Neotecha SAPRO - Tank Valve Type ST for aseptic sampling with bottles Installation, Operation and Maintenance Instruction

Similar documents
Neotecha SAPRO - Aseptic sampling valve SV Installation, operation and maintenance instructions

Neotecha Tank Sampling Sapro

Neotecha Tank Sampling Sapro

ph measuring system for hygienic process applications

Constant Pressure Crude Oil Container Model CPCCP

Instructions for Use. Plug connection for medical gas and vacuum acc. to DIN / 2004 DIN EN ISO

Reduce pressure zone device suitable for high and medium hazard rated applications Flanged end connections

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

Instruction presentation Special valves

FITTINGS AND VALVES Düker Plunger Valve RKV Type 7015 Professional control of flow rates, reservoir levels and pressures Safe, hygienic and durable!

Differential pressure controller (PN 16) AHP - return mounting, adjustable setting

Pressure relief valve DHV 712 DN 65-80: 0,5-10 bar, DN : 0,3-4 bar, DN 100: 0,5-6 bar

Premium Series. Economical Price & Reliable Performance. High Quality. Rust-Resistant. Long Life. Tamper-Resistant locking Screw.

CAST IRON SAFETY VALVE TYPE 6301

REPAIR PART NUMBER NUMBER DESCRIPTION 0137 ELE 0136 ELE 12 VOLT SOLENOID

Ball valve HKSF-W100. Ball valve HKSF-W100. RMA Kehl GmbH & Co. KG Oststrasse 17 D Kehl / Germany

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

HOVAP SERIES 9600 BASICFLOW PROCESS VALVES INSTRUCTION MANUAL

Model VR6 System. Installation, Operation & Maintenance

Pressure Reducing Valve DMV 755

Pressure Reducing Valve DMV 755 set range: 1,0-9,0 bar

Operating instructions - Translation of the original -

G type DUCTED EXHAUST SAFETY VALVE 2871 AND 288X SERIES. Model/Ref:

Contents. 1. General information BROEN BALLOMAX Steel Ball Valves Approvals Quality Management...2

Pressure Reducing Valve DMV 750 set range: bar

Tank Blanketing Pressure Regulators RHPS Series

MAINTENANCE PROCEDURE FOR X 650

Sanitary pneumatic retractable fitting for use with Unical 9000 and Uniclean 900 automated control systems

Solenoid valves VZWF, force pilot operated

Pressure Reducing Valve DMV 750

Pressure relief valve DHV 716 set range: 0,5-10,0 bar

Pressure Relief Valve DHV 718

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

1700ESL Kamvalok Coupler

Installation Instructions

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN Original manual

CUT OFF TOOL MODEL: CAT113

Crosby style JCE Safety Valve Installation, Maintenance and Adjustment Instructions CROSBY

Type ACE97. Introduction. Installation. P.E.D. Categories. Specifications. Overpressure Protection. Installation Guide English May 2002

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Mounting and operating instructions EB 2530 EN. Self-operated Pressure Regulator. Pressure Reducing Valve Type M 44-2

ATS430 turbidity sensor Retractable insertion assembly

Pressure relief valve DHV 725 set range: 0,2-10,0 bar

Spare parts list. ( ) Spare parts list HD 3.0/20 C Ea

STAINLESS STEEL LOW PRESSURE REDUCING VALVE LPRV ELITE

Installation, Operation and Maintenance Manual for Back Pressure Regulator

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

AVP-F. p setting range (bar) Code No. 003H6200

Reduce pressure zone device suitable for high and medium hazard rated applications BSP screwed connections

Atmospheric relief valve type 1100 Installation and maintenance instructions

Pressure Reducing Valve

Pressure/Vacuum Diaphragm Valve PROTEGO UB/SF. deflagration- and endurance burning-proof

Armatures for analytical probes

ITT RICHTER CHEMIE-TECHNIK

Series 44 Self-operated Pressure Regulators. Type 44-1 B Pressure Reducing Valve. Type 44-6 B Excess Pressure Valve

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

Purge Star & Ring Purge Systems

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

VSR modulating valves with membrane actuator

Parts Manual STRIKER Striker 2840 Parts List - PN Printed in USA 03/18/08

Mounting and Operating Instructions EB 2172 EN. Series 43 Temperature Regulators. Type 43-5 and Type Type Type 43-6

A soft sealed atmospheric safety valve with large flow capacity and soft seat tightness for steam condenser and turbine applications

STAINLESS STEEL OVERFLOW VALVE TYPE 417

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

CFC2. Float Control For Closed Tanks MODEL. Specifications. Dimensions (In Inches) Reservoir Connection 1" NPT (Typ. Both Ends 2 Places)

// ADapters GUIDelines

Type 2080 FALTENBALGVENTIL BELLOWS VALVE VANNE A SOUFFLET. Operating Instructions. Bedienungsanleitung Manuel d utilisation

6301 TYPE CAST IRON SAFETY VALVES

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Description Total stroke Maximum Recommended Page in mm gripping force in N workpiece weight in kg

Spare parts list. ( ) Spare parts list HD 3.0/20 C Ea

Manual Actuated Boiler Blowdown Valves

Hydraulic Punch Drivers

Hydraulic comparison test pump Models CPP1000-X, CPP1600-X

Manual retractable fitting with safe-to-operate rotary drive

ATV 90 Y-12 YOUTH 2-STROKE RED (A2004ATB2BUSR) Page 1 of 52 A-ARM, FLOOR PANEL, AND BUMPER ASSEMBLY

Differential pressure controller (PN 25) AVP - return and flow mounting, adjustable setting

Operating instructions for KLINGER SCHÖNEBERG ball valves type RK-Chemoball, type RK-Proball and type RK-Ecoball

Tank Blanketing Regulators

K Page 1 of 48 Air Intake - Pg. 17

BSA6T and BSA64T Stainless Steel Bellows Sealed Stop Valves Installation and Maintenance Instructions

ASSEMBLY & USE MANUAL

Instruction Manual Contact Pressure Vacuum Gauge

CLASS TLP Pressure Reducing Valve

Materials : Stainless steel ASTM A351 CF8M

Differential pressure and flow controller (PN 16) AVPQ - return mounting, adjustable setting

DISPENSING VALVE MODEL VDP100 INSTRUCTION MANUAL

Flexi-pro Aerosol Delivery System (FADS) Manual

DS05C,D,G Dial Set Pressure Regulating Valves

Size : Ends : Min Temperature : Max Temperature : Materials : Cast iron body

MUELLER. A Wall Type. Indicator Post. Reliable Connections. General Information 2. Technical Data/ Dimensions 3. Installation 4-5.

Telefon (+45) Telefax (+45)

Rp [ ] Sv min. [ ] C215QP-B 15 1/ C215QP-D 15 1/ C220QP-F 20 3/ [ ]

Pilot valves for pilot operated servo valves

Tank Blanketing Regulators

Spare parts list ( ) Spare parts list HD 2.3/15 C Ed/ 3.0/20

TYPE A Male camlock-female BSPP

Fisher Lever Lock Handlever Actuators for POSI SEAL A81 Rotary Valves

BR 27g. Operating, assembly and maintenance instructions for discontinuous sampling valve. 1. Design, operation and dimensions:

Transcription:

7 Before installation these instructions must be fully read and understood 1.4 Illegal technical changes It is illegal to do mechanical changes of the design of the Sapro tank valve, or to use parts of other manufacturers for repairs, otherwise Neotecha can not guarantee for the safety of the valve. 1.5 Warning of misuse The Sapro tank valve with all connection pipes and actuators can never be used to climb up or to stand on it. 1.6 Danger sources Chemical external The sleeve on the actuator should not be cleaned with solvents, Use alcohol or a mild detergent. Valves for aseptic sampling of liquid mediums stored in tanks. Please read this installation, operation and maintenance instruction carefully before you handle the tank valve! 1 Safety Please read these instructions carefully. 1.1 General danger potential Not paying attention to this instruction Incautious use or misuse Unqualified personnel or staff 1.2 Use as directed The Sapro tank valve is used for aseptic sampling from tanks into bottles. The pressure range is between no pressure (about 0,1 bar) and 10 bar maximum. All material in contact with the product is stainless steel 1.4435 (316L). The soft seals are FPM (FKM). 1.3 Restricted use The valve can be used only with products when the materials used (1.4435 and FPM) are compatible. Mechanically Enough space is needed to handle the tank valve, to prevent of squeezing the hands while operating the valve Electrically To prevent static charges, the entire tank valve is made of conductive metal. Thermically With a sample medium temperature range between -10 C to 160 C, the surface of the tank valve may assume the same temperature. It is therefore necessary to wear gloves to avoid injuries when operating the valve. Dynamically The valve is equipped with a dead man handle (spring return). The handle will spring back by itself when let go. This can cause injuries to the operators hand. Therefore do not let the handle go when enough sample is taken. 1.7 Requests for the operators To ensure proper and consistent sample taking the operator is to be trained as per users sample taking instructions www.valves.emerson.com 2017 Emerson. All rights reserved. VCIOM-01982-EN 15/09

2 General data 2.1 Dimensions ø 35 Tank wall Sample medium side Valve I-clamp Drains ± 405 ± 300 Tri-clamp ± 251 Overall height: 322 (dummy) / 410 (bottle adapter) 2.2 Technical data Pressure range: 0.1 bar to 10 bar Temperature range: -10 C to 160 C Bottle connection: ISO GL45 For use with bottles: 100 ml to 2000 ml Size I-clamp: 3 Cherry Burrell Size Tri-clamp: 2½ 3 Transportation and storage 3.1 Transportation The Sapro tank valve is shipped in a box. Please leave the unit in this box until it is ready to be installed. 3.2 Storage conditions The Sapro tank valve has to be stored in a dry, dust-free place until it s installed on the tank. 4 Installation The tank has to be empty before any installation or repair of the valve can take place! Preparation The tank needs a 3 I-Line connection flange, welded to the outer surface of the tank wall. The tank valve with the wet part is mounted with the I-clamp according following illustration: I-flange welded to the tank Tank wall Tank valve cpl. I-clamp 2

5 Detailed description of the tank valve and actuator 5.1 Illustration Tank valve (wet part) Reservoir Pressure gage V1 C1 V3 Drain 3 V2 V4 Drain 5 Handlever V5 Drain 4 Bottle adapter Dummy Sampling bottle 5.2 Connections and valves C1: Customer connection (Swagelok ⅛ ) for: a. Steam inlet to sterilize the sample passage in the actuator and bottle adapter in pos.1. b. Inlet for any cleaning liquids before taking samples, depending on the medium to sample. V1: Ball valve to control the inlet of the customer connection C1. V2: By opening this valve the alcohol from the reservoir fills the inner part of the tank valve body. V3: Drain valve to control the filling of the tank valve body from the reservoir when no samples are taken. The body is only filled entirely when the alcohol flows through this valve down drain 3. V4: Drain valve for stem blocking media. By taking an alcohol sample at this point the state of blocking media and the stem seal can be monitored. V5: Actuator drain valve. Prior to sterilizing the sample passage this valve must be opened and closed again when sterilizing process is completed. Reservoir: The purpose of the reservoir is to ensure a constant head of blocking media (alcohol) around the operator stem. Bottle adapter: Aseptic sampling adapter for bottles with ISO GL45. Pressure gage: Pressure gage to control the pressure coming from the customer connection C1. 3

5.3 Description of the bayonet positions When bottle adapter is connected to pos. 2 the valve on the adapter is open. Pos. 2 Pos. 1 In this position steam can be put through the product passage and exit through the drain valve V5. 5.4 Description of the sample / blocking medium sides Sample side Blocking side Sample medium Blocking medium (alcohol) Bearing and seal Bottle adapter valve open 6 Maintenance 6.1 Cleaning Though all parts of the Sapro tank valve are made of SS, they can be cleaned with normal cleaning liquid. The s can also be cleaned with normal cleaning liquid. 6.2 How to check the sample / blocking medium seal The bearing is filled with the blocking medium (alcohol) coming from the reservoir and can be checked by opening V3 to take a sample of the alcohol. If the alcohol is polluted with the sample medium, the seal needs to be exchanged. 4

6.3 Exchange of the seat seal and The tank has to be empty before any installation or repair of the valve can take place! The valve seat, seat seal and the can be dismantled by removing the four screws of the Tri-clamp flange. The defective parts can now be exchanged, please check all surfaces for scratches. When you screw on the Tri-clamp flange again, please attach importance to the gap (only about 0.3 mm) between the Tri-clamp flange and the tank valve is the same all around. I-clamp Seat seal Valve seat Tri-clamp flange Tank valve Screws 7 3D Explosion drawings 7.1 Tank valve with actuator Seat seal Valve seat Screw cap Reservoir Body Roller Handlever with grip Tank valve (wet part) Tri-clamp flange Bearing with s Shaft Spring support Locking pin with chain s s Flange Bearing sleeve Coil springs Bayonet Bearing Tie rod Bonnet Protective sleeve (Detail view) s 5

7 7.2 Dummy 7.3 Bottle adapter X-ring X-ring Interlocking screw Pins Adaptor top Pins Spindle seal X-ring Plug Operation pin Valve spindle Body Threaded insert Coil spring Grip Grip Grips Screw Nipple Bottle adapter Diaphragm Washer Sampling bottle Screw 2017 Emerson. All rights reserved. 6