Ripple tank with LED light source. Ripple tank with LED light source, complete

Similar documents
Ripple Tank with LED-light source, complete

Wave phenomena in a ripple tank

RIPPLE TANK - with rippler & kit

MKIII RIPPLE-STROBE TANK

Digital Melting Point Apparatus

Technical Note. Determining the surface tension of liquids by measurements on pendant drops

Chapters 25: Waves. f = 1 T. v =!f. Text: Chapter 25 Think and Explain: 1-10 Think and Solve: 1-4

User Manual PTW-DensiX. D / Di/Zi

Instruction Manual BA90 Precision Barometer

WAVES. Pulses are disturbances or a single wave motion. A continuous production of pulses will give rise to a progressive wave (wave train).

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic

EcoCell Vertical Electrophoresis Unit

INSTRUCTION MANUAL FOR USE OF BABY WALKER FLOWER

(Supplementary) Investigation Waves in a Ripple Tank

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Waves. Kevin Small or

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Instruction Manual Contact Pressure Vacuum Gauge

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

OPERATING AND MAINTENANCE MANUAL GOLD LINE EXTRACTION CLEANERS M14-M26

Fingertip Pulse Oximeter

Vibration isolation system 1VIS10W. User manual

Mechanical waves Electromagnetic waves

Ripple Tank Exploring the Properties of Waves Using a Ripple Tank

Assembly and instruction manual for the Ripple Tank no

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Manual Refractometer PCE-ALK

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

Thermochemistry/Calorimetry. Adiabatic coefficient of gases Flammersfeld oscillator LEC 02. What you need: What you can learn about

Outline Chapter 7 Waves

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

SAUTER FK Version /2016 GB

Ripple Tank. Equipment:

Waves. Name and Surname: Class: L E A R N I N G O U T C O M E. What are waves? Why are waves formed?

Installation and Use Manual

SVEA II GSM Connector User Manual

Manual. Kingpad mc-32 Edition. Bus system control pad for installation in the Graupner mc-32 transmitter. No Copyright Graupner/SJ GmbH

Pre AP Physics: Unit 7 Vibrations, Waves, and Sound. Clear Creek High School

Ripple Tank: Instruction Manual

NATURE AND PROPERTIES OF WAVES P.1

QPL... Pressure Switches. Siemens Building Technologies HVAC Products

PORTABLE WORK LIGHT MODEL NO: CL6FS OPERATION INSTRUCTIONS PART NO: GC0315

BMR 2100XL Operating Instructions Page 1 of 13

Operating instructions Installation instructions To be kept in the vehicle!

Pendant Drop Measurements

KERN EG/EW Version /02

Instruction Manual EMA 84 Digital Pressure Gauge

Declaration of Conformity as per Directive 97/23/EC

WIKA INSTRUMENT CORPORATION

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e

EC214 EC215 - EC215R Bench Conductivity Meters

Unit 7: Waves and Sound

HI 2314 HI 2315 HI 23151

Installation and Maintenance Instruction Manual

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies.

ROMPA SUPA SPARKLE BUBBLE TUBE 20288

Installation & Maintenance Instructions TEC-11N TIMER. Analogue electronic timer 02/09

LED 14W POLO BULKHEAD WITH MICROWAVE SENSOR

Item # in1 Rotating Table. Inch BILLIARDS AIR HOCKEY TABLE TENNIS

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

Contents. Operating instructions. Pressure switch, heavy-duty version Model PSM-520

Agilent 1220 Infinity II LC Mobile Upgrade Kit

METEOR LSR36500 LIGHTING UNDERWATER POOL LIGHT. Operating Instructions

LVU2800 Series. Ultrasonic Level Transmitter

! Warning, refer to accompanying documents.

Achat 112 mid-high-range speaker. user manual

Operating Instructions Temperature Measuring System MF420-5T-100

Siphon vessel / Priming aid

Introduction to Waves

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Product instruction manual Easymount Sign Laminators EM-S1400C, EM-S1600C, EM-S1400H, EM-S1600H

SkillGuide. User Guide. English

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00

What is a wave? A wave is a disturbance that transfers energy from place to place.

icreasepro Creaser Operators Manual

Ripple Tank P CONTENTS: DEMONSTRATION SETUP INSTRUCTIONS:

Declaration of Conformity as per Directive 97/23/EC

Pull-rope emergency stop switch Type HENEX and SEGEX OPERATING INSTRUCTIONS

19 Which row shows an example of a transverse wave and an example of a longitudinal wave?

Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment

RIPPLE TANK CAT NO. PH 0767A

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Waves and Sound. Honors Physics

Instruction Manual for Configura Cushionair Portable Pump

micro CA-100 Inspection Camera

Operating Instructions in compliance with Pressure Equipment Directive 2014/68/EU. FAS Brass Check Valve RDL

Theatre Control Panel. Instructions for Use. January 2017 ASF-29 (1)

Operating instructions for variable area flow meter with needle valve. Model: UVR/UTR

Period: Date: 1. A single disturbance that moves from point to point through a medium is called a. a. period b. periodic wave c. wavelength d.

NovelBright LED Illuminator

INTRODUCTION TO WAVES. Dr. Watchara Liewrian

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

SPH4U UNIVERSITY PHYSICS

TrekMill USER'S MANUAL MM5050

Wave a repeating disturbance or movement that transfers energy through matter or space

Air Track Rail Air Track Rail, Basic Equipment

DIVER NAVIGATION BOARD. RJE International, Inc. RJE International Feb 1, TAC-100D Revision 10

INSTRUCTION MANUAL. January 23, 2003, Revision 0

Transcription:

Ripple tank with LED light source Ripple tank with LED light source, complete 11260-02 11260-88 PHYWE Systeme GmbH & Co. KG Robert-Bosch-Breite 10 37079 Göttingen Germany Telefon +49 (0) 551 604 0 Fax +49 (0) 551 604 107 E-mail info@phywe.de Internet www.phywe.com Operating instructions The unit complies with the corresponding EC guidelines Fig. 1: Ripple tank with LED light source, complete 11260-88 TABLE OF CONTENTS 1 SAFETY PRECAUTIONS 2 PURPOSE AND DESCRIPTION 3 FUNCTIONAL AND OPERATING ELEMENTS 4 HANDLING 5 NOTES ON OPERATION 6 TECHNICAL DATA 7 SCOPE OF SUPPLY 8 ACCESSORIES 9 LITERATURE 10 WARRANTY 11 WASTE DISPOSAL 1 SAFETY PRECAUTIONS Attention! Only use the instrument for the purpose for which it was designed. Only use the instrument in dry rooms in which there is no risk of explosion. Before applying mains voltage, ensure that the earth lead of the experimental set-up is properly connected to the mains earth lead. The mains plug is only to be plugged into a mains socket which has an earth lead. Do not use an extension cable without earth lead, as this would negate the protective effect. Check that your mains supply voltage corresponds to that given on the type plate fixed to the supplied power supply. Assemble the experimental set-up in such a way that the mains switch and/or plug connector of the instrument are freely accessible. Keep the ventilation slots of the experimental set-up free of any covering. Do not attempt to open the instrument. Take care that no liquids or objects enter in through the ventilation slots. Do not connect any other devices to the ripple tank. Caution: Separate the instrument completely from the electric mains before loosening, exchanging or removing any cable connection! Separate the instrument completely from the mains before filling or emptying the wave tray! Longer observation of the wave image under stroboscopic illumination can trigger epileptic seizures in endangered people. Do not stare directly into the glowing LED on purpose. Carefully read these operating instructions completely before operating this instrument. This is necessary to avoid damage to it, as well as for user-safety. 1

2 PURPOSE AND DESCRIPTION The ripple tank serves for the generation and study of surface waves in water. It enables not only experiments that are characteristic for surface waves in water to be carried out, but also experiments that are analogous to those carried out on light waves. Many important wave optics laws can be clearly and simply demonstrated by means of water waves. The following describes how the ripple tank functions and how water wave portrayal is accomplished. Water waves are generated by a mechanical vibration generator that is integrated in the ripple tank (loudspeaker principle). This generator can cause either a single plug or several plugs to be dipped into the water in the wave tray for the generation of circular waves, or a slide to be dipped in for the generation of flat waves with a frequency of between 5 Hz and 60 Hz. Periodic dipping causes surface waves to be created. A green LED is fixed to the bottom of the ripple tank housing. This illuminates the wave tray from below and, due to the optical principle of shadow projection (Fig. 2), produces a representation of the water waves that is rich in contrast. Either continuous light or stroboscopic light can be used. Stroboscopic light allows the propagation velocity of the water waves to be slowed down to a complete standstill. Details on the use of stroboscopic light are given in section 4 Handling. Screen Wave tray 3 FUNCTIONAL AND OPERATING ELEMENTS 3.1 Control panel The ripple tank apparatus is equipped with a control panel with 14 keys (Fig. 3) and an LCD screen. In the following, an introduction is given to the adjustment possibilities and the wave generator required for the generation of water waves. Fig. 3: The 14 keys control panel is situated on the top of the housing of the ripple tank. 1 Vib Switches the vibration generator(s) on and off. 2 LED Switches back and forth between the following settings of the LED: Continuous light Stroboscopic light Off 3 f Pressing (+) increases, and pressing ( ) reduces the frequency within the 5 Hz and 60 Hz range. LED Fig. 2: The shadow projection principle Light from a point light source passes through the wave tray and is partially subject to deflection according to the structure of the surface waves. This deflection generates a bright-dark pattern on a screen. When the drawing table that is supplied by default is placed on the ripple tank, and a sheet of white paper is placed on the table, then the water waves can be observed on the sheet of paper. No focussing optical elements are used in the portrayal of the waves. The creation of the portrayal results exclusively from the diffraction properties of the water surface. Wave troughs make the light from the LED divergent, water crests make it convergent. Because of this, the wave troughs can be recognised as dark stripes on the observation surface (sheet of paper) and the wave crests as bright stripes. The result of this principle is a wave image that is sharp and rich in contrast. It must be noted here though, that a good image contrast is only given at small wave amplitudes. At larger amplitudes, the light rays emitted by the LED combine far before they reach the observation screen and diverge from each other again on the further long way to it. The result is the disappearance of the wave image. For this reason, the wave generator is equipped with an amplitude control element which ensures that the optimal generator amplitude can be set for each individual experiment. 4 Δf A difference in frequency between the vibration generator and the stroboscopic light can be set here. The two frequencies are equal in the default state, in which the waves are depicted as a stationary wave under stroboscopic light. A frequency difference can be set with the (+) and ( ) keys to obtain a slow motion portrayal of the waves. 5 Δφ A phase difference between the integrated vibration generator and an additional, external vibration generator (11260-10) can be set here. In the default state, the generator oscillates in phase. A phase difference of between 0 and 360 can be set in steps of 15. 6 Amplitude Pressing (+) increases the amplitude of the vibration generator (4 steps), pressing ( ) reduces it. 7 Pulse Pressing this key gives the vibration generator a single pulse which causes the plug to be dipped into the water just once. On releasing the key, the plug is brought back to the rest position. If the external vibration generator (19) should be connected, it will also be operated. 2

8 Reset Switches the vibration generator and the LED off. All other settings stay at their current value to allow a quick restart of the ripple tank. 17 13 14 9 Cal int. For calibration of the integrated vibration generator. The amplitude of the exciter can be set in a range from 80 % to 120 % independent of the external vibration generator. If the amplitudes should vary, this enables a synchronisation of the amplitudes of the integrated and the external vibration generator. First press the Cal int. key. Immediately after that use the (+) and ( ) keys of the Amplitude selector (6) for finetuning. The calibration is kept even after a reset (8) or after unplugging the power supply (11). 10 Cal ext. Same as (9) for calibration of the external vibration generator. The amplitude of the external exciter can be set independently of that of the internal exciter in a range from 80 % to 120 %. First press the Cal ext. key. Immediately after that sue the (+) and ( ) keys of the Amplitude selector (6) for finetuning. 3.2 Further functional elements In addition to the control panel, the ripple tank has the following further functional elements (see Fig. 4, 5 and 6): 16 Fig. 5: Integrated vibration generator (15) with wave tray (13), holding rod (14), and plug (17). 14 Adjusting screws, 4 pieces For horizontal alignment of the wave tray. 15 Integrated vibration generator Serves to hold each type of exciter. By loosening the two adjusting screws in the base, the exciter can be rotated in plane. 16 Holding rod Serves to mount and position the respective wave generator. 17 Wave generator Serves to generate waves by dipping into the water while the integrated vibration generator is switched on. 15 18 18 Tensioning screw For raising and lowering the wave generator to a suitable position in the water of the wave tray. 11 12 19 External vibration generator (accessory: 11260-10) Can be used as second generator and is necessary for the Doppler effect experiment. 19 Fig. 4: Socket for the power supply (11) and sockets for the connection of the external vibration generator (12). 11 Connecting socket 12 V For connection of the ripple tank apparatus to an external power supply (supplied by default). The socket is on the side of the housing. 12 Connecting sockets, external vibration generator An external vibration generator with 4 mm plugs can be connected here. The polarity of the connection is of importance. To check polarity, the function Pulse (7) can be used. If both the internal and external vibration generators move down when the button Pulse is pressed, both will move in phase. Otherwise the polarity of the external vibration generator can be changed or the two vibrators work with a phase difference of 180. 13 Wave tray Waves are generated in the water filled in the wave tray. Plastic foam has been stuck to the inner sides of the wave tray. This allows an almost complete absorption of the water waves that hit the sides of the wave tray and thereby avoids unwanted wave reflections. Fig. 6: External vibration generator (11260-10) with accessories. 4 HANDLING This section describes how to put the water wave apparatus into operation and cause water waves to be generated. Please read this section carefully to avoid failure or improper operation. Note: To protect the apparatus from damage, the vibration generator turns off automatically after 10 minutes of continuous operation. Experiments are not expected to take longer time in this operation mode. 3

4.1 Putting the ripple tank into operation First connect the connecting socket (11) of the ripple tank via the provided power supply to the mains. The LCD screen lights up and the ripple tank is ready to use. 4.2 Filling the wave tray First, the empty wave tray has to be placed on the ripple tank. Afterwards, it can be carefully filled with water. The wash-bottle supplied can be used to fill the wave tray. Fill the tray about three quarters high. After this, to have reflection of water waves at the sides suppressed, moisture the whole side-foam with a finger so that water can wet it. Use the adjusting screws (14) to align the wave tray horizontally. To do so, we recommend that you look flat across the water surface and use the height of the water level at the different sides of the tray as reference. Now use the screws to bring the tray to about the same height on all sides and so to a horizontal position. Fix the holding rod with slide to the exciter and bring it to the lower end of the wave tray. The wave tray must now be correctly adjusted, as the slide must be exactly evenly immersed into the water. If this is not done precisely, the wave image will be distorted and unclear. See the information given on the correct adjustment of the wave tray in section 4.2 Filling the wave tray. The slide must also be set to be horizontal. Therefore, fix it as horizontally as possible to the integrated exciter and tighten the clamping screw on the exciter further until the slide is held slightly above the water surface. Use the water surface as reference to adjust the slide to the horizontal position. Afterwards, turn the clamping screw on the exciter further until the slide is dipped into the water (about 1 to 2 mm). Set a frequency between 18 Hz and 25 Hz and choose a small amplitude as well as continuous light. A wave image should be obtained that is similar to that shown in Fig. 8. Attention! When lifting the filled tray, the water tends to slop! Pay attention that no water enters the device. We recommend using the wash-bottle or e.g. a sponge to finally remove the water while the wave tray is still resting on the ripple tank. 4.3 Cleaning The cover plate and the wave tray of the ripple tank consist of acrylic glass. Please perform cleaning of the cover plate and the wave tray with suitable (i.e., non-abrasive) tissues without aggressive chemicals added. 4.4 Generation of circular waves Fix the holding rod (14) with a single plug (17) to the integrated generator (15) and bring it to the middle of the wave tray. Turn the clamping screw on the generator until the plug is dipped into the water in the wave tray. Set to continuous light, to a frequency of between 5 Hz and about 20 Hz and to a small amplitude (1 or 2). Prepare the observation surface by placing a sheet of white paper on the drawing table situated over the wave tray. Darkening the room could be useful for better visibility of the wave image. Should the wave image be distorted or unclear, vary the frequency f and/or the amplitude until a wave image is obtained that is similar to that shown in Fig. 7. Fig. 8: Flat waves generated by the plane exciter. Should the wave image be distorted or unclear, vary the frequency, the amplitude or the immersion depth. When no improvement in the wave image results, it could be possible that the slide is not correctly adjusted and therefore does not dip evenly into the water. In this case, align the slide as described above. 4.6 Usage of the stroboscopic light The use of stroboscopic light is only instructed in those individual experiments in which it is required to reach the experimental target. All other experiments should first be carried out without stroboscopic light because stroboscopic illumination considerably changes the impression of the image, partially. Interference patterns can be far more impressively seen when continually projected than when under stroboscopic illumination. In this context it must be considered that for continuous light, the human eye is only capable of recognizing the wave image in the lower frequency range. Because of this, we recommend that only frequencies of up to about 30 Hz be used under continuous light. Stroboscopic light should be used with shorter wavelengths (higher frequencies), so that wave propagation is slowed down and made visible to the human eye. Fig. 7: Circular waves generated by the plug-shaped exciter. 4.5 Generation of flat waves The generation of flat waves is somewhat more difficult, because it requires an exact alignment of the wave tray with the slide used. It could be necessary to add 1 to 2 drops of a soap solution to the water for better wetting. Do not exceed this amount, however, as then turbidity and foaming would occur and cause deterioration in the quality of the image. 5 NOTES ON OPERATION This high-quality instrument fulfils all of the technical requirements that are compiled in current EC guidelines. The characteristics of this product qualify it for the CE mark. This instrument is only to be put into operation under specialist supervision in a controlled electromagnetic environment in research, educational and training facilities (schools, universities, institutes and laboratories). This means that in such an environment, no mobile phones etc. are to be used in the immediate vicinity. The individual connecting leads are each not to be longer than 2 m. 4

The instrument can be so influenced by electrostatic charges and other electromagnetic phenomena that it no longer functions within the given technical specifications. The following measures reduce or do away with disturbances: Avoid fitted carpets; ensure potential equalization; carry out experiments on a conductive, earthed surface, use screened cables, do not operate high-frequency emitters (radios, mobile phones) in the immediate vicinity. This instrument complies with Class A, Group 1 of the EN 55011 norm and it is only permitted to operate unlimitedly outside of residential areas. Should electromagnetic disturbances, despite the limitations of usage to a technical room of a school or other place of education, occur in a surrounding residential area, then it can be demanded of the operator that he or she carries out appropriate measures (e.g. screening, longer distance away from sensitive equipment, shorter operating principles, as short as possible connecting cables etc.) at his or her own cost. 6 TECHNICAL DATA (typical for 25 C) Operating temperature range 5...40 C Relative humidity < 80 % Frequency 5 60 Hz Vibration generator amplitude 1 4 Stroboscope frequency (Δf) 2.5 2.5 Hz Phase difference 0 360 Line voltage 110 240 V Mains frequency 50/60 Hz LED luminous flux, max. 285 lm Dimensions (mm) 300 x 370 x 330 (W x L x H) Tray area (mm) 280 x 210 (W x L) Weight approx. 5.7 kg 7 SCOPE OF SUPPLY The Ripple tank with LED light source, complete (11260-88) is supplied with the following components: Basic unit 11260-02 Power supply 12 V DC / 2 A 12151-99 Accessory set for ripple tank 11260-12 Drawing table for ripple tank 11260-13 Wave tray for ripple tank 11260-14 8 ACCESSORIES Various experiments, such as the Doppler effect experiment, or demonstration of experiments via a mirror, require accessories. The following are available: External vibration generator for ripple tank 11260-10 Demonstration set with mirror for ripple tank 11260-30 9 LITERATURE Demonstration Experiments: Experiments with the ripple tank 16040-02 10 WARRANTY We guarantee the instrument supplied by us for a period of 24 months within the EU, or for 12 months outside of the EU. Excepted from the guarantee are damages that result from disregarding the Operating Instructions, from improper handling of the instrument or from natural wear. The manufacturer can only be held responsible for the function and technical safety characteristics of the instrument, when maintenance, repairs and alterations to the instrument are only carried out by the manufacturer or by personnel who have been explicitly authorized by him to do so. 11 WASTE DISPOSAL The packaging consists predominately of environmentally compatible materials that can be passed on for disposal by the local recycling service. Should you no longer require this product, do not dispose of it with the household refuse. Please return it to the address below for proper waste disposal. PHYWE Systeme GmbH & Co. KG Customer Service Robert-Bosch-Breite 10 D 37079 Göttingen Germany Phone +49 (0) 551 604-274 Fax +49 (0) 551 604-246 Fig. 9: Scope of supply of ripple tank with LED, complete 11260-88. 5