INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Similar documents
AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Introductory Lab: Vacuum Methods

TELEDYNE HASTINGS INSTRUMENTS

TELEDYNE HASTINGS INSTRUMENTS

Design Features. General Description

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Exercise 2-2. Second-Order Interacting Processes EXERCISE OBJECTIVE DISCUSSION OUTLINE. The actual setup DISCUSSION

Model 601CV. Mass Flow Controller. Product Features:

Mass Flow Meter (MFM) for Gases

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Mass Flow Controller (MFC) for Gases

The Discussion of this exercise covers the following points:

Micro Channel Recuperator for a Reverse Brayton Cycle Cryocooler

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

The Discussion of this exercise covers the following points:

living with the lab control of salinity 2012 David Hall

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

Lab #1 Pressure: Bubblers and Water Balloons CEE 331 Fall 2003

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

control of salinity Thanks to David Hall, from LWTL

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

Mass Flow Controller (MFC) for Gases

800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

The Ins and Outs of I/P Transducers

living with the lab control of salinity 2012 David Hall

CHARACTERISTICS OF LEAK DETECTION BASED ON DIFERENTIAL PRESSURE MEASUREMENT

The Experts in Vacuum Solutions

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

UNDERSTANDING, ACCELERATED

Instructions for SMV 3000 Multivariable Configuration (MC) Data Sheets

WP2 Fire test for toxicity of fire effluents

Experiment 8: Minor Losses

Pressure Measurements

Mass Flowmeters. Mass Flowmeters for Gases

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter

TC Series FOR THE MASS FLOW CONTROL OF ALL KIND OF GASES THERMAL MASS FLOW CONTROLLER OUTLINE FEATURES OPERATION PRINCIPLE

Aera Industrial MFC Product Guide. Innovative Control at Work

SPD Pressure Sensor Families

Module 2, Add on Lesson Depth Sensor. Teacher. 90 minutes

Best Practice Guide, Servomex 2700

BS12N Piezoresistive OEM Pressure Sensor

FPG8601 Force Balanced Piston Gauge

Mass Flow Meter (MFM) for gases

Anemometry. Anemometry. Wind Conventions and Characteristics. Anemometry. Wind Variability. Anemometry. Function of an anemometer:

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

CONTROL and INSTRUMENTATION

Mass Flow Controller (MFC) for Gases

FUNDAMENTALS OF PRESSURE REGULATORS ROBERT BENNETT MANAGER OF TRAINING ELSTER AMERICAN METER

HumiSys HF High Flow RH Generator

Digital Mass Flow Meters and Controllers for Gases

THE IMPLEMENTATION OF TOROIDAL THROAT VENTURI NOZZLES TO MAXIMIZE PRECISION IN GAS FLOW TRANSFER STANDARD APPLICATIONS

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

CORESTA RECOMMENDED METHOD N 6

TELEDYNE HASTINGS INSTRUMENTS PRELIMINARY BULLETIN. Models HFM-D-301 & 305, HFC-D-303 & 307

EASIDEW TRANSMITTER with Current Source Output

BS12 Piezoresistive OEM Pressure Sensor

Characterizers for control loops

TC Series FOR THE MASS FLOW CONTROL OF ALL KIND OF GASES THERMAL MASS FLOW CONTROLLER GENERAL FEATURES OPERATION PRINCIPLE

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

THE INSIDE STORY OF THE F-SERIES:

Cover Page for Lab Report Group Portion. Pump Performance

Multiphase MixMeter. MixMeter specification. System Specification. Introduction

Gerald D. Anderson. Education Technical Specialist

How to specify a product. Process Sensors and Mechanical Instruments

AMS 2710 PCB pressure sensor module with V output

Instruction and Maintenance Manual

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

Expert Hydrostatic Level Transmitters

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Auto-Zero Calibration Technique for Pressure Sensors

Truck-mounted Mass Flow Metering for LPG Delivery

Succeeding with Production Air Leak Testing Methods. Paul Chamberlain President, CEO

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Instrumentation & Data Acquisition Systems

Best Practice for Calibrating LTH Conductivity Instruments

Electronic Volume Correctors

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Generating Calibration Gas Standards

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate.

A centrifugal pump consists of an impeller attached to and rotating with the shaft and a casing that encloses the impeller.

Mass Flow Controllers

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

Spring Loaded Regulators Constant Loaded Regulators Twin Parallel Flow Regulators Field Service Regulators

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

The Estimation Of Compressor Performance Using A Theoretical Analysis Of The Gas Flow Through the Muffler Combined With Valve Motion

Transcription:

TELEDYNE HASTINGS TECHNICAL PAPERS INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE Proceedings of FEDSM 98 1998 ASME Fluids Engineering Division Summer Meeting June 21-25, 1998, Washington, DC FEDSM 98-5292 ABSTRACT A new sensor is presented for measuring the mass flow of gases. The sensor has an inherently linear response and minimal pressure drop, and responds to changes in flow rate in less than 300 milli-seconds. The sensitivity of the sensor to changes in the ambient temperature and other sources of error has been minimized. It provides a linear output from zero to 25 standard cm 3 / min (SCCM) 1 of nitrogen, which is 200% to 500% greater than the operating range of previous thermal mass flow sensors. Detailed description of the sensor design, theoretical discussion of the measurement method, and performance data are presented. NOMENCLATURE C P = specific heat at constant pressure C v = specific heat at constant volume P = electrical power supplied to either heater P d = electrical power supplied to downstream heater P u = electrical power supplied to upstream heater R 1 = value of fixed resistor in series with either heater T ambient = ambient gas temperature T heater = heater temperature T m = mean gas temperature T m,mid = mean gas temperature at axial midpoint of tube T m,1 = mean gas temperature at tube inlet T m,2 = mean gas temperature at tube outlet SCCM = standard cubic centimeters per minute 1 Note that the mass flow rate is expressed in SCCM by industry convention. This value expresses the mass flow rate as its equivalent volumetric flow rate at standard conditions of 0 o C and 101.325 kpa. The actual mass flow rate in SI units is obtained by multiplying the SCCM value by the gas density at standard conditions. Lawrence W. Eget and William. J. Alvesteffer Teledyne Brown Engineering: Hastings Instruments, P. O. Box 1436, Hampton, Virginia 23661 USA Phone: (757) 723-6531 Fax: (757) 723-3925 V 1 V 2 m p q conv q i,d q,i,u q t,d q t,u v x = voltage at top of either bridge = voltage across either resistive heater coil = mass flow rate of the gas = pressure of the gas = rate of convective heat transfer = rate of conduction by downstream insulation = rate of conduction by upstream insulation = rate of conduction by tubing at downstream = rate of conduction by tubing at upstream = volume of the gas = distance along the tube INTRODUCTION Many methods for measuring gas flow are reviewed by Goldstein (1983). These measurement techniques can be classified either as volumetric flow measurements or mass flow measurements. Volumetric flow measurements are often undesirable and inaccurate because the volume of a gas is directly proportional to the absolute gas temperature, and inversely proportional to the absolute pressure applied to the gas volume. Consequently, when a given number of molecules of a particular gas is passed through a volumetric flow meter at different operating conditions, the meter will indicate different flow rates. The output of a volumetric flow meter can be corrected by taking into account the temperature and pressure of the metered gas, but these additional measurements increase the complexity and cost of the volumetric flow meter. Mass flow measurements, however, are substantially independent of variations in the temperature and pressure of the gas stream. Of the available methods for measuring the mass flow rate of a gas, thermal transfer sensors are the most common. These sensors all have some means of heating th

gas flow, since heat transferred to a flowing fluid is known to be proportional to the mass flow rate of that fluid. Conventional sensors will then obtain some temperature measurement which is indicative of the heat transfer rate, and is thus characteristic of the mass flow rate. One common measurement approach is to maintain one aspect of the heater, such as the heater power, voltage, or current constant. The output from this approach is based on the temperature change of the gas at points upstream and downstream from the heater. Other conventional approaches utilize coils of temperature sensitive resistance wire which serve as combination heating elements and temperature monitors. The resistance of the wire will increase with an increase in temperature, which again is used to indicate the heat transfer rate, which is directly proportional to the mass flow rate. Since these traditional sensing methods all rely on temperature measurements to indicate the mass flow rate, they introduce new temperature dependent errors in the mass flow signal. Changes in the temperature of the operating environment or in the ambient temperature of the sensed gas typically alter the conventional flow meter output by 0.05% to 0.1% per degree Celsius. This is a significant error considering the thermal mass flow meter commonly achieves 1% to 2% full scale error when run at stable temperature conditions. It is easy to see that the conventional thermal mass flow sensor, while more desirable than volumetric flow sensors, does have limitations. The driving motivation for the work presented in this paper is thus to provide a new type of thermal mass flow sensor which can overcome the limitations of previous sensors. SENSOR DESCRIPTION A cross section of the sensor is shown in Fig. (1). The sensor consists of two coils of resistance wire with a high temperature coefficient of resistance (3500 ppm/ o C) wound around a stainless steel tube with internal diameter of 0.6604 mm and 7.62 cm length. Each coil is 1.372 cm in length, and they are separated by 1.27 mm distance. These two identical resistance wire coils are used to heat the gas stream, and are symmetrically located upstream and downstream on the sensor tube. Insulation surrounds the sensor tube and heater coils with no voids around the tube to prevent any convection losses. The ends of this sensor tube pass through an aluminum block and into the stainless steel sensor base. This aluminum block thermally shorts the ends of the sensor tube and maintains them at ambient temperature. There are two coils of resistance wire that are wound around the aluminum block. The coils are identical to each other, and are symmetrically spaced on the aluminum ambient block. These coils are wound from the same spool of wire that is used for the sensor heater coils so they have the same resistivity and the same temperature coefficient of resistance as the sensor heater coils. The number of turns is controlled to have a resistance that is 10 times larger than the resistance of the heater coils. Thermal grease fills any voids between the ambient temperature block and the sensor tube to ensure that the ends of the sensor tube are thermally tied to the temperature of this aluminum block. Aluminum has a very high thermal conductivity which ensures that both ends of the sensor tube and the two coils wound around the ambient block will all be at the same temperature. This block is in good thermal communication with the stainless steel sensor base to ensure that the ambient block is at the same temperature as the main instrument block, and therefore the same temperature as the incoming gas stream. This allows the coils wound on the aluminum block to sense the ambient gas temperature. Two identical Wheatstone bridges are employed, as shown in Fig. (2). Each bridge utilizes an ambient temperature sensing coil and a heater coil. The heater coil and a constant value series resistor comprise the first leg of the bridges. The second leg of each bridge contains the ambient sensing coil and two constant value series resistors. These Wheatstone bridges keep each heater temperature at a fixed value of 48 o C above the ambient sensor temperature through the application of closed loop control, and the proper selection of the constant value bridge resistors. THEORY OF OPERATION The heat convected to or from a fluid is proportional to the mass flow of that fluid. From Incropera and De Witt (1990) we know that the rate at which the heat is transferred is governed by Eq. (1): dq dx m d ( = C T + pv ) dx conv v m As in conventional thermal mass flow meters, an assumption is made that the fluid is a perfect gas. By making the perfect gas assumption and integrating over the total length of the sensor tube, Incropera and De Witt develop an expression for the total tube heat transfer rate which is: q = m C ( T T ) conv p m, 2 m, 1 Since the constant differential temperature sensor has 2 heater coils symmetrically spaced on the sensor tube, it is convenient to consider the upstream and downstream heat transfer modes separately. The electrical power supplied to (1) (2)

either of the heater coils will be converted to heat, which can be dissipated by radiation, conduction, or convection. The radiation term is negligible due to the low temperatures used by the sensor, and because the sensor construction preferentially favors the conductive and convective heat transfer modes. The thermal energy of each heater will then be dissipated by conduction down the stainless steel sensor tube, conduction to the insulating foam, plus the convection due to the mass flow of the sensed gas. This relationship between the power input to the heater and the heat dissipation mechanisms is thus shown in Eq. (3) for the upstream heater, and Eq. (4) for the downstream heater. P = q + q + m C ( T T ) u t, u i, u p m, mid m, 1 P = q + q + m C ( T T ) d t, d i, d p m, 2 m, mid Because great care is taken to wind the resistive heater coils symmetrically about the midpoint of the tube, it is assumed that the heat conducted along the sensor tube from the upstream heater will be equal to the heat conducted through the tube from the downstream heater. Similarly, the heat conducted from the upstream and downstream coils to the foam insulation surrounding them is assumed to be equal, based on the symmetry of the sensor construction. By equating these conduction terms, Eq. (3) and Eq. (4) can be combined to give: (3) (4) the tube surface temperature at the axial midpoint is very close to the operating temperature of the heater coils. The mean temperature of the gas stream is then approximately the same as the heater temperature. By approximating the mean gas temperature as the heater temperature, we can write Eq. (7): ( ) P P = 2 mc T T (7) u d p heater ambient The value of the constant pressure specific heat of a gas is virtually constant over small changes in temperature. By maintaining both heaters at the same, constant temperature difference above the ambient gas stream temperature, the difference in heater power is a function only of the mass flow rate. Fluctuations in ambient gas temperature which cause errors in conventional mass flow sensors are avoided; The resistance of the ambient sensing coil changes proportionally with the ambient temperature fluctuations, causing the closed loop control to vary the bridge voltage such that the heater resistance changes proportionally to the ambient temperature fluctuation. The power supplied to each of the 2 heater coils is easily obtained by measuring the voltage across the heater, shown as V 2 on Fig.(2), and the voltage across the fixed resistor R 1. Since R 1 is in series with the heater R H they have the same current flowing through them. The electrical power supplied to a given heater is then calculated: Pu m C p( Tm, mid Tm, 1) = Pd m C p( Tm, 2 Tm, mid) (5) P = ( V ) V V 1 2 2 R 1 (8) Since the sensor tube inlet and outlet are linked by an aluminum ambient bar, the high thermal conductivity of the bar provides a thermal short, constraining the ends of the sensor tube to be at equal surface temperature. Moreover, the tube ends and the aluminum ambient bar have intimate thermal communication with the main flow passageway prescribed by the main stainless steel flow meter body. This further constrains each end of the sensor tube to be equal to the ambient gas temperature. Incorporating these restrictions into Eq. (5), and combining like terms, Eq. (6) can be written. ( 2 2 ) Pu Pd = mcp Tm mid T (6), ambient With a constant differential temperature applied to each heater coil and no mass flow through the sensor the difference in heater power will be zero. As the mass flow rate through the sensor tube is increased, heat is transferred from the upstream heater to the gas stream. This heat loss from the heater to the gas stream will force the upstream bridge control loop to apply more power to the upstream heater so that the 48 o C constant differential temperature is maintained. The gas stream will increase in temperature due to the heat it gains from the upstream heater. This elevated gas stream temperature causes the heat transfer at the downstream heater to gain heat from the gas stream. The heat gained from the gas stream forces the downstream bridge control loop to Since the length of each heater section is nearly 21 times greater than the inside tube diameter, the mean gas temperature at the tubes axial midpoint is approximately equal to the tube surface temperature at that point. Recall that the outside of the sensor tube is well insulated from the surroundings, therefore

apply less power to the downstream heater coil in order to maintain a constant differential temperature of 48 o C. DESCRIPTION OF EXPERIMENTAL METHOD Data is presented for the accuracy, temperature coefficients, response time to a step change in flow, and pressure drop of the sensor. In all of the experiments the sensors under test were mounted on identical flow meter bodies, through which the main gas flow was passed. The accuracy test were conducted by passing dry nitrogen through a pressure regulator, and into a flow control valve, which was used to vary the mass flow rate of the gas. From the valve, the nitrogen passed through a flow reference, and then into the units under test. The reference used was a commercially available calibration instrument which indicates the mass flow rate with ±0.2% accuracy. All of the instruments under test were configured such that all of the gas flow passed through the sensor tube. Tests of the sensor temperature coefficient were conducted using a computer controlled environmental chamber. The sensors under test were placed inside the environmental chamber, while the flow reference was located outside the chamber where it was unexposed to thermal changes. Leak free gas lines passed through an access port in the chamber wall, connecting the flow reference to the test sensors. These gas lines were made of copper with sufficient length such that the test gas achieved the temperature of the environmental chamber before entering the test sensors. The chamber was cycled from 56 to -10 degrees Celsius while maintaining the humidity level as low as possible. The sensors were operated at 20% of their calibrated full scale for 1.5 hours after each step change in chamber temperature. After this temperature stabilization time the mass flow rate through the sensors was cycled from zero to full scale. Sensor output, temperature, and output of the mass flow reference were recorded after 4 minutes of stabilization time at each flow set point. Response time tests were conducted using dry nitrogen. The nitrogen was passed through a pressure regulator and into a bleed valve, which was used to adjust the gas flow rate. After the bleed valve, the gas was routed into a two position solenoid valve. The solenoid valve had minimal internal volume and was used to trigger a step change in flow by diverting the nitrogen between its 2 outlets. A sensor was connected to each outlet of the solenoid valve so that both paths had equal flow resistance. A mechanical rough pump was connected to the downstream end of the sensors to prevent the gas from compressing during the step change from the no flow to full flow condition. A calibrated oscilloscope collected the sensor outputs. Sensor pressure drop versus flow rate was measured for dry nitrogen at 22 o C. The sensors were mounted on a flow meter base which was configured so that all of the gas flow passed through the sensor. A calibration instrument which indicates the mass flow rate with ±0.2% accuracy was placed upstream of the test sensor. One side of a calibrated differential pressure gauge was connected between the flow reference and the inlet to the test sensor. The other side of the differential pressure gauge was connected to the outlet of the device under test, which was vented to atmosphere. Another calibrated pressure gauge was used to measure the atmospheric pressure. TEST RESULTS A graph of the sensor output versus mass flow rate is shown in Fig. (3). Both heaters require a power of 110 mw at zero flow, which is an output of 0 mw. With gas flow through the sensor the heater power diverges producing an output with typical sensitivity of 2.35 mw / SCCM. The output is inherently linear for flow rates up to 25 SCCM, having less than 0.75% deviation from a terminal based line, as shown in Fig. (4). In contrast, existing mass flow sensors must remain below 5 SCCM to 10 SCCM, depending on the design, in order to achieve a similar degree of linearity. Information on the thermal stability of the constant differential temperature sensor is shown in Fig. (5) and Fig. (6). Figure (5) gives the percentage change in full scale output versus ambient temperature. The maximum deviation of the full scale output from the original full scale value is equivalent to a full scale thermal coefficient of 0.03% per degree Celsius. The percentage change in the zero flow output versus the ambient temperature is shown in Fig. (6). At the zero flow condition maximum thermal coefficient of the sensor is 0.052% per degree Celsius. A graph of the typical sensor response to a step change in flow rate is shown in Fig. (7). Rise time for the sensor is less than 300 milli-seconds. Settling time occurs in under 450 milliseconds, with less than 2% overshoot. The sensor has an inherently linear pressure drop over its 25 cm 3 / min measurement range, as shown in Fig. (8). The typical differential pressure required per cm 3 / min of mass flow is 5.6 Pa. Existing sensors typically require a 50 Pa pressure difference in order to drive 1 cm 3 / min through the sensor. CONCLUSION The constant differential temperature sensor described in this paper provides superior performance in accuracy, measurement range, response time, and thermal stability. It is also less obtrusive to the measured gas flow than are conventional sensors, due to its minimal pressure drop. ACKNOWLEDGMENTS The authors would like to acknowledge Dr. Robert Ash of Old Dominion University for his contributions.

REFERENCES Goldstein, Richard J., 1983, Fluid Mechanics Measurements, Hemisphere Publishing Corporation, New York Incropera, F.P., and De Witt, D.P., 1990, Fundamentals of Heat and Mass Transfer, John Wiley and Sons Inc., New York 0.150 0.140 0.130 0.120 0.110 0.100 Heater Power 0.090 0.080 0.070 0.060 0.050 0.040 B-8, Upstream B-8, Downstream B-8, Differential Power B-3, Upstream B-3, Downstream B-3, Differential Power B-4, Upstream B-4, Downstream B-4, Differential Power 0.030 0.020 0.010 0.000 0 5 10 15 20 25 Mass Flow Rate (SCCM) Fig. (3): Power Required To Maintain Constant Differential Heater Temperature Versus Mass Flow Rate V 1 R 1 R 2 Fig. (1): Cross Sectional View of the Constant Differential Temperature Mass Flow Sensor V 2 V 3 R 3 R h R a Fig. (2): Control Structure For Maintaining The Heated Resistor R h At Constant Differential Temperature

0.75% B-8, FS=25 SCCM B-8, FS=5 SCCM B-3, FS=5 SCCM 110% Output Non-Linearity as Per Cent of Full Sca 0.50% 0.25% 0.00% -0.25% B-3, FS=25 SCCM B-4, FS=5 SCCM Sesnor Output As Percentage Of F 100% 90% 80% 70% 60% 50% 40% 30% D ifferen tial P ow er 102% 98% -0.50% 20% 10% -0.75% 0% 20% 40% 60% 80% 100% Mass Flow Rate As Per Cent of Full Scale Flow Rate 0% 0 1 2 3 4 5 6 7 8 Time (seconds) Fig. (4): Percentage Deviation In Sensor Output From A Terminal Based Line Versus Percent Of Full Scale Flow Rate Fig. (7): Response Of The Sensor Output To A Step Change In Flow From 0 To 25 SCCM Of Nitrogen 1.5% 1.5% 1.0% B-3, B B-4, C B-5, D B-6, E B-8, G 1.0% 0.5% B-3, B B-4, A B-5, F B-6, E B-8, G Output Change as % Origin 0.5% 0.0% Output Change as % Origin 0.0% -0.5% -1.0% -1.5% -0.5% -2.0% -2.5% -1.0% -20-10 0 10 20 30 40 50 60 Temperature (C) -20-10 0 10 20 30 40 50 60 Temperature (C) Fig. (5): Sensor Full Scale Output Variation Versus Ambient Gas Temperature Fig. (6): Sensor Output Variation At Zero Flow Versus Ambient Gas Temperature

160 B-5 140 B-8 B-3 B-6 Sensor Pressure Drop During Op 120 100 80 60 40 B-4 20 0 0 5 10 15 20 25 30 F lo w R a te (c m ^ 3 / m in ) Fig. (8): Sensor Pressure Drop Versus Flow Rate Of