Determination of Zn using Atomic Absorption with Multiple Standard Additions

Similar documents
Determination of Zn using Atomic Absorption with Multiple Standard Additions

Determination of Sodium using Atomic Emission

Ultra-Low Copper Concentrations Determined by Rotating Disk Electrode Stripping Voltammetry

University of New Haven DEPARTMENT OF CHEMISTRY and CHEMICAL ENGINEERING

PerkinElmer Aanalyst100

VOLUMETRIC TECHNIQUES

CHEMISTRY FACULTY LABORATORY SAFETY CONTRACT CENTRAL CAMPUS

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Lab Equipment ANALYTICAL BALANCE

Lab: Safety Due: Lab # 0

Gas Laws. Figure 1: Experimental Set-up with Leveling Bulb. GCC CHM 151LL: Gas Laws GCC, 2019 page 1 of 8

SOP for use of the SX19 Stopped-Flow Instrument

Check-In, Lab Safety, Balance and Volumetric Glassware Use, Introduction to Statistics, and MSDS Familiarization

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Student Information & Laboratory Safety Contract

Any laboratory is equipped with specific tools, equipment,

Gas Laws. 2 HCl(aq) + CaCO 3 (s) H 2 O(l) + CO 2 (g) + CaCl 2 (aq) HCl(aq) + NaHCO 3 (s) H 2 O(l) + CO 2 (g) + NaCl(aq)

The use of the analytical balance, and the buret.

Armfield Distillation Column Operation Guidelines

Safety Rules for the Science Lab

Adaptations of Desert Plants

Experiment #2. Density and Measurements

Flinn Scientific's Student Safety Contract

The Ideal Gas Constant

1. Determining Solution Concentration

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

Standard Operating Procedure (SOP)

SOLUBILITY OF A SOLID IN WATER

Measuring Mass and Volume

Vapor Pressure of Liquids

Physics Experiment 17 Ideal Gas Law Qualitative Study

Gas Laws. 2 HCl(aq) + CaCO 3 (s) H 2 O(l) + CO 2 (g) + CaCl 2 (aq) HCl(aq) + NaHCO 3 (s) H 2 O(l) + CO 2 (g) + NaCl(aq)

longitudinal wave Figure 1 amplitude transverse wave Figure 2

Wolfson Campus Natural Science Department. Biology Laboratory. Student Safety Contract

Safety rules must be followed

Standard Operating Procedure Porosity Analyzer - The Micromeritics Tri-Star II (surface area) with Micrometrics VacPrep 061 (Sample Degas System)

CHM250 Calibration and Measurement Lab. Balance Calibration

LABORATORY INVESTIGATION

EXPERIMENT 1 BASIC LABORATORY TECHNIQUES AND TREATMENT OF DATA MEASUREMENTS

Take Home Test Score:

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials

Standard Operating Procedure (SOP)

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

SOP: Buck Scientific BLC-20P HPLC Operation

Guide to Safety in the Science Classroom

CHM 2045L Physical Properties

Equipment Operating Procedure Glove Box

Salt Lowers the Freezing Point of Water

SOP Standard Operating Procedure. Etching All Materials

Location and Use of Safety Equipment. Extinguisher Shower Fire Blanket

Vapor Pressure of Liquids

The Gas Laws: Boyle's Law and Charles Law

UNH Department of Forensic Sciences

BASIC RULES OF USAGE, OLIS DSM-20 CD Management The instrument is currently managed by the Analytical Biophysics Core Facility.

SCIENCE LABORATORY SAFETY INSTRUCTIONS & CONTRACT

SOLUBILITY OF A SOLID IN WATER

Katy Independent School District Science Safety Rules

EXPERIMENT 2. Laboratory Procedures INTRODUCTION

Eric Sheagley, Lab Supervisor Fall, 2015

Density and Stress in Plastics Mary V. McCrary

acrolein, acetaldehyde and acetone( cm -1 ); methanol (1306 cm -1 ); ethylene (949 cm -1 ); and isoprene (893 cm -1 ).

BASIC LABORATORY TECHNIQUES (Revised )

How Environment-friendly Is Your Home?

Using the Akta Prime plus October 22, 2012

Cover Page for Lab Report Group Portion. Head Losses in Pipes

UNIVERSITY OF HOUSTON-CLEAR LAKE ENVIRONMENTAL, HEALTH AND SAFETY DEPARTMENT

CHM 100 / Introductory Laboratory Experiment (r10) 1/11

MINIMUM SAFETY REGULATIONS

FLUORESCENCE DETERMINATION OF OXYGEN

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Hands-On Experiment Density and Measurement

General Safety Rules

Variable Temperature Operation

Roy G. Biv Charles W. McLaughlin

PRE LABORATORY ASSIGNMENT: Lab Section Score: /10 READ THE LAB TEXT BEFORE ATTEMPTING THESE PROBLEMS!

TEMPERATURE S RELATIONSHIP TO GAS & VAPOR PRESSURE

High School Safety Acknowledgment Form

RUN TIME: 50 MIN PL02 Page 1 of 7 POGIL LAB EXERCISE 02 What Makes the Lab a Safe and Effective Place to Do Chemistry?

SCIENCE EXPERIMENTS ON FILE Revised Edition Tornado Model. Safety Please click on the safety icon to view safety precautions.

Safety & Class Contract

Blood Spatter Inquiry

CHM Introductory Laboratory Experiment (r17sd) 1/13

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

LABORATORY TECHNIQUES. Pouring Liquids

In case of emergency or accident: 1. Call the instructor immediately. Describe the nature of the accident or injury.

EXPERIMENT 12 GAS LAWS ( BOYLE S AND GAY-LUSSAC S LAW)

BASIC LABORATORY TECHNIQUES (Revised )

Molecular Materials Research Center Beckman Institute California Institute of Technology

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Vapor Pressure of Liquids

Department of Chemistry

Boyle s Law: Pressure-Volume Relationship in Gases

Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice. Christine Slater PhD

SAFETY ORIENTATION. University of Evansville Department of Chemistry

Laboratory Safety Rules Holt High School (from )

Read over Techniques #2, 4, 5, 6, and 9 in the Demonstrations of Nine Practical Lab Techniques booklet.

Natural Reaction Boosters C. V. Krishnan

Transcription:

Determination of Zn using Atomic Absorption with Multiple Standard Additions 1. Purpose This procedure will determine the concentration of zinc at the parts-per-million level using flame atomic absorption and the multiple standard addition quantitation method. 2. Background Flame atomic absorption (FAA) spectroscopy is a relatively common technique used primarily for the determination of metals dissolved in solution. Like other forms of absorption spectroscopy, a lamp is used with a monochromator to provide a specific wavelength of light. This light is passed through a sample chamber containing the analyte. The greater the concentration of analyte, the more light it absorbs as predicted by Beer s Law (A=abc). Thus, absorbance is directly related to the concentration of the analyte. In FAA, the source of light is a hollow cathode lamp (HCL) made of the same element as the analyte of interest, thus providing spectral energy that is exactly tuned to the specific wavelength(s) absorbed by the analyte in its atomic state. The sample chamber is not a cuvette, but instead is the slot burner head of the instrument. The HCL light passes directly above the slit of the burner head and is absorbed by the atoms made in the flame. The flame s energy is what desolvates, vaporizes, and then atomizes the analyte, necessary if one is to do atomic spectroscopy. In this laboratory, you will be given a sample container containing a zinc solution. You will dilute the solution to 250-mL to prepare your zinc unknown solution. You will also prepare a Zn standard with a concentration of about 2-ppm and then prepare a series of standard addition solutions in which increasing aliquots of your 2-ppm standard are added to aliquots of your unknown to create a series of spiked unknown solutions. You will measure the absorbance of the resulting five solutions to construct a multiple standard addition calibration curve. Using the least-squares analysis program in Excel, you will determine the concentration of the diluted unknown and then use that value to determine the Zn concentration in the original 250-mL unknown solution. 3. Materials and Equipment Zn standard metal Con HCl Assorted glassware Varian Atomic Absorption Spectrophotometer 4. Safety Protective eyewear must be worn at all times. Concentrated hydrochloric acid (con HCl) is very caustic and must be handled with care. Notify your instructor of any spills and wash immediately should you come in contact with the HCl. AA_3.4.docx Page 1 9/28/2016

The internal parts of the FAA are extremely hot and will cause severe burns on contact. Do not touch the burner head assembly for any reason. Notify your instructor if there is any problem with the instrument. 5. Experimental Method 5.1 Preparation of the standard Using a small weighing boat, accurately weigh sufficient pure granular zinc to produce a 500.0- ml of solution with a concentration of about 1000-ppm zinc. While the solution does not have to be exactly 1000.0-ppm, the actual concentration should be calculated and recorded to four significant figures. Using a wash bottle, carefully transfer the granular zinc to a 250 ml beaker with ~5-10 ml of nanopure water. In the fume hood, add about 10 ml con HCl using a small graduated cylinder and place a watch glass on top of the beaker containing the zinc and HCl. There should be sufficient water from the transfer to dilute the acid enough so it does not react too fast. Place the beaker on a hot plate and heat until the metal completely dissolves; there will be no more bubbling from the oxidation of the Zn. Remove the beaker and allow it to cool for a few minutes and then quantitatively transfer the solution to a 500-mL volumetric flask. Dilute to the mark, mix, and label this solution the "Zn stock standard solution. The stock standard must be diluted (accurately) to about 2-ppm concentration. It is too concentrated to do this accurately in a single step and so a serial dilution must be performed resulting in one or more intermediate solutions. When the final dilution has been made to prepare the 2-ppm standard solution, label it Zn working standard. 5.2 Preparation of the unknown Obtain a zinc unknown from the instructor and quantitatively transfer the sample to a 1-L volumetric flask, dilute to volume and label it "Zn unknown solution. 5.3 Preparation of five unknown solutions spiked with the 2-ppm working standard Prepare the following five solutions each in a 100-mL volumetric flask (transferred to small plastic bottles as necessary). Each of the solutions will contain 10.00-mL of the unknown Zn solution along with increasing aliquots (0 20 ml) of the 2-ppm working standard. Dilute each volumetric flask to the mark with nanopure water and mix thoroughly. Record your data in tabular form as below, including two additional columns labeled ppm Standard and Abs. Solution aliquot zinc working standard aliquot unknown zinc 1 0 ml 10.00 ml 2 5.000 ml 10.00 ml 3 10.00 ml 10.00 ml 4 15.00 ml 10.00 ml 5 20.00 ml 10.00 ml AA_3.4.docx Page 2 9/28/2016

Take the following to the instrument room and find the instrument: 1. All 5 standard addition solutions 2. Large beaker of nanopure water to use as a blank 3. Kim wipes (may be already located near the instrument) 4. Notebook 5.4 Analysis using the Varian SpectrAA 200 spectrometer Turn on the instrument by pressing the red switch on the lower left corner (Figure 1& 2a). Place the capillary tube into a large beaker filled with nanopure water (Figure 2b). Figure 1: Varian SpectrAA 200 Atomic Absorption Spectrophotometer Figure 2a: Red power switch Figure 2b: Capillary tube inserted into a beaker of water AA_3.4.docx Page 3 9/28/2016

Open the valves for the acetylene gas and compressed air. The acetylene gas cylinder is located to the far left of the instrument. Turn the gas on by turning the flower-shaped main valve (only!) at the top of the cylinder counter-clockwise (see Figure 3). Do not touch the pressure adjustment valve below the pressure gages (see the instructor if you do). Figure 3 Acetylene gas cylinder and regulator Figure 4 Compressed air valve The compressed air valve (which may already be open) is located on the wall behind the instrument. The valve is open when handle parallels the tubing (Figure 4). Turn on the computer, if it has not already been running. The computer should automatically start Varian SpectrAA 100/200 application upon startup. If the instrument was already on, open the Varian SpectraAA 100/200 shortcut on the desktop. When the program is loaded, you will see the main menu as shown in Figure 5. Figure 5: Varian SpectrAA application If you haven t done so already, turn on the instrument by lipping the switch on the lower left corner of the instrument (Figure7, next page). Once you have turned on the instrument, you will get a warning on the computer. Ignore the message and click "OK". Ignite the lame by the following procedure. AA_3.4.docx Page 4 9/28/2016

First, conirm that the capillary tube is in a beaker containing nanopure water. Switch the gas control knob (located front right of the instrument see Figure 6) from "OFF" to "AIR" setting. As the compressed air exits the burner head, it will make a hissing sound as soon as you switch the knob to air setting. Figure 6: Gas (air) and acetylene flow control Figure 7: Varian SpectrAA Power switch and ignite button Next, slowly turn the acetylene low knob (Figure 6) counter-clockwise. Simultaneously, press and hold the black ignite button (Figure 7) located next to the main power switch to ignite the lame on the burner. During this time, the gas low will be much louder due to the lame-starting assembly. Once the lame has been ignited, adjust the acetylene low so that it is suficient to maintain the lame (approximately 1 ml/s). The lame is visible through the window above the sample introduction port (Figure 7). It should appear green in color through the protective shield. If you open the shield, the lame should be blue and only be a couple millimeters in height ( fuel-lean, as opposed to fuel-rich which typically is much higher with yellow AA_3.4.docx Page 5 9/28/2016

Next, on the computer, click "Worksheet" button on SpectrAA 100/200 application (Figure 8). Figure 8 Click "New From Template" button. You should see window with available template. Choose "Zinc by AA Chm466" and click "OK" (Figures 9 and 10). Figure 9: Load Worksheet select "New From Template" Figure 10: Select " Zinc by AA Chm466" Next, you will be prompted to name your worksheet. Enter the name of the ile in the Name box. Record this name in your lab notebook. Enter your name in Analyst box and click "OK" (Figures 11 and 12). Figure 11 Figure 12: Name your worksheet, and enter your name At this point, the computer should show the screen in Figure 13 on the next page. AA_3.4.docx Page 6 9/28/2016

Figure 13: Worksheet for data collection. You are now ready to begin the analysis. Click the "Read" button on the left side to start. You should see a dialog box named "Analysis Checklist". Verify that these settings are correct and document the slit width, lamp position, and fuel rate in your notebook. Ask your instructor for help checking the slit width (the lamp and slit width adjustment lever are beneath the door to the left of the lame. Click "OK" to continue. Wait for a few seconds while the instrument will make necessary adjustments to itself. Once the instrument is ready, you can blank the instrument with nanopure water. Remember to wipe off the capillary with a Kim wipe between all readings. Next, the instrument will ask for sample #1. Remove the blank, place the capillary in sample #1 and click "OK". The instrument will acquire absorbance data from the sample solution for approximately 15 seconds. You can watch the progress at the top right of the program screen. Do not remove the capillary tubing until the data acquisition is complete. Once it is done reading for 15 seconds, remove the capillary tube from the solution, wipe the tube, then place the capillary tube in the nanopure water. After the instrument has taken the measurement for sample #1, on the computer move the cursor to the cell for sample #2 and put the capillary in sample #2. Click "Read" button and repeat the same process for sample #2. It is important that the cursor be on the cell of the sample to be read. Otherwise, the data will not be recorded with the correct sample. Continue the process above for all of the standard solutions. Record the absorption values in your notebook. Once you are done, make sure to put the capillary tube in nanopure water. When all of the solutions have been read, click "Stop" button to stop the analysis. Ensure that all of the absorbance values are documented in your notebook. It is important that both your solutions are all analyzed on the same day. The instrument response can change day to day; therefore, running all solutions at the same time will produce the most accurate result. Print out the data by the following procedure. Exit the worksheet: Go to the "Exit" menu and select "Return to Main Index". In the Main Index, click "Report" button. Refer to Figure 8 for a screen shot of the Main Index. AA_3.4.docx Page 7 9/28/2016

Figure 14: Select your worksheet Figure 15: Print your report Select your report ile (Figure 14). Click the "4. Report" tab. Click "Print Preview " button to preview the report (Figure 15). If the preview looks satisfactory, print your result by clicking the "Print " button on the right-hand side. When your data has printed, click the close button. Double-check that the data print-out and the values in your notebook exactly agree. Exit the SpectrAA application. Shut off the acetylene gas: Turn the acetylene knob on the instrument clockwise. Close the main valve on the acetylene tank. Turn off the compressed air: Switch the gas control knob to the off position. Turn off the inhouse compressed air. Turn off the instrument power. 6. Data Analysis/Calculations Each of the five solutions contains the same amount of unknown but different amounts of the 2- ppm standard. Thus, the absorbance of Solution 1 is due entirely to the Zn in the diluted unknown while the signals from Solutions 2-5 are due to the combination of the unknown and the standard spikes. To determine the concentration of the diluted unknown, use the Excel least-squares software program (available on the balance room computers as well as the computers in the Chemistry Computer lab). Open the program and enter your data for the five solutions, using the ppm of Zn from the standard in each of the five solutions (e.g., Solution 1 will have a value of 0 ppm). At that point, run the least-squares function and enter your name as prompted. Enter 0 for the unknown value. This graph will allow you to determine the concentration of the diluted unknown (though it will be a negative value since this is multiple standard addition). In your notebook, note the quantitation method being used. Use this value to determine the concentration of the unknown in the original 250-mL volumetric flask to two decimal places. AA_3.4.docx Page 8 9/28/2016

Instrumental Analysis Laboratory Safety Rules A. Instructions: Carry out all manipulations in accordance with instructions and the safety rules and procedures given herein. B. Eye Protection: All students and staff working in the laboratory must wear safety glasses at all times. If a student needs to be reminded more than three times to wear goggles, she/he will be dismissed from lab for the remainder of the day, and will not be given an opportunity to make up the work. C. Apparel: The clothes you wear in lab are an important part of your safety equipment, and should offer protection from splashes/spills. Closed toed shoes (sneakers are fine), Full-length pants or a full-length skirt, and A shirt that completely covers your torso (i.e. at minimum, a t-shirt). In other words, you must NOT wear shorts to lab. You must NOT wear flip-flops, sandals, or crocs. You must NOT wear tank tops, halter tops, spaghetti-strap tops, or low cut jeans to lab. Exposed abdomens, hips, and backs are not safe in the lab. D. Gloves: Gloves are an important part of personal protection. Gloves will be available at all times in the laboratory. Your instructor will require their use when appropriate. E. Food: Food, drinks, and gum are not allowed in lab. None at all, not even water bottles. F. Sanitation Issues: Be sure to wash your hands before leaving lab, before you eat anything outside of lab, and before you answer your cell phone. G. Music: Individual headphones are not allowed. Your may choose to play music for the entire class. H. Cell Phones and Other Electronic Devices: Cellular phones and other electronic devices that you do not need to perform your laboratory work should be put away. I. Other: All students are explicitly prohibited from: 1. conducting any unauthorized experiments. 2. removing chemicals or apparatus from the laboratory for any reason. 3. working in the lab alone, or at other than regularly scheduled lab periods. 4. smoking in the laboratory or within 20 feet of any doorway. 5. impeding movement in aisles or through doorways with bags, skateboards, etc.