Endura AZ40 Transmitter Main board spares kit

Similar documents
Endura AZ40 Sensor assembly Smart electronics PCB

Endura AZ40 Sensor assembly High temperature filter assembly kit

Aztec ADS430 EZCLEAN Compressed air supply unit

ATS430 turbidity sensor Retractable insertion assembly

C1960. Multi-recipe profile recorder/controller. Measurement made easy

9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs

ABB MEASUREMENT & ANALYTICS DATA SHEET series Dissolved oxygen systems

The benefits of the extended diagnostics feature. Compact, well-proven, and flexible

ABB MEASUREMENT & ANALYTICS DATA SHEET. Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer

9408 series Dissolved oxygen systems

Navigator 600 Silica analyzers

Model NDA Differential pressure transmitter. Deltapi N Series A complete range of pneumatic pressure transmitters

ABB MEASUREMENT & ANALYTICS DATA SHEET. AP100 ph/redox (ORP) cartridge sensors with water-wash option

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter

PositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible

MANUAL DIRECT PURGE OPTION. UNION Instruments GmbH CWD2005 PLUS. General information, safety standards and regulations for direct purge option

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

AP100 series ph/redox (ORP) cartridge sensors with water-wash option

ABB measurement & analytics data sheet. JAA, JAF and JAG Pneumatic indicating controllers

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

Endura AZ25 series probe Combustion oxygen monitor

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

AUTOMATIC GAS MANIFOLDS

INSTALLATION, OPERATING & MAINTENANCE MANUAL WATER DRAIN VALVE

Pressure Dump Valve Service Kit for Series 2300 Units

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

TFS230 FLOOR STRIPPER OPERATION & MAINTENANCE

Dissolved Oxygen Cartridge Electrode Systems

Service ISO17025 explained

FPD212 Dual-chamber orifice fitting

Supplementary Operator s Manual 42/24-14 EN Rev. 3

FPD175 Compact orifice carrier assembly

RPS900W Redundant Power Supply. Installation Guide.

SPECIFICATION // DATA SHEET WATER DRAIN VALVE AM SENSORS LTD AMS-S-084/DS/ (ISSUE 1)

AK100 Series ATEX Compliant Gas Analyzer System for Hydrogen-cooled Alternators

Instruction Manual BA90 Precision Barometer

GAS FUEL VALVE FORM AGV5 OM 8-03

ABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas

Pressure Dump Valve Service Kit for Series 3000 Units

2600T Series Pressure Transmitters Calibration Frequency Determination

CoriolisMaster mass flowmeter Diagnostics, verification and proof test

ABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

ABB MEASUREMENT & ANALYTICS DATA SHEET. AK100 Series ATEX compliant gas analyzer system for hydrogen-cooled alternators

Rotary Abreast Automatic Teat Sprayer User Manual. Technical Manual Revision 0.01

Topline Peristaltic Dosing Pump for PAC Installation, Commissioning and O&M Manual

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

AREA VALVE SERVICE UNITS

Navigator 500 Dissolved oxygen analyzer. Measurement made easy. Accurate and reliable measurement of dissolved oxygen in high purity water

Service Calibration are you doing it properly?

Measurement & Analytics Wet gas monitoring

S265 Remote seals with capillary tube. 2600T Series Pressure Transmitters. Engineered solutions for all applications

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

EL227CL/NC Polyurethane resin/hardener composition

Models NAE Pneumatic differential pressure transmitters with remote diaphragm seals

OPERATOR S MANUAL SPREADER. CSS, VNQ [For Combine Harvester AW82V] Original instructions

GENUINE CARGO NET INSTALLATION AND USER S INSTRUCTIONS

Autocalibration Systems for In Situ Oxygen Analyzers

Digital Melting Point Apparatus

KMW3000C. User Manual. Water Chiller for Xcalibur and Gemini systems. Version 1.6, September 2010

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

ABB MEASUREMENT & ANALYTICS DATA SHEET. Endura AZ20 oxygen monitor Combustion gas analysis

2600T Series Pressure Transmitters Calibration Frequency Determination. Pressure Measurement Engineered solutions for all applications

PURGEX X Purge Controller D Version

2600T Series Electronic Transmitters Models 264_268HS/NS 264GS/AS, 264HF/NF

UsER manual for Watersens ph -REDOX

HANDBOOK. Squeeze Off Unit SOU 250. Please refer any queries to:

Instruction Manual Contact Pressure Vacuum Gauge

Instructions for Use

SF 6 Product Guide. Gas Analysis Instruments and Accessories

MoveRoll Conveyor Operating and Maintenance Manual

Service Bulletin WRSPUDF Ride-On Roller Screeds LPG Hose Replacement

/2004 US/CA

DK ton, Double-acting, Die-type Crimping Tool

combustion efficiency monitoring General Purpose Applications Flue gas oxygen analyzers Combustion & Environmental Monitoring An Company

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

Risk Analysis and Hazard Reduction Report V1.0

BGA158 Series CE Approved Class A Shutoff Gas Valve

HANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to:

Half Facepiece Respirator Assembly 5000 Series, Dual Cartridge, Organic Vapor/P95, Disposable

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

ACV-10 Automatic Control Valve

Budget Range Operators Handbook

222 Schwinn Recumbent Exercise Bike Parts List Full Size Hardware Chart Product Illustration Assembly Instructions

Installation and Service Manual

2-1- Shipping Receiving Lifting Location on Site Installation Stages 6

MODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS

: Fireproofing Blanket

Stand-Alone Bubble Detection System

Operating Instructions Part No

BGA158 Series CE Approved Class B Shutoff Gas Valve

AutoChanger Installation & User Guide Issue 2

PLAQUE MFM0680 UNDERWATER LIGHT. Operating Instructions

abc OPERATING, INSTALLATION & MAINTENANCE MANUAL FOR G6 GAS SAMPLER

Dissolved Oxygen Systems

Transcription:

Instruction replacement procedure INS/ANAINST/018 EN Rev. A Endura AZ40 Transmitter Main board spares kit Kit reference: AZ400 767 Measurement made easy 1 Introduction This publication details replacement procedures for the AZ40 transmitter main board (AZ400 767) fitted to Endura AZ40 transmitters. Before carrying out any procedures, read Section 3, page 2. These procedures must be carried out by a suitably-trained technician. Kit contains: Transmitter main board assembly This publication Endura AZ40 Sensor assembly Main board spares kit (part no. INS/ANAINST/018 EN) 2 Tools / Items required Transmitter door key (supplied with transmitter) Operating instruction OI/AZ40-EN* Anti-static strap T8 x 40 Torx screwdriver Small flat-bladed screwdriver Small crosshead screwdriver Medium crosshead screwdriver *Operating instruction OI/AZ40-EN contains mandatory safety information and can be downloaded via the link (above) or by scanning this code:

3 Health & Safety 3.1 Safety precautions Be sure to read, understand and follow the instructions contained within this document before and during use of the equipment. Failure to do so could result in bodily harm or damage to the equipment. Installation, operation, maintenance and servicing must be performed: by suitably trained personnel only in accordance with the information provided in this document in accordance with relevant local regulations 3.2 Potential safety hazards 3.2.1 Process conditions and requirements Environmental conditions High air / equipment / structure temperatures, poor air quality and adverse environmental conditions may be present when the process is running. It is recommended that the process is shut down before performing these procedures. The process must be cool enough to enable shutdown, disconnection and removal of the sensor in a safe manner and in accordance with relevant local regulations. Appropriate PPE, including mask and goggles must be worn when preparing the process for these procedures. 3.2.2 Endura AZ40 sensor fibrous material in probe assembly WARNING Serious damage to health Fibrous material The sensor and probe assemblies (standard and high temperature versions) contain fibrous material that can be a health hazard if airborne. The material, predominantly aluminosilicate refractory fibres, AS 142844-00-6. Refractory ceramic fibres (RF) are classified as: ategory 1B carcinogen under regulation (E) No 1272/2008 the classification, labelling and packaging regulations. ategory 2B carcinogen by inhalation by The International Agency for Research on ancer (IAR). When removing the sensor cover and subsequent maintenance activities, exposure to the airborne fibres could occur. ABB have conducted air sampling assessments within the breathing zone of the operator and have identified that an exposure limit of 1 fibre / cubic centimetre is unlikely to occur. Exposure to any carcinogen must be kept as low as reasonably practicable. Appropriate PPE defined below, must be worn when working with probe assemblies (all installation, replacement, maintenance procedures): A face fit tested, half mask conforming to EN140 (or equivalent) with a level 3 particulate filter conforming to EN 143 (or equivalent). Disposable protective coveralls in accordance with Type 5 ISO 13982-1:2004 (or equivalent). Goggles and gloves. 3.2.3 Endura AZ40 sensor / probe installation to pressurized process DANGER Serious damage to health / risk to life Pressurized equipment do not install / remove / the sensor / probe if the process is at positive pressure Installation, operation, maintenance and servicing of pressurized equipment must be performed: by suitably trained personnel only in accordance with the information provided in this document in accordance with relevant local regulations when process conditions are suitable to allow enough to enable installation / maintenance 2 INS/ANAINST/018 EN Rev. A Endura AZ40 Transmitter Main board spares kit

3.2.4 Endura AZ40 sensor high operational temperature on exposed parts High temperature on exposed surfaces During operation, exposed sensor surfaces can reach 200 (392 F). Ensure suitable PPE is available and is worn before handling the sensor. Do not touch exposed surfaces until the sensor / probe is cool enough to handle with PPE. 3.2.6 Endura AZ40 analyzer electrical To ensure safe use when operating this equipment, the following points must be observed: up to 240 V A may be present. Ensure the supply is isolated before removing the terminal cover normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and / or temperature Safety advice concerning the use of the equipment described in this document or any relevant Material Safety Data Sheets (where applicable) can be obtained from the ompany, together with servicing and spares information. Endura AZ40 transmitter weight The transmitter weighs 7.6 kg (17 lb) and must be mounted in accordance with the information supplied in Operating instruction OI/AZ40-EN. Suitable lifting equipment must be available when installing / removing the transmitter from the mounting. Fig. 3.1 High temperature points on exposed sensor surfaces during operation 3.2.5 Endura AZ40 sensor weight The sensor weighs 9.0 kg (20 lb). When fitted with a probe / filter assembly, the combined sensor / probe weight is dependent on probe length / type plus filter option refer to Operating instruction OI/AZ40-EN for weight details. The sensor / probe assembly must be mounted in accordance with the information supplied in Operating instruction OI/AZ40-EN. Suitable lifting equipment must be available when installing / removing the sensor / probe from the process. Endura AZ40 Transmitter Main board spares kit INS/ANAINST/018 EN Rev. A 3

4 Isolating the transmitter 4.1 Isolating the transmitter Referring to Fig. 4.1. 1. Isolate transmitter A from incoming mains powers supplies B. DANGER Serious damage to health / risk to life The transmitter must be isolated from mains powers supplies before carrying out this procedure. 5 Disconnecting transmitter cables Referring to Fig. 5.1: 1. Unlock transmitter door A using the transmitter door key and open the transmitter door. 2. Use a small crosshead screwdriver to unscrew 4 (captive) terminal cover screws B and remove transmitter terminal cover. ustomer-supplied mains isolator Fig. 4.1 Isolating the transmitter from incoming mains power supplies Fig. 5.1 Accessing the main board 4 INS/ANAINST/018 EN Rev. A Endura AZ40 Transmitter Main board spares kit

Referring to Fig. 5.2: 3. Use a small flat-bladed screwdriver to disconnect wires at terminal blocks A and B and any connections made at terminal blocks. Line in Neutral in Earth Line out Neutral out Earth Digital inputs Relays Analog outputs 1 2 3 4 5 6 1 2 3 4 omms (Modbus connections) D+ (white) D (green) SN (screen) DI1 DI2 DI3 DI4 A OM D OM V+ (red) O M (black ) ck) 1 OM 2 TX/RX 3 TX/RX+ 4 TX 5 TX+ + + + + Fig. 5.2 Disconnecting transmitter mains, signals, relays and digital inputs Endura AZ40 Transmitter Main board spares kit INS/ANAINST/018 EN Rev. A 5

6 Replacement procedure Referring to Fig. 6.1: 1. Use a T8 x 40 Torx screwdriver to unscrew 6 Torx screws A and remove processor / display board cover B. 2. Disconnect the flexi-cable from the processor display board D. Referring to Fig. 6.2: 3. Unplug 2 solenoid valve connectors A and use a medium crosshead screwdriver to unscrew and remove earth stud connector screw and washer B. Disconnect the earth wire and retain washer and screw for re-use. SD Industrial ard Fig. 6.1 Disconnecting the flexi-cable from the transmitter display board Fig. 6.2 Unplugging solenoid valve connectors and disconnecting earth wire at earth stud 6 INS/ANAINST/018 EN Rev. A Endura AZ40 Transmitter Main board spares kit

Referring to Fig. 6.3: 4. Use a small crosshead screwdriver to unscrew and remove 6 main board retaining screws A. Retain for re-use. Remove the main board from the transmitter enclosure. 5. Disconnect the flexi-cable B from back of the main board. Retain the cable for re-use. Fig. 6.3 Removing the main board from the transmitter cubicle 6. onnect the existing flexi-cable to the back of the new main board and fit the new main board in the reverse order of steps 1 to 5. 7. Re-make cable connections refer to Section 5, page 4. 8. Prepare the transmitter for operation by reversing the isolation procedure refer to Section 4, page 4. 9. Refer to OI/AZ40-EN to restore the analyzer to full operational condition. Endura AZ40 Transmitter Main board spares kit INS/ANAINST/018 EN Rev. A 7

ontact us ABB Limited Process Automation Howard Road, St. Neots ambridgeshire, PE19 8EU UK Tel: +44 (0)870 600 6122 Fax: +44 (0)1480 213 339 Mail: enquiries.mp.uk@gb.abb.com ABB Inc. Process Automation 125 E. ounty Line Road Warminster, PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 Note We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. opyright 2016 ABB All rights reserved 3KXA722444R2001 INS/ANAINST/018 EN Rev. A 02.2016 www.abb.com/measurement Sales Service