PQAO Training 2017 Karl Tupper, San Luis Obispo APCD

Similar documents
5/30/2014. PQAO Station Operations sub-committee. Phil Wagner, presenter CA Air Resources Board

Thermo Environmental 48C Carbon Monoxide QA Plan Section I Electronic Calibration Branch Responsibilities

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Air Monitoring Directive Chapter 7: Calibration

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

STATE OF CALIFORNIA AIR RESOURCES BOARD AIR MONITORING QUALITY ASSURANCE VOLUME V AUDIT PROCEDURES MANUAL FOR AIR QUALITY MONITORING APPENDIX E

Environmental Protection Division

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Thermo Environmental 48i Trace Level Carbon Monoxide - QA Plan Section I. Electronic Calibration Branch Responsibilities

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

8.1 Calibration of gas sensors

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

Generating Calibration Gas Standards

TELEDYNE MODEL T200UP NITROGEN DIOXIDE (NO-NO2-NOx) MONITORING SYSTEM Section I. Electronic Calibration Branch (ECB) Responsibilities

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Iowa Exceedances of the National Ambient Air Quality Standards, 2013 (Reported Through October 8 th ) Iowa DNR Ambient Air Monitoring Group

RAM 4021 Operation Manual

Standard Operating Procedure

Application Note AN-107

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Sr. No. General points CPCB Comments 1. Print option, acknowledgement of submission for CPCB OCEMS Protocol

Section I: ECB Responsibilities Sulfur Dioxide Standard Operating Procedure Revision 10 November 1, 2016

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

PERFORMANCE SPECIFICATION PROPELLANT, HYDROGEN

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Thermo Environmental Model 146C Calibrator Certification. Section Revision No September 17, 2014 Page 1 of 54

RAM Operation Manual

RAM 4021-DPX Operation Manual

Section : Electronics and Calibration Branch (ECB) Responsibilities

Gas Diluter Portable Gas Dilution System User s Manual

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

Transfeu WP2 Subtask Development of small-scale test method for fire effluents

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Section II: Operator Responsibilities Sulfur Dioxide Standard Operating Procedure. Revision 12. Effective Date: November 1, 2016

OPERATING PROCEDURE. Number: V2. Authors. Approvals

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Monterey Bay Unified Air Pollution Control District Silver Cloud Court, Monterey, CA 93940

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

SOUTH COAST AIR QUALITY MANAGEMENT DISTRICT METHOD INSTRUMENTAL ANALYZER PROCEDURES FOR CONTINUOUS GASEOUS EMISSION SAMPLING

Pressure Regulators. Operating Instructions. Instrumentation

A. Evaluation of the written program

NOTICE OF AVAILABILITY FOR PUBLIC REVIEW AND COMMENT 2011 AIR MONITORING NETWORK PLAN

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

Calibration Requirements for Direct Reading Confined Space Gas Detectors

INDUSTRIAL HYGIENE GROUP Standard Operating Procedure. TVA-1000 PID/FID Portable Toxic Vapor analyzer. 1 of 9. Contents

911A Summa Canister Sampler

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Continuous PM Monitoring Update MARAMA Monitoring Committee Workshop December 2016

6200E Installation Requirements

APTI 464. Analytical Methods for Air Quality Standards. Laboratory Exercise Manual. March 2010

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Safety Data Sheet. Zero Air 1 PRODUCT AND COMPANY IDENTIFICATION. Manufacturer NorLab a division of Norco 898 W. Gowen Rd.

TECHNICAL INFORMATION

Air Quality Program. July G (rev. 7/04)

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

December 16, I. Introduction

SAFETY AND HEALTH STANDARD CONFINED SPACE ENTRY REQUIREMENTS Effective Date: 06/10/10 Standard: 16.2 Document Number: KUCSH0038 Rev: 07

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

EN ISO/IEC Technical Requirements

Verification Of Calibration for Direct-Reading Portable Gas Monitors

Aeroqual AirCal 1000 User Guide

4.0 Audit Procedure for the MetOne SASS Speciation Sampler

HiQ laboratory gas generators.

Using the UltraRAE. Firmware 2.35

Thermo Scientific Model 146i Principle of Operation

21700 Nordhoff St. Chatsworth, CA Phone

Working in Gas Hazard Areas

Niltox Breathing Air Purifier Range

RGC-IR Remote Gas Calibrator for IR400

Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012

Met One E-BAM Particulate Monitor

Chapter 5. Sulfur Dioxide Measurement Principles and Calibration Procedures. Introduction

Protocol Gas Verification Program (PGVP)

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Respiratory Protection

STANDARD OPERATING PROCEDURES

CALIBRATION. empanelled laboratories following standard procedures and using certified reference standards.

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, ARGON

Guidelines for Real Time Sampling & Analyses

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

GAS DETECTION: CALIBRATION GAS

SUBSTATION MAINTENANCE ELECTRICAL OPERATING PROCEDURE

Safety Regulations and Procedures GENERAL SAFETY Confined Spaces - S New 11/96. To provide guidance for the safe entry of confined spaces.

FV Flash Vessel Installation and Maintenance Instructions

HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

SL G3. Warning

Truck-mounted Mass Flow Metering for LPG Delivery

IDEM Indiana Department of Environmental Management Office of Land Quality P.O. Box 6015 Indianapolis, IN OLQ PH: (317)

PERFORMANCE SPECIFICATION PROPELLANT PRESSURIZING AGENT, NITROGEN

Transcription:

PQAO Training 2017 Karl Tupper, San Luis Obispo APCD 1

Right now: Introduction, Big Picture This afternoon: Demo: Ozone Analyzer Calibration Some additional slides (nitty gritty) More time for Q&A Session: Station Operations & Documentation B Room: Mountain Vista 2 Times: 2:35 to 3:15 and again 4:20 to 5:00 2

Calibration is defined as: The comparison of a measurement standard, instrument, or item with a standard or instrument of higher accuracy FINAL to detect and quantify inaccuracies and to report or eliminate those inaccuracies by adjustment. (EPA QA Handbook Vol II) Verification : comparison done, but no AS IS adjustment made We ll focus on gas instruments, but concepts apply to PM and meteorology, too. 3

Calibrations should be carried out at the field monitoring site by allowing the analyzer to sample test atmospheres containing known pollutant concentrations The analyzer should be warmed up: at least several hours but preferable overnight Warm-up periods (per API) : 2 hours (O3) 24 48 hours (NOx) 4-7 days (TCO) 4

During the calibration, the analyzer should be operating in its normal sampling mode, including: Filters / scrubbers /conditioners Through as much of the ambient air inlet system as is practicable. Ideally, through the probe. Keep this in mind when: Designing new station Purchasing calibration equipment 5

EPA QA Vol II, Section 12.3: Installation Relocation Repairs or Service Interruption Upon any indication of potential malfunction At least every 6 or 12 months (see below) Station shutdown Minimum frequency of calibration (EPA QA Vol II, App D) Every 6 months Or annually, IF zero/span check is done daily Keep track of when calibrations are happening! 6

Audits: No calibrations within 30 days before an audit (CARB QA) Ambient Conditions: Pollution: not during pollutant episodes, (i.e.,o 3 not in afternoon) O 3 : Not during O 3 season Met / Particulates: not during high wind conditions or adverse weather conditions due to safety 7

EPA requirements (EPA QA Vol II. App D): Ozone, SO 2, Carbon Monoxide: Zero 4 Upscale Points NO, NO 2, NO x : Zero 4 Upscale Points for NO & NO x 3 NO 2 Titration Points 8

No EPA/ARB requirements, but: Highest point should be 80% of full scale and >NAAQS (EPA QA Vol II) Keep limits of calibrator (i.e. MFC ranges) in mind Good idea to bracket audit levels ARB proposed 2017 audit levels (ppm): 9

EPA audit levels: Regular analyzers: 3 yellow levels Trace analyzers: 2 pink & 2 yellow 10

Must be certified and traceable EPA QA Vol II, Section 12.1.2 CARB Standards Lab: http://www.arb.ca.gov/aaqm/qa/stdslab/stdslab.htm Check expiration dates: Cal Gas Tanks Mass Flow Controllers Ozone transfer standard (6 months) Flow Standards (1 year) etc Properly warm-up/equilibrate: Flow Standards Ozone transfer standard etc 11

Zero Air Generator is part of calibration system Per EPA s TSA of ARB, we should: Verify that our zero is clean Document ZAG maintenance In depth presentations on this topic: Wednesday, 10:35 11:00 Yousef Hameed Zero Air Guidance: Introduction. Wednesday, 1:45 & 3:30, Mountain Vista 2 David Roque Zero Air Generator Certification. 12

General practice is to allow concentrations to stabilize (plateau) for at least 10-15 minutes Teledyne API Analyzers: Stability should be <0.5 ppb (non-trace level) 15 min NO NO 2 NO x Time 13

Usual stuff: name, data, initials/signature Information for standard(s): Serial number(s) Calibration/certification date(s) Concentrations, correction factors, etc. Information for instrument: Serial number As-Is and Final slope/intercept/correction factor Calibration info: Calibration points Instrument response 14

EPA says (QA Vol II sec 12.2 & App D; 40 CFR 50 Apps A, C, D or F): All points within 2% of calibration scale For NO 2 : NO x convertor efficiency > 96% ARB says (SOPs): Overall % Accuracy : 100% * (Σtrue i Σresp i ) / Σtrue i Ozone: <2% NO/NO 2 /No x : <5%, convertor efficiency > 96% Check your SOPs and QAPPs!! 15

Whatever happens: Document Document Document!! Double check calibration equipment, data logger, and calculations Corrective Action: Rezero and/or Respan Fix or replace instrument Perform calibration/verification prior to placing monitor back online. Inform upstream data users Effect on Data: Depends Note: post-processing of data no longer accepted 16

17

18

PQAO Training 2017 Afternoon Session Karl Tupper, San Luis Obispo APCD 19

Response plateauing Purging regulators Warm-up period 20

4 NO/NO x points 3 NO 2 points 1 zero 15 min NO NO 2 NO x Time 21

NO NO 2 NO x 22

NO NO 2 NO x 23

24

1. Attach the regulator to the cylinder valve. 2. Open the cylinder valve to flow gas into the regulator, then immediately close the cylinder valve. 3. Increase the regulator outlet pressure several PSI. 4. Then open the regulator outlet valve to vent the gas trapped in the regulator. 5. Close the regulator outlet valve. 6. Repeat steps 2 through 5 (4-5 times). CAUTION: Many gas mixtures contain harmful or toxic substances. Suitable vents must be used to avoid breathing these substances. 25