Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Similar documents
ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

Warnings: Notes: Revised: January 8,

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

MJB-3 Mask Aligner Property of TAU MNCF August 2012

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

Standard Operating Manual

SOP for Karl Suss MJB3 #1 Mask Aligner

Warnings: Notes: Revised: October 5, 2015

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date:

Karl Suss MJB4 Mask Aligner

COBILT CA-800 Mask Aligner Equipment Operation

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Operation of the mask aligner MJB-55

Photolithography. Operating Instructions

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Karl Suss MA6 Mask Aligner SOP

OAI Model 200 Tabletop Mask Aligner Portland State University

Quintel Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

University of Minnesota Nano Fabrication Center

OPERATION OF THE DIMPLER

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

EE 432 Lab 3 PMOS source/drain lithography and diffusion

Standard Operating Manual

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

University of Minnesota Nano Fabrication Center

COMELEC C-30-S Parylene Coating System

Columbus Instruments

Revised: June 7, 2017

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

UNLOADING PROCEDURE OF AN AIRFLOW CONTAINER

VERTICAL AIR COMPRESSORS

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

March CS-1701F Reactive Ion Etcher

Eco Snow CO2 Clean System Operation

RH800 & 2000 BASIC OPERATIONS

University of MN, Minnesota Nano Center Standard Operating Procedure

2 Switching off ECRIS 3. 4 Acquire Beam Spectrum 4. 6 Vent Instrument Chamber 5. 7 Pump down Chamber 5. 8 Baking out Chamber 5.

VERTICAL AIR COMPRESSORS

Usage Policies Notebook for Parylene Coating System

R E D I C O N T R O L S

Operating Procedures for Metal Evaporator I

Equipment Operating Procedure Glove Box

Introduction to Pneumatics

[USING THE VITROBOT April 9, Using the Vitrobot

Table of contents. Table of contents 1. Installation 2. Introduction ( Display, monitor ) 3. Introduction ( Operating elements ) 4

200L BELT DRIVEN AIR COMPRESSOR

Oerlikon Sputtering Evaporator SOP

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

SETUP GBW/GBWE SETUP. upper or lower position. Placing the lock in the upper position will provide 1 foot (305 mm) more height.

Standard Operating Manual

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Instruction Manual Otoflash G171 UV-flash-device for Light Curing

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

Nordiko Metal Sputtering System Standard Operating Procedure

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13

LAMINAR FLAME SPEED ANALYSIS USING PIV (PARTICLE IMAGE VELOCIMETRY)

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

University of MN, Minnesota Nano Center Standard Operating Procedure

Underwater Housing for Canon G7X Mark II

G7S Hand Pump Owner s Manual

24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

50L BELT DRIVEN AIR COMPRESSOR

SEM LEO 1550 MANUAL. I. Introduction. Generality

Mobile High Pressure Compressor Unit for Compressing Air and Breathing Air. PE 250-TE with compressor control (optional equipment)

MODEL 269DF-LCF-LCF/WS FILLING SYSTEM

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Approved by Principal Investigator Date: Approved by Super User: Date:

MMRC PVD Evaporator Instructions

SCS Field Revision No. 1.0 June 2001 STANDARD OPERATING PROCEDURE (SOP) FOR FIELD OPERATION OF THE SEQUENTIAL CYCLONE SAMPLER (SCS)

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

B. Assemble the Method-25 sample train with a heated probe/filter.

OPERATING MANUAL LO-CROSS-FLOW CF M

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

UNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No

Usage Policies Notebook for Trion RIE / ICP Dry Etch

STANDARD OPERATING PROCEDURE VORTEX-GENIE 2

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

PDY TON HYDRAULIC CRIMPING TOOL WARNING

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

BASIC RULES OF USAGE, OLIS DSM-20 CD Management The instrument is currently managed by the Analytical Biophysics Core Facility.

NCTU NFC Lithography-Related Tools Operation. Manual

1)! DO NOT PROCEED BEYOND THIS MARK

Autotrol Performa ProSoft

ShellPa. Standard Mechanical Cell Stretch System Model No: NNMS Serial #: User Manual

COMPRESSOR REPLACEMENT PROCEDURE (R-134A)

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

Basic ICP Operating Procedures

Underwater Housing for Panasonic Lumix DMC-ZS60, DMC-TZ80

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

24L AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: RANGER 7/240 PART NO: LS0913

Inside Front cover This page will remain blank.

1/4 SHEET PALM SANDER

Transcription:

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Start-up Procedure: 1. Start the grey pump in the service corridor behind the aligner 2. Open the valve for the Nitrogen gas 3. Plug in the power for the air conditioning and HEPA filter unit 4. In the Turn Table group put the switches in single and cylinder work positions. 5. in the shutter release control group put the switches in non release and with alignment positions 6. In the exposure group put thee switch in manual off position. 7. Plug in the power cord for the PPC and power supplies 8. Start the PPC by pushing the power button 9. Turn on the power supply for the mercury lamp 10. Press the start button on the mercury lamp power supply and hold it for 1-3 seconds (you should hear a sharp sound from the power supply) 11. Wait for 30-40 minutes for the mercury lamp to warm up.

Exposing: 1. Turn on the power for the yellow alignment light. 2. Place your mask in the mask holder and turn on mask vacuum 3. Put your sample in the centre of the sample holder 4. Press the start button to rotate the stage, the expose indicator should now be on. Remove any previously exposed samples. 5. Move the wafer focus to zero position 6. Focus on your sample 7. Focus the mask on your sample using the mask focus 8. Make a rough alignment of your pattern 9. Move wafer focus to the 100 μm position 10. Align your sample 11. Set the exposure time on the timer using the outer scale 12. Flip the switch for release to automatic release position. 13. Move the microscope to the left edge to start exposure. 14. If you have more samples to expose, return to point 3 15. Flip the Turn table switch to Cylinder no work position 16. Flip to release switch to non release position 17. Press the Start button to rotate the stage. 18. Remove your sample 19. Put the microscope back in a centred position 20. Turn of the alignment lamp. 21. When exposing next sample start from 1

Press here to open Mask holder Open for mask loading Alignment light Mask/Dust cover Mask loaded Focus

Mask Focus Wafer Focus

System shut down: 1. Turn of the vacuum for the mask and remove the mask 2. Move the microscope to a centred position if this is not already done. Not doing this can damage the automatic release mechanism that starts the exposure. 3. Place the silicon wafer on the mask holder to prevent dust from falling into the optical column. 4. Turn of the yellow alignment light if this is not already done 5. Turn of the mercury lamp power supply 6. Turn of the power for the PPC 7. Unplug the power cords for the power supplies and the PPC. 8. Wait for 30 min and then turn of the nitrogen flow and the grey pump in the service corridor. 9. Unplug the air conditioning/hepa unit. General information and operating tips If you can t focus on your sample (i.e. it is too low or too high) you can slowly turn the Pressure valve on the right side of the PPC to raise and lower the stage. If the stage doesn t move, reduce the pressure to minimum and slowly start to increase it. At some point the sample holder will start to move up and down in a pulsing movement, this indicates that the pressure is just below the lower threshold. By slowly increasing the pressure from this point you should be able to move the stage. Please do note that the pressure and position of the stage also has an upper limit. Pressure Control

In cylinder no work mode the stage can be rotate time and again but it will not raise the stage and it will not be possible to start an exposure, this mode is good for unloading the last sample in a series. In Cylinder work mode the stage can only be rotated once and then the PPC will go into exposure mode. If you don t want to expose for some reason but want to rotate the stage to remove your sample press the Stop button once to enter emergency stop mode and end the expose mode, then press it again to end the emergency stop mode. In non release mode the system will not start an exposure if the microscope is moved to the left edge; this can be good while aligning so you don t accidentally start an exposure while looking into the microscope. In automatic release mode the system will start the exposure when the microscope reaches the left edge; this is how you normally start the exposure. When you move the wafer focus you will also lose some fine alignment, therefore always do your fine alignment after using the wafer focus. Not using the wafer focus will result in poor resolution. This has no effect if your smallest structure is 20 or so microns, but it is necessary to achieve higher resolution. The mercury lamp of the aligner only generates a uniform dosage in the central part of the sample holder make sure this is where your pattern is. If your mask is much bigger than the holder and you worry about scratching patterns on the edge there is a piece of clean room paper cut so that it will cover the edge of the holder to prevent scratching. Don t forget to cover the column when turning of system or the optical part will get dirty. Don t forget to unplug the cables or some of the transformers and power stabilizers will overheat. Please turn of the yellow light while not aligning since this generates quite a lot of heat and don t need to heat up before use. Let the mercury lamp cool down before turning of the pump, this will greatly increase its life time. If you accidentally turn of the mercury lamp let if cool down for at least 40 minutes before turning it back on, otherwise it could be damaged or explode and the resulting mercury fumes are very toxic. Check if someone else wants to use the system after you before turning of the mercury lamp.

If leaving for lunch leave the mercury lamp on, it takes more damage from being turned on then from running for 1 hour. If leaving for a longer period of time, turn it of, it has a limited life time (1000 hours or so). If you notice that the exposure times start to increase a lot (3-5 times longer than usual), call someone who is responsible for the system, the mercury lamp might be getting old and has to be changed or it might just be out of alignment. If the lamp is getting old the risk of explosion increases.