Spring Uses in Exoskeleton Actuation Design

Similar documents
EXPERIMENTAL STUDY OF EXOSKELETON FOR ANKLE AND KNEE JOINT

An investigation of kinematic and kinetic variables for the description of prosthetic gait using the ENOCH system

REPORT DOCUMENTATION PAGE

3 people 3 unique lifestyles 3 advanced foot solutions

Body Stabilization of PDW toward Humanoid Walking

Conceptual Design of an Energy Efficient Transfemoral Prosthesis

empower Reclaim your power. Information for technicians empower Ottobock 1

Actively Controlled Lateral Gait Assistance in a Lower Limb Exoskeleton*

Artifacts Due to Filtering Mismatch in Drop Landing Moment Data

An autonomous, underactuated exoskeleton for loadcarrying augmentation

Using GPOPS-II to optimize sum of squared torques of a double pendulum as a prosthesis leg. Abstract

Microprocessor Technology in Ankle Prosthetics

Gait analysis for the development of the biped robot foot structure

C-Brace Orthotronic Mobility System

Empower. Reclaim your power. Information for technicians. Empower Ottobock 1

-Elastic strain energy (duty factor decreases at higher speeds). Higher forces act on feet. More tendon stretch. More energy stored in tendon.

Controlling Walking Behavior of Passive Dynamic Walker utilizing Passive Joint Compliance

ZMP Trajectory Generation for Reduced Trunk Motions of Biped Robots

Simulation of the Hybtor Robot

Power Assessment of the Human Ankle during the Stance Phase of Walking for Designing a Safe Active Prosthesis in Below-Knee Amputees

Toward a Human-like Biped Robot with Compliant Legs

A Bio-inspired Behavior Based Bipedal Locomotion Control B4LC Method for Bipedal Upslope Walking

Inertial compensation for belt acceleration in an instrumented treadmill

Brief Biomechanical Analysis on the Walking of Spinal Cord Injury Patients with a Lower Limb Exoskeleton Robot

A MODIFIED DYNAMIC MODEL OF THE HUMAN LOWER LIMB DURING COMPLETE GAIT CYCLE

A New Approach to Modeling Vertical Stiffness in Heel-Toe Distance Runners

A CONTINOUS ROTARY ACTUATION MECHANISM FOR A POWERED HIP EXOSKELETON

Computer Aided Drafting, Design and Manufacturing Volume 26, Number 2, June 2016, Page 53. The design of exoskeleton lower limbs rehabilitation robot

A Novel Gear-shifting Strategy Used on Smart Bicycles

HRC adjustable pneumatic swing-phase control knee

Dynamically stepping over large obstacle utilizing PSO optimization in the B4LC system

The technique of reciprocal walking using the hip guidance orthosis (hgo) with crutches

Mechanical Knee Prescription Guide

Adaptation to Knee Flexion Torque Assistance in Double Support Phase

Motion Control of a Bipedal Walking Robot

Learning Energy Efficient Walking Based on Ballistics

Velocity-dependent reference trajectory generation for the LOPES gait training robot

A Passive Exoskeleton with Artificial Tendons

Finite-State Control of Powered Below-Knee Prosthesis with Ankle and Toe

Kungl Tekniska Högskolan

Centre for Autonomous Systems

Gait Analyser. Description of Walking Performance

GROUND REACTION FORCE DOMINANT VERSUS NON-DOMINANT SINGLE LEG STEP OFF

Simulation-based design to reduce metabolic cost

+ t1 t2 moment-time curves

INTERACTION OF STEP LENGTH AND STEP RATE DURING SPRINT RUNNING

Toward a Human-like Biped Robot with Compliant Legs

A Biomechanical Approach to Javelin. Blake Vajgrt. Concordia University. December 5 th, 2012

Steffen Willwacher, Katina Fischer, Gert Peter Brüggemann Institute of Biomechanics and Orthopaedics, German Sport University, Cologne, Germany

IAC-04-IAA DESIGN OF A HIGH-TENSION ELASTICALLY DEFORMING SPACE TETHER DEPLOYER

G-EOL. Discover the simplicity of gait therapy intended for daily use

Increasing ankle push-off work with a powered prosthesis does not necessarily reduce metabolic rate for transtibial amputees

Dynamic Warm up. the age of the athlete current physical condition and prior exercise experience

Normal and Abnormal Gait

Myths and Science in Cycling

PURPOSE. METHODS Design

Ecole doctorale SMAER Sciences Mécaniques, Acoustique, Electronique, Robotique

A NEW GOLF-SWING ROBOT MODEL UTILIZING SHAFT ELASTICITY

Learning Energy Efficient Walking with Ballistic Walking

Powered Bipeds Based on Passive Dynamic Principles

Effects of Ankle Stiffness on Gait Selection of Dynamic Bipedal Walking with Flat Feet

Biomechanics and Models of Locomotion

The Starting Point. Prosthetic Alignment in the Transtibial Amputee. Outline. COM Motion in the Coronal Plane

Serve the only stroke in which the player has full control over its outcome. Bahamonde (2000) The higher the velocity, the smaller the margin of

KICKBIKE Your key to optimum sports performance

USA Track & Field Heptathlon Summit- November

OPTIMAL TRAJECTORY GENERATION OF COMPASS-GAIT BIPED BASED ON PASSIVE DYNAMIC WALKING

Analysis of ankle kinetics and energy consumption with an advanced microprocessor controlled ankle foot prosthesis.

Posture influences ground reaction force: implications for crouch gait

Available online at Prediction of energy efficient pedal forces in cycling using musculoskeletal simulation models

CHAPTER IV FINITE ELEMENT ANALYSIS OF THE KNEE JOINT WITHOUT A MEDICAL IMPLANT

A MATHEMATICAL MODEL TO DETERMINE THE TORQUE FOR A PROSTHETIC LEG-LAGRANGIAN EQUATION

Skippy: Reaching for the Performance Envelope

From Passive to Active Dynamic 3D Bipedal Walking - An Evolutionary Approach -

Analysis of Backward Falls Caused by Accelerated Floor Movements Using a Dummy

Define terms and compute basic physics problems related to sprinting

1B1 Meridium. Reclaim your way. Information for practitioners. Meridium Ottobock 1

LOCOMOTION CONTROL CYCLES ADAPTED FOR DISABILITIES IN HEXAPOD ROBOTS

In this course you will learn the following

Walking Simulator Mechanism

Experimental Modeling and Control of Pneumatic Cylinders for Robotic Applications

Managing and Recycling Human Energy: A Mechanical Redesign of the UCSC Lower Limb Exoskeleton. Rachel Rieger, Jacob Rosen

A Neuromuscular Model of Human Locomotion and its Applications to Robotic Devices

Spider Robot for Motion with Quasistatic. Force Constraints

The Hurdle Events. Jeff Martin IATCCC Clinic. Indianapolis, Indiana. 5 Myth s about Hurdling

PERCEPTIVE ROBOT MOVING IN 3D WORLD. D.E- Okhotsimsky, A.K. Platonov USSR

Active Orthosis for Ankle Articulation Pathologies

Humanoid Robots and biped locomotion. Contact: Egidio Falotico

Assessments SIMPLY GAIT. Posture and Gait. Observing Posture and Gait. Postural Assessment. Postural Assessment 6/28/2016

Journal of Chemical and Pharmaceutical Research, 2016, 8(6): Research Article. Walking Robot Stability Based on Inverted Pendulum Model

Joint Torque Evaluation of Lower Limbs in Bicycle Pedaling

Design of Variable-Damping Control for Prosthetic Knee based on a Simulated Biped

Florida. Published in: IEEE ICRA. possibly. reversed. augmentationn or. endurance) only connection them unsuitable for applications.

The Incremental Evolution of Gaits for Hexapod Robots

Compliance for a cross four-bar knee joint

Locomotion Concepts. Autonomous Mobile Robots. Concepts Legged Locomotion Wheeled Locomotion. Autonomous Systems Lab. Zürich. Localization.

Normal Gait and Dynamic Function purpose of the foot in ambulation. Normal Gait and Dynamic Function purpose of the foot in ambulation

Identify and explain how specific kinematic and kinetic elements relate to the absolute speed technical model

Emergent walking stop using 3-D ZMP modification criteria map for humanoid robot

HIP-KNEE control for gait assistance with Powered Knee Orthosis

Transcription:

211 IEEE International onference on Rehabilitation Robotics Rehab Week Zurich, ETH Zurich Science ity, Switzerland, June 29 - July 1, 211 Spring Uses in Exoskeleton Actuation Design Shiqian Wang, Wietse van Dijk Biomechanical Engineering Dept. Delft University of Technology (TU Delft) Delft, the Netherlands shiqian.wang@tudelft.nl Herman van der Kooij Biomechatronics and Rehabilitation Technology University of Twente Enschede, the Netherlands Biomechanical Engineering Dept. Delft University of Technology, Delft, the Netherlands h.vanderkooij@utwente.nl Abstract An exoskeleton has to be lightweight, compliant, yet powerful to fulfill the demanding task of walking. This imposes a great challenge for the actuator design. Electric motors, by far the most common actuator in robotic, orthotic, and prosthetic devices, cannot provide sufficiently high peak and average power and force/torque output, and they normally require high-ratio, heavy reducer to produce the speeds and high torques needed for human locomotion. Studies on the human muscle-tendon system have shown that muscles (including tendons and ligaments) function as a spring, and by storing energy and releasing it at a proper moment, locomotion becomes more energy efficient. Inspired by the muscle behavior, we propose a novel actuation strategy for exoskeleton design. In this paper, the collected gait data are analyzed to identify the spring property of the human muscle-tendon system. Theoretical optimization results show that adding parallel springs can reduce the peak torque by 66%, 53%, and 48% for hip flexion/extension (F/E), hip abduction/adduction (A/A), and ankle dorsi/plantar flexion (D/PF), respectively, and the rms power by 5%, 45%, and 61%, respectively. Adding a series spring (forming a Series Elastic Actuator, SEA) reduces the peak power by 79% for ankle D/PF, and by 6% for hip A/A. A SEA does not reduce the peak power demand at other joints. The optimization approach can be used for designing other wearable robots as well. Keywords-exoskeleton; actuation; spring; SEA; PEA; wearable robots I. INTRODUTION By 25, more than 25, people in the United States and about 2.5 million people all over the world, according to a conservative estimate, were living with spinal cord injury (SI) induced paralysis [1]. And the number increases every year. Research has been conducted to mitigate the negative effects on the patient s personal and social life. Some exoskeletons have been designed to help the SI patients or stroke survivors walk again. ReWalk, claims to be the first commercially viable upright walking assistance tool, and enables wheelchair users with lower-limb disabilities to stand, walk, and even climb stairs [2]. Berkeley Bionics claims that with their newly developed elegs system, a walking speed of.89 m/s is attainable [3]. Both ReWalk and elegs weigh about 2 kg, which is relatively lightweight, but for both systems the users cannot free their hands and have to use crutches to maintain stability and compensate for the power insufficiency at the actuators. The REX robotic exoskeleton from REX bionics has five actuated degrees of freedom (DoFs) at each leg, and enables the user to walk, make turns, and even climb stairs and slopes [4]. Based on observation, the walking speed of the REX exoskeleton is slower than elegs and its weight is about 38 kg. All three exoskeletons use electric motors due to their excellent controllability and ease of supplying power, compared with other actuators such as hydraulic and pneumatic actuators. However, in most of the current designs, the relative low power and torque density of the electric motor result in either a heavy or a functionally limited system. Other actuators with high power and force densities either are difficult to control or too noisy to wear, or require auxiliary energy supply [5]. The goal of this paper is to investigate the potential reduction in power and torque demands on actuators (mainly electric motors) by using series elastic actuation (SEA) and parallel elastic actuation (PEA) in the exoskeleton (Fig. 1). Only linear spring will be considered at this stage. Both SEA and PEA have been used for walking robots and wearable robots. SEAs, due to their low output impedance, high force fidelity and excellent controllability, and energy storing capability [6], are the most common actuators in robotic, orthotic, and prosthetic devices. RoboKnee and Legged-Robot from Yobotics [7][8], the Robotic Tendon from Arizona State University (ASU) [11], and Powered Ankle-Foot Prosthesis from MIT [12], use linear series elastic actuators (L-SEA) to power the knee or the ankle. Rotary series elastic actuator (R- SEA) is used as well, e.g., the mobility assist exoskeleton from Florida Institute for Human and Machine ognition (IHM) [13] and eseaj from University of Twente [14]. In [11], researchers claimed that, in theory, adding a series spring to the ankle actuator could reduce the peak motor power from 25W to 77W for an 8 kg person and a much lighter motor could be used. Fewer applications of PEA than SEA are employed. For M Figure 1. Parallel Elastic Actuation (PEA, left) and Series Elastic Actuation (SEA, right) M 978-4244-9861/11/$26. 211 IEEE 891

example, parallel elasticity is employed in the Powered Ankle- Foot Prosthesis from MIT [12] [16] and the ViSHaRD4 robotic arm from TU Munich [15]. In the Ankle-Foot Prosthesis, parallel elastic actuation is formed. The parallel spring reduces the load borne by the torque source (SEA, in this case) and improves the walking metabolic economy [16]. In general, adding a series elastic element reduces the peak power demand on the motor. Therefore the motor size can be reduced. The direct purpose of adding a parallel elastic element is to reduce the torque demand on torque source (e.g., the gear and motor system). In fact, both the motor size and gear ratio can be reduced. Reduced torque cuts down the input current and copper loss (I 2 R) and reduces energy consumption. In this paper, we first present the gait data analysis in section II and then explain the optimization methods for SEA and PEA in section III. The optimization results are represented and discussed in section IV. II. GAIT DATA ANALYSIS AND RESULTS A. Experiment setup and data processing Eight subjects (four females and four males), with no gait abnormality, were instructed to walk along a 15-m walkway at different speeds (slow.8 m/s and normal 1.2 m/s). The subjects are 24.3±1.4 year old, 18±.1 cm tall, and weigh 68.8±12.1 kg (mean value ± standard deviation). 3D kinematic data from the movements and ground reaction forces from force plates were recorded simultaneously, with the VION MX system (Oxford metrics, Oxford, United Kingdom), and two force plates (AMTI, Watertown, Mass., United States). The raw data were first post-processed (e.g. filtering the highfrequency noise), and inverse dynamics was performed to obtain the joint torques, and finally the joint powers were calculated. Vel (deg/s) Due to the repetitive and cyclical property of walking - Sagittal RHIP.8m/s 1.2m/s 2-2 Sagittal RKNE movement, some commonly used techniques were employed to process the data [17]. Intra- and inter-subject averaged profiles have been used. The logic behind this is that measurements from different subjects show very similar trends and patterns. In the following section, only averaged joint angle, velocity, torque, and power are shown. By assuming the gait data can be linearly scaled according to subject s body weight, torques and powers were normalized by the subject s body weight (mass). The time has been normalized to the gait cycle time. A gait cycle is divided into phases according to the placement of a foot and the interaction between foot and ground. Four events, heel strike (HS), foot flat (FF), heel off (HO), and toe off (TO) divide the gait cycle into four consecutive phases accordingly, early stance, mid-stance, late-stance, and swing phase. In the exoskeleton we are developing, four degrees of freedom (DOFs) will be actuated: the hip flexion/extension (F/E), hip abduction/adduction (A/A), knee flexion/extension (F/E), and ankle dorsi/plantar flexion (D/PF). B. Temporal profiles The processed gait data for two walking speeds (slow walking.8 m/s, solid line, and normal walking 1.2 m/s, dash line) are shown in Fig. 2. Similar gait profiles (i.e., kinematic and kinetic patterns) were observed for both slow and normal walking. Peak powers and torques increase as walking speed increases. The late-stance time (interval between HO and TO) increases with speed. Gait data ranges at.8 m/s (TABLE I), were scaled to a body weight of 13 kg. 13 kg reflects the maximal weight for subjects who will be included and the maximal weight of the exoskeleton (3 kg). The peak torque and power, especially those at the ankle, indicate strong demand on the motor power and the torque capacity of the gear and motor. Slow/normal walking gait data in sagittal plane were compared to the gait data collected by D. Winter [17]. Similar - -2 Sagittal RANK 2-2 -4-6 Frontal RHIP Ang (deg) FF FF FF 2 1 5 FF -2 1-4 FF -6 FF FF -5 FF Torq (Nm/kg).5 -.5 1.5 -.5.5 -.2 -.4 -.6 Power (W/kg).5 -.5.2.4.6.8 1.5 -.5.5.2.4.6.8 1 2 1.2.4.6.8 1.2 -.2.2.4.6.8 1 Figure 2. Inter-subject averaged gait data. Torques and powers are normalized to body mass; the time is normalized to gait cycle time; Sagittal and Frontal indicate human anatomical planes; RHIP = right hip; RKNE = right knee; RANK = right ankle; the same in other figures. 892

patterns were observed.. Joint torque-angle curves To obtain the stiffness characteristics of individual joints, joint torques were plotted against joint angles (Fig. 3). The stiffness K is the tangent of the torque-angle curve. Κ = dτ/dθ J. (1) where T is the joint torque and θ J is the joint angle. TABLE I. GAIT DATA RANGES FOR A 13 KG SUBJET WALKING AT.8 M/S Sagittal plane Frontal plane Quantities HIP (F/E) Knee (F/E) Ankle (D/PF) Hip (A/A) Min Max Min Max Min Max Min Max Angle Velocity [deg./s] -66 114-218 275 54 14-48 29 Joint Angle [deg.] -9.5 24.6-6.3 2.6-7. 13.5-5. 4.9 Joint Torque [Nm] -37.1 8.6-24.2 94.4 96 13.3-97. 2.3 Joint Power [W] -39.3 43.9 14 34.3-95 217.2-28. 36.4 During walking, the hip functions as a nonlinear spring in sagittal plane. From HS to HO (early stance and mid-stance phase) the hip spring is loaded, and from HO to end of the stride (late-stance and swing phase), the hip is unloaded. The loading and unloading curves overlap each other. The knee functions as a linear spring during early- and mid-stance phase. And the spring characteristic is altered after the heel is lifted. The ankle works as a linear (or very close to) spring during stance phase. After the heel is lifted (HO), the spring stiffness is changed slightly. The torque-angle relation suggests that parallel springs (either linear or nonlinear) could be used to store the energy..5 -.5 D Sagittal RHIP D 1 2 D -.5.5.8(m/s) Sagittal RANK B B D A A.8(m/s) -5 5 1 A B A B 1.5 D Sagittal RKNE B B A A -6-4 -2 D -.2 -.4.8(m/s) Frontal RHIP -.6.8(m/s) -5 5 Figure 3. Torque-angle curves. At different walking speeds (.8 m/s and 1.2 m/s), torque-angle curves differ slightly, but exhibit similar patterns. A A B B By comparing the torque-angle curves at these two walking speeds, we can make a preliminary conclusion that, when walking speed varies in the range of.8~1.2 m/s, speed influences on joint stiffness are relative small. The spring stiffness is not necessarily to be adaptive to speed for slow and normal walking. III. OPTIMIZATION METHOD Before explaining the optimization method, the assumptions we made are discussed. A. Assumptions It is assumed that the exoskeleton will totally mimic the human natural gait, so the gait data when wearing the exoskeleton can be obtained by multiplying the normalized joint torques and powers with the total weight of the wearer and the exoskeleton. Furthermore, the exoskeleton is to help spinal cord injured (SI) patients to walk. To avoid introducing too much psychological burden to the patients, we are aiming at relative low speed, that is,.8 m/s. In this paper we first focus on the speed of.8 m/s, and then we explain the speed dependency in section IV.B. For both PEA and SEA, linear springs are considered. The benefit of using linear springs is investigated. To increase the controllability of individual joints, bi-articular and multiarticular actuation like in [9][1], are not considered. The other assumptions on the actuation are Frictions due to motor and transmission (e.g., gears) are neglected. Since motor copper and iron losses are not included, efficiencies of the motor and gears are not included in the optimization process; Parallel springs can be engaged and disengaged when necessary; to avoid chattering, engagement and disengagement occurs when the spring is at rest (zero spring force); The inertias of the motor and gears are not included. Thus the power required for accelerating the motor and gear rotating parts is not included in the optimization; The motor requires (the same amount of) energy for both motoring and braking modes. B. Optimization method for SEA The main purpose of using a series spring is to store the energy and reuse it during power-demanding moments during gait, for example, during push off at the ankle, so that the peak power demand on the motor can be reduced. In a SEA, the motor power P m can be calculated using the following formulas [11], P m T T = T θ J +, (2) K S 893

where T is the joint torque and T is the time derivative of joint torque; θ is the joint velocity; K J S is the spring stiffness of the series spring; θ is the motor velocity (input side of the spring, m output side of the gears). The objective is to find a spring stiffness that minimizes the peak motor power ( S) = max ( m), (, ) f K ( K + S ) f K P min. S. Optimization method for PEA The primary goal of adding a linear parallel spring is to reduce the peak torque demand on the motor and gears during gait. To reduce the peak torque by fitting a line (linear spring), the least absolute deviations (LAD) method would be the right choice. However, considering the possibly unstable or multiple solutions of LAD, we choose the least squares (LS) method to fit the torque-angle curve to find the optimal spring stiffness. The LS method tries to distribute the torque evenly around the spring torque-angle curve. Further, for most problems (with no outliers in the data set), the solutions of LAD and LS are very close. The objective function is N ( P) = [ i P( θj, i θe) ] f K T K min i = 1 ( ) f, ( KP ), KP + where T i is the joint torque at the ith time instant; θ Ji, is the joint angle; K P is the parallel spring stiffness; θ E is the spring engagement angle. When the joint reaches this angle, the spring torque is zero, that is, the spring is at rest. IV. RESULTS AND DISUSSION This section shows the optimization results, i.e. the torque reduction, power or energy reduction. The discussion will be carried out joint by joint. A. Optimization results and discussions 1) Series elastic actuation (SEA) At the ankle, when the spring stiffness increases from zero to infinity, the peak motor power first drops rapidly and then rises relative slowly to gait power 217 W (Fig. 4 upper left plot). The minimum indicates that the optimal spring (stiffness 782 Nm/rad) reduces the peak motor power from 217 to 45 W (by 79%), and the rms power from 61.7 to 13.3 W (by 78%). The close-to-zero motor speed during mid-stance (Fig. 4 upper right plot) suggests that the motor tries to hold the position of the spring and the spring is absorbing energy. The motor torque is exactly the same as the ankle torque during gait, under the assumption given in section III.A. Although series elasticity 2, (3) (4) Peak motor power(w) 5 4 3 2 RANK, Sagittal 217 [782, 45.1] 1 2 1 3 1 4 Spring stiffness.(nm/rad) - -2.2.4.6.8 1 1.2 1.4-2 Gait data motor of SEA -4.2.4.6.8 1 1.2 1.4 can reduce the power and energy demands, it cannot reduce the high demand on motor torque. It requires a relatively heavy torque motor or a large gearbox (spindle, harmonic drive, gear sets, etc.) with a high gear ratio to generate and strong structure to bear the 2 Nm peak torque. The same optimization method is applied to other joints. For hip F/E and Knee F/E, no obvious reduction in peak power and rms power is observed. At the hip in frontal plane, adding a series spring can reduce the peak motor power from 36.3 to 14.5W (by 6%) and the rms motor power from 12.2 to 6.18W (by 49%). 2) Parallel elastic actuation (PEA) When a parallel spring is added to the joint, the motor only needs to deliver part of the joint torque, that is, the residual torque, which is the difference between the required joint torque and spring torque. The torque and power profiles at the hip are shown in Fig. 5. By fitting a line to the torque-angle curve, we find the optimal (in a least-square sense) spring stiffness. At the hip, the parallel spring reduces the peak motor torque from 8.6 to 27.2Nm (by 66%) and from 97.1 to 45.9 Nm (by 53%), and the rms motor power from 24.8 to 12.5W (by 5%) and from 12.2 to 6.73W (by 45%), in sagittal plane and frontal plane, respectively. At the knee, the spring stiffness changes dramatically after heel off (HO). When implementing a spring with fixed spring stiffness, the motor has to produce a large torque in swing phase (larger than natural gait torque) to counteract the spring to move the lower leg. There is no benefit of adding a parallel spring during swing. Similarly, if a parallel spring is added to the ankle, the ankle motor has to counteract the spring during late stance and swing phase (Fig. 6 middle plot), since the spring stiffness varies and even becomes zero during swing. The conflict can be solved by adding multiple springs with different stiffness values, which work alternately. That will introduce complex and heavy mechanism (see e.g., [18]). For Power (W) Speed (rad/s) 2 3 2 rms61.7 217 rms13.3 45.1 -.2.4.6.8 1 1.2 1.4 Figure 4. SEA for the ankle. Dash line: collected gait data scaled to a total weight 13 kg; solid-dot-line: the motor speed, torque and power. Optimal spring reduces the motor power by reducing the motor speed, and the motor torque is the same as joint torque 894

Power (W) 8 6 4 2-2 -.2.2.4.6 Angle (rad) 5 gait spring the above reason, we constrained the design to use only one spring. For the knee, no such single spring can achieve a large reduction in the motor torque and power. Hence, no spring is suggested for the knee. For the ankle, a unidirectional parallel spring is suggested. The spring engages when the ankle starts dorsiflexion. The unidirectional parallel spring reduces the peak motor Power (W) RHIP, Sagital gait,rms 34.7 motor,rms 12.6-5.3.6.9 1.2 5 27.2-5 - 5 gait spring -.2 -.1.1.2.3 Angle (rad) 2-8.6 gait,rms 24.844 37.9 motor,rms 12.5-5.3.6.9 1.2 gait,rms 84.9 motor,rms 32.4 12-2 97.2.4.6.8 1 1.2 1.4 3 2 gait,rms 61.7 motor,rms 23.9 -.1 -.5.5.1 Angle (rad) 217 85.1 -.2.4.6.8 1 1.2 1.4 Figure 6. Unidirectional parallel spring for the ankle. PEA reduces the motor torque and power; after adding the spring, the peak motor torque occurs during swing, where the motor has to overcome the spring force. 2-2 -4-6 -8 5 gait spring RHIP, Frontal -5-45.9 gait,rms 58.7 motor,rms 22.3 - -97.1.3.6.9 1.2 4 2 36.3-2 -21.6 gait,rms 12.2 motor,rms 6.73-4.3.6.9 1.2 Figure 5. PEA for hip F/E (left column) and A/A (right column); dash line: collected gait data; solid-dot-line: motor torque and power. Both motor torques and powers are reduced by adding a linear parallel spring; torque from 197 to 12Nm (by 48%) and rms motor power from 61.7 to 23.9W (by 61%). When a spring is added, the peak motor torque occurs during swing, the motor has to fight against the spring to dorsiflex the foot and maintain foot clearance (Fig. 6). 3) Series elastic actuation vs. parallel elastic actuation According to the above analysis of the results, at the hip, in sagittal plane, the use of a PEA is suggested, since parallel spring reduces both the torque and power demand on motor and gears. At the knee, no spring is suggested. At the hip in frontal plane and at the ankle, both SEA and PEA can achieve large power reduction, but parallel springs can reduce the torque, while series springs cannot. Hence PEA is more favorable for all these four powered DoFs. 4) Series elastic actuator for parallel elastic actuation A parallel spring reduces the torque demand on the motor and gears. A series spring can reduce the peak motor power. We will investigate whether the use of a series elastic actuator to replace the motor in parallel elastic actuation can further improve the energy efficiency. Using the optimal parallel spring stiffness and residual torques that the motor has to deliver, the optimization process for series elastic actuation is performed for all the joints. For example, the peak motor power for the hip is plotted against the series spring stiffness (Fig. 7). The series spring only achieves marginal reduction in the peak motor power (from 37.9 to 34.6W) at the hip in sagittal plane. At other joints, the series spring achieves either no or marginal improvement. In conclusion, the SEA cannot improve the energy economy further. However, the SEA can lower the mechanical output impedance at high frequency (by decoupling the motor mass from the load) and improve the performance of force control. B. Variable stiffness Previous discussion focuses on one walking speed (.8m/s). In this section, the sensitivity of optimal spring stiffness to speed and weight are briefly discussed. The dependencies will be illustrated using the ankle actuation. 1) Weight dependency The optimal spring stiffness is linearly related to the total weight of the wearer and the exoskeleton, both for PEA and SEA (Fig. 8), under the assumption given in section III.A. 2) Speed dependency Since we only obtained two sets of gait data, the conclusion we draw can only be based on these two data sets. At the ankle, Peak motor power (W) 5 RHIP, Sagittal (549,34.6) (1e+4,37.7) 1 2 1 3 1 4 Series spring stiffness (Nm/rad) Figure 7. Using a series elastic actuator in parallel elastic actuation only achieves marginal benefit in terms of peak power reduction. 895

Spring stiffness(nm/rad) 8 7 6 RANK, Sagittal, SEA spring 5.8(m/s) 4 8 9 11 12 13 Weight of human+exo.(kg) Spring stiffness(nm/rad) 6.8(m/s) 5 8 9 11 12 13 Weight of human+exo.(kg) the series spring stiffness reduces as the walking speed increases from.8 to 1.2 m/s, In contrast, the parallel spring stiffness increases as the walking speed increases (Fig. 8). For a constant walking speed, the optimal spring stiffness can be tailored to a specific user (according to his/her body weight), which only requires offline (manual) adjustment of the spring. Increase/decrease in walking speed requires online (motorized) adjustment of the spring stiffness to achieve the maximal torque and power reduction. The implementation of variable stiffness spring likely ends up with complicated and heavy actuators (see e.g. [19]). For a total weight of 8kg, when the speed increases from.8 m/s to 1.2 m/s, the optimal parallel spring stiffness increases from 512Nm/rad to 547Nm/rad, that is, a sensitivity of (547-512)/512 = 6.8% (Fig. 8 right plot). The optimal parallel spring stiffness is relatively insensitive to speed changes. One constant stiffness parallel spring for both speeds is not the optimal solution, but it can reduce the motor torque and power compared to a joint without any parallel spring. V. ONLUSION In this paper, the benefit of adding elasticity to exoskeleton actuation is investigated. Two types of elastic actuations, series elastic and parallel elastic actuation, are discussed. PEA reduces the peak torque by 66%, 53%, and 48% for hip flexion/extension (F/E), hip abduction/adduction (A/A), and ankle dorsi/plantar flexion (D/P), respectively, and the rms power by 5%, 45%, and 61%, respectively. SEA reduces the peak motor power by 79% for ankle dorsi/plantar flexion and by 6% for hip ab/adduction; but SEA doesn t reduce the joint torques. To summarize, we suggest using parallel elastic actuation to power the hip and the ankle. At the knee, no spring is suggested. Using a series elastic actuator as the torque source in parallel elastic actuation will not further improve the power efficiency. However, series elastic actuation can lower the mechanical output impedance at higher frequencies (e.g. due to impact) and improve the performance of force control. 9 8 7 RANK, Sagittal,PEA spring Figure 8. Optimal series and parallel spring stiffness is linearly related to total weight; walking speed affects the optima slightly. REFERENES [1] Rick Hansen Spinal ord Injury Registry (RHSIR), Spinal cord injury facts and statistics, http://rickhansenregistry.org/page192.htm, 26. [2] ARGO Medical Technology Ltd., http://www.argomedtec.com/ [3] Berkeley Bionics, http://berkeleybionics.com/exoskeletons-rehabmobility/ [4] REX Bionics, http://www.rexbionics.com/ [5] H. Herr, R. Kornbluh, "New horizons for orthotic and prosthetic technology: artificial muscle for ambulation," Proc. of SPIE - The International Society for Optical Engineering, Smart Structures and Materials 24 - Electroactive Polymer Actuators and Devices (EAPAD), San Diego, A, USA, Vol. 5385, No. 1, pp. 1-9, 24. [6] G.A. Pratt, M.M. Willianmson, Series Elastic Actuator, Proc. of IEEE Int. onf. on Intelligent Robots and Systems, Pittsburgh, PA, USA, Vol. 1, pp. 399-46, 1995. [7] J. E., Pratt, B. T., Krupp,. J., Morse, S. H., ollins, "The RoboKnee: an exoskeleton for enhancing strength and endurance during walking," Proc. of IEEE Int. onf. on Robotics and Automation, New Orleans, LA, USA, Vol. 24, Issue 3, pp. 243-2435, 24. [8] J.E., Pratt, B.T., Krupp, "Series elastic actuators for legged robots," Proc. of SPIE - The International Society for Optical Engineering, Vol. 5422, pp. 13544, 24. [9] K., Endo, H., Herr, Human walking model predicts joint mechanics, electromyography and mechanical economy, Proc. of Int. onf. on Intelligent Robots and Systems, St. Louis, USA, pp. 4663-4668, 29. [1] A.J., van den Bogert, "Exotendons for assistance of human locomotion," BioMedical Engineering Online, Vol. 2, art. No. 17, 23. [11] K.W., Hollander, R., Ilg, T.G., Sugar, D., Herring, "An efficient robotic tendon for gait assistance," Journal of Biomechanical Engineering, Vol. 128, No 5, pp. 788-791, 26. [12] S.K., Au, H, Herr, J., Weber, E.., Martinez-Villalpando, "Powered ankle-foot prosthesis for the improvement of amputee ambulation," Proceedings of Annual Int. onf. of the IEEE Engineering in Medicine and Biology, Lyon, France, pp. 32-326, 27. [13] H. K., Kwa, J. H., Noorden, et al., "Development of the IHM Mobility Assist Exoskeleton," IEEE Int. onf. on Robotics and Automation, Kobe, Japan, pp. 2556-2562, 29. [14] Lagoda,., Schouten, A.., Stienen, A. H. A., Hekman, E. E. G., van der Kooij, H., "Design of an electric series elastic actuated joint for robotic gait rehabilitation training," IEEE RAS and EMBS Int. onf. on Biomedical Robotics and Biomechatronics (BioRob), Tokyo, Japan, pp. 26-29, 21. [15] Scheint, M., Sobotka, M., Buss, M., "Optimized parallel joint springs in dynamic motion: omparison of simulation and experiment," IEEE RAS and EMBS International onference on Biomedical Robotics and Biomechatronics (BioRob), Tokyo, 21, pp. 485-49. [16] S. K. Au, J. Weber, H. Herr, "Powered Ankle Foot Prosthesis Improves Walking Metabolic Economy," IEEE Trans. on Robotics, Vol. 25, No. 1, pp. 51-66. [17] D.A. Winter, Biomechanics and motor control of human movement, New Jersey: John Wiley & Sons, Inc., 4 th ed., 29. [18] A. ullell et al., "Biologically based design of an actuator system for a knee ankle foot orthosis," Mechanism and Machine Theory, Vol. 44, No. 4, pp 86 872, 29. [19] Eiberger, O., Haddadin, S., et al., "On joint design with intrinsic variable compliance: derivation of the DLR QA-Joint," Proc. of IEEE Int. onf. on Robotics and Automation (IRA), Anchorage, AK, USA, 21, pp. 1687694. 896