Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Similar documents
Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Practical approach and problems in in-situ RGA calibration

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN

The Helium Leak Detector

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

MIL-STD-883K w/change 2 METHOD SEAL

Residual Gas Analysis Systems for Industry

Total Pressure Measurement Using Residual Gas Analyzers

VIC offers a variety of Calibrated Gas Leaks

METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE)

APPLICATION NOTE. GC Integrated Permeation Device

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Go/no-go test for the acceptance of components for unbaked vacuum sectors

Chapter 6 Leak detection

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Advanced Test Equipment Rentals ATEC (2832)

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

ISO/TC 112/WG 2 N 49. ISO TS QMS WD v6. Date of document Expected action Comment Due Date Background

PROTOCOL FOR COMPRESSED AIR PROFICIENCY TESTING (CAPT) PROGRAM SAMPLE ANALYSIS

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

ONEIDA RESEARCH SERVICES, Inc.

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

CORESTA RECOMMENDED METHOD N 6

Continuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions

Standard Operating Procedure for H 2 S and SO 2 Converter

Bonding Reliability Testing for Wafer Level Packaged MEMS Devices

Hiden QIC Series Gas Analysers Real-time gas analysis systems

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

Gases and welding distributors continue

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Phosgene Sensor. Sensoric COCl2 3E 1

Smart Water Application Technologies (SWAT) TM

What I have learned about SF 6 gas testing.a Practical explanation

Applications Note: Use of "pentane equivalent" calibration gas mixtures

Major Design Features

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

This test shall be carried out on all vehicles equipped with open type traction batteries.

Dissolved Gas Analysis - Past, Present and Future

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES

Gun Vacuum Performance Measurements February 2003

AMT-Ex Dewpoint Transmitter

at NIST: ultra-low outgassing rates

The latest automotive systems require innovative leak test methods and fixturing.

ISO INTERNATIONAL STANDARD

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

In-Water Mass Spectrometry for Characterization of Light Hydrocarbon Seeps and Leaks

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Gas Mixture Two Components. Gas Mixtures Two Components

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, NITROGEN

The instrument should have a mass range from 2 to 2000 m/z or better Interface

STANDARD OPERATING PROCEDURES

LESER Deutschland Standard Functional Tightness Test (Cryo) Content

FLUORESCENCE DETERMINATION OF OXYGEN

High Performance Mass Spectrometers

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer

On-Stream Tightness Testing of Vacuum Process Installations

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

Quadruple mass spectrometers (transducer type)

CalCheck. Instrument User Manual V1.8

Ultimate Performance in Gas Chromatography

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

LEAP CO 2 Laboratory CO 2 mixtures test facility

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

STANDARD OPERATING PROCEDURES

Aug 17, 2001 LAB MANUAL MAXIMUM SPECIFIC GRAVITY (RICE VOIDS TEST) OF PAVING MIXTURES AASHTO Designation T 209 (Mn/DOT Modified))

analytical bulk flow distance FLOW level pressure temperature industrial communication

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Application Note AN-107

Gas Sampling in the DST

AUK Environmental Sensor. Determining the refractive index of the air. AUK Environmental Sensor

Leak Checking Large Vacuum Chambers

Misuse of Combustible Gas Meters

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Mass Spec will not Autotune

APPENDIX 1. PART 10: Leak Detection Inspector, Level 1, 2 and 3 CERTIFICATION SCHEME FOR PERSONNEL. 3 rd Edition, February 2016

Subpart E. Pressure Testing

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

Mass Flow Controllers

ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Barrier Development and Evaluation Methodology. D.S. Musgrave 1 1 Thermal Visions, Inc., Granville, USA

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Gaseous oxygen technical grade Specification

DEFENSE LOGISTICS AGENCY LAND AND MARITIME P.O. BOX 3990 COLUMBUS, OHIO

Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements

ASTM UPDATE to COQA. COQA Meeting Feb San Antonio. RJ (Bob) Falkiner SR (Steve) David GJ (Jay) Grills W (Wayne) Kriel

DSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

A Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide

Transcription:

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The purpose of this test method is to quantitatively measure the relative concentration of the internal vapor content, including water vapor, in gas filled hermetically sealed devices using a mass spectrometry technique. 2.0 Scope This test method is applicable to all hermetic (cavity) devices fabricated with internal pressures approximating one atmosphere. Any device size may be tested if the device has an internal volume within the volume range where the instrument measurement specifications can be demonstrated for the cavity size of that device. Results of this test reflect the internal vapor content in the device at the time of the test. Results depend on device history and device preconditioning. It is the responsibility of the submitter to specify any additional device preconditioning that the test lab must perform prior to conducting the test and, to provide adequate suppporting information to the laboratory regarding the purpose and intended use of the test data. Although this protocol is destructive, it may be used on electrical or mechanical reject devices provided that hermeticity has not been jeopardized and the devices are fully representative of the manufacturing process and environmental conditions per the purpose and intent of the test. This test method may be used as a Pass/Fail test to qualify production lots or as a process quality monitor per submitter specified test conditions. Client-specific test protocals may be designed to achieve optimum data accuracy and sensitivity for ongoing product monitoring programs. This test method can be a powerful engineering tool when device test temperature and device prebake/pre-conditioning conditions are changed. It can be used to evaluate the sources of moisture and other vapors found in hermetic devices, determine the thermal breakdown temperatures of organic materials and, can evaluate the effectiveness of hermetic seal screening tests. 3.0 Apparatus The apparatus to conduct internal vapor analysis shall consist of: 3.1 a mass spectrometer capable of: 3.1.1 detecting mass ions in the minimum continuous mass range from 2 Atomic Mass Units (AMU) to 150 AMU; 3.1.2 unit mass resolution of spectra; 3.1.3 the ability to measure and control the peak shape tuning and peak-to-peak resolution over the minimum continuous mass range;

3.1.4 spectra acquisition in continuous scan mode in minimum increments of 0.1 AMU; 3.1.5 sufficient ion signal intensity and ion signal dynamic range to meet System Measurement Requirements per Section 4.0 of this document. A minimum of three (3) decades of dynamic range of the ion signal intensity is required. 3.2 a heating system to maintain the device casing at a specified minimum temperature during the test. The specified minimum temperature shall be 100 o C ±5 o C unless otherwise specified; 3.3 a puncture mechanism capable of piercing a hole in the device without compromising the integrity of the test; 3.4 a vacuum system capable of achieving the System Measurement Specifications per Section 4.0; 3.5 a heated transfer path, which shall be maintained at 100 o C ± 5 o C, capable of transferring the device gases from the device to the mass spectrometer chamber; 3.6 pressure sensor(s) to provide a means of measuring the pressure in the transfer path and in the mass spectrometer chamber; 3.7 a leak valve or equivalent mechanism to control the flow of gases from the device to the mass spectrometer chamber; 3.8 a calibration system to provide a means of calibrating the mass spectrometer for each of the commonly detected gases (i.e., water vapor, nitrogen, oxygen, argon, carbon dioxide, hydrogen, helium, where applicable) at concentration ranges and volumes typical of the devices to be tested and, in a manner that r e p r oduces actual device testing protocol; 3.9 a moisture generator system capable of generating moisture concentrations in the range specified in Paragraph 4.5 and a dewpoint analyzer capable of measuring dewpoint in the same range to an accuracy of ± 0.2 o C or better. Calibration of this device shall occur at least once every year to NIST traceable standards or other internationally recognized standards; 3.10 profile calibrated oven(s) or equivalent to precondition devices at a specified temperature as needed prior to hot insertion into the device holder. 4.0 System Measurement Specifications The measurement system of the mass spectrometer shall meet the following measurement specifications and shall be demonstrated on an as needed basis. 4.1 Sufficient dynamic range of the ion signal intensity to directly measure the unsaturated amplitude of all mass peaks within the minimum continuous mass range. 4.2 Detection of 100 ppmv of oxygen, argon, carbon dioxide, hydrogen and helium in a nitrogen balance for the volume range of devices to be tested. 4.3 Quantitative analysis of oxygen, argon, carbon dioxide, hydrogen, helium, methane, ammonia and fluorocarbon in a nitrogen balance to a reproducible accuracy of ± 5% of the true value or ±100 ppmv, whichever is greater (e.g., if the true value of oxygen is 20.95%, then the accuracy must be ± 1.05% which is 5% of 20.95%). 4.4 Quantitative analysis of moisture at 5000 ppmv in a nitrogen balance to a reproducible accuracy of ±10%.

4.5 Validation of a linear calibration curve for moisture between the concentration levels of 2500 ppmv and 7500 ppmv in a nitrogen/moisture mixture. The system shall demonstrate an average error that is less than 500 ppmv using a minimum of nine (9) data points over this range. The data shall be generated in three groupings of three data points, each at the median and extremes of the range (ref. Figure 1). Figure 1: 3-Point Calibration Curve in Triplicate 10,000 7,500 5,000 2,500 0 0 2,500 5,000 7,500 10,000 Generat ed Moist ure (ppmv ) Generated Reported Absolute Moisture Moisture Error (ppmv) (ppmv) (ppmv) 2,500 2,396 104 2,500 2,628 128 2,500 2,452 48 5,000 5,077 77 5,000 4,876 124 5,000 5,199 199 7,500 7,682 182 7,500 7,598 98 7,500 7,456 44 Average Error: 112 4.6 Quantitative analysis of all other gases not specified in Paragraphs 4.2 through 4.5 shall be performed assuming a unit calibration factor using the primary peak. 5.0 Calibration 5.1 At a minimum, weekly calibration shall be performed for nitrogen, oxygen, argon, hydrogen, helium and carbon dioxide per Paragraph 4.3. Air, at room temperature, may be used as an acceptable standard for the calibration of oxygen and argon, if calibrated relative to nitrogen. NIST traceable standards or other internationally recognized standards shall be used for the calibration of hydrogen, helium and carbon dioxide. 5.2 At a minimum, yearly calibration shall be performed for methane, ammonia and fluorocarbon per 4.3. 5.3 A daily calibration shall be performed for moisture at 5000 ppmv at the volume ranges representative of devices to be tested on that instrument on that day. This calibration shall have a reproducible accuracy of ±10%. 5.4 Unless otherwise calibrated, on an as needed basis, unit calibration shall be assumed for the remaining inorganic and organic gases. 6.0 Information to be Submitted to the Analytical Facility The following information shall be provided by the Submitter to the Analytical Facility along with the devices to be tested: 6.1 the number of devices to be tested; 6.2 where applicable, device failure criteria which typically specifies concentration of specific gases (e.g., moisture) in ppmv or percent; 6.3 the estimated internal pressure of the device; 6.4 the presence of polymeric materials, organics or desiccants in the cavity of the device;

6.5 prebake conditions, if any; Note: An optional prebake may be performed on hermetic devices immediately preceding the Internal Vapor Analysis test. Prebake conditions may be dependent on the engineering objectives, testing protocol, reference specifications and device material temperature limits. Typical prebake conditions are 16-24 hours at 100 C, 16-24 hours at 85 C or 150 hrs at 150 C. 6.6 the device minimum test temperature if other than 100 o C ± 5 o C; 6.7 the preferred puncture site on the device as well as the thickness and internal clearance at the selected site; if critical. 7.0 Device Handling and Preparation 7.1 Devices shall be inspected visually for obvious defects that may prevent adequate testing of the device. 7.2 The exterior of the device must be free of process materials (e.g., conformal coating) and contamination so as not to interfere with the test protocol. The exterior of the device may be cleaned if warranted. Any mechanical or chemical cleaning shall be performed in a manner so as not to compromise the integrity of the test. 7.3 Leads of the device may be bent or trimmed when permitted and only in a manner that does not compromise the integrity of the package seal. 7.4 If specified, the device shall be preheated prior to insertion in the mass spectrometer for testing. Preconditioning shall be reported on the test report. 7.5 The device may be thinned by abrasion at the designated puncture site to facilitate puncturing. Thinning must be done immediately prior to insertion into the mass spectrometer. 8.0 Test Procedure 8.1 Devices shall be prepared for testing per the device handling and preparation procedure in Section 7.0. 8.2 At the conclusion of any preconditioning as required, devices shall be removed individually from the oven. The puncture site shall be thinned as needed and the device loaded immediately into the device mounting fixture. If preconditioning is not required, the device shall be thinned as needed and then loaded into the device mounting fixture. 8.3 The mass spectrometer chamber shall be pumped down until background spectra and pressure levels are sufficient to achieve the System Measurement Specifications per Section 4.0. The background levels of all volatile species for which calibration has been conducted shall be undetectable or sufficiently low so as not to impair system detection limits or accuracy requirements. 8.4 The background spectra for the minimum mass range of 2 AMU to 150 AMU shall be acquired.

8.5 The device shall then be punctured and the following properties of the released gases shall be measured immediately: 8.5.1 a mass spectra scan(s) to measure the internal vapor content of the device at the time of puncture. The mass spectra shall be generated per the minimum capabilities as detailed in Section 3.1. 8.5.2 the increase in transfer path pressure as the gases are released from the punctured device. Note: The measured pressure indicates the relative amount of gas in the device. If the measured pressure deviates from the norm for that device type, the measured pressure may indicate, among other things, that: (1) the device package was not hermetic; (2) the device was sealed under reduced pressure or partial vacuum; (3) the puncture was not fully accomplished; or (4) the device may not have been mounted properly in the test fixture. 8.6 The laboratory shall interpret the data and make a determination as to the following: 8.6.1 that the test was or was not a valid test, using the pressure reading cited in Paragraph 8.5.2 and other system performance indicators; 8.6.2 the proportions (by volume) of all the released gases with data corrected for system-dependent matrix effects, when applicable. 8.6.3 the occurrence of significant unidentified peaks in the mass spectra which shall be reported as unknowns. 8.7 Test results shall be recorded on a report and shall include the following data and notes: 8.7.1 the proportions (by volume) of all the released gases or, if the test of an individual device is invalid, the device test attempt shall be reported as a "No Test" with an explanation; 8.7.2 prebake conditions, as performed; 8.7.3 any test-related information that may impact the interpretation of the test results. 8.8 A copy of the test report, raw data and related calibration data sufficient to requantitate the test results shall be archived. February, 2002 Oneida Research Services, Inc.