MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS

BRITISH COMPRESSED GASES ASSOCIATION Registered office: 4a Mallard Way, Pride Park, Derby, UK. DE24 8GX Company Number: 71798, England

Compressed Medical Oxygen Medicinal gas, compressed

Dialflow Regulator. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use

General Safety Information

Pressure Regulators. Instructions for Use

Liquid Medical Oxygen Medicinal gas, cryogenic Package Leaflet: Information for the User

Pressure Regulator. Instructions for Use

Medical Gas Data Sheet (MGDS) Medical air

Home Oxygen Therapy in NHS Scotland. Level 1 Education for Patients & Carers

Oxygen Dialflow Meter. Instructions for Use

Micro Dial-Flowmeter. Instructions for Use

Synthetic medical air.

Peninsula Dental Social Enterprise (PDSE)

Medical Gas Data Sheet (MGDS) Medical helium. BOC: Living healthcare

BOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare

Presence PR, PRC & PRE

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

Medical oxygen/carbon dioxide mixtures

User manual. ZX trolley.

Technical Information Sheet

Lung Function medical gas mixture 4.

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Medical Gas Data Sheet (MGDS) Medical air. BOC: Living healthcare

Medical Gas Data Sheet (MGDS) Medical helium. BOC: Living healthcare

Lung Function medical gas mixture 1.

Glasgow City Council EDUCATION SERVICES. Technician Support Service

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Oxygen Demand Valve. Instructions for Use

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Information about oxygen

PACKAGE LEAFLET: INFORMATION FOR THE USER. Medical Oxygen/Carbon Dioxide Mixtures Oxygen, Carbon Dioxide, Medicinal gas, compressed

SERIES-O REGULATOR. Instructions for use

BCGA GUIDANCE NOTE 17

Analgesic Demand Valve. Instructions for Use

Domiciliary Oxygen Switch. Instructions for Use

Safety information. Working safely with the GENIE hydrogen cylinder

Complex Care Hub Manual: Oxygen Therapy

Safety information. Working safely with the GENIE hydrogen cylinder

9% helium/35% oxygen/ nitrogen medical gas mixture.

BROCKTON AREA MULTI-SERVICES, INC. MEDICAL PROCEDURE GUIDE. Date(s) Reviewed/Revised:

CODE OF PRACTICE 47 THE SAFE USE OF INDIVIDUAL PORTABLE OR MOBILE CYLINDER GAS SUPPLY EQUIPMENT

User Information Sheet 028

SUMMARY OF PRODUCT CHARACTERISTICS

APC Pressure Reducer with Sensor

MANUAL GAS MANIFOLDS

SAFE HANDLING OF CYLINDERS Information for Customers Handling Gas Cylinders

10% carbon dioxide/oxygen medical gas mixture.

Cutting and shaping materials using thermal cutting equipment

5% carbon dioxide/oxygen medical gas mixture.

Working Environment Safety Assessment

1. NAME OF THE MEDICINAL PRODUCT. Medicinal Air, Air Liquide 100%, medicinal gas, compressed. 2. QUALITATIVE AND QUANTITATIVE COMPOSITION

AUTOMATIC GAS MANIFOLDS

Being discharged from hospital with oxygen

Risk Assessment. A guide to the safe storage and use of oxygen and dissolved acetylene equipment

Valve with Integrated Pressure Regulator for medical OXYgen

Glasgow City Council EDUCATION SERVICES. Technician Support Service

Home Oxygen Therapy. Delivering the best in care. UHB is a no smoking Trust

SAFETY PRINCIPLES FOR PRESSURE REGULATORS FOR MEDICAL OXYGEN CYLINDERS

DSEAR- Storage, Use and handling of gas cylinders guidance.

Risk Management Department. Compressed Gas Safety

Bauman Machine Inc. COMPRESSED GAS CYLINDER SAFETY

Safe Handling of Pressurized Gases and Cylinders. Dr. Michael Hanisch

This written Compressed Gas Plan is kept at the corporate office and in the written Safety & Health Program.

TECHNICAL INFORMATION SHEET 17. Revision 1 : 2016 MODEL RISK ASSESSMENT FOR THE MANUAL HANDLING OF GAS CYLINDERS

USE AND ASSEMBLY INSTRUCTIONS

Decant Filling of Gas Cylinders

Home oxygen therapy. Oxygen Team Patient Information Leaflet

Using propane cylinders safely

OXY FUEL SYSTEM SET UP & SHUT DOWN PROCEDURE CHECKLIST ACETYLENE FUEL GAS

Firesafe Flowmeter. Instructions for Use

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

BS ml Bottle Sampling Unit

Savard Labor & Marine Staffing, Inc. Compressed Gas Safety Program Rev3/14

Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet.

General Precautions for handling compressed gases

ATI Skills Modules Checklist for Oxygen Therapy

Technical Specification

ACCUCRAFT UK LTD Unit 4, Long Meadow Industrial Estate, Pontrilas, Herefordshire. HR2 0UA Tel:

TECHNICAL INFORMATION SHEET 18

Welcome to the Specialized Medical Services Respiratory training webinar series!

OXYGEN ACETYLENE WELDING & CUTTING EQUIPMENT

Rev. B. Operating Instructions

How to Order Home Oxygen Part A HOOF

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Compressed medical oxygen.

1.1. Any ASA member who meets the following criteria will be allowed unsupervised access to ASA oxygen system

Safe Work Instructions For Oxygen Acetylene Equipment

Oxygen concentrators. A user guide.

Compressed medical oxygen.

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

CITY AND COUNTY OF DENVER CR&CF RISK UNIT Compressed Gas Safety Standard

MODULE 2 BRAZING CYLINDER TRANSPORT, STORAGE AND EQUIPMENT

UTMB RESPIRATORY CARE SERVICES GUIDELINES Respiratory Care Services Medical Gas Cylinder Storage, Use and Transport

REFERENCE GUIDE ON MEDICINAL GAS CYLINDER VALVES

SUMMARY OF PRODUCT CHARACTERISTICS

80 Litre Suction Oil Drainer

AND OPERATION INSTRUCTIONS

Transcription:

BRITISH COMPRESSED GASES ASSOCIATION Registered office: 4a Mallard Way, Pride Park, Derby, UK. DE24 8GX Company Number: 71798, England www.bcga.co.uk TECHNICAL INFORMATION SHEET 36-2017 MEDICAL GASES. THE SAFE HANDLING AND USE OF GAS CYLINDERS FITTED WITH VALVES WITH INTEGRATED PRESSURE REGULATORS Background Anyone required to handle and use medical gas cylinders should have an understanding of how to do so safely and are to have received appropriate training. For patients at home the training is normally provided by the Homecare oxygen provider. In a healthcare facility environment it is the responsibility of the employer to ensure that their staff have the appropriate training and can operate the equipment correctly and safely. This should include specific training in fire prevention. Whilst oxygen is not flammable, it does support combustion, and where high concentrations of oxygen may be present the consequences of a fire can be devastating to both staff, patients and third parties. Medical oxygen gas cylinders are predominantly fitted with a valve with an integrated pressure regulator making it a safer and easier method for administering gas to the end user. This type of valve is increasingly being fitted to cylinders containing other medical gases. This enables gas to be delivered to the patient without having to connect an additional pressure regulator to the gas cylinder, therefore avoiding the need for the user having to make a high pressure gas connection to the cylinder. The valves on these gas cylinders comply with standards such as BS EN ISO 10524, Part 3, Pressure regulators for use with medical gases. Pressure regulators integrated with cylinder valves. The standard allows for two types of valve to be used. Either you will have a valve that has a separate on/off valve and flow selector, or a valve that combines the on/off valve and flow selector in a single control mechanism. Understand which option is fitted and follow your gas suppliers instructions for the safe use of the cylinder. This guidance details best practice for using and handling gas cylinders fitted with an integrated pressure regulator and its associated equipment safely. The guidance below should be followed: 1. WARNING: Do not smoke or use naked flames near medical gases. 1 of 5

2. Ensure that medical gas cylinders are kept free of oils or greases. When handling gas cylinders ensure your hands are clean. If you have used an alcohol gel on your hands, make sure that it has fully evaporated before handling the cylinder. When creams are necessary, only use approved water based creams. 3. Check you have the correct medical gas for the patient; read the label and take heed of the safety information. If you require additional information ask your healthcare professional for the patient information leaflet and/or contact the gas supplier. 4. Ensure that the gas within the cylinder is within its expiry date; this is detailed on the batch label. Always use the cylinder with the earliest expiry date first. 5. Check there is sufficient gas in the cylinder for the required therapy by checking the contents gauge. 6. If the cylinder is less than a quarter full, consider replacing with a full cylinder. If a patient is being transferred check there is sufficient gas for the whole journey (allowing for changes in patient demand and possible delays). Have spare cylinders available, as necessary. 7. Medical gas cylinders are fitted with a tamper evident seal to indicate that the cylinder has not been used. When using the cylinder for the first time, check the tamper evident seal is intact. If the tamper evident seal is not intact do not use the cylinder and contact your gas supplier for further advice. Before use remove the tamper evident seal. 2 of 5

8a. Ensure the flow selector is set to zero before opening the valve or connecting any equipment. 8b. Where the flow selector is integrated with the shut-off valve ensure it is selected to the off position. 9. Remove any protection cap, or open any fixed cover from the outlet connection. Ensure the outlets are free from any contamination. If there is evidence of contamination do not use. Option 1 Tube connection If using the flow outlet, connect the tubing to the fir tree outlet, making sure that it is securely fitted. Option 2 Probe connection If using the Schrader outlet, insert the probe into the outlet and check that it is securely fitted. NOTE: When engaged a click will be heard. 10. Open the valve slowly. Ideally the cylinder should be held upright while the valve is opened, with the valve outlet pointing away from people and softfurnishing, such as bedding. 11a. Option 1 Tube connection Select the prescribed flow on the flow selector. Check there is a flow of gas from the end of the tubing. 3 of 5

11b. Option 2 Probe connection Check for leaks. Listen for a hissing sound whilst the gas is flowing. If there is a leak, close the valve, allow any excess gas to vent. If the leak seems to originate from the cylinder or valve, stop using and contact the gas supplier. If the leak is from the equipment, remove, check for damage, replace as necessary, then refit the equipment. Check for leaks before use. 12. To prevent cylinders from falling over always place them in an appropriate position, where available use an appropriately designed holder or trolley. In a homecare environment it is acceptable to lie cylinders on the floor taking care to prevent a trip hazard, for example, by positioning close to a wall. 13. When using with a patient in a bed in a healthcare facility, the cylinder should be secured in an upright position next to the bed. It is advisable not to place the cylinder on the bed near the patient. If there is no alternative, then make sure that the cylinder is opened and checked for leaks before placing on the bed. 14. Whilst the cylinder is in use ensure that the gas is flowing; monitor the condition of the patient. Check the contents gauge periodically to ensure there is sufficient gas for the therapy. 15. If you need to temporarily remove the mask or cannula, even for a short period of time, turn the flow selector to 0 or off. 16. When a supply of medical gas is no longer required for the patient or when the cylinder is empty, remove the equipment from the patient before disconnecting and removing or replacing the cylinder. For a replacement cylinder repeat steps 3 to 14. 4 of 5

17a. Option 1 Tube connection For temporary removal the equipment may remain connected. When treatment is no longer required remove the equipment, unless it is required for emergency use. When the cylinder is empty, close the cylinder valve. Always close the valve slowly, do not use excessive force. Allow time for the equipment to vent. Turn the flow selector to 0 or off. Replace any protective caps provided for the valve outlet. 17b. Option 2 Probe connection When not in use or when the cylinder is empty, close the cylinder valve. Always close the valve slowly, do not use excessive force. Momentarily open the flow selector to allow any residual gas to vent and then remove the Schrader probe. Replace any protective caps provided for the valve outlet. 18. Check the contents gauge. Assess if the cylinder contains sufficient gas for further use. As necessary, replace the cylinder. Return empty cylinders to your store for replacement by the gas supplier. 19. If a fault is identified with the gas cylinder or valve assembly: Remove the cylinder from use; Store separately from other cylinders; Label (indicating the suspected fault); Contact the gas supplier. NOTE: The gas supplier may require details such as the bar code number, batch number and a description of the suspected fault. For more information: British Standards Institute (BSI) British Compressed Gases Association (BCGA) Medicines & Healthcare products Regulatory Agency (MHRA) www.bsigroup.co.uk www.bcga.co.uk www.mhra.gov.uk This Technical Information Sheet may be freely reproduced, except for advertising, endorsement or commercial purposes. Please acknowledge the source as the British Compressed Gases Association. All rights reserved. BCGA. 5 of 5