Sample purification procedure for confirmation of 3 -hydroxy stanozolol by gas chromatography/high resolution mass spectrometry

Similar documents
May 1, By John Heim,Doug Staples

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Longterm Detection of Anabolic Steroid Metabolites in Urine

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

WADA Technical Document TD2016EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Retention Time Locking: Concepts and Applications. Application

Methylprednisolone detection in urine following local and oral administrations

WADA Technical Document TD2018MRPL

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

A simple and rapid ESI-LC-MS/MS method for simultaneous screening of doping agents in urine samples

SIMPLE, EFFICIENT AND SIMULTANEOUS DETERMINATION OF ANABOLIC STEROIDS IN HUMAN URINE

WADA Technical Document TD2018EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

1. INTRODUCTION CH 3 CH3

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

WADA Technical Document TD2010EAAS. Endogenous Anabolic Androgenic Steroids Testing, Reporting and Interpretive Guidance

LCMSMS Solutions For Steroid Analysis

SPECIFIC DETECTION OF ANDROST-4-ENE-3,17-DIONE

WADA Technical Document TD2017NA

Extended Application Note

Experiment GC : Analysis of BTEX by GC-FID

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Rapid and Reliable Detection of Dissolved Gases in Water

CORESTA RECOMMENDED METHOD Nº 67

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Renzo BAGNATI Mario Negri Institute of Pharmacological Research- IRCCS Milano - Italy

Gas chromatographic determination of 1,4-dioxane at the

GC-MS(/MS) investigations on long-term metabolites of 17-methyl steroids

Introduction. Saurabh Dubey 1, * and Alka Beotra 2

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

GCMSD-Headspace Analysis SOP

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

no peaks missing peaks too small peaks increasing retention decreasing retention

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Automated Determination of Dissolved Gases in Water Anne Jurek

RayBio Human vwf ELISA Kit

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Agilent Dimension Software for ELSD User Manual

M/s. Thermo fisher Scientific

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Introduction. Experimental

RECENT ADVANCES IN DOPING ANALYSIS (4)

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES

EVALUATION OF PROFICIENCY TESTING PROGRAM, SETTING PRIORITIES FOR THE FUTURE.

SUPPORTING INFORMATION

Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

RECENT ADVANCES IN DOPING ANALYSIS (2)

Chemical Analysis for Methyltestosterone

SMART 4 Sports Panel. Drug Free Sport 1 7/13/07

APPLICATION NOTE. GC Integrated Permeation Device

Setting a New Standard in Flash Chromatography Performance

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Improved Performance in Capillary Electrophoresis using Internal Standards

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Hair analyses. Hair analysis beyond workplace testing. Ia: Fatal steroid abuse. Identification of stanozolol in hair

RayBio Human TNF-alpha ELISA Kit

Sports drug testing an analyst s perspective

Gas chromatographic and mass spectrometric analysis of conjugated steroids in urine

Measurement reliability in horseracing laboratory operation

The instrument will process the samples one by one till all the samples have been cleaned up.

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

HUMAN IL6 KITS PROTOCOL

TRACE LEVEL DETERMINATION OF PER- AND POLYFLUOROALKYL SUBSTANCES (PFASs) IN WATER USING THE AGILENT 6460 LC/MS/MS

Studies on the determination of chlorotestosterone and its metabolites in bovine urine

MSD 96-Well MULTI-ARRAY CRP Assay

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell

Conducting an XFe Assay in an Hypoxia Chamber at 3% O ²

Carrier Gases in Capillary GC

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

N Korea Japan China Thailand Taiwan Philipines EP-5 EP km. Source: Google Maps and

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994

RayBio Human IFN alpha/beta R2 ELISA Kit

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

Transcription:

Author proof done Research Article Sample purification procedure for confirmation of 3 -hydroxy stanozolol by gas chromatography/high resolution mass spectrometry S. Ahi, I. M. Reddy, A. Beotra, R. Lal, S. Jain, T. Kaur ABSTRACT Dope Control Center, Sports Authority of India, J. N. Stadium, New Delhi, India Received: 07.08.2007 Revised: 18.04.2008 Accepted: 23.08.2008 Correspondence to: Alka Beotra E-mail: drabeotra@rediffmail.com Objective: To improve the detection limit of 3 -hydroxy-stanozolol by using the double extraction procedure, specific for basic drugs. Materials and Methods: The urine samples were spiked in four replicates with 3 -hydroxystanozolol at different concentrations of 1, 2, 5 and 10 ng/ml, processed by two different methods and injected into gas chromatography/high resolution mass spectrometry (GC- HRMS) instrument. The data was analyzed by comparing the recovery values and the ion match criterion of the two procedures. Results: The results show that the recovery percentage and ion match criterion of 3 -hydroxystanozolol at lower concentrations has a significant improvement when Solid phase extraction was performed, instead of Liquid-liquid extraction in the second extraction procedure. Conclusion: The sample preparation procedure using Oasis-MCX cartridges allows confirmation of 3 -hydroxy-stanozolol at the minimum required performance limit (MRPL) decided by the World Anti Doping Agency. This procedure may be used for the confirmation of suspicious samples found in routine testing, as it efficiently fulfills the ion-matching criterion laid down by the World Anti Doping Agency. KEY WORDS: 3 -hydroxy-stanozolol, gas chromatography/high resolution mass spectrometry, purification, World Anti Doping Agency Introduction Since 1967, the use of performance enhancing drugs is banned in sports. Anabolic Steroids (AAS) have been included in the list banned by the International Olympic Committee (IOC) in 1976. Since the year 2002, the World Anti Doping Agency (WADA) has taken over and made more stringent guidelines for the testing of drugs by doping labs. [1] Stanozolol (17α-methyl-17β-hydroxy-5α-androstane- [3, 2, C] pyrazole), first synthesized in 1959, [2] is one of the most misused anabolic steroids in sports. [3] Stanozolol is metabolized to a large extent and the main metabolic product in urine is mono hydroxylated, 3 -hydroxy-stanozolol. Stanozolol and its metabolites are different from most anabolic steroids and are particularly difficult to detect in urine. These compounds have a poor gas chromatographic behavior and the measured concentrations are generally very low. In order to ensure that all doping control laboratories can detect the presence of prohibited substances uniformly, WADA has established a minimum detection capability for testing methods called minimum required performance limits (MRPL). The MRPL for anabolic steroids is 10 ng/ml, except clenbuterol, nandrolone, stanozolol, methyl testosterone and methandione, for which the detection limit is 2 ng/ml. In view of the above, it is very crucial to strengthen the testing protocol of the lab to improve the detection limit of these five drugs. We have already reported improvement in the detection limit of clenbuterol and epimetendiol with SPE cartridges. [4] The identification of stanozolol by various extraction procedures has been shown in literature. [5-6] The sample preparation procedures vary from Solid phase extraction and Liquid-liquid extraction [7] to Immunoaffinity chromatography. [8] However, irrespective of the methods of sample preparation, the identification of stanozolol and its metabolites has proved to be problematic, particularly in cases of low concentrations, due to interfering matrix peak. Hence, it is required to improve sample extraction procedure, which would ensure removal of interfering matrix, thereby allowing the WADA criteria of confirmation by ion match to be fulfilled. The objective of the present paper is to develop a confirmation procedure for Stanozolol by comparing its efficacy with the existing procedure. 164

Materials and Methods Reference standards, chemicals and reagents The reference standard of 3 -hydroxy stanozolol was purchased from the National Measurement Institute (Sydney, Australia). The derivatizing reagents N-Methyl- N-Trimethylsilyltrifluroacetamide, Iodo trimethylsilyl and Dithioerythritol were purchased from Sigma (St-Louis, Missouri, USA). The organic solvents and reagents were of HPLC grade. β-glucuronidase (E.coli) enzyme was purchased from Roche Diagnostics Corporation (Indianapolis, USA). Amberlite XAD2 (polystyrenedivinyl benzene) resin was purchased from Sigma (St-Louis, MO, USA) and solid phase extraction Oasis MCX cartridges from Waters Corporation (Massachusetts, USA). Deionised water was prepared on a Milli Q laboratory plant (Millipore, Bedford, USA). Stock standard solution was prepared in methanol, at a concentration of 1000 ng/ml, and was stored at 4 C, until used. Instrumentation and conditions The chromatographic system consisted of an Agilent 6890 series GC (Agilent Technologies, Waldron, Germany) equipped with 7679 series automatic liquid sampler. Ultra 1, fused silica (0.2 mm x 17 m x 0.11 µm) (Agilent Technologies, USA) column was used for good separation. The parameters like oven program, carrier flow and injection port temperature are presented in Table 1. Mass spectrometric analyses were conducted using highresolution mass spectrometer (JEOL JMS-700, Tokyo, Japan). The main working parameters of the mass spectrometer are summarized in Table 2. The gas chromatographic system and high-resolution mass spectrometer were both controlled using Table 1 Gas chromatography working parameters Instrument Agilent 6890 GC, 7679 automatic liquid sampler Injection mode Automatic split, split ratio 11:1 Injection volume 2 µl Injection port temperature 280 C Carrier fl ow 120 kpa helium at 180 C (constant pressure) Oven program 180 C for 1 min, 229 C at 3 C/min, 300 C at 40 C/min, fi nal hold for 3 min. Column Ultra 1, fused silica, 0.2mm x 17m x 0.11µm Table 2 High-resolution mass spectrometer-working parameters Instrument JEOL JMS-700 M station Tune Auto tune Tune peak width 0.5 amu Acquisition mode Selective ion monitoring Ion current 300 µa Ionizing energy 70 ev Resolution 10000 Workstation Unix workstation JMS-700 M Station software (JEOL JMS-700, Tokyo, Japan). Sample extraction procedure Solid phase extraction is very efficient and more versatile because it requires less organic solvent, it is easy to wash out interfering substances and it does not form emulsions. Extraction procedure 1: Extraction procedure 1 is the existing procedure, which is being used in our lab for testing of anabolic androgenic steroids. [9,10] Two/four ml of urine, based on the specific gravity, is applied on the pre-prepared XAD2 columns. The columns are washed with two ml of water to eliminate most of the water-soluble urinary constituents, which have not been absorbed on the solid support. The free and conjugated steroids are then eluted with 2.5 ml of methanol. The eluted effluent is evaporated under nitrogen stream at 60 C and the residue is dissolved in one ml of phosphate buffer (ph 7.0). Hydrolysis is performed with 50 µl of β-glucuronidase enzyme for one hour, at 60 C. The sample is alkalized by adding K 2 CO 3 and is extracted with 5 ml of distilled diethyl ether. After centrifugation, the organic phase is separated and dried with Na 2 SO 4, 50 µl of internal standard 17α-methyl testosterone (Conc. 500 ng/ml) is added. It is then evaporated under Nitrogen stream at 60 C. Extraction procedure 2: The first step of elution of drug from XAD2 resin and the hydrolysis step is similar to that of the Procedure 1. Two/four ml of urine, based on the specific gravity, is applied on the pre-prepared XAD2 columns. The columns are washed and the drug is eluted with 2.5 ml of methanol. The eluted effluent is evaporated and hydrolysis is performed with 50 µl of β-glucuronidase enzyme for one hour, at 60 C. 10 µl of 5N HCl is added to hydrolysate to make it acidic. Oasis MCX cartridges are conditioned with 2 ml of methanol and water. The acidified effluent is applied on the Oasis MCX cartridges and washed with 2 ml HCl and methanol. The basic drug, 3 - hydroxy stanozolol is eluted with 2 ml methanol containing 1% ammonium hydroxide. Internal standard is added and evaporated as mentioned in Procedure 1. The dried residue is again re-dissolved in 200 µl of methanol and dried again followed by derivatization. Derivatization The dried residue of Procedures 1 and 2 are dissolved in 50 µl of MSTFA/Iodo-TMS/Dithioerythritol (1000/2/2:v/v/w) mixture. The mixture is incubated at 60 C, for 15 minutes and then transferred into vials. 2 µl is injected into GC/HRMS. Results The identification criterion of stanozolol was as per the fragmentation ions 560.3650, 546.3493 and 545.3410. The detection of 3 -hydroxy stanozolol with Extraction Procedure 2 [Figure 1] on the GC/HRMS showed improved recovery, as compared to Extraction procedure 1 [Figrue 2]. Hence, Extraction Procedure 2 was validated as per the WADA International Standard of Laboratories (ISL) [11] requirements. Method validation Calibration and quality control samples Internal standard, working and quality control solutions were prepared by diluting its stock solution with methanol. Calibration standards of various concentrations were prepared, processed and analyzed to draw a calibration curve. Quality 165

Table 3 Spiked urine samples showing calculated concentrations and recovery percentages of 3 -hydroxy stanozolol Spiked concentration Concentration recovered Recovery percentage (%) (ng/ml) Mean ± SD (ng/ml) Mean ± SD Extraction procedure 1 Extraction procedure 2 Extraction procedure 1 Extraction procedure 2 1 0.8605 ± 0.0332 1.1722 ± 0.1466 71.59 ± 2.77 97.52 ± 12.20 2 1.1314 ± 0.1837 1.5486 ± 0.2751 53.40 ± 8.67 73.09 ± 12.98 5 3.1388 ± 0.6618 2.7209 ± 0.3534 70.60 ± 14.89 61.20 ± 7.95 10 6.7958 ± 1.0828 5.7967 ± 0.6229 66.41 ± 10.58 56.64 ± 6.08 control samples were prepared at concentrations 2 ng/ml for the minimum required performance limit (MRPL), required by WADA. [10] Calibration curve and linearity A three-point calibration curve was constructed in the range of 1-10 ng/ml (1, 2, 10 ng/ml). Linearity was assessed by a weighted (1/x) least squares regression analysis method. The calibration curve had a correlation coefficient (r 2 ) of 0.9999 [Figure 3]. The acceptance criterion for accuracy and precision was followed as per international rules. Table 3 summarizes the calculated concentrations of spiked samples of 3 -hydroxy stanozolol in human urine. Limit of detection (LOD), signal to noise ratio Limit of detection for 3 -hydroxy stanozolol is measured by spiking 1 ng/ml of drug in the drug free urine sample. Limit of detection (LOD) values were calculated as the analyte concentration providing a signal to noise value equal to three, as determined by the Agilent MSD proprietary Chemstation software. Six spiked urine samples were extracted and analyzed for a few days, to cover the various performances for the instrumentation. Selectivity Ten blank urine samples obtained from different sources and pretreated by Extraction Procedure 2 were analyzed to check for possible chemical and chromatographic interferences. The selected ion chromatogram profiles did not show the presence of any significant signal (S/N<3) at the typical retention time of the analyte of interest, indicating that the method is selective for 3 -hydroxy stanozolol. Precision and accuracy The quality control samples were randomized daily, processed and analyzed at the end of the batch. The within batch, between batch precision and accuracy were determined by analyzing six sets of quality control samples. The acceptance criteria of within and between batch precision and accuracy was calculated as per international rules. The precision of the method was expressed as relative standard deviation and accuracy of the method was expressed in terms of bias (percentage deviation from true value). Intra assay precision (%) and accuracy were evaluated by extracting and analyzing three replicates of urine samples spiked at the MRPL concentration, performed by three different operators. Figure 1: Mass spectrum of 2 ng/ml Spiked 3 -hydroxy stanozolol from Extraction procedure 2 Sample Name: MSIM0017 Comment: 2ng ext-oasis Compound Name: 3 OH- Stanozolol 166

Total recovery Spiked urine samples were prepared by adding adequate volume of the corresponding methanolic working standard solution to 2 ml of negative urine, yielding final concentrations of 1, 2, 5 and 10 ng/ml respectively. The quantitative ion 560.3650 was taken for calculation of recovery percentage. Total recovery percentage of 3 -hydroxy stanozolol was calculated by comparing the mean peak areas of three processed spiked samples with the mean peak areas of unprocessed direct reference standard solutions of same concentrations. The recovery percentage of Extraction Procedure 1 was found to range between 53.4 to 71.59% and of Extraction Procedure 2 in the range of 56 to 97% for four of the concentrations, ranging between 1-10ng/mL [Table 3]. The extraction recovery percentages were found to be within the limits and were acceptable. Internal standard recovery was also found to be good and agreeable, as it was consistent, precise and reproducible. Ion match criteria The samples processed by Extraction Procedure 1 did not meet the ion match criteria of WADA at 1 and 2 ng/ml, whereas Extraction Procedure 2 showed conformance to ion match criteria at all the concentrations (1, 2, 5 and 10 ng/ml). Comparative ion matching criteria of 3 -hydroxy stanozolol by Extraction Procedures 1 and 2 is shown in Tables 4 and 5. Doping control samples The sample preparation procedure described was applied to a few doping control samples tested positive for stanozolol misuse. The results are illustrated in Table 6, which shows better recovery when the samples are processed by Extraction Procedure 2. Table 4 Ion matching criteria of 3 -hydroxy stanozolol (conc. 1 ng/ml) by extraction procedure 1 RT ISTD RRT RRT% Peak areas Relative abundance (R.A) Difference of R.A (D.R.A) m/z m/z m/z m/z m/z m/z m/z m/z m/z 560.365 546.349 545.341 560.365 546.349 545.341 560.365 546.349 545.341 Direct standard 19.55 18.35 1.06 100 201735 35483 110372 100 17.5 54.7 - - - Sample-1 19.55 18.35 1.06 100 108780 000 19949 100 000 18.3-17.5 36.3 Sample-2 19.55 18.35 1.06 100 82343 69064 40078 100 83.8 48.6 - -66.2 06.0 Sample-3 19.55 18.35 1.06 100 90803 000 000 100 000 000-17.5 54.7 Sample-4 19.55 18.35 1.06 100 91389 000 000 100 000 000-17.5 54.7 WADA Ion matching criteria Ion allowance RT ISTD RRT RRT% Retention time of drug Retention time of internal standard Relative retention time Relative retention time percentage RT/RRT ± 0.2 min or 99 to 101 % R.A > 10 % D.R.A R.A > 50 % = 10 % R.A 25 % to 50 % = ± (R.A x 0.2) % R.A < 25 % = ± 5 % Table 5 Ion matching criteria of 3 -hydroxy stanozolol (conc. 1 ng/ml) by extraction procedure 2 RT ISTD RRT RRT% Peak areas Relative abundance (R.A) Difference of R.A m/z m/z m/z m/z m/z m/z m/z m/z m/z 560.365 546.349 545.341 560.365 546.349 545.341 560.365 546.349 545.341 Direct standard 19.55 18.35 1.06 100 201735 35483 110372 100 17.5 54.7 - - - Sample-1 19.55 18.35 1.06 100 198580 31689 114524 100 15.9 57.6-1.6-2.9 Sample-2 19.55 18.35 1.06 100 853167 184953 507876 100 21.6 59.5 - -4.0-4.8 Sample-3 19.55 18.35 1.06 100 322149 48550 207848 100 15.0 64.5-2.5-9.8 Sample-4 19.55 18.35 1.06 100 395780 75524 241067 100 19.0 60.9 - -1.4-6.1 WADA Ion matching criteria Ion allowance RT ISTD RRT RRT% Retention time of drug Retention time of internal standard Relative retention time Relative retention time percentage RT/RRT 0.2 min or 99 to 101% R.A > 10% D.R.A R.A > 50 % = 10% R.A 25% to 50% = (R.A x 0.2)% R.A < 25% = 5% 167

Figure 2: Mass spectrum of 2 ng/ml Spiked 3 -hydroxy stanozolol from extraction procedure 1 Sample Name: MSIM0021 Comment: 2 ng ext-xad 2 Compound Name: 3 OH-stanozolol Discussion The objective of the present study was to improve extraction procedure of 3 -hydroxy stanozolol to achieve good recovery percentage and to achieve acceptable ion matching criteria laid down in WADA International Standard of Laboratories. The separation of drug was performed on gas chromatography and detection was done by high-resolution mass spectrometer. The lowest possible detection of 3 -hydroxy stanozolol is 1 ng/ml on GC/HRMS and was achieved with the present method. The total recovery percentages of 3 -hydroxy stanozolol were found to range from 53 to 71% with Extraction Procedure 1 and 56 to 97% with Extraction Procedure 2. The extraction recovery percentages were found to be within the limits and agreeable. Internal standard recovery is also found good and acceptable, as it was consistent, precise and reproducible. The first extraction procedure concentrates the drug in polystyrene divinyl benzene resin column and further extracts it through liquid-liquid extraction. However, the second Extraction Procedure involves the use of solid phase Table 6 Spiked urine samples showing calculated concentrations of 3 - hydroxy stanozolol with extraction procedure 1 and 2 Positive sample Concentration recovered Mean ± SD (ng/ml) Extraction procedure 1 Extraction procedure 2 Athlete-1 2.093 ± 0.007 3.504 ± 0.185 Athlete-2 1.725 ± 0.479 2.292 ± 0.166 Figure 3: Typical calibration curve of spiked sample of 3 -hydroxy stanozolol Ratio 6 5 4 3 2 1 0 0 2 4 6 8 10 12 Concentration (ng/ml) extraction with oasis cartridges. The elution of the drug was done with the alkaline organic solvent, which shows extraction recovery between 56 and 97%. The recovery percentages of the drug from both the extraction procedures were found to be in conformity at concentrations 5 and 10 ng/ml. However, recovery of drug at 1 and 2 ng/ml showed significant improvement by extraction Procedure 2. To declare a positive finding, it is required to detect a drug in athlete s urine and also to fulfill ion match criteria wherein three characteristic ions of presumptive positive sample should match with reference standards. [11] The ion match criterion was not being fulfilled with the first extraction procedure at lower concentrations, which was the reason for planning the present work. However, the analysis of data revealed that the peak height and area for quantitation 168

of ion 560.3650 was good with Extraction Procedure 1 but relative peak area and heights of 545.3410 and 546.3493 ions did not show conformance to WADA requirements, [11] which may be due to matrix interferences. The ion match criteria with Extraction Procedure 1 and 2 is shown in Tables 4 and 5, which clearly shows removal of matrix interferences by Extraction Procedure 2, thereby fulfilling WADA criteria of ion match. This is essentially required to make it legally strong, which is required for doping labs. Conclusion The present study could successfully achieve detection, good percentage recovery and fulfillment of ion matching criteria of 3 -hydroxy stanozolol at MRPL as specified by WADA, using the second Extraction Procedure protocol, with detection level as low as 1 ng/ml. Excellent detection limits enable the long-term detection of stanozolol misuse in doping control. This procedure may be used for the confirmation of suspicious samples found in routine testing. References 1. The World Anti Doping Code. The 2007 prohibited list international standard. Available from: http://www.wada-ama.org/rtecontent/document/list_2007.pdf. 2. Haynes RC Jr. Goodman and Gilman. In: Rall G, Nies, Taylor, editors. The pharmacological basis of therapeutics. McGraw-Hill International Editions, New York: 1996. p. 1459-60. 3. Horning S, Schanzer W. Steroid Screening using GC-HRMS. Recent Advances in Doping Analysis (14). Sport und Buch Strau, Koln 1996. p. 261-70. 4. Reddy MI, Beotra A, Jain S, Kaur T, Lal R. PuriÞcation of urine samples to improve detection limit of anabolic agents. Indian J Pharmacol 2007;39:38-41. 5. Fubholler G, Koch A, Mareck U, Thomas A, Schmechel A, Sigmund G, et al. New sample preparation procedure for the conþrmation of stanozolol and its metabolites by means of GC/MSMS and LC/ESI-MSMS. Recent Advances in doping analysis (14). Sport and Buch Straub, Koln 2006. p. 229-38. 6. Ayotte C, Goudreault D, Charlebois A. Testing for natural and synthetic anabolic agents in human urine. J Chromatogr B Biomed Appl 1996;687:3-25. 7. McDowell RD, Pearce JC, Murkitt GS. Liquid-solid sample preparation in drug analysis. J Pharm Biomed Anal 1986;4:3-21. 8. Ferrara SD, Tedeschi L, Frison G, Brusini G, Castagna F, Bernardelli B, et al. Drug of abuse testing in urine: Statistical approach and experimental comparison of immunochemical and chromatographic techniques; J Anal Toxicol 1954;18: 278-91. 9. Ueki M, Jain S, Ikekita A, Beotra A, Okano M, Sato M, et al. Dope testing during the 1st Afro-Asian games in India. J Sports Traumatol Allied Sports Sci 2004;5: 79-89. 10. Beotra A, Jain S, Kaur T. Dope testing during the 1st Indo-Pak games in India. Recent Advances in Doping Analysis (13). Sport und Buch Strau, Koln 2005. p. 457-60. 11. The World Anti Doping Code, International Standard for Laboratories. Version 4.7 April 2007. Announcement PSG College of Pharmacy National Seminar on Cancer-Biology and Development of New Chemical Entities Nov 20 th -22 nd, 2008 Venue PSG College of Pharmacy, Coimbatore Contact Prof. A. K. Chandrasekharan, Principal Dr. M. Ramanathan, Prof & Head, Dept of Pharmacology PSG College of Pharmacy, Peelamedu, Coimbatore - 4. Tel.: 0422 4345841 Mobile No.: 98949 74476 Fax No.: 0422 2594400 E-mail akchandrasekharan@gmail.com, muthaiahramanathanin@ yahoo.co.in cancerbiol@ psgpharma.ac.in Website www.psgpharma.ac.in 169