

Implementing IEC Standards for Safety Instrumented Systems

ACQUITY UPLC Sample Manager - Fixed Loop Operator s Overview and Maintenance Information

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

LENNOX SLP98UHV DIAGNOSTIC CODES

Twin & Triple Control Concealed Thermostatic Shower Valve

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

SEMPEO SQA Unit Code FP2M 04 Maintaining mechanical devices and equipment

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

CGMP KITS PROTOCOL ASSAY PRINCIPLE

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

Transcription:

DXH Troubleshooting FlowCharts WARNINGS AND PRECAUTIONS Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate instrument. Beckman Coulter, INC., urges its customers to comply with all national health and safety standards such as the use of barrier protection. This may include, but it is not limited to, protective eyewear, gloves and suitable laboratory attire when operating or maintaining this or any other automated laboratory equipment. INTENTION FOR USE This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any other manual.

Daily Checks Background fails Still fails which parameter? WBC RBC HGB PLT DIFF NRBC RETIC Zap Apertures x3 Clear appropriate Apertures (WBC or RBC) through Dx Tools>Maintenance tab Clean Apertures, Perform HGB Blank Verification, Likely causes: - Microbubbles in sweepflow lines - Electronic noise, or -Diluent contamination Prime Sweepflow Rinse Mix Chamber Go to Dx Tools>VCSn tab, select Fluidic>Rinse mix Chamber>Start Perform in order, as needed: - Flush Flow Cell (no cleaner) Both, (Repeat 5x) - Flush Flow Cell with cleaner (5 mins) - Clear Flow Cell Aperture 3x Clean Apertures, Shutdown for 30 mins, Still fails Run a cassette load of samples. Repeat Daily Checks Still fails When was the diluent last changed? Prime diluent Still fails Coulter Customer Support Center Run cassette of samples, Shutdown for 30 mins, Diff, NRBC still fails Retic still fails Prime the Retic reagents (Dx tools>common Services tab) 2

QC 6C, Retic-X Which results are affected? CBC Parameters? (WBC, HGB, RBC, PLT, DIFF, NRBC, RETIC Parameters (%) Refer to QC 1 Refer to QC 2

QC 1 CBC parameters (WBC, RBC, HGB, MCV, PLT, MPV) Out of Range? R Flags? Voteouts? Parameter affected in same way for each level (ie all running H or L; or showing a bias in same direction, but within range)? Reagent/s changed recently? All levels of QC affected and results confirmed by running new vials of each Are the other levels showing a bias on the same side of the mean? New lot? Try another box of reagent t Resolved Resolved Coulter Customer Support Center to advise of findings and request replacement Trending noted? Sudden shift? (except for MCV which is expected to trend up over time) - Has calibration factor been changed recently? - Has PMI been done recently? - Has FSE visited for an issue and made adjustments (changed hardware etc)? Check calibration and adjust as needed Pass Pass Pass Possible dilution issue?hardware - Clean outside of BSV - Clear appropriate Apertures (in Dx Tools>Maintenance) - Perform repeatability Fail Repeat repeatability Fail - Perform Clean Apertures (in Dx Tools>Maintenance) - Perform repeatability Fail Repeat repeatability Fail New vial gives same result? New lot of QC? Lab can set own target (as per IFU) as long as CVs are within performance specifications. Contact Beckman Coulter Customer Support Center to advise of findings Coulter Customer Support Center 4

QC 2 Out of Range? R Flags? Voteouts? DIFF, NRBC, RETIC Reagent/s changed recently? All levels affected? Try new vial of same level. Issue persists? New lot? Try another box of reagent Issue persists R flags? Results out of range or no results? Vial related. System is okay to run. Coulter Customer Support Center to advise of findings and request replacement Issue resolved Perform Rinse Mix Chamber Go to Dx Tools>VCSn tab, select Fluidic>Rinse mix Chamber>Start When complete, Finish Diagnostics Results in range? Are patient samples giving R flags? System is okay to run (Refer to 6C QC IFU> Expected Results section) New kit lot of QC? Coulter Customer Support Center to advise of findings Perform in order, as needed: - Flush Flow Cell (no cleaner) Go to Dx Tools>Maintenance tab, select Flush Flow Cell.>Choose Both >Start (Repeat 5x) - Flush Flow Cell with Cleaner for 5 mins - Clear Flow Cell Aperture When complete, Finish Diagnostics Issue persists 5

SAMPLES Which results are affected? All Parameters? (WBC, HGB, RBC, PLT, DIFF, NRBC, RETIC) CBC parameters only? (WBC, HB, RBC, PLT) AMTC module parameters only? (DIFF, NRBC, RETIC) Refer to Samples 1 Refer to Samples 2 Refer to Samples 3

SAMPLES 1 All Parameters (WBC, HGB, RBC, PLT, DIFF% #, NRBC% #, RETIC% #) Check the Sample Aspiration Probe - has it been recently replaced? Check the aspiration line connected at the top of the probe. Check the probe is seated & screwed in correctly. Ver 1.3 (vember 2016) Possible blocked Sample Probe OR Sample Aspiration Syringe /Motor issue Is blood being pulled through to the Blood Sample Valve? Is Sample Aspiration Syringe moving up & down freely? Signs of leaking around aspiration syringe? Replace Sample Aspiration Probe Confirm which results are affected (cannot be all) and refer to appropriate flowchart for further troubleshooting Check thumbscrew at bottom of syringe is finger tight. - If already tight, syringe needs replacing. Coulter Customer Support Center - If loose, tighten and check results again Could be faulty syringe motor or syringe itself. Coulter Customer Support Center Beckman Coulter Technical Training 7

SAMPLES 2 CBC parameters only (WBC, HGB, RBC, PLT) Is the BSV moving during the analysis cycle? Go to Dx Tools>CBC tab, select Mechanical>Cycle BSV. Does the BSV respond? Coulter Customer Support Center Voteouts? (Partial or Total) Which parameter/s? ALL Plt only - Clean outside of BSV -Perform Cycle BSV Still no CBC results? Is Sample Aspiration Syringe moving up & down freely? WBC or RBC only HGB only WBC & HB RBC & PLT Check PLT histogram spike at 2fL? Possible causes: - microbubbles in sweepflow line, prime sweepflow - electronic noise, check cover in place - possible diluent contamination, change diluent Voteouts? (Partial or Total) result Perform Zap Apertures x3 Voteouts persist Is HGB lamp on when analyser is powered on? Perform Clean Apertures and HGB Blank Verification Problem persists Clear appropriate Apertures (RBC or WBC) (Dx tools >Maintenance tab) 3x Clean Apertures (Dx tools >Maintenance tab) - Check when DxH Lyse was last changed and replace if coincides with onset of the problem. Contact Beckman Coulter if issue not resolved. - Possible dilution issue (if both parameters are affected in the same direction (ie H or L) - Check BSV and appropriate pump for movement - Clean outside of BSV - Perform Cycle BSV from Dx Tools> Maintenance - Unresolved? Coulter Customer Support Center Could be faulty syringe motor or syringe itself. Check thumbscrew at bottom of aspiration syringe is finger tight and look for signs of leaking - If loose, tighten and check results again - If already tight, syringe needs replacing. - Coulter Support Center Voteouts persist Voteouts persist Coulter Support Ver 1.3 (vember 2016) Beckman Coulter Technical Training 8

SAMPLES 3 Check SAM. Is sample being dispensed to the chambers? Is the aspiration Syringe moving up & down freely? ALL no results, low events, R flags AMTC parameters only* (DIFF, NRBC, RETIC) Perform in order, as needed: DIFF only NRBC only RETIC only Pathway or Reagent specific *te if you observe %, but not absolute values, then the issue is with the 1 o parameter used for calculating the abs value (eg WBC for DIFF, RBC for RETIC). DV or Flow Cell Issue Cycle DV Go to Dx Tools>CBC tab, select Mechanical>Cycle DV. Does the DV respond? Coulter Customer Support Center Still no results - Flush Flow Cell (no cleaner) Go to Dx Tools>Maintenance tab, select Flush Flow Cell.>Choose Both >Start (repeat 5x) - Flush Flow Cell (with Cleaner) for 5 mins (repeat if needed) - Clear Flow Cell Aperture Check counts (% and # affected?), flags (R?) and error messages Perform Rinse Mix Chamber Go to Dx Tools>VCSn tab, select Fluidic>Rinse mix Chamber>Start. If not resolved, proceed to next step. Low events, Low event rate, Required parameters for generating dataplots missing, R flags High event rate, R flags Unlock DV Go to Dx Tools>CBC tab, select Mechanical>Unlock DV. Does the DV respond? Possible blockage inside DV or input line of Flowcell Perform in order, as needed: - Flush Flow Cell (no cleaner) Go to Dx Tools>Maintenance tab, select Flush - Flow Cell.>Choose Both >Start (repeat 5x) - Repeat Flush Flow Cell with Cleaner for 5 mins (repeat if needed) - Clear Flow Cell Aperture Check reagents. - Near end of open expiry? - Nearly empty? - Changed recently? Change relevant reagent pack Ver 1.3 (vember 2016) Coulter Customer Support Center Issue persists Beckman Coulter Technical Training 9