Radnoti Liver Perfusion System

Similar documents
ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

Aerobic Respiration. Evaluation copy

INSTRUCTION MANUAL. January 23, 2003, Revision 0

PRODUCT SHEET. Order probe only as RXPROBE02

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

User s Guide. CO2 Controller for miniature incubators

User s Guide. Pressure Controller for liquid delivery systems

User s Guide. Vacuum Controller for liquid delivery systems

UsER manual for Watersens ph -REDOX

LogIT Dissolved Oxygen Probe

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

PH01 Perfusion Cannula Manual

Instruction Manual Updated 7/26/2011 Ver. 2.2

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

User s Guide. CO2 & O2 Controller for miniature incubators

STERILIZABLE DISSOLVED OXYGEN SENSORS

12mm Dissolved Oxygen Sensor Care and Use Instructions

MEMBRANE CHAMBER MC. Innovative Engineering for Science

AP100 series ph/redox (ORP) cartridge sensors with water-wash option

B. Assemble the Method-25 sample train with a heated probe/filter.

1020 Industrial Drive, Orlinda, TN fax

256 Pneumatic Pressure Indicator

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Unisense In Situ UniAmp Connector System

HI 2314 HI 2315 HI 23151

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Non-Membrane Disinfection Sensors Instruction Manual

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

EC214 EC215 - EC215R Bench Conductivity Meters

SS32L Dissolved O 2 Probe

Installation Procedure for your Disco Double Kit, by Airgun Lab, LLC.

The Ideal Gas Constant

Dissolved Oxygen in Water. Evaluation copy. Table 1. Temperature Range ( C) Trout Smallmouth bass Caddisfly larvae

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Cell Disruption System

QUIK-jet Micro-Injection System

OxyProbe II Instruction Manual

Techcon Systems TS1258 Pressure Pot

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

QUICK START GUIDE TO DRINKING WATER MONITORING

Vinmetrica Dissolved Oxygen System User Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

Free Chlorine Sensor For additional information, please visit our website at

BPS-2, Rev TABLE OF CONTENTS

Dissolved Oxygen Sensor

Bulletin TCR-104 & 109 Filling and adding to the Glycol pressure system

DISSOLVED OXYGEN SENSOR BT34i

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

PO Box 370 Phone: Swansea NSW 2281 Fax: AUSTRALIA Mobile: Web:

Met One E-BAM Particulate Monitor

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Columbus Instruments

Experiment C-6 Gas Solubility

UltRo Dual Flow Reverse Osmosis Water System

Manual Solvent Purification System. Innovative Technology, Inc.

Constant Pressure Crude Oil Container Model CPCCP

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

EcoCell Vertical Electrophoresis Unit

Heavy Duty Dissolved Oxygen Meter

Gas Laws. Figure 1: Experimental Set-up with Leveling Bulb. GCC CHM 151LL: Gas Laws GCC, 2019 page 1 of 8

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Calcimeter Instruction Manual

Heat Engine. Reading: Appropriate sections for first, second law of thermodynamics, and PV diagrams.

Installation, Operation, and Maintenance Manual

Multi-Probe Sampling System User Guide & Manual

Operating Instructions

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

Gas Pressure Sensor (Order Code GPS-BTA)

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

Durham Geo-Slope Indicator S-510 Triaxial Cell

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

Vapor Pressure of Liquids

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

OPERATING INSTRUCTIONS

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

Objective To identify a pure liquid substance using the physical properties of solubility, density, and boiling point.

GAS CONVERSION KIT Installation Instructions Range/Rangetop

Digital Vacuum Regulator

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

RD(H)20/25 Pressure-Reducing Regulator User Manual

Vapor Pressure of Liquids

USER GUIDE, VOL.2.0 MYOGRAPH SYSTEM - 112PP/114P

The Decomposition of Hydrogen Peroxide

Vacuum grease can be applied to Warner chambers by use of a small artist s dotting brush or a syringe. Both approaches are described below.

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

EXPERIMENT 8 Ideal Gas Law: Molecular Weight of a Vapor

Glove Box Installation Manual

Paddle Bar Replacement

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

MARTINI. instruments. THANK YOU FOR CHOOSING. Sales and Technical Service Contacts:

Quick Start Guide LIQ-QSG-499AOZ, Rev M July Rosemount 499AOZ. Dissolved Ozone Sensor

Cover Page for Lab Report Group Portion. Head Losses in Pipes

CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

1. Introduction Why use the Flow-Through Cell? Component Identification... 1

YSI Model 63. Handheld ph, Conductivity, Salinity and Temperature System. Operations Manual

Transcription:

Radnoti Liver Perfusion System 130003 Offset Offset Gain Gain Radnoti 2006

Description Qty Part # A Base only, for 4-bar stand 1 159950-B4 B Stabilizer Bar only, for 4-Bar stand 2 159950-C4 C Rod 24 Long Stainless Steel 5 159950-24 D Single Ring Clamp (Membrane oxygenator) 2 130144-RC E Single Ring Clamp (in/out flow manifold ele+sep) 2 159953-2 F Single Ring Clamp (Bubble trap) 2 120149-RC G Single Ring Clamp (Liver perfusion reservoir w/j 100ml) 1 159953-100 H Single Ring Clamp (500ml reservoir) 1 120141-0 I Stand Clamp Medium (Red) 7 159952-11 J Water-Jacketed Reservoir 500ml 1 120142-0 K Oxygenating Bubbler with 2 inlets 1 140144-0 L Liver Perfusion Chamber w/opt port and lid 1 670101 M Liver Perfusion Reservoir w/j 100ml 1 1581-100 N Bubble Trap Compliance Q.D. 2 130149 O Inflow Manifold Triple Elect+Septa 1 120170-3 P Outflow Manifold Triple Elect+Septa 1 120171-3 Q Membrane Oxygenator (Small) 1 130144-MH R Luer Manifold 4-Port 2 120165-4 S Heater Bath Circulator Pump 1 170051-A T Peristaltic Pump 1 170100-A U Easy Load Pump Head 2 170110 V Pressure Transducer 2 159905 W Water-Jacketed Flex Tube Assembly 36 2 130155-36 48 1 130155-48 60 1 130155-60 X Q.D. Tygon Tubing (Water-Jacketed) 10 3 120159-10 15 5 120159-15 7 120159-20 30 1 120159-30 48 1 120159-48 Y Inner Tygon Tubing (Perfusate, gas) 10 3 120157-10 4 120157-20 24 1 120157-24 40 2 120157-40 48 2 120157-48 AA Manifold 5 Port Q.D. 2 120161 BB Silicone Tubing 2 130144-188-20 24 1 130144-188-24 40 1 130144-188-40

Offset Gain (A) 4-Bar Stand (B) Stabilizer (C) S.S. Support Rod (D) Single Ring Clamp (E) Single Ring Clamp (F) Single Ring Clamp (G) Single Ring Clamp (H) Single Ring Clamp (I) Stand Clamp (J) Water-Jacketed Reservoir (K) Oxygenating Bubbler (L) Liver Perf Chamber (M) Liver Perf Reservoir (N) Bubble Trap (comp) Q.D (O,P) Inflow and Inflow Manifold (S) Heater Bath Circulator Pump (Q) Membrane Oxygenating (R) 4-Port Luer Manifold (T) Peristaltic Pump (U) Easy-Load Pump Head (AA) Manifold 5 port QD (W) Flex Tube Assembly (X) Q.D. Tygon Tubing (V) Pressure Transducer (Y) Inner Tygon Tubing

Fig 1 Fig 2 1 2 3 4 Fig 3 Place base on lab bench where it will be used. Screw in 24 SS rods into base. Slide a Stabilizer bar onto the rods about 4 from the base. Slide 2 stand clamps onto rods 1 and 2 onto rod 2. Fig 3 Fig 4 Slide the other Stabilizer bar onto the rods about 2 from the top. Srew in a 24 SS rod into rod 3. Slide a Stand Clamp onto rod 3 and 4 just above the stabilizer bars. Slide a Stand Clamp on rod 3 36 from the base. 1 Place the ring clamps for the bubble traps(f) in the stand clamp on top of rod 3 and the hole all the way to the right on the lower stabilizer. Place the ring clamps for the oxygenator(d) in the lower stand clamps on rods 1& 2. place the ring clamps for the resevoirs (G & H) in the upper stabilizer.place the ring clamps for the inflow and outflow manifolds (E) in the lower stabilizer.

Offset Gain Offset Gain Fig 5 Place transducers into the Stand clamps on top of rods 1 & 2. Place Bubble Traps into the ring clamp on top of rod 3 and ring clamp all the way to the right on the lower stabilizer. Put the membrane oxygenator in the ring clamps on rods 1 & 2. place inflow and outflow manifolds in the ring clamps in the lower stabilizer. Place the.5 liter resevoir and 100ml resevoir in the ring clamps on the upper resevoir WARNING: Mounting the glass components on ring clamps (careful not to over tighten)

Flex Tube Assembly Chart 60 36 36 48 Optional Flow Meter

Perfusate Connection Chart *Silicone tubing: Slip tubing over female end of 1 way stopcock *24 15 * *40 10 * waste 40 waste Optional Flow meter 000101 10 48 40

Gas Connection 24 48

Water Jacketed Connection Chart 10 60 Water return to pump 30 15 15 15 10 10 15 15 48 Water from pump

RADNOTI QUICK DISCONNECT INSTRUCTIONS For Reference only A Slide Cap-With hole onto the water line. (Tygon tubing #120159) B Push the plastic sleeve insert into the end of the Tygon tubing with the flared end first. (Catalog #120160) C Be sure to use 5/16 x 3/16 inch Tygon tubing. (Catalog #120159) D MOISTEN TIP OF TYGON TUBING before insertion into the threaded glass. CAUTION: To avoid breakage, do not over tighten the screw cap. (C) (A) (D) (B) Plastic Insert Replacement Part List: Catalog No. Description 120160 Plastic sleeve insert, 25/Pkg 120159 Tygon tubing 5/16 x 3/16 ID, 50 ft/min 160196 Drilled cap for W/J.Q.D.

Instruction Manual Radnoti 159905 Micro Pressure Transducer With Amplifier Setup: Connect 8 pin power supply connection to amplifier. Connect output BNC connector to recording device. Plug sensor connector into amplifier. Calibration: Make sure to let the amplifier warm up for about 10 minutes before calibration. The output of the amplifier is 0 to about 3.2 volts full scale. This allows for precise calibration of 10mv/ mmhg. 1. Connect a pressure calibration bulb and gauge to the pressure transducer. 2. Turn the Gain control fully clockwise. 3. Zero the amplifier to about 000 on the readout using the Offset control. 4. Inflate the system to 150mmHG and hold. 5. Turn the gain control until the display shows 150. 6. Deflate to zero and readjust to 000 7. Inflate again to 150mmHg and readjust Gain. 8. When pressure is at zero the readout should show 000 and when at 150 the readout should show 150. Handling: The tip of the fitting is filled with a pressure transferring gel. It is important not to disturb this gel as it may degrade the performance of the transducer. It is very important not to drop the fitting on its open end. At the tip of the fitting there is a small glass housing that may be damaged.

MI-710 Micro-combination ph Microelectrodes Operating Instructions The microelectrode is ready to use. Carefully unwind the tape and remove the probe from the protective glass tube. The white translucent sleeve, which covers the hole, should be moved down the glass barrel slightly to ventilate the reference chamber and replaced after each use. The use of a separate Reference Electrode is not required. Calibration The probe can be held by hand when measuring ph. It has been standardized in two (2) ph buffers. Follow the procedure recommended by the manufacturer of your ph meter for calibrating your ph meter with our microelectrode. Optimum Response Time: Optimum response time will be obtained after the probe has been exercised in two (2) buffer solutions. Fill one beaker with a ph 4 buffer or equivalent and a second beaker with a ph 7 buffer or equivalent. Hold the microelectrode and touch the ph 4 buffer surface allowing 5 seconds for equilibration, and then touch the ph 7 buffer surface in the same manner. Do this several times. It is not necessary to rinse the probe because the carry-over is negligible. Signal Generation ph Electrodes Ideally a ph electrode will output a voltage of 0mV in a ph 7 buffer. This can very by +/- 50mV based on the individual ph electrode. The Nernst equation tells us that a ph buffer 4 should be 160mV greater (more positive) than the reading in a ph 7. The reading in a ph10 should be -160mV less (more negative than that of a reading in a ph7 buffer. The 160 comes from being 3(7-4)*59(Nernst value at 20 degrees)=168mv for 100% slope and 160 is slightly less than 100% theoretical. Handling The outer reference chamber of the microelectrode can be refilled when necessary with the 3 M KCl solution, using the bulb pipette. The reference chamber should be filled slowly. Be careful not to apply pressure against the inner glass capillary tube. Cleaning When using the microelectrode in solutions containing protein, it should be soaked in an enzyme cleaning solution such as Terg-a-zyme (Alconox, Inc.) or a chromic/sulfuric acid glass cleaning solution for a couple of minutes after each use to remove the protein from the glass and the reference junction. This will prolong the useful life of the microelectrode. Storage Always clean the microelectrode before storing: Long-term (over 2 weeks): Return the probe to its original container and prepare it in the same condition in which you received it. Usually this means simply moistening the sponge located in the bottom of the protective glass tube with ph 4 buffer. Short-term: The probe can be left in an acid ph buffer solution (ph 4).

MI-730 O2 Microelectrodes Operating Instructions The MI-730 Electrode Kit contains the following: 1 Oxygen Electrode with membrane housing 6 Replaceable membrane housings 1 Bottle of electrolyte 1 Filling tip 1 Set of instructions 5 Polishing pads Electrode Assembly and Preparation 1. Remove the acrylic housing from the Oxygen electrode by unscrewing it from the electrode body. This housing, with affixed Teflon membrane, does not contain the necessary oxygen electrolyte solution. 2. Add oxygen electrolyte solution to the acrylic housing. This is accomplished by attaching a filling tip to the bottle of electrolyte and adding the electrolyte to the housing to a minimum height of 6 mm. Gently place the bubble-free end of the filling tip against the Teflon membrane and release the electrolyte to the minimum height. 3. Insert the housing on to the Oxygen electrode, being careful not to trap any air bubbles near the electrode tip. The housing is screwed clockwise into the body of the electrode until it stops. Check the tip of the electrode for proper seating of the housing against the electrode. The electrode is properly seated if the electrode protrudes slightly beyond the end of the housing. Calibration Calibration of the electrode requires the use of two standard gases with percent values that are close to the percent values of Oxygen to be measured. Common values used are 0% Oxygen for zeroing and 21% Oxygen (Ambient Air) for sloping or gain. A. Calibration for gas samples: When samples to be measured are gaseous, then calibration should be performed with humidified gases. Two possible setups for calibrating the electrode are shown below. Keep the tip of the electrode as far as possible from the surface of the water. The bubbling rate of the gas should be slow (3-6 bubbles per second). Although bubbling at a faster rate will flush the chamber more quickly, it will also cause a cooling effect on the electrode. 1. Bubble the 0% gas through the chamber and adjust the zero of the meter after a stable reading is obtained. It may take up to 15 minutes to de-gas the chamber of contaminants, however bubbling the gas vigorously will cause droplets to collect on the tip of the electrode. This will make the electrode

2. Bubble the 21% gas (or any other percent value you decided to use for your application) through the chamber until a stable reading is obtained. Adjust the calibration control to 21%. This procedure of alternating between the two gases should be continued until you become confident of stability and reproducibility. The electrode is now ready to use. B. Calibration for liquid samples To decrease calibration time, two separate calibration chambers should be used. One for the 0% gas and another for the sloping gas such as 21%. When setting up the calibration chambers initially, it will take up to 30 minutes to flush each chamber to obtain a steady state oxygen level and a constant temperature. Again the bubbling rate should be carefully regulated (3-6 bubbles per second) so that both calibrating liquids are at the same temperature. Calibrating standards and samples must be at the same temperature for accurate Oxygen measurements. 1. Immerse the tip of the electrode into the 0% standard and adjust the zero of the meter after a stable reading is obtained. 2. Remove the electrode from the first standard and place it into the second standard. Adjust the calibration control to the value of the second standard (ex: 21%) Alternate between the two standards until you become confident of stability and reproducibility. The electrode is now ready to use. Handling When necessary, the membrane of the electrode can be replaced by following the assembly procedure above. When removing and replacing a membrane as well as when calibrating or making measurements, be careful not to apply pressure against the internal electrode. Any excessive pressure against the internal electrode can cause the electrode to crack rendering it useless and unrepairable. Cleaning When using the electrode in solutions containing protein, the electrode should be soaked in an enzyme cleaning solution such as Terg-a-zyme (Alconox, Inc.) after each use for a couple of minutes to remove the protein from the membrane surface. This will prolong the useful life of the membrane. Storing Always clean and rinse the electrode before storing: Long-term (over 1 month): Remove the membrane housing from the electrode. Rinse the internal electrode with distilled water and pat dry. Place a new, unfilled membrane housing over the internal electrode and attach loosely (Do not seat completely). This membrane will serve to keep the dust off of the electrode tip. Short-term: The electrode can be left in room air with membrane housing still attached.

Output Conversion Formula for conversion of percent oxygen to solubility in moles/liter: S= (a/22.414) x (760-p)/760) x (r%/100) S = solubility of gas in moles per liter a = absorption coefficient of gas at temperature P = vapor pressure of water at temperature r% = actual reading in percent Oxygen Temperature (T) vs. absorption coefficient (a) for oxygen in water: T a T a 5.04287 19.03161 6.04180 20.03102 7.04080 21.03044 8.03983 22.02988 9.03891 23.02934 10.03802 24.02881 11.03718 25.02831 12.03637 26.02783 13.03559 27.02736 14.03486 28.02691 15.03415 29.02649 16.03348 30.02608 17.03283 35.02440 18.03220 40.02306 T = Degrees Celsius a = Absorption Coefficient

Radnoti 2006 Radnoti Pressure Sensor with Amplifier 159905

159905 Setup: Connect 8 pin power supply connection to amplifier. Connect output BNC connector to recording device. Remove protective cover from sensor. Plug sensor connector into amplifier. Calibration: Make sure to let the amplifier warm up for about 10 minutes before calibration. The output of the amplifier is 0 to about 3.2 volts full scale. This allows for precise calibration of 10mv/mmHG. 1. Connect a pressure calibration bulb and gauge to the pressure transducer. 2. Turn the Gain control fully clockwise. 3. Zero the amplifier to about 000 on the readout using the Offset control. 4. Inflate the system to 150mmHG and hold. 5. Turn the gain control until the display shows 150. 6. Deflate to zero and readjust to 000 7. Inflate again to 150mmHg and readjust Gain. 8. When pressure is at zero the readout should show 000 and when at 150 the readout should show 150. Handling: The tip of the fitting is filled with a pressure transferring gel. It is important not to disturb this gel as it may degrade the performance of the transducer. It is very important not to drop the fitting on its open end. At the tip of the fitting there is a small glass housing that may be damaged. 227 West Maple Ave Monrovia, CA. 91016 - Phone: 626 357-8827 Toll Free: 800 428-1816 Fax: 626 303 2998

159905 227 West Maple Ave Monrovia, CA. 91016 - Phone: 626 357-8827 Toll Free: 800 428-1816 Fax: 626 303 2998