Quality Control of Pulmonary Diagnostic Systems

Similar documents
Independent Health Facilities

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

EasyWarePro Message Numbers

EasyWarePro Message Numbers

Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction

EasyOne Pro LAB Measurement Technology Background

PULMONARY FUNCTION TESTING CORE

Single-Breath Diffusing Capacity for Carbon Monoxide Instrument Accuracy Across 3 Health Systems

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Understanding the Issues Affecting the Accuracy of Measuring Oxygen Consumption Utilizing Metabolic Carts

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

Public Assessment Report Scientific discussion. Lung test gas CO (He) AGA, 0.28%, inhalation gas, compressed (carbon monoxide, helium) SE/H/1154/01/MR

A Pulmonary Function Product of Rare Distinction

HARD (Gas Dilution) HARDER (Plethysmography) EASIEST (MiniBox+)

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

Thermo Environmental 48C Carbon Monoxide QA Plan Section I Electronic Calibration Branch Responsibilities

Body Box 5500 Series Pulmonary Function Testing Systems

Stikstofuitwas en Heliumdilutie. Eric Derom Universitair Ziekenhuis Gent

VIC offers a variety of Calibrated Gas Leaks

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

Medical Instruments in the Developing World

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

OPERATING PROCEDURE. Number: V2. Authors. Approvals

MasterScreen PFT. Powered by SentrySuite

Improve Process Reliability

Susan Blonshine, RRT, RPFT, AE-C, FAARC

Transferpette. Lightweight precise rugged reliable!

Unit 15 Manual Resuscitators

Central scientific research automobile and automotive engine institute

DXH Troubleshooting FlowCharts

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Part 1: Inspiratory and expiratory pressures

INSPECTION OF RECOVERY BOILERS

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Draeger Fabius GS Pre-Use Check

Transferpette. Lightweight precise rugged reliable!

Test lungs from IMT Analytics simulate a range of realistic ventilation situations during training.

Instruction Manual No. 742, 8/98

Recommendations on Checking Anaesthesia Delivery Systems

Instruction Manual 742 5/1/2009. Eclipse Ratio Regulators ES Series Version 1

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

SF 6 Product Guide. Gas Analysis Instruments and Accessories

Applications Note: Use of "pentane equivalent" calibration gas mixtures

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

Errors in Monitoring. BWH Clinical Conference 10/06/04. Copyright 2004, James H Philip, all rights reserved.

INSTRUCTOR RESOURCES

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Technical Procedure for General Laboratory Equipment

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

2017 ERS/ATS standards for single-breath carbon monoxide uptake in the lung

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Model Description

Boyle s Law: Pressure-Volume Relationship in Gases

DeZURIK Double Block & Bleed (DBB) Knife Gate Valve Safety Manual

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

DeZURIK. KSV Knife Gate Valve. Safety Manual

Human Respiration Laboratory Experiment By

Proficiency Testing Corrective Action Checklist: PT PROVIDER: PT EVENT: TEST:

Application Note AN-107

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Calibration Requirements for Direct Reading Confined Space Gas Detectors

OHD L-47. Guidelines for Resolving Differences in Test Results

PQAO Training 2017 Karl Tupper, San Luis Obispo APCD

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

Cartridge Based & Point-Of-Care Blood Gas Analyser QAP Performances & Quality Management Systems.. Issues

RAM 4021 Operation Manual

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

Modeling Diffusion Rates of a Gas in an Enclosed Space

VT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features

February 22, Dear Jerry Masiello:

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing

Quantitative Analysis of Hydrocarbons by Gas Chromatography

GASES FIRE & RESCUE SPECIALTY. Gases & Equipment. Medical Gases Liquid & High Pressure Oxygen U.S.P. Nitrogen N.F. Carbon Dioxide U.S.P.

Supplemental Body Plethysmograph Service Guide Version/Revision 1.A ComPAS ver. 1.9

TECHNICAL BULLETIN. TriControls. Control Materials for Liquid QC and Calibration Verification

A review of best practices for Selection, Installation, Operation and Maintenance of Gas meters for Flare Applications used for Managing facility

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

AUTOMATIC FLOW CARTRIDGES. Competition Analysis Not All Automatics Are Created Equal

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

DeZURIK. KGC Cast Knife Gate Valve. Safety Manual

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Lab 1. Adiabatic and reversible compression of a gas

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

RAM Operation Manual

SL G3. Warning

ASTM UPDATE to COQA. COQA Meeting Feb San Antonio. RJ (Bob) Falkiner SR (Steve) David GJ (Jay) Grills W (Wayne) Kriel

Data Mining Data is logged every 3 hours 11 parameters : 1. Number of call 2. Correspondents and calls pattern 3. Call duration 4. Circle of friend 5.

Topics For Discussion Acadiana Gas Measurement School

Gas Laws: Boyle s and Amonton s Laws Minneapolis Community and Technical College v.9.08

Transcription:

Quality Control of Pulmonary Diagnostic Systems James P. Sullivan, BA, RPFT Pulmonary Diagnostic Laboratories Memorial Sloan Kettering Cancer Center New York, New York QA, QC and QI: what s the difference? Quality Control: systematic use of methods to ensure the device conforms to standards. Quality Assurance: does the device meet requirements? Quality Improvement: programs to improve or enhance the performance of a device or procedure. Continuous Quality Improvement: ongoing QI programs. Why are QA and QI Important? If it moves air, we can get a number 1

Mechanical Devices: Calibration Syringes: Can be used for both calibration and control for flow volume, gas dilution and DLco. Mechanical Simulator Devices: For flow, VTG and DLco. Calibration vs. Control: Calibration: create device correction factors Control: verify accuracy of correction factors Human Controls: Human Biologic Standard (HBS): Also referred to as bio QC Should be healthy non smokers asthmatics need not apply Both mechanical and human controls should be run when troubleshooting equipment problems, and again to confirm proper (accurate) operation after repair. Human Controls: pros Requires no additional equipment or set up time. Works on all measurements. Checks the entire system. The absolute best way to really learn how to instruct your patients and operate your system. Almost all labs have a suitable HBS QC subject. 2

Human Controls: cons Not as reproducible and consistent; up to 5% variance is acceptable, which is not good enough Not all technologists are suitable subjects for HBS QA Diurnal variations may limit reliability Technologist resistance Mechanical Devices: pros All labs have a calibration syringe. Highly reproducible and consistent. Mechanical Devices: cons May be costly and complicated. May not check the entire system. What if the device itself is faulty? How is the accuracy of the device checked? 3

The ideal ongoing QA program will be a combination of both human and mechanical controls. Calibration Syringes: 3 L is the common standard. All labs should have at least two 3 L syringes, one to calibrate and one to verify. 5 L is very desirable, because this volume more closely approximates more patients. 7 L is best. Syringes should be serviced (cleaned, lubricated and calibrated) as per manufacturer s recommendations. Syringe connectors should be replaced regularly. Calibration Syringes: FVC: perform FVL at clinically relevant flows; should return syringe volume +/ 3%, i.e., 2.91 3.09 L using 3 L syringe. FRC He or N 2 : perform FRC measurement with different volumes; should return syringe volume plus deadspace. D L co: D L co should be essentially zero; this procedure essentially checks for leaks in the system. VA should be syringe volume plus deadspace. 4

Morgan Scientific VTG Simulator: SensorMedics Plethysmograph Simulator Isothermal Lung Analog Ruppel s Manual of Pulmonary Function Testing, 9 th Edition 5

Hans Rudolph: Flow Volume Simulator and Resistance Standards: Hans Rudolph D L co Simulator: Reference Literature All pulmonary laboratories should have various reference literature available, including, but not limited to: Department policy and procedure manuals Vendors operating manuals All installation records, including initial control studies Corrective action logs ATS/ERS Statements and Lab Manual AARC CPGs Recent pulmonary diagnostics text (Ruppel/Mottram, Wanger, Wasserman, etc.) 6

Human Biologic Standard Controls Collect at least 20 sets of ATS compliant data for each control parameter. To be done on properly maintained and functioning analyzer after a successful calibration. Usual control parameters are FVC, FEV 1, FRC, TLC, DLco and VA. Attempt to perform the baseline studies at different times of the day to better duplicate real world conditions. Human Biologic Standard Controls Calculate mean and SD for each parameter. Enter data onto spreadsheet for calculation and graphing. In a normal data distribution: 65% of data will be within ±1 SD of the mean, 95% of data will be within ±2 SD of the mean, 99% of data will be within ±3 SD of the mean. 7

Human Biologic Standard Controls General QA Rules: ±2 SD are considered warning limits. Values between 2 and 3 SD indicate an error and the procedure should be repeated. Values beyond ±3 SD are unacceptable and troubleshooting procedures should commence. FVC Lung Volumes 8

DLco Westgard Rules 9

FVC This method uses both the tracer gas and carbon monoxide in the DLco mixture to calculate the VA. This is done by calculating the ratios of the expired to inspired gases (both the tracer and CO). A calibration syringe s volume at full extension is always the same. Calibration syringes dilute; they do not diffuse. The expected result for DLco at all stroke volumes is zero. 10

The expected result for VA (hereafter called target volume) is: the overall syringe volume, plus the adapter deadspace, plus the instrument deadspace (including the filter, if used). This is true no matter what the starting syringe volume is. If the stroke starts with the syringe completely closed, the test gas will be diluted only by the deadspace of the syringe, connector and instrument. If the syringe plunger is pulled out halfway before it is attached to the device, the dilution will be 50% plus the deadspace of the syringe, connector and instrument. VA should be the same whether the expired to inspired CO or tracer ratio is used. IVC ( ATPS) expired gas inspired gas This is the formula for the target volume (ATPS): We can calculate the target volume by measuring: the inspired volume, and, the difference between the expired and expired gases. If the IVC is too low or high, i.e., a flow measurement issue, the VA will be off in the same direction. If the IVC is within range and the VA is high a leak or a malfunctioning valve permitting excess dilution. If the VA calculated by tracer dilution gives an accurate VA result and the VA measured by CO dilu on does not problem with the CO analyzer. If the VA measured by CO is accurate and the tracer measurement is not problem with the tracer analyzer. 11

Initial work on this method began in 2011. At least two systems from CareFusion, MGC Diagnostics, Morgan Scientific and nspire were used. The 5 L syringe on the Hans Rudolph DLco Simulator was used. Advantages over Hans Rudolph DLco Simulator: More sensitive in detecting analyzer alinearity. Much less expensive and complex. Can better detect flow and individual analyzer problems. Known Limitations: The syringe must be removed from the system between trials and completely flushed (same with the HR device). Deadspace of syringe must be known. Second collar that is adjustable is needed on syringe shaft. Syringe must be precisely set for each volume. The greater the gas dilution, the less sensitive this method is. 12

Work to be done: Development of an adapter that increases mixing of test and deadspace gases in the syringe. Spreadsheet with correction factors for current pulmonary systems completed. Questions or Comments? James P. Sullivan, BA, RPFT Supervisor, Pulmonary Diagnostic Laboratories Memorial Sloan Kettering Cancer Center New York, New York (212) 639 8399 sullivaj@mskcc.org 13