Manual Oxygen Analyzer A4009T

Similar documents
JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

Instruction and Maintenance Manual

Autocalibration Systems For In Situ Oxygen Analyzers

FTC130 Transmitter. Operating Manual

Sensoric 4-20 ma Transmitter Board Operation Manual

Columbus Instruments

SAPCON. User Manual. Capacitance Continuous Level Indicator. . Introduction. . General Description. . Principle of Operation. .

BUBBLER CONTROL SYSTEM

Best Practice Guide, Servomex 2700

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Electronic gas volume corrector model DGVC-04

Instruction and Maintenance Manual

GAS FUEL VALVE FORM AGV5 OM 8-03

CONTROL and INSTRUMENTATION

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

OSAT Series. Intrinsically Safe Infrared Temperature Sensor Operator's Guide

Hot extractive multi-channel gas analyser for continuous process and emissions monitoring.

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

CONDUCTIVITY POINT/S LEVEL SWITCH JAYCEEAQUA 4000 SERIES

Technical Data Sheet MF010-O-LC

EPP4 Pressure Regulator Basic G 1/4"

FLAMMABLE GAS SENSOR/TRANSMITTER

Transmitter CS 21 Operation Manual

Installation, operating and maintenance Instructions for Seemag bypass level indicator

MODEL GT820 OXYGEN SENSOR

BUBBLER CONTROL SYSTEM

Autocalibration Systems for In Situ Oxygen Analyzers

T EK-SUB 4800C 19 mm Submersible Level Transmitter

Model EXAxt AV550G. Averaging Oxygen Analyzer

Stand-Alone Bubble Detection System

Overview. Front Panel: Keypad and Display

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

CO, NO, NO 2, NO x, SO 2, CH 4, HCl, CO 2 & H ² O. Extractive low cost, low maintenance devices for continuous process and emissions monitoring.

Model PDT Dewpoint Transmitter

ACV-10 Automatic Control Valve

SSL. Smart Valve Positioner.

High-performance submersible pressure transmitter For level measurement Model LH-20

Climatic Independent Level Sensor LAR

OPTIBAR P 1010 C Technical Datasheet

Hydro-Control V User Guide

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Instruction Manual Differential Pressure Transmitter

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

Pneumatic high-pressure controller Model CPC7000

Transmitter CS 21 Operation Manual

Operating Instructions Part No

SSR. Smart Valve Positioner.

RPS900W Redundant Power Supply. Installation Guide.

Datasheet: K-30 ASCII Sensor

Pneumatic high-pressure controller Model CPC7000

High-performance submersible pressure transmitter For level measurement Model LH-20

L (up to 25 m) Electronic part. working temp.: Ñ IP 66 / IP67. Mounting bracket ÐÑ. Sensor type. working temp.:

SIL Safety Manual. ULTRAMAT 6 Gas Analyzer for the Determination of IR-Absorbing Gases. Supplement to instruction manual ULTRAMAT 6 and OXYMAT 6

SonoMeter 30 Energy Meters

combustion efficiency monitoring General Purpose Applications Flue gas oxygen analyzers Combustion & Environmental Monitoring An Company

DPI 610/615 Series. Druck Portable. Pressure Calibrators. GE Sensing. Features. Setting the Standard for Portable. Pressure Calibrators

User manual CF8-D/W-IN

Supplementary Operator s Manual 42/24-14 EN Rev. 3

[Instruments for vacuum measurement, checking and adjustment] 3

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

TH-1800-T TRX-700-CNG Converter

Quick Installation Guide Doc nr.: C Date:

BS12 Piezoresistive OEM Pressure Sensor

High-performance submersible pressure transmitter For level measurement Model LH-10

Pressure Automated Calibration Equipment

ATM/GR. Electronic pressure switches CUSTOMER BENEFITS

Moisture Image Series 1. Service Manual

Features. Description

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

combustion efficiency monitoring Flue gas oxygen analyzers General Purpose Applications Combustion & Environmental Monitoring

Manual of SF6 Comprehensive Tester

GasSense NDIR User Manual

Gas density monitor With integrated transmitter Model GDM-100-TI

Controller for boilers Galan - regulator for management of heating elements and circuit SolarSentinel-DBTW User guide

INSTRUCTION MANUAL Pressure Relief Device LPT

SG33KTL-M Quick Installation Guide. 1 Unpacking and Inspection

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

User Manual. asense VAV. CO 2 / temperature sensor with built-in general purpose controller

AUTOMATIC GAS MANIFOLDS

BS12N Piezoresistive OEM Pressure Sensor

SG36KTL-M Quick Installation Guide. 1 Unpacking and Inspection

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller

WF STEUERUNGSTECHNIK GMBH. INFORMATION and TECNICAL DESCRIPTION

Instruction Manual. Biogas Analyzer InCa Bio 04

Smart performance with innovative and ever-strong coil drive even under harsh working environments

Mass Flow Controller (MFC) for Gases

Manual Weighingblock VB2 series and Uniscale

USER MANUAL. Intelligent Diagnostic Controller IDC24-A IDC24-AF IDC24-AFL IDC24-F IDP24-A * IDP24-AF * IDP24-AFL * IDP24-F * 1/73

Intrinsically safe solenoid valves on/off controls - ATEX certification

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

Programmable Submersible Level Transmitters

AMT-Ex Dewpoint Transmitter

User Manual. asense miii. / CO sensor with built-in general purpose controller

HumiSys HF High Flow RH Generator

ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL

Model 601CV. Mass Flow Controller. Product Features:

Malema Sensors M-2700 Integrated Ultrasonic Flow Meter

Transcription:

Manual Oxygen Analyzer A4009T Revision 01-EN DIVIZE B.V.- info@divize.com November 25, 2014

Contents 1 Measuring principle 3 2 Safety 4 3 Installation 5 3.1 Mounting................................................ 5 3.1.1 Analyzer............................................. 5 3.1.2 Sensor.............................................. 5 3.2 Electrical connections.......................................... 6 3.2.1 Power supply.......................................... 8 3.2.2 Relays.............................................. 8 3.2.3 Analog Outputs......................................... 8 3.2.4 Digital inputs.......................................... 8 3.2.5 Sensor.............................................. 8 4 Settings 10 4.1 Control system............................................. 10 4.2 Settings................................................. 10 4.2.1 Setting values.......................................... 11 4.2.2 Menu-structure......................................... 12 4.2.3 Parameters........................................... 12 4.2.3.1 Alarms........................................ 12 4.2.3.2 Analog Output.................................... 13 4.2.3.3 Version Hard- & Software.............................. 14 4.2.3.4 Calibration...................................... 14 5 Maintenance 15 5.1 Analyzer................................................. 15 5.2 Sensor.................................................. 15 5.3 Calibration................................................ 15 5.3.1 Setting Zero Gas and Span Gas................................ 15 5.3.1.1 Zero Gas....................................... 15 5.3.1.2 Span Gas....................................... 16 5.3.2 Manual Calibration....................................... 17 5.3.2.1 Zero Gas....................................... 17 5.3.2.2 Span Gas....................................... 18 Manual A4009T-O2 analyser R01-EN 1 / 33

CONTENTS 6 Failure analysis 19 6.1 Status display.............................................. 19 6.2 Alarm display.............................................. 19 6.3 Failure.................................................. 20 6.4 Storing.................................................. 21 7 Automatic Calibration 22 7.1 Installation................................................ 22 7.1.1 Connection diagram automatic calibration valves........................ 23 7.1.2 Electrical connection for automatic calibration........................ 23 7.2 Conguration.............................................. 24 7.2.1 Setting the auto calibration parameters............................. 24 7.2.1.1 Selecting the automatic calibration menu...................... 24 7.2.1.2 Auto cal parameters................................. 24 7.2.1.3 Ebable-/disable automatic calibration........................ 25 7.2.1.4 Setting the zero gas time............................... 25 7.2.1.5 Setting the Span Gas Time............................. 26 7.2.1.6 Setting the maximum dierence from the calibration gas.............. 26 7.2.2 Starting automatic calibration.................................. 26 Appendix 1 : Dimension 28 Appendix 2: Specication 29 General................................................. 29 Power Supply.............................................. 29 Digital Inputs.............................................. 30 Relay Contacts............................................. 30 Analog Output............................................. 31 Sensor.................................................. 32 Manual A4009T-O2 analyser R01-EN 2 / 33

Chapter 1 Measuring principle The sensor measures the dierence in oxygen content between a reference and a sample gas. As reference gas normal ambient air is used with a oxygen content of 20,94%. The sensor consist of 4 main parts. Pumpingcel (3), nernstcel (4) heater (5) and diusion-chamber (6). The sample gas enters the diusion chamber. Based on the dierence in oxygen content of the gas in the reference chamber (2) and diusion chamber (6) a voltage is generated. When the oxygen content of both chambers is equal the generated voltage is 450 mv. The output of the dierential amplier (7) is zero in this case. When the oxygen content is not equal oxygen molecules can be added or removed by the pumping cell (3) In this case the output of the dierential amplier is not zero. The dierence is evaluated by the PID controller and sent to the pumping cell (3) The oxygen content is adjusted until the dierence is zero. The current to the pumping cell is a measure for the measured oxygen content and is indicated on the display (9). The reference air for the sensor is supplied through the wiring sleeve on the B4009Sxx sensors, therefore the wiring may not be obstructed. On the B4009Txx sensors the reference air is supplied through the metal casing. Pressure dierence between the sample gas line and the reference air supply will inuence the transport of oxygen molecules and therefore inuence the measurement. The measurement may be compensated for this pressure dierence. Figure 1.1: Schematic display of the oxygen sensor. Manual A4009T-O2 analyser R01-EN 3 / 33

Chapter 2 Safety The analyzer may only be installed and controlled by skilled personnel. Follow the instructions in this manual. Verify the electrical connections before applying power to the unit. The analyzer is only suitable for installation in a safe environment, the analyzer may not be installed in a explosion hazardous area unless sucient protection is in place. The analyzer is intended for the measurement of gas with low oxygen content. Gas with low oxygen content is dangerous to humans. It is necessary to make sure there is sucient ventilation while working with the analyzer. Manual A4009T-O2 analyser R01-EN 4 / 33

Chapter 3 Installation 3.1 Mounting 3.1.1 Analyzer. The analyzer housing is designed to be mounted inside a (control) panel. The control elements and display are easy accessible at the front, the wiring can be connected at the backside inside the panel. The analyzer shall be mounted in a clean place, away from dust, moisture and oil. The analyzer shall be mounted in a by maintenance personnel easy accessible location. The ambient temperature at the mounting location shall not exceed 50 C. 3.1.2 Sensor The sensor shall be mounted in the gas ow that needs to be analyzed. The mounting assembly M4009R01 can be used for mounting the sensor. The sensor shall be placed in a sucient ventilated area as the reference air for the sensor is the surrounding air. The measurement accuracy may be inuenced when sucient ventilation is not available. The sensor will be hot during operation. it may be necessary to provide a heat shield to protect personnel. The sensor must be mounted in a 90 degree angel to the gas ow see gure 3.1. To prevent moisture entering the sensor the sensor shall be mounted straight upwards if possible. If this is not possible the sensor may be mounted in an angle of maximum 80 degrees to the vertical, as shown in gure 3.2. The sensor must be protected from water. The sensor cable shall be mounted stress free as this is the supply for reference air to the sensor. The sensor may not be connected to a power supply other then the analyzer as this may damage the sensor. Manual A4009T-O2 analyser R01-EN 5 / 33

3.2. ELECTRICAL CONNECTIONS Figure 3.1: Mounting the sensor, maximum angle of 80 degrees from vertical Figure 3.2: Mounting sensor 90 deg to gasow 3.2 Electrical connections On the backside of the analyzer, the electrical connections are provided. These connections are provided with detachable screw connectors. The connectors can easily be pulled from the analyzer for dis-assembly and assembly of the analyzer. Manual A4009T-O2 analyser R01-EN 6 / 33

3.2. ELECTRICAL CONNECTIONS Figure 3.3: Electrical connections on the backside of the analyzer Con Symbol Beschrijving 1 +24VDC Power supply positive (+) 2 0VDC Power supply negative (-) 11 C Relay contact Common Analyser failure Alarm 12 NC Relay contact Normally Closed Analyser failure Alarm 14 NO Relay contact Normally Open Analyser failure Alarm 21 C Relay contact Common Oxygen Low-low Alarm 22 NC Relay contact Normally Closed Oxygen Low-low Alarm 24 NO Relay contact Normally Open Oxygen Low-low Alarm 31 C Relay contact Common Oxygen Oxygen high-high Alarm 32 NC Relay contact Normally Closed Oxygen high-high Alarm 34 NO Relay contact Normally Open Oxygen high-high Alarm 41 C Relay contact Common Oxygen hoog alarm of Automatic calibration Span gas 44 NO 51 C Relay contact Common Oxygen low alarm or Automatic calibration zero gas 54 NO 61 Analog output 1 positive (+) 62 Analog output 1 negative (-) 63 Analog output 2 positive (+) 64 Analog output 2 negative (-) 71 Digital input 1positive (+) 72 Digital input 1 negative (-) 73 Digital input 2positive (+) 74 Digital input 2 negative (-) 81 H+ Sensor Heater + 82 H- Sensor Heater - 83 UN Sensor voltage 84 VG Sensor Virtual reference 85 IP Sensor current 86 IA Sensor current reference Table 3.1: Electrical connections to the analyzer Manual A4009T-O2 analyser R01-EN 7 / 33

3.2. ELECTRICAL CONNECTIONS 3.2.1 Power supply The power supply shall be connected to terminal 1 (+) and 2 (-). A stabilized DC supply of 24 VDC is required for reliable operation of the analyzer. The power supply shall be capable of delivering at least 1,3 Ampere. 3.2.2 Relays The analyzer consist of 5 relay contacts, 3 changeover contacts and 2 normally open contacts. All relay contacts are fail-safe, this means the relay is energized when there is no failure. In case of a failure the relay will be deenergised. Both normally open contacts serve a double function. When automatic calibration is selected the contacts are used to control the calibration gas valves. When automatic calibration is disabled the contacts are used for the high and low oxygen alarms. More information about calibration can be found in chapter 7 3.2.3 Analog Outputs The analyzer is equipped with 2 independent analog 4-20mA outputs. Figure 3.4: Connection diagram for Analog output 4-20mA Both analog outputs are passive, a power supply shall be attached to operate the outputs. The analog outputs represent the measured oxygen content. The range can be selected through the settings menu. see 4.2.3.2 3.2.4 Digital inputs The analyzer is equipped with 2 digital inputs. The inputs can be activated by connecting a 24VDC power supply. Digital input 1 activates the alarm suppression function. When the input is high the alarms are suppressed. Digital input 2 activates the automatic calibration routine. 3.2.5 Sensor The sensor is connected to the analyzer by means of the connection cable K4009A01. There are 2 types of sensors which shall be connected in dierent ways. The type of sensor is indicated on the sticker on the sensor. Connection of B4009S01 and B4009S02 sensor. Manual A4009T-O2 analyser R01-EN 8 / 33

3.2. ELECTRICAL CONNECTIONS Sensor Cable Connector Wire Connection A4009T Red 1 1-brown 85 Yellow 2 2-red 84 White 3 3-orange 81 Grey 4 4-yellow 82 Black 5 5-green 86 6 6-blue 83 Table 3.2: Connection B4009Sxx sensor Connection of the B4009T01 sensor Sensor Cable Connector Wire Connection A4009T 1 1-brown 86 White 2 2-red 85 Yellow 3 3-orange 81 Grey 4 4-yellow 83 Blue 5 5-green 82 Black 6 6-blue 84 Table 3.3: Connection B4009Txx sensor Manual A4009T-O2 analyser R01-EN 9 / 33

Chapter 4 Settings 4.1 Control system Figure 4.1: Front side analyzer. Button functions + Next / More Previous/Less v Select option/conrm x Back/Escape Table 4.1: Button functions. 4.2 Settings The user menu provides access to several settings. For example, Alarm levels, analog output ranges and calibration gases can be set through the menu. Manual A4009T-O2 analyser R01-EN 10 / 33

4.2. SETTINGS 4.2.1 Setting values Display < High High > O2 25.0% val < High High > O2 25.0% < High High > O2 15.0% < High High > O2 15.0% < High High > O2 15.0% val Select through the navigation the parameter that needs to be changed, adjustable settings can be found in this chapter. Conrm you want to change the selected parameter by pressing the <V> button With the <+> and <-> button the selected value can be changed. In the example the rst digit of the number is adjusted By pressing the <V> button the change is conrmed. If available the cursor will jump to the next digit. By pressing the <X> button the change can be canceled. The described procedure need to be repeated for all digits. After changing the last digit, the change is conrmed by pressing the <V> buton. Now the parameter is changed and stored. Manual A4009T-O2 analyser R01-EN 11 / 33

4.2. SETTINGS 4.2.2 Menu-structure The settings are available through the menu structure shown below: See gure 4.2 20.79 % <v> < Alarms > <v> <x> <x> see 4.2.3.1 Alarms <+>_ <->^ < Analog Out > <v> see 4.2.3.2 Analog output <x> <+>_ <->^ < Version > <v> HW build 0948 <x> SW build 0210 <+>_ <->^ < Calibration > <v> see 4.2.3.4 Calibration <x> <press v for 10 seconds>_ < Auto Calib > <v> see 7.2.1.2 Autocal parameters OFF <x> Figure 4.2: Main menu. 4.2.3 Parameters 4.2.3.1 Alarms Display Parameter MIN MAX TYP < High High > O2 25.0% val <+>_ <->^ < High > O2 24.0% val <+>_ <->^ < Low > O2 02.0% val <+>_ <->^ < Low Low > O2 01.0% val When the measured oxygen content is higher then the value set as high-high alarm, the high-high oxygen alarm will be activated. When the measured oxygen content is higher then the value set as high alarm, the high oxygen alarm will be activated. When the measured oxygen content is lower then the value set as low alarm, the low oxygen alarm will be activated. When the measured oxygen content is lower then the value set as low-low alarm, the low-low oxygen alarm will be activated. 0.0 99.9 25.0 0.0 99.9 24.0 0.0 99.9 2.0 0.0 99.9 0.0 Manual A4009T-O2 analyser R01-EN 12 / 33

4.2. SETTINGS 4.2.3.2 Analog Output The analog output represents the measured oxygen content as a current from 4 to 20 ma. This current can be evaluated by a display, controller, recorder, PLC or DCS system, in order to monitor the oxygen content in a remote location. The connected equipment shall be congured so the measured oxygen content is equal to the settings in the A4009T oxygen analyzer. Figure 4.3: Analog output characteristic Based on the settings for 4 ma (MIN) and 20 ma (MAX) the A4009T oxygen analyzer will calculate the current supply to the 4-20 ma loop based on the measured oxygen content. It is also possible to set the value for (MIN) higher then the value for (MAX) When the MIN value is equal to the MAX value the analog output is disabled. Display Parameter MIN MAX TYP < Analog 1 MIN > O2 05.0% val <+>_ <->^ < Analog 1 MAX > O2 21.0% val <+>_ <->^ < Analog 2 MIN > O2 05.0% val <+>_ <->^ < Analog 2 MAX > O2 21.0% val When the measured oxygen content is equal or lower then the set value the output current wil be 4 ma for output 1. When the measured oxygen content is equal or lower then the set value the output currrent wil be 20 ma for output 1. When the measured oxygen content is equal or lower then the set value the output currrent wil be 4 ma for output 1. When the measured oxygen content is equal or lower then the set value the output currrent wil be 20 ma for output 2. 0.0 99.9 0.0 0.0 99.9 25.0 0.0 99.9 0.0 0.0 99.9 25.0 Manual A4009T-O2 analyser R01-EN 13 / 33

4.2. SETTINGS 4.2.3.3 Version Hard- & Software Display Parameter MIN MAX TYP Hard- en software version - - - HW Build 0948 SW Build 0210 4.2.3.4 Calibration Display Parameter MIN MAX TYP < Zero Gas > O2 01.50% val Oxygen content of the zero calibration gas. 0.0 99.9 1.50 <+>_ <->^ < Span Gas > O2 20.94% val Oxygen content of the span calibration gas. 0.0 99.9 20.94 <+>_ <->^ < Calib. Zero > Start manual zero gas calibration. - - - <+>_ <->^ < Calib. Span > Start manual span gas calibration. - - - Manual A4009T-O2 analyser R01-EN 14 / 33

Chapter 5 Maintenance 5.1 Analyzer The analyzer does no require much maintenance. The control panel shall be cleaned regularly with a soft cloth. 5.2 Sensor The sensor does not require a lot of maintenance during it's life span. Check the sensor for contamination and clean if necessary. In case of severe contamination the sensor shall be replaced. Check the wiring to the sensor, check if the protection sleeve to the wiring is not pinched as this is the reference air supply to the sensor. 5.3 Calibration Depending on the usage, the analyzer need to be calibrated regularly. Before calibration is started the sensor need to be heated at a stable temperature. "Zero Calibration" shall be done with a gas mixture with certied composition. "Span gas" shall be instrument air with 20.94% oxygen; (Normal ambient air) The pressure for the "Zero" and "Span" gas shall be equal to the measured gas. 5.3.1 Setting Zero Gas and Span Gas 5.3.1.1 Zero Gas The oxygen concentration in the zero gas shall be set prior to calibration. Manual A4009T-O2 analyser R01-EN 15 / 33

5.3. CALIBRATION Display 20.94 % Measurement display Press <V> and press <-> next < Calibration > Calibration option mainmenu. Press <V> and <-> <-> next < Zero Gas > O2 01.50% val Zero Gas option Calibration menu. In this menu, the oxygen percentage of the zero gas can be selected. < Zero Gas > O2 01.50% val See chapter 4.2.1 for a description of changing the value.... < Zero Gas > O2 01.50% val Exit the screen by pressing <X> several times until the measured oxygen content apears in the display. 5.3.1.2 Span Gas Display 20.94 % Start Press <V> and <-> < Calibration > Calibration main menu Press <V> and <+> < Span Gas > O2 20.94% val Span Gas value Calibration menu < Span Gas > O2 20.94% val See chapter 4.2.1 for a description of the procedure for changing the value.... < Span Gas > O2 20.94% val The value is set. Go back to the screen with oxygen measurement by pressing <X> several times. Manual A4009T-O2 analyser R01-EN 16 / 33

5.3. CALIBRATION 5.3.2 Manual Calibration 5.3.2.1 Zero Gas Display 20.94 % Start. Press <V> and <-> < Calibration > Calibration main menu Press <V> < Calib. Zero > Select zero gas calibration and press <V> Zero Gas O2 01.50% val The oxygen content of the reference gas which shall be used is displayed. Conrm by pressing <V> Zero Measure O2 01.53% val The display shows the current measured value. Wait until the value is stable and conrm by pressing <V>. Press UP For 3s Press <+> for 3 seconds to complete the calibration. Calibration Done Calibration is completed! Go back to the screen with oxygen measurement by pressing <X> several times. Manual A4009T-O2 analyser R01-EN 17 / 33

5.3. CALIBRATION 5.3.2.2 Span Gas Display 20.94 % Start. Press <V> and <-> < Calibration > Calibration main menu Press <V> < Calib. Span > Select span gas calibration by pressing <+> and press <V> Span Gas O2 20.94% val The oxygen content of the reference gas which shall be used is displayed. Conrm by pressing <V> Span Measure O2 20.94% val The display shows the current measured value. Wait until the value is stable and conrm by pressing <V>. Press UP For 3s Press <+> for 3 seconds to complete the calibration. Calibration Done Calibration is completed! Go back to the screen with oxygen measurement by pressing <X> several times. The analyzer has an automatic calibration function. This is documented in chapter 7 Manual A4009T-O2 analyser R01-EN 18 / 33

Chapter 6 Failure analysis 6.1 Status display Display Sensor Heating Description The sensor is heating. This message is displayed when the temperature of the sensor is to low during start up of the analyser. Auto cal Zero The automatic calibration sequence is running. This message is displayed during the zero calibration. The message will only appear when automatic calibration is selected. Auto cal Span The automatic calibration sequence is running. This message is displayed during the span calibration. The message will only appear when automatic calibration is selected. 6.2 Alarm display Display Cause Solution High High Alarm O2 20.00% The measured oxygen content is higher then the preset alarm value for the "High High"-alarm. Check the measured oxygen content. Modify the alarm value if required. High Alarm O2 18.00% The measured oxygen content is higher then the preset alarm value for the"high"-alarm Check the measured oxygen content. Modify the alarm value if required. Low Alarm O2 02.00% The measured oxygen content is lower then the preset alarm value for the "Low"-alarm Check the measured oxygen content. Modify the alarm value if required. Low Low Alarm O2 01.00% The measured oxygen content is lower then the preset alarm value for the"low Low"-alarm Check the measured oxygen content. Modify the alarm value if required. See 4.2.3.1 for an explanation about setting the alarm values. Manual A4009T-O2 analyser R01-EN 19 / 33

6.3. FAILURE 6.3 Failure Display Cause Solution Heater Failure Heater current to low. Check sensor wiring/connections. Replace sensor. Analog Out 1 Failure Analog Out 2 Failure Auto Cal Failed O2 20.89% Fail, difference too large No supply voltage on analog output. See "Analog Out 1 Failure". The dierence between the oxygen content measured during automatic calibration and the set calibration value is to high. The dierence between the oxygen content measured during calibration and the set calibration value is to high. In case the analog output is not used, check if the output is disabled by setting the MIN and MAX value in the Analog Out menu at the same value. Check the wiring/connections. Check the external power supply. See "Analog Out 1 Failure" Check the duration for opening the span gas valve (See 7.2.1.5) and zero gas valve (See 7.2.1.4). Check the Zero- and/or Spancalibrationgas- supply. Check the position of valves and regulators in the supply and discharge lines for the calibration gas. Check the value for zerogas (see 5.3.2.1) and spangas. (see 5.3.2.2) Check the setting for the max dierence. (see 7.2.1.6) Check the sensor for contamination and correct mounting. Check the Zero- and/or Spancalibrationgas- supply. Check the position of valves and regulators in the supply and discharge lines for the calibration gas. Check the value for zerogas (see 5.3.2.1) and spangas. (see 5.3.2.2) Check the setting for the max dierence. (see 7.2.1.6) Check the sensor for contamination and correct mounting. Manual A4009T-O2 analyser R01-EN 20 / 33

6.4. STORING 6.4 Storing Symptom Cause Solution No indication on the display Oxygen indication is not correct.. 4-20 ma value on analog output is wrong. No reaction on display when a button is pressed. No power supply to Display/measurement unit.. Display/measurement unit defect. Insucient gas ow to the sensor. Sensor is contaminated. Wrong calibration. Conguration of the analog output is not correct. Signalreceiver in the 4-20 ma loop is congured wrong. Load on the 4-20 ma circuit is to high. Button has no function in the current menu. Button is defect. Check the power supply. Check the wiring/connections Check the sensor cable for correct wiring and damage. Contact your distributor Check the gas ow to the sensor. Check th position of valves and regulators. Remove the sensor and check for contamination. In case of severe contamination, repalce the sensor and check lters in the supply line. Calibrate the analyser. Contact your distributor. Check the conguration of the Analog output. Check the setting of the Min and Max value in the Analog output menu. Check the conguration of the signalreceiver. Does this match the settings in the oxygen analyser. Check the power supply on the analog output. Contact your distributor. Check if the button has a function in the menu. Contact your distributor. Manual A4009T-O2 analyser R01-EN 21 / 33

Chapter 7 Automatic Calibration The automatic calibration allows for periodic calibration of the analyzer without human interaction. In this way the it is guaranteed the analyzer measures correctly. 7.1 Installation. To use the automatic calibration function 2 extra valves are required to switch on and o the span and zero gas supply to the sensor. Also an electrical 24VDC signal is required to start the automatic calibration procedure. Warning: Make sure the process where the analyzer is used allows automatic calibration. automatic calibration the analyzer is not available. The 4-20 ma current loop output and alarm contacts will keep there function during automatic calibration. During the Manual A4009T-O2 analyser R01-EN 22 / 33

7.1. INSTALLATION. 7.1.1 Connection diagram automatic calibration valves. Figure 7.1: Process connections automatic calibration. 7.1.2 Electrical connection for automatic calibration Figure 7.2: Electrical connections automatic calibration. Manual A4009T-O2 analyser R01-EN 23 / 33

7.2. CONFIGURATION 7.2 Conguration Before the automatic calibration function can be used it must be congured. 7.2.1 Setting the auto calibration parameters. 7.2.1.1 Selecting the automatic calibration menu. Display 20.94 % When the measurement is displayed: Press the <+> button for 10 seconds until the "Automatic calibration"menu appears. < Auto Calib. > OFF 7.2.1.2 Auto cal parameters Display Parameter MIN MAX TYP Enable automatic calibration OFF ON OFF < Auto Calib > OFF <+>_ <->^ <Zero Gas Time > 0010 s Time the zero gas valve is opened. (sec) 0 9999 10 <+>_ <->^ <Span Gas Time > 0010 s Time the span gas valve is opened. (sec) 0 9999 10 <+>_ <->^ <Difference MAX> O2 02.00% val Maximum dierence between the measured oxygen content during calibration and the set values. 0.0 99.9 2.0 Manual A4009T-O2 analyser R01-EN 24 / 33

7.2. CONFIGURATION 7.2.1.3 Ebable-/disable automatic calibration Display < Auto Calib. > OFF Auto Calib. OFF Auto Calib. ON < Auto Calib. > ON Select "Auto Calib." option in the "automatic calibration"-menu (see 7.2.1.1) Press <V> to change the value. Press <-> to disable automatic calibration press <+> to enable automatic calibration. Conrm the change by pressing the <V> button The value is changed 7.2.1.4 Setting the zero gas time. Display <Zero Gas Time > 0010 s Zero Gas Time 0010 s... Zero Gas Time 0010 s <Zero Gas Time > 0010 s Select the "Zero Gas Time" option in the "automatic calibration"-menu (see 7.2.1.1) Press <+> to select zero gas time and press <V> to change the value. See chapter 4.2.1 for a description how to change the value. Press the <V> button. The value is changed. Manual A4009T-O2 analyser R01-EN 25 / 33

7.2. CONFIGURATION 7.2.1.5 Setting the Span Gas Time Display <Span Gas Time > 0010 s Span Gas Time 0010 s... Span Gas Time 0010 s <Span Gas Time > 0010 s Select the "Span Gas Time" option in the "automatic calibration"-menu (see 7.2.1.1) Press <+> <+> to select span gas time and press <V> to change the value. See chapter 4.2.1 for a description how to change the value. Press the <V> button. The value is changed. 7.2.1.6 Setting the maximum dierence from the calibration gas. Display <Difference MAX> O2 02.00% val Difference MAX O2 02.00% val... Difference MAX O2 02.00% val <Difference MAX> O2 02.00% val Select the "Dierence Max" option in the "automatic calibration"-menu (see 7.2.1.1) Press the <V> button to change the value. See chapter 4.2.1 for a description how to change the value. Press the <V> button. The value is changed. 7.2.2 Starting automatic calibration. Automatic calibration is started by applying a 24VDC signal to digital input 2. The signal must be present for at least 1 second. see 7.1.2 Calibration sequence is as indicated in the ow diagram below. Manual A4009T-O2 analyser R01-EN 26 / 33

7.2. CONFIGURATION Figure 7.3: Flow diagram automatic calibration Manual A4009T-O2 analyser R01-EN 27 / 33

Appendix 1: Dimension Manual A4009T-O2 analyser R01-EN 28 / 33

Appendix 2: Specication General # Parameter Value Unit Remark 1 Measuring range 0-21 % O 2 2 Acuracy < 0.2 % O 2 3 Linearity < 0.2 % O 2 4 Stabilisatietijd < 4 s 5 Reaction time < 1 s 6 Warm up time 5 min 30 min for full acuracy 7 Storage temperature -20-70 C 8 Operating temperature 0-70 C 9 Max humidity 96 % non condensing 10 Protection IP20 11 Dimensions 96 x 96 x 163 mm l x w x d 12 Weight 920 gr Tabel 7.1: Analyzer specication. Power Supply Parameter Min Nom Max Unit Remark 1 Supply Voltage 19 24 32 VDC 2 Current 0.2 0.7 1.3 A Tabel 7.2: Power Supply Manual A4009T-O2 analyser R01-EN 29 / 33

Appendix 2: Specication Digital Inputs # Parameter Value Unit Remark 1 Count 2 pcs 2 Type Optical Isolated - 3 Nominal voltage 24 VDC 4 Maximum voltage 32 VDC 5 Logic '1' > 7.5 VDC 6 Logic '0' < 4.5 VDC 7 Isolation voltage 1000 VAC 50/60 Hz 8 Isolation resistance > 1 MΩ Tabel 7.3: Parameters digital input Relais Contacts # Parameter Value Unit Remark 1 Count 2/3 pcs 2 Type Relais SPST-NO/SPDT - 3 Max voltage 250 VAC 32 VDC 4 Max. current 0.4 AAC 1.0 ADC 5 Isolation voltage 1000 VAC 50/60 Hz 6 Isolation resistance > 1 MΩ 7 Function High oxygen alarm - (*1 High-high oxygen alarm - Low oxygen alarm - (*1 Low-low oxygen alarm - Analyserfailure - Auto Cal ref gas valve - (*1 Auto Cal zero gas valve - (*1 Tabel 7.4: Parameters relay contacts. *1) When the automatic calibration is activated the high and low oxygen alarm contacts are no longer available. These contacts are used for controlling the calibration gas valves. Manual A4009T-O2 analyser R01-EN 30 / 33

Appendix 2: Specication Analog Qutputs # Parameter Value Unit Remark 1 Count 2 pcs 2 Type 4-20mA 2-wire current loop 3 Range Programmable 4 Function Measured O 2 % 5 Minimum supply voltage 7 VDC 6 Maximum supply voltage 32 VDC 7 Current 4-20 ma 8 Isolation voltage 1000 VAC 50/60 Hz 9 Isolation resistance > 1 MΩ 10 Protection Wrong connection Overvoltage Tabel 7.5: Parameters analog output. Manual A4009T-O2 analyser R01-EN 31 / 33

Appendix 2: Specication Sensor # Parameter Value Unit Remark 1 Supplier DIVIZE 2 Type B4009S01 3 Order reference B4009S01 4 Sample gas temperature 0-450 C 5 Sample gas pressure Ambient 6 Sensortype Planar ZrO 2 dual cell 7 Heater 8 Weight 9 Dimensions 10 Cable length 11 Max temp housing 12 Isolation resistance Integrated 120 g without cable 84 x 22Ø mm 800 mm C Depending on sample gas temperature 12.1 Heater Sensor > 1 MΩ 12.2 Heater Housing > 1 MΩ 12.3 Housing Sensor > 1 MΩ @ 25 C temp. 13 Mounting 13.1 Thread 13.2 Mounting torque 14 Electrical connections M18 x 1.5 40-60 Nm 14.1 Number of wires 6 14.2 Connection cablel K4009A01 Tabel 7.6: Sensor specication Manual A4009T-O2 analyser R01-EN 32 / 33

DIVIZE b.v. Boezemweg 23e 2641KG Pijnacker The Netherlands Tel: +31 (0)88-7733733 Fax: +31 (0)88-7733737 www.divize.com info@divize.com Revision 01-EN : November 25, 2014