Validity of wrist-worn consumer products to measure heart rate and energy expenditure

Similar documents
EXAMINING THE RELABILITY AND VALIDITY OF THE FITBIT CHARGE 2 TECHNOLOGY FOR HEART RATE DURING TREADMILL EXERCISE

HHS Public Access Author manuscript Int J Cardiol. Author manuscript; available in PMC 2016 April 15.

Current State of Commercial Wearable Technology in Physical Activity Monitoring

Using Hexoskin Wearable Technology to Obtain Body Metrics During Trail Hiking

Journal of Exercise Physiologyonline

JEPonline Journal of Exercise Physiologyonline

Safety Behavior for Cycling : Application Theory of Planned Behavior

The Optimal Downhill Slope for Acute Overspeed Running

How Running Barefoot (Xero Shoes) and Running Shod (Montrail Shoes) Effects Percent of Maximum Heart rate - A Case Report

TRAINING WITH! PHYSICLO RESISTANCE GEAR. Testing & Validation

Does wearing a wrist guard affect the site of wrist fracture in snow sports?

Fitbit s 100+ Billion Hours of Resting Heart Rate User Data Reveals Resting Heart Rate Decreases After Age 40

How Running Barefoot (Xero Shoes) and Running Shod (Montrail Shoes) Effects Percent of Maximum Heart rate - A Case Report

Traffic Parameter Methods for Surrogate Safety Comparative Study of Three Non-Intrusive Sensor Technologies

The effects of a suspended-load backpack on gait

Development of methods to objectively identify time spent using active and motorised modes of travel to work: how do self-reported measures compare?

Validation of a Step Test in Children Ages 7-11

Initial Mortality of Black Bass in B.A.S.S. Fishing Tournaments

Towards determining absolute velocity of freestyle swimming using 3-axis accelerometers

Validation of an Inertial Sensor System for Swing Analysis in Golf

The Singapore Copyright Act applies to the use of this document.

Effect of walking speed and placement position interactions in determining the accuracy of various newer pedometers

Activity profiles in adolescent netball: A combination of global positioning system technology and time-motion analysis

Validity and Reproducibility of the Garmin Vector Power Meter When Compared to the SRM Device

Modeling Pedestrian Volumes on College Campuses

ScienceDirect. Long-distance, short-distance: triathlon. One name: two ways.

Comparison of Three Tests for Assesing the Aerobic Aptitude to the Elite Swimmers

Assessment of an International Breaststroke Swimmer Using a Race Readiness Test

Objective Physical Activity Monitoring for Health-Related Research: A Discussion of Methods, Deployments, and Data Presentations

Changes in transition times in Ironman Hawaii between 1998 and 2013

Validity of Four Activity Monitors during Controlled and Free-Living Conditions

NATIONAL STEPS CHALLENGE TM SEASON 4 STEP UP TO TAKE OFF WITH SINGAPORE AIRLINES GROUP CHALLENGE FREQUENTLY ASKED QUESTIONS

Academic Grant CPR process monitors provided by Zoll. conflict of interest to declare

Procedia Engineering Procedia Engineering 2 (2010)

March Madness Basketball Tournament

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

Evaluating the Influence of R3 Treatments on Fishing License Sales in Pennsylvania

Smart Water Application Technologies (SWAT)

Clinical Study Synopsis

Impact Points and Their Effect on Trajectory in Soccer

Empirical Analysis of Speeding Behaviour and Determining. Speed Limits for Bicycles. Abstract

Time-motion and heart-rate characteristics of adolescent female foil fencers

Hydroacoustic surveys of Otsego Lake s pelagic fish community,

TEST REPORT. ACCORDING TO IES LM For. Samsung Electronics Co., LTD. 1,Samsung-Ro,Giheung-Gu,Yongin-City,Gyeonggi-Do 17113, Korea

Exercise Energy Expenditure Estimation Based on Acceleration Data Using the Linear Mixed Model

March Madness Basketball Tournament

Article Title: The Validity of Microsensors to Automatically Detect Bowling Events and Counts in Cricket Fast Bowlers

Effects of Age and Body Mass Index on Accuracy of Simple Moderate Vigorous Physical Activity Monitor Under Controlled Condition

VALIDITY OF SELECTED CARDIOVASCULAR FIELD-BASED TEST AMONG MALAYSIAN HEALTHY FEMALE ADULT. S. H. Azmi 1,*, and N. Sulaiman 2

Correlation analysis between UK onshore and offshore wind speeds

Applying Hooke s Law to Multiple Bungee Cords. Introduction

PROTOCOL FOR COMPRESSED AIR PROFICIENCY TESTING (CAPT) PROGRAM SAMPLE ANALYSIS

Determining bicycle infrastructure preferences A case study of Dublin

Equation 1: F spring = kx. Where F is the force of the spring, k is the spring constant and x is the displacement of the spring. Equation 2: F = mg

Effects of directionality on wind load and response predictions

APPROACH RUN VELOCITIES OF FEMALE POLE VAULTERS

MEMORANDUM. Investigation of Variability of Bourdon Gauge Sets in the Chemical Engineering Transport Laboratory

Validation of PiezoRx Pedometer Derived Sedentary Time

How Accurate Is Your Activity Tracker? A Comparative Study of Step Counts in Low-Intensity Physical Activities

Evaluation of pedestrians speed with investigation of un-marked crossing

Individualising conditioning programs for large tactical groups

Analysis of energy systems in Greco-Roman and freestyle wrestlers participated in 2015 and 2016 world championships

Relationship between Aerobic Training and Testosterone Levels in Male Athletes

The impact of freediving on psychomotor performance and blood catecholamine concentration

WALK WITH A DOC 2017 EVALUABILITY ASSESSMENT

Is lung capacity affected by smoking, sport, height or gender. Table of contents

Accuracy and Precision of High-Speed Field Measurements of Pavement Surface Rutting and Cracking

Comparison of Four Feedback Methods Used to Help Improve Swimming Relay Exchanges - A Pilot Study

SCREENING OF TOPOGRAPHIC FACTOR ON WIND SPEED ESTIMATION WITH NEURAL NETWORK ANALYSIS

Sontek RiverSurveyor Test Plan Prepared by David S. Mueller, OSW February 20, 2004

Flip-flop footwear with a moulded foot-bed for the treatment of foot pain: a randomised controlled trial

Bicycling Simulator Calibration: Proposed Framework

Abstract. Ulrik Röijezon, PhD and Paul Arden, MCCI

A COMPARATIVE STUDY ON RESPIRATORY PARAMETERS BETWEEN SHORT DISTANCE AND LONG DISTANCE SWIMMERS

ETB Pegasus equipment was used in the following paper

Leg Power in Elite Male Fencers: A Comparative Study among the Three Competitive Disciplines

Active mobility and health: Insights from the PASTA Project

At What Extent Physical Training Can Minimize Gender Differences in Male and Female Trained Swimmers? - A Comparative Analysis

CARDIOVIT AT-104 ergospirometry

THE INFLUENCE OF SLOW RECOVERY INSOLE ON PLANTAR PRESSURE AND CONTACT AREA DURING WALKING

Influence of Angular Velocity of Pedaling on the Accuracy of the Measurement of Cyclist Power

E. Agu, M. Kasperski Ruhr-University Bochum Department of Civil and Environmental Engineering Sciences

Highway & Transportation (I) ECIV 4333 Chapter (4): Traffic Engineering Studies. Spot Speed

Simulation analysis of the influence of breathing on the performance in breaststroke

INTERACTION OF STEP LENGTH AND STEP RATE DURING SPRINT RUNNING

Weather and Cycling in Dublin : Perceptions and Reality

Smart Health Walking Digital Pedometer And Heart Rate Watch Manual

AFG FITNESS APP OWNER S MANUAL AFG MANUEL DU PROPRIÉTAIRE DU TAPIS ROULANT AFG MANUAL DEL PROPIETARIO DE LA CAMINADORA

Evaluation of the ACC Vehicles in Mixed Traffic: Lane Change Effects and Sensitivity Analysis

Changes in a Top-Level Soccer Referee s Training, Match Activities, and Physiology Over an 8-Year Period: A Case Study

Analysis of Acceleration Value based on the Location of the Accelerometer

COMPARISON OF PHYSIOLOGICAL VARIABLES AND BODY COMPOSITION VARIABLES AMONG WALKING TYPES ON TREADMILL

The effect of different backpack loading systems on trunk forward lean angle during walking among college students

Effect of the Grip Angle on Off-Spin Bowling Performance Parameters, Analysed with a Smart Cricket Ball

GENETICS OF RACING PERFORMANCE IN THE AMERICAN QUARTER HORSE: II. ADJUSTMENT FACTORS AND CONTEMPORARY GROUPS 1'2

The Incidence of Daytime Road Hunting During the Dog and No-Dog Deer Seasons in Mississippi: Comparing Recent Data to Historical Data

A CROSS-SECTIONAL ANALYSIS OF SKILL RELATED PHYSICAL FITNESS COMPONENTS OF KAYAKING AND ROWING PLAYERS

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

WindProspector TM Lockheed Martin Corporation

EN-MOTION TREADMILL. EN-Motion: the ultimate PHYSIO treadmill

Transcription:

Brief Communication Validity of wrist-worn consumer products to measure heart rate and energy expenditure DIGITAL HEALTH Digital Health Volume 4: 1 7! The Author(s) 2018 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/2055207618770322 journals.sagepub.com/home/dhj Robert S. Thiebaud 1, Merrill D. Funk 2, Jacelyn C. Patton 1, Brook L. Massey 1, Terri E. Shay 1, Martin G. Schmidt 1 and Nicolas Giovannitti 1 Abstract Introduction: The ability to monitor physical activity throughout the day and during various activities continues to improve with the development of wrist-worn monitors. However, the accuracy of wrist-worn monitors to measure both heart rate and energy expenditure during physical activity is still unclear. The purpose of this study was to determine the accuracy of several popular wrist-worn monitors at measuring heart rate and energy expenditure. Methods: Participants wore the TomTom Cardio, Microsoft Band and Fitbit Surge on randomly assigned locations on each wrist. The maximum number of monitors per wrist was two. The criteria used for heart rate and energy expenditure were a three-lead electrocardiogram and indirect calorimetry using a metabolic cart. Participants exercised on a treadmill at 3.2, 4.8, 6.4, 8 and 9.7 km/h for 3 minutes at each speed, with no rest between speeds. Heart rate and energy expenditure were manually recorded every minute throughout the protocol. Results: Mean absolute percentage error for heart rate varied from 2.17 to 8.06% for the Fitbit Surge, from 1.01 to 7.49% for the TomTom Cardio and from 1.31 to 7.37% for the Microsoft Band. The mean absolute percentage error for energy expenditure varied from 25.4 to 61.8% for the Fitbit Surge, from 0.4 to 26.6% for the TomTom Cardio and from 1.8 to 9.4% for the Microsoft Band. Conclusion: Data from these devices may be useful in obtaining an estimate of heart rate for everyday activities and general exercise, but energy expenditure from these devices may be significantly over- or underestimated. Keywords Photoplethysmography, physical activity, fitness trackers, activity monitors, Fitbit Submission date: 30 August 2017; Acceptance date: 8 March 2018 Introduction The development of wearable technology to track both heart rate and energy expenditure has improved over the past few years due to the use of photoplethysmography. Data from these devices may facilitate healthy behaviors such as increased physical activity. 1,2 However, if the information collected from wearable technology is not accurate, the usefulness of these devices is limited. Photoplethysmography monitors heart rate by using light emitting diodes and a photo diode. 3 Shorter wavelengths (green light) are emitted into the skin to help minimize motion artifacts, but it does not penetrate skin depth as well as longer wavelengths. 3 Despite using shorter wavelength light, motion artifacts are still present with this technology, so different algorithms are used to further decrease motion artifacts. 3 If heart rates are monitored accurately, they could be used to track exercise intensity and improve the estimation of energy expenditure. For proprietary reasons, many technology companies do not reveal how they validate their technology or which variables they use to estimate energy expenditure, so it is difficult for consumers to 1 Department of Kinesiology, Texas Wesleyan University, USA 2 Department of Health and Human Performance, University of Texas Rio Grande Valley, USA Corresponding author: Robert S. Thiebaud, 1201 Wesleyan Street, Fort Worth, TX 76105, USA. Email: rthiebaud@txwes.edu Creative Commons Non Commercial CC BY-NC: This article is distributed under the terms of the Creative Commons Attribution-Non- Commercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https:// us.sagepub.com/en-us/nam/open-access-at-sage).

2 DIGITAL HEALTH know how valid different devices are and how they compare with other devices. Few studies have investigated the validity of wrist-worn devices that use photoplethysmography to measure both heart rate and energy expenditure. 4 7 Some studies have exclusively examined heart rate and found sufficient accuracy during treadmill walking and running, 8,9 while others found less accurate readings. 10,11 Many different devices and protocols have been used in previous studies, therefore it is important to build sufficient evidence to provide consumers with valuable information on the validity of popular devices. The purpose of this study was to determine the validity of three common wrist-worn consumer monitors at measuring heart rate and energy expenditure during walking, jogging and running. We hypothesized that the devices would be more accurate in estimating heart rate compared to energy expenditure. Methods Participants Twenty recreationally active males and two females participated in this study (mean (SD): age ¼ 22 (3) years, height ¼ 1.73 (0.09) m and weight ¼ 75.9 (10.2) kg). Participants were told about the nature, purpose, details and any risks associated with the experiment, and each participant gave written informed consent. The University s Institutional Review Board approved the protocol of the research study. Exercise protocol Wrist-worn monitors were randomly placed on subjects wrists with a maximum of two monitors on one wrist. Devices were randomly placed on the wrists to avoid any bias that may be produced by placing devices in the same place each time. A possibility of less accurate readings may occur with more than one monitor on a wrist, but monitors were placed on the wrist based on the manufacturer s instructions. Other studies have also used similar procedures to test the validity of these devices. 6 8 Participants exercised on a treadmill at 3.2, 4.8, 6.4, 8 and 9.7 km/h for 3 minutes at each speed with no rest between speeds. These speeds were chosen to reflect various intensities that the general healthy population may experience, and these speeds have been used in other studies. 8,11 The duration of 3 minutes was chosen to allow heart rate to reach steady state at each intensity. Other studies have also used 3 5-minute stages. 7,8,11 Heart rate was measured using electrocardiography (three-lead electrocardiogram (ECG), Quinton Õ Q-Stress, version 4.5, Cardiac Science, Bothell, WA, USA). Energy expenditure was measured using a metabolic cart system (Trueone 2400 Õ metabolic measurement system, Parvomedics, Sandy, UT, USA). Consumer wrist-worn monitors Fitbit Surges (Fitbit Inc., San Francisco, CA, USA), Microsoft Bands (Microsoft Corp., Redmond, WA, USA), and TomTom Cardios (TomTom Inc., Burlington, MA, USA) were placed on the wrist and set to treadmill mode if available and according to the manufacturers recommendations. Statistical analysis The average heart rate and energy expenditure recorded during the 3 minutes at each speed were used for analysis. Pearson correlations measured associations between criterion variables and wrist-worn monitors. Spearman s rank correlation coefficients were used for any variables that were not normally distributed. Statistical significance was set at an level of < 0.01. The criterion measure for heart rate was the ECG and for energy expenditure was the metabolic cart. Mean bias was calculated by subtracting the wrist-worn device from the criterion and 95% limits of agreement were also calculated. For equivalence testing, 95% precision was assumed if the wrist-worn monitors 90% confidence intervals were within an equivalence zone that was between 10% of the criterion mean for energy expenditure and 5% of the criterion mean for heart rate. 12 Mean absolute percentage error (MAPE) ((monitor criterion)/criterion 100%) provided a general measurement error for monitors. Results Heart rate Wrist-worn monitors overestimated heart rates compared to the criterion for all speeds except for the Fitbit Surge, which underestimated heart rate at 8 and 9.7 km/h (Table 1). The MAPE varied from 2.17 to 8.06% for the Fitbit Surge, from 1.01 to 7.49% for the TomTom Cardio and from 1.31 to 7.37% for the Microsoft Band (Table 1). The equivalence zones for heart rate are found in figure 1. Energy expenditure The Fitbit Surge overestimated energy expenditure at each speed, while the TomTom Cardio overestimated energy expenditure at 3.2, 4.8 and 6.4 km/h and underestimated energy expenditure at 8 km/h and 9.7 km/h. The Microsoft Band underestimated energy expenditure

Thiebaud et al. 3 Table 1. Heart rate. 3.2 km/h Heart rate (bpm) Fitbit Surge TomTom Cardio Microsoft Band ECG Mean (SD) 89 (11) 85 (10) 89 (10) 84 (10) Mean bias (SD) 5 (19) 1 (9) 6 (8) LoA 70 108 76 94 73 106 Correlation 0.57 a 0.89 a 0.62 a MAPE (%) 6.5 (13.13) 1.01 (1.38) 7.37 (10.75) Equivalence zones 85 91 81 87 b 85 91 76 92 c,80 88 d 4.8 km/h Mean (SD) 100 (13) 98 (14) 97 (8) 91 (9) Mean bias (SD) 7 (10) 7 (12) 5 (10) LoA 80 120 75 120 77 116 Correlation 0.53 0.55 a 0.35 MAPE (%) 8.06 (12.04) 7.49 (13.41) 6.59 (11.86) Equivalence zones 95 103 92 100 94 99 82 100 c,87 96 d 6.4 km/h Mean (SD) 114 (11) 116 (13) 114 (10) 112 (9) Mean bias (SD) 2 (8) 5 (9) 3 (8) LoA 98 130 100 133 98 130 Correlation 0.69 a 0.76 a 0.63 a MAPE (%) 2.17 (7.68) 4.5 (7.45) 2.46 (7.07) Equivalence zones 110 116 b 111 119 109 115 b 100 122 c, 106 117 d 8 km/h Mean (SD) 132 (13) 141 (12) 141 (11) 135 (10) Mean bias (SD) 4 (8) 6 (11) 5 (12) LoA 117 147 120 162 118 164 Correlation 0.82 a 0.52 0.37 MAPE (%) 2.77 (6.04) 4.38 (8.42) 4.19 (9.15) Equivalence zones 127 135 137 144 137 143 b 122 150 c, 129 143 d 9.7 km/h Mean (SD) 150 (15) 157 (13) 156 (13) 155 (13) (continued)

4 DIGITAL HEALTH Table 1. Continued. Heart rate (bpm) Fitbit Surge TomTom Cardio Microsoft Band ECG Mean bias (SD) 5 (9) 2 (6) 2 (13) LoA 133 167 146 167 132 181 Correlation 0.84 a 0.91 a 0.53 MAPE (%) 3.35 (5.51) 1.11 (5.52) 1.31 (8.05) Equivalence zones 145 153 153 160 b 153 160 b 140 171 c, 147 163 d Values are mean (SD). ECG: electrocardiogram; LoA: 95% limits of agreement; MAPE: mean absolute percentage error. a p-value < 0.01. b Indicates that values are within the 5% equivalence zone of the electrocardiogram. c Indicates 10% equivalence area. d Indicates 5% equivalence area. Table 2. Energy expenditure. Energy expenditure (kcal) Fitbit Surge TomTom Cardio Microsoft Band MetCart 3.2 km/h Mean (SD) 8.7 (2.3) 6.2 (2.2) 5.4 (1.6) 6.1 (1.0) Mean bias (SD) 2.7 (2.0) 0.1 (1.8) 0.7 (1.7) LoA 4.9 12.5 2.7 9.6 2 8.8 Correlation 0.52 0.57 a 0.19 MAPE (%) 44.5 (33.0) 0.4 (31.6) 9.4 (27.9) Equivalence zone 7.9 9.2 5.4 6.7 4.8 5.7 5.5 6.7 4.8 km/h Mean (SD) 27.8 (5.4) 19.3 (5.2) 16.2 (4.0) 17.1 (2.9) Mean bias (SD) 10.3 (4.4) 2.2 (3.7) 0.9 (4.3) LoA 19.2 36.4 11.9 26.6 7.8 24.7 Correlation 0.53 0.70 a 0.24 MAPE (%) 61.8 (27.5) 12.0 (22.2) 3.6 (24.1) Equivalence zones 25.9 29.0 17.5 20.4 14.8 17.1 15.4 18.8 6.4 km/h Mean (SD) 51.0 (8.1) 34.4 (9.1) 32.1 (6.2) 33.4 (5.7) Mean bias (SD) 17 (6.3) 1.0 (6.6) 1.3 (7.7) (continued)

Thiebaud et al. 5 Table 2. Continued. Energy expenditure (kcal) Fitbit Surge TomTom Cardio Microsoft Band MetCart LoA 38.6 63.4 17.8 47.4 47.2 Correlation 0.55 a 0.66 a 0.15 MAPE (%) 52.7 (21.8) 2.5 (20.3) 1.8 (22.3) Equivalence zones 48.2 52.7 31.1 36.4 b 29.9 33.4 30.1 36.7 8 km/h Mean (SD) 80 (11.6) 49.5 (12.7) 59.7 (10.5) 57.9 (10.0) Mean bias (SD) 20.7 (9.8) 9.5 (9.8) 1.1 (12.8) LoA 60.8 99.2 29.7 68.4 34.6 84.7 Correlation 0.53 0.60 a 0.20 MAPE (%) 37.0 (19.1) 16.4 (16.8) 4.0 (21.6) Equivalence zones 75.8 82.5 45.0 52.2 56.3 62.3 b 52.1 63.7 9.7 km/h Mean (SD) 112.7 (16.1) 66.7 (17.0) 96.1 (16.4) 90.7 (15.1) Mean bias (SD) 21.7 (13.3) 24.0 (13.4) 5.4 (17.4) LoA 86.6 138.7 40.5 92.9 61.9 130.3 Correlation 0.60 a 0.56 a 0.37 MAPE (%) 25.4 (15.7) 26.6 (15.0) 7.5 (18.6) Equivalence zones 106.9 116.2 60.7 70.3 90.3 99.6 b 81.6 99.8 Values are mean (SD). MetCart: metabolic cart; LoA: 95% limits of agreement; MAPE: mean absolute percentage error. a p-value < 0.01. b Indicates values are within the equivalence zone of the metabolic cart. at 3.2, 4.8 and 6.4 km/h and overestimated energy expenditure at 8 and 9.7 km/h (Table 2). The MAPE varied from 25.4 to 61.8% for the Fitbit Surge, from 0.4 to 26.6% for the TomTom Cardio and from 1.8 to 9.4% for the Microsoft Band (Table 2). The equivalence zones for energy expenditure are found in figure 2. Discussion The main findings from this study were that wrist-worn monitors produce more accurate readings for heart rates compared to energy expenditure. However, the accuracy of the devices may be influenced by the intensity. When comparing the accuracy of heart rates from the wrist-worn monitors to the ECG readings, the highest correlations were at the fastest speeds for the Fitbit Surge (r ¼ 0.84) and TomTom Cardio (r ¼ 0.91), while the highest correlation for the Microsoft Band was at 6.4 km/h (r ¼ 0.63). Stahl et al. 8 performed a similar study to ours and found higher correlations than we did for the TomTom Cardio (r ¼ 0.959) and Microsoft Band (r ¼ 0.956), although they used the average heart rates throughout the entire exercise protocol to determine their correlations. In another study, Gillinov et al. 10 found concordance correlations of 0.88 between the TomTom Surge and ECG leads. Part of the reason for the lower correlations between heart rates in the current study may be due to a smaller sample size, a different criterion measure used (ECG vs. polar heart

6 DIGITAL HEALTH 3.2 km/h 9.7 km/h Microsoft band Microsoft band TomTom TomTom FitBit surge FitBit surge 60 70 80 90 100 110 120 130 140 150 160 170 Heart rate (bpm) Figure 1. Heart rate equivalence testing to evaluate agreement between devices and electrocardiogram (ECG) at 3.2 and 9.7 km/h. Dashed lines represent the 5% equivalence zones for ECG and solid lines represent 90% confidence intervals for different devices. Microsoft band TomTom FitBit surge 50.0 60.0 70.0 80.0 90.0 100.0 110.0 120.0 Total energy expenditure (Kcal) Figure 2. Equivalence testing for total energy expenditure at 9.7 km/h to evaluate agreement between devices and metabolic cart. Dashed lines represent the 10% equivalence zones for the metabolic cart and solid lines represent 90% confidence intervals for different devices. rate monitor) and that we correlated heart rates at each speed instead of an overall heart rate. When examining the MAPE, other studies have found similar results to the current study. For example, Stahl et al. 8 found that MAPE varied from 0.97 to 5.71% for the TomTom Cardio and from 3.06 to 8.39% for the Microsoft Band, while Gillinov et al. 10 found MAPE of 6.2% for the TomTom Cardio. Similarly, we found that MAPE varied from 1.01 to 7.49% for the TomTom Cardio and from 1.31 to 7.37% for the Microsoft Band, with the lower MAPE found at the faster speeds. Shcherbina et al. 6 also found a larger percent error during walking compared to running for the Fitbit Surge and Microsoft Band. Overall, it appears that the faster the heart rate due to increasing speed, the more accurate the devices become.

Thiebaud et al. 7 Although heart rate was fairly accurate using these monitors, energy expenditure varied much more and did not necessarily correlate with heart rate. For example, when examining the total energy expenditure, MAPE was greater than 20% for the Fitbit Surge and TomTom Cardio, while that of the Microsoft Band was only 7.5%. This confirms other studies that used similar wrist-worn monitors in that they do not accurately measure energy expenditure. 5 7 Though it is unclear how each device calculates the energy expenditure, it does not appear that monitoring heart rate concurrently creates an accurate measure of energy expenditure. One factor that could have impacted the results was the use of a treadmill mode. Two of the devices had a treadmill mode while one device did not. This may have limited some of the ability to accurately measure heart rate and energy expenditure in the device without a treadmill mode. The impact of this on the results is unclear and future studies should determine how using different modes influences the measurement of heart rate and energy expenditure. In addition, because participants only exercised through the multiple speeds once, the reliability of the measurements is unclear. One limitation of the study was the small sample size. A smaller sample size can lead to a lack of uniformity and can decrease statistical power. Despite the small sample size and large variation in the current study, it appears that the results follow a pattern similar to those of other studies investigating these devices. 8,10 Conclusions Wrist-worn monitors report more accurate heart rates than energy expenditure during treadmill exercise. However, the accuracy of the devices for measuring heart rate may not yet be high enough for use in a research setting or for athletes who use heart rate measurement to reach precise heart rates for training purposes. Data from these devices may be useful in obtaining an estimate of heart rate for everyday activities and general exercise, but caution should be taken when using energy expenditure from these devices as the calories may be significantly over- or underestimated. Acknowledgements: We would like to thank the participants who volunteered for this study. Contributorship: RT and MF researched literature and conceived the study. RT, MF and JP were involved in protocol development and gaining ethical approval. RT, JP, BM, MS, TS and NG were involved in patient recruitment and data collection. RT and MF were involved in data analysis. RT wrote the first draft of the manuscript. All authors reviewed and edited the manuscript and approved the final version of the manuscript. Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article. Ethical approval: The ethics committee of Texas Wesleyan University approved this study (Research Ethics Committee number: SP170021). Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work was supported by the Sam Taylor Fellowship Award. Guarantor: RT. Peer review: This manuscript was reviewed by two individuals. The author(s) have elected for these individuals to remain anonymous. References 1. Lewis ZH, Lyons EJ, Jarvis JM, et al. Using an electronic activity monitor system as an intervention modality: A systematic review. BMC Public Health 2015; 15: 585. 2. Patel MS, Asch DA and Volpp KG. Wearable devices as facilitators, not drivers, of health behavior change. JAMA 2015; 313: 459 460. 3. Zhou C, Feng J, Hu J, et al. Study of artifact-resistive technology based on a novel dual photoplethysmography method for wearable pulse rate monitors. J Med Syst 2016; 40: 56. 4. Parak J, Uuskoski M, Machek J, et al. Estimating heart rate, energy expenditure, and physical performance with a wrist photoplethysmographic device during running. JMIR Mhealth Uhealth 2017; 5: e97. 5. Wallen MP, Gomersall SR, Keating SE, et al. Accuracy of heart rate watches: Implications for weight management. PLoS One 2016; 11: e0154420. 6. Shcherbina A, Mattsson CM, Waggott D, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. JPersMed2017; 7: 3. 7. Dooley EE, Golaszewski NM and Bartholomew JB. Estimating accuracy at exercise intensities: A comparative study of self-monitoring heart rate and physical activity wearable devices. JMIR Mhealth Uhealth 2017; 5: e34. 8. Stahl SE, An H, Dinkel DM, et al. How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough? BMJ Open Sport Exerc Med 2016; 2: e000106. 9. Delgado-Gonzalo R, Parak J, Tarniceriu A, et al. Evaluation of accuracy and reliability of PulseOn optical heart rate monitoring device. Conf Proc IEEE Eng Med Biol Soc 2015; 2015: 430 433. 10. Gillinov S, Etiwy M, Wang R, et al. Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise. Med Sci Sports Exerc 2017; 49: 1697 1703. 11. Wang R, Blackburn G, Desai M, et al. Accuracy of wristworn heart rate monitors. JAMA Cardiol 2017; 2: 104 106. 12. Chowdhury EA, Western MJ, Nightingale TE, et al. Assessment of laboratory and daily energy expenditure estimates from consumer multi-sensor physical activity monitors. PloS one 2017; 12: e0171720.