Aeris Performance 2. Product Manual

Similar documents
Element DS. Product Manual

Wave Comfort 2. Product Manual

AllPro Foot. Posterior Mount Product Manual

AllPro Foot. Product Manual

Aeris Activity. Product Manual

K2 Foot. Product Manual

PRODUCT MANUAL. patent pending

Kinterra Instructions for Use Product Number: ROM-00-XXXXX-XX

Kintrol Instructions for Use Product Number: VS4

Foot System Prosthetist Manual

MOTION FOOT MX. Manual

1C11 Terion K2. Confidence with every step. Information for Practitioners

The Starting Point. Prosthetic Alignment in the Transtibial Amputee. Outline. COM Motion in the Coronal Plane

Your Life. Your Adventure. Triton Family of Products

Instructions for 1324 AP / AL Stance Flexion 5 Bar Mechanical Knee

1B1 Meridium. Reclaim your way. Information for practitioners. Meridium Ottobock 1

Niagara Foot. Dynamic Performance at Every Step

Normal Gait and Dynamic Function purpose of the foot in ambulation. Normal Gait and Dynamic Function purpose of the foot in ambulation

SYMBIONIC LEG 3 INTRODUCTION KEY FEATURES ÖSSUR DYNAMIC SOLUTIONS

RAIZE. from Hosmer PRODUCT MANUAL

Life Without Limitations

WalkOn product range. Dynamic Ankle-Foot Orthoses. Information for specialist dealers

Ankle biomechanics demonstrates excessive and prolonged time to peak rearfoot eversion (see Foot Complex graph). We would not necessarily expect

Foot mechanics & implications on training, posture and movement

Otto Bock and Springlite Feet

Meridium. Reclaim your way. Information for clinicians

PROPRIO FOOT INTRODUCTION KEY FEATURES ÖSSUR DYNAMIC SOLUTIONS

Axis of rotation is always perpendicular to the plane of movement

Gait. Kinesiology RHS 341 Lecture 12 Dr. Einas Al-Eisa

Dynamic Warm up. the age of the athlete current physical condition and prior exercise experience

Dynamix Ankle Foot Orthoses Range

Mechanical Knee Prescription Guide

Purpose. Outline. Angle definition. Objectives:

C-Brace Orthotronic Mobility System

Stability Trainer. New! Balance Products. Rocker and Wobble Boards. New!

SECTION 4 - POSITIVE CASTING

Plié 3 Microprocessor Knee

Normal and Abnormal Gait

ASSESMENT Introduction REPORTS Running Reports Walking Reports Written Report

Case Study: Chronic Plantar Heel Pain/Plantar Fasciopathy. CASE STUDY PRESENTATION by Resonance Podiatry and Gait Labs

SPORTS FEET. Break through your barriers

video Purpose Pathological Gait Objectives: Primary, Secondary and Compensatory Gait Deviations in CP AACPDM IC #3 1

ATHLETES AND ORTHOTICS. January 29, 2014

PRO-FLEX. Less load, more dynamics

Sample Biomechanical Report

Assessments SIMPLY GAIT. Posture and Gait. Observing Posture and Gait. Postural Assessment. Postural Assessment 6/28/2016

A7HLE71CO PHYSICAL THERAPY

Gait Analysis at Your Fingertips:

Terion 1C10. Back on track. 1C10 Terion Ottobock 1

3 people 3 unique lifestyles 3 advanced foot solutions

video Outline Pre-requisites of Typical Gait Case Studies Case 1 L5 Myelomeningocele Case 1 L5 Myelomeningocele

The DAFO Guide to Brace Selection

Synchronised Swimming. Figure Grade 2

Unity Sleeveless Elevated Vacuum System

Recent Advances in Orthotic Therapy for. Plantar Fasciitis. An Evidence Based Approach. Lawrence Z. Huppin, D.P.M.

KICKBIKE Your key to optimum sports performance

Increasing ankle push-off work with a powered prosthesis does not necessarily reduce metabolic rate for transtibial amputees

Troubleshooting and Correcting DAFO Fit Problems

Guide to the Sulcus Stick

C-Brace Reimbursement Guide

Routine For: Total Knee Arthroplasty (All)

1. Hip flexion Muscles: Iliopsoas (psoas major + iliacus)

C-Brace Reimbursement Guide

The Mechanics of Modern BREASTSTROKE Swimming Dr Ralph Richards

Saucony Liberty ISO. Overview (from Website) Shoe Specifications: (from website) Disclaimer.

engineered products Foldable Topside Creeper Safety First VIDEO INSTRUCTIONS ARE AVAILABLE ON OUR WEBSITE AT:

The Problem. An Innovative Approach to the Injured Runner. Dosage. Mechanics. Structure! Postural Observations. Lower Quarter Assessment

Synchronised Swimming. Figure Grade 1

1C30 Trias. Moving in harmony. Information for users. 1C30 Trias Ottobock 1

12/4/2010 3:10 / 3:40

Belton High School Marching Fundamentals 2016

AEROBIC GYMNASTICS Code of Points APPENDIX II Guide to Judging Execution and Difficulty

USER GUIDE ROLLING WALKERS INSTRUCTIONS FOR USE JANUARY, Maximum User Weight: 170kg (Do not exceed this weight)

Running Gait Mechanics. Walking vs Running. Ankle Joint Complex Sagittal Plane. As speed increases, when has walking ended and running begun?

Before you start! Is iwalk 2.0 right for you?

Running from injury 2

Analysis of ankle kinetics and energy consumption with an advanced microprocessor controlled ankle foot prosthesis.

Giovanni Alfonso Borelli Father of Biomechanics

Beginner Skier Progression -Class Levels One and Two Never-ever skiers to skiers making advanced wedge turns

Empower. Reclaim your power. Information for technicians. Empower Ottobock 1

INSTALLATION INSTRUCTIONS. Parts List. Tools Required. Before You Begin. Installation. Customer Information BICYCLE ATTACHMENT JUL.

SETTING THE HANDLE HEIGHT ON THE ROLLATOR

CHAPTER IV FINITE ELEMENT ANALYSIS OF THE KNEE JOINT WITHOUT A MEDICAL IMPLANT

to decrease this force without compromising alignment. Four methods of heel modification are frequently used to decrease destabilizing forces at heel

USER MANUAL. All rights reserved

The Unified Theory of Foot Function

SPINNER RIDE GETTING STARTED GUIDE. Welcome to a personalized fitness experience for your members

empower Reclaim your power. Information for technicians empower Ottobock 1

USER MANUAL

From early rehab to full recovery RHEO KNEE XC. go anywhere

Loading your garment onto the applicator. Helpful tips. Please ensure that your garment is not inside out before you begin.

Final Assembly Instructions Bikes with Quill Stems

SASK-A-POLE OWNERS AND USERS MANUAL

As a physiotherapist I see many runners in my practice,

Tripod Series OPERATOR S MANUAL

Lumbar Decompression Surgery: Correction of foot drop

Afoot. What s VASYLI. Biomechanical foot function: a Podiatric perspective. Introduction: Anatomical structure and function. Continued on Page 2

10/22/15. Walking vs Running. Normal Running Mechanics. Treadmill vs. Overground Are they the same? Importance of Gait Analysis.

Gait analysis through sound

Synchronised Swimming. Skill Level 5

Transcription:

Aeris Performance 2 Product Manual

Instructions The Aeris Performance 2 is designed to be maintenance free. The foot is water resistant; however, if the foot is submerged in water, the foot and foot shell should be rinsed with fresh water and dried immediately. The Aeris Performance 2 should be inspected every 6 months for signs of abnormal wear and that the attachment/alignment screws are secure. Instructions to the Practitioner Please review the indications, contraindications, and FAQ sections of the manual before use of the foot. These instructions should be read prior to fitting and followed to ensure the proper integration of the Aeris Performance 2 into the patient s prosthetic system. The foot stiffness is based on weight and activity level. Please provide accurate patient information so that the appropriate foot may be selected. Product Specifications Indications Moderate to active transtibial or transfemoral amputees as defined by functional K3 and K4 levels Unilateral or bilateral patients Patients that would benefit from greater energy return and multi-axial function Patients weighing up to 330 lbs. (150 kg) Contraindications Build height below 6.75 in. (17 cm) Patients weighing over 330 lbs. (150 kg) Product Specifications Patient weight: Up to 330 lbs. (150 kg) Foot weight: 18.35 oz. (520 g) Build height: 6.75 in. (17 cm) Functional level: K3 K4 Durable; meets ISO-22675 standard 2

Warranty 36 months from date of patient fitting Foot Shell (sold separately) - 6 months from date of patient fitting The Aeris Performance 2 has been designed and manufactured for specific patient weights. Failure to follow the weight guidelines and/or overload conditions caused by the patient, such as heavy lifting, high impact sports, or abusive activities that would otherwise damage the natural limb, may void the warranty. Satisfaction Guarantee 60 days from date of patient fitting Installation 1.0 - Product Description The Aeris Performance 2 foot uses two carbon composite elements that conform to terrain while storing and releasing energy during gait. The foot is designed to be used with most pyramid receiver devices (Figure 1). The pyramid dome for the foot is permanently attached to the Pylon Main Spring Lower Foot Plate pylon (main and top) spring and should not be removed. Pylon Top Spring Split Toe Design Figure 1 2.0 Static (Bench) Alignment Standard bench alignment techniques may be used for the Aeris Performance 2 foot. Before aligning, the initial heel height should be established. The foot is designed for a 3/8 in. or 1 cm heel height. The initial heel height can be established with a simple spacer under the heel. The top of the pyramid should be parallel with the work surface before proceeding with alignment. A backward leaning pylon indicates that the heel height is too low and will make late stance rollover difficult. 3

Transtibial Bench Alignment 5 The socket should be set in the proper amount of inset found in the evaluation. When using an integrated shuttle lock/distal attachment component, the plum line from the bisection of the socket at the proximal brim in the frontal and sagittal plane should bisect the ankle pyramid. When using separate suspension and attachment components, the foot may be slightly inset 1 12 mm depending on the limb Figure 3 length. Short limb lengths are set with very little inset of 2 3 mm and longer limb lengths may tolerate a greater varus thrust at 10 12 mm. The longitudinal axis of the foot will be outset approximately 5 by aligning the medial border of the foot with the line of progression. Transfemoral Bench Alignment Standard TKA alignment can be utilized with the trochanter line bisecting the distal ankle. The knee is set 3-6 mm posterior to the TKA line. A plum line may also be used from the proximal sagittal socket bisection, falling 10 mm anterior to the knee axis (or through the knee axis for an SNS unit) and the 10 15 mm posterior to the mid-foot. The knee and foot should both be aligned with 5 of toe out and external rotation respectively. The socket should also be set in the proper adduction angle of 7 12 and the flexion angle should be 5 more than the hip flexion contracture if present. 2.1 - Dynamic Alignment The foot is flexible and conforms to uneven terrain. This characteristic may mask forefoot loading anomalies during static alignment that may then become more noticeable during dynamic alignment. Small alignment changes will smooth the transition from heel to toe and optimize gait. Patient feedback during this process is essential. Adjustments of the plantar and dorsiflexion angles will help the patient achieve a smooth transition from heel to toe. The pylon should remain vertical in the frontal plane throughout gait. 4

Check for smoothness of gait and ground contact throughout the stance phase of gait. If the tibial progression is slowed from heel strike to midstance, or the heel compression is too great, dorsiflexion of foot may correct this problem. If the socket progresses rapidly forward from heel strike to midstance or the heel seems too hard, plantarflexion of the foot may solve this problem. If the foot progresses too quickly from midstance to toe loading, increased plantarflexion may solve this issue. If the foot hesitates from midstance to toe loading, dorsiflexion may be indicated. Check to make sure the pylon is vertical throughout gait. If there is a medial lean, tighten the proximal medial screw; if there is a lateral lean, tighten the proximal lateral screw. Special Considerations Due to the torsion compliance of the foot, it is very important to establish the proper external rotation of the foot in relationship to the socket. If the foot rotates too far internally or externally, it may feel less stable to the patient. Making an external rotation adjustment during the dynamic alignment should fine-tune the stability of the forefoot and improve the rollover characteristics of the foot. 5

3.0 - Foot Shell Installation and Removal The foot features a unique cosmetic foot shell that is flexible and durable. Use care in the installation and removal of the foot shell to maintain its appearance and durability. NOTE: Never use a sharp edged tool such as a screwdriver to install or remove the foot shell. Installation Pull the Spectra sock provided onto the foot from toe to heel, pulling excess material to the ankle so that it does not bunch under the Figure 4 heel or toe of the foot. Heel Lock Insert the forefoot into the foot shell as far as possible. Set the heel on a supportive surface with the toe up and push the shell onto the foot until the toe is in position. Rotate the foot side to side to allow the foot shell to slide onto the heel Push the foot shell up onto the heel or, if necessary, insert a shoehorn into the foot shell and allow the heel to slide down a shoehorn into the heel lock. IMPORTANT: The heel of the lower foot plate must slide into the heel lock in the foot shell for proper alignment and to secure the foot in the foot shell (Figure 4). Removal Place the foot on the bench so that the heel is hanging over the edge of the bench. Apply downward force to the top portion of the foot shell at the heel and the heel plate should pop out of the heel lock, allowing removal of the foot shell by hand. If the foot shell is too tight, a smooth edged shoe horn may be used to disengage the heel lock. 6

Frequently Asked Questions What can the practitioner do if the heel or toe is too soft or too firm? The heel and toe rollover resistance may be fine tuned during the dynamic alignment (section 2.1) by plantarflexing or dorsiflexing the forefoot. How long should the foot shell last? The foot shell is designed to provide realistic appearance and maximum performance of the Aeris Performance 2. The life of the foot shell will depend on level of activity and degree to which it is protected from wear and damage with socks and shoes. Can the pyramid be removed so that the foot can be bolted directly to the socket or other attachment? No, the pyramid attachment should not be removed or altered in any way. Doing so will void the warranty and could put the patient at risk of injury. Daily Care for the Patient Patients should clean the prosthetic foot shell with a soft cloth and a soap and water solution and should inspect the shell for the presence of sand or other debris weekly. The foot shell may also be cleaned with rubbing alcohol (70%). Do not use acetone. It will damage the foot shell. If the foot performance changes or if it makes noise, the patient should immediately contact his or her practitioner. 7

www.fillauer.com Fillauer LLC 2710 Amnicola Highway Chattanooga, TN 37406 423.624.0946 Fax 423.629.7936 2017 Fillauer LLC M058/06-29-12/08-16-17