Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Similar documents
AKTA pure 25 New Owner s Intro

Operating Instructions Multi-Channel Vacuum Degasser

The instrument will process the samples one by one till all the samples have been cleaned up.

SDS 9414i HPLC Pump. 1. Standard Accessories Stainless Steel Version PEEK Version Safety Instructions...

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

SOR-100 Solvent Racks

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Columbus Instruments

Degassing in HPLC and Related Technologies. Operator s Manual

6900 Maintenance Instruction System Flush

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

Filtration FILTRATION PRODUCT TIPS

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

Degasser DG 2.1S User Manual V6880 HPLC

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

VACCUPERM Vacuum chlorine gas dosing systems

Flow meter. bellow vaporizer. APL valve. Scavenging system

H 2. Sensor Series Take no risk! Install a Hydrogen Sensor in your Gaschromatograph

The New PHD 4 Portable Helium Detector

UT 3000 Mercury Ultratracer

215 Liquid Handler Up-Grade Manual

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

Thermo Scientific Model 146i Principle of Operation

PH01 Perfusion Cannula Manual

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Degassing ERC KK Saitama, Japan

OPERATION MANUAL NTF-15

BSS 2 Bottle Sampling System. Instruction manual Version 1.8

ISM-3 User Manual PBI D 04/2013

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

ÄKTAprime. User Manual

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

Using the Akta Prime plus October 22, 2012

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Constant Pressure Inlet (CCN) Operator Manual

Peak Biotech - Company & Product Presentation

MANUAL DIRECT PURGE OPTION. UNION Instruments GmbH CWD2005 PLUS. General information, safety standards and regulations for direct purge option

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

Type C-CUT. Quickstart. English Deutsch Français

LC-SPE. Organizer 3.0 User Manual. Innovation with Integrity. Version 001 NMR

Keep the balance in sight

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Your local gas generation partner. Precision series Modular gas generation solution for GC.

490 Micro GC Site Preparation Checklist

Digital Melting Point Apparatus

Oxymax COS61/COS61D. Technical Information

6800 Maintenance Instruction System Flush Procedure

OPERATION MANUAL NTF-60 Plus

Instruction Manual Dräger MSI P7 and MSI P7 plus


DIAPHRAGM VACUUM PUMPS AND COMPRESSORS

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Customer Responsibilities

ALU-CD 12, ALU-CD 25, ALU-CD 35, ALU-CD

User Manual. Quantos Automated Dosing Liquid Module

GE Healthcare. ÄKTAprime plus. Operating Instructions. Original instructions

SOP: Buck Scientific BLC-20P HPLC Operation

Topline Peristaltic Dosing Pump for PAC Installation, Commissioning and O&M Manual

Membrane covered optical CO 2 sensor

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Used with Dissolution Tester for Fraction and Flow Measurements

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

Mass Flow Meter (MFM) for Gases

OXY 3690 MP. Measuring transducer for air oxygen

HumiSys HF High Flow RH Generator

Operating Manual ALU 7 - ALU 10 - ALU 20 - ALU 26 ALU 35 - ALU 60 - ALU 100

Mass Flow Controller (MFC) for Gases

Manual of SF6 Comprehensive Tester

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

On-Stream Tightness Testing of Vacuum Process Installations

BASIC PHYSICS APPLIED TO ANAESTHESIOLOGY

Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Overview. Front Panel: Keypad and Display

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Type 131. Roller base. Information and Operating Instructions for Dewar Vessels Type 131 up to 135 boxed shaped KGW-ISOTHERM

Traysealer T 100. Tray packaging in the smallest space

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

USER GUIDE. Fog Generator. ATMOSPHERE APS Series Haze Generator. MAX APS Series. No: , Revision A/d

UsER manual for Watersens ph -REDOX

THE UNIVENTOR 400 ANAESTHESIA UNIT

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

GasSense NDIR User Manual

Customer Responsibilities

IBEDA GAS MIXERS HIGHEST PRECISION FOR YOUR WELDING AND CUTTING PROCESSES

RIGGTEK-Info, Page 1 of 16 DissoPrep X8 FAQ s

6890 GC Site Preparation

THE POWERFUL LEAK DETECTOR FOR ALL APPLICATIONS IN POWER PLANTS

h2000 plus HyDROGEN LEAk DETECTOR

2B Technologies, Inc.

Transcription:

Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com

2

Manual Degasser DDG-75 3 Table of content Shipping List... 4 Chapter 1: Safety Instructions... 4 Chapter 2: Specifications... 5 Chapter 3: Instrument description... 5 Chapter 4: Installation... 8

4 Shipping List 1 pc Degasser (2, 3 or 4-channels) 1 pc Manual 1 pc Power cable 2 pcs PTFE tube (1.5 x 3.0 mm, 1m) for each installed degassing channel with fitting Chapter 1: Safety Instructions The manufacturer does not warrant for any defects or damage resulting from incorrect operation and maintenance, non-observance of the manual s instructions and negligence during installation. Before putting the instrument into operation, read carefully the manual and should there be any further questions, please get in contact with your supplier. 1.1 General Electrical Hazards 1. Check actual line voltage to confirm that the Set voltage (on the rear side; voltage selector and fuse carrier) of the instrument is correct. 2. Before changing the instrument s voltage or before changing defective fuses, disconnect the instrument from all power sources. 3. The instrument has to be plugged into grounded wall sockets only. 4. This instrument can be used only with other instruments which comply with the general safety regulations. 1.2 General Precautions 1. In order to avoid any damages, perform periodic leak checks on all installed supply lines. 2. The instrument is only allowed to be used for applications with specifications described under chapter 3. 3. For flammable and/or toxic solvent, follow a regulated approved waste disposal program. Never dispose of such products through the municipal sewage system. 4. The instrument is suitable for operation between 10 C and 35 C surrounding temperature. 5. For trouble-free and reliable operation, it is recommended to use only filtrated or particlefree samples for the degassing process. 6. The instrument shou1d be cleaned with appropriate cleaning agents.

Manual Degasser DDG-75 5 Chapter 2: Specifications Degassing method: With an applied vaccum, dissolved gases are continuously removed through a semi-permeable membrane Flowrates: max. 10 ml/min for each degassing channel Efficiency: 0.5 ppm oxygen at 0.5 ml/min Parts with solvent contact: PTFE and PEEK Analog output (for documentation of actual vacuum): 0-1 V for 0-999 mbar Error messages: LED neutral relay contact Volume per channel: 8 ml Power supply: 230/110 V 50/60 Hz Dimensions: 110 x 150 x 350 mm Weight: 5.6 kg Part numbers: 2-channels wo/display #945105 3-channels wo/display #945106 4-channels wo/display #945107 2-channels w/display #945155 3-channels w/display #945156 4-channels w/display #945157 Chapter 3: Instrument description The degassing of solvents is very important in liquid chromatography, as oxygen and other in the solvent dissolved gases affect the analysis for various reasons: through volume contraction (gradient mixing), different pressure ratio (in front and behind the column) and caused by different temperatures, the dissolved gases are set free in course of time. This will influence the reliability of the analysis (pump side) as well as the detection efficiency (measuring cell). Therefore, dissolved gases have to be removed from the solvent. The best known methods are ultrasonics and vacuum processing, flushing of the solvent with a stream of helium and vacuum degassing devices. The ultrasonic method is suitable for degassing organic eluents but is not adequate enough for hydrous solutions. For these kind of solutions, the solvent has to be degassed in a suction bottle placed under vacuum. But both of these methods are only short-term solutions as after the degassing process, the solvent will be continuously enriched with gases taken from the air. Furthermore, for gradient elution, the degassing quality is not sufficient enough. The known methods are ultrasonics and batch vacuum processing, flushing of the solvent with a stream of helium and continuous vacuum degassing devices. Continuously usable and quite simple to handle is the method of flushing the solvent with a stream of helium. Hereby, the stable remaining He-molecules are displacing all other gases. An equal maintained gas pressure buffer prevents the helium from escaping and gradient elutions can be carried out without any problems. The same solvent quality is achieved with the usage of a vacuum device. The great advantage when using the vacuum device is, that no further costs concerning the supply and consumption of helium arise. Furthermore, for some analysis methods, e.g. detection in the lower UV range or with RI-detection, the vacuum degassing method allows an Operation with lowest possible detection limits due to more smooth and stable base line runs.

6 3.1 Operation Principle In a vacuum chamber, the solvent to be degassed is distributed into several, parallel arranged, semi-permeable membrane chambers. The integrated membrane coils (each consists of about 10 PTFE capillaries with a length of approx. 10 m) is located in a vacuum chamber. When a high vacuum (more than -850 mbar) is applied, only the in the liquid dissolved gases are diffusing through the thin membrane into the vacuum chamber. The vacuum pump is controlled with a pressure switch, so a vacuum is constantly applied. With a total solvent volume of 8 ml per degassing channel, the degasser achieves an extremely low remaining gas concentration of 0.5 to 5 ppm oxygen, depending on the degassers flow rate (0 to 10 ml/min.). 3.2 Front panel The vaccum degasser is available for degassing 2, 3 or 4-channels. On the left side there are 2, 3 or 4 inlets of the solvents. On the right side there are 2, 3 or 4 outlets of the solvents (see Fig. 1 and 3). The LED display indicates: Green (Pump): The vacuum is below the set value. The integrated vaccum pump is in operation until the set value is reached. Yellow (Error): The set vacuum is not achieved after some time (i.e the pump is still working). If this error is displayed for a longer period of time, the pump will switch of automatically. Red (Power): Shows of the instrument is swith on. Fig. 1: Front panel

Manual Degasser DDG-75 7 3.3 Rear panel At the rear panel, the mains connection, the analog pressure output/the error message contacts are installed. Also, the fitting for directing possibly existing solvent traces to the laboratory s venti1ation can be found at the rear panel. 3.4 Maintenance When using the degasser for solvents with a high salt concentration, the system should be flushed with destillated water occasionally. Fig. 2: Rear panel #1: Power Supply #2: Power Switch #3: Voltage selector amd fuse carrier #4: Tube connection #5: Interface The tubing connection is for removing toxic solvents. A tubing should be installed to direct the solvent into the ventilation. The interface (9-pin Sub-D) allows the documentation of the actual pressure output and also the error messages: Pin 1 Analog pressure (0..1V) 2 not connected 3 Error relay (normally open) 4 Error relay (normally closed) 5 not connected 6 Analog pressue (0..1V) 7 not connected 8 Error relay COM 9 Error relay COM

8 Chapter 4: Installation The degasser is only allowed to be installed on the suction side of a HPLC pump or with gradient elution, positioned in front of the mixer. With the help of the delivered PTFE capillaries, each single solvent has to be connected to the fittings positioned at the right and left side of the coded channel. Through one of the capillaries, the solvent is transported from the reservoir into the degasser and the degassed solvent is leaving the degasser through the other capillary which is connected to the analysing system. The degassing efficiency is independent of the flow direction, however, it is recommended to have uniform connections for all of the solvent inlets and outlets. Before putting the degasser in operation, all connected capillaries have to be filled with the eluent. This is done after the installation with the help of the delivered syringe. The syringe is connected at the pump s relief valve and the solvent is sucked on as long as there is none air bubbles remaining in the capillaries. Already after a warm-up time of approx. 3 to 5 minutes, the degasser is ready for operation. Low pressure gradient (Fig.4): This shows an example of flow diagramm which can be applied for the degassing process with low pressure gradient operation. Here, both channels are connected to separate degasser channels. Multi-channel high pressure system (Fig. 5): Each channel of the degasser is connected with the suction valve of an HPLC pump. Intensive degassing system (Fig. 5): The degassing quality can be improved by coupling two or several membrane channels. Fig. 3: Degasser front panel

Manual Degasser DDG-75 9 Fig. 4: Low pressure gradient system with installed degasser Fig. 5: Multi-channel high pressure system Fig. 6: Intensive degassing process

10