Standards of Practice Emergency Shower/Eyewash Installation, Use, Testing & Maintenance

Similar documents
1.0 PURPOSE 2.0 SCOPE 3.0 DEFINITIONS. ANSI: American National Standards Institute. CCC: Chemical Control Centre

Standards of Practice Personal Protective Equipment

EMERGENCY WALL MOUNTED HAND OPERATED DELUGE SHOWER

Design Requirements for Eyewash and Safety Shower Units

ENVIRONMENTAL HEALTH AND SAFETY STANDARD OPERATING PROCEDURES

Environmental Health & Safety Policy & Procedure #35

GUARDIAN EQUIPMENT Emergency Eye Wash and Shower Equipment. ANSI Z358.1 Compliance Checklist

ANSI Z Compliance Checklist

Emergency Eyewash & Shower Testing Program

Emergency Shower & Eyewash. Training Guide

EMERGENCY EYEWASH & SHOWERS

EMERGENCY EYEWASH STATIONS & SAFETY SHOWERS GUIDELINES

Cal Poly Eye Wash and Safety Shower Inspection Program June 20, 2018

EMERGENCY EYEWASH & SHOWER STATION TRAINING AND MAINTENANCE REQUIREMENTS

PEDESTAL MOUNTED EMERGENCY EYE & EYE/FACE WASH

ANSI / ISEA Z Compliance Checklist

Laboratory Safety Guideline Emergency Eyewash and Safety Showers

ANSI / ISEA Z Compliance Checklist

WALL MOUNTED EMERGENCY EYE & EYE/FACE WASH HAND OPERATED

EYE WASH STATIONS & SAFETY SHOWERS GUIDELINES

Product Manual B-Safety ClassicLine & PremiumLine Emergency Eyewash and Eye/Face Wash Equipment

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

Installation. Heat Trace Combination Drench Shower and Eyewash Unit (includes 120 V and 208/240 V versions) S19-304GA

Laboratory Chemical Hygiene Plan

Model Portable Eyewash

Compressed Gas Cylinders - Guideline

SAFETY EMERGENCY SAFETY SHOWERS AND EYEWASH STATIONS TABLE OF CONTENTS 1. INTRODUCTION SCOPE & PURPOSE DEFINITIONS...

Hazard Communication Program

Laboratory Safety Plan

LABORATORY FUME HOOD SPECIFICATION AND MAINTENANCE PROGRAM

ALBUQUERQUE PUBLIC SCHOOLS RESPIRATORY PROTECTION PROGRAM

SAFETY TRAINING LEAFLET 06 CARBON DIOXIDE

COMPLETION OF PROCEDURE ASSESSMENT FORM (COSHH RELATED) GUIDANCE NOTES (Version 3)

Model TWBS.SH Tempered Water Blending System

AED Program 2017 January 2017

CONTROL OF HAZARDOUS ENERGY

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

Standards of Practice Respiratory Protection: Appendix C

SOP #: Pressure Vessels

Operations: HSE. Health and Industrial Hygiene. Laboratory Safe Work Practice

BY ORDER OF THE SECRETARY OF THE AIR FORCE AIR FORCE OCCUPATIONAL SAFETY AND HEALTH STANDARD SEPTEMBER Safety

Student Shop Safety. University of Tennessee Safety Guide GS-005. Purpose. Scope and Applicability. Definitions and Abbreviations.

Installation. S19-300T (top supply) S19-300B (bottom supply)

All Multi-Tech Non-Phthalate Color Inks MATERIAL SAFETY DATA SHEET

Risk Management Resources

EM385 Construction Safety and Health Certification Course Section 12 Control of Hazardous Energy

Hydrofluoric Acid. Syracuse University Laboratory Standard Operating Procedure. I. Purpose: Physical and Chemical Properties: Potential Hazards:

MANUFACTURING TECHNICAL INSTRUCTIONS - SAFETY FALL HAZARD CONTROL REQUIREMENTS SERIES & NO. SMI - 157

Hazardous Energy Control

Standard Operating Procedure

Title: Benzene Awareness Effective Date: 11/14/2014 Control Number: THG_0037 Revision Number: 1 Date: 10/22/2015 Annual Review Completed: 5/13/2015

SUNY OSWEGO ENVIRONMENTAL HEALTH AND SAFETY

Product Information Report Lockout/Tagout Hazardous Energy Control

Instructions for Installation, Use and Maintenance FSI Safety Tank Shower Systems

PROCEDURES FOR REPAIRS TO ASME NV STAMPED PRESSURE RELIEF DEVICES OF NUCLEAR SAFETY RELATED PRESSURE RELIEF VALVES

Title: Abrasive Blasting Effective Date: 11/17/2014 Control Number: THG_0034 Revision Number: 1 Date: 10/22/2015 Annual Review Completed: 5/13/2015

PURPOSE OF THE POLICY

COLLEGE OF ENGINEERING LABORATORY SAFETY MANAGEMENT SYSTEM (SMS) & HAZOP PROTOCOL January 2018

SOUTH LYON FIRE DEPARTMENT Manual of Procedures 403

LOCKOUT/TAGOUT PROGRAM

Work Health and Safety Risk Management Procedures

Lockout - Tagout. Control of Hazardous Energy OSHA Standard

Laboratory Safety Training. Environmental Health and Safety

ASSOCIATED DOCUMENTS:

Installation, Operation and Maintenance Manual for Back Pressure Regulator

CONFINED SPACE WRITTEN PROGRAM

Control of Hazardous Energy. Environmental Health and Safety

LOCK-OUT/TAG-OUT (LO/TO) SAFETY PROGRAM

HIGH-IMPACT. Leader s Guide

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

1. The Middletown Public Schools will have defibrillators in each school building.

THERM-O-MIX STATION/WWM TEMPERED WATER SUPPLY

UTAS EH&S INSTRUCTION No. 036 June 2012 Revision No. 3. Purging of Flammable Gas Equipment and/or Piping (Explosion Hazard)

SUNY OSWEGO ENVIRONMENTAL HEALTH AND SAFETY

BCF WORK INSTRUCTION Respirator Inspection/Cleaning/Maintenance/Repair and Storage

Lab-Specific Standard Operating Procedure (LSOP)- Pyrophoric Chemicals

Pressure Vessel Program

ASSE International Product (Seal) Listing Program

PRESSURE SYSTEMS POLICY

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

Cold Weather Safety Program

Laboratory Safety Review Checklist One Shields Ave Davis, CA Phone: (530) Fax: (530)

Lock Out/Tag Out Control of Hazardous Energy

PART Q CONTROL OF HAZARDOUS ENERGY (LOCKOUT-TAGOUT)

A Guide to Eyewash and Safety Shower Facilities

Cabinet. 1.0 POLICY It is the policy of Calvin College to comply with the MIOSHA Respiratory Protection Standard Part 451.

Laboratory Standard Operating Procedure: Sensitizers

Hazardous Waste Training Plan. Supersedes: 02/15/16 (Rev.02) Preparer: Owner: Approver: EHS Team Member EHS Team Member EHS Manager

Vehicle-mounted meters, pump supplied

Compressed Gases and Cryogens

LABORATORY SAFETY INSPECTION

ASME and PED Requirements for Drum Level Instrumentation. Presented by: Amir Hedayatnia Clark-Reliance Corp.

-JHA- Job. For Science and Engineering. Hazard Assessment


T130/T131/T147. Economy Basket Stretcher with Lifting Kit

Laboratory Standard Operating Procedure: Specimen Preservatives

MEDICAL AIR Effective date: 02/80 MAINTENANCE Revision date : 08/12

CLUB RISK ASSESSMENT TEMPLATE

Equipment Operation Procedures

PRESCOTT SCHOOL DISTRICT NO South A Street Prescott, WA The Control of Hazardous Energy (Lockout/Tagout)

Transcription:

Standards of Practice Emergency Shower/Eyewash Installation, Use, Testing & Maintenance Document Authority: EHS Coordinator Document Custodian: EH&S Coordinator Effective Date: 09.01.15 Issuing Dept: College of Agriculture EH&S Next Review Date: Document Number: 09.01.21 Control Tier: II ESEW001 I. PURPOSE To provide uniform requirement for the emergency shower and eyewashes performance, use, testing and maintenance procedures. II. SCOPE K-state university laboratories, non-laboratory areas, such as art and science studios, maintenance facility and other shops. III. CONSEQUENCES OF DEVIATION This procedure serves as an essential aspect of having an operational ready emergency shower or eyewash station for faculty, staff and students. Ignoring this procedure could result in serious injuries or fatalities. IV. RESPONSIBILITY 1. Environmental Health & Safety Work with departments to identify the area(s) requiring emergency showers and eyewash stations Provide standards for installing the type of emergency showers and eyewash stations, and testing and maintenance instructions 2. Facility maintenance Make any necessary repairs for the emergency showers and eyewash stations regarding the inspection results. 3. Departments Departments should verify the staff instructed the location, use the emergency showers and eyewash stations when need Departments or other designated parties should inspect their emergency showers and eyewash units once a week. Departments should contact facility maintenance for repair service when the emergency shower and eyewash station are not working properly as expected. Department laboratories are responsible for purchase and get appropriate installation for the emergency showers and eyewash station based on the work requirements. Page 1 of 5

V. REQUIRED LOCATION(S) Any area in laboratories involving corrosive materials, including but not limited to battery charging areas, and any laboratory using hazardous chemicals must have an emergency shower and eyewash plumbed to the building water supply line. VI. EQUIPMENT INSTALLATION REQUIREMENTS 1. All emergency showers and eyewash station must be purchased and installed to comply with the latest ANSI Z358.1 American National Standard for Emergency Eyewash and Shower Equipment. 2. All new installations must deliver tepid flushing fluid. Personal wash units shall be protected from freezing and shall not be exposed to ambient temperatures exceeding 38 (100 ). 3. All emergency showers should be assembled and installed by the manufacturer s instructions and the flushing fluid delivery requirements, the location should be no more than 10 seconds for the users to reach. The pattern of the shower should be at least 208.3 cm (82 in.) for the top of the flushing fluid column, and not more than 243.8cm (96 in.) from the surface on which the user stands. The center of the spray shall be at least 40.6 cm (16 in.) from any obstruction. Be connected to a supply of flushing fluid per the manufacturer s installation instructions to produce the required spray pattern for a minimum period of 15 minutes. Deliver tepid flushing fluid When the plumbed eye/face wash is installed, its performance shall be verified in accordance with the following procedures When the self-contained eye/face wash is installed, its installation shall be verified in accordance with manufacturer s instructions. VII. TESTING Each part of the combination unit shall be certified individually and when activated procedures. Plumbed eye/face washes and self-contained Eye/Face Washes should be certified by rules from the Testing Procedures for Certification. Determine that flushing fluid is substantially dispersed throughout the pattern. The flushing fluid column pattern shall be at least 208 cm (82 in.) and not more than 243.8cm (96 in.) from the surface on which the user stands. Measure the diameter of the flushing fluid pattern 152.4 cm (60 in.) above the surface on which the user stands. The diameter shall be a minimum of 50.8 cm (20 in.). Throughout the 15-minute test, verify that the flow rate is a minimum of 75.7 liters per minute (20 gpm). VIII. MAINTENANCE AND TESTING 1. Manufacturers shall provide operation, inspection and maintenance instructions with emergency shower equipment. Instructions shall be readily accessible to maintenance and training personnel. Page 2 of 5

2. Plumbed emergency showers shall be activated weekly for a period long enough to verify operation and ensure that flushing fluid is available. 3. Self-contained emergency showers shall be visually checked weekly to determine if flushing fluid needs to be changed or supplemented. Such inspection shall be conducted in accordance with manufacturer s instructions 4. Employees who may be exposed to hazardous materials shall be instructed in the location and proper use of emergency showers. 5. All emergency showers shall be inspected weekly to assure conformance with installation requirements of this standard. IX. RELATED FORMS: Inspection form (Appendix 1) RELATED DOCUMENTS: ANSI Z358.1-2014 American National Standard for Emergency Eyewash and Shower Equipment Appendix 1 Weekly/Monthly Inspection Log Date Function properly Y/N Inspector Initial Additional Comments Page 3 of 5

X. SAFETY CONSIDERATIONS A1. Personal Wash Unit In order to keep the eye from serious injury is to follow the instruction for workers to get the minimum injury the possible. A2. Signage Visible signage should be placed on or immediately adjacent to the shower/eyewash station. A3. First Aid Practices The work place should provide professional training and develop the common sense in employees to get safety equipment while working with hazardous, and the usage of emergency showers and eyewash stations. A4. Waste Disposal The department should give appropriate disposal of waste from operating the emergency eyewash and shower equipment. A5. Personal Protective Equipment Emergency eyewash and shower equipment is not a replacement for primary protection devices. The personal protective equipment should be provided in the working area. Workers should be able to use when they need. XI. INSTALLATION CONSIDERATION B1. Supply Lines Installation should be in accordance with proper practices and procedures. B2.Water Capacity ANSI has reference standard to ensure that the testing for certification purposes is consistent and reproducible. It is the responsibility of the designer and owner to ensure proper flushing fluid delivery. The testing of emergency eyewash and shower equipment is to be conducted at normal facility operating pressures. Excess flow pressure can injure the user or damage the equipment. B3. Valve Operation To ensure safety, the control valve should remain open to allow the user to use both hands for keeping the eyes open. B4. Alarm Devices In order for users to understand the terms of applying the safety equipment, the alarm system with warning lights would be able to indicate whether the device is in operation. Page 4 of 5

B5. Placement of Emergency Eyewash and Shower Equipment The emergency shower and eyewash equipment should be immediately functionally working after installation. B6. Delivered Flushing Fluid Temperature Temperatures in excess of 38 (100 ) have proven to be harmful to the eyes and can enhance chemical interaction with the skin and eye tissue. For cold flushing fluid temperatures, a temperature of 16 (60 ) is suitable for the lower parameter for tepid flushing fluid without causing hypothermia to the equipment user. B7. Weekly or Monthly Activation for Plumbed Emergency Eyewash and Shower Equipment Departments are responsible for performing inspection for the emergency showers and eyewash station for each lab. Also, each lab should report any repair issue for emergency eyewash and shower to ensure work place safety. Revision Log REVISION DATE REVISION NO. REVISION AUTHORITY NATURE OF REVISION 08.14.15 0 EH&S Coordinator Date of Original Document Issuance Controlled documents are maintained electronically. Printed documents are UNCONTROLLED. Prior to relying on a printed document, verify that it is current. Page 5 of 5