CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

Similar documents
Brit. J. Anaesth. (1968), 40, 666. REBREATHING CIRCUITS A Review

Breathing Circuits. Product training

QED-100 Clinical Brief

Selecting and Connecting Breathing Systems

TRICHLOROETHYLENE CONCENTRATION FROM A " BOYLE " TYPE ANAESTHETIC APPARATUS. Department of Anaesthetics, Welsh National School of Medicine, Cardiff

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

J.A. BAIN AND W.E. SPOEREL~

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

CLINICAL ANAESTHESIA AND THE MULTIPLE-GAUZE CONDENSER-HUMIDIFIER

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems

Breathing Systems. Professor Khalid Bashir

THE PERFORMANCE OF A WALTON FIVE ANAESTHETIC MACHINE

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

HALOTHANE AND REBREATHING. Research Department of Anaesthetics, Royal College of Surgeons of England

Monitoring, Ventilation & Capnography

have expressed widely varying opinions on this topic (Catterall and Snow, 1960; Catterall, 1960;

Flow meter. bellow vaporizer. APL valve. Scavenging system

Anesthesiology Research and Scholarship. Follow this and additional works at:

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Recommendations on Checking Anaesthesia Delivery Systems

DURING the course of certain investigations it became

Let s talk about Capnography

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS

Guidelines on Monitoring in Anaesthesia

IT is frequently desirable to be able to determine the

Historically, rebreathing anesthesia systems

Sign up to receive ATOTW weekly -

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

(Received 16 January 1946)

Effect of Fresh Gas Flow on Isoflurane Concentrations during Low-flow Anaesthesia

Figure 1. A schematic diagram of the human respiratory system.

W. D. A. SMITH Research Department of Anaesthetics, Royal College of Surgeons of England, London

Human gas exchange. Question Paper. Save My Exams! The Home of Revision. Cambridge International Examinations. 56 minutes. Time Allowed: Score: /46

Medical Instruments in the Developing World

Section 14: Equipment failure

GENERAL ANAESTHESIA IN OPHTHALMIC SURGERY. Portsmouth Group of Hospitals

Anaesthesia breathing circuits

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION

Standard Operating Procedure

Checking a Breathing System

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Mechanical Ventilation

AUTOVENT 4000 VENTILATOR

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

Practical considerations in anaesthetising exotic species. By Keith Simpson BVSc MRCVS AMIIE(Electronics).

O-Two Self-Study Guide. Equinox Relieve

Medical Ventilators. Ryan Sanford Daniel Tajik

KOALA. Disposable Adult Anaesthetic Circuits for Low Flow Anaesthesia KO.S5.22MM. Adult Anaesthetic Circuits. Medical Pty Ltd

27/10/58. Background and Rationale. Background and Rationale. Background and Rationale. Objective. Background and Rationale

Anaesthetic infrastructure and supplies

Operation and Maintenance of the EPV200 Portable Ventilator

Breathing Process: Inhalation

Day-to-day management of Tracheostomies & Laryngectomies

RESPIRATORY MUSCLES IN HEALTH AND EMPHYSEMA *

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

Homecare Pneumology NEONATOLOGY ANAESTHESIA INTENSIVE CARE VENTILATION Step beyond the future

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

administration of 5% carbon dioxide to an infant has any chemical effect in

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

PROBLEM SET 8. SOLUTIONS April 15, 2004

Canine Endotracheal Intubation

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Exploring the relationship between Heart Rate (HR) and Ventilation Rate (R) in humans.

International Journal of Science, Environment and Technology, Vol. 5, No 5, 2016,

throughout. The constant-flow respiration was administered through a intravenously at appropriate intervals (in addition to the general

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

The use of pre-mixed nitrous oxide and oxygen in dental anaesthesia

APPARATUS Supplementary oxygenation with the laryngeal mask airway: a comparison of four devices

Bird Mark 7 Ventilator Manual

RESPIRATORY REGULATION DURING EXERCISE

Cardiovascular and respiratory adjustments to exercise

It is a product of proteins broken down in the mammal. It is exchanged for oxygen which is taken into the blood.

The Merlin Small Animal Ventilator by Vetronic Services Ltd. Contents page. Introduction.2. Features of Merlin. 3

Unit 15 Manual Resuscitators

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

Chapter 37: Pulmonary Ventilation. Chad & Angela

LISTADO DE CIRCUITOS DE PACIENTE

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

CALCULATING SOLUTIONS AND DRUG DOSE RATES. % solution = (weight (g) 100) volume of solution (ml)

Gas exchange measurement in module M- COVX

CHAPTER 3: The respiratory system

Respiration. Figure 22: Schematic representation of the respiratory system

PERFORMANCE OF NEEDLE VALVES

Cyril M. Conway MB BS FFARCS

(Received 9 September 1940)

THE literature on this subject, which was reviewed recently (CAMPBELL, doses of amytal, and in addition received A.C.E. mixture during the

Welcome to Maxstar Medical

GASEOUS HOMEOSTASIS AND THE CIRCLE SYSTEM

Update in Anaesthesia 31

CHAPTER 3: The cardio-respiratory system

The SAV04 Small Animal Ventilator. Contents Page

Hospital and Transport for Controlled Breathing

TWO MORE RESPIRATORS

Emergency Transport and Ventilation

Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia

Douglas and Haldane(2) has shown that the oxygen determinations. since it forms the basis of the "Coefficient of Utilisation" (Krrogh) and

Recommended Minimum Facilities for Safe Anaesthetic Practice in Operating Suites

Oxygen convulsions are believed by many workers to be caused by an accumulation

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the

Transcription:

Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre, British Oxygen Company Ltd. IN the past few years increasing attention has been paid to the elimination of carbon dioxide from the inspired anaesthetic mixture. This is important because of the effects of carbon dioxide on the respiratory pattern and depth, on cardiac rhythm, and on ganglionic and myoneural blockade With semiclosed systems of anaesthesia many attempts have been made to calculate, on a theoretical basis, the minimum gas flow rate necessary to eliminate rebreathing, or to reduce it to an acceptable level. The work of Molyneux and Pask (1951), Mapleson (1954), Woolmer and Lind (1954), has been important in this field. From their publications it can be seen that there is no general agreement either on the minimum flow rate necessary or on the most effective semiclosed system. The conclusions of Mapleson, derived from theoretical considerations, were broadly in agreement with those of Woolmer and Lind, derived from the use of a mechanical "patient", on the following points. (1) Permissible carbon dioxide concentrations in the inspired gases can be realized by the use of Magill's semiclosed system (Mapleson System "A" fig. 1), and a flow rate of fresh gas of 7 1/min. BAG FIG. 1 Mapleson system " A ' 196

CARBON DIOXIDE ELIMINATION 197 (2) At this flow rate the Magill attachment produces more effective carbon dioxide elimination than does the "Tpiece" system (Mapleson "E", fig. 2). The results obtained with a mechanical " patient" by Bracken and Sanderson (1955) on the Magill attachment, are also consistent with the conclusions of Mapleson. It was considered that it would be of interest to investigate these points in clinical anaesthesia, and to compare with them the results obtained with another semiclosed system, namely the Mapleson System "B", figure 3. There are difficulties in carrying out carbon dioxide analyses because the concentration at the patient's lips varies continuously during the breathing cycle; and any analysis apparatus which cannot respond to this rapidly changing concentration must inevitably give an average reading which is of little value and can be meaningless. In addition it is extremely difficult to collect identical portions of a number of respirations into a composite sample large enough for analysis. The infrared analyser is capable of following these variations in concentration, six virtually instantaneous analyses being taken per second. It introduces, however, a difficulty of its own. As a sample flow rate of approximately 2 1/min is required, the clinical level will be upset if this gas flow FRESH GASES >' ATMOSPHERE* FIG. 2 Maplsson system " E ' BAG FIG. 3 Mapleson system " B '

198 BRITISH JOURNAL OF ANAESTHESIA is removed from the circuit for any appreciable time. The solution of this problem was to return the sample to the system after it had been passed through the analyser. The return point must be " downstream " of the sampling point for the direction of flow being investigated (e.g. for inspired analyses the sample must be returned to the patient's lungs). This method of analysis has been previously described by Bracken and Sanderson, and in more detail by de Clive Lowe (to be published), who has also carried out some comparative tests using a technique similar to that employed in the present investigation. METHOD All investigations were carried out on adult patients. All patients received premedication with either papa veretum 21 mg and scopolamine 0.48 mg or pethidine 100 mg and scopolamine 0.48 mg. The cases were unselected except that surgery was at sites below the level of the clavicle, which allowed ready interchange of the various systems under investigation. For the most part relaxant drugs were not used except for initial intubation. Induction was effected in all cases with intravenous thiopentone. The patient was given suxamethonium or gallamine triethiodide, then ventilated with oxygen. A No. 7 (French) gum elastic catheter and a cuffed endotracheal tube of the maximum permissible size were then introduced into the trachea. The tip of the gum elastic catheter lay approximately 1 cm nearer the carina than the tip of the cuffed tube. A small T-junction was inserted between the endotracheal connection (i.e. the angle-piece) and the catheter mount. Through the side tube of this T- junction it was possible to extract a sample of respired gases. This sample was passed through the infrared analyser and returned to the trachea distal to the endotracheal cuffed tube via the gum elastic catheter. Whichever system was under investigation, anaesthesia was, in each case, maintained with a nitrous oxide oxygen mixture of a constant 5 1./2 1. composition, augmented by intermittent injections of a 5 per cent solution of thiopentone. Most patients also received between 25 and 75 mg of pethidine during anaesthesia. In an attempt to keep the tidal volume as large as possible all patients were maintained on as light a plane of anaesthesia as was compatible with the needs of surgery. As soon as spontaneous respiration had returned to a clinically satisfactory level, sampling was commenced, and this could be maintained for as long as desired. The systems under investigation were not applied in any set order but each patient had all three systems applied during the same operation; and no system was changed for another until the results of the carbon dioxide estimations were constant. In a few instances, to suit surgical needs, injections of suxamethonium chloride were made during operation, making artificial respiration necessary. In these circumstances results of further sampling were not recorded until spontaneous respiration had returned and the carbon dioxide level was again constant. RESULTS In a total of 28 cases investigated, 347 readings were taken, and the results of these are shown below:

CARBON DIOXIDE ELIMINATION 199 Circuit employed No. of readings Standard deviation Carbon dioxide % in inspired gases. Mean of readings. per cent 0.29 1. Magill attachment Mapleson " A " 2. Ayres " T-piece " Mapleson " E " 3. Mapleson " B " 107 0.26 131 0.67 1.59 109 0.88 1.75 The results using the Mapleson " E " require some comment. A hose volume of 200 ml was used at first and was found to be too small in some cases, since the inspired gases were diluted with air causing a lighter level of anaesthesia. The hose was therefore replaced by one of 320 ml capacity, and this proved, clinically, to be satisfactory in all cases. Doubling of the input to 14 1/min appeared to produce readings comparable with those on the Magill attachment at 7 1/min but insufficient readings were obtained for this result to be conclusive. These mean figures were examined statistically using Student's " t" test. The difference between the mean volume obtained with the Mapleson "A" system (0.29 per cent) and either the Mapleson "B " system (1.75 per cent) or Mapleson " E " system (1.59 per cent) was found to be statistically highly significant, as would be anticipated on inspection. However, again as would be expected, the difference between the mean value for the Mapleson "B " system (1.75 per cent) and that for the Mapleson " E " system (1.59 per cent) was found not to be statistically significant when compared with the variability of the readings obtained. This means that we have failed to detect any significant difference between the Mapleson "B" and Mapleson "E" systems with respect to their carbon dioxide elimination. DISCUSSION From the above results it can be seen that the readings fall into the pattern predicted by Mapleson. At a flow rate of 7 1/min the Magill attachment was more effective in clearing the expired carbon dioxide from the system than either the " T-piece " or Mapleson " B " systems. The results also compare in general with those of Woolmer and Lind using a flow rate of 7 1/min with the Magill and " T- piece " systems. Their results show that with the Magill attachment and the "artificial patient" the inspired carbon dioxide had a concentration of under half of that found by us with infrared analysis and an actual patient; in the case of the "T-piece" system the figure for the " artificial patient" was slightly over one and a half times higher than in the case of the real patient. The quantitative discrepancy in these results may be related to the relative volumes of the corrugated expiratory tubing and dead spaces in their system and ours. In the present investigation, while the " T-piece " used was clinically satisfactory, there may still have been some dilution of the inspired gases with air. Also, the effect of variation in wave

200 BRITISH JOURNAL OF ANAESTHESIA form, rate and depth of respiration must not be forgotten. It was noted clinically that when using the Mapleson " A " system the conduct of anaesthesia was smoother than with other systems which returned higher concentrations of carbon dioxide to the patient. It must be emphasized that the investigation described in this article is entirely clinical, and the results apply only to the particular systems used. Therefore no deductions can be made as to whether any of the three systems would be similarly effective when used with children or infants. No measurement was made of the tidal volume at the time of sampling, nor was any measurement made of the resistance required to " blow off " the valve on the Mapleson " A " and Mapleson "B " systems; however, it has been shown that the effect of this last variable is insignificant in elimination of carbon dioxide (Woolmer and Lind, 1954; Bracken and Sanderson, 1954). In general the experiment was designed to interfere as little as possible with those conditions normally found in day-to-day work. It is of interest that our clinical results compare closely with two theoretical examinations of the problem. SUMMARY Measurements were made under clinical conditions of the percentage of inhaled carbon dioxide returned to anaesthetized patients by three different types of semiclosed system commonly used in this country. The measurements were carried out with a continuous infrared carbon dioxide anlyser. The results obtained compared closely with theoretical results obtained by other workers. CONCLUSIONS It may be claimed from these tests that, in adults, under the conditions used, the Mapleson " E " (T-piece) and Mapleson " B " systems return a high concentration of carbon dioxide to the patient. The inspired concentrations from these two circuits are comparable with each other, and are both greater than that from the Mapleson "A" system (Magill attachment), under the same conditions. ACKNOWLEDGMENTS We wish to acknowledge the receipt of a small grant in aid of this work from the East Grinstead Anaesthetic Research Trust. We are grateful to Mr. C. R. Mc- Laughlin for help in the preparation of this article One of us (A.B.) wishes to thank the British Oxygen Company Ltd. for permission to take part in this publication, and we are all indebted to this company for their generous loan of the analyser. REFERENCES Bracken, A., and Sanderson, D.M. (1955). Brit. J. Anaesth., 27, 428. Mapleson, W. W. (1954). Brit. J. Anaesth., 26, 323. Molyneux, L., and Pask, E. A. (1951). Brit. J. Anaesth., 23, 81. Woolmer, R.. and Lind. B. (1954). Brit. J. Anaesth., 26, 316.