Solvent Resistant Stirred Cells for Ultrafiltration and Filtration Applications

Similar documents
Stirred Ultrafiltration Cells

Stirred Ultrafiltration Cells Models 8003, 8010, 8050, 8200, User Guide

Stainless Steel 142 mm Filter Holder YY

User Guide. Amicon Stirred Cells. Cat. No. UFSC05001, 50 ml UFSC20001, 200 ml UFSC40001, 400 ml

Model 2000 High-performance Ultrafiltration Cell

OVERVIEW METCELL SPECIFICATIONS. Lab scale nanofiltration equipment The METcell Range

Dispensing Pressure Vessels

Multi-Probe Sampling System User Guide & Manual

Installation, Operation, and Maintenance Manual

Geotech 1.66 Reclaimer

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

Accu-Tab Systems 1000 Series by Axiall Corporation

Instruction Manual Updated 7/26/2011 Ver. 2.2

AIR/OVER HYDRAULIC JACK 20 TON

CF-SERIES HIGH FLOW REGULATORS

AND OPERATION INSTRUCTIONS

Accu-Tab Systems 2000 P Series by Axiall Corporation

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER

Hydraulic Punch Drivers

HAYWARD FLOW CONTROL Series PBV Back Pressure Valve and Series RPV Pressure Relief Valve INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS

1020 Industrial Drive, Orlinda, TN fax

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

Pressure Injection Cell Operator s Manual

Instruction Manual Updated 5/26/2009 Ver. 2.0

EcoCell Vertical Electrophoresis Unit

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

Installation, Operation, and Maintenance Manual

Directions for Use. Vivaflow 50 50R

453 Series Steam Heated Vaporizing Regulator

3M Liqui-Cel EXF-4x13 and 4x28 Series Membrane Contactors

Accu-Tab PowerBase 3012AT by Axiall Corporation

Agilent 1290 Infinity Pump Head Maintenance

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

accidents which arise due to non-observance of these instructions and the safety information herein. SPECIFICATIONS

3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

SIX R (P ) AND SEVEN R (P ) POSITION CYLINDERS Service Information

Pressure Relief Valve Instruction Manual

Installation, Operation, and Maintenance Manual

User Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References

Geotech 3.5 Reclaimer Installation and Operation Manual

Constant Pressure Crude Oil Container Model CPCCP

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

BS Series Basket Strainer

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Techcon Systems TS1258 Pressure Pot

Cocoa Patio Pond with Lit Spillway

Model ASSEMBLY and OPERATING INSTRUCTIONS

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Geotech 1.66 Reclaimer Installation and Operation Manual

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

SINGLE STAGE ULTRA-HIGH PRESSURE REDUCING REGULATOR INSTALLATION AND OPERATION INSTRUCTIONS. Certified ISO 9001

3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

2 Sentry MCL Installation, Operation & Maintenance

INSTRUCTION MANUAL GAS TRAK

accidents which arise due to non-observance of these instructions and the safety information herein. SPECIFICATIONS

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

GETZ EQUIPMENT INNOVATORS PART No. 4G59678 CARBON DIOXIDE MINI PUMP FOR EXTINGUISHER SERVICE (Revised )

SSFU SUPER SPRAYFAST UNIVERSAL ADHESIVE APPLICATOR

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

Push-To-Set Intermittent Suction Unit (PTS-ISU) Instructions for Use

Integrity Testing LifeASSURE BNA045 and BNA065 Series Filter Cartridges

Model No Product Name Handpiece Only Complete System

Binks SG-2 Plus TM 2 QT. PRESSURE CUPS Model No (Standard) Model No (with OscillatingAgitator)

DynaPak Gas Sampler S Y S T E M S U P P O R T M A N U A L DP-2000

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064

Accu-Tab PowerBase 3070AT by Axiall Corporation

Assembly and Installation Procedures

30T A/Manual Hydraulic Shop Press

CP-D/CP-M Circular Polarizer for Phoroptor. User s Guide

DuraSite 500(1) Portable Sample Vessel User's Guide. Version (1) = 90 Duro Wetted O-Rings

Rev. 1. Cryogenics Thermoflasks. 2122, 2123, 2124, 2129, and 2130

TITAN FLOW CONTROL, INC.

G7S Hand Pump Owner s Manual

DynaPak 2000L System Support Manual Version

Welker Constant Pressure Cylinders With Welker Solid Indicator, Syringe T-Handle, or Vortex Mixer

CHEMICAL INDUCTION KIT (Pt.No Issue 6, October 2015)

APCO ASU-SCAV & ASU-CAV SINGLE BODY COMBINATION AIR VALVES

STERILIZABLE DISSOLVED OXYGEN SENSORS

Quick Start Guide LIQ-QSG-499ACL-03, Rev I July Rosemount 499ACL-03. Monochloramine Sensor

3M Liqui-Cel EXF-10x28 Series Membrane Contactor

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

Sterile Visual Flow Indicator

HAYWARD FLOW CONTROL FLV SERIES, PVC/CPVC FILTER VESSEL INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

RG2200 BC Integrated Octopus

Neotecha Tank Sampling Sapro

A Quick Start Guide for the ER 4123D CW-Resonator

ROPV R40 E Series User Manual

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Hygienic Bubble Trap Installation and Operations Manual

Acid- Rite ph Adjustment System 2500 by Axiall, a Westlake Company

DuraSite 3000 Sample Vessel User's Guide. Version Rev-1

Transcription:

Solvent Resistant Stirred Cells for Ultrafiltration and Filtration Applications Cat. No. XFUF 047 01 (47 mm) Cat. No. XFUF 076 01 (76 mm) User Guide

Notice The information in this document is subject to change without notice and should not be construed as a commitment by Millipore Corporation. Millipore Corporation assumes no responsibility for any errors that may appear in this document. This manual is believed to be complete and accurate at the time of publication. In no event shall Millipore Corporation be liable for incidental or consequential damages in connection with or arising from the use of this manual. Cat. No. EU424/U, Rev. A, 09/15/2002 1994, 2002 MILLIPORE CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA. ALL RIGHTS RESERVED. THIS BOOK OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHERS. Millipore, Amicon, and Biomax are registered trademarks of Millipore Corporation. Vaseline is a registered trademark of Chesebrough-Pond s, Inc. Ultracel is a trademark of Millipore Corporation.

Solvent Resistant Stirred Cells User Guide 1 Introduction The Millipore Solvent Resistant Stirred Cell is constructed from stainless steel and borosilicate glass. It is designed to offer a rapid and efficient method for concentrating or desalting samples of up to 300 ml under high pressure. Principle A solution is held over an ultrafiltration membrane in the pressurized cell. Solvent and solutes less than the membrane s molecular weight cutoff pass through the membrane as filtrate; solutes greater than the molecular weight cutoff are retained and concentrated within the cell. Applications Use the Millipore Stirred Cell for the following applications: Removal or concentration of macrosolutes Removal or exchange of buffers, solvents, and microsolutes Features The Millipore Solvent Resistant Stirred Cell offers the following features: Minimum number of components for simple assembly and easy cleaning Borosilicate glass cylinder and a high-quality (316L) stainless steel base and top plate for resistance to high temperatures and corrosion (See the section of this guide titled Resistance to Chemical Agents. ) Wide, heavy base to ensure cell stability Safety valve, set to a nominal 6-bar (90 psi) pressure limit Shallow membrane holder for easier membrane insertion and removal Availability of a full range of membranes with different molecular weight cutoffs, including cellulosic and polysulfone membranes (See the section of this guide titled Membranes Available from Millipore. ) Negligible filtrate holdup volume (1 ml), because the membrane support structure provides an efficient drainage path for filtrate Removable PTFE-coated stirring bar shaped to maximize filtration efficiency 1

2 www.millipore.com Exploded View Solvent Resistant Stirred Cell A Filler/pressure relief cap and O ring H O-ring seal B Safety valve and O ring I Stirring bar C Glass cylinder J Stirring bar grip, grip D Funnel screw and washer E Bottom plate K Pressure inlet port F Filtrate port L Pressure tubing G Filtrate tube M Cross knob 2

Solvent Resistant Stirred Cells User Guide 3 Unpacking To unpack the Solvent Resistant Stirred Cell, take the following steps. 1. Remove the contents from the packaging and place the cell on a clear, level surface (retain the packaging for future transport). 2. Unscrew the three cross knobs from the support pillars and lift off the top plate assembly and the O- ring. 3. Remove the glass cylinder and O-ring from the base. Operating Guidelines Use Millipore membranes. They are manufactured to exacting industrial standards. Follow the instructions supplied with the membranes for cleaning and storage, as well as for buffer or solvent compatibility. Prefilter or centrifuge any solution containing particulate matter that may interfere with filtration, such as cell debris or precipitates. Pre-dilute the solution when a viscous agent (such as sucrose or glycerine) is to be removed, because highly viscous solutions filter slowly. To maximize the recovery of retained substances, continue stirring for a few minutes after pressure release to resuspend the polarized layer on the membrane surface. To minimize the filtrate holdup volume, use the shortest possible length of filtrate tubing. Samples larger than 300 ml may be concentrated by connecting the pressure supply tubingto an auxiliary reservoir. The following reservoir is available from Millipore: RC800 Mini-Reservoir (Millipore Catalogue Number 6028). An auxiliary reservoir may also be used for a diafiltration operating mode. The cell is connected to a pressurized auxiliary reservoir of diafiltrate solution with the desired microsolute concentration. In this mode, the diafiltrate solution rapidly replaces the filtrate because the volume of solution in the cell, and therefore the concentration of macrosolute, remain constant. PLEASE NOTE: An auxiliary reservoir is not supplied with the Stirred Cell. Resistance to Chemical Agents The components of the Stirred Cell are resistant to most solvents. However, DO NOT USE the Millipore Stirred Cell with strong acids (ph<1), alkalies (ph>12), or oxidizing agents. The use of the Stirred Cell with certain aldehydes, ketones (such as acetone), aliphatic ethers, or aliphatic esters may reduce the life of the fluorocarbon O-rings while causing them to swell. For other solvent sensitivities, please consult a standard text or contact Intersep, Inc. CAUTION: Please refer to specific membrane operating instructions for membrane buffer or solvent and temperature compatibility (see the section of this guide titled Membranes Available from Millipore ). 3

4 www.millipore.com Assembly for Operation To assemble the Solvent Resistant Stirred Cell before use, take the following steps. 1. Insert a membrane disc into the base, shiny side up. It is recommended that you first wet the membrane with the buffer or solvent in use, in order to seat the membrane flat and evenly in the base. CAUTION: Handle the membrane disc by its edge to avoid scratching and contaminating the surface. 2. Gently position the O-ring on the membrane disc. 3. Place the glass cylinder upright on the O-ring. 4. Ensure that the stirring bar is centrally positioned in its grip. 5. Carefully position the top plate assembly on top of the glass cylinder. 6. After ensuring that the glass cylinder and two O-rings are located between the base and top plate, fully clamp the cell by screwing the cross knobs onto the three support pillars. NOTE: In order to ensure a proper seal between the cylinder and the O-ring in the lid, it is important to move the lid evenly down the cylinder, to seat the lid properly, and to turn each knob slowly and equally. To do so, clamp and release the top plate by 1/2 turns of each cross knob in sequence. 7. Make sure the filtrate port is firmly screwed into the base. Push on the filtrate tubing. 8. Unscrew and remove the filler or pressure relief cap and introduce the sample directly into the cell, using the funnel if required. NOTE: Some ultrafiltration membranes may be protected with trace amounts of glycerine or azide, Screw and Washer which may be washed through by running the Stirred Cell with a small volume of buffer or solvent immediately before use with the sample. 9. Replace the cap, screwing it on as far as it will go. Operating Procedure To operate the Solvent Resistant Stirred Cell, take the following steps. 1. Place the cell centrally on a magnetic stirring table. 2. Push the pressure-supply tubing firmly into the pressure inlet port and connect the tubing to a regulated gas pressure source. NOTE: Nitrogen is recommended. The use of compressed air may cause large ph shifts by inducing the dissolution of carbon dioxide or the oxidation of sensitive solutions. 3. Pressurize the system to 6.0 bar (90 psi). CAUTION: Do not exceed 6.0 bar. Before operating the Stirred Cell with hazardous or especially valuable materials, Millipore recommends that you check the system pressure to ensure that all components have been properly assembled. 4. Turn on the stirring table and adjust the stirring rate so that the depth of the vortex is no more than one-third that of the stirred solution level. 5. Leave the solution on the activated stirring table until the retentate has been reduced to the required volume. 4

Solvent Resistant Stirred Cells User Guide 5 Shutdown Procedure 1. Turn off the nitrogen pressure source. CAUTION: Do not attempt to unscrew the filler or pressure relief cap before turning off the gas pressure source. 2. Slowly unscrew the cap to vent pressure before disconnecting the pressure supply tubing. CAUTION: Rapid pressure loss can cause the solution to froth and overflow. 3. Turn off the stirring table. 4. Remove the cap and pipette; pour out or add solution as desired. 5. Disassemble the cell and wash all components with soapy water or acetone. Rinse thoroughly with water. Sterilization (Advised Methods) To sterilize the Solvent Resistant Stirred Cell, use any of the following methods. Method Heat: Gas: Liquid: Requirements Autoclave at 120 C for 30 minutes Ethylene oxide 70% ethanol, 50% isopropanol, 5% formalin CAUTION: Clamp the cell loosely when autoclaving. Maintenance Replace the O-rings if they show damage or wear. Check the stirring bar periodically for burrs or rough edges that could damage the membrane. CAUTION: The borosilicate glass cylinder must be replaced if cracked or chipped. Storage If the cell is not to be required for several weeks, leave it disassembled to extend the life of the O-rings. 5

6 www.millipore.com Troubleshooting If little or no filtrate is obtained: a. Make sure that the system is pressurized. b. Adjust the stirring rate to ensure the resuspension of any polarized layer on the membrane surface. c. If the solution is highly viscous because microsolutes are present, dilute the solution to increase the flow rate. d. Make sure that the shiny side of the membrane faces upward. e. Check the filtrate port for blockage. f. Check that the correct membrane type is being used. If the filtrate rate is abnormally high: a. Make sure that no part of the stirring bar touches the membrane surface. b. Check that the correct membrane type is being used. c. Check the membrane surface for lesions, scratches, or roughness. If a cell leaks: a. Check that the O-rings are properly placed. b. Make sure the cell is firmly clamped and that the cap and four cross knobs are screwed on fully. c. Check the O-rings for nicks or cuts. CAUTION: Do not use Vaseline jelly, which may damage the membrane. 6

Solvent Resistant Stirred Cells User Guide 7 Specifications The Solvent Resistant Stirred Cell has the following specifications. 47 mm 76 mm Cell capacity 75 ml 300 ml Stirred minimum volume 2.5 ml 10 ml Membrane diameter 47 mm 76 mm Effective membrane area 15 cm 2 40 cm 2 Filtrate holdup volume 0.3 ml 1 ml Dimensions Base diameter 80 mm 110 mm Height 122 mm 165 mm Pressure Limits Maximum operating pressure 6 bar (90 psi) 6 bar (90 psi) Safety valve setting 6 bar (90 psi) 6 bar (90 psi) 7

8 www.millipore.com Component Materials The components of the Stirred Cell are as follows. Component Base plate assembly Top plate assembly Glass cylinder Stirring bar grip Stirring bar coating O-ring Pressure tube Filtrate tube Funnel Material 316L stainless steel 316L stainless steel Borosilicate glass PTFE PTFE Fluorocarbon Nylon Silicone rubber HDPE 8

Solvent Resistant Stirred Cells User Guide 9 Product Ordering Information Please call your nearest Millipore office to order any of the following components or replacement parts. See the section titled Technical Assistance for information about contacting Millipore Corporation. Component Quantity Catalogue No. 47 mm 76 mm Filter or pressure relief cap and O-ring seal 1 XFUF 047 10 XFUF 076 10 Safety valve and O-ring seal 1 XFUF 0SP 01 XFUF 0SP 01 Glass cylinder 1 XFUF 047 11 XFUF 076 11 Stirring bar grip 1 XFUF 0SP 02 XFUF 0SP 02 Grip screw (M4 6) 1 XFUF 0SP 03 XFUF 0SP 03 Grip screw washer (M4) 1 XFUF 0SP 04 XFUF 0SP 04 Stirring bar 1 XFUF 047 12 XFUF 076 12 O-ring seal for glass cylinder 1 XFUF 047 13 XFUF 076 13 O-ring seal for filler or pressure relief cap 1 XFUF 047 14 XFUF 076 14 O-ring seal for safety valve 1 XFUF 0SP 05 XFUF 0SP 05 Pressure inlet port 1 XFUF 0SP 06 XFUF 0SP 06 Pressure tube 2 M XFUF 0SP 07 XFUF 0SP 07 Filtrate tube 0.5 M XFUF 0SP 08 XFUF 0SP 08 Cross knobs (M6) 3 XFUF 047 15 XFUF 076 15 Funnel 1 XFUF 047 16 XFUF 076 16 Glass cylinder holding clips 1 XFUF 0SP 09 XFUF 0SP 09 9

10 www.millipore.com Membranes Available from Millipore A full range of Millipore membranes for use with Stirred Cells is available. The following table lists catalog numbers for each membrane type. Qty. 14 mm 25 mm 44.5 mm 63.5 mm 76 mm 90 mm 150 mm Ultracel Amicon Ultrafiltration Discs, Regenerated Cellulose YM1 1,000 NMWL 100 40422 10 13312 13322 13332 13342 YM3 3,000 NMWL 100 40423 5 13351 13361 10 13412 13422 13432 13442AM YM10 10,000 NMWL 100 40424 5 13451 13461AM 10 13612 13622 13632 13642 YM30 30,000 NMWL 100 40423 5 13651 13661AM 10 13712 13722 13732 13742 5 13751 13761AM 10

Solvent Resistant Stirred Cells User Guide 11 Membranes Available from Millipore, continued Qty 25 mm 44.5 mm 47 mm 63.5 mm 76 mm 90 mm 150 mm Ultracel PL Ultrafiltration Discs, Regenerated Cellulose PLAC 1,000 NMWL 10 PLAC O25 1O PLAC 43 1O PLAC O47 1O PLAC O62 1O PLAC O76 1O 5 PLAC O90 05 PLAC 150 05 PLBC Discs, 3,000 NMWL 10 PLBC 025 1O PLBC O43 1O PLBC O47 1O PLBC O62 1O PLBC O76 1O 5 PLBC O90 1O PLBC 150 05 PLCC Discs 5,000 NMWL 10 PLCC 025 1O PLCC O43 1O PLCC O47 1O PLCC O62 1O PLCC O76 1O 5 PLCC O90 05 PLCC 150 05 PLGC Discs, 10,000 NMWL 10 PLGC 025 1O PLGC O43 1O PLGC O47 1O PLGC O62 1O PLGC O76 1O 5 PLGC O90 05 PLGC 150 05 PLTK 30,000 NMWL 10 PLTK 025 1O PLTK O43 1O PLTK O47 1O PLTK O62 1O PLTK O76 1O 5 PLTK O90 05 PLTK 150 05 PLHK 100,000 NMWL 10 PLHK 025 1O PLHK O43 1O PLHK O47 1O PLHK O62 1O PLHK O76 1O 5 PLHK O90 05 PLJK 150 05 Biomax PB Ultrafiltration Discs, Polyethersulfone PBCC 5,000 NMWL 10 PBCC 025 1O PBCC O43 1O PBCC O47 1O PBCC O62 1O PBCC O76 1O 5 PBCC O90 05 PBCC 150 05 PBTK 10,000 NMWL 10 PBTK 025 1O PBTK O43 1O PBTK O47 1O PBTK O62 1O PBTK O76 1O 5 PBTK O90 05 PBTK 150 05 PBQK 50,000 NMWL 10 PBQK 025 1O PBQK O43 1O PBQK O47 1O PBQK O62 1O PBQK O76 1O 5 PBQK O90 05 PBQK 150 05 11

12 www.millipore.com Membranes Available from Millipore, continued Qty 25 mm 44.5 mm 47 mm 63.5 mm 76 mm 90 mm 150 mm Biomax PB Ultrafiltration Discs, Polyethersulfone PBHK 100,000 NMWL 10 PBHK 025 1O PBHK O43 1O PBHK O47 1O PBHK O62 1O PBHK O76 1O 5 PBHK O90 05 PBHK 150 05 PBMK 300,000 NMWL 10 PBMK 025 1O PBMK O43 1O PBMK O47 1O PBMK O62 1O PBMK O76 1O 5 PBMK O90 05 PBMK 150 05 PBVK 500,000 NMWL 10 PBVK 025 1O PBVK O43 1O PBVK O47 1O PBVK O62 1O PBVK O76 1O 5 PBVK O90 05 PBVK 150 05 Amicon PM Ultrafiltration Discs, Polyethersulfone AMICON Discs, PM10 10,000 NMWL 10 13112 13122 13132 13142 5 13151 13161 AMICON Discs, PM30 30,000 NMWL 10 13212 13222 13232 13242 5 13251 13261 Amicon YC05 Ultrafiltration Discs, Cellulose Acetate Amicon Discs, YC05 500 NMWL 10 13012 13022 13032 13042 5 13051 13061 Maximum operating pressure for Millipore Stirred Cells is 6 bar. Polysulfone membranes have excellent chemical resistance and hydrophobicity with high flux rates and a low fouling characteristic. The regenerated cellulose membranes have outstanding chemical resistance. They can be used in the presence of aliphatic and aromatic hydrocarbons, esters, halogenated hydrocarbons, alcohols, and ethers. Hydrophilicty, high flux rates, and low fouling characteristics are maintained. 12

Solvent Resistant Stirred Cells User Guide 13 Technical Assistance For more information, contact the Millipore office nearest you. In the U.S., call 1-800-MILLIPORE (1-800-645-5476). Outside the U.S., see your Millipore catalogue for the phone number of the office nearest you or go to our web site at www.millipore.com/offices for up-to-date worldwide contact information. You can also visit the tech service page on our web site at http://www.millipore.com/techservice. Millipore Corporation is pleased to provide internet access to Material Safety Data Sheets (MSDS) for its products that contain hazardous materials. To obtain any MSDS documents that may be associated with this product, go to the MSDS page of our website at http://www.millipore.com/msds.nsf/home. Statement Regarding Compliance with the Pressure Equipment Directive, 97/23/EC Millipore Corporation certifies that this product complies with the European Pressure Equipment Directive, 97/23/EC of 29 May 1997. This product is classified under Article 3 3of thepressure Equipment Directive. It has been designed and manufactured in accordance with sound engineering practice to ensure safe use. The product is accompanied by user instructions and bears markings to permit identification of Millipore Corporation as the manufacturer or authorized representative of this product within the European Community. In compliance with Article 3 3of thepressure Equipment Directive, this product does not bear the CE mark. 13

14 www.millipore.com Standard Warranty Millipore Corporation ("Millipore") warrants its products will meet their applicable published specifications when used in accordance with their applicable instructions for a period of one year from shipment of the products. MILLIPORE MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED. THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The warranty provided herein and the data, specifications and descriptions of Millipore products appearing in Millipore's published catalogues and product literature may not be altered except by express written agreement signed by an officer of Millipore. Representations, oral or written, which are inconsistent with this warranty or such publications are not authorized and if given, should not be relied upon. In the event of a breach of the foregoing warranty, Millipore's sole obligation shall be to repair or replace, at its option, the applicable product or part thereof, provided the customer notifies Millipore promptly of any such breach. If after exercising reasonable efforts, Millipore is unable to repair or replace the product or part, then Millipore shall refund to the customer all monies paid for such applicable product or part. MIL- LIPORE SHALL NOT BE LIABLE FOR CONSEQUENTIAL, INCIDENTAL, SPECIAL OR ANY OTHER INDIRECT DAMAGES RESULTING FROM ECONOMIC LOSS OR PROP- ERTY DAMAGE SUSTAINED BY ANY CUSTOMER FROM THE USE OF ITS PRODUCTS. 14